QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

Similar documents
DRI METHAQUALONE APPLICATION BECKMAN COULTER AU5800

CEDIA OPIATE APPLICATION BECKMAN COULTER AU5800

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator

ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System

The Adiponectin Turbidimetric Immunoassay Reagent Kit

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

Morinaga Mouse C-peptide ELISA Kit

Insulin (Porcine/Canine) ELISA

Bovine Insulin ELISA

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Porcine/Canine Insulin ELISA

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

EliKine Free Thyroxine (ft4) ELISA Kit

The NGAL Turbidimetric Immunoassay Reagent Kit

Mouse C-Peptide ELISA Kit

Toxoplasma gondii IgM ELISA Kit

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum

HbA1c (Human) ELISA Kit

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

IgG Antibodies To Toxoplasma Gondii ELISA Kit Protocol

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

CYTOMEGALOVIRUS (CMV) IgM ELISA Kit Protocol

IV2-113E Use by. Invitron Glargine ELISA Kit REF LOT IVD. Definitions. English. For in-vitro diagnostic use. Instructions for use.

Mouse Ultrasensitive Insulin ELISA

Thyroid Stimulating Hormone (S-TSH) Thyroid Stimulating

human Total Cathepsin B Catalog Number: DY2176

Mouse GLP-2 EIA. Cat. No. KT-374. For the quantitative determination of GLP-2 in mouse serum or plasma. For Research Use Only. 1 Rev.

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Human HIV (1+2) antigen&antibody ELISA Kit

Rubella virus IgG ELISA Kit

Mouse C-peptide ELISA

Mouse C-peptide ELISA

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

Serum Amyloid A ELISA

Mouse serum Insulin 200 tests

Insulin ELISA. For the quantitative determination of insulin in serum and plasma.

Toxoplasma gondii IgM ELISA Kit

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures.

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Human Cytomegalovirus IgM ELISA Kit

Human Cytomegalovirus Virus (CMV) IgG ELISA Kit

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

Nori Rabbit IL-2 ELISA Kit DataSheet

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

H.Pylori IgG

Procine sphingomyelin ELISA Kit

H.Pylori IgG Cat # 1503Z

Instructions for Use. IA2 Antibody ELISA. Enzyme Immuno Assay for the Quantitative Determination of Antibodies against IA-2 in Serum.

Canine Thyroid Stimulating Hormone, TSH ELISA Kit

Apolipoprotein A-1 ELISA

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System.

Thyroxine (T4) Human ELISA Kit

TSH Receptor Autoantibody ELISA

Clinical Chemistry Specific Proteins Presentation

PERIOSTIN ELISA CONTENTS

Cotinine (Mouse/Rat) ELISA Kit

Rat Leptin-HS ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN

Insulin Rodent (Mouse/Rat) Chemiluminescence ELISA

Triiodothyronine (T3) ELISA

This package insert contains information to run the Prealbumin assay on the ARCHITECT c Systems and the AEROSET System.

Mercodia Porcine C-peptide ELISA

RealLine HIV quantitative Str-Format

Technical Bulletin No. 161

Rat Hemoglobin A1c (HbA1c) Kit Instructions

25(OH) Vitamin D ELISA (BD-220BA), 192 Tests

High-density Lipoprotein Cholesterol (HDL-C) Assay Kit

For In Vitro Diagnostic Use

Rat Leptin ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN

LDL (Human) ELISA Kit

STELLUX C-peptide Chemiluminescence ELISA

Instructions for Use. Tg Antibody ELISA

Toxoplasma gondii IgM (Toxo IgM)

YK052 Mouse Leptin ELISA

Total Acid Number in petroleum products by automatic titration

Vitamin A-E Serum HPLC Analysis Kit

RayBio Human Granzyme B ELISA Kit

Total Thyroxine ELISA (T4)

Mercodia Porcine C-peptide ELISA

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Mouse C3 (Complement Factor 3) ELISA Kit

Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit

DIAGNOSTIC AUTOMATION, INC.

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

ab MitoTox Complex I OXPHOS Activity Microplate Assay

See external label 2 C-8 C Σ=96 tests Cat # 3171Z. Free Estriol. Cat # 3171Z. Enzyme Linked Immunosorbent Assay

25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL

ACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man Coulter HIV-1 p24 ELISA May 21, 2004

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems.

STANDARD OPERATING PROCEDURE TITLE: TOTAL IGE EIA (HYTEC 288)

Mouse Leptin ELISA Kit Instructions

Technical Bulletin No. 162

Cortisol (Sheep) ELISA Kit

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

LC-MS/MS Method for the Determination of Tenofovir from Plasma

INSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY

CoQ10(Coenzyme Q10) ELISA Kit

Transcription:

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole blood on automated clinical chemistry analyzers. The results obtained are used as an aid in the management of kidney, liver, and heart transplant patients receiving tacrolimus therapy. This in vitro diagnostic device is intended for clinical laboratory use only. For In Vitro Diagnostic Use Only Purpose The information provided in this application sheet is intended as a supplement to the package insert. Refer to the package insert for information on intended use, reagent storage, reagent preparation, specimen collection, specimen preparation, specimen storage, quality control, and additional performance data. Ordering Information Item QMS Tacrolimus Assay QMS Tacrolimus Calibrator Set More Diagnostics Rap/Tac/CsA Control Level Q More Diagnostics Rap/Tac/CsA Control Level 1 More Diagnostics Rap/Tac/CsA Control Level 2 More Diagnostics Rap/Tac/CsA Control Level 3 Size R1 18 ml, R2 12 ml, Extraction Reagent 50 ml 6 levels, 4 ml Calibrator A 2 ml Calibrators B-F Beckman Coulter Reorder Number A53727 A53728 4 x 4 ml B48482 4 x 4 ml B51007 4 x 4 ml A53712 4 x 4 ml A53713 AU Bottle 15 ml 63165 AU Bottle 30 ml 63094 Technical Support For Technical Support, please contact your local Beckman Coulter Representative. Reagent Storage Refer to the package insert for information on reagent storage. P/N A53727 Page 1 of 13

Instructions For Use Procedure for Analyzer Refer to the operator s manuals for information on analyzer operation. Before use, invert several times, avoiding the formation of bubbles. Dispense R1 reagent and R2 reagent into appropriate AU bottles as shown in the table below: AU Reagent Bottle QMS Tacrolimus Assay Kit R1 Compartment R2 Compartment Anti-Tacrolimus Monoclonal Antibody R1 One Bottle (30 ml) Tacrolimus-Coated Microparticles R2 One Bottle (15 ml) Warning: These reagents have to be programmed to fixed positions. Do not use the Thermo reagent bottles directly on the AU analyzer. Significant interference from QMS Tacrolimus into the Microalbumin (OSR6167) and Urine/CSF Albumin (B38858/B46435) assays has been observed due to reagent carryover in random access analyzers. Setup the recommended contamination parameters below: No. PRECEDING TEST NAME Type FOLLOWING TEST NAME AU480 Contamination Parameters Type REAGENT PROBE CLEANER KIND WASH COUNT EFFECTIVE OF WATER CLEANING SAME USE MIXER 1 TAC R1 MALB/UALB R1 CLN 100%* 1 No Yes No 2 TAC R1 MALB/UALB R2 CLN 100%* 1 No Yes No CUVETTE AU680 Contamination Parameters EFFECTIVE PRECEDING FOLLOWING REAGENT PROBE WASH SAME USE No. OF WATER TEST NAME TEST NAME CLEANER KIND COUNT CLEANING MIXER CUVETTE 1 TAC MALB/UALB CLN 100%* 1 No Yes No No. PRECEDING TEST NAME FOLLOWING TEST NAME AU5800 Contamination Parameters REAGENT PROBE CLEANER KIND WASH COUNT EFFECTIVE OF WATER CLEANING MIXER SAME USE CUVETTE 1 TAC MALB/UALB CLN 100%* 1 No Yes No * CLN OSR0001/ODR2000, replenish cleaner on a daily basis or as required. Note: For the AU5800 it is recommended to separate MALB and UALB from TAC by using designated rings, if possible. Page 2 of 13

Results and Data Interpretation Specimen Preparation Results for samples will be printed in ng/ml. Refer to the package insert for the complete specimen preparation. Due to sample stability, it is recommended not to exceed a maximum of 24 extracted samples per run. The product insert can be found at the Thermo Fisher website: www.thermoscientific.com/diagnostics Calibration Use the QMS Tacrolimus Calibrator Set. The calibrators are prepared like patient samples. The value on the bottle is the value to use in the parameters below. These are lot number specific and should be updated when calibrator lot numbers change. Page 3 of 13

Application Parameters Parameters The following tables outline the QMS Tacrolimus Assay chemistry parameters on the Beckman AU480, AU680, and AU5800 analyzers. QMS TACROLIMUS, AU480 Specific Test Parameters General LIH ISE Range Test Name: TAC Type: Serum Operation: Yes Sample Volume 10.0 L Dilution 0 L OD Limit Pre-Dilution Rate 1 Min. OD -2.00 Max. OD 3.00 Reagents Volume: R1(R1-1) 125 L Dilution 0 L Reagent OD limit: R2 (R2-1) 75 L Dilution 0 L First Low -2.00 High 3.00 Last Low -2.00 High 3.00 Dynamic Range Low 1.0 High 30.0 Correlation Factor A 1 B 0 Wavelength: Pri. 700 nm Sec. None nm Factor for Maker A 1 B 0 Method: FIXED1 Reaction slope: + Onboard Stability # Days # Hour Measuring Point 1: First 20 Last 27 LIH Influence Check # Measuring Point 2: First Last Lipemia Linearity: % Icterus No Lag Time: No Hemolysis Specific Test Parameters General ISE Range Test Name: TAC Type: Serum Value/Flag: # Level L: # Level H: # Specific Ranges: From To Panic Value Sex Year Month Year Month Low High Low High 1. # # # # # # # # # 2. # # # # # # # 3. # # # # # # # 4. # # # # # # # 5. # # # # # # # 6. # # # # # # # 7. No demographics # # 8. Not within expected values # # Unit ng/ml Decimal Places # Page 4 of 13

QMS TACROLIMUS, AU480, continued Calibration Specific General ISE Test Name: TAC Type Serum Use Serum Cal. Calibration Type: 6AB Formula: POLYGONAL Counts: 2 <Calibrator Parameters> Slope Check - Calibrator OD Conc Factor Range Low High Allowable Range Check Point 1: # * -2.0 3.0 Reagent Blank Point 2: # * -2.0 3.0 Calibration Point 3: # * -2.0 3.0 Advanced Calibration Point 4: # * -2.0 3.0 Operation No Point 5: # * -2.0 3.0 Interval (RB/ACAL) Point 6: # * -2.0 3.0 Point 7: Point 8: Point 9: Point10: <Point Cal. For Master Curve> No. of Correction Points Use Master Curve Lot Calibration Calibrator OD Conc OD Range Stability Low High Reagent Blanks # Day # Hour Point 1: Calibration # Day # Hour Point 2: MB Type Factor: 1-Point Calibration Point None With CONC-0 # User defined * Lot specific calibrator values Page 5 of 13

QMS TACROLIMUS, AU680 Specific Test Parameters General LIH ISE Range Test Name: TAC Type: Serum Operation: Yes Sample Volume 10.0 L Dilution 0 L OD Limit Pre-Dilution Rate 1 Min. OD -2.00 Max. OD 3.00 Reagents Volume: R1(R1-1) 125 L Dilution 0 L Reagent OD limit: R2 Volume 75 L Dilution 0 L First Low -2.00 High 3.00 Last Low -2.00 High 3.00 Dynamic Range Low 1.0 High 30.0 Common Reagent Type None Name Correlation Factor A 1 B 0 Wavelength: Pri. 700 nm Sec. None nm Factor for Maker A 1 B 0 Method: FIXED1 Reaction slope: + Onboard Stability # Days # Hour Measuring Point 1: First 20 Last 27 LIH Influence Check # Measuring Point 2: First Last Lipemia Linearity: % Icterus No Lag Time: No Hemolysis Specific Test Parameters General ISE Range Test Name: # Type: Serum Value/Flag: TAC Level L: # Level H: # Specific Ranges: From To Panic Value Sex Year Month Year Month Low High Low High 1. # # # # # # # # # 2. # # # # # # # 3. # # # # # # # 4. # # # # # # # 5. # # # # # # # 6. # # # # # # # 7. No demographics # # 8. Not within expected values # # Unit ng/ml Decimal Places # Page 6 of 13

QMS TACROLIMUS, AU680, continued Calibration Specific General ISE Test Name: TAC Type Serum Use Serum Cal. Calibration Type: 6AB Formula: POLYGONAL Counts: 2 <Calibrator Parameters> Slope Check - Calibrator OD Conc Factor Range Low High Allowable Range Check Point 1: # * -2.0 3.0 Reagent Blank Point 2: # * -2.0 3.0 Calibration Point 3: # * -2.0 3.0 Advanced Calibration Point 4: # * -2.0 3.0 Operation No Point 5: # * -2.0 3.0 Interval (RB/ACAL) Point 6: # * -2.0 3.0 Point 7: Point 8: Point 9: Point10: <Point Cal. For Master Curve> No. of Correction Points Use Master Curve Lot Calibration Calibrator OD Conc OD Range Stability Low High Reagent Blanks # Day # Hour Point 1: Calibration # Day # Hour Point 2: MB Type Factor: 1-Point Calibration Point None With CONC-0 # User defined * Lot specific calibrator values Page 7 of 13

QMS TACROLIMUS, AU5800 Parameters Specific Test Parameters General LIH ISE HbA1c Calculated Test Range Test Name: TAC Type: Serum Operation Yes Sample Volume 10.0 L Dilution 0 L OD Limit Pre-Dilution Rate 1 Diluent Bottle # Min.OD -2.00 Max.OD 3.00 Rgt. Volume R1(R1-1) 125 L Dilution 0 L Reagent OD Limit R1-2 L Dilution L First Low -2.00 High 3.00 Last Low -2.00 High 3.00 R2(R2-1) 75 L Dilution 0 L Dynamic Range Low 1.0 High 30.0 Common Rgt. Type None Name Correlation Factor A 1 B 0 Wavelength Pri 700 nm Sec. None nm Factor for Maker A 1 B 0 Method FIXED1 Reaction Slope + Onboard Stability Period # Day # Hour Measuring Point1 1 st 20 Last 27 LIH Influence Check # Measuring Point2 1 st Last Lipemia Linearity Limit % Icterus Lag Time Check No Hemolysis Parameters Specific Test Parameters General LIH ISE HbA1c Calculated Test Range Test Name: TAC Type: Serum Value/Flag: # Level Low # High # Specific Ranges: From To Sex Year Month Year Month Low High 1. # # # # # # # 2. # # # # # # # 3. # # # # # # # 4. # # # # # # # 5. # # # # # # # 6. # # # # # # # 7. Standard demographics # # 8. Not within expected values # # Panic Value Low # High # Unit ng/ml Decimal Places # Page 8 of 13

QMS TACROLIMUS, AU5800 Parameters Calibrators General ISE Calibration Specific Calibration Parameters Test Name: TAC Type Serum Cuvette. Use Serum Cal. Calibration Type: 6AB Formula: POLYGONAL Counts: 2 <Calibrator Parameters> Range Calibrator OD Conc Low High Slope Check - Point 1: # * -2.0 3.0 Point 2: # * -2.0 3.0 Allowance Range Check Point 3: # * -2.0 3.0 Point 4: # * -2.0 3.0 Reagent Blank Point 5: # * -2.0 3.0 Calibration Point 6: # * -2.0 3.0 Point 7: Advanced Calibration Point 8: Operation No Point 9: Point 10: Interval (RB/ACAL) <Point Cal. For No. of Correction Points Use Master Curve Lot Calibration Master Curve> OD Range Calibrator OD Conc Low High Stability Point-1 Reagent Blank # Day # Hour Point-2 Calibration # Day # Hour MB Type Factor: 1-Point Calibration Point None with Conc-0 # User defined * Lot specific calibrator values Page 9 of 13

Absorbance Results and Data Interpretation Performance Data Refer to the QMS Tacrolimus assay kit package insert for additional information on results and data interpretation. Example Calibration Curve, Tacrolimus (AU480): 0.25 Calibration Curve 0.20 0.15 0.10 0.05 0.00 0 5 10 15 20 25 30 Value (ng/ml) Page 10 of 13

Precision These degrees of precision and equivalency were obtained in typical testing procedures on an AU system and are not intended to represent the performance specifications for this reagent. Four levels of control samples were tested in replicates of 2, twice per day for 20 days, total N = 80 on the AU480/AU5800. The results are presented in the following table: Controls Control 1 Control 2 Control 3 Control 4 AU480 Mean (ng/ml) Within-Run SD (ng/ml) Within-Run CV (%) Total SD (ng/ml) Total CV (%) Mean (ng/ml) Within-Run SD (ng/ml) Within-Run CV (%) Total SD (ng/ml) Total CV (%) 3.6 10.5 17.6 25.9 0.12 0.29 0.24 0.45 3.4 2.7 1.3 1.7 0.26 0.70 0.63 0.95 7.2 6.6 3.6 3.7 AU5800 3.7 10.8 17.9 25.8 0.16 0.23 0.97 1.09 4.3 2.2 5.4 4.2 0.22 0.60 1.07 1.23 6.0 5.6 6.0 4.8 Three whole blood pools comprised of samples obtained from patients taking Tacrolimus and three negative whole blood pools spiked with Tacrolimus were tested in replicates of 2, twice per day for 20 days, total N = 80 on the AU680. The results are presented in the following table: Controls Spiked A Spiked B Spiked C Patient Pool A Patient Pool B Patient Pool C AU680 Mean (ng/ml) Within-Run SD (ng/ml) Within-Run CV (%) Total SD (ng/ml) Total CV (%) 3.0 10.0 20.9 3.2 10.4 24.2 0.15 0.19 0.39 0.13 0.23 0.52 4.9 1.9 1.9 4.1 2.2 2.1 0.21 0.36 1.05 0.20 0.37 1.11 7.1 3.6 5.0 6.2 3.6 4.6 Page 11 of 13

Linearity Eleven levels of calibrator and calibrator blends were run against a single calibration curve and the linearity calculated on the AU480 and AU5800. Ten levels of diluted calibrators were run against a single calibration curve and the linearity calculated on the AU680. The analytical range for this assay is 1.0 to 30.0 ng/ml. Error flags will appear for samples recovering above or below the assay range. The Tacrolimus assay recovered between 97 110% of expected values on the AU480. The Tacrolimus assay recovered between 96 103% of expected values on the AU680. The Tacrolimus assay recovered between 100 109% of expected values on the AU5800. Limit of Quantitation Six levels of low end known samples were run across 5 days, 5 replicates per level and their limit of quantitation calculated. The observed LoQ for Tacrolimus on the AU480 was 1.0 ng/ml. The observed LoQ for Tacrolimus on the AU680 was 0.9 ng/ml. The observed LoQ for Tacrolimus on the AU5800 was 0.8 ng/ml. Accuracy and Correlation One hundred and seven blood samples were assayed with the QMS Tacrolimus Assay on the Beckman Coulter AU480 and tested with reference method AU680. One hundred and eight blood samples were assayed with the QMS Tacrolimus Assay on the Beckman Coulter AU5800 and tested with reference method AU680. A Deming s Analysis yielded the following: AU480 = 1.00*(AU680) 0.08 with a correlation coefficient of 0.999. AU5800 = 1.02*(AU680) + 0.23 with a correlation coefficient of 0.998. Two hundred and sixty-six samples were assayed on the AU680 and compared to the LC-MS/MS method. A Deming s Analysis yielded the following: AU680 = 1.13*(LC-MS/MS) + 0.54 with a correlation coefficient of 0.965. Page 12 of 13

Additional Information Important Since Beckman Coulter does not manufacture the reagent or perform quality control or other tests on individual lots, Beckman Coulter cannot be responsible for the quality of the data obtained which is caused by performance of the reagent, any variation between lots of reagent, or protocol changes by the Manufacturer. Shipping Damage Please notify your Beckman Coulter Clinical Support Center if this product is received damaged. 2016 Thermo Fisher Scientific, Inc. All rights reserved. AU Series Systems are the registered trademarks of Beckman Coulter. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. EC REP Microgenics GmbH, Spitalhofstrasse 94, 94032 Passau, Germany Tel: +49 (0)851-88 6890/Fax: +49 (0)851-88 68910 End Page 13 of 13