COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION

Similar documents
HammerFUZE HAMMERTOE COMPRESSION SYSTEM SURGICAL TECHNIQUE

Contents. Chapter 1 4 Chapter Chapter Chapter Chapter 5 15

CANNULINK. Intraossous Fixation System SURGICAL TECHNIQUE

Implantable K-wire SURGIC A L T ECHNIQUE

Competitive Overview. 1.K-Wires. Arrowhead Med Technologies

Integra. DigiFuse Cannulated Intramedullary Fusion System SURGICAL TECHNIQUE

Zimmer Small Fragment Universal Locking System. Surgical Technique

Correction System. Surgical Technique

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

PHALINX. Hammertoe Fixation SURGICAL TECHNIQUE

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique

Correction System. Surgical Technique

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

Surgical Technique. Clavicle Locking Plate

Foot & Ankle. Smart Toe II. Intramedullary Implant. Operative Technique. Foot & Ankle

Technique Guide Hammertoe Correction System

The Vilex FUZETM. Dual Thread Screw & Intramedullary Nail in One Implant. The Ultimate TTC Arthrodesis Internal Fixator

SURGICAL TECHNIQUE. Alpine Cementless Hip Stem

DHS Plate 135º. DHS Platte 135º

Foot and Ankle Technique Guide Proximal Inter-Phalangeal (PIP) Fusion

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

Surgical Technique. Olecranon Locking Plate

Surgical Technique. Anterolateral and Medial Distal Tibia Locking Plates

DuaFit. Proximal Interphal angeal Impl ant

System. Humeral Nail. Surgical Technique

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

SURGICAL TECHNIQUE GUIDE

Surgical Technique. Distal Humerus Locking Plate

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

Arthrodesis of the Ankle

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

SMV Scientific Bone Plate and Screw System Surgical Technique

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System.

Surgical Technique. Proximal Humerus Locking Plate

Surgical Technique International Version. Clavicle Locking Plate

Surgical Technique. Calcaneal Locking Plate

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

Surgical Technique. 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates

LMH. Minimal Bone Resection. Lesser Metatarsal Head Implant. Thin, low-profile design for minimal bone resection

Flower Medium Headless & Cannulated Screws

HELIOS h i p s y s t e m

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Clinical Evaluation Surgical Technique

ORTHOLOC 3Di. Foot Reconstruction System Midfoot Fusion Plate SURGICAL TECHNIQUE

CORRECTING HAMMERTOE DEFORMITIES UTILIZING AN INTRAMEDULLARY DEVICE: Case Reports

RFS. Resorbable Fixation System SURGICAL TECHNIQUE

Astrid Modular Forefoot System. Surgical Technique

3.5 mm Locking Attachment Plate

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

DYNAMIC, TRANSVERSE COMPRESSION. Low Profile, Anatomic Design, Type II Anodized. Mechanical Compression Designed to Stimulate the Fusion Process

PediNail Pediatric Femoral Nail

Surgical Technique. Customer Service:

Forefoot/Midfoot Plating System. Surgical Technique

Zimmer Trabecular Metal Ankle Interpositional Spacer and Trabecular Metal Ankle Fusion Spacer

MPJ Implant System. Patents Pending Worldwide

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

PipTree. Arthrodesis of the PIP joint

SpeedTip CCS 2.2, 3.0

Surgical Technique. Targeter Systems Overview

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Fixed and Variable Geometry Total Shoulder Arthroplasty

Zimmer Natural Nail System

Distal Radius Plate 2.4/2.7 dorsal and volar

Lesser MPJ Hemi Implant

Biomet Large Cannulated Screw System

Conventus CAGE PH Surgical Techniques

MEDICAL ADVANCED TECHNOLOGY FOOT SURGERY: COMPRESSION AND VARISATION STAPLE SYSTEM FOOT SURGERY STAPLE FIXATION

Lapidus Arthrodesis System Instructions for Use

CrossTIE. PIP Arthrodesis Implant. surgical technique ordering information PEEK IMPLANTS PEEK IMPLANTS ALLOGRAFT IMPLANTS CROSSTIE INSERTER

The Flower Four Corner Fusion Plate

Surgical Technique. Forearm Fracture Solutions

SURGICAL TECHNIQUE GUIDE: JONES FRACTURE USING THE PRECISION JONES FRACTURE SCREW SYSTEM

Technique Guide. 6.5 mm Midfoot Fusion Bolt. For intramedullary fixation of the medial column of the foot.

Surgical Technique Guide

Clinical. Solutions. Synthes Solutions. Foot and Ankle.

Distal Radius Plate Instrument and Implant Set. Discontinued December 2017 DSUS/TRM/0916/1063(1)

Technique Guide. 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System.

TITANIUM TIBIAL NAIL SySTEM

Hand Fracture System. Surgical Technique

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

AcUMEDr. Olecranon Threaded Compression Rod

Foot & Ankle. EasyClip. Osteosynthesis Compression Staples. Foot & Ankle

S U R G I C A L T E C H N I Q U E David A. McQueen, MD Return to Menu

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

ORTHOLOC 3Di. Foot Reconstruction System SURGIC AL TECHNIQUE

VA LOCKING CALCANEAL PLATES 2.7

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Surgical Technique. Locking Small Fragment Overview

Surgical Technique 4.5/8.5MM BEAMING SYSTEM. Customer Service:

LCP Distal Tibia Plate

Flexible Fragment Fixation. Surgical Technique

Transcription:

C H A P T E R 3 0 COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION Scott R. Roman, DPM INTRODUCTION The most common fixation for proximal interphalangeal fusion is provided by a smooth 1.1 mm (0.045") or 1.6 mm (0.062") Kirschner wire (K-wire) placed in an antegrade manner through the middle and distal phalanges while proximal interphalangeal joint extension and distraction are maintained. It is then placed in retrograde fashion into the proximal phalanx. Pin fixation is necessary for 4-6 weeks after surgery (1). The pins are capped to prevent the sharp ends of the pins from catching on objects such as the patient s bed sheets. (Figure 1) Although percutaneous K-wires are effective, nonunions can be quite common. Other treatment challenges associated with K-wire fixation include migration and loss of fixation (2). Issues such as pin tract infections as well as difficult postoperative management by patients make alternative fixation methods desirable (3). Having used percutaneous K-wires to provide fixation to the resected proximal interphalangeal joint, surgeons seek a means of providing fixation that is more stable than a K-wire as well as to mitigate the difficult postoperative management required by patients receiving this type of fixation. This article compares test results of the K-wire and an intramedullary device with a new design. Figure 1. Postoperative stabilization with K-wires. IMPLANT DESIGN CRITERIA In planning and research prior to design of the implant, the surgeon design team identified performance objectives for the new implant that would offer clinical advantages over current implant alternatives. These design imperatives included: Stable fixation Keep completely intramedullary with no parts of the implant exposed outside the skin. Maintain the initial compression applied at insertion. Resist the rotational and pull-out forces affecting the lesser toes. Simple implant design Avoid introducing potential stress-risers and the need to assemble intraoperatively by developing a one-piece design. Provide anatomic sizing to fit a broad range of patient anatomy. Fit the patient s natural toe anatomy by providing a neutral and 10 plantar angle. Minimize bone removal with a low-profile design. Uniform strength characteristics throughout the implant. Ensure the product material has a proven long term biocompatibility so it can be safely implanted permanently. Will not require removal like the K-wire. Efficient technique Create an operative technique similar to that for K-wire fixation with little additional complexity, such as drilling and tapping necessary to accommodate a threaded device. Remove little or no bone when preparing the bone for implant insertion. Proven material Must be biocompatible, clinically proven, extensively studied and commonly used for permanently implanted medical devices.

CHAPTER 30 149 Bending and fatigue characteristics able to endure the forces exerted on the lesser toes without having to add unnecessary material volume. Capable of being efficiently manufactured to ensure reproducible results. Does not require special handling or storage. Simple instrumentation Instrumentation resembles those with which surgeons are familiar. Emphasize simplicity to ensure reproducible results. IMPLANT DESIGN The Arrow-LOK Digital Fusion System (Arrowhead Medical Device Technologies, Collierville, TN) includes an implant (Figure 2) designed to improve upon the performance of the K-wire, the accepted standard for fixation of osteotomies, arthrodesis, and reconstruction in the lesser toes following corrective procedures. The surgeon-designers maintained the same basic surgical technique as the one to which they were well accustomed. The device provides more stable fixation than the K-wire and reduces the incidence of complications that can be caused by pin tract infections. Since the Arrow-LOK is intended to be permanently implanted, the surgeon avoids having to remove the device. Machined from a single piece of 316L stainless steel, the Arrow-LOK is a solid, 1.5 mm (0.059") diameter shaft with a three-dimensional arrow shape at each end. The implants are available in 3 lengths (16 mm, 19 mm, 22 mm) to fit the lesser toes of a wide range of patients. It also comes in 2 different angles: one with a neutral orientation, the other with a 10 plantargrade angle (Figure 3). PERFORMANCE TESTING The Arrow-LOK implant was compared to the K-wire in a series of performance tests. The results of the tests confirmed that the design characteristics of the Arrow-LOK improve upon the fixation performance of the K-wire (Table 1). Dynamic Four-Point Bend Testing Four-point bend testing was conducted by the Georgia Institute of Technology s Mechanical Properties Research Laboratory in accordance with the methods prescribed in ASTM F 564-02, Standard Specification and Test Methods Figure 2. Arrow-LOK Digital Fusion System Implant (19 mm). 3-dimensional arrow 3.5 mm wide 3.0 mm long 40% F138-08 (316L) stainless steel 1.5 mm (0.059 ) diameter 60% Available in lengths of 16 mm, 19 mm, and 22 mm Straight and angled options Figure 3. Arrow-LOK implant design. Table 1 10 plantar bend TESTING SUMMARY TABLE Test Dynamic Four Point Bend Testing Resistance to Pullout Rotational Stability Results Equivalent to the K-wire in fatigue testing Over 20 times more resistant to double-sided pull-out than a K-wire. Over 10 times more resistant to rotational forces causing initial deviation than the K-wire

150 CHAPTER 30 for Metallic Bone Staples. In this test, the Arrow-LOK device with a 3.5 mm diameter arrow, 1.5 mm shaft diameter and 50 mm in length was compared to the K-wire (1.6 mm diameter X 50 mm long). An Arrow-LOK implant 50 mm in length was tested since this represented the worst case as the bending moment increases with length. Results: The Arrow-LOK implant produced comparable test results to a K-wire in tests determining the force displacement response and the stress displacement response. The peak maximum and minimum stresses for the Arrow-LOK implant were also comparable to those of the K-wire. The small variance in the results was expected and is attributed to the difference in the diameter between the 2 devices (0.10 mm). The Arrow-LOK implant exhibited equivalent appearance to the K-wire after being subject to 250,000 cycles in fatigue testing, showing no signs of cracking or any other indications of fatigue. Conclusion: The Arrow-LOK is equivalent to the K-wire in fatigue testing. There are no stress-risers introduced into the Arrow-LOK as a result of any of its design features. Resistance to Pull-out Testing was conducted to evaluate the force needed to pull the 3.5mm diameter Arrow-LOK and a 0.062" (1.6 mm) K-wire out of polyurethane block and bone models. Results: The mean force required for initial deviation of the Arrow-LOK was 21.9 Newtons (N) compared to Table 2 PULL-OUT TESTING SUMMARY RESULTS 0.062" (1.6mm) Kwire Arrow-LOK Single-Sided Pullout: Mean Force Causing 1.5N 13.9N Initial Deviation Single-Sided Pullout: Mean Force Causing 2.1N 29.9N Complete Pull-out Double-Sided Pullout: Mean Force Causing 1.9N 21.9N Initial Deviation Double-Sided Pullout: Mean Force Causing 2.9N 62.1N Complete Pull-out 1.9N for the K- wire. Complete pull-out of the devices tested required a mean of 62.1N and 2.9N respectively (Table 2). Conclusion: The Arrow-LOK device is significantly superior to the K-wire with respect to the resistance to pull-out forces. Rotational Stability Testing was conducted to compare the rotational forces necessary to produce movement in a polyurethane bone implanted with either the 3.5 mm diameter Arrow-LOK or a 0.062" (1.6 mm) K-wire. Results: The mean force required for initial deviation of the Arrow-LOK was 19.3 Newtons (N) compared to 1.6N for the K-wire. Rotation to a point greater than 20 degrees required a mean of 29.6N and 3.2N respectively. The data indicates the Arrow-LOK exhibited significantly greater resistance to rotational forces than the K-wire (Table 3). Conclusion: The Arrow-LOK is significantly superior in its resistance to rotation than the K-wire. SURGICAL TECHNIQUE: INSERTION The surgical technique recommends preoperative templating to ensure the intramedullary canal can accommodate the 3.5 mm width of the implant. Proper use of instrumentation does not remove cancellous bone but rather compacts it against the cortical bone to create a stable foundation of compacted cancellous bone against the implant. After gaining exposure to the joint, resect the head of the proximal phalanx and the base of the intermediate phalanx bones. Table 3 ROTATIONAL TESTING SUMMARY RESULTS 0.062" (1.6mm) Kwire Arrow-LOK Mean Force Causing 1.6N 19.3N Initial Deviation Mean Force Causing 3.2N 29.6N Deviation >20

CHAPTER 30 151 Figure 4. Figure 5. Figure 6. Figure 7. Ream Create a pilot hole in the intramedullary canals of both the proximal (Figure 4) and intermediate (Figure 5) phalanx with the Arrow-LOK Reamer. The reamer is a 152 mm long 316L stainless steel wire with a diameter of 1.6 mm. It has a smooth trocar tip that creates a pilot hole without removing bone. The 5 mm graduations from the tip serve as a depth gauge for implant insertion. Broach Prepare the pilot hole on both the proximal (Figure 6) and intermediate (Figure 7) phalanges with the Arrow-LOK Broach. Broach each phalanx to the depth estimated during preoperative planning. Make note of the depth marking to which the broach was inserted in each phalanx. The Arrow-LOK Broach prepares the pilot hole that was just created by the reamer to create a shape in the intramedullary canal that will accept the three-dimensional arrowhead tip of the Arrow-LOK implant. The Broach conserves bone by compacting the cancellous bone of the phalanx to enhance the engagement of the Arrow-LOK implant s arrowhead tip. Insert and Compress Grasp the Arrow-LOK implant with the Arrow-LOK Insertion Tool at the depth corresponding to the broaching depth of the proximal phalanx. The Arrow-LOK Insertion Tool firmly grasps the Arrow-LOK implant preventing both rotation and translation to ensure the implant is placed securely in the broached pilot hole. Insert the Arrow-LOK into the proximal phalanx (Figure 8), locking it into the intramedullary canal. Grasp the digit and insert the distal aspect of the implant into the entry portal prepared in the intermediate phalanx (Figure 9) locking the implant into the intramedullary canal (Figure 10). Release and remove the Insertion Tool. Grasping the digit firmly, compress the two bones together to advance the implant both proximally and distally to its final locked position. Close the wound in accordance with generally accepted surgical technique.

152 CHAPTER 30 Figure 8. Figure 9. Figure 10. OPTIONAL: SURGICAL TECHNIQUE FOR REMOVAL In the event that the Arrow-LOK requires removal, standard instrumentation common to most surgical centers and hospitals facilitates the procedure. The following is an overview of the recommended technique for removal of the Arrow-LOK. 1) Access the surgical site through generally accepted surgical technique. Expose the Arrow-LOK implant with resection of the bone or fibrous tissue using either hand or power instrumentation. 2) Cut the Arrow-LOK implant using a wire cutter (Figure 12) exposing the shaft of the implant into the surgical site on both the proximal and intermediate phalanges (Figure 13). 3) Resect the bone around the implant shaft utilizing a side cutting burr on power instrumentation or a 3.5 mm cannulated drill bit. Note that the 3.5 mm cannulated drill bit is the same width as the head of the 3.5 mm Arrow-LOK implant. 4) Figure 11. Postoperative radiograph of the Arrow- LOK. The implant shaft is now exposed (Figure 14). 5) Grasp the exposed implant shaft with surgical needle nose pliers or a straight hemostat (Figure 15). Pull the implant from surgical site. CONCLUSION The most common fixation for proximal interphalangeal fusion is provided by a smooth K-wire. Although percutaneous K-wires have been used extensively for years, complications are well-documented. Surgeons have attempted to overcome the shortcomings of K-wires with a variety of surgical techniques, designs and materials, each presenting its own compromises and trade-offs. After

CHAPTER 30 153 Figure 12. Figure 13. Figure 14. considering the key objectives of proximal interphalangeal fixation, surgeons developed an implant combining the benefits of K-wires with a design offering more stable fixation. The Arrow-LOK Digital Fusion System exhibits the potential to offer surgeons an alternative solution to proximal interphalangeal fixation that will provide patients a more effective clinical outcome than current treatment alternatives. Figure 15. REFERENCES 1. Campbell s Operative Orthopaedics, ed. S. Terry Canale, MD (Philadelphia, Mosby, Inc., 2003), 4060. 2. Harmonson JK, Harkless L. Operative procedures for the correction of hammertoe, claw toe, and mallet toe: a literature review. Clin Podiatr Med Surg 1996;13:211-20. 3. Clinical Policy Bulletin: Hammertoe Repair. Aetna, Inc. November 13, 2009.http://www.aetna.com/cpb/medical/data/ 600_699/0636.html (June 2010) p3