Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Similar documents
Leila E. A. Nichol Royal Surrey County Hospital

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

EPID dosimetry confi guration and pre-treatment IMRT verifi cation

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Volumetric Modulated Arc Therapy - Patient Specific QA

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

IMRT QA: Point Dose Measurements or 2D Array?

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams

Follow this and additional works at:

Small field diode dosimetry

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates

Verification of Advanced Radiotherapy Techniques

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

REPLACING PRETREATMENT VERIFICATION WITH IN VIVO EPID DOSIMETRY FOR PROSTATE IMRT

IMRT point dose measurements with a diamond detector

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

EORTC Member Facility Questionnaire

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Optimized Varian asi portal dosimetry: development of datasets for collective use

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Feasibility of portal dosimetry for flattening filter-free radiotherapy

Limits of Precision and Accuracy of Radiation Delivery Systems

EPID dosimetry for pretreatment quality assurance with two commercial systems

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Portal dosimetry in wedged beams

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

Eric E. Klein, Ph.D. Chair of TG-142

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

Patient and Linac QA - present and the future

IBA Dosimetry Company Presentation. Patient Safety & Quality Control

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Hampton University Proton Therapy Institute

SunCHECK Patient Comprehensive Patient QA

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

Radiotherapy In-vivo Treatment Verification: Multi-center analysis for more than 6 million fractions delivered

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Activity report from JCOG physics group

iviewdose Confidence and assurance in dose delivery

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Future upcoming technologies and what audit needs to address

Pre-treatment verification for IMRT / IGRT and Rotational therapy

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

A new transmission methodology for quality assurance in radiotherapy based on radiochromic film measurements

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

DELIVERED PATIENT DOSE

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Inhomogeneity correction and the analytic anisotropic algorithm

Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea

Application(s) of Alanine

Unrivaled, End-to-End

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Gamma analysis dependence on specified low-dose thresholds for VMAT QA

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

Online Care, Online Control, Online Confidence

Agenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry

IMRT pre-treatment verification basic approaches

Diode calibration for dose determination in total body irradiation

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate

Efficient Dosimetry for Proton Therapy

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

Comparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha, Niko Papanikolaou, Sotirios Stathakis a Department of Radiation Oncology, University of Texas Health Science Center at San Antonio, TX, USA stathakis@uthscsa.edu Received 05 November, 2014; accepted 19 December, 2014 The purpose of this study is to evaluate the use of the Dosimetry Check system for patient-specific IMRT QA. Typical QA methods measure the dose in an array dosimeter surrounded by homogenous medium for which the treatment plan has been recomputed. With the Dosimetry Check system, fluence measurements acquired on a portal dosimeter is applied to the patient s CT scans. Instead of making dose comparisons in a plane, Dosimetry Check system produces isodose lines and dose-volume histograms based on the planning CT images. By exporting the dose distribution from the treatment planning system into the Dosimetry Check system, one is able to make a direct comparison between the calculated dose and the planned dose. The versatility of the software is evaluated with respect to the two IMRT techniques step and shoot and volumetric arc therapy. The system analyzed measurements made using EPID, PTW seven29, and IBA MatriXX, and an intercomparison study was performed. Plans from patients previously treated at our institution with treated anatomical site on brain, head & neck, liver, lung, and prostate were analyzed using Dosimetry Check system for any anatomical site dependence. We have recommendations and possible precautions that may be necessary to ensure proper QA with the Dosimetry Check system. PACS number: 87.55.Qr, 87.56.Fc Key words: dosimetry check, IMRT QA, QA techniques, dosimetry I. INTRODUCTION An early technique used for patient-specific quality assurance (QA) was combining an ionization chamber and film. Ion chambers provide excellent stability, linear response to absorbed dose, small directional dependence, beam-quality response independence, and traceability to a primary calibration standard. (1,2) Films are convenient to use, provide excellent spatial resolution, and acts as a permanent record of the integrated planar dose distributions. (3) By using a water-equivalent phantom with the capability of supporting an ion chamber and films, one could verify both absolute and relative dose for an intensity-modulated radiation therapy (IMRT) treatment with any analyzing software. Two-dimensional (2D) arrays provide a quick and efficient means of performing patientspecific QA by eliminating the need for film processing. These detectors typically employ diode or ion chamber array which are calibrated to provide both absolute and relative dose measurements immediately after irradiation. Although electronic portal imaging dosimeter (EPID) was designed to provide patient position verification, interest in portal dosimetry is growing due to high spatial resolution, large imaging area, stability, dynamic range, and real-time acquisition of images. (4,5,6) Typically, the a Corresponding author: Sotirios Stathakis, Department of Radiation Oncology, University of Texas Health Science Center San Antonio, 7979 Wurzbach Rd., San Antonio, TX 78229, USA; phone: (210) 450 1010; fax: (210) 450 1076; email: stathakis@uthscsa.edu 329 329

330 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 330 use of an EPID for patient-specific QA is limited to 2D dose distributions. The Dosimetry Check system investigated in this study has taken this approach one step further by employing a pencil beam algorithm in order to calculate dose from fluence measurements taken with the EPID. (7) It can be assumed that some dose differences will arise from the pencil beam algorithm used in Dosimetry Check and the more sophisticated algorithms used in TPS, but it is believed that these differences should be no greater than 5%. (7) II. MATERIALS AND METHODS In the first part of the study, linearity of Dosimetry Check system with dose delivered is studied using a 10 cm 10 cm 6 MV photon beam in order to ensure that the system operates as desired. All measurements were taken with a 6 MV photon beam produced by a Varian Novalis TX linac (Varian Medical Systems, Palo Alto, CA) equipped with a HD120 MLC system. Pinnacle 3 TPS (Philips Medical Systems, Andover, MA) used at our institution employs a convolution/superposition algorithm. Frame rate dependence of the system is studied with an integrated image for step-and-shoot IMRT plans versus a cine-mode continuous image for the arc therapy plans. The dependence of the Dosimetry Check system on the detectors was analyzed by comparing EPID based measurements against other QA detectors including PTW seven29 ion chamber array (PTW, Freiburg, Germany) and IBA MatriXX ion chamber array (IBA Dosimetry, Bartlett, TN). The amorphous-si (asi1000) EPID is an array of 1024 768 a-si based photodiodes arranged in a 40 30 cm 2 active area. The PTW seven29 array consists of 729 vented ionization chambers (0.125 cc active volume) arranged in a matrix of 27 27 chambers with center-to-center spacing of 10 mm. (8) The IBA I mrt MatriXX array comprises of 1020 vented ion chambers (0.08 cc active volume) arranged in a matrix of 32 32 chambers with center-to-center spacing of 7.62 mm. (9) As a part of this study, treatment plans from five different anatomical sites (brain, head and neck, liver, lung, and prostate) were chosen to check for any anatomical site dependence of the EPID-based Dosimetry Check analysis. For dosimetric verification, gamma analysis was performed using 3% dose difference, 3 mm distance-to-agreement criteria, with a 5% dose threshold upon normalizing to the prescription dose. A. Verification of linearity Linearity of the QA dosimetric system allows for the usage of a single integrated image to calibrate all fields. To verify the linearity of the detectors, five fields of size 10 cm 10 cm with increasing MUs are irradiated on the EPID, seven29, and MatriXX. For the purposes of this study, both the seven29 and MatriXX were positioned on the couch at a source-to-detector distance (SDD) of 100 cm. However, EPID dosimeter was set at the highest possible setting of 110 cm SDD, allowing us to protect the electronics of EPID. This necessitates the Dosimetry Check software to make the distance corrections for differences in SDD. The integrated images were then imported into the Dosimetry Check system and the signal values for each image were plotted against the MUs to establish the calibration file. B. Software evaluation The Dosimetry Check software was used to evaluate patient-specific QA on the two IMRT techniques step and shoot, as well as smart arc with measurements based on EPID set to 110 cm SDD. With step-and-shoot IMRT plans, it is only necessary to have one integrated image per field. In a smart arc plan, a continuous image technique allows the EPID to collect multiple integrated images at a frame averaging selected by the user. A balance between the timer resolution and the number of images was achieved using a frame averaging that gives an image within every five degrees of gantry rotation. Frame averaging is selected according to the gantry rotation speed and the degree of modulation. It is possible that a loss of signal may

331 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 331 occur in the duration between two consecutive image acquisitions and, therefore, an integrated image of each arc is also taken. The Dosimetry Check system applies a correction factor of the ratio of the sum of the signal from the continuous images (acquired at every 50 gantry rotation) and the single integrated image (acquired for the whole arc) during dose reconstruction. Gamma analysis was then used to compare the measured data with the reference data using a 3%/3 mm criteria and a 5% dose threshold after normalizing to the prescription dose. The Dosimetry Check system uses integrated images of radiation fields in order to calculate dose intended to be delivered. The X-ray intensity map measured with EPID is converted to relative monitor units by using a calibration file. The calibration files are created by taking the central axis value of a 10 10 cm 2 image with a known amount of monitor units (MU). This process is repeated with increasing MUs to study the linearity of Dosimetry Check system. (10) Once the integrated images are calibrated, they are transformed into incident intensity fluence by spatial filtering with a deconvolution kernel. (11) The fluence fields can then be back-projected (shown in Fig. 1) and the dose is calculated using a pencil beam algorithm (shown in Fig. 2). Fig. 1. Back-projection and pencil beam calculation. Fig. 2. Pencil beam algorithm used in dosimetry check.

332 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 332 The pencil beam was developed from the commissioning data and the Monte Carlo calculated polyenergetic kernel. (7) After importing the CT scans, plan file, structure file, and dose file from the treatment planning software (TPS), a direct comparison between computed dose and planned dose can be made not only in a plane but in the entire body or specific organs of the patient. The Dosimetry Check system also provides the user with isodose lines, point doses, dose-volume histograms (DVH), and gamma analysis in order to make these direct comparisons. C. Detector dependence Fifteen step-and-shoot IMRT plans from various treated anatomical sites were chosen from patients previously treated at our institution. Integrated images were taken for each field for the 15 plans with all three detectors, with setup as shown in Fig. 3. Each field was delivered with the gantry set to upright position and SDD of 110 cm for the EPID and 100 cm for both the PTW seven29 and IBA I mrt Matrixx. A 10 cm 10 cm field with 100 MUs was delivered for calibration to account for the daily variations of the linac. The images were then imported into the Dosimetry Check system along with the CT scans, structure file, plan file, and dose distribution file exported from the TPS. The Dosimetry Check system uses the imported fluence fields (shown in an example in Fig. 4) to calculate the absolute dose, and isodose curves (example shown in Fig. 5) to compare with the reference data. Gamma analysis was used to compare the measured data with the reference data using the above-mentioned criteria. Bland- Altman test is performed in analyzing the agreement between two methods of measurement. In this study, pairwise comparison between EPID and seven29, EPID and MatriXX, and seven29 and MatriXX were performed using Bland-Altman tests to identify the agreement and the presence of any bias. Fig. 3. Picture of the setup: (a) EPID, (b) PTW seven29 array, and (c) IBA I mrt MatriXX. Fig. 4. Calibrated fluence fields for (a) EPID, (b) PTW seven29, and (c) IBA I mrt MatriXX.

333 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 333 Fig. 5. Isodose lines in the reference data from TPS (green) and calculated with Dosimetry Check system (magenta) on the same patient with (a) asi1000 EPID, (b) PTW seven29, and (c) IBA I mrt MatriXX. D. Anatomical site dependence The dose algorithm used in the Dosimetry Check system is a pencil beam algorithm which differs from the convolution/superposition algorithm widely used in the TPSs. Differences in these algorithms may lead to discrepancies in the dose calculated with the Dosimetry Check system and may depend on the level of heterogeneity of the anatomical site. In the anatomical site-dependence study, EPID-based measurements were collected on 23 step-and-shoot IMRT plans from five different anatomical sites (brain, head & neck, liver, lung, and prostate). Dose calculated by Dosimetry Check system was compared with the TPS reference dose using 3D gamma analysis. III. RESULTS In the linearity study, five measurements were taken at each MU setting to verify the consistency in the measurements. The mean values were plotted and a linear fit with the regression calculated was applied to each set of data, as shown in Fig. 6. The R 2 value for each plot was 1, clearly showing the dose response of each detector is linear. The standard deviation (SD) in the measurements is < 1%, except for seven29 and MatriXX where it is 3.7% and 1.3% at 2MUs, respectively. Mean percent SD values decrease from 1.74% to 0.05% with increasing MUs, showing more consistency across the three detectors. In evaluating the Dosimetry Check software on step-and-shoot IMRT plans and arc therapy plans, ten plans of each type from patients previously treated at our institution were measured. While integrated images were acquired in the step-and-shoot IMRT plans, continuous images were measured in the arc therapy plans treatment type. Gamma analysis between the measured and the reference plan dose yielded mean ± SD of 96% ± 1% for the step-and-shoot IMRT plans. The agreement is slightly higher at 98.9% ± 0.6% for the arc therapy plans, although no statistically significant difference was observed between the two treatment techniques using a two-tailed Student s t-test (p > 0.1). In the detector dependency study, integrated images were taken for each field of the 15 stepand-shoot IMRT plans with all three detectors. These plans correspond to various anatomical sites from patients treated earlier at our institution. The measured detector fluence data and the patient CT scans were imported into the Dosimetry Check system to calculate DVH of various organs. Figure 7 shows the DVH calculated by Dosimetry Check system and the reference DVH from the TPS for a head/neck IMRT plan. Gamma analysis was used to compare the measured and reference data using 3%/3 mm criteria and a 5% dose threshold was used after normalizing to the prescription dose. The gamma values for the 15 plans are listed, along with the mean and SD value, in Table 1. Three Bland-Altman tests were performed in order to compare the measured data and results are shown in Fig. 8. Each Bland-Altman test performed a pairwise

334 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 334 Fig. 6. Output signal linearity plots against MUs in a 10 cm 10 cm 6 MV open beam for (a) EPID, (b) seven29, and (c) MatriXX. Fig. 7. Dosimetry Check system calculated DVH comparing the asi1000 EPID (solid), PTW seven29 (dashed), IBA I mrt Matrixx (dashed/dotted), with reference DVH from TPS (dotted) for a head/neck plan. Table 1. Gamma values calculated by Dosimetry Check system with fluence from EPID, seven29, and MatriXX on 15 IMRT plans. Anatomical Site EPID seven29 Matrixx Abdomen 98.85% 98.08% 94.90% Brain 1 97.77% 95.09% 93.51% Brain 2 94.03% 94.30% 95.03% Brain 3 99.26% 98.45% 97.70% Head and Neck 1 92.51% 92.68% 90.62% Head and Neck 2 98.59% 93.74% 96.33% Head and Neck 3 96.86% 96.48% 95.56% Liver 1 99.09% 97.93% 98.22% Liver 2 99.62% 98.90% 99.13% Liver 3 99.46% 97.61% 97.13% Lung 1 93.71% 92.19% 93.64% Lung 2 95.66% 92.34% 93.38% Lung 3 96.12% 95.59% 93.59% Prostate 1 93.85% 92.60% 92.61% Prostate 2 96.28% 94.82% 93.68% Average 96.78% 95.39% 95.00% SD 2.41% 2.40% 2.34%

335 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 335 Fig. 8. Results from Bland-Altman test analysis comparing gamma values between (a) EPID PTW seven29, (b) EPID MatriXX, and (c) PTW seven29 MatriXX. comparison between the gamma values calculated using imported fluence from two of the detectors. The bias ± SD values were 0.01391 ± 0.01357 and 0.01775 ± 0.01379 for comparisons between the EPID and seven29 and MatriXX, respectively. The bias value for the comparison between the 2D arrays was 0.003847 ± 0.01515. In the anatomical site dependency study, 23 IMRT plans were chosen from patients previously treated at our institution. The Dosimetry Check system calibrated the imported fluence fields collected as integrated images on EPID along with the reference data from the TPS to calculate the absolute dose, isodose curves, and DVHs. Shown in Fig. 9 are the reference and calculated isodose curves and DVHs for five treatment plans brain, head/neck, liver, lung and prostate. The gamma values for the 23 plans are listed in Table 2 and the average SD values are plotted according to anatomical site in Fig. 10. Small anatomical site dependence was observed. Of the sites investigated, liver and brain patients showed the strongest agreement between calculated and reference data, with mean gamma values of 99.15% and 97.38%, respectively. The remaining sites had mean gamma values between 94.49% and 95.16%. Of note is the lung plan where an over estimation of dose to the PTV in the lung is observed in the pencil beam algorithm, as shown in Fig. 9(d).

336 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 336 Fig. 9. DVH and isodose lines in the reference data from TPS (green) can be compared with those calculated by the Dosimetry Check system (magenta) for: (a) brain, (b) head and neck, (c) liver, (d) lung, and (e) prostate treatment plans.

337 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 337 Table 2. Dosimetry Check system calculated gamma values on 23 IMRT patient plans to various anatomical sites. Anatomical Site Gamma Brain 1 97.77% Brain 2 94.03% Brain 3 99.26% Brain 4 97.10% Brain 5 98.72% Head and Neck 1 96.41% Head and Neck 2 98.59% Head and Neck 3 92.99% Head and Neck 4 90.59% Head and Neck 5 96.86% Liver 1 98.75% Liver 2 99.09% Liver 3 98.81% Liver 4 99.62% Liver 5 99.46% Lung 1 93.71% Lung 2 95.66% Lung 3 96.12% Prostate 1 93.85% Prostate 2 96.28% Prostate 3 93.79% Prostate 4 94.91% Prostate 5 93.60% Fig. 10. Breakdown of gamma analysis values from the EPID Dosimetry Check system calculated data according to the anatomical site on the 23 IMRT plans. Error bars represent 1 SD. IV. DISCUSSION The first aspect of the software we investigated was the ability to perform patient-specific quality assurance with different treatment techniques. The linear dose response characteristic of each detector allowed us to take a single image for calibration for each patient QA. A calibration image was taken on the day of making patient-specific QA measurements instead of using a calibration file, in order to take into account the daily output variation of the linac. In our institution we currently employ both step-and-shoot treatments, as well as smart arc treatments, and the Dosimetry Check software was able to perform adequately for both treatments. The difference between testing these two treatment techniques lies in the image acquisition mode. A slightly higher gamma value is observed in the arc plans, although the difference is negligible. The next aspect that was investigated is the ability of the system to use fluence maps from various detectors for patient-specific QA. The Dosimetry Check system is not restricted to

338 Narayanasamy et al.: Evaluation of Dosimetry Check software for IMRT QA 338 work only with EPID, but allows fluence maps from a number of different detectors. A group of 15 step-and-shoot IMRT patient plans were chosen from various anatomical sites and fluence maps were taken with three separate detectors, the asi1000 EPID, PTW seven29, and IBA I mrt Matrixx. Gamma analysis and Bland-Altman tests showed that all three detectors performed adequately, with negligible differences. In the anatomical site dependence test, 23 IMRT plans from five anatomical sites with varying levels of inhomogeneity were investigated. Inherent differences between a pencil beam algorithm employed in Dosimetry Check system and the superposition/convolution algorithm used in Pinnacle 3 TPS can result in dose discrepancies between the calculated and reference dose. The pencil beam algorithm used in the Dosimetry Check system is not expected to compute dose accurately at the internal interfaces, especially when compared with a convolution/ superposition algorithm. (7) Pencil beam algorithms are known to overestimate dose in low density inhomogeneity areas. (12) In one of the lung plans shown in Fig. 9, the dose calculated for the PTV is higher using a pencil beam algorithm compared to the reference DVH. While some anatomical site dependence does occur, our investigation showed that the differences are small and the Dosimetry Check system was able to perform well in QA for all anatomical sites. Caution is recommended with the use of this system when inhomogeneity exists, and it is recommended to take a close look at the DVH and isodose lines. V. CONCLUSIONS Dosimetry Check system is a very versatile software in its ability to perform patient-specific QA for various treatment techniques, anatomical sites, and detectors that can be used. Most other QA techniques can only perform a gamma analysis in a single plane. By employing its own dose calculation algorithm, the dose can be calculated in three dimensions, and the isodose lines and DVH can be generated for comparison against the reference data. Another advantage in employing a pencil beam algorithm is the ability to calculate dose in less time. An arc treatment has over 100 images and dose calculation time ranging from a few seconds up to a minute per image, could become a concern. Overall, the Dosimetry Check system provides a versatility and wealth of knowledge that can provide a strong base for a clinical patient-specific QA program. REFERENCES 1. Low D, Moran JM, Dempsey JF, Dong L, Oldham M. Dosimetry tools and techniques. Med Phys. 2011;38(3):1313 38. 2. Spezi E, Angelini AL, Romani F, Ferri A. Characterization of a 2D ion chamber array for the verification of radiotherapy treatments. Phys Med Biol. 2005;50(14):3361 73. 3. Pai S, Das IJ, Dempsey JF, et al. TG-69: Radiographic film for megavoltage beam dosimetry. Med Phys. 2007;34(6):2228 58. 4. Warkentin B, Steciw S, Rathee R, Fallone BG. Dosimetric IMRT verification with a flat-panel EPID. Med Phys. 2003;30(12):3143 55. 5. Van Esch A, Depuydt T, Huyskens DP. The use of an asi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. Radiother Oncol. 2004;71(2):223 34. 6. Wendling M, Louwe RJ, McDermott LN, Sonke JJ, van Herk M, Mijnheer BJ. Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry model. Med Phys. 2006;33(2):259 73. 7. Math Resolutions, LLC. Dosimetry Check: quality control system for radiation therapy conventional linac and tomotherapy, pre-treatment and exit-transit dosimetry. Columbia, MD: Math Resolutions; 2010. 8. PTW-Freiburg. User manual: 2D-Array Seven29 and Array interface. Frieburg: PTW; 2013. 9. IBA Dosimetry. I mrt MatriXX: user s guide. Schwarzenbruck, Germany: IBA; 2013. 10. McCurdy BM, Luchka K, Pistorius S. Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device. Med Phys. 2001;28(6): 911 24. 11. Renner WD, Norton K, Holmes T. A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction. J Appl Clin Med Phys. 2005;6(4):22 39. 12. Knoos T, Ahnesjo A, Nilsson P, Weber L. Limitations of a pencil beam approach to photon calculations in lung tissue. Phys Med Biol. 1995;40(9):1411 20.