Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Similar documents
Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

The white blood cell differential (WBC-diff) technologies

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

Sysmex Journal International Vol.9 No.2 (1999)

Clinical Utility. Clinical Utility of the Automated IG-Count Xavier Troussard

XN series. Case interpretation. Gebruikersdag Vlaanderen- 6 oktober 2016

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD

Customer Information Literature List White Blood Cells

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

Performance evaluation of Celltac G: a new automated hematology analyzer

ADx Bone Marrow Report. Patient Information Referring Physician Specimen Information

INFECTION/ INFLAMMATION

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital

LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP)

Identification and Enumeration of Nucleated Red Blood Cells in Peripheral Blood

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT

LEANING Away Waste and Improving Patient Care with Advances in Hematology

Flow cytometry leukocyte differential : a critical appraisal

Differential Blood Smear H3

HumaCount 5D. Outperforming 5-part Hematology System. CoreLab DX. Hematology

XN-SERIES. XN Technology and Case Studies

Combining. and New Diagnostic. to Help Clinicians Achieve. Patient Outcomes at. per Healthcare Encounter

XN-SERIES. Hematology Case Studies: Every Picture Tells a Story

Sysmex Journal International Vol.26 No.1 (2016)

10/30/2015. XN Case Studies: Every Picture Tells a Story

Lavender Top Management SUCCESS BEYOND FINANCES

Detection of important abnormalities of the differential

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision

Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network

A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter

Biochemical and Hematological Markers of Inflammation Accurately Predict Sepsis and its Severity in ICU Patients

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies

The Differential Cell Count

HISTOLOGY VIRTUAL LABORATORY BLOOD AND LYMPHATICS SYSTEM

Pseudobasophilia in Pediatric Age Group

HSCT MANAGEMENT WHITE PAPER. Managing stem cell apheresis effectively

Use of the XE-2100 in a Patient with Cold Auto-immune Hemolytic Anemia

5/1/2017 DISCUSSION POINTS. Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN

We noted a problem with the use of the hematology

Notes for the 2 nd histology lab

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana

Validation of Criteria for Smear Review Following Automated Blood Cell Analysis in Ain Shams University Laboratory

Customer Information Literature List Platelets

Cbc with differential

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist

Evaluation of the QBC Star centrifugal three-part differential haematology system

2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts

Literature List White Blood Cells

Hematopathology Case Study

rev bras hematol hemoter. 2014;36(3): Revista Brasileira de Hematologia e Hemoterapia Brazilian Journal of Hematology and Hemotherapy

Performance evaluation and relevance of the CellaVision TM DM96 system in routine analysis and in patients with malignant hematological diseases

Comparative Evaluation of Two Flowcytometric Analysers as Diagnostic Tools for the Automated Detection of Malaria

SIMTI Servizi Srl. Original Article

Ordering Physician CLIENT,CLIENT. Collected REVISED REPORT

Marked pseudoreticulocytosis in a leukemic dog

Differential Blood Smear H3

Automated Hematology Analyzer. Differentiate with the XT-4000i

Hematology & Coagulation Practicum Objectives CLS - 647

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom

Morfologia normale e patologica

HEMATOLOGIC MORPHOLOGY- AECOM HEMATOLOGY COURSE

3/31/2017 OBJECTIVES CASE STUDY #1 MANUAL REVIEW. Hematology Case Studies: Every Picture Tells a Story

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Impact of Integrating Rumke Statistics to Assist with Choosing Between Automated Hematology Analyzer Differentials vs Manual Differentials

Autoanalyzer Generated Spurious Basophilia in Adolescents and Adults

EDUCATIONAL COMMENTARY MORPHOLOGIC ABNORMALITIES IN LEUKOCYTES

of s and s Disclosure

Usefulness of NEUT-X determination in routine diagnostic procedures: application to myelodysplastic syndromes. F. Cymbalista.

Brief Communication Diagnostic Hematology

Literature List Body Fluids

It s Easier Than you Think How to Implement the Advanced Clinical Parameters

3. Blood Cell Histograms:

Collected: , PM Sent: , PM Received: , PM Preliminary: , PM. Notification Status: COMPREHENSIVE DIAGNOSIS

By Dr. Mohamed Saad Daoud

Introduction to Haematology. Prof Roger Pool Department of Haematology University of Pretoria

Collect and label sample according to standard protocols. Gently invert tube 8-10 times immediately after draw. DO NOT SHAKE. Do not centrifuge.

Role of immature reticulocyte fraction in evaluation of aplastic anemia in cases of pancytopenia

Performance Evaluation of NS-hema21t Automated Hematology Analyzer and Comparison of the Hematological Parameters with Sysmex XT1800i

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How?

A HemoCue Novel Technology Product for 5-Part Differentiation of Leukocytes Point-of-Care

The Evolving Role of Reticulated Platelets

CELL-DYN 3700 Strength in Technology, Proven Reliability

SH A CASE OF PERSISTANT NEUTROPHILIA: BCR-ABL

VETERINARY HEMATOLOGY ATLAS OF COMMON DOMESTIC AND NON-DOMESTIC SPECIES COPYRIGHTED MATERIAL SECOND EDITION

Clinical use of flow cytometry

Please contact Customer Service with any questions. Ph: option 5

4. TEXTBOOK: ABUL K. ABBAS. ANDREW H. LICHTMAN. CELLULAR AND MOLECULAR IMMUNOLOGY. 5 TH EDITION. Chapter 2. pg

Myelodysplastic Syndrome Case 158

Peripheral Blood Smear: Diagnostic Clues and Algorithms

Transcription:

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes A. M. CENCI *1, M. MACONI *2, and B. CASOLARI *1 *1 Laboratory of Clinical Pathology, Ospedale S. Agostino, Modena, Italy. *2 Laboratory of Clinical Pathology, Azienda Ospedaliera Arcispedale Santa Maria Nuova, Reggio Emilia, Italy. Immature Granulocytes (IG) are normally not present in peripheral blood. Their presence is an important indicator of enhanced bone marrow activation. The present study evaluated the diagnostic performance of the Sysmex XT-2000i automated hematology analyzer in the detection of myelocytes and/or metamyelocytes in patients affected by non-haematological disease, in comparison with manual microscopy review (NCCLS H20-A) and also in comparison with Sysmex XE-2100, used as a predicate device. Sysmex XT-2000i shows good specificity and sensitivity in the detection of low, medium and high concentrations of immature myeloid cells. The diagnostic performance of the XT-2000i IG and LS flags is very high and comparable to the results obtained with Sysmex XE-2100. These results suggest that Sysmex XT-2000i would be an excellent second analyser for larger laboratories concerned with harmonization of instrumentation (in particular as a back-up for Sysmex XE-2100), or as a primary instrument for medium-sized routine or STAT laboratories. Key Words Automated Hematology Analyzer, XT-2000i, Immature Granulocytes, NCCLS H20-A (Sysmex J Int 15 : 1-6, 2005) Received 13 Jan, 2005; Accepted 20 Jan, 2005 INTRODUCTION The presence of immature granulocytes (IG) in peripheral blood is potentially important information indicating enhanced bone marrow activation. The promyelocyte, myelocyte and metamyelocyte stages of myeloid maturation may be correlated with systemic inflammatory stress or leukaemic reactions. The detection of IG is also important in distinguishing patients with haematological diseases [chronic myeloproliferative disorders (CMPD) and myelodysplastic syndromes (MDS)] from patients with infection. Early detection of bacteraemia and sepsis facilitates timely initiation of antimicrobial therapy and reduces morbidity and mortality in these situations. While there is no true gold standard for the detection of sepsis and blood culture has some problems resulting from poor collection technique 1,2) and in vitro bacterial growth, clinicians have tried to use other laboratory parameters to predict bacteraemia and sepsis. So far, quantitation of leucocytes, including neutrophils, is important in patients who might have an infection, because of the acute inflammatory response to bacterial infection 3-5). Preterm and neonatal blood samples commonly contain immature blood cells of particularly variable morphology and thus frequently need to be differentiated. During pregnancy the absolute count of segmented neutrophils approximately doubles, reflecting not a true increase in granulopoiesis, but rather increased demargination of the circulating neutrophils 6). The precise automated detection and enumeration of immature granulocytes would be a useful parameter to predict the presence of these physiological and pathological states. Sysmex XT-2000i (Sysmex Corporation, Kobe, Japan) is a new fully automated haematology analyser with a throughput of 80 samples per hour. The analyser can provide up to 30 parameters including CBC, a complete 5-part leucocyte differential, reticulocyte parameters and optical (fluorescence) platelet counts 7). These measurements are performed with the same technologies and reagents used by Sysmex XE-2100, a top of the range haematology analyser with a throughput of 150 samples per hour that, in addition, performs nucleated red blood cell (NRBC) and the haematopoietic progenitor cell (HPC) counts in two additional channels: the NRBC and the IMI (Immature Information) channel 8). Both analysers detect the presence of abnormal or immature WBCs providing a list of suspect messages: Blasts?, Immature Gran?, Left Shift?, Atypical Lympho?, Abn Lympho/ Blasts?. These suspect messages are generated from abnormal cell locations on the WBC/DIFF scattergram. Sysmex XE-2100 also combines pattern abnormalities from the IMI scattergram (not present on XT-2000i). Our study focusses on the diagnostic performance of the Immature Gran? (IG) and Left Shift? (LS) flags gen- 1

erated by Sysmex XT-2000i in the detection of myelocytes and/or metamyelocytes, with normal morphology, in patients affected by non-haematological disease (e.g. infection, inflammation, post-surgery/bleeding, pregnancy). The results, in terms of sensitivity, specificity and efficiency of the IG and LS flags, are presented in comparison with those from Sysmex XE-2100, used as a predicate device. MATERIALS AND METHODS Haematology analysers Sysmex XT-2000i performs the WBC count and the 5- part differential in two different channels by flow cytometry using a semiconductor laser. In the WBC/BASO channel a specific reagent lyses the red cells and selectively suppresses degranulation of basophils, resulting in their separation from other forms of WBCs. The combination of the side-scattered light (inner complexity of the cells) and the forward-scattered light (size) are presented in the WBC/BASO scattergram (Fig. 1) where the WBC and basophil counts are estimated. In the WBC/DIFF channel, after lysis of the red cells, a specific polymethine fluorescence dye enters the WBCs and stains the DNA and RNA therein. Lymphocytes, monocytes, eosinophils and neutrophils + basophils are then classified and counted in the DIFF scattergram (Fig. 2) according to their side-scattered light (inner complexity of the cells) and side fluorescence intensity (proportional to their content of nucleic acid). Abnormal and immature WBCs are located in specific areas of the DIFF scattergram 9) (Fig. 3), flagged by the following suspect messages: Blasts?, Immature Gran? (IG), Left Shift? (LS), Atypical Lympho? and Abn Lympho/ Blasts?. The immature granulocytes are classified as cells with higher fluorescence intensity than neutrophils, because their DNA content is greater than the latter. The IG flag of XT-2000i is mainly related to the presence of metamyelocytes and myelocytes. The LS flag is mainly related to the presence of band cells and metamyelocytes. Sysmex XE-2100 performs the leucocyte count and 5- part differential with the same technology and the same reagents as Sysmex XT-2000i 10). XE-2100 is, however, equipped with an additional specific channel (IMI) that provides information on the presence of immature granulocytes, blasts and haematopoietic progenitor cells by a combination of radio frequency and direct current (RF/DC) technologies. This channel uses a haemolytic reagent that selectively protects immature cells. Both Sysmex XT-2000i and Sysmex XE-2100 offer immature granulocyte counts (IG % and IG #) as research parameters. With a new software release XE-PRO and XE-IG Master the IG count can now be reported with Sysmex XE-2100 both in absolute terms and as a percentage of total WBC 11). In this article we do not present any comparison between the automated IG count and manual microscopy, because the main focus of this study was to assess the predictive value of the myeloid flags IG and LS of XT-2000i in detecting the presence of paraphysiological immature granulocytes in patients without haematological malignancy. Blood specimen selection and processing 822 samples were included in this study. All venous blood specimens were collected into K 2 -EDTA anticoagulant, kept at room temperature, and analysed on both instruments within 1 hour of their arrival in the laboratory. The blood specimens were divided into three different groups using the following criteria: 1 st group - 178 samples from outpatients and inpatients (e.g. cardiology, orthopaedics, surgery, internal medicine, pregnancy), taken during a preliminary evaluation of Sysmex XT-2000i at the time of its installation 2 nd group - 600 samples, from outpatients and inpatients, randomly chosen from the routine workload on six different days (100 samples each day) excluding those from onco-haematology wards. 3 rd group - 44 samples, selected from patients with normal WBC count (lower than 10.0 10 9 /L) and minimal haematological abnormalities on manual microscopy review for the presence of 2.0% or less of metamyelocytes and/or myelocytes and/or more than 5% of band cells. Manual microscopy review The clinical sensitivity and specificity of the IG and LS Fig. 1 WBC/BASO scattergram Fig. 2 WBC/DIFF scattergram Fig. 3 Abnormal cell location 2

suspect flags of Sysmex XT-2000i were assessed by comparison with microscopy WBC differential counts as described previously 12-14) and the performance of both analysers was compared. The reference differential WBC counts were performed by microscopy examination of 400 cells on May-Grünwald-Giemsa stained preparations by two qualified medical doctors each counting 200 cells on separate smears according to the NCCLS H20-A protocol 15). For the 1 st and the 2 nd group of specimens, the samples were considered positive by microscopy review if they showed more than 2% metamyelocytes and/or myelocytes and/or more than 10% band cells. For the 3 rd group of specimens, as stated previously, all samples were selected and considered positive by microscopy review if they showed from 0.5% to 2% metamyelocytes and/or myelocytes and/or more than 5% band cells. None of the positive samples selected in this study were positive by microscopy review for the presence of promyelocytes or blasts. STATISTICAL ANALYSIS To compare the results from XT-2000i, XE-2100 and the reference procedure, truth tables were constructed to determine: - True Positive [TP] (samples positive at microscopy review flagged by the instrument with the IG and / or the LS flag) - True Negative [TN] (samples negative at microscopy review and not flagged by the instrument with the IG and / or the LS flag) - False Positive [FP] (samples negative at microscopy review but flagged by the instrument with the IG and/or the LS flag, or samples positive at the microscopy review but flagged by the instrument with other flags such as Blasts? in addition to IG and/or LS flag) - False Negative [FN] (samples positive at microscopy review but negative or flagged by the instrument with flags other than IG or LS) The predictive values of IG and LS flags for the presence of immature granulocytes after microscope review were determined as follows: Sensitivity (%) = [TP / (TP + FN)] 100 Specificity (%) = [TN / (TN + FP)] 100 Efficiency (%) = [(TP + TN) / (TP + TN + FP + FN)] 100 Comparison between XT-2000i and XE-2100 for the IG and LS flags generated was determined as follows: - Number of negative samples in agreement (concordance): number of samples without IG or LS on both analysers - Number of positive samples in agreement (concordance): number of positive samples flagged with only IG and/or LS on both analysers - Number of negative samples in disagreement (discordance): number of negative samples flagged with IG and/or LS only by one analyser - Number of positive samples in disagreement (discordance): number of positive samples flagged with IG and/or LS only by one analyser. RESULTS In a preliminary study of Sysmex XT-2000i, 178 samples were compared to Sysmex XE-2100 to assess the sensitivity and specificity of the IG and LS suspect flags to the presence of more than 2% myelocytes and/or metamyelocytes on microscopy examination. Only 6 out of 178 samples (3.37%) had a WBC count higher than normal (10.0 10 9 /L to 15.0 10 9 /L). The comparison of IG and LS flags generated by the two instruments is shown in Table 1. The discrepancies that were present in the 11 discordant samples are described in Table 2. The results of sensitivity, specificity and efficiency for both analysers compared with reference microscopy review are shown in Table 3. In the 2 nd group we reviewed the occurrence of the IG and LS (alone or combined) in our normal routine workload. In 600 samples, randomly chosen from the routine workload of six different days, only 11 (1.83 %) had WBC counts higher than normal (from 10.0 10 9 /L to 15.0 10 9 /L). With both analysers we found 18 out of 600 samples flagged with IG and/or LS. By microscopy review we confirmed that 9 samples were true positives for minimal myeloid abnormalities and 7 samples were true positive for the presence of 4% or more myeloid immature cells. 2 samples were false positive on both analysers. In both these samples we found platelet clumps. The 582 samples not flagged with IG and/or LS on both analysers were negative by microscopy review. Thus, we found a full agreement between XT-2000i and XE-2100 in this group of patients. The results of the diagnostic performance of both analysers are described in Table 4. Finally, the sensitivity of the IG and LS suspect flags was studied on a 3rd group of 44 samples, selected from patients with a normal WBC count (less than 10.0 10 9 /L) and presenting minimal haematological abnormalities on manual microscopy review (0.5% to 2.0% metamyelocytes and/or myelocytes and/or more than 5% of band cells). Examples of scattergrams obtained by Sysmex XT- 2000i in two samples from this patient group are shown in Figs 4 and 5. The comparisons of IG and LS flags generated by the two analysers in this group of 44 samples are shown in Table 5. The results of the diagnostic performances of both instruments in this group of patients are shown in Table 6. Despite the number of false positive and false negative results for the IG and LS flags obtained in this group of patients, the high sensitivity of both analysers should be underlined in the detection of very low concentrations of myelocytes and metamyelocytes when all suspect flags are considered (40 of 44 for XT-2000i, 37 of 44 samples for XE-2100). Grouping all 822 samples together, the overall diagnostic performance of the IG and LS flags in the detection of immature myeloid cells is shown in Table 7. 3

Table 1 XT-2000i vs XE-2100 : comparison of IG and LS flags on the 1 st group of 178 samples Concordance in negative samples 130 (73.03 %) Concordance in positive samples 37 (20.79 %) Discordance in negative samples 2 (1.12 %) Discordance in positive samples 9 (5.06 %) Table 2 Discordance in the 1 st group of 178 samples TP XT-2000i ; FP on XE-2100 6-4 samples flagged with Blasts+IG on XE-2100-2 samples flagged with Blasts+IG+At. Lymph on XE-2100 FP XT-2000i ; TN on XE-2100 2-2 samples flagged with IG on XT-2000i TP XT-2000i ; FN on XE-2100 1-1 sample without IG / LS on XE-2100 FN XT-2000i ; TP on XE-2100 2-2 samples flagged with Blasts? on XT-2000i Table 3 Diagnostic performance of IG and LS flags on the 1 st group of 178 samples True Positive (TP) 44 39 True Negative (TN) 130 132 False Positive (FP) 2 6 False Negative (FN) 2 1 Sensitivity 95.6% 97.5% Specificity 98.5% 95.7% Efficiency 97.8% 96.1% Table 4 Diagnostic performance of IG and LS flags on the 2 nd group of 600 samples True Positive (TP) 16 16 True Negative (TN) 582 582 False Positive (FP) 2 2 False Negative (FN) 0 0 Sensitivity 100% 100% Specificity 99.6% 99.6% Efficiency 99.6% 99.6% Fig. 4 WBC/BASO and DIFF scattergrams from XT-2000i WBC = 9.8 10 9 /L. Microscopy review: 1 myelocyte; 1 metamyelocyte /100 WBC 4

Fig. 5 WBC/BASO and DIFF scattergrams from XT-2000i WBC = 7.4 10 9 /L. Microscopy review: 1 myelocyte /100 WBC Table 5 XT-2000i vs XE-2100 : comparison of IG and LS flags on the 3 rd group of 44 samples XT-2000i XE-2100 IG 14 5 IG + LS 4 2 LS 4 1 IG / LS + other flags 13 20 Negative for IG / LS 9 (5 samples flagged with Blast?, 4 samples without any flag) 16 (9 samples flagged with Blasts?, 7 samples without any flag) Table 6 Diagnostic performance of IG and LS flags on the 3 rd group of 44 samples True Positive (TP) 22 8 True Negative (TN) 0 0 False Positive (FP) 13 20 False Negative (FN) 9 16 Sensitivity 71% 33.3% Specificity N.A. N.A. Efficiency 50% 18.2% Table 7 Diagnostic performance of IG and LS flags on the 822 total samples True Positive (TP) 82 63 True Negative (TN) 712 714 False Positive (FP) 17 28 False Negative (FN) 11 17 Sensitivity 88.2% 78.8% Specificity 97.7% 96.2% Efficiency 96.6% 94.5% 5

DISCUSSION Quantitative analysis of immature granulocytes is useful for good clinical management of patients affected by haematologic diseases (chronic myelogenous leukaemia, chronic myeloproliferative disorders, myelodysplastic syndromes) or patients presenting with infection. The IG compartment includes promyelocytes, myelocytes and metamyelocytes. Several studies have assessed the clinical usefulness of quantitative and qualitative changes in IG for predicting infection in paediatric and adult populations associated, or not, with other laboratory data (total WBC count, C-reactive protein (CRP), etc). Myeloid progenitor cells were significantly higher in infected than in uninfected patients 16). Sysmex XT-2000i shows good specificity and sensitivity in the detection of low, medium and high concentrations of immature myeloid cells. The diagnostic performance of the XT-2000i IG and LS flags in the detection of immature myeloid cells is very high and comparable to the results obtained with Sysmex XE-2100, even without the immature information (IMI) channel. The use of fluorescence flow cytometry for the 5-part differential analysis may be the reason for this performance and for the very good agreement with the manual reference method. These results suggest that Sysmex XT-2000i would be an excellent second analyser for larger laboratories concerned with harmonization of instrumentation (in particular as a back-up for Sysmex XE-2100 since the same technology and reagents are used), or as a primary instrument for medium-sized routine or STAT laboratories. References 1) Wilson ML : Blood cultures: introduction. Clin Lab Med, 14:1-7, 1994. 2) Lorian V, Amaral L : Predictive value of blood cultures [published correction appears in Infect Control Hosp Epidemiol.1992; 13:693] Infect Control Hosp Epidemiol, 13: 293-294, 1992. 3) Wenz B, et al. : The clinical utility of the leukocyte differential in emergency medicine. Am J Clin Pathol, 86:298-303, 1986. 4) Shapiro MF, Greenfield S : The complete blood count and leukocyte differential count: an approach to their rational application. Ann Intern Med, 106: 65-74, 1987. 5) Cornbleet PJ : Clinical utility of the band count. Clin Lab Med, 22:101-136, 2002. 6) Hiss RG : Blood disorders in pregnancy. Laros RK, Ed. Lea & Febirger, Philadelphia, 1986. 7) Langford K, et al. : Performance evaluation of the Sysmex XT- 2000i automated haematology analyzer. Laboratory Hematology, 9:29-37, 2003. 8) Briggs C, et al. : Performance evaluation of the Sysmex XE-2100 automated haematology analyser. Sysmex J Int, Vol 9, No. 2:113-119, 1999. 9) Inoue H : Overview of Automated Hematology Analyzer XE-2100. Sysmex J Int, Vol 9, No. 1: 58-64, 1999. 10) Van der Meer W., MB Dinnissen J., H de Keijzer M. : Evaluation of the Sysmex XT-2000i, a new automated haematology analyser. Sysmex J Int, Vol 12, No. 2: 71-75, 2002. 11) Bruegel M, et al. : Reference values for Immature Granulocytes in healthy blood donors generated on the Sysmex XE-2100 automated haematology analyser. Sysmex J Int, Vol 14, No. 1: 5-7, 2004. 12) International Committee for Standardization in Haematology: protocol for type testing equipment and apparatus used for haematology analysis. J Clin Pathol, 31:275-279, 1992. 13) Weber N, Mast BJ, Houwen B : Performance evaluation of hematology analyzers: an outline for clinical laboratory testing. Sysmex J Int, Vol. 5:103-113, 1995. 14) Buttarello M, et al. : Il conteggio dei granulociti immaturi sull analizzatore Sysmex XE-2100. Rivista di Medicina di Laboratorio, Vol.5 suppl. at nr.3: 269, 2004. 15) National Committee for Clinical Laboratory Standards (NCCLS): Reference leukocyte differential count (proportional) and evaluation of instrumental methods. NCCLS Approved Standard NCCLS document H20-A, NCCLS; Villanova PA, 1992. 16) Selig C, Northdurft W : Cytokines and progenitor cells of granulocytopoiesis in peripheral blood of patients with bacterial infection. Infect Immun, 63:104-109, 1995. 6