Measurement of Scattered Radiation Dose Around Radiology Unit at Dr. Saiful Anwar Hospital, Malang

Similar documents
Radiation Protection- Cath lab

Radiation Safety for New Medical Physics Graduate Students

2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry

Radiation Safety in the Catheterization Lab

Table of Contents. Introduction 3. Background 4

Radiation Safety Bone Densitometer

Lessons learned from accidents in conventional external radiotherapy

Dosimetric Consideration in Diagnostic Radiology

The x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone

Facing the challenges: IAEA activities to protect the patients

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

TheRadiationDoseMeasurementoftheComputedTomographyScantoEnsuretheSafetyforMedicalStaffinDr.SardjitoGeneralHospitalYogyakartaIndonesia

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Now offering a Real-time dose measurement system featuring RaySafe i3. Powered by

PAGE 1 OF 5 HEALTH, SAFETY & ENVIROMENTAL MANUAL PROCEDURE: S560 Radiation Safety REV /14/2012

RaySafe i3. Visualize your radiation exposure in real time.

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

RADIATION SAFETY. Junior Radiology Course

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Managing the imaging dose during image-guided radiation therapy

Now offering a subscription based real-time dose measurement system featuring RaySafe i3. Powered by

in developing institutional policies, procedures, and /or protocols. The Canadian Society of

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

Site Planning and Design of PET/CT Facilities. Melissa C. Martin, M.S., FACR, FAAPM AAPM Annual Meeting, Orlando, FL August 2, 2006

"The Good Side of Radiation: Medical Applications"

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Why radiation protection matters?

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S

Introduction. Chapter 15 Radiation Protection. Advisory bodies. Regulatory bodies. Main Principles of Radiation Protection

Background Radiation in Radiography Departments and Who Wears Thermoluminescent Dosimeter (TLD) Badges

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

PRINCIPLES AND METHODS OF RADIATION PROTECTION

Module Rhodes

45 Hr PET Registry Review Course

Upon successful completion of the course, the student should be competent in the following tasks:

RADIATION PROTECTION SERVICE DEPARTMENT OF WELLBEING, SAFETY AND HEALTH

Practice and Risk at Medical Facilities in Agency Operations

REPORT: MEASUREMENT OF X-RAY DOSE SECURITY CHECK BERGEN AIRPORT FLESLAND (AVINOR)

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED *

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

Radiation Safety Training Module: Diagnostic Radiology Radiation Protection in Diagnostic Radiology

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

TM on the New Dose Limits for the Lens of the Eye: Implications and Implementation

Mapping the ASRT Objectives for Radiation Protection (47 Objectives) and Radiation Biology (21 Objectives) to this Text

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

IORT with mobile linacs: the Italian experience

QUANTIFICATION OF THE RISK-REFLECTING STOCHASTIC AND DETERMINISTIC RADIATION EFFECTS

The Difference Between Old and New Intervention Effect to Decrease Pain on Frozen Shoulder

Medical Use of Radioisotopes

Radiation Protection: A Review

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

NCRP PAC 4: Radiation Protection in Medicine. Overview. NCRP and PAC 4. Donald L. Miller, MD FSIR FACR Center for Devices and Radiological Health FDA

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Quiz True/False: Large amounts of radiation to insects will cause them to mutate!

Evening session 1 Stakeholder platform opportunity 18:30-20:00 IAEA International Atomic Energy Agency

Understanding Radiation Therapy. For Patients and the Public

Upper Gastrointestinal (GI) Tract X-ray (Radiography)

Douglas J. Simpkin, Ph.D. Aurora St. Luke s Medical Center Milwaukee, Wisconsin. www.

TIN1.1 3/10/06 10:31 AM Page 1. Section 1 The Basics

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1

Survey of Radiation Dose Levels in Patients in X-Ray Units of Some Selected Hospitals in Jos Metropolis

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Radiation Protection in Laboratory work. Mats Isaksson, prof. Department of radiation physics, GU

A Standardised Approach to Optimisation

Krueger-Gilbert Health Physics, Inc.

Nuclear Radiation Today

Quality and Safety of Patient Healthcare and Personalized Medicine Using Ionizing Radiation

Application of international standards to diagnostic radiology dosimetry

Results of dose control and measurement plans applied for SPEAR3 commissioning year (FY04)

15-17 October 2014, BfS Neuherberg. WHO International Consultation:

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

RPR 48B. APPLICATION TO USE DIAGNOSTIC EXTERNAL RADIATION IN HUMAN RESEARCH STUDIES

RADIATION SAFETY REFRESHER TRAINING FOR AUGUSTA UNIVERSITY USERS OF RADIOACTIVE MATERIAL

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PERSONNEL MONITORING AND DOSIMETRY POLICIES

A study of indoor radon levels in Iraqi Kurdistan Region, Influencing factors and lung cancer risks

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

Trends in Occupational Exposure in Malaysia

Procedural radiation exposure of interventional cardiologists and radiologists Kuipers, G.

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

CT Radiation Risks and Dose Reduction

When worker safety is mission critical. Thermo Scientific EPD TruDose Electronic Dosimeter

REPORT OF THE TECHNICAL MEETING: THE NEW DOSE LIMIT FOR THE LENS OF THE EYE IMPLICATIONS AND IMPLEMENTATION VIENNA, 2-4 OCTOBER 2012

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

Children's (Pediatric) Voiding Cystourethrogram

Breast Tomosynthesis. What is breast tomosynthesis?

Bureau of Laboratory Quality Standards

Krueger Gilbert Health Physics, Inc.

CONTROLLABLE DOSE: A discussion on the control of individual doses from single sources. Roger H Clarke

ARRT Specifications Radiation Exposure & Monitoring

JMSCR Vol 06 Issue 12 Page December 2018

Transcription:

The 2nd International Conference on Vocational Higher Education (ICVHE) 2017 The Importance on Advancing Vocational Education to Meet Contemporary Labor Demands Volume 2018 Conference Paper Measurement of Scattered Radiation Dose Around Radiology Unit at Dr. Saiful Anwar Hospital, Malang Novita Rosyida and Ekaratri Noor W Vocational Program, University of Brawijaya, Malang, East Java, Indonesia Corresponding Author: Novita Rosyida novitarosyida@ub.ac.id Received: 8 June 2018 Accepted: 17 July 2018 Published: 8 August 2018 Publishing services provided by Knowledge E Novita Rosyida and Ekaratri Noor W. This article is distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use and redistribution provided that the original author and source are credited. Selection and Peer-review under the responsibility of the 2nd ICVHE Conference Committee. Abstract The measurement of radiation dose in diagnostic radiology and radiotherapy is considered to be some critical factor+s in optimizing radiation protection for healthcare practitioners, patients and public. This article presents the overview of an average radiation dose measurement in public area around Radiology Unit in Dr. Saiful Anwar Hospital (RSSA), Malang. The critical radiation dose recommendation from BAPETEN is below 50 msv/year for healthcare practitioners and below 5 msv/year for public area. This radiation dose measurement was done outside the CT Scan room, radiotherapy Cobalt-60, brachytherapy and fluoroscopy room, and X-ray room, using RGD 27091 survey meter. The measurement of radiation background was held while all instruments producing radiation were in standby position. The average value of radiation background around radiology unit was 0.83 µsv/h. The lowest dose rate measurement was 0.6 µsv/h at CT scan area, and the highest dose rate measurement was 3.57 µsv/h, which was in front of the brachytherapy unit s door. The measurement results show that the scattered radiation dose around radiology unit is lower than the critical radiation dose recommendation from BAPETEN. Keywords: radiation, scattering, radiation dose 1. Introduction Medical radiation is commonly used in clinical diagnosis and treatment; however, improper use can result in serious damage to the human body [1, 2]. The use of X-ray and gamma ray has extended and become very important in both diagnostic radiology imaging and radiotherapy. Nowadays, radiology seems to be a necessary procedure in diagnosis and treatment of patients. X-ray examinations have several clinical advantages. In a lot of ways, it has preserved its priority over other diagnostic methods. In most cases, X-ray is the first choice of the diagnostic algorithm examination. In fact, X-ray examination is still the most frequently used modality [3]. How to cite this article: Novita Rosyida and Ekaratri Noor W, (2018), Measurement of Scattered Radiation Dose Around Radiology Unit at Dr. Saiful Anwar Hospital, Malang in The 2nd International Conference on Vocational Higher Education (ICVHE) 2017 The Importance on Advancing Vocational Page 669 Education to Meet Contemporary Labor Demands, KnE Social Sciences, pages 669 678. DOI 10.18502/kss.v3i11.2796

Scattered photons are radiations that change in direction as a result of interaction with some media. These scattered photons are detrimental to contrast the image, reduce accurate representations of human anatomy, and increase the patients and workers dose [4, 5]. The scattered radiation occurs when the photon scatters to the side or is reflected 90 from the initial photon beam. The back scattered radiation is when the photon scatters is in a backward direction from the initial primary beam, at either 180 or at various angles [4]. Ionizing radiation has, since its discovery, proved to be able to bring the human population tremendous benefits when used in medicine, but as ionizing radiation is associated with risks due to stochastic and deterministic effects, it is necessary to consider the protection of patients from potential harm [2]. Unnecessary exposure of patients can arise from medical procedures that are not justified for a specified objective, application of procedures to individuals that are not justified on the basis of their conditions, and medical exposures that are not appropriately optimized for the situation in which they are used. This can lead to unnecessary risks due to stochastic effects. Unintended exposure of patients can arise from unsafe design or use of medical technology, leading to deterministic effects. Radiation protection of patients is intended to protect from both unnecessary and unintended exposures [1]. The critical radiation dose recommendation from BAPETEN referring to ICRP is below 50 msv/year for healthcare practitioners and below 5 msv/year for public area [6]. 2. Methods The scattered radiation dose measurement was performed in radiology unit at dr. Saiful Anwar Hospital (RSSA). This measurement was done outside the CT Scan room, radiotherapy Cobalt-60, brachytherapy and fluoroscopy room, and X-ray room, using Rontgen Gamma Ray Dosimeter (RGD 27091). RGD 27091 is battery-operated dosimeter and the dose rate meter provided with an ionization chamber. It is used for measuring continuous roentgen and gamma rays. This device is applicable for measuring pulsed roentgen rays in the dose rate mode as well. Its high measuring sensitivity, low dependence on energy and direction of the probe allow for reliable measurements. Besides its main application as a precision radiation protection dosimeter, the device can also be used as a dosimeter for high dose rates because it has wide measuring range. The measurement for all rooms was done in the fixed point, that is usually crossed by medical staff, radiographer or the patient guardian or is usually called public area. DOI 10.18502/kss.v3i11.2796 Page 670

The dose value in each point is made from the average of five repetitions from measurement in the same point but different patient, because the measurement was done simultaneously while the source of radiation operated. Measurement of background radiation was held while the source of radiation in standby position. The point of background radiation measurement was near the source of radiation inside the room, and the value of background radiation was the average value of measurement in each room. 3. Result The scattering radiation doses at different point in radiology unit of dr. Saiful Anwar Hospital (RSSA) were measured. By using RGD 27091 dosimeter, the scattered radiation doses at the public area (outside diagnostic room) were detected. The average indoor background radiation in radiology unit RSSA was 0,83 µsv/h. The measured background radiation in Brachytherapy and fluoroscopy unit was 1 µsv/h, in radiotherapy Cobalt-60 was 0.7 µsv/h, in the CT Scan unit was 1.01 µsv/h and in the X-ray unit was 0.63 µsv/h. The results of measurement at several point are as follow: 3.1. CT scan room CT Scan room is located in the first floor of radiology unit at the back office of radiology. The measurements are located in six point, see Figure 1: Point A is located in the backside of the door linked the operator room. Point B is located in the operator room. Point C is located in the entrance door. Point D is located in the east side radiation source linked to the radio diagnostic corridor. Point E is located in the south side of radiation source. Point F is located in the west side of the room s wall. The result of measurement at CT scan area can be seen in Table 1. The highest measurement is 1.85 µsv/h at point A and the lowest is 1.13 at point E. 3.2. Result of dose rate measurement in CT scan area DOI 10.18502/kss.v3i11.2796 Page 671

Figure 1: Point of measurement at CT Scan area. T 1 No. Position Dose Rate (µsv/h) 1 A 1.85 2 B 1.48 3 C 1.36 4 D 1.38 5 E 1.13 6 F 1.21 3.3. Radiotherapy Cobalt-60 The dose rate measurement in radiotherapy unit is located in second room of radiotherapy Cobalt-60. The source of radiation comes from gamma rays. The point of measurement is as seen in Figure 2. Point A is located in the operator room. Point B is in the west side of the room s wall. Point C is located in the north side of the room s wall. Point D is located in the east side of the room s wall. DOI 10.18502/kss.v3i11.2796 Page 672

Figure 2: Point of measurement at radiotherapy Cobalt-60 area. The result of measurement from gamma rays scattering in radiotherapy unit can be seen in Table 2. 3.4. Result of dose rate measurement in radiotherapy Cobalt-60 area T 2 No Position Dose Rate (µsv/h) 1 A 1.068 2 B 1.066 3 C 1.032 4 D 1.066 DOI 10.18502/kss.v3i11.2796 Page 673

3.4.1. Brachytherapy and fluoroscopy area The source of fluoroscopy comes from X-ray radiation. At this area, the measurement was done in six points, see Figure 3. Point A is located in entrance room, point B is located in operator room, point C is located in the north side of the room s wall, point D is located in the east side of radiation source, point E is located in east side, point F located in the south side. The background radiation was measured inside the room at point E, and the result of measurement was in 1 µsv/h. Figure 3: Point of measurement at brachytherapy and fluoroscopy area. The result of dose rate measurement in brachytherapy and fluoroscopy unit can be seen in Table 3. 3.5. Result of dose rate measurement in brachytherapy and fluoroscopy area 3.5.1. X-ray area The measurement of X-ray area was done in five points, see Figure 4., point A is located in the operator room, point B is located in the west entrance, point C is located in the DOI 10.18502/kss.v3i11.2796 Page 674

T 3 No. Position Dose Rate (µsv/h) 1 A 3.57 2 B 0.64 3 C 0.75 4 D 1.27 5 E 0.97 6 F 0.75 north side of the room s wall, point D is located in the east side of the room s wall. Point E is located in the south entrance. Figure 4: Point of measurement X-ray area. The result of dose rate measurement in X-ray area can be seen in Table 4. 3.6. Result of dose rate measurement in X-Ray area DOI 10.18502/kss.v3i11.2796 Page 675

T 4 No. Position Dose Rate (µsv/h) 1 A 3.57 2 B 0.64 3 C 0.75 4 D 1.27 5 E 0.97 6 F 0.75 4. Discussion According to the limits of exposure radiation for medical staff mandated by the BAPE- TEN referring to ICRP, the effective radiation dose should not exceed 100 msv in five years and the effective dose in a single year should not exceed 50 msv and 5 msv for public area [1, 6]. The construction of radiology unit in RSSA were made from brick and cement with the thickness of 20 cm for X-ray, and 30 cm thickness for gamma ray. The entrance door and operator door were coated using 2 mm lead, and for the window were coated using 5 mm glass lead. The construction was based on radiation protection for medical staff and public area. The dose rate measurement outside CT Scan room was about 1.13 to 1.85 µsv/h. The lowest value was point E that was located in the operator desk, while the highest value was point A that was located in the door linked to the operator room. Point A was the nearest position from the source of radiation. The public area around CT Scan room was located at point C and D (Figure 1). The result from measurement around radiotherapy room was about 1.032 to 1.068 µsv/h. In this area, the dose rate was lower than other area of which the source of radiation is gamma ray. It means that the protection of radiation in this area is good for public service. The dose rate measurement around brachytherapy and fluoroscopy area was about 0.64 3.57 µsv/h. The lowest value in this area was at point B (the operator desk) and the highest value was at point A at the entrance door (Figure 3). At the X-ray area, all of the measurement shows that the value was below 1 µsv/h. At this area, the public area was point A, B and E (Figure 4). However, radiation exposure is a sensitive topic among medical personnel and public area. Fortunately, according to the present results, the value of scattered radiation dose around radiology unit in RSSA is about 0.6 3.57 µsv/h lower than ICRP and BAPETEN recommendation. The value of scattered radiation was attenuated to the background DOI 10.18502/kss.v3i11.2796 Page 676

radiation level. Additionally, since radiological technicians wear lead clothing during exposure, they need not be concerned. In the present work, the experiments were conducted in an operating room, and the scattered radiation doses from exposure in real situations were measured. The data of scattered radiation attenuation in a space when being exposed were established, which will enrich professional knowledge of radiation safety for clinical radiological technicians and provide useful data for radiological students. 5. Conclusions The scattered radiation doses around radiology unit of dr. Saiful Anwar Hospital were detected using RGD 27091 dosimeter. During the experiment, the area that was investigated is around CT scan unit, brachytherapy and fluoroscopy unit, radiotherapy unit and X-ray unit. The source of radiation is not only from X-ray but also from gamma ray. The average indoor background radiation in radiology unit RSSA is 0.83 µsv/h. The value of scattered radiation around radiology unit is 0.6 3.57 µsv/h lower than ICRP and BAPETEN recommendation. Acknowledgment The author would like to thank the staff of the Vocational Program, Brawijaya University and the Radiological staff at RSSA Malang for all support given to the study. References [1] J. L. Heron, R. Padovani, I. Smith and R. Czarwinski, Radiation Protection of Medical Staff, European Journal of radiology, vol. 76, pp. 20-23, 2010. [2] O. Holmberg, J. Malone, M. Rehani, D. McLean and C. Renate, Current issues and actions in radiation protection of patients, European Journal of Radiology, vol. 76, pp. 15-19, 2010. [3] H.-W. Chiang, Y.-L. Liu, T.-R. Chen, C.-L. Chen and H.-J. Chiang, Scattered Radiation Doses Absorbed by Technicians at Different Distances From X- Ray Exposure: Experiments on Prosthesis, Bio-Medical Materials and Enginering, vol. 26, pp. S1641- S1650, 2015. [4] M. N. W and F. K, The Dependence of Scattered Radiation Dose to Personnel on Technique Factor in Diagnostic Radiology, The British Journal of Radiology, vol. 65, DOI 10.18502/kss.v3i11.2796 Page 677

pp. 44-49, 1992. [5] A. T. Naji and M. S. Jaafar, Reduction of Backscattered Radiation in an X-Ray Room by Using Fabricated Iron Steel Grid, Advances in Material Physics and Chemistry, vol. 3, pp. 191-193, 2013. [6] BAPETEN, SK BAPETEN No. 01/Ka-BAPETEN/V-99 tentang Kesehatan Terhadap Radiasi Pengion, BAPETEN, Jakarta, 1999. DOI 10.18502/kss.v3i11.2796 Page 678