Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method

Similar documents
Biomechanical Analysis of Jaw Bone with Cyst Using CT-Image Based Finite Element Method

Stress Evaluation of Lower Limbs with Hip Osteoarthritis and Hip Arthroplasty

Effect of the femoral stem size on femur bone quality towards THR

Topology optimisation of hip prosthesis to reduce stress shielding

Biomechanical Study of the Effects of Balloon Kyphoplasty on the Adjacent Vertebrae

Finite Element Analysis of Cemented Hip Arthroplasty: Influence of Stem Tapers

Dynamic Finite Element Analysis of Impulsive Stress Waves Propagating from the Greater Trochanter of the Femur by a Sideways Fall

ME/BioE C176 Final Exam, Fall 2001

BIOMECHANICAL EFFECTS OF BALLOON KYPHOPLASTY ON TREATED AND ADJACENT NON-TREATED VERTEBRAL BODIES: PRE AND POST OPERATIVE EVALUATION

An effort is made to analyse the stresses experienced by the human femur. In order

Metha Short Hip Stem System

Research experimental of hipbone arthroplasties using tensometric models

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION

Prediction Of Pathological Fracture In The Proximal Femora With Bone Lesions Using A CT Scan-based Finite Element Method

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Journal of Biomechanical Science and Engineering

A Useful Reference Guide for the Stem Anteversion During Total Hip Arthroplasty in the Dysplastic Femur

Bone mechanics bone strength and beyond

Anabolic Therapy With Teriparatide Indications Beyond Osteoporosis

APPLICATION OF COMPOSITE FRACTURE MECHANICS TO BONE FRACTURE ANALYSIS USING ABAQUS XFEM

Finite Element Analysis of Femoral Neck Stress in Relation to Pelvic Width

Aesculap Trilliance Triple Tapered Polished Hip Stem

FINITE ELEMENT ANALYSIS : AN EFFECTIVE TOOL FOR PROSTHESES DESIGN

(c) B. Ravi, IIT Bombay 1

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

Collars on Cementless Femoral Stems Make More of a Contribution to Primary Stability When Subjected to Adverse Mechanical Environments

S. N. Khan 1, R. M. Warkhedkar 2, A. Shyam 3, B. R. Patil 4

OPTIMIZATION OF SKELETAL HIP IMPLANT CROSS-SECTIONS

5A* ODEP. Conserving bone. Preserving soft tissue. Restoring biomechanics. rating

PLR. Proximal Loading Revision Hip System

[ICESTM-2018] ISSN Impact Factor

INTERNATIONAL JOURNAL OF MECHANICAL ENGINEERING AND TECHNOLOGY (IJMET)

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

Robotic-Arm Assisted Surgery

Illustrative exercises for the lectures

Design and Analysis of Dental Surgical Fixture

Doctoral Dissertation

Modélisation par éléments finis des effets des médicaments sur le remodelage osseux : Approche mécanobiologique

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

Analysis of Stress and load Distribution on Hip and Knee Joint after Unilateral Total Hip Arthroplasty

Plain Film CT. Principal Modality (2): Case Report # [] Date accepted: 15 March 2014

Engineering Approaches to Age-related Diseases

NUMERICAL MODELING OF ACETABULAR RECONSTRUCTION CAGES MADE OF CFRP AND ESTIMATION OF MECHANICAL BEHAVIOR FOR BONE AND CAGE

Number Crunching: How and When Will Numerical Models Be Used in the Clinical

Beyond BMD: Bone Quality and Bone Strength

Comparative Study of Fixation Devices for Intertrochanteric Fractures

FEA Analysis of Neck Sparing Versus Conventional Cementless Stem

FINITE ELEMENT ANALYSIS TO QUANTIFY THE EFFECT OF SURGERY IN PROXIMAL FEMUR

R/F. Can T-smart Tomosynthesis Improve Diagnostic Accuracy on THA Component Stability? 1. Abstract

Finite Element Analysis of a Novel High Flexion Knee (Ti UHMWPE) used in Total Knee Arthroplasty

BIOLOGY and BIOMECHANICS OF NORMAL & OSTEOPOROTIC BONE

BIOMECHANICS OF THE SPINE AND IMPLANT FIXATION

Cementless femoral stem type SF

The role of the calcar femorale in stress distribution in the proximal femuros4_

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

HIP JOINT CONTACT PRESSURE DISTRIBUTION PRIOR AND AFTER TRIPLE OSTEOTOMY

Assessing Bone Quality in Terms of Bone Mineral Density, Buckling Ratio and Critical Fracture Load

A further enhanced classic. Wagner SL Revision Hip Stem

Optimum implant geometry

CONTRACTING ORGANIZATION: Spaulding Rehabilitation Hospital, Boston, MA 02129

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases

Introduction to Biomedical Engineering

Ethan M. Braunstein, M.D. 1, Steven A. Goldstein, Ph.D. 2, Janet Ku, M.S. 2, Patrick Smith, M.D. 2, and Larry S. Matthews, M.D. 2

Analysis of an Early Intervention Tibial Component for Medial Osteoarthritis

Abstract. 1 Introduction

Sectra Orthopaedic Solutions

CLINICAL PAPER / ORTHOPEDIC

Age changes in the human femoral neck and progression to failure. Functional Adaptation in the Femoral Neck

Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant

The Mechanical Behavior of Bone Cement in THR in the Presence of Cavities

Surgical Technique. Hip System

Biomechanical Investigations on the Total Knee Arthroplasty (TKA)

Beyond BMD: Bone Quality and Bone Strength

Bare Bones. Use of Systemic Osteoporosis Drugs in Select High Risk Patients Undergoing Spine Surgery

Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur

Effect of Bone Composition and Apparent Density on Inhomogeneity in Energy Dissipation during Tension

Klinische und radiologische Ergebnisse einer «isoelastischen» Pfanne

HELIOS h i p s y s t e m

ISSN: [NCIME: October 2016] Impact Factor: 4.116

Segmental Fracture Of A Cemented Femoral Stem - A Case Report And Review Of Litrature

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis

Predicting the Position of the Femoral Head Center

Application of Rapid Prototyping for Development of Custom Made Orthopedics Prostheses: dddddd An Investigative Study ddddddddd dddddddddd

Effect of different load conditions on a DHS implanted human femur

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

CT Imaging of skeleton in small animals. Massimo Marenzana

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code

Vasu Pai D orth, MS, National Board [orth],mch, FRACS, FICMR Total Hip Arthroplasty

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013

3A ODEP. Preserving the natural anatomy and restoring patient biomechanics

FINITE ELEMENT ANALYSIS OF KNEE IMPLANT AND PROTOTYPE FABRICATION USING RPT PROCESS

Optimum implant geometry

Clinical Application of Computed Radiography in Orthopedic Surgery

MECHANICAL PROPERTIES OF A SINGLE CANCELLOUS BONE TRABECULAE TAKEN FROM BOVINE FEMUR

Biomechanics of Solids

Optimizing function Maximizing survivorship Accelerating recovery

Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring

Transcription:

Research Article Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method Mitsugu Todo * Research Institute for Applied Mechanics, Kyushu University, Kasuga, Japan * Corresponding Author: Dr. Mitsugu Todo, Research Institute for Applied Mechanics, Kyushu University, 6-1 Kasuga-koen, Kasuga 816-8580, Japan, Tel/Fax: +81-92-583-7762; E-mail: todo@riam.kyushu-u.ac.jp Received: 24 May 2018; Accepted: 05 June 2018; Published: 11 June 2018 Abstract In the present study, medical CT images were utilized to develop 3D-computational models of hip joints and femoral bones. The finite element analysis was then performed to characterize the biomechanical problems associated with the bone models. The FEA results clearly exhibited dramatic changes of stress and SED profiles due to osteoarthritis and hip arthroplasty. Furthermore, a damage modelling was also introduced into the FEA to characterize the fracture problems of femoral bone with prostheses under a downstairs step and sideways fall conditions. Micro-damage formations are well predicted in the vicinity of the prostheses for bone cases. It is thus confirmed that CT-FEA can be very useful to understand the biomechanical problems in the field of orthopaedics. Keywords: Computed tomography; Finite element method; Bone mineral density; Osteoarthritis; Bone fracture 1. Introduction Computed tomography (CT) scan makes use of X-ray images to produce cross-sectional images of specific areas of human body. These sliced images can be utilized to construct the 3D structure of the area by layering each images in the direction perpendicular to the slicing. For example, 3D structures of skeleton can be developed by selecting bone parts from the CT images. The intensity of each image is directly related to the bone mineral density (BMD) distribution and a bright region usually corresponds to a higher BMD. It has been established that a simple linear relationship can be assumed between the CT values (Hounsfield unit) and BMD; therefore, it is possible to easily estimate the BMD distribution on each image. Furthermore, it is also possible to construct a 3D computational model of the bone with BMD distribution for finite element analysis (FEA) [1, 2]. Such CT-image based EFA (CT-FEA) has widely been utilized to characterize biomechanical problems associated with orthopaedic [2-6] and dental sciences [7, 8], in which diseases related to bones are mainly treated. Bones mainly consist of organic and inorganic mineral components such as collagen and carbonate apatite. It has been 34

believed that the apatite determines the stiffness and strength of the bone structures, and therefore, the bone mineral density (BMD) can be a controlling parameter in the biomechanical analysis of skeletal structures. Interactive problems between bone structures and metal implants have also been investigated by CT-FEA since such problems are strongly related to mismatching of elastic modulus between the material and bone tissue and/or stiffness between the implant and bone structure. In this paper, two different biomechanical problems associated with hip arthroplasty are presented. One is biomechanical analysis of stress shielding, which is one of the most important mechanical problems associated with orthopaedic metal implants (Figure 1a). In this problem, bone tissue surrounding the metal implant such as a stem of hip prosthesis is gradually absorbed due to the mismatching of stiffness between the bone and stem. The reason is thought to be the reduction of mechanical stimulus to the bone due to the small deformation of stem with high stiffness. Another problem is fracture of bone associated with hip prostheses as shown in Figure 1b. CT-FEA was applied to analyse those two kinds of problems. (a) Stress shielding (b) Bone fracture Figure 1a-1b: Biomechanical problems associated with hip joint. 2. Finite Element Analysis 2.1 Three dimensional modelling of hip joint An example of 3D hip joint model constructed from CT images of 52 years old patient is shown in Figure 2. The CT images were provided by Kyushu University Hospital. This female patient has been suffered by osteoarthritis of right hip joint. The computational modelling was conducted by using Mechanical Finder software (RCCM, Inc.). Distribution of bone mineral densities (BMD) were estimated from CT values of each of the images by assuming a linear relation between the CT values and BMD. Then, the elastic modulus of a finite element was evaluated from the average value of BMD of the element on the basis of Keyak s model [1]. The elastic modulus distributions of the right and left hip joints are shown in Figure 3. It is worth noting that inhomogeneous realistic distributions of modulus were clearly reproduced in the 3D finite element models. A virtual model of total hip arthroplasty (THA) can be developed by using the current hip joint model and CAD images of hip prosthesis. Osteotomy was applied to the hip joint model so that the hip prosthesis can properly be 35

inserted and placed into the femur and pelvis. The schematic procedure of the 3D THA modelling is shown in Figure 4a. A simple boundary condition corresponding to a stance situation was introduced into the THA model as shown in Figure 4b. Distributed load of 1800 N was applied to the top of the pelvis. Figure 2: 3D hip joint model constructed from CT images. Figure 3: Distribution of elastic modulus. (a) THA modelling procedure (b) Boundary condition Figure 4a-4b: Modelling of total hip arthroplasty. 36

In order to analyse bone fracture associated with hip arthroplasties, a whole femoral model was also developed from CT images of a 54 years old male. Resurface hip arthroplasty (RHA) and THA were then treated to the femoral model as shown in Figure 5. As the boundary conditions, two different situations such as downstairs step and sideway fall were assumed and simplified conditions were applied to the model as shown in Figure 6. A damage model was also introduced into the FEA code in order to predict bone fracture by damage accumulation. It was assumed that each of the finite elements could be damaged under both tensile and compressive situations. Under tensile condition, the bone tissue was assumed to be a linear elastic and brittle so that the maximum principal stress controlled the tensile fracture. The tensile fracture took place when the maximum principal stress reached its critical value. On the other hand, the bone tissue was assumed to be an elastic-plastic body under compression, and therefore, the equivalent stress controlled the yielding behaviour, while the compressive fracture occurred when the minimum principal strain reached its critical value. The elastic modulus of the tensile and compressive fractured elements was set to be the minimum value of the elastic moduli in the model. Figure 5: RHA and THA model for bone fracture analysis. Figure 6: Boundary conditions for bone fracture analysis. 37

3. Results and Discussion 3.1 Hip joint analysis Distributions of the strain energy density (SED) on the cross-sectional areas of the normal, abnormal (OA) and THA femur are shown in Figure 7. It is seen that SED is severely concentrated in the neck of the OA femur due to severe deformation of the femoral head and neck. On the contrary, SED is extensively reduced in the THA femur as a result of stress shielding. The kind of SED reduction corresponds to the reduction of mechanical stimulus to the bone tissue, resulting in activating bone resorption behaviour and reduction of BMD. Such SED reduction is clearly shown in Figure 8a where SED is plotted along the line from the lesser trochanter to the lower edge of the femoral models. Drastic reduction of SED is observed in the THA femur, while severe SED concentration appears in the abnormal model. Such SED concentration may accelerate the OA and deformation of the femoral head and neck. Schematic illustration is presented in Figure 9. Old bone is resorved by osteoclasts and cavities are created. Then, osteoblasts try to create new bone tissue consisting of type I collagen and carbonate apatite crystals. Reduction of mechanical stimulus tends to break such balanced remodelling cycle. In the field of biomaterials, it has been considered that one of the most effective ways to avoid the stress shielding is to reduce the elastic modulus of the stem material, Ti alloy. Currently, the modulus of Ti alloy has successfully been reduced up to around 50 GPa without sacrificing other mechanical properties. The effect of such modulus reduction on the SED distribution in the femur is shown in Figure 8b. SED effectively increased due to the reduction of modulus, suggesting improvement of mechanical stimulus. Figure 7: SED distribution on the cross-sections of femoral models. 38

Strain energy density (kj/m 3 ) J Surg Res 2018; 1 (2): 034-041 80 70 60 50 40 30 20 10 Abnormal Normal THA 0 A B (a) Comparison of three models (b) Effect of modulus reduction Figure 8a-8b: SED profiles on the line downward from the lesser trochanter (a) and from the greater trochanter (b). Figure 9: Schematic of bone remodelling behaviour. 3.2 Fracture analysis Distribution of failure elements in the RHA and THA models under two different loading conditions are shown in Figure 10. Under the downstairs, the RHA model shows damage formation along the interface between the head and bone. Elemental damages are also observed in the vicinity of the stem. This damage accumulation may be related to loosening of the implant or fracture of femoral neck. In the THA model, damages are generated in the top region of bone close to the stem, suggesting that loosening of stem may occur due to such damage accumulation. Under the sideway falls, most of damages are generated in the great trochanter at which the load is applied directly. Furthermore, extensive damages are also observed in the femoral neck of the RHA model, suggesting that neck fracture can take place under the sideway falls condition. The fracture loads of the RHA and THA models under the two conditions are compared with that of the intact bone model in Figure 11. Obviously, under the downstairs step, the fracture loads of the HA models are higher than that of the intact, indicating that the HAs tend to ease the risk of bone fracture. On the contrary, under the sideway falls, the fracture loads of the HA models become lower than that of the intact, suggesting that the fracture risk increases due to the HAs. 39

(a) Downstairs step (b) Sideway falls Figure 10: Distribution of failure elements under two conditions. Figure 11: Comparison of fracture load. 4. Conclusions CT-image based finite element analysis was applied to analyse the biomechanical problems associated with hip joint and hip arthroplasty. The results obtained are summarized as follows: Stress shielding effect in the vicinity of metal stem due to total hip arthroplasty can be reproduced as reduction of the strain energy density distribution. Stress shielding may be able to overcome due to increasing mechanical stimulus by reducing the elastic modulus of the metal stem. 40

Fracture problems of bone and loosening of implants may be able to analyse CT-FEA with the damage model. CT-FEA is also useful to assess the fracture risk of bone. References 1. Keyak JH, Rossi SA, et al. Prediction of femoral fracture load using automated finite element modelling. Journal of Biomechanics 31 (1998): 125-133. 2. Bessho M, Ohnishi I, et al. Prediction of proximal femur strength using a CT-based nonlinear finite element method: differences in predicted fracture load and site with changing load and boundary conditions. Bone 45 (2009): 226-231. 3. Mazlan MH, Todo M, et al. Effect of cage insertion Orientation on stress profiles and subsidence phenomenon in posterior lumbar inerbody fusion. Journal of Medical and Bioengineering 5 (2016): 93-97. 4. Mazlan MH, Todo M, et al. Biomechanical Effects of Balloon Kyphoplasty on Treated and adjacent nontreated vertebral bodies: Pre and Post-operative Evaluation. ARPN Journal of Engineering and Applied Sciences 10 (2015): 6197-6202. 5. Abdullah AH, Todo M. Effects of hip arthroplasties on bone adaptation in lower limbs: A computational study. Journal of Bioscience and Medicines 3 (2015): 1-7. 6. Abdullah AH, Todo M. Stress evaluation of lower limbs with hip osteoarthritis and hip arthroplasty. Journal of Medical and Bioengineering 4 (2015): 100-104. 7. Arahira T, Todo M, et al. Assessment of Mechanical Stability and Safety for Fully Edentulous Maxilla with Dental Implants. Journal of Solid Mechanics and Materials Engineering 5 (2010): 526-538. 8. Albogha MH, Kitahara T, Todo M, et al. Maximum principal strain as a criterion for prediction of orthodontic mini-implants failure in subject-specific finite element models. Angle Orthodontist 86 (2016): 24-31. Citation: Mitsugu Todo. Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method. Journal of Surgery and Research 1 (2018): 34-41. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC-BY) license 4.0 41