Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using CLEAN UP CCX2 and LC-MS/MS

Similar documents
Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using 500 MG CLEAN UP CCX2 and LC-MS/MS

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS

Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS

Identification and Quantification of Psychedelic Phenethylamine 2C Series using SPE and LC-MS/MS

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

Analysis of Limonin in Citrus Juice Using QuEChERS and LC-MS/MS

UCT STYRE SCREEN INNOVATION THROUGH CHEMISTRY

Determination of 78 Banned or Controlled Racing Industry Drugs in Horse Urine Using SPE and LC-MS/MS

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Determination of Chlorophenoxyacetic Acid and Other Acidic Herbicides Using a QuEChERS Sample Preparation Approach and LC-MS/MS Analysis

Application Note. Abstract. Authors. Pharmaceutical

Determination of Pharmaceutical Residues in Bovine Milk via LC MS/MS Following Solid Phase Extraction

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Rapid and Robust Detection of THC and Its Metabolites in Blood

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Abstract. Introduction

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Performance of an ultra low elution volume 96-well plate

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

UCT SELECTRA C18 HPLC COLUMN INNOVATION THROUGH CHEMISTRY

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

Systematic Evaluation of Solid Phase Extraction (SPE) Chemistries for the Determination of Acidic, Neutral, and Basic Drugs

Surviving Matrix Effects Experiments. Grace van der Gugten St. Paul s Hospital, Vancouver, BC, Canada

The Raptor HILIC-Si Column

Strata -X Polymeric SPE

Pesticide Residue Analysis in Whole Milk by QuEChERS and LC-MS/MS

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

Application Note. Author. Abstract. Introduction. Food Safety

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Removal of Purple Pigmentation from Cannabis using QuEChERS Extraction, ChloroFiltr dspe Clean-up and LC-MS/MS

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Fig. 1: Chemical structure of arachidonic acid COOH CH 3

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Institute of Toxicology Clinical Toxicology and Pharmacology Berliner Betrieb für Zentrale Gesundheitliche Aufgaben, Berlin

Dienes Derivatization MaxSpec Kit

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

LC/MS/MS Analysis of Metabolites of Synthetic Cannabinoids JWH-018 and JWH-073 in Urine

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS

E XCEL LENCE JUST GOT BETTER UCT FORENSICS CLEAN SCREEN XCEL SPE COLUMNS

Analysis of Pesticide Residues, Mycotoxins and Potency in Cannabis using QuEChERS Extraction, ChloroFiltr dspe Cleanup and LC-MS/MS

Authors. Abstract. Introduction. Environmental

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples

EASIMIP TM PATULIN Product Code: P250 / P250B

SPE Cartridges. All Finneran Products Are Certified. Conditioning Sample application Wash Elution

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation

MACHEREY-NAGEL. Chromatography application note. Determination of pyrrolizidine alkaloids. Abstract. Introduction

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Phenomenex roq QuEChERS Kits. Overview. Spinach Facts

What You Can t See Can Hurt You. How MS/MS Specificity Can Bite Your Backside

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

Enrichment of Phospholipids from Biological Matrices with Zirconium Oxide-Modified Silica Sorbents

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

UCT SELECTRA AQUEOUS C18 HPLC COLUMNS

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis

Finally, a Solid Phase Extraction product for Drugs of Abuse

Determination of Ractopamine and Clenbuterol in Beef Samples

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Supplementary Information. Effects of Perfluorooctanoic Acid on Metabolic Profiles in Brain and Liver of Mouse by a

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results

LCMS Vitamin D3 Method development

Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

GlycanPac AXH-1 Columns

Catecholamines and neurotransmitter. Comprehensive tools for routine and research

The Ephedra Story: An Automated SPE Procedure and Analysis for the Determination of Ephedra and Ephedra-like Compounds in Dietary Supplements

Sample Concentration and Analysis of Human Hormones in Drinking Water

Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system

APPLICATIONS TN A Comparison of Various Kinetex C18 Phases USP: L1

HVA, VMA and 5-HIAA in Urine

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Transcription:

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using CLEAN UP CCX2 and LC-MS/MS UCT Part Numbers CUCCX256 Clean Up CCX2 (C8 + Carboxylic Acid) 500mg / 6mL SPE Cartridge SLPFPP100ID21-3UM Selectra PFPP HPLC column 100 x 2.1 mm, 3 μm SLPFPPGDC20-3UM Selectra DA guard column 10 x 2.0 mm, 3μm SLGRDHLDR Guard Column Holder Summary: Catecholamines are hormones produced by the cells in the interior region of the adrenal glands. These hormones function in the control of heart rate, metabolism and blood pressure (1). The primary catecholamines are dopamine, norepinephrine, and epinephrine. Metanephrine and normetanephrine are the 3- methoxy metabolites of epinephrine and norepinephrine, respectively. Elevated catecholamine levels are commonly associated with stress; however, in patients with pheochromocytoma or paragangliomas, these hormones are released into the body without being triggered by any outside factors (2). Pheochromocytoma is a condition in which a neuro-endocrine tumor develops in the medulla of the adrenal glands. It originates from the chromaffin cells and results in uncontrolled secretion of epinephrine and other catecholamines. The classic symptoms of pheochromocytomas are those attributable to excess adrenaline production. Patients often experience episodes of sweating, headache, hypertension and heart palpitations. Pheochromocytomas are rare and usually benign; however, in about 10% of cases, these tumors can be malignant (2). When a tumor of chromaffin cells occurs outside of the adrenal glands it is known as a paraganglioma. These tumors produce the same effects as pheochromocytomas and can form anywhere in the body. Patients who are thought to have a pheochromocytoma are often diagnosed following biochemical tests that measure the levels of catecholamines and metanephrines. Metanephrines are stable metabolites of catecholamines that are cosecreted directly with catecholamines from these tumors. While catecholamines only last a short time in the body before they are metabolized, metanephrines last much longer making them a more informative marker to indicate an abnormality (3). Catecholamine and metanephrine levels may be measured in either plasma or in urine that has been collected over a 24-hour period. Generally, plasma tests are preferred due to the enhanced sensitivity of this assay, but there is elevated risk of reporting false positives utilizing this matrix. The 24-hour urine test is considered to be much more conclusive. The criterion used to deem a sample positive for pheochromocytoma is metanephrine values twice the average upper limit (1). By utilizing UCT s CLEAN UP CCX extraction columns along with a Selectra

By utilizing UCT s CLEAN UP CCX extraction columns along with a Selectra PFPP HPLC column, excellent sample clean up and analyte separation was achieved. This was demonstrated by the observation of good recovery for all five of the hormones analyzed. Dopamine Epinephrine Norepinephrine pka = 9.27 pka = 8.9 pka = 8.85 Metanephrine Normetanephrine pka =9.25 pka = 9.06 Figure 1: Catecholamines studied for the purpose of this application Sample Preparation: To 1 ml of urine add internal standard(s) and 100 µl of 6M HCl. Incubate for 20 minutes at 90ºC for 20 minutes. Neutralize ph with NaOH (~75 µl) and 3mL of Acetate buffer (ph ~7.0). Mix/Vortex SPE Procedure: 1. Condition Column a) 1 x 3 ml MeOH b) 1 x 3 ml D.I. H 2O c) 1 X 3 ml Acetate Buffer (ph 7) 2. Apply Sample a) Load at 1-2 ml/minute 3. Wash cartridge a) 1 3 ml D.I. H 2O. b) 1 3 ml 50:50 MeOH:ACN c) Dry cartridges for ~10 minutes under a high vacuum. 4. Elution a) Elute with 1 3 ml MEOH containing 5% formic acid b) Evaporate the sample to dryness under a gentle stream of nitrogen at <40 C. A 1% HCl in MEOH solution should be used to prevent volatization by the formation of the hydrochloric salt of the drugs. c) Reconstitute in 100 µl of mobile phase

LC-MS/MS Parameters: HPLC system MS system Instrumentation Agilent 1200 Binary Pump SL API 4000 QTRAP (MS/MS) HPLC column UCT Selectra PFPP, 100 2.1 mm, 3 µm Guard column UCT Selectra PFPP, 10 2.1 mm, 3 µm Column temperature 50 C Flow rate Injection volume 10 µl 300 µl/min Mobile Phase Gradient % Mobile Phase A % Mobile Phase B Time (min) (0.1% Formic Acid in Water) (0.1% Formic Acid in MEOH) 0.0 99 1 2.0 99 1 3.0 0 100 5.0 0 100 5.2 99 1 10.0 99 1 Chromatogram -- Catecholamine Standard 100ng/mL MRM Transitions Analyte Retention Time (min) Q1 Q3 1. Norepinephrine 1.19 170.4 152.0 2. Epinephrine 1.45 184.0 166.0 3. Normetanephrine 1.85 166.9 135.0 4. Dopamine 2.12 153.9 136.9 5. Metanephrine 3.12 180.0 165.0

Results: ANALYTE Recovery (n=3) 75 ng/ml 500 ng/ml 2000 ng/ml Epinephrine 86% 88% 84% Norepinephrine 101% 91% 79% Metanephrine 79% 78% 63% Normetanephrine 49% 42% 34% Dopamine 97% 91% 86% Discussion: The structures and pka values of epinephrine, norepinephrine, metanephrine, normetanephrine and dopamine make them ideal candidates for clean-up via cation exchange solid phase extraction (SPE). Two chemically different SPE columns with various combinations of wash and elution solvents were evaluated for the optimization of this procedure. UCT s benzenesulfonic acid strong cation exchange column and corresponding methodology allowed for excellent recovery of the metanephrines, however, it greatly reduced the recovery of the catecholamines regardless of the wash/elution solvent combination. Previous literature suggests that the use of highly basic elution solvents containing ammonium hydroxide lead to degradation of the catecholamines, which was noted in the extraction results. In an attempt to improve recovery, a 2% trimethylamine / 98% methanol solution was substituted for elution purposes and no improvement was noted. Subsequently, UCT s CLEAN UP CCX was selected. This is a carboxylic acid, weak cation exchange, sorbent. It allowed for adequate recovery of the metanephrines without compromising the stability of the catecholamines. Samples were loaded at ph 7, ensuring that the functional groups of both the analytes of interest and the sorbent were fully charged. The charged state allowed for strong ionic interactions between the sorbent and the catecholamines. The ionic interaction permitted the use of an extensive range of wash solutions. Ultimately, it was determined that a D.I. H 2 O wash followed by an acetonitrile: methanol (50:50) wash best removed unwanted matrix and other endogenous interferences. For elution purposes, an acidic solution (5% formic acid in methanol) sufficiently lowered the ph of the sorbent. This neutralized the carboxylic acid groups on the sorbent and disrupted the ionic bonds with the analytes of interest. The analytes were now able to be efficiently removed from the SPE column.

Conclusions: 1. Both catecholamine and metanephrine levels need to be monitored. This is done for both the diagnosis of a patient with pheochromocytoma, and also to monitor treatment for patients who are already known to have the condition. 2. By utilizing UCT s CLEAN UP CCX extraction columns and corresponding methodology, total urine catecholamine and metanephrine levels can be monitored simultaneously, reducing both analyst time and instrument time. 3. UCT s PFPP analytical column allowed for the simultaneous separation of the extremely polar catecholamines and metanephrines. Of the various HPLC phases tried (pentafluorophenylpropyl, polyaromatic, aqueous C18, and Diol) and conditions attempted, the Selectra PFPP (pentafluorophenylpropyl) column was the only one that produced acceptable baseline separation and peak shape for the quantitation of all analytes. 4. It is strongly recommended to use matrix-matched calibration curves, which include isotopically labeled internal standards to compensate for any remaining matrix that is not removed via the extraction procedure. References: 1. "Catecholamines in Urine." WebMD. WebMD, 20 June 2012. Web. 15 June 2015. 2. "Pheochromocytoma: MedlinePlus Medical Encyclopedia." U.S National Library of Medicine. U.S. National Library of Medicine, 15 June 2015. Web. 16 June 2015. 3. "Pheochromocytoma." - Mayo Clinic. N.p., 02 May 2014. Web. 22 June 2015. 8108-03-01 UCT, LLC 2731 Bartram Road Bristol, PA 19007 800.385.3153 215.781.9255 www.unitedchem.com Email: methods@unitedchem.com UCT, LLC 2018 All rights reserved