Department of Laboratories St. Louis, MO 63110

Similar documents
Procedure. Applies To: UNM Hospitals Responsible Department: Rapid Response Laboratory Revised: 3/2015

Controls (positive and negative) for quality assessment-located in central laboratory Note: See central lab UA procedure for QC. Read strips visually.

Clinical Laboratory Science: Urinalysis

Urine Reagent Strip (Blood - Ketone - Glucose - Protein - ph - Bilirubin)

URINE REAGENT STRIPS FOR URINALYSIS

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

Sensatest 1 PARAMETER: BLOOD, GLUCOSE AND PROTEIN URINE REAGENT STRIPS (1 PARAMETER) Test for Urine Occult Blood

URINE DIPSTICK AND SULPHOSALICYLIC ACID TEST. Špela Borštnar UREX 2015, Ljubljana, Slovenia

LABORATORY 6: The Complete Urinalysis with Automation

EXERCISE 11: MACROSCOPIC AND CHEMICAL EXAMINATION OF URINE

Urinalysis and Body Fluids CRg. Unit Outline. Chemical Exam of Urine. Unit 3. Reagent strip manufactures

Management of Central Venous Access Devices. Blood Glucose Monitoring

BCH472 [Practical] 1

Fraser Health Authority Accu-chek Inform II: Training Script

LABORATORY 2: Macroscopic Urinalysis

Standard Operating Procedure

Clinical Test Report. of DUS10 (Urine Reagent Strips) Effective Date: April DFI Lab. Dong-Ai Hospital Medical Center: Clinical Pathology

LABORATORY 5: The Complete Urinalysis

CHEMSTRIP 7 - URINE TEST STRIP

Year 1 MBChB Clinical Skills Session Urinalysis

SOLUTION TEST STRIPS

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

Accu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014

Type Description Date Version Performed By Notes

CJ Shuster A&P2 Lab Addenum Urinanalysis 1. Urinanalysis

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient.

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Detection and Estimation of Some Abnormal Constituents. Amal Alamri

Tips for Training New Customers

Precision ceedpro. Online Training for. Glucometer & Capillary Collection

PHYSICAL PROPERTIES AND DETECTION OF NORMAL CONSTITUENTS OF URINE

It s not just water! What is Urinalysis?

Taking a dip into urinalysis

CHEMSTRIP 10 Urine Test Strips * READ POLICY PRIOR TO STARTING TUTORIAL

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised:

Precision Xtra Training

Evaluation of VACUETTE Urine CCM Tubes for Urine Strip Testing

References Required document for Laboratory Accreditation by the College of American Pathologists.

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory

CLINITEK Novus Automated Urine Chemistry Analyzer

Urine Dipstick Medical Mobile Unit: Patient Test Procedure

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Charging the AccuChek Inform II on the Base Unit. Highlights of New AccuChek Inform II Whole Blood Glucose System

Supporting information on optical spectrum of LEDs

JOHN H. STROGER HOSPITAL OF COOK COUNTY ANNUAL TRAINING HEMOCCULT

LeadCare BLOOD LEAD ANALYZER. Quick Reference Guide

A. History Urinalysis is the oldest lab test still being performed today

Protn-Latex. For Determination of Protein in Latex. (Cat. # Latex, Latex) think proteins! think G-Biosciences

Standard Operating Procedure

Assisting in the Analysis of Urine. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved.

Influenza A IgG ELISA

Routine urine examination

TITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter

Lifescan Glucose Meter

URINANLYSIS. Pre-Lab Guide

Urinalysis (Macroscopic( Chemical Tests) ) Background. Ishihara Color Blindness Tests 12/22/2012. Mohammad Reza Bakhtiari DCLS, PhD

CentriVet GK Blood Glucose & Ketone Monitoring System

QuickTox Drug Screen Dipcard (with and without Adulteration Tests)

HBeAg and HBeAg Ab ELISA Kit

QuickTox Drug Screen Dipcard (with and without Adulteration Tests)

How to interpret your urine sample results

The Urinary System. Lab Exercise 38. Objectives. Introduction

ANTIBODY SCREENING by Uni-Gold Recombigen HIV

Physical Characteristics of

Dear Prodigy System Owner:

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.

SARASOTA MEMORIAL HOSPITAL

Summary of Significant Changes at this Revision

CoaguChek S System. Quick Reference Guide. This is a CLIA-waived System. Coagulation Testing with Fresh Whole Blood

How to interpret your urine sample results

Urisys 1100 Urine Analyzer. Policies and Procedures. for use with Urisys 1100 Urine Analyzer. This is a CLIA-Waived System

Helicobacter pylori Antigen Test

Application Note. Light Microscopic Analysis of Urine ZEISS Primo Star and ZEISS Axio Lab.A1

PROFESSIONAL MEDICAL PRODUCTS

2. Before Testing Monitor Checker Test...10 Inserting Lancets into Lancing Device...11 Quality Control Testing...13

Written By: Joann O Connor Date: 01/27/2017 Effective Date: 02/22/2017 1

Title: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.

Influenza A & B Screening

DETERMINE SPECIMEN REQUIREMENTS HIV-1/2 Ag/Ab Combo TESTING PROCEDURE

Instructions for Use. Welcome!

Instructions for Performing In-Office Lab Tests

Blood Glucose & Ketone Monitoring System

Roche CoaguChek XS Plus INR

ACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man Coulter HIV-1 p24 ELISA May 21, 2004

HARBOR-UCLA MEDICAL CENTER Torrance, California DEPARTMENT OF EMERGENCY MEDICINE POLICY AND PROCEDURE MANUAL

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

TABLE OF CONTENTS GENERAL INFORMATION... 1

To provide you with necessary knowledge and skills to accurately perform 3 HIV rapid tests and to determine HIV status.

Point of Care Testing. Bedside Glucose Meter

Human HIV (1+2) antigen&antibody ELISA Kit

Summary of Significant Changes at this Revision. Items Required

For In Vitro Diagnostic Use

Cotinine (Mouse/Rat) ELISA Kit

Items in the package:

HIV-1 p24 ELISA Kit. Catalog Number KA assays Version: 06. Intended for research use only.

HSV-1 IgM ELISA. Catalog No (96 Tests) For Research Use Only. Not for use in Diagnostic Procedures.

Transcription:

Page 1 of 8 PROCEDURE: URINALYSIS ON THE CLINITEK STATUS+ ISSUE DATE: August 2010 REVISION DATE: REVIEWED DATE: May 2017 PRINCIPLE: The Siemens Clinitek Status Analyzer is a portable instrument powered by batteries or by an electrical outlet for bench top use. It is for in vitro diagnostic use in the semi-quantitative or qualitative detection of protein, blood, leukocytes, nitrite, glucose, ketone, ph, specific gravity, bilirubin, and urobilinogen in urine samples. The system utilizes the reflection from 6 LED s to quantify urine constituents. The light is converted into electrical impulses, which are processed by the instrument s microprocessor and converted into clinically meaningful results. Siemens Multistix reagent strips are intended for use to assist diagnosis in kidney function, urinary tract infections, carbohydrate metabolism (e.g. diabetes mellitus), and liver function. The strips also measure physical characteristics, including acid-base balance and urine specific gravity. Test results can be used along with other diagnostic information to rule out certain disease states and to determine if microscopic evaluation is needed. Multistix reagent strips are ready to use upon removal from the bottle and the entire reagent strip is disposable. The reagent strips are for in vitro diagnostic use. They have been determined to be nonhazardous under the guidelines issued by OSHA in 29 CFR 1910.1200(d). EQUIPMENT/SUPPLIES 1. Seimens 10SG MultiStix 2. Clinitek Status + 3. Quantrimetrix Urine Quality Control Dropper Plus Level 1 and 2. Chemical Principles of the Reagent Strips Test Name Glucose Bilirubin Protein ph Blood Ketone Chemical Principle Glucose oxidase catalyzes the formation of gluconic acid and hydrogen peroxide from the oxidation of glucose. Peroxidase catalyzes the reaction of hydrogen peroxide That couples 4-aminoantipyrine and 4-methylcatechol to create an orange to dark red color reaction. Bilirubin couples with diazotized dichloroaniline in a strongly acid medium. Colors range through various shades of tan. At a constant ph, the development of any green color is due to the presence of protein (Protein error-of-indicators principle). Colors range from yellow for negative through yellow-green and green to green-blue for positive reactions. The double indicator principle gives a broad range of colors covering the entire urinary ph range. Colors range from orange through yellow and green to blue. Hemoglobin catalyzes the reaction of diisopropylbenzene dihydroperoxide and 3,3',5,5'-tetramethylbenzidine. Colors range from orange through green; very high levels of blood may cause the color development to continue to blue. Acetoacetic acid reacts with nitroprusside. Colors range from pink, for a negative reading, to maroon for a positive reading.

Page 2 of 8 Urobilinogen In a modified Ehrlich reaction, -diethylaminobenzaldehyde in conjunction with a color enhancer reacts with urobilinogen in a strongly acid medium to produce a pink-red color. Nitrite Leukocyte Specific Gravity Nitrate (derived from the diet) is converted to nitrite by the action of Gram negative bacteria in the urine. At the acid ph of the reagent area, nitrite in the urine reacts with -arsanilic acid to form a diazonium compound. This diazonium compound couples with 1,2,3,4-tetrahydrobenzo(h)quinolin-3-ol to produce a pink color. Esterases in granulocytic leukocytes catalyze the hydrolysis of the derivatized pyrrole amino acid ester to liberate 3-hydroxy-5-phenyl pyrrole. This pyrrole then reacts with a diazonium salt to produce a purple product. pka changes occur for certain pretreated polyelectrolytes in relation to ionic concentration. In the presence of an indicator, colors range from deep bluegreen in urine of low ionic concentration through green and yellow-green in urines of increasing ionic concentration STORAGE: Store Multistix 10 reagent strips at room temperature (15-30 o C) and out of direct sunlight. All unused strips must remain in the original bottle. Transferring to any other container may cause reagent strips to deteriorate and become nonreactive. Do not use the strips after their expiration date. Do not remove the desiccant from the bottle. Protection against exposure to light, heat, and ambient moisture is mandatory to guard against altered reagent activity. Avoid touching the test areas of the reagent strip. Discoloration or darkening of the reagent areas may indicate deterioration, discard the deteriorated strips. Quantrimetrix Urine Quality Control is stored in EU Department at room temperature and is stable for 30 days. New Quality control vials are supplied by the Point of Care testing area in the Core lab. These are stored refrigerated in 2N54 and need to be brought to room temperature to run quality control. SPECIMEN 10 ml (optimal) freshly voided, well mixed urine. Do Not centrifuge urine. Minimum volume = 1mL. Specimen must be labeled with patient identification. Collect the urine in a clean, dry, covered container. If the urinalysis can not be performed within 30 minutes, the urine should be refrigerated. Refrigerated specimens must be analyzed with 8 hours. Refrigerated urines must be brought to room temperature before analysis. UNACCEPTABLE SPECIMENS: 1. Any specimen left at room temperature more than 2 hours. 2. Any refrigerated specimen greater than 8 hours old. 3. Grossly bloody specimens should be sent to the core laboratory for analysis. 4. Specimens with urine preservatives. If an unacceptable specimen is received, note the reason for rejection and the physician notified in Wellsoft and request a new, acceptable specimen from the patient.

Page 3 of 8 PROCEDURE - QUALITY CONTROL 1. At the main Select screen, touch QC Test/Due QC. QC Due will appear when 24 hours have passed since the last Quality Control analysis. 2. Next screen select: QC Strip Test (Required). Required will be on the screen when 24 hours have passed since the last Quality Control analysis. 3. Screen then prompts for Operator ID. User the scanner( Yellow Button) and scan your SLCH ID barcode. Press Enter. Users when competency has been completed will be scanned in as users. This will not allow anyone that has not been trained to enter the system. 4. Screen prompt then requires the operator to enter their name. Use the alpha keyboard and enter your name. Press Enter. 5. Control 1 Level 1 appears on the Screen. Press Enter Lot Number and Expiration date. 6. Enter Level 1 (White Cap) Lot Number using the alpha key pad. The current lot number is posted by the Clinitek Status. Press Enter. 7. Enter Control Level 1 Expiration Date. Found on vial under Lot number. The current lot/expiration date is posted by the Clinitek Status. Press Enter. 8. Press Enter new Lot and Expiration date of MultiStix. 9. Enter Lot number on Siemens MultiStix 10SG. Lot number is found on the vial and is posted by the Clinitek Status. Press Enter. 10. Strip expiration screen appears. Using the arrow key enter the year and month. Expiration date is on the vial of MultiStix 10SG and is posted by the Clinitek Status. Press enter. 11. START button will appear. BEFORE you press start, take one MultiStix out of the vial and have the Level 1 QC ready to dose the strip. Once you press START you have 8 seconds to apply Quality Control 1 drop to each pad, starting at the bottom of the strip. This is the Glucose pad (aqua) color. Remove excess urine. Blot strip by turning on edge against a paper towel. Place strip on the strip guide on the Clinitek Status reader table. 12. At the end of the 8 Second countdown, the test table and strip will automatically be pulled into the instrument. WARNING: Do not push or pull the test table. 13. The analyzer will perform an automatic calibration and finish analyzing the sample. 14. Once the analysis is complete the Results screen will be displayed. The instrument will PASS or FAIL the quality control. QC Pass select the PRINT button and results will be printed. Place in the envelope marked Quality Control next to the instrument. Results will be entered manually in the Cerner Lab system. You do not have to manually record the results from the quality control. The instrument knows the status of pass/fail and will not allow patient testing until successful quality control has been analyzed.

Page 4 of 8 15. Select Done. The Clinitek then prompts you for Control 2.. Select Enter Lot and expiration date. Press Enter. 16. Enter Control lot number for Control 2 (Blue Cap) and expiration date. This is posted by the Clinitek. Same screens as for Control 1. 17. Strip Lot: Select: Use last lot is this is the same lot number of MultiStix used for running Control 1. Press Enter. 18. START. Press only when you are ready to apply Control 2 to MultiStix. Same procedure as in #11 above. 19. The results will appear once the anlysis is completed. The instrument will PASS or FAIL the quality control. QC PASS select PRINT button and results will be printed. Place in the envelope marked Quality Control next to the instrument. See #14 above. 20. Select Done. You will now be able to perform patient testing when both levels of Quality Control have passed. 21. When one or both levels of the Quality Control fail the instrument will prompt you to repeat the failed QC test. Repeat and follow the instrument screens. Remember to check to make certain that you are running Level 1 when prompted by the instrument, if you mix up the Quality Control it will not pass. Quality Control fails contact the Core Lab/Point of Care and send specimens to the Core Lab until Quality Control can be resolved. When Instrument failures occur specimens must be sent to the Core Lab for testing. You must put the strip on the test table in the correct position or it will not assay the strip. When this happens it pulls the strip in and it is not analyzed but is returned to the original position. You must get a new strip and repeat the assay. PROCEDURE PATIENTS 1. At the main screen, touch Strip Test. The Operator ID screen will appear. 2. Scan operator ID badge, press Enter. 3. Select Enter New Patient. 4. On the Alpha Key Board type in the Patient s name. Press Enter. 5. Patient ID screen will appear, Scan the patient s MRN bar code. Press Enter. 6. Press START Button. Once you touch the START button you have 8 seconds to dip the test strip into urine specimen and place it on the test table.

7. Dip test strip in urine specimen. Remove excess. Blot strip by turning on edge against absorbent orange wipe or paper towel. 8. Place the strip on the instrument test table. DO NOT PUSH OR PULL THE TEST TABLE. 9. At the end of the 8 second countdown, the test table and strip will automatically be pulled into the instrument. 10. The analyzer will perform an automatic calibration and finish analyzing the sample. Page 5 of 8 11. When analysis is complete, the RESULTS screen will be displayed. Results will automatically print. Record results in WELLSOFT. Result tapes should be placed in the shred for HIPPA. 12. Remove the test strip and dispose of it in biohazard container. 13. Touch done to complete the test and return to the main Select screen. CALIBRATION The Clinitek Status Analyzer performs a self test and calibration each time it is turned on. In addition, the analyzer performs an automatic calibration each time a test is run. The while calibration bar (on the test Table) provides NIST traceable calibration. SYSTEM START UP AND MAINTENANCE The system is turned on by pressing the on/off button located at the front of the instrument. The analyzer automatically runs a system diagnostic check during which it performs a series of electronic, signal and memory checks, as well as ensures there is sufficient battery voltage to operate when necessary. Daily Maintenance (Performed by the EU staff or Point of Care Staff) A. Wipe off instrument. B. Clean feed table. Remove the feed table insert by grasping the front end of the insert on each side with a thumb and finger. Clean the insert with a gauze or kimwipe wet using distilled water. Weekly Maintenance (Performed by the EU staff or Point of Care Staff) A. Wipe the feed table with a kimwipe dampened with distilled water. Replace the feed table insert the same way as when removing it. Periodic Maintenance (Performed by the EU staff or Point of Care Staff) A. Change the paper. See Operator s Manual for instructions. B. Extensive cleaning of feed table insert when excessively dirty. Remove the feed table insert as described under the daily maintenance procedure.

Page 6 of 8 NOTE: Do not use any material that will scratch the calibration chip. Do not use solvents of any kind to clean the feed table, insert, or chip. If the calibration chip becomes damaged, the feed table insert must be replaced. LIMITATIONS: 1. Substances that cause abnormal urine color may affect integrity of reagent areas on strip. The color development may be masked or a color produced that could be interpreted as a false positive. Results should include a notation of urine color and a comment about the possible false positive results. 2. Protein: visibly bloody urine may cause elevated results. 3. Blood: Capoten (captopril) may reduce the sensitivity. Certain oxidizing contaminants, such as hypochlorite, may produce false positive results. Microbial peroxidase associated with urinary tract infection may cause a false positive reaction. 4. Leukocytes: Elevated glucose concentrations (> 3 g/dl) may cause decreased test results. The presence of some cephalasporin antibiotics or high concentrations of oxalic acid may also cause decreased test results. Tetracycline may cause decreased reactivity, and high levels of the drug may cause a false negative reaction. Positive results may occasionally be due to contamination of the specimen by vaginal discharge. 5. Nitrites: Pink spots or pink edges should not be interpreted as a positive result. A negative result does not rule out significant bacteriuria. False negative results may occur with shortened bladder incubation of the urine, absence of dietary nitrate, or the presence of nonreductive pathological microbes. 6. Glucose: Ketone bodies reduce the sensitivity of the test; moderately high ketone levels (40 mg/dl) may cause false negatives for specimens containing small amounts of glucose (75-125 mg/dl), but the combination of such ketone levels and low glucose levels is metabolically improbable in screening. 7. Ketone: False trace results may occur with highly pigmented urine specimens or those containing large amounts of levodopa metabolites. Compounds such as mesna (2-mercaptoethane sulfonic acid) that contain sulfhydryl groups may cause false positive results or an atypical color reaction. 8. Bilirubin: Indican (indoxyl sulfate) can produce a yellow-orange to red color response that may interfere with the interpretation of a negative or positive reading. Metabolites of Lodine may cause false positive or atypical results. Atypical colors (colors that are unlike the negative or positive color blocks shown on the color chart) may indicate that bilirubin-derived bile pigments are present in the urine sample and may be masking the bilirubin reaction. 9. Urobilinogen: The test pad may react with interfering substances such as p-aminosalicylic acid and sulfonamides. Atypical color reactions may be obtained if formalin is present. Strip reactivity increases with temperature; the optimum temperature is 22-26 o C. The test is not a reliable method for the detection of porphobilinogen. 10. Bilirubin and urobilinogen are very unstable when exposed light. 11. Contamination of the urine specimen with skin cleaners containing chlorhexidine may affect protein,

Page 7 of 8 bilirubin and specific gravity test results. Results that do not match the patients clinical picture, a specimen should be sent to the Core Lab for verification. Correlations with the Atlas and Siemens Manual Dipstick are performed quarterly. Status instrument is down specimens should be sent to the Core lab. REFERENCE RANGE: Specific gravity values normally range between 1.008 and 1.022 depending on fluid intake Analyte Normal ph 5 8 Protein Glucose Qual. Ketones Qual. Bilirubin Qual. Urobilinogen 0.2-1.0 Ehrlich units/dl Nitrites Leukocyte Esterase REFERENCES: 1. Siemens Multistix Reagent Strips Product Insert, AN30516C, Rev. 4/99. 2. Freeman, J.A. and M.F. Beeler: Laboratory Medicine - Clinical Microbiology, Lea and Febiger, Philadelphia, PA, 1974. 3. Graff, S.L., A Handbook of Routine Urinalysis, J.B. Lippincott 4. Henry, John, Clinical Diagnosis and Management by Laboratory Methods, W.B. Saunders Co., 16 th ed., 1979. 5. Race, G.L. and M.G. White: Basic Urinalysis, Harper and Row, Hagerstown, MD, 1979. 6. Siemens Medical Solutions Diagnostics Clinitek Status Analyzer Operator s Manual 13287. rev. T, 2008-05.

Page 8 of 8 Donna Walck Written By: Technical Supervisor Medical Director Date Date Annual Review: Medical Director/Designee Date Location of hard copy location(s): Core Lab Office See Test Methods / Test Locations (POC.1.0002.0) Date Archived: