DEVELOPMENT AND VALIDATION OF A REVERSE-PHASE HPLC METHOD FOR ANALYSIS OF TEMOZOLOMIDE IN A CAPSULE FORMULATION

Similar documents
RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms

Pelagia Research Library

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS TOLBUTAMIDE IN PURE AND PHARMACEUTICAL FORMULATIONS

Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by HPLC Method

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page:

Pelagia Research Library

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

International Journal of Pharma and Bio Sciences

Scholars Research Library

SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

Reverse Phase HPLC Analysis of Atomoxetine in Pharmaceutical Dosage Forms

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

Simultaneous Estimation of Gemcitabine Hydrochloride and Capecitabine Hydrochloride in Combined Tablet Dosage Form by RP-HPLC Method

Development and Validation for Simultaneous Estimation of Sitagliptin and Metformin in Pharmaceutical Dosage Form using RP-HPLC Method

RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form

ISSN (Print)

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article

DEVELOPMENT AND VALIDATION OF NEW HPLC METHOD FOR THE ESTIMATION OF PALIPERIDONE IN PHARMACEUTICAL DOSAGE FORMS

Hyderabad, India. Department of Pharmaceutical Chemistry, Glocal University, Saharanpur, India.

International Journal of Pharma and Bio Sciences DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET

ISSN: ; CODEN ECJHAO E-Journal of Chemistry 2011, 8(3),

Journal of Pharmacreations

Scholars Research Library. Der Pharmacia Lettre, 2016, 8 (6): (

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

J Pharm Sci Bioscientific Res (4): ISSN NO

A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide in Bulk and Pharmaceutical Dosage Forms

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

World Journal of Pharmaceutical Research

RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR THE ESTIMATION OF BACLOFEN IN BULK AND PHARMACEUTICAL DOSAGE FORMS

36 J App Pharm Vol. 6; Issue 1: 36-42; January, 2014 Rao et al., 2014

New RP-HPLC Method for the Determination of Fludarabine in Pharmaceutical Dosage Forms

Development and Validation of Stability Indicating HPTLC Method for Estimation of Seratrodast

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

Tentu Nageswara Rao et al. / Int. Res J Pharm. App Sci., 2012; 2(4): 35-40

A HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC ASSAY FOR LERCANIDIPINE HYDROCHLORIDE

Scholars Research Library

Development of a Validated RP-HPLC Method for the Analysis of Citicoline Sodium in Pharmaceutical Dosage Form using Internal Standard Method

Journal of Chemical and Pharmaceutical Research

DEVELOPMENT OF RP-HPLC METHOD FOR ESTIMATION OF DROTAVERINE HYDROCHLORIDE IN PHARMACEUTICAL FORMULATIONS

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

Development and Validation of RP-HPLC Method for the Estimation of Gemigliptin

Development and validation of RP-HPLC method for simultaneous estimation of gliclazide and metformin in pure and tablet dosage form

Method Development and Validation for the Estimation of Saroglitazar in Bulk and Pharmaceutical Dosage Form by RP-HPLC

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ESTIMATION OF LACOSAMIDE IN BULK AND ITS PHARMACEUTICAL FORMULATION

Development and validation of related substances method for Varenicline and its impurities

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD ESTIMATION OF TOLVAPTAN IN BULK PHARMACEUTICAL FORMULATION

International Journal of Medicine and Pharmaceutical Research. International Journal of Medicine and Pharmaceutical Research

Isocratic Reversed Phase Liquid Chromatographic Method Validation for the Determination of Cilostazol in Pure and Formulations

Available online at Scholars Research Library

Corresponding Author:


VALIDATED RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF DEXAMETHASONE AND GRANISETRON IN COMBINED DOSAGE FORMS

Airo International Research Journal ISSN : Volume : 7 October 2015

Research Article. ISSN Available online at 746

A stability indicating RP-HPLC method for simultaneous estimation of darunavir and cobicistat in bulk and tablet dosage form

Simultaneous estimation of Metformin HCl and Sitagliptin in drug substance and drug products by RP-HPLC method

S. G. Talele, D. V. Derle. Department of Pharmaceutics, N.D.M.V.P. College of Pharmacy, Nashik, Maharashtra, India

Stress Degradation Studies And Validation Method For Quantification Of Aprepitent In Formulations By Using RP-HPLC

Pankti M. Shah et al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; 07-16

Development and validation of liquid chromatographic method for trazodone hydrochloride

MEDAK DIST. ANDHRA PRADESH STATE, INDIA. Research Article RECEIVED ON ACCEPTED ON

Available online Research Article

RP- HPLC and Visible Spectrophotometric methods for the Estimation of Meropenem in Pure and Pharmaceutical Formulations

Estimation of Etoricoxib in Tablet Dosage form by RP- HPLC using Internal Standard with Emphasize on Specificity Parameter Method

210 J App Pharm Vol. 6; Issue 2: ; April, 2014 Mustafa et al, 2014

ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF SOLID DOSAGE FORM OF ANTINEOPLASTIC DRUG IMATINIB MESILATE BY RP HPLC.

Research Article Simultaneous Estimation of DL-Methionine and Pyridoxine Hydrochloride in Tablet Dosage Form by RP-HPLC

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR CILOSTAZOL IN TABLET DOSAGE FORM

Estimation of Zanamivir Drug present in Tablets using RP-HPLC Method

NOVEL RP-HPLC METHOD. B.Lakshmi et a. concentration range KEY INTRODUCTION. Diltiazem is used to. . It works by of contractionn of the. (dilate).

Estimation of zolmitriptan by a new RP-HPLC method

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article

REVERSE PHASE HPLC METHOD FOR THE ANALYSIS OF ALFUZOSIN HYDROCHLORIDE IN PHARMACEUTICAL DOSAGE FORMS

Method Development And Validation Of Losartan Potassium And Hydrochlorothiazide In Comined Dosage Form By RP-HPLC

Research Article. A novel validated RP-HPLC-DAD method for the estimation of Lenvatinib Mesylate in bulk and pharmaceutical dosage form

Analytical Method Development and Validation for the Estimation of Imatinib Mesylate and its Impurity in Pharmaceutical Formulation by RP-HPLC

Development and validation of UV-visible spectrophotometric method for estimation of rifapentine in bulk and dosage form

Estimation of Emtricitabine in Tablet Dosage Form by RP-HPLC

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF ETORICOXIB AND THIOCOLCHICOSIDE IN PHARMACEUTICAL DOSAGE FORMS

Stability indicating RP-HPLC method development and validation of Etizolam and Propranolol hydrochloride in pharmaceutical dosage form

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research

International Journal of Drug Research and Technology

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

CHAPTER 3: ANALYTICAL METHOD DEVELOPMENT AND VALIDATION

Research and Reviews: Journal of Pharmacy and Pharmaceutical Sciences

International Journal of Advances in Pharmaceutical Analysis IJAPA Vol. 5 Issue 4 (2015) 85-92

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ESTIMATION OF FEBUXOSTAT IN BULK AND TABLET DOSAGE FORM

ANALYTICAL TECHNIQUES FOR THE QUALITY OF ANTI- AIDS DRUGS

Scholars Research Library. Der Pharmacia Lettre, 2015, 7 (5):44-49 (

Asian Journal of Research in Chemistry and Pharmaceutical Sciences

Development and Validation of Stability Indicating HPLC method for Determination of Eperisone HCL in Bulk and in Formulation

Volume 2 (6), 2014, Page CODEN (USA)-IJPRUR, e-issn: International Journal of Pharma Research and Health Sciences

Sujatha and Pavani et.al. Indian Journal of Research in Pharmacy and Biotechnology ISSN: (Print) ISSN: (Online)

Validation of Changes to the USP Assay Method for Ibuprofen Tablets

SIMPLE AND VALIDATED RP-HPLC METHOD FOR THE ESTIMATION OF CARBOPLATIN IN BULK AND FORMULATION DOSAGE FORM. Subhashini.Edla* B.

Transcription:

Int. J. Chem. Sci.: 11(2), 2013, 1055-1063 ISS 0972-768X www.sadgurupublications.com DEVELOPMET AD VALIDATIO OF A REVERSE-PHASE HPLC METHOD FOR AALYSIS OF TEMOZOLOMIDE I A CAPSULE FORMULATIO B. MOHAMMED ISHAQ a, K. VAITHA PRAKASH * and G. KRISHAMOHA b Department of Pharmaceutical Analysis, SSJ College of Pharmacy, Gandipet, HYDERABAD (A.P.) IDIA a Research Scholar, Research and Development, Jawaharlal ehru Technological University, Kakinada, KAKIADA (A.P.) IDIA b Centre for Pharmaceutical Sciences, IST, JTU-H, HYDERABAD (A.P.) IDIA ABSTRACT Temozolomide is an antineoplastic agent with activity against a broad spectrum of murine tumors. A high-performance liquid chromatographic method was developed and validated for the analysis of temozolomide in capsule formulation. In the present work, chromatographic separation of temozolomide was achieved on Waters HPLC with a Zorbax column (4.6 x 150 mm, 3 μm) using sodium phosphate buffer ph 7: methanol 30:70 v/v at a flow rate of 0.8 ml/min and detection UV wavelength of 287 nm. The injection volume was 20 μl and the column temperature was maintained at 28 C. The method was validated according to ICH requirements. The validation characteristics included accuracy, precision, specificity, linearity, and analytical range. The standard curve was found to have a linear relationship (r 2 = 0.99) over the analytical range of 20-60 μg/ml. The detection limit was 0.01µg/mL and the quantification limit was 0.03 µg/ml. The recovery/accuracy of the method ranged from 50 to 150%. Data collected in this study were analyzed using JMP statistical software package by one-way analysis of variance (AOVA). The assay was shown to be specific, accurate, precise, and reliable for use in routine analysis of Temozolomide capsules. Key words: Temozolomide capsules, Antineoplastic agent, Zorbax column, High-performance liquid chromatography; Sodium phosphate buffer. ITRODUCTIO Temozolomide (8-carbamoyl-3-methyl-imidazo- [5, 1-d]-1,2,3,5-tetrazin-4-(3H)-one (Fig. 1) is an alkylating agent of the imidazotetrazine derivatives that exhibits broad- * Author for correspondence; E-mail: prakash.karanam@gmail.com; Ph.: +91-9885355562

H C 3 1056 K. V. Prakash et al.: Development and Validation of. spectrum antitumor activity against murine tumors 1. The literature is silent on development an assay method using HPLC, though a number of reports exist on procedures for its determination from biological fluids, such as plasma and urine 1-5. The purpose of this work was to develop and validate an RP-HPLC method, employing UV detection, which would allow for analysis of the Temozolomide its capsule formulation without the need for an extraction. O H 2 O Fig. 1: Chemical structure of temozolomide EXPERIMETAL Materials Temozolomide was received gratis from Taj Pharmaceuticals Ltd. (Mumbai, India) and was used as received. Sodium hydroxide and HPLC grade methanol were purchased from LOBA Chemie Pvt. Ltd. (Mumbai, Maharashtra) and Merck (Darmstadt, Germany), respectively. Ultra-pure water was obtained from ELGA (Bucks, UK) water purification unit. Waters total recovery vials (Waters, Milford, MA, USA) were of glass type1, class A with 950 μl maximal injectable volume. All other chemicals were of analytical reagent grade. Instrumentation The HPLC system consisted of a Waters 2690 separation module (Waters, Milford, MA, USA) coupled with a UV detector. A Zorbax C18 (4.6 x 150 mm, 3 μm, Make: Zorbax) column, at 28 ± 2 C, was used for the separations. The injection volume was 20 μl and the analytical wavelength was 287 nm. Instrument control, data acquisition, processing and peak symmetry was completed with the Empower 2 Software (Waters, Milford, MA, USA). An isocratic chromatographic method was employed for the separation of the API from the capsule excipients. The conditions of the method were, sodium phosphate buffer ph 7: methanol (30 : 70 v/v) as mobile phase with a flow rate of 0.8 ml/min. Total run time was 6 min.

Int. J. Chem. Sci.: 11(2), 2013 1057 System suitability The HPLC system was equilibrated with the initial mobile phase composition, followed by six injections of the same standard. These six consecutive injections were used to evaluate the system suitability on each day of method validation. Standard preparation The primary stock solution of temozolomide was prepared by analytically weighing approximately 10 mg of temozolomide and diluting to a final volume of 10 ml with the mobile phase, further 0.4 ml of the above stock solution was taken into a 10 ml volumetric flask and dilute upto the mark with mobile phase resulting in a stock solution at or near 40 μg/ml and filtered through 0.45 µm membrane filter. Serial dilutions were performed, with mobile phase, to make five standards, ranging from 20 to 60 μg/ml. These standards were then stored at 4 C and replaced every 2 weeks. Three standard calibration curves were injected periodically throughout the HPLC procedure, in order to establish linearity of detector response to analyte concentration. Preparation of 40 μg/ml capsule formulation solution Analytically weighed portions of equivalent weights of 20 temozolomide capsules were diluted with 10 ml of mobile phase. Samples were sonicated (Bransonic-1510R, Bransonic Ultrasonics Corporation, Danbury, CT) for 10 min. further 0.4 ml of the above solution was diluted to 10 ml with mobile phase and filtered through 0.45 µm membrane filter. Aliquots were then transferred to HPLC vials for injection. Statistical analysis Data collected in this study were analyzed using JMP statistical software package by one-way analysis of variance (AOVA). Univariate linear regression analysis using least square method was applied to test the fitted model. Correlation coefficient was calculated and the results of the statistical analysis were considered significant if their corresponding p-values were less than 0.05. RESULTS AD DISCUSSIO Development and optimization of isocratic HPLC conditions Initial assay parameters were based on physical-chemical properties of temozolomide. Due to the octanol water partition coefficient (Log P) -2.8 of the parent compound, a C18

1058 K. V. Prakash et al.: Development and Validation of. column was selected for development. Based on the Log P and the melting point, 212 C, temozolomide has a calculated water solubility of 5.09 e + 00 g/l. 6 Therefore a reversed phase assay was deemed most appropriate for initial testing. A UV scan of temozolomide showed a maximal absorbance at 287 nm. A typical chromatogram of standard preparation (40 μg/ml) and a chromatogram of capsule dosage form were shown in Fig. 2 and 3. 0.60 0.50 2.829 0.40 UA 0.30 0.20 0.10 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 Minutes Fig. 2: Chromatograph of standard temozolomide preparation 0.60 2.834 0.50 0.40 UA 0.30 0.20 0.10 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 Minutes Fig. 3: Chromatograph of temozolomide capsule dosage form Preliminary method development was conducted on a C18 4.6 x 150 mm, 3 μm column, operated at 28 ± 2 C. This column provides efficient and reproducible separations of nonpolar compounds while minimizing solvent usage. Consequently, it was selected for method development and remains the column utilized in the validated assay. Preliminary

Int. J. Chem. Sci.: 11(2), 2013 1059 method development of suitable isocratic conditions to resolve temozolomide, on the C18 column, was conducted with sodium phosphate buffer ph 7.0 : methanol 30 : 70 v/v as the mobile phase. A mobile phase of sodium phosphate buffer ph 7.0 : methanol 30 : 70 v/v was found to provide a reproducible, baseline resolved peak with an average peak tailing factor of 1.59. Based on the pressure limitations of the column and the HPLC system, flow rates were maximized to decrease assay time without adversely affecting the system or the column. Typical pressure throughout the method ranges from 2750 to 3250 psi. The length of time in each portion of the assay was varied to accommodate a 6-min run time. Method validation Linearity To determine linearity, five standard solutions of increasing concentration were injected in triplicate. The calibration curve was prepared by plotting the area under the response from the detector (AUC) line versus the concentration and analyzed through least squares regression (Table 1). The assay was found to be linear in the range from 20 to 60 μg/ml. Calibration curves were linear with an average correlation coefficient of r 2 0.999. Calibration plot was shown in Fig. 3. The difference between the experimental and calculated concentrations is represented by the relative error (R. E.); the low absolute value of the R.E. shows the absence of scatter of the data from the linear best-fit line (Fig. 4 and Table 1). 3000000 Peak area 2500000 2000000 1500000 1000000 y = 41063 x + 305685 2 R = 0.9993 500000 0 0 10 20 30 40 50 60 70 Concentration ( μg/ml) Fig. 4: Linearity plot of HPLC method for analysis of temozolomide capsules

1060 K. V. Prakash et al.: Development and Validation of. Table 1: Linearity of HPLC method for analysis of temozolomide capsules S. o. Linearity level Concentration Area 1 I 20 µg/ml 1121401 2 II 30 µg/ml 1529276 3 III 40 µg/ml 1978755 4 IV 50 µg/ml 2344717 5 V 60 µg/ml 2766815 Accuracy and precision Correlation coefficient 0.9993 The accuracy of this method was determined as the percentage of the theoretical drug recovered from the temozolomide capsules. Mean drug recoveries was 100.5%, with standard deviation of 0.86. Accuracy was determined by the three QC standards and evaluated for 3 days as an average drug content percentage. This data convey that the assay is highly accurate based on the recovery of drug from the capsule matrix. The method precision was established by five injections of the standard QC sample at a 40 μg/ml concentration level for the intra-day precision and on 3 days for the intermediate precision. Precision was expressed as a coefficient of variation percentage (CV %) of the analyte peak. Result for the accuracy of temozolomide is summarized in Table 2. Results for the intra- and inter-day precision are summarized in Table 3. Table 2: Accuracy of HPLC analysis of temozolomide capsules (n = 3) % Concentration (at specification level) Average area Amount added (mg) Amount found (mg) % Recovery Mean recovery 50% 2006872 * 5.0 5.0 100.0% 100% 4014113 ** 10.0 10.0 100.0% 100.5% 150% 6104804 * 15.0 15.2 101.4% * n = Average area of six replicate samples ** n = Average area of three replicate samples

Int. J. Chem. Sci.: 11(2), 2013 1061 Table 3: Precision of HPLC analysis of temozolomide capsules Injection Intraday precision Area Intraday precision Injection-1 1871423 1869365 Injection-2 1876279 1868938 Injection-3 1874529 1861814 Injection-4 1879273 1867522 Injection-5 1873436 1866552 Average 1874987 1866837 Standard deviation 2973.1 3023.9 % RSD 0.2 0.2 Limit of quantitation and detection The limit of detection (LOD) and limit of quantitation (LOQ) of this method were determined from the standard deviation of the response of a known concentration of temozolomide, as per ICH Q2B guidelines 7,8. The LOD for this assay, calculated from three times the noise level of the response, is 0.01 µg/ml. The LOQ for this assay, calculated from ten times the noise level of the response, is 0.03 µg/ml. Specificity and selectivity The selectivity of the assay for temozolomide is shown in the chromatographs in Fig. 2 and 3. The placebo shows no detector response near retention time of temozolomide: 2.85 min, the specificity of this method has been confirmed with Ultra violet detector and analyzed with Empower Manger Software. A library spectrum was used to confirm peak identity and the absence of any impurity that possesses a different UV spectrum from temozolomide. These results support the specificity and selectivity of this assay for temozolomide. Assay A mass of not less than 20 capsules were prepared to a fine, uniform particle size powder using a sieve. After calculating the average capsule content weight, a composite equivalent to the 10 mg was accurately weighed and quantitatively transferred into a 100 ml volumetric flask. Approximately, 30 ml of diluent was added; the solution was sonicated for

1062 K. V. Prakash et al.: Development and Validation of. 10 min, the flask was equilibrated to room temperature, carefully filled to volume with the diluent, and mixed well. A portion of the solution was filtered through a 0.45 mm membrane filter, discarding the first 2-3 ml of the filtrate. A portion of the filtered sample (2.0 ml) was diluted into a 50 ml volumetric flask with the mobile phase and mixed well. Twenty micro liters of standard and sample solutions were injected into an injector of HPLC, peak area of standard amount of drug and the sample were computed. The Percentage purity of the temozolomide in capsules was determined by following formula: Where: AT WS DT P Avg. wt Assay % = 100 AS DS WT 100 Label claim AT = Peak Area of Temozolomide obtained with test preparation AS = Peak Area of Temozolomide obtained with standard preparation WS = Weight of working standard taken in mg WT = Weight of sample taken in mg DS = Dilution of Standard solution DT = Dilution of sample solution P = Percentage purity of working standard The results of assay of temozolomide capsules were shown in Table 4. Table 4: Assay of temozolomide capsules S. o. Sample area % Assay 1 1871423 99.80 2 1876279 100.06 3 1874529 99.97 4 1879273 100.22 5 1873436 99.91 6 1874987 99.99 Average assay 100 STD 0.14 % RSD 0.14

Int. J. Chem. Sci.: 11(2), 2013 1063 COCLUSIO An optimized RP-HPLC assay, employing UV detection, has been developed and validated for temozolomide in a capsule formulation. The isocratic RP-HPLC assay allows for the separation of temozolomide and the capsule components, without the need for an extraction procedure. The linearity, precision, accuracy, reproducibility, and selectivity nature of this assay have been established, and have been deemed sufficient to support regular quality control analysis of capsules of temozolomide. ACKOWLEDGEMET The authors express sincere thanks to Management of Annamacharya College of Pharmacy for their cooperation in the present research work. REFERECES 1. H. Kim, P. Likhari, D. Parker, P. Statkevich, A. Marco, C.-C. Lin et al., J. Pharmaceut. Biomed. Analy., 24, 461-468 (2001). 2. F. Shen, L. A. Decosterd, M. Gander, S. Leyvraz, J. Biollaz and F. Lejeune, J. Chromatogr. B, 667, 291-300 (1995). 3. M. Andrasi, R. Bustos, A. Gaspar, F. A. Gomez and A. Klekner, J. Chromatogr. B, 878, 1801-1808 (2010). 4. H. Ki Kim, C.-C. Lin, D. Parker, J. Veals, J. Lim, P. Likhari et al., J. Chromatogr. B, 703, 225-233 (1997). 5. F. Shen, L. A. Decosterd, M. Gander, S. Leyvraz, J. Biollaz and F. Lejeune, J. Chromatogr. B, 667, 291-300 (1995). 6. S. Preissner, K. Kroll, M. Dunkel, C. Senger, G. Goldsobel, D. Kuzman, S. Guenther, R. Winnenburg, M. Schroeder and R. Preissner, ucleic Acids Res., (2010) Jan. 38 (Database issue): D237-43. Epub. (2009). 7. ICH Q2B: Validation of Analytical Procedures: Methodology (1997) pp. 1-8. 8. FDA: Guidance for Industry, Analytical Procedures and Methods Validation, August (2000). Accepted : 11.04.2013