Managing the imaging dose during image-guided radiation therapy

Similar documents
Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

Concept for quantifying the dose from image guided radiotherapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Collapsed Cone Convolution 2D illustration

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Radiation Safety in the Catheterization Lab

1. Murphy, Martin J., et al. "The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75.

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

SRS Uncertainty: Linac and CyberKnife Uncertainties

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

8/3/2016. Outline. Site Specific IGRT Considerations for Clinical Imaging Protocols. Krishni Wijesooriya, PhD University of Virginia

In-Room Radiographic Imaging for Localization

In-Room Radiographic Imaging for Localization

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Flattening Filter Free beam

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

Biological Effects of Ionizing Radiation Module 8 - AAPM/RSNA Curriculum. Basic Radiation Biology

Motion gating and tracking techniques: overview and recent developments

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

I. Equipments for external beam radiotherapy

Patient dose assessment of CT perfusion scanning at the RSCH

Imaging Rotation. University of Michigan Department of Radiation Oncology Division of Radiation Physics. Resident:

CT Radiation Risks and Dose Reduction

Credentialing for the Use of IGRT in Clinical Trials

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Subject: Image-Guided Radiation Therapy

Why radiation protection matters?

Why is CT Dose of Interest?

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

Sarcoma and Radiation Therapy. Gabrielle M Kane MB BCh EdD FRCPC Muir Professorship in Radiation Oncology University of Washington

Dosimetric Consideration in Diagnostic Radiology

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

Overview of Advanced Techniques in Radiation Therapy

IMAGING DOSES IN RADIATION THERAPY FROM KILOVOLTAGE CONE-BEAM COMPUTED TOMOGRAPHY


Radiation Protection- Cath lab

SBRT REQUIRES: STEREOTACTIC BODY RADIOTHERAPY STEREOTACTIC BODY RADIOTHERAPY (SBRT) (SBRT) What s s in a name? Stereotactic Body Radiotherapy

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Implementation of advanced RT Techniques

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT

typical IMRT fraction time and expand high definition radiotherapy anywhere in the body with the widest range of motion of the

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

7/31/2018. Research Activities: Imaging for RT Image Guidance for Pediatric Radiotherapy. Chia-ho Hua, PhD. Outline

Medical Fluoroscopic Exposures and Incidents

Learning Objectives. New Developments in Radiation Therapy Targeting. Respiration-Induced Motion. Targeting Uncertainty in RT

SBRT I: Overview of Simulation, Planning, and Delivery

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

EXACTRAC HIGHLY ACCURATE PATIENT MONITORING

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic

Uncertainties and Quality Assurance of Localization and Treatment in Lung SBRT Jing Cai, PhD Duke University Medical Center

Radiation related cancer risk & benefit/risk assessment for screening procedures

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1

Facing the challenges: IAEA activities to protect the patients

Image Guided Proton Therapy and Treatment Adaptation

Vascular & Interventional Education Days Thomas M Griglock, Ph.D., DABR Chief Diagnostic Imaging Physicist, OHSU

Mapping the ASRT Objectives for Radiation Protection (47 Objectives) and Radiation Biology (21 Objectives) to this Text

Has radiotherapy the potential being focal?

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

An audit of radiation dose of 4D CT in a radiotherapy department

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Advances in external beam radiotherapy

Radiation Quantities and Units

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

TESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

8/2/2012. Transitioning from 3D IMRT to 4D IMRT and the Role of Image Guidance. Part II: Thoracic. Peter Balter, Ph.D.

Computed tomography Acceptance testing and dose measurements

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2

Tumor Tracking Current & Future Developments. Josh Evans, Ph.D. Virginia Commonwealth University Richmond, VA

Corporate Medical Policy

Innovative RT SBRT. The variables with REQ in superscript are required.

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Skyscan 1076 in vivo scanning: X-ray dosimetry

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System

Transcription:

Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA

Imaging during radiotherapy Radiographic image guidance has emerged as the new paradigm for patient positioning, target localization, and external beam alignment in radiotherapy. The AAPM has recognized the importance of imaging dose management by supporting a task group report on the subject: Murphy M, Balter J, Balter S, BenComo J, Das I, Jiang S, Ma C, Olivera G, Rodebaugh R, Ruchala K, Shirato H, Yin F, The management of imaging dose during image-guided radiotherapy, Report of the AAPM Task Group 75, Medical Physics 34(10): 4041 4063, 2007.

Outline Imaging procedures for radiotherapy Dose summation Dose evaluation Dose reduction Summary

Uses of radiographic imaging in Imageguided radiation therapy (IGRT) Precise daily positioning of the patient before treatment MV portal imaging dual kv planar imaging in-room fan-beam and cone-beam CT Intra-fraction motion monitoring kv radiography and fluoroscopy Daily plan adaptation fan-beam and cone-beam CT

BrainLab Exactrac x-ray

Varian On-board Imager (OBI)

CT on rails

CyberKnife

Cone-beam CT

Some IGRT scenarios daily pre-treatment CTs for 30 fractions: 60-400 msv two pairs of MV portal images daily for 30 fractions: 40-400 msv two minutes of daily kv fluoroscopy for 30 fractions: 40-120 msv 100 dual kv planar images daily for 5 fractions: 10-100 msv

Summing the imaging dose during IGRT

Difficulties in determining the imaging dose during IGRT Data for the dose delivered by the various radiographic imaging modalities being used during radiation therapy are presently scattered widely through the literature, making it difficult to estimate the total dose that the patient will receive during a particular treatment scenario. IGRT systems often are configured differently than diagnostic imaging setups.

Issues in summing the doses from different modalities different imaging scenarios distribute dose in fundamentally different ways, making it difficult to sum dose in a radiobiologically consistent manner: e.g., planar kv imaging dose attenuates rapidly along the line of sight; CT dose is uniformly distributed through the patient planar kv dose is evaluated as entrance (skin) dose in air kerma, without scattering; axial kv dose (CT) is evaluated as computed tomography dose index (CTDI), with scattering for kv, air kerma and absorbed dose are essentially the same; for MV they are not the same in regions of electronic disequilibrium (e.g., air/tissue boundaries). kv surface dose buildup layer is very thin; MV surface buildup layer is much deeper; thus skin doses are very different

Summing doses Because of the differing qualities of kv, planar, CT, and MV exposures, the doses should only be compared and summed in units of effective dose, which represents the approximate biological detriment associated with a given integral dose

Effective dose The concept of effective dose (or effective dose equivalent) was introduced to provide a common framework for evaluating the biological detriment of exposure to ionizing radiation via any means. From Jacobi (Radiat Environ Biophys 12, 1975): the mean absorbed dose from a uniform whole-body irradiation that results in the same total radiation detriment as from the non-uniform, partial-body irradiation in question.

Effective dose E = T w T H T millisieverts (msv) where the H T are the average organ doses to tissue T for a particular exam and the w T are tissue weighting factors that represent the relative radiation sensitivities of the organs. Unlike local absorbed dose, effective dose is a volume integral and thus depends on both the beam fluence and imaging area

Evaluating the imaging dose

Interpreting risk Two categories of risk: deterministic risk e.g., skin injury, cataracts has an approximate threshold that can be observed on an individual basis. stochastic risk e.g, the increased risk of a secondary cancer is probabilistic and is extrapolated from population-based data. Stochastic risk is estimated from the total effective dose

Examples of deterministic risk Effects Threshold Time of onset early transient erythema 2000 mgy 2 24 hours Temporary epilation 3000 mgy 1.5 weeks main erythema 6000 mgy 3 weeks permanent epilation 7000 mgy 3 weeks dermal necrosis 15,000 mgy > 52 weeks eye lens opacity > 1000 mgy > 5 years cataract (debilitating) > 5000 mgy > 5 years

Examples of stochastic risk from IGRT concomitant dose 70 year old male with prostate cancer CT for planning (8.2 msv) 30 daily portal image pairs (30x1.3 msv) 0.2% risk of secondary cancer 30 year old female patient w/ cervical cancer Planning CT (8.2 msv) 30 daily CTs for positioning (30x8.2 msv) 1.25% risk of secondary cancer

Difficulties in interpreting risk Diagnostic imaging and image-guided surgery introduce ionizing radiation, while IGRT adds it to an already considerable therapeutic exposure. Increased imaging dose during IGRT can reduce normal tissue exposure to the treatment beam, thus reducing overall concomitant dose.

Difficulties in interpreting risk not everyone has the same sensitivity children are 10 times more sensitive than adults; girls are more sensitive than boys, women have different organ sensitivities than men. Not everyone is in the same risk category a seventy year old man undergoing image-guided prostate treatments is in an entirely different risk situation than a 15 year old undergoing image-guided radiosurgery for an AVM on the spinal cord.

Comparing imaging and therapeutic effective doses In addition to the primary therapeutic dose deposited in the target volume we have Primary concomitant dose from beams transiting normal tissue Secondary concomitant dose from internal and external scatter, linac leakage, etc

Evaluating imaging dose by comparison to therapeutic dose Effective dose is the established way to measure and compare radiation dose in terms of its stochastic risk Imaging dose is delivered in standard formats Therapeutic dose is delivered in highly variable patient-specific scenarios There has not yet been much done to calculate effective doses for therapy Therefore it is not yet feasible to make precise quantitative comparisons of imaging versus therapy dose

Example dose comparison Estimate of 854 msv effective dose from scattered internal beam and external leakage and scatter for a 70 Gy prostate treatment - AA Cigna et al, Dose due to scattered radiation during external radiotherapy: a prostate cancer case history, Radiation Protection Dosimetry 108 (1): 27-32, 2004. Compare to 350 msv from 35 daily diagnostic quality pelvic CT scans for image-guided adaptive radiotherapy

Reducing the imaging dose

Dose reduction In general, for imaging during IGRT one should not assume that diagnostic quality procedures and images are necessary in all applications

Dose reduction Effective dose is a volume integral, so reducing fluence and/or area helps Planar imaging FOV for patient setup can be collimated down to the region of interest CT scan length can be reduced to the region of interest Digital tomosynthesis - CT acquired over a limited angular arc reduces CTDI

Temporal dose reduction Fluoroscopy for target tracking doesn t always require video frame rates (30 frames/second) or a continuous exposure - e.g., respiration monitoring can be done at 2-4 frames per second via pulsed fluoroscopy

Fluence reduction Image registration for patient setup, contour transfer, and dose summing can be fairly insensitive to noise, so the technique can be reduced below diagnostic quality

Reduction by modality change Intra-fraction monitoring via kv fluoroscopy can be replaced by cine-mode MV portal imaging using the treatment beam This can work for 3D conformal therapy, where the beam aperture is fixed for each field Cannot work for dynamic IMRT, where the beam aperture is constantly changing

Summary Imaging systems and procedures for IGRT are often configured differently than for diagnostic exams, resulting in different doses IGRT can combine several different imaging procedures for a particular patient Estimation of the total concomitant dose must recognize the variations in dose deposition by using effective dose as the common measure Risk evaluation in IGRT is fundamentally different than in diagnostic radiology because the imaging dose is added to a high therapeutic dose There is a need for estimates of effective dose for the treatment procedures to enable evaluation of imaging dose in context IGRT procedures do not always require diagnostic quality images and thus allow a variety of dose reduction strategies