Effectiveness of compensating filters in the presence of tissue inhomogeneities

Similar documents
Heterogeneity Corrections in Clinical Trials

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Dosimetric advantage of using 6 MV over 15 MV photons in conformal therapy of lung cancer: Monte Carlo studies in patient geometries

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity correction and the analytic anisotropic algorithm

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Hampton University Proton Therapy Institute

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Tolerance levels of CT number to electron density table for photon beam in radiotherapy treatment planning system

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Unrivaled, End-to-End

Additional Questions for Review 2D & 3D

3D heterogeneous dose distributions for total body irradiation patients

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

ph fax

A new geometric and mechanical verification device for medical LINACs

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Dose differences between the three dose calculation algorithms in Leksell GammaPlan

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Contribution of CT to Quantitative Radiation Therapy Planning

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Eric E. Klein, Ph.D. Chair of TG-142

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Leila E. A. Nichol Royal Surrey County Hospital

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

A comparative dosimetric analysis of the effect of heterogeneity corrections used in three treatment planning algorithms

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

Quality assurance in external radiotherapy

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

SBRT Credentialing: Understanding the Process from Inquiry to Approval

The bootstrap method to improve statistical analysis of dosimetric data for radiotherapy outcomes

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy


Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm.

Dosisverifikation mit Delta 4 Discover während der Behandlung

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Comparison of bulk electron density and voxel-based electron density treatment planning

Anthropomorphic breast phantoms for quality assurance and dose verification

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD

Collapsed Cone Convolution 2D illustration

Diode calibration for dose determination in total body irradiation

Electron therapy Class 3: Clinical procedures

A spiral phantom for IMRT and tomotherapy treatment delivery verification

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

TREATMENT PLANNING: CLINICAL ELECTRONS John A. Antolak, Ph.D. Prepared for the AAPM Therapy Physics Review Course Austin, TX July 19 20, 2014 PURPOSE

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Limits of Precision and Accuracy of Radiation Delivery Systems

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Radiochromic film dosimetry in water phantoms

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

Follow this and additional works at: Part of the Medicine and Health Sciences Commons

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional

Activity report from JCOG physics group

A Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Credentialing for the Use of IGRT in Clinical Trials

A simple technique for craniospinal radiotherapy in the supine position

The Physics of Oesophageal Cancer Radiotherapy

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Method to plan, administer, and verify supine craniospinal irradiation

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Protection of the contralateral breast during radiation therapy for breast cancer

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Review of TG-186 recommendations

5/28/2015. The need for MRI in radiotherapy. Multiparametric MRI reflects a more complete picture of the tumor biology

Commission, Ganakbari, Savar, Dhaka, Bangladesh

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

A translational couch technique for total body irradiation

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

IN VIVO IMAGING Proton Beam Range Verification With PET/CT

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

Biases affecting tumor uptake measurements in FDG-PET

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Accounting for center-of-mass target motion using convolution methods in Monte Carlo-based dose calculations of the lung

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Effectiveness of compensating filters in the presence of tissue inhomogeneities A. Sethi,* L. Leybovich, N. Dogan, and G. Glasgow Department of Radiation Oncology, Loyola University Medical Center, Maywood, Illinois 60153 Received 24 October 2002; accepted for publication 25 April 2003 CT based 3D treatment planning systems 3DTPS can be used to design compensating filters that, in addition to missing tissue compensation, can account for tissue inhomogeneities. The use of computer-driven systems provides a practical, convenient, and accurate method of fabricating compensating filters. In this work, we have evaluated a commercially available PAR Scientific DIGIMILL milling machine linked with FOCUS 3DTPS. Compensating filters were fabricated using refined gypsum material with no additives. Thus, filters were of manageable dimensions and were not sensitive to common machining errors. Compensating filters were evaluated using a homogeneous step phantom and step phantoms containing various internal inhomogeneities air, cork, and bone. The accuracy of two planning algorithms used to design filters was experimentally evaluated. The superposition algorithm was found to produce better agreement with measurements than the Clarkson algorithm. Phantom measurements have demonstrated that compensating filters were able to produce a uniform dose distribution along the compensation plane in the presence of tissue inhomogeneity. However, the dose variation was greatly amplified in planes located beyond the inhomogeneity when a single compensated beam was used. The use of parallel-opposed compensated beams eliminated this problem. Both lateral and depth-dose uniformity was achieved throughout the target volume. 2003 American College of Medical Physics. DOI: 10.1120/1.1582251 PACS number s : 87.53. j, 87.66. a Key words: compensating filters, tissue inhomogeneity, dose uniformity INTRODUCTION Three-dimensional compensating filters are often used to provide corrections for missing tissue and/or tissue inhomogeneities encountered by a photon beam. 1 6 However, with a compensating filter, a uniform dose distribution can be achieved in only a selected compensation plane CP, which is usually positioned at the target center. In planes located above and below the CP, the filter overcompensates and undercompensates respectively, resulting in nonuniform dose distribution in these planes. The degree of dose nonuniformity increases with the increasing distance from the CP. When a photon beam passes through an inhomogeneous medium, the dose nonuniformity is significantly increased. The automated systems, which consist of a 3D treatment planning system 3DTPS linked with a compensator-milling machine, have considerably increased the use of compensating filters. Compared to the older manual methods 7,8 of designing filters, computer-driven systems 9 13 provide improved compensation for surface irregularities better precision and continuous versus discrete shape profile. In addition, the CT based filter design allows for the compensation of tissue inhomogeneities. We have evaluated compensating filters produced by a system consisting of PAR Scientific DIGIMILL S&S Par Scientific, Inc., Brooklyn, NY and FOCUS 3DTPS Computerized Medical Systems, Inc., St. Louis, MO. Filters were designed to compensate for surface irregularities and 209 1526-9914Õ2003Õ4 3 Õ209Õ8Õ$17.00 2003 Am. Coll. Med. Phys. 209

210 Sethi et al.: Effectiveness of compensating filters in the... 210 FIG. 1. Color Cross sectional view of phantoms phantom A on left, phantom B on right used in this work. Film positions are shown as dotted lines. various internal inhomogeneities air, cork, and bone for a range of photon energies. Both Clarkson 14 and superposition 15 algorithms were tested in designing compensating filters and evaluating the resultant dose distribution. All filters were fabricated using refined gypsum material 16 18 without any additives. METHODS AND MATERIALS Compensating filters were evaluated using two specially fabricated acrylic step phantoms. Phantom cross section dimensions were 18 cm 13 cm Fig. 1. The thickness of the phantom was 15 cm. The first phantom (A) had a three-step profile on both sides, whereas the second phantom (B) had a three-step profile on only one side. Each step was 2 cm high and 2 cm deep. Each phantom had a rectangular cavity (5.5 8 cm 2 ) that could be filled with bone ( 1.65 gm/cm 3 ), cork ( 0.3 gm/cm 3 ), or Styrofoam ( 0.01 gm/cm 3 ). Cork was used to model lung tissue and the Styrofoam was used to simulate an air cavity. For measurements with a homogeneous phantom, the cavity was filled with acrylic material ( 1.18 gm/cm 3 ). Phantom A was designed to test compensation for both surface irregularities and internal inhomogeneities. Phantom B, on the other hand, allowed for separately studying the effects of surface irregularities and internal inhomogeneities. The phantoms were irradiated with either a single lateral or parallel-opposed lateral fields. Co60 Theratron-1000, MDS Nordion, Kanata, Ontario, Canada, 6 MV, and 18 MV Varian 2100C, Varian Corporation, Palo Alto, CA photon beams were used in this work. Field size was 16 10 cm 2. Treatment plans with and without compensating filters were developed for a homogeneous phantom cavity filled with acrylic and for phantoms with different types of inhomogeneities. To evaluate dose compensation in the presence of an asymmetrically located inhomogeneity, the cavity was shifted laterally by 1 cm from the phantom center Fig. 1. Compensating filters made of refined gypsum were constructed using an automated fabrication system. Refined gypsum with no additives was selected as the filter material because of its relatively low density ( 2.0 gm/cm 3 ). This density is large enough to limit filter thickness to 5.5 cm within accessory mount clearance and provide compensation for up to 15 cm of missing tissue. At the same time, the density is sufficiently low to make the filter transmission insensitive to common fabrication errors ( 1 mm). In our work with gypsum filters, these fabrication errors resulted in 1% transmission error. Similar fabrication errors, if the filters were made of Lipowitz metal alloy Cerrobend, 9.4 gm/cm 3 ), would have caused 4.4% change in transmission. 9 The location of the compensation plane was varied as follows: center of phantom, and at 4 cm laterally from the phantom center Fig. 1. The beam isocenter was always placed in the compensation plane. This allowed us to evaluate the influence of the location of inhomogeneity on dose compensation when the cavity was located in front of or behind the compensation plane. Each phantom could accommodate both film and ionization chamber measurements Fig. 1. Dose distributions were measured in three sagittal planes: right sagittal RS, center sagittal CS,

211 Sethi et al.: Effectiveness of compensating filters in the... 211 TABLE I. Dose variation in percentage for a right lateral uncompensated beam incident on phantom A. Co60 6MV 18MV Medium RS CS LS RS CS LS RS CS LS Homo. 13.5 28.0 26.0 11.1 19.5 17.8 7.9 11.4 11.7 Bone 16.4 39.0 30.0 11.0 19.8 14.8 7.4 11.7 9.5 Cork 16.0 46.0 58.0 12.8 30.2 38.2 9.0 14.2 26.8 Air 14.9 46.0 67.0 12.4 31.0 45.2 8.4 14.5 29.7 and left sagittal LS. The separation between planes was 4 cm. Radiographic XV2 films Eastman Kodak Co., Rochester, NY were used for the measurement of dose distributions. CMS Dynascan Computerized Medical Systems, Inc., St. Louis, MO film scanning system was used to scan films. Film optical density was converted to dose using appropriate depth-corrected H&D curves. The measured dose distributions were compared with FOCUS TPS calculations. An ionization chamber was positioned in the center sagittal plane CS and was used for absolute point dose verification. Superposition and Clarkson algorithms were used for compensating filter design and dose calculations. Finally, inhomogeneity corrected treatment plans with and without compensating filters were developed for several patients with lung tumors. A four-field AP/PA and parallel opposed oblique beam arrangement was used. This beam arrangement was chosen as it is commonly used to produce uniform dose distributions for plans without heterogeneity corrections. The impact of compensating filters on target dose uniformity was evaluated. RESULTS Dose profiles along the central vertical axis in each sagittal plane were evaluated, since the highest dose gradient is expected along this axis. Table I shows the dose variation for a single right lateral beam with no compensating filter and the cavity filled with different materials. As seen in this table, the dose variation ranged from 7.4% to 67% depending on energy, inhomogeneity, and measurement planes. For comparison, in the case of a flat homogeneous phantom with no surface irregularities left lateral beam incident on phantom B), the dose variation for all beam energies was limited to 2%. In general, the variation in dose decreased with increasing photon energy. Except for the bone inhomogeneity, the smallest dose variation was seen for the homogeneous phantom in all planes for all photon energies. For 6 and 18 MV beams, dose variations in the RS and CS planes were similar for bone and homogeneous phantoms. In the LS plane, the dose variation was 2 3% lower for the case of bone filled cavity. In contrast, for the Co60 beam, the lowest dose variation was seen for the homogeneous phantom in all planes. Dose distributions also showed a significant dependence on the location of the measurement plane. In the planes located beyond the cavity e.g., LS plane, the dose variation increased with decreasing density of cavity material. In the plane passing through the cavity CS plane, the dose variation was similar within 1% for cork and Styrofoam at all beam energies. In the phantom TABLE II. Dose variation in percentage for a right lateral compensated beam incident on phantom A. Co60 6MV 18MV Medium RS CS LS RS CS LS RS CS LS Homo. 5.3 2.3 8.4 3.0 3.4 6.5 12.4 0.4 4.2 Bone 14.0 4.5 7.0 5.4 2.8 5.4 10.2 2.3 2.6 Cork 15.0 4.2 19.3 10.6 5.4 16.5 15.4 4.6 15.5 Air 16.7 5.0 23.2 12.2 4.8 18.7 14.0 6.5 20.0

212 Sethi et al.: Effectiveness of compensating filters in the... 212 FIG. 2. Color Dose profiles in the RS, CS, and LS planes for a right lateral 6 MV compensated beam incident on phantom A with cork cavity. containing cork or Styrofoam, the dose inhomogeneity was significantly higher than that for the bone phantom. In the planes located upstream of the cavity e.g., RS plane the dose variation changed only slightly ( 3%) with the cavity material. Also, for all photon energies and cavity materials, the RS plane dose variation was significantly lower than that in the CS and LS planes. The introduction of a compensating filter significantly improved dose uniformity in the plane of compensation, however large dose nonuniformities were seen in other planes. Table II shows results obtained for a single compensated beam. The plane of compensation was the CS plane. In this plane, the dose uniformity was greatly improved for all cavity materials. In general, the dose variation was lowest for homogeneous phantom ( 3.5%) and was higher for cork and Styrofoam filled phantoms up to 6.5%. With the addition of the filter, the dose variation in planes beyond the cavity LS plane was significantly reduced for all energies and cavity materials. However, large dose nonuniformities still remained in both RS and LS planes. Figure 2 compares dose profiles in the three sagittal planes for a single compensated 6MV beam. The cavity material is cork. In addition to providing a uniform dose in the CS plane, the compensator reduced dose variation in the LS plane by more than twice. In the RS plane, however, little change in dose profile was observed. Figure 3 compares calculated and measured dose profiles in the RS plane for a parallel-opposed 6 MV compensated beams. Calculations using the superposition algorithm more closely agreed with the measured data than Clarkson calculations. The deviation between measured and calculated profiles was up to 7% for Clarkson algorithm and up to 2% for the superposition algorithm. Furthermore, at points corresponding to cavity interface and abrupt shape changes in the phantom, better than 1% agreement was seen between superposition calculations and the measured profile. Therefore a good overall agreement between calculated superposition and measured profile shapes was observed. Similar results not shown were obtained for other measurement planes and beam energies. Thus, all calculations reported in this work were performed using the superposition algorithm. FIG. 3. Color A comparison between measured and calculated dose profiles using Clarkson and superposition algorithms in the RS plane for a pair of compensated parallel-opposed 6 MV beams incident on phantom A.

213 Sethi et al.: Effectiveness of compensating filters in the... 213 TABLE III. Dose variation in percentage for a pair of parallel-opposed compensated beams incident on phantom A. Co60 6MV 18MV Medium RS CS LS RS CS LS RS CS LS Homo. 2.3 2.5 2.3 1.4 2.5 1.4 5.0 0.7 5.0 Bone 4.0 3.0 4.0 3.1 2.2 4.0 3.7 1.4 4.0 Cork 2.0 2.1 2.0 4.2 4.0 3.5 5.0 2.7 5.0 Air 2.0 2.8 2.0 2.0 2.5 4.0 4.3 3.5 5.0 When parallel-opposed compensated beams were used, the dose variation in all planes decreased to 5% for all photon energies and cavity materials Table III. Figure 4 illustrates the dose variation in all three sagittal planes for parallel-opposed compensated 6 MV beams incident on phantom A with cork inhomogeneity. The isocenter was located in the CS plane. In this case, the dose was uniform to within 4.5% in all planes. Results obtained for parallel-opposed lateral beams incident on the asymmetrically shaped phantom B filled with cork are shown in Table IV. In this case, quite adequate dose uniformity was obtained with only one compensating filter, provided the compensated beam faced the irregular surface and the plane of compensation was positioned beyond the inhomogeneity. This is illustrated in Table IV row 5. Here only the right lateral field was compensated and the plane of compensation was the LS plane Fig. 1. When the superposition algorithm was used, the absolute dose measured in the CS plane with an ionization chamber was in good agreement within 1.5% with calculations. Figure 5 compares target dose profiles for a lung patient planned with and without compensators. All treatment plans included tissue inhomogeneity corrections. When no compensating filters were used, the dose in the target varied from 15% to 5%. The addition of compensating filters resulted in reduction of dose variation to less than 5%. Based on treatment plans with inhomogeneity corrections, the mean lung dose was underestimated by 6% in the plans without corrections. When compensating filters were added, the mean and maximum lung doses decreased by 5% and became practically equal to the doses obtained with treatment plans developed assuming the patient to be a homogeneous medium. DISCUSSION Historically, compensating filters have been used as missing tissue compensators to correct for surface irregularities. However, the CT based 3DTPS are also capable of correcting dose distributions for inhomogeneities. Therefore, compensating filters may be designed to compensate for both patient surface irregularities and internal inhomogeneities. The increased use of compensating filters is facilitated by readily available computer-driven compensator fabrication systems linked with 3DTPS. In our department, a system consisting of PAR scientific milling machine and FOCUS TPS is being used. FIG. 4. Color Calculated dose profiles superposition algorithm in the RS, CS, and LS planes for parallel-opposed compensated 6 MV beam arrangement incident on phantom A.

214 Sethi et al.: Effectiveness of compensating filters in the... 214 TABLE IV. Dose variation in percentage for a pair of parallel-opposed 6 MV beams incident on phantom B with cork inhomogeneity. Beam configuration Isocenter plane 6MV RS CS LS RLAT with CF; LLAT without CF CS 4.2 5.3 4.8 Parallel-opposed compensated CS 5.2 1.4 4.5 RLAT with CF; LLAT without CF RS 7.0 11.8 10.6 Parallel-opposed compensated RS 1.5 2.5 2.5 RLAT with CF; LLAT without CF LS 3.9 1.0 2.0 Parallel-opposed compensated LS 4.2 1.0 2.1 In a 3DTPS, the choice of dose algorithm dictates the accuracy of calculated dose distribution in the presence of tissue inhomogeneities. Compared to Clarkson calculations, the superposition algorithm has been shown to provide better agreement with measured data, especially in the interface region. 19 This can be explained by the ability of the superposition algorithm to more accurately model changes in scatter radiation due to tissue inhomogeneities and surface irregularities. In the superposition algorithm, which is based on collapsed cone dose calculation, 20 the energy deposition kernels are distorted to account for variations in electron density in each voxel. Our results confirmed that compared to Clarkson algorithm, the superposition algorithm indeed produced better agreement with measured dose profiles. The dose variation reduced from 7% Clarkson to 2% superposition. The agreement between measured and calculated doses was especially improved in the regions corresponding to sharp changes in phantom surface and tissueinhomogeneity interface. The selection of refined gypsum mixture with no additives for fabrication of compensating filters was based on its density (2.0 gm/cm 3 ). This density represents a good compromise between competing requirements for an ideal filter material: the density should be large enough to provide a compensator of manageable dimensions, and yet it should be small enough that common errors in filter fabrication ( 1 mm) do not produce measurable errors in dose distribution. A parallel-opposed beam configuration is commonly used to compensate for dose variation with depth. However, the presence of surface irregularities and/or tissue inhomogeneities creates dose variations in planes perpendicular to the beam. When a compensating filter is added to the beam, a uniform dose distribution within 2 3% is achieved in only one plane the plane of compensation, CP. Overcompensation reduced dose and undercompensation increased dose occur in planes above and below the CP respectively. When an inhomogeneity is present, these dose variations are amplified. Indeed, when an inhomogeneity extended beyond the compensation plane CP, the dose variation in planes beyond CP increased by a factor of 2 3 compared to homogeneous medium. The dose variation was more pronounced at lower photon energies and for low-density materials cork and Styrofoam, which represent lung and air filled cavities in patients. FIG. 5. Color Target dose profiles with inhomogeneity corrections for a lung patient treated with 6 MV four-field arrangement.

215 Sethi et al.: Effectiveness of compensating filters in the... 215 When a pair of compensated parallel-opposed beams is used, regions of overcompensation due to one beam may overlap with regions of undercompensation from the opposing beam. Thus, the use of compensators in parallel-opposed beams should produce lateral dose uniformity in addition to depth dose uniformity. Indeed as our measurements have demonstrated, the degree of dose nonuniformity in planes located 4 cm away from CP was reduced from as high as 23% to 2% when compensators were introduced in Co60 parallel-opposed beams. Similar effect was observed with 6 MV and 18 MV beams. There are several publications concerning the use of compensating filters. 9 13 However, in most publications, the impact of inhomogeneities was not considered at all. In others, only cork and Styrofoam were studied. In many publications, only one energy usually 6 MV photons was considered and no comparison between different algorithms was carried out. Furthermore, dose homogeneity was studied only in the compensation plane. In addition, most of the compensators considered in literature were non-gypsum based. In our paper, a comprehensive study that considered a large range of inhomogeneities and beam energies was carried out. Also, the ability to provide volumetric dose compensation was evaluated. Our study also compared the accuracy of superposition and Clarkson algorithms in designing compensators. The choice of gypsum material in our work was based on a number of factors, e.g., simple compensator fabrication, reproducible and uniform density filters, cost-effectiveness cheapest material available for compensating filter, and suitable density for most clinical applications. However, the relatively low density of gypsum precludes its use in the IMRT. CONCLUSIONS Compensating filters may be effectively used for restoring dose uniformity in the presence of surface irregularities and internal inhomogeneities. Computer-driven systems provide a practical, convenient, and accurate method for compensating filter fabrication. The use of refined gypsum results in manageable size filters that are not sensitive to commonly encountered machining errors. The introduction of compensating filters in parallel-opposed beams provides lateral dose compensation throughout the target volume. The accuracy of algorithms available in a 3DTPS for designing compensating filters should be experimentally evaluated. *Email address: asethi@lumc.edu 1 F. Ellis, E. J. Hall, and R. Oliver, A compensator for variation in tissue thickness for high energy beams, Br. J. Radiol. 32, 421 1959. 2 E. J. Hall and R. Oliver, The use of standard isodose distributions with high energy radiation beams the accuracy of a compensator technique in correcting for body contours, Br. J. Radiol. 34, 43 1961. 3 J. Van de Geijn, The construction of individualized intensity modifying filters in cobalt 60 teletherapy, Br. J. Radiol. 38, 865 1965. 4 A. L. Boyer, Design of compensating filter, in Advances in Radiation Therapy Treatment Planning, edited by A. E. Wright and A. L. Boyer American Institute of Physics, New York, 1983, pp. 185 220. 5 D. M. Robinson and J. W. Scrimger, Megavoltage photon beam dose reduction with retracted tissue compensators, Phys. Med. Biol. 32, 1031 1987. 6 L. F. Reinstein, New approaches to tissue compensation in radiation oncology, in Advances in Radiation Oncology Physics, edited by J. A. Purdy American Institute of Physics, New York, 1992, pp. 535 572. 7 W. R. Hendee and C. E. Garciga, Tissue compensating filters for Cobalt 60 teletherapy, Am. J. Roentgenol., Radium Ther. Nucl. Med. 99, 939 1967. 8 F. M. Khan, V. C. Moore, and D. J. Burns, An apparatus for the construction of irregular surface compensators for use in radiotherapy, Radiology 90, 593 1968. 9 P. A. Jursinic, M. B. Podgorsak, and B. R. Paliwal, Implementation of a three-dimensional compensation system based on computed tomography generated surface contours and tissue inhomogeneities, Med. Phys. 21, 357 1994. 10 A. S. Beddar, C. Thomason, and P. M. K. Leung, Description and evaluation of a new 3-D computerized treatment planning dose compensator system, Med. Dosim 19, 227 1994. 11 P. S. Basran, W. Ansbacher, G. C. Field, and B. R. Murray, Evaluation of optimized compensators on a 3D planning system, Med. Phys. 25, 1837 1998. 12 G. S. Mageras, R. Mohan, C. Burman, G. D. Barest, and G. J. Kutcher, Compensators for three-dimensional treatment planning, Med. Phys. 18, 133 1991.

216 Sethi et al.: Effectiveness of compensating filters in the... 216 13 W. D. Renner, T. P. O Connor, and N. M. Bermudez, An algorithm for design of beam compensators, Int. J. Radiat. Oncol., Biol., Phys. 17, 227 1989. 14 J. R. Cunningham, P. N. Shrivastava, and J. M. Wilkinson, Program IRREG-calculation of dose from irregularly shaped radiation beams, Comput. Programs Biomed. 2, 192 1972. 15 T. R. Mackie, J. W. Scrimger, and J. J. Battista, A convolution method of calculating dose for 15MV x-rays, Med. Phys. 12, 188 1985. 16 K. J. Weeks, B. A. Frass, and K. M. Hutchins, Gypsum mixtures for compensator construction, Med. Phys. 15, 410 1988. 17 V. R. Arora and K. J. Weeks, Characterization of gypsum attenuators for radiotherapy dose modification, Med. Phys. 21, 77 1994. 18 G. P. Glasgow, N. Dogan, A. Sethi, and J. Stanton, Refined gypsum effective attenuation coefficients for Co-60, 6-, 10-, and 18-MV X-rays, Phys. Med. Biol. 48, 63 2003. 19 M. Miften, M. Wiesmeyer, A. Kapur, and C.-M. Charlie Ma, Comparison of RTP dose distributions in heterogeneous phantom with BEAM Monte Carlo simulation system, J. Appl. Clin. Med. Phys. 2, 21 2001. 20 A. Ahnesjo, Collapsed-cone convolution of radiant energy for photon dose calculations in heterogeneous media, Med. Phys. 16, 577 1989.