Integra. Spider and Mini Spider Limited Wrist Fusion System SURGICAL TECHNIQUE

Similar documents
Surgical Technique Carpal Fusion

The Wrist Fusion Set. Stainless Steel and Titanium TECHNIQUE GUIDE. Instruments and implants approved by the AO Foundation

2.4 mm Variable Angle LCP Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Wrist Fusion Instrument and Implant Set.

Surgical Technique. DISCLOSURE: This device is not approved for sale in the U.S.A. Customer Service:

VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Clover Staple. Surgical Technique

Distal Radius Plate Instrument and Implant Set. Discontinued December 2017 DSUS/TRM/0916/1063(1)

AcUMEDr. MoDULAR HAND SYSTEM

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

WristMotion Wrist Hemiarthroplasty System Instructions for Use

Acutrak 2 Headless Compression Screw System Micro, Mini, and Standard Screws. Supplemental Use Guide Four Corner Fusion

ReMotion Total Wrist System

Surgical Technique. SLIC Screw System

Integra. ADVANSYS Plating System SURGICAL TECHNIQUE

Integra. PyroCarbon Lunate SUTURE ANCHOR SURGICAL TECHNIQUE

Integra. Freedom Wrist Arthroplasty System SURGICAL TECHNIQUE

Acu-Loc Wrist Spanning Plate System. Surgical Technique

Wrist Arthritis & Partial Wrist Fusion

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

2.4 mm Variable Angle Locking Intercarpal Fusion System

Chapter 19. Arthroscopic Bone Grafting for Scaphoid Nonunion. Introduction. Operative Technique. Radiocarpal and Midcarpal Exploration

Maestro. Wrist Radial Fracture System. Surgical Technique. Knees Hips Extremities Cement and Accessories PMI Trauma Technology

Distal Radius Plate 2.4/2.7 dorsal and volar

Technique Guide Wrist Hemiarthroplasty System

TABLE OF CONTENTS. Limited intercarpal joint arthrodesis (including isolated capitolunate, two-column and four-corner fusions)

Integra. SafeGuard Mini Carpal Tunnel Release System SURGICAL TECHNIQUE

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Integra. Silicone PIP Implant SURGICAL TECHNIQUE

Surgical Technique. Forearm Fracture Solutions

TIBIAXYS ANKLE FUSION

Surgical Technique Guide

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Integra. PyroSphere System SURGICAL TECHNIQUE

3. Ulno lunate, Ulno triquetral ligament. Poirier: Between RSC &LRL. 5. Dorsal intercarpal ligament

Surgical Technique Guide

MICRONAIL. Intramedullary Distal Radius System SURGICAL TECHNIQUE

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

Locking Radial Head Plates

Integra. Katalyst Bipolar Radial Head System SURGICAL TECHNIQUE

Acu-Loc Wrist Plating System. Surgical Technique

WINSTA-R. Distal Radius System

Visualize, stabilize, mobilize. Wristore * Distal Radius Fracture Fixator Abbreviated Surgical Technique

JuggerKnot Soft Anchor 1.0 mm Mini. Scapholunate Ligament Repair/Reconstruction. Brochure and Surgical Technique

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

Arthrodesis System 2.0/2.3, 2.5

Introduction. The wrist contains eight small carpal bones, which as a group act as a flexible spacer between the forearm and hand.

The Flower Four Corner Fusion Plate

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Low Bend Distal Tibia Plates

Modular Ulnar Head surgical technique. Transforming Extremities

SPEED TM Hand & Wrist System. Procedure Manual

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

Zimmer Small Fragment Universal Locking System. Surgical Technique

DYNAMIC, TRANSVERSE COMPRESSION. Low Profile, Anatomic Design, Type II Anodized. Mechanical Compression Designed to Stimulate the Fusion Process

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

VariAx TM Distal Radius Locking Plate System

LCP Distal Humerus Plates

Silicone PIP, MCP & MCP-X (PreFlex)

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005)

PediLoc 3.5mm and 4.5mm Bowed Femur Plate Surgical Technique

Technique Guide. 6.5 mm Midfoot Fusion Bolt. For intramedullary fixation of the medial column of the foot.

upper extremity fixation system

Olecranon Locking Plate II

Integra. Silicone MCP Joint Prosthesis SURGICAL TECHNIQUE

CANNULINK. Intraossous Fixation System SURGICAL TECHNIQUE

SURGICAL TECHNIQUE GUIDE PROTEAN. radial head plate. As described by: Jorge L. Orbay, M.D. Miami Hand & Upper Extremity Institute Miami, Florida

Surgical Technique. Proximal Humerus Locking Plate

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

Wrist Fusion System SURGICAL TECHNIQUE

Instrument and Implant for wrist fracture

AcUMEDr. FoREARM ROD SYSTEM

Percutaneous Scaphoid Fixation: A Volar Approach

SURGICAL TECHNIQUE GUIDE PROTEAN. radial head plate. As described by: Jorge L. Orbay, M.D. Miami Hand & Upper Extremity Institute Miami, Florida

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

Distal Radius and Distal Ulna Plates System Self-Tapping Spherical Locking Screw Self-Tapping Conical Locking Screw Cortex Screw

A New Benchmark in Tibial Fixation

Carpal rows injuries!

Hoffmann II Compact External Fixation System. Modular System for

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Ascension. Silicone MCP surgical technique. surgical technique Ascension Silicone MCP

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Surgical Technique 4.5/8.5MM BEAMING SYSTEM. Customer Service:

Bipolar Radial Head System

Integra. DigiFuse Cannulated Intramedullary Fusion System SURGICAL TECHNIQUE

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Acu-Loc 2 Wrist Plating System

Surgical Technique. Wrist Plating System

The Flower Medial Column Fusion Plate

3.5 mm LCP Olecranon Plates

Locking Ankle Plating System. Surgical Technique

PediLoc Tibia Surgical Technique

Surgical Technique. Olecranon Locking Plate

Transcription:

Integra Spider and Mini Spider Limited Wrist Fusion System SURGICAL TECHNIQUE

Table of contents Description... 02 Indications... 02 Contraindications... 02 Surgical Technique... 03 Spider Introduction-Four Corner Fusion... 03 Step 1: Skin Incision and Exposure... 03 Step 2: Reduction/Provisional K-wire Fixation... 04 Step 3: Rasping... 05 Step 4: Denude the Cartilage... 05 Step 5: Bone Grafting... 05 Step 6: Application of Spider Plate... 06 Step 7: Compression of the Four Bones... 07 Step 8: Postoperative Care... 07 Mini Spider Introduction-STT Fusion... 08 Step 1: Skin Incision and Exposure... 08 Step 2: Provisional K-wire Fixation... 09 Step 3: Rasping... 09 Step 4: Denude the Cartilage... 10 Step 5: Bone Grafting... 10 Step 6: Application of Mini Spider Plate...10 Step 7: Postoperative Care...11 Instrument Set... 12 Ordering Information... Back Cover

Description The Spider and Mini Spider Limited Wrist Fusion System consist of two stainless steel plates, featuring an 8-hole and a 6-hole design. Both plates have a unique three-dimensional design optimized for four-corner and other limited wrist fusion procedures. The Spider plates sit just below the dorsal aspect of the carpus and can be utilized in various locations in the wrist. Indications Osteoarthritis Rheumatoid arthritis Post-traumatic or degenerative wrist arthritis Complex fractures of the wrist Revision of failed partial wrist fusion Carpal instability Contraindictions Use of this product is contraindicated in the presence of any of the following: Severe tendon, neurological or vascular deficiencies that may compromise the affected extremity Any concomitant disease that may compromise the function of the plate Infection See package insert for full prescribing information 2

Surgical Technique As the manufacturer of this device, Integra LifeSciences Corporation does not practice medicine and does not recommend this or any other surgical technique for use on a specific patient. The surgeon who performs any procedure is responsible for determining and using the appropriate technique in each patient. Caution: Federal law restricts this device to sale by or on the order of a physician or practitioner. Spider Introduction: Four Corner Fusion The Spider Limited Wrist Fusion Plate (8-holes) can be utilized in various locations throughout the wrist depending on the procedure being performed. The plate has been used successfully for: Luno-triquetro-capito-hamate (four corner) fusion Scaphocapitate (SC) fusion Radioscapholunate (RSL) fusion Scapholunacapitate (SLC) fusion Scaphotrapezium trapezoid (STT) fusion Preparation of the site to be fused and the specific techniques are fairly similar regardless of where the fusion is desired. The larger Spider plate s design is optimized for four-corner limited arthrodesis. The plate can also be utilized for other limited wrist fusion procedures for which the plate is appropriately sized. Step 1 Skin Incision and Exposure 1-1 A longitudinal incision is made centered over the dorsal wrist or a transverse incision can be utilized (Figure 1-1). Dissection is carried down through the subcutaneous tissues taking care to protect the sensory branch of the radial and ulnar nerve fibers. 1-1 Alternatively, a separate longitudinal incision can be made between the fourth and fifth compartments, or an anatomical dorsal ligament-sparing approach can be used for direct exposure to the four-bone region. These techniques are also utilized in patients who undergo a four-corner fusion for stability alone where the scaphoid does not require excision. 3

1-2 Exposure for SLAC Deformities That Require Scaphoid Excision Exposure is undertaken through the third dorsal compartment transposing the extensor pollicis longus tendon radially. The incision is taken from the third compartment, between the second and fourth compartments, distally through the capsule exposing the scaphoid. The scaphoid is generally removed using a rongeur, taking care to protect the radioscaphocapitate ligaments. A radial styloidectomy may be indicated. Exposure of the four-bone region encompassing the lunate, triquetrum, capitate, and hamate can be undertaken through this incision (Figure 1-2). 1-2 Step 2 Reduction/ Provisional K-wire Fixation 2-1 After appropriately exposing the capitate, hamate, lunate, and triquetrum, any rotational instability is reduced and temporary percutaneous K-wire fixation is accomplished, make sure to keep the 1.1 mm (.045 ) K-wires as volar as possible within the four bones, it is especially important to ensure complete reduction of the lunate (Figure 2-1). 2-1 Reduction of Lunate in Visi Deformity Reduce lunate into a neutral position. 1.6 mm (.062 ) K-wire (Palmar - flexed) Dorsal Side Volar Side Lateral Views 2-2 Provisional fixation in the volar half of the carpal bones being fused tends to produce less toggle in the carpus while rasping is performed. This provides a more uniform, recessed defect to accept the plate and bone graft (Figure 2-2). 2-2 Provisional Fixation Place provisional fixation with K-wire from the ulnar carpus to transfix the bones to be fused. The K-wire should be placed into the volar 1/2 of the carpus to avoid impingement with the rasp for the Spider plate. K-wires Step 3 Rasping Ulnar Side Dorsal View Radial Side Dorsal Side Volar Side Lateral View (Rasp dorsally,fixation is volar.) 3-1 3-1a The Spider rasp is centered at the four-corner junction and used to rasp down, at least flush with, or even below the bone surface of the dorsal aspect of the carpus (Figure 3-1 and 3-1a). 3-1 The defect from the Spider rasp exactly complements the size, shape and contour of the Spider Plate. 3-1a Rasping can be done manually or with a power attachment. 4

Step 4 Denude the Cartilage 4-1 A small rongeur is used to denude the cartilage between the four bones down through subchondral bone to good cancellous bone (Figure 4-1). 4-1 Step 5 Bone Grafting 5-1 Autogenous cancellous bone graft is taken from either the Lister s tubercle of the distal radius, or from the iliac crest less commonly, and then packed between each of the joint surfaces and the junction of the four-corner fusion at the bottom of the rasped defect (Figure 5-1). 5-1 Note: Excised Scaphoid is not recommended for grafting. 5

Step 6 Application of the Spider Plate 6-1 6-2 After appropriate packing of the bone graft, the Spider plate is then aligned to allow maximum screw placement in each of the four bones (Figure 6-1). Two screws can be placed within each of the four bones. Using the Spider plate holder/drill guide instrument, ensure that the plate is at or below the level of the remaining carpus (recessed). Confirm proper alignment and ensure adequate recession of the plate, utilizing gross inspection and fluoroscopic imaging. Holding the plate in appropriate alignment and using the drill guide, a 1.5mm drill bit is used to place an index screw. Place this screw into the lunate. Utilize the drill guide to perform the first drilling of the carpal bone. This allows alignment of the plate and appropriate centralization of the drill through one of the holes in the Spider plate (Figure 6-2). Do not tighten the first screw placed into the lunate; simply post the screw. Drill a second hole for screw placement in a diametric orientation to the lunate screw to balance the plate upon initial fixation. The remaining holes in the Spider plate can be drilled utilizing the drill guide to allow several degrees of freedom in the placement of the screws themselves. Initially, place one screw in each of the four bones, utilizing the 8mm, 10mm, 12mm and 14mm length, with the 2.4mm diameter cancellous self-tapping screws to fix the plate down. 2.8mm cancellous self-tapping screws can be utilized as salvage screws through the plate should problems occur with the 2.4mm screws that involve lack of purchase in the carpus. 6-1 6-2 6

Step 7 Compression of the Four Bones 7-1 7-2 Screws should then be tightened in a sequential fashion, compressing together the four bones (Figure 7-1). Each screw should be of appropriate length with care given to avoid protrusion to the pisotriquetral joint (Figure 7-1 and 7-2). Intraoperative fluoroscopy is undertaken to ensure appropriate placement of the plate and screws. Range of motion testing is undertaken to ensure excellent stability and lack of impingement of the fusion plate. Additional bone graft can be packed within the center portion of the four-corner region through the plate itself without difficulty. 7-1 Step 8 Postoperative Care Irrigation of the wound and sequential repair of the capsule and retinacular structures are undertaken. After skin closure, a short-arm splint is placed allowing early active finger range of motion. 7-2 After the sutures have been removed at approximately one week, a removable splint can be placed to allow early active range of motion exercises or a short-arm cast can be placed for three to four weeks of protection. Radiographs should be taken on a sequential basis to ensure appropriate fusion of the four-corner region prior to allowing return to normal activities. 7

Mini Spider Introduction: Scaphotrapezium Trapezoid (STT) Fusion The Mini Spider Limited Wrist Fusion Plate (6-holes) can be utilized in various locations throughout the wrist for specific fusion procedures. The plate has been used successfully for: Scaphotrapezium-trapezoid (STT) fusion Scaphocapitate (SC) fusion Luno-triquetro-capito-hamate (4 corner) fusion Preparation of the site to be fused and the specific techniques are fairly similar regardless of where the fusion is desired. The Mini Spider plate has been optimized for the scaphotrapezium-trapezoid arthrodesis. However, the plate can be utilized for other limited wrist fusion procedures for which the plate is appropriately sized. Step 1 Skin Incision and Exposure 1-1 A dorsal longitudinal incision is made over the wrist in line with the scaphotrapezium-trapezoid joint, extending proximally in line with 1-1 the Extensor Pollicis Longus to Lister s tubercle (Figure 1-1). The extensor retinaculum is incised and released distal to Lister s tubercle. Alternately, a transverse incision can be utilized for joint exposure. A radial styloidectomy may be indicated. Scaphoid Reduction If the scaphoid is subluxed, reduction to a normal anatomic position is facilitated by introducing an instrument such as a hemostat under the distal pole of the scaphoid in line with the radioscaphocapitate ligament. Using the hemostat, derotate the scaphoid, bringing the proximal articulation of the scaphoid into a normal anatomic alignment with the radial fossa. Provisional pinning is utilized to maintain reduction throughout the procedure. 8

Step 2 Provisional K-wire Fixation 2-1 The provisional K-wire fixation maintains the reduction, provides a resistance to the rasp, prevents toggling of the bones, and produces a more uniform defect for the plate. Two 1.1mm (.045 ) K-wires are inserted parallel through the scaphoid into the capitate to fixate the scaphoid. K-wires are positioned to avoid the rasped defect and they should not engage the radiocarpal joint (Figure 2-1). Additional K-wires may be inserted, if necessary, fixating the trapezium to the trapezoid and/or the trapezoid to the scaphoid. 2-1 Step 3 Rasping 3-1 The Mini Spider rasp is centered over the STT joint and used to rasp down, at least flush with or even below the bone surface of the dorsal aspect of the carpus. The defect from the Mini Spider rasp exactly matches the size, shape and contour of the Mini Spider plate. Rasping can be done manually or with a power attachment (Figure 3-1). 3-1 9

Step 4 Denude the Cartilage 4-1 A small rongeur is used to denude the cartilage between the three bones down through subchondral bone to good cancellous bone. To achieve optimal cancellous bone contact for fusion, denuding of the cartilage is taken down to the volar aspect of the STT joint. Alternately, prior to denuding of the interstices, pilot holes are drilled into each of the STT bones utilizing the Mini Spider plate and one screw into each of the bones. This facilitates anatomic alignment prior to denuding the interstices and packing of the bone graft (Figure 4-1). 4-1 Step 5 Bone Grafting 5-1 Autogenous cancellous bone graft is taken from Lister s tubercle or from the iliac crest. The graft is packed between each of the joint surfaces and the junction of the STT fusion down through the volar aspect of the rasped defect. It is important that no bone graft has extruded into the scaphocapitate joint space (Figure 5-1). 5-1 Step 6 Application of the Mini Spider Plate 6-1 Place the Mini Spider plate into the rasped defect. Ensure that the plate is at least flush or recessed relative to the level of the remaining dorsal carpal cortex. Confirm proper alignment and adequate recession of the plate utilizing fluoroscopy (Figure 6-1). 6-1 10

Step 6 (continued) Application of Mini Spider Plate 6-2 Using the Mini Spider plate holder/drill guide instrument, rotate the plate for optimal alignment, allowing for two screws in each of the STT bones (Figure 6-2). 6-2 6-2a For hole placement, use either end of the plate holder/drill guide (Figure 6-2 and 6-2a). Anatomic considerations may permit only one screw in the trapezoid. Proper screw depth and placement should not compromise the 1st CMC joint, or the capitate articulation. Tighten all the screws in a balanced fashion. Intraoperative fluoroscopy is undertaken to ensure appropriate placement of the plate and screws. Range of motion testing is also performed. 6-3 Additional bone graft can be packed into the center portion of the STT fusion through the plate (Figure 6-3). 6-3 Step 7 Postoperative Care Irrigation of the wound is undertaken and sequential repair of the capsule is performed. The distal aspect of the extensor retinaculum may be transposed under the EPL and utilized as a flap at the distal radius. After skin closure, a short-arm splint is placed allowing early active finger range of motion. After the sutures have been removed at approximately 10-14 days, either a removable splint can be used to allow early range of motion exercises, or a short-arm cast can be placed for three to four weeks of protection. Radiographs should be taken on a sequential basis to ensure appropriate fusion of the STT region prior to allowing return to normal activities. 11

Spider and Mini Spider Limited Wrist Fusion System Instrument Set P O N M L K J A B C D E F G H I Associated Instruments A - Spider Rasp Brush B - Spider Rasp (Disposable) C - T-Handle D - Mini Spider Rasp (Disposable) E - Spider Plate Holder/Drill Guide F - Mini Spider Plate Holder/Drill Guide G - Drill Guide H - 2.5mm Hex Driver I - Drill Bits (Disposable) J - Bone Screws K - Mini Spider Plate L - Spider Plate M - Screw Holding Forceps N - Quick Couple Driver O - Depth Gauge P - 3.5mm Tap 12

Integra Spider and Mini Spider - Limited Wrist Fusion System Implants Catalog Number Description 07-0005 Spider Limited Fusion Plate 07-0006 Mini Spider Limited Fusion Plate Implants 2.4mm Bone Screws Catalog Number Description 07-2408 2.4mm Bone Screw 8mm 07-2410 2.4mm Bone Screw 10mm 07-2412 2.4mm Bone Screw 12mm 07-2414 2.4mm Bone Screw 14mm Implants 2.8mm Bone Screws Catalog Number Description 07-2810 2.8mm Bone Screw 10mm 07-2814 2.8mm Bone Screw 14mm 07-2818 2.8mm Bone Screw 18mm Instruments Catalog Number Description 60-0704 2.0mm Drill Bit Solid 60-0709 1.5mm Drill Bit Solid 60-0715 Spider Plate Holder 60-0716 Cleaning Brush 60-0718 T-Handle Quick Couple 60-0719 Mini Spider Plate Holder 60-0720 Spider Rasp 60-0721 Mini Spider Rasp 60-0722 1.5/2.0mm Drill Guide 60-0724 2.5mm Hex Driver AO Couple 60-0730 AO Quick Coupling Handle 60-0740 Tap 3.5mm Tray Catalog Number Description 07-5020 Sterilization Case/Caddy WFS 07-5000 Instrument Set Component Materials 8-Hole and 6-Hole Plate: Stainless Steel Screws: Stainless Steel For more information or to place an order, please contact: Integra n 311 Enterprise Drive, Plainsboro, NJ 08536 877-444-1122 USA n 609-936-5400 outside USA n 866-800-7742 fax integralife.com Spider, Mini Spider, Integra and the Integra logo are trademarks of Integra LifeSciences Corporation. 2011 Integra LifeSciences Corporation. All rights reserved. Printed in the USA 1K ER4820-08/11