Outline. QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences. Frame Rate Limitation. US Imaging Range Equation.

Similar documents
Hands-On Ultrasound Physics and Quality Control Workshop. Outline

Routine Quality Assurance Cookbook

Debbie Childs RDMS, RVT Sonographer Murphy Medical Center Murphy, NC

8/3/2016. Diagnostic Ultrasound Imaging Quality Assurance. Purpose. Information on US QA. James A. Zagzebski 1, Ph.D. Zheng Feng Lu 2, Ph.D.

Ultrasound Accreditation Program Requirements

Current Ultrasound Quality Control Recommendations and Techniques

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRAINING PHANTOMS

Ultrasound Accreditation and Regulatory Compliance from A to Z

TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

Ultrasound Physics & Terminology

Medical Physics 2.0. Ultrasound 2.0. N. J. Hangiandreou, Ph.D. 1 P. L. Carson, Ph.D. 2 Z. F. Lu, Ph.D. 3. Mayo Clinic 2. University of Michigan 3

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Underwater Acoustic Measurements in Megahertz Frequency Range.

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Descriptions of NDT Projects Fall 2004 October 31, 2004

Quality Control Program Of Real Time Medical Ultrasound Machines In Sudan

AUStrian Test kit. Type : Basic. developed by

Joon-Il Choi, MD, Pyo Nyun Kim, MD, Woo Kyoung Jeong, MD, Hyun Cheol Kim, MD, Dal Mo Yang, MD, Sang Hoon Cha, MD, Jae-Joon Chung, MD

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

Quality Assurance of Ultrasound Imagers: Procedures, Expectations, and Philosophies

Optimizer 1425A. User s Guide. Ultrasound Image Analyzer for Doppler and Grey Scale Scanner. België / Belgique. Nederland.

Hands-on. Physics Workshop. 2-day Workshop June 1-2, 2018 Ann Arbor, MI. Program Director: Brian Fowlkes, PhD

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Physical Principles of Ultrasound

Introduction. The goal of TRUS QA is to ensure your system can do all of this accurately.

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

A new method of sonograph lateral resolution measurement using PSF analysis of received signal

CT Quality Control Manual FAQs

Breast Tomosynthesis. What is breast tomosynthesis?

Pipeline Technology Conference 2007

Example 1: How would you evaluate this transducer? What is the most common problem found in ultrasound QC testing? Uniformity is Subjective

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

Automatic quality control processing for detection of element/channel dropout in ultrasound transducers

Ultrasound Physics & Doppler

Preamble (disclaimer)

Medical Diagnostic Imaging

Ultrasound in Medicine

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions.

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

A quality control program for MR-guided focused ultrasound ablation therapy

ACR MRI Accreditation: Medical Physicist Role in the Application Process

Ultrasound Physics and Knobology Alan Macfarlane. Consultant Anaesthetist Glasgow Royal Infirmary

CONTENTS. Test Number cpd Tanya Reynolds (Nat. Dip. Diag. Rad., B. Tech. Diag. Rad., B. Tech. Ultrasound)

Employer s Unit of Competence Ultrasonic testing of materials, products and plant

DIGITAL IMAGE PROCESSING IN ULTRASOUND IMAGES

Ultrasonic Testing Level I:

Mammography Quality Control: A Refresher

Introduction to Biomedical Imaging

Real Time 2D Ultrasound Camera Imaging: A Higher Resolution Option to Phased Array Bob Lasser Randy Scheib Imperium, Inc.

ULTRASOUND. OB/Gyn (Core) Ultrasound PIEZOELECTRIC EFFECT. Principles of Ultrasound Physics and Instrumentation. Nathan Pinkney, BS, CDOS

Ultrasound Principles cycle Frequency Wavelength Period Velocity

CLINICAL PRIVILEGE WHITE PAPER

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals

AAPM Annual Meeting. ACR Accreditation Update in CT

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

w. D. Jolly, F. A. Bruton, and C. Fedor

High incidence of defective ultrasound transducers in use in routine clinical practice

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions (Revised: July 31, 2017)

Ultrasound in Anesthesia: Applying Scientific Principles to Clinical Practice

COURSE DESCRIPTION FOR NONDESTRUCTIVE TESTING

Basic Physics of Ultrasound and Knobology

2015 ARDMS Sonography Principles & Instrumentation Job Task Analysis Summary Report

Concepts of Imaging and Knobology

Basic Ultrasound Physics Board Review Questions

Please follow all instructions carefully.

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Evaluation of the Quality of Thick Fibre Composites Using Immersion and Air- Coupled Ultrasonic Techniques

Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound

Terminology Tissue Appearance

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Dr Emma Chung. Safety first - Physical principles for excellent imaging

ACR Accreditation Update in MRI

Update on MQSA and Mammography Accreditation

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector

10 Years Experience in Industrial Phased Array Testing of Rolled Bars

Performance of phased array and conventional ultrasonic probes on the new ISO reference block

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red)

Lesson 03: Sound Wave Propagation and Reflection. This lesson contains 15 slides plus 14 multiple-choice questions.

Mammography. What is Mammography? What are some common uses of the procedure?

Ultrasonic Phased Array Testing of Complex Aircraft Structures

CRACK DETECTION AND PIPELINE INTEGRITY SOLUTIONS

Performance Evaluation of Ultrasound Systems: Advanced Methods and Applications

Optimising your Doppler settings for an accurate PI. Alison McGuinness Mid Yorks Hospitals

DC-6. Diagnostic Ultrasound System

Nuclear Associates

Confident and Conclusive Diagnosis with Flexible Solutions

Developments in Ultrasonic Inspection II

Ultrasonic arrays are now widely used in underwater sonar

ADVANCE ULTRASONIC INSPECTION

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound

Transcription:

QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences Zheng F. Lu, PhD Radiology Department Columbia University Email: zfl1@columbia.edu Outline General overview of basic ultrasound physics QA/QC guidelines provided by professional bodies Accreditation programs QC testing procedures and phantoms Discussion: efficacy of ultrasound QA/QC program References US Imaging Range Equation D = ct/2 D is the depth where the echoes are generated; c is sound propagation speed in the media; t is the time delay between the pulse emission and echo reception. D Frame Rate Limitation The frame rate is ultimately limited by the sound propagation speed (c). The maximum frame rate is determined by the following relationship, where D is the maximum depth of the field of view, and N is the total number of acoustic lines per frame. Frame Rate max = 2 c ND Dynamic Range Block Diagram of an Ultrasound Imager Dynamic range is the ratio of the largest to smallest signals that an instrument or a component of an instrument can respond to without distortion. Signal Out Saturation Dynamic Range Noise Echo Signal In Transducer Beam Former Transmitter Digital storage & transfer Receiver amplify & Process Memory scan converter Hard Copy Monitor 1

Gain Vs Power Increasing either gain or power increases signal amplitude and image brightness throughout the field Increasing output power also increases the signal to noise ratio; increasing gain increases the signal level as well as the noise level. Therefore, increasing the power will improve penetration but increasing the gain will not. Increasing output power also increases exposure to the patient; increasing gain does not. the minimum distance between two objects positioned along the axis of the beam where both are displayed as separate structures. Spatial Pulse Length (SPL) Axial Resolution Axial - Resolution SPL 2 Lateral Resolution the minimum distance between two objects positioned perpendicular to the axis of the beam where both are displayed as separate structures. DOPPLER SHIFT f D Considering the Doppler angle θ: 2v cos( θ ) f c f D 0 Lateral Resolution is determined by the beam profile. f0 θ f r Goal of Ultrasound QA/QC Program Aliasing To make sure a system is set up correctly and performs to specified standards. To maintain the consistency of the performance. To reveal problems at its earliest stage before it severely interferes with the clinical practices. 2

Three Levels of Ultrasound QC Testing Level 1 Frequent QC testing: Operator s QC Level 2 Semi-annual or annual QC testing: Physicist s QC Level 3 Acceptance testing / baseline set-up American Association of Physicists in Medicine (AAPM) Ultrasound Task Group No. 1 Real-time B-mode ultrasound quality control test procedures By MM Goodsitt et al Medical Physics. 25(8):1385-406, 1998 Aug. American Institute of Ultrasound in Medicine (AIUM) Quality Assurance Manual for Gray Scale Ultrasound Scanners (Stage 2) edited by E. Madsen AIUM, Laurel, MD, 1995 The Institute of Physical Sciences in Medicine (IPSM) Report No. 71 Routine Quality Assurance of Ultrasound Imaging Systems edited by R. Price York: ISPM, 1995 American Institute of Ultrasound in Medicine (AIUM) Performance Criteria and Measurements for Doppler Ultrasound Devices: Technical Discussion 2 nd Edition The Institute of Physical Sciences in Medicine (IPSM) Report No. 70 Testing of Doppler Ultrasound Equipment edited by PR Hoskins, SB Sherriff and JA Evans AIUM, Laurel, MD, 2002 York: ISPM, 1994 3

ACR Accreditation www.acr.org Breast Ultrasound Accreditation Program (including Ultrasound-guided Breast Biopsy) General Ultrasound Accreditation Program Obstetrical Gynecological General Vascular Combination of the above AIUM Accreditation www.aium.org Ultrasound practices in various fields: Breast Obstetrical and Trimester-Specific Obstetrical Gynecological Abdominal/General ACR Accreditation QC Requirements A QA program should be in place; The QA program must be directed by a medical physicist or by the supervising physician; Routine QC testing must be done at least semiannually; Do the same tests, monitor changes over time and take corrective actions when necessary; Testing results, corrective action, and the effects of corrective action must be documented and maintained on site. There is currently no ACR-preferred single phantom. ACR Recommended Semi-Annual QC Tests for Breast Ultrasound Accreditation Maximum depth of visualization Distance accuracy (vertical and horizontal) Uniformity Electrical-mechanical cleanliness condition Anechoic void perception Ring down Lateral resolution Quality Control Checklist ACR Recommended Technologist s QC Tests for Breast Ultrasound Accreditation Adherence to universal infection control procedures for each biopsy All transducers should be cleaned between patients Distance calibration - quarterly Gray scale photography - quarterly ACR Required Semi-Annual QC Tests for General Ultrasound Accreditation System sensitivity and/or penetration capability Image uniformity Photography and other hard copy recording Low contrast object detectability (optional) Assurance of electrical and mechanical safety Vertical and horizontal distance accuracy (recommended only when the program is initiated for a scanner) 4

Ultrasound QC Phantoms B-mode: 1. Multi-purpose or general purpose tissue/cyst phantom 2. Low contrast resolution phantom 3. Spherical lesion phantom 4. Beam profile and slice thickness phantom 5. Prostate QC phantom Doppler-mode: 1. Flow phantom 2. String phantom Ultrasound Phantom Suppliers ATS Labs, Bridgeport, CT www.atslabs.com CIRS, Norfolk VA www.cirsinc.com Gammex RMI, Madison WI www.gammex.com Nuclear Associates, Hicksville NY www.inovision.com/nuclear_associates Research Laboratories Desiccation: Water-based Phantom Material Monitoring Desiccation (moisture loss) Monitor phantom weight loss regularly Rejuvenation requirements provided by manufacturer Rejuvenation recommended usually after a 10-15 gram weight loss Rejuvenation accomplished by injecting a water-alcohol fluid through the polyurethane scanning surface Rejuvenation: Water-based Phantom Materials Rejuvenation Procedures and Hints Follow manufacture s recommended rejuvenation procedures Remember: 1 gram = 1 cc = 1 ml Remove air bubbles from syringe prior to injection Pick a remote location for injection - seal with super glue Scan the phantom with a vertical column and a horizontal row of reflectors; The digital caliper readout on screen is checked against the known distance between reflectors; Distance Accuracy Criteria: Vertical: 1.5% of the actual distance or 2 mm, whichever is greater. Horizontal: 3% of the actual distance or 3 mm, whichever is greater. System Sensitivity/Penetration The maximum depth of visualization is determined by comparing the gradually weakening echo texture to electronic noises near the bottom of the image. Do this test with the same settings and monitor the changes over time. 5

System Sensitivity/Penetration This test should be done with following settings: maximum transmit power, proper receiver gain and TGC that allows echo texture to be visible in the deep region, transmit focus at the deepest depth. Image Uniformity Adjust the TGC controls and other sensitivity controls to obtain an image as uniform as possible Inspect the image to detect any kinds of vertical or radially oriented streaks dropouts reduction of brightness near edges of the scan brightness transitions between focal zones Soft and/or Hard Copy Recording I The shades of gray, weak, and strong echo texture should be optimized and consistent between the image display on the ultrasound scanner and the photographic hard copies or soft copy displays on the workstation in the reading room. For quick follow-up testing, the Grayscale bar pattern on the clinical image display can be used. Soft and/or Hard Copy Recording II Use the SMPTE test pattern and other patterns if they are available on the ultrasound scanner. Workstation monitor display should be included in QC tests. Soft and/or Hard Copy Recording III Low Contrast Object Detectability Film processor QC needs to be done daily. Darkroom fog test needs to be done at least semi-annually. Scans of a low contrast resolution phantom can reveal the low contrast object detectability which is an optional test on the ACR semi-annual QC test list for general ultrasound accreditation. 6

Assurance of Electrical and Mechanical Safety The equipment related physical checks are also worth attention. These include: checking for any cracks or delamination in the transducers, noting for loose and frayed electric cables, loose handles or control arms, checking the working condition of the wheels and the wheel locks, making sure all the accessories (such as VCR, printer, etc.) are fastened securely to the system, Making sure the image monitors are clean and the air filters are not too dusty. Dead Zone (Ring Down) A group of reflectors consisting of fibers are placed at different separations from the top of the phantom (~ 1-8 mm). As the transducer scans across the top, the distance from the transducer to the first reflector completely imaged is equal to the dead zone (ring down) distance. Axial and Lateral Resolution 1.3 mm 3.4 mm Spatial resolution may be evaluated by either of the following ways: Count the number of the pins distinguished without overlap in lateral and axial direction; thus the minimum target spacing is documented; Measure the size of the pin in lateral and axial direction. Spherical Lesion Phantom 2 and 4 mm diameter spherical targets; Low scatter level; Target centers are coplanar. Courtesy of Dr. J. A. Zagzebski, UW-Madison Prostate Ultrasound QC Methods Distance Accuracy Template overlay Courtesy of Dr. J. Kofler, Mayo Clinic Ultrasound Doppler QC Testing Doppler QC tests include Doppler signal sensitivity, Doppler angle accuracy, Color display and Gray-scale image congruency, Range-gate accuracy, flow readout accuracy. 7

Doppler Velocity Accuracy: Variations among 8 clinical units each with transducers of 3.5 MHz, 5 MHz and 7 MHz. Flow Rate (ml/min) 3.5MHz 5 MHz 7 MHz Average 100 5.7% 17.4% 8.2% 10.4% 237 10.2% 7.7% 8.2% 8.7% 398 6.7% 10.6% 4.1% 7.1% 611 8.0% 6.9% 9.2% 8.0% 800 10.2% 13.9% 7.1% 10.4% 1093 13.0% 13.0% 12.5% 12.9% AVG 9.0% 11.6% 8.2% 9.6% Review of A QA/QC Program in a period of 7 years Group I (28 units with age of 5 years or less) 101 reports with 67 deficiencies: Image uniformity test Mechanical checks Image display soft / hard copy quality Poor spatial and contrast resolution Doppler QC tests (sensitivity, color congruency) Software problem Maximum depth of visualization reduction 30% 27% 21% 9% 6% 4% 3% Review of A QA/QC Program (cont.) in a period of 7 years Discussion Group II (10 units with age of over 5 years) 28 reports with 67 deficiencies: Poor spatial and contrast resolution, dead zone Mechanical checks Maximum depth of visualization reduction Image display soft / hard copy quality Software problem Image uniformity test Doppler QC tests (sensitivity, color congruency) 31% 25% 18% 15% 6% 3% 2% The guidelines for ultrasound QA/QC programs are descriptive. Maintaining an ultrasound QA/QC program is straightforward and effective in identifying deficiencies. Ultrasound phantoms are needed for evaluating performance and testing consistency. However, we need to know that the tissue mimicking material can only mimic some of the properties of tissue in real clinical situations. Discussion (cont.) Assessment of the efficacy and relevance of ultrasound QA procedures is rare. The guidelines need to be updated periodically as ultrasound technology develops. References J Zagzebski, US quality assurance with phantoms. In Categorical Course in Diagnostic Radiology Physics: CT and US Cross-Sectional Imaging, Edited by L. Goldman and B. Fowlkes, 2000, Oak Brook, IL: Radiological Society of North America, pp. 159-170. J Zagzebski and J Kofler, Ultrasound Equipment Quality Assurance, in Quality Management in the Imaging Sciences, ed. By J Papp, 2002, St. Louis, Mosby, pp. 207-215. SC Metcalfe and JA Evans, A study of the relationship between routine ultrasound quality assurance parameters and subjective operator image assessment, The British Journal of Radiology, 65: 570-575, 1992. 8

References (cont.) A Goldstein, slice thickness measurements, J Ultrasound Med, 7: 487-498, 1988. NJ Dudley, K Griffith, G Houldsworth, M Holloway, MA Dunn, A review of two alternative ultrasound quality assurance programmes, European J. of Ultrasound, 12: 233-245, 2001. NM Donofrio, JA Hanson, JH Hirsch, WE Moore, Investigating the efficacy of current quality assurance performance tests in diagnostic ultrasound, Journal of Clinical Ultrasound, 12: 251-260, 1984. J. Kofler, R. Kruger, Z. Lu, B. Fowlkes and M. Holland, Clinical Ultrasound Physics: a workbook for physicists, residents and students, Medical Physics Publishing, 2001. 9