2D Free-Breathing Dual Navigator-Gated Cardiac Function Validated Against the 2D Breath-hold Acquisition

Similar documents
Relationship of Number of Phases per Cardiac Cycle and Accuracy of Measurement of Left Ventricular Volumes, Ejection Fraction, and Mass

Impact of the ECG gating method on ventricular volumes and ejection fractions assessed by cardiovascular magnetic resonance imaging

Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration SINE Wave Curve On Magnetic Resonance Imaging

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2012 April 1.

Analysis of Residual Coronary Artery Motion for Breath Hold and Navigator Approaches Using Real-Time Coronary MRI

Original Research. Girish Narayan, MD, 1 * Krishna Nayak, PhD, 2 John Pauly, PhD, 3 and Bob Hu, MD 2,4

Cardiac magnetic resonance (CMR) imaging is widely considered

Using Radial k-space Sampling and Steady-State Free Precession Imaging

Normal Human Left and Right Ventricular Dimensions for MRI as Assessed by Turbo Gradient Echo and Steady-State Free Precession Imaging Sequences

Raja Muthupillai, PhD. Department of Diagnostic and Interventional Radiology St. Luke s Episcopal Hospital. Research Support: Philips Healthcare

CME. Original Research

Objectives. CMR Volumetric Analysis 8/25/11. CMR Volumetric Analysis Technique. Cardiac imaging plane acquisition. CMR Volumetric Analysis

ABSTRACT INTRODUCTION

Fulfilling the Promise

Measurement of Left Ventricular Velocities: Phase Contrast MRI Velocity Mapping Versus Tissue-Doppler-Ultrasound in Healthy Volunteers

A copy can be downloaded for personal non-commercial research or study, without prior permission or charge

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine

Disclosure of Interests. No financial relationships to disclose concerning the content of this presentation or session.

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview

Velocity Vector Imaging as a new approach for cardiac magnetic resonance: Comparison with echocardiography

General Cardiovascular Magnetic Resonance Imaging

Global left ventricular circumferential strain is a marker for both systolic and diastolic myocardial function

MR Advance Techniques. Cardiac Imaging. Class IV

Objectives 8/17/2011. Challenges in Cardiac Imaging. Challenges in Cardiac Imaging. Basic Cardiac MRI Sequences

How much are atrial volumes and ejection fractions assessed by cardiac magnetic resonance imaging influenced by the ECG gating method?

Echocardiographic Assessment of the Left Ventricle

CHAPTER. Quantification in cardiac MRI. This chapter was adapted from:

NIH Public Access Author Manuscript J Magn Reson Imaging. Author manuscript; available in PMC 2013 August 01.

Applying a Level Set Method for Resolving Physiologic Motions in Free-Breathing and Non-Gated Cardiac MRI

Cardiovascular magnetic resonance artefacts

Automated Identification of Minimal Myocardial Motion for Improved Image Quality on MR Angiography at 3 T

Influence of Velocity Encoding and Position of Image Plane in Patients with Aortic Valve Insufficiency Using 2D Phase Contrast MRI

MR Advance Techniques. Vascular Imaging. Class II

Functional Chest MRI in Children Hyun Woo Goo

Previous talks. Clinical applications for spiral flow imaging. Clinical applications. Clinical applications. Coronary flow: Motivation

MRI protocol for post-repaired TOF

Reproducibility and Accuracy of Echocardiographic Measurements of Left Ventricular Parameters Using Real-Time Three-Dimensional Echocardiography

Fully Refocused Gradient Recalled Echo (FRGRE): Factors Affecting Flow and Motion Sensitivity in Cardiac MRI

Free-Breathing Phase Contrast MRI with Near 100% Respiratory Navigator Efficiency Using k-space-dependent Respiratory Gating

Spiral Coronary Angiography Using a Blood Pool Agent

Strain-encoded cardiac magnetic resonance imaging: a new approach for fast estimation of left ventricular function

CME. Original Research

Volumetric Late Gadolinium-Enhanced Myocardial Imaging With Retrospective Inversion Time Selection

Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance

Cardiac gating calibration by the Septal Scout for magnetic resonance coronary angiography

MR coronary artery imaging with 3D motion adapted gating (MAG) in comparison to a standard prospective navigator technique

Ventricular function assessment using MRI: comparative study between cartesian and radial techniques of k-space filling

Automatic cardiac contour propagation in short axis cardiac MR images

Contijoch et al. Journal of Cardiovascular Magnetic Resonance (2015) 17:37 DOI /s

How We Perform Delayed Enhancement Imaging

Cardiac Cine MR-Imaging at 3T: FLASH vs SSFP

Multiple Gated Acquisition (MUGA) Scanning

Amol S Pednekar 1*, Hui Wang 2, Scott Flamm 3, Benjamin Y. Cheong 4 and Raja Muthupillai 4

Cardiac Function Evaluation with Cine MRI of the Heart

Real-Time Black-Blood MRI Using Spatial Presaturation

Rapid Quantitation of High-Speed Flow Jets

MRI evaluation of left ventricular volume in children and adults during inscanner

Self-Gated Cardiac Cine MRI

Noncoronary Cardiac MDCT

UK Biobank. Imaging modality Cardiovascular Magnetic Resonance (CMR) Version th Oct 2015

Automated Volumetric Cardiac Ultrasound Analysis

Magnetic Resonance Angiography

Tools for cardiovascular magnetic resonance imaging

Impact of Papillary Muscles in Ventricular Volume and Ejection Fraction Assessment by Cardiovascular Magnetic Resonance

ORIGINAL RESEARCH. Reproducibility of Myocardial T 1 and T 2 Relaxation Time Measurement Using Slice-Interleaved T 1 and T 2 Mapping Sequences

Three-dimensional Wall Motion Tracking:

1Pulse sequences for non CE MRA

Coronary Magnetic Resonance Angiography 1.5T Techniques

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2008 May 13.

Analysis of left ventricular mass, volume indices and ejection fraction from. SSFP cine imaging: a comparison of semi-automated, simplified manual,

Cardiac MRI in Small Rodents

Can SCMR CMR protocol recommendations

Jihye Jang 1,2 Steven Bellm 1 Sébastien Roujol 1 Tamer A. Basha 1,3 Maryam Nezafat 1,4 Shingo Kato 1 Sebastian Weingärtner 1,5 Reza Nezafat 1 *

Cardiovascular MR Imaging at 3 T: Opportunities, Challenges, and Solutions 1

Normal values for cardiovascular magnetic resonance in adults and children

Non-Contrast MRA. How and When 1996! Why Non-Contrast MRA? Angiography: What are our goals? Inflow Techniques Differences in excitation hx

MultiTransmit 4D brings robustness to 3.0T cardiac imaging

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE

A Magnetic Resonance Imaging Method for

Phase-Sensitive Inversion Recovery for Detecting Myocardial Infarction Using Gadolinium-Delayed Hyperenhancement

Tissue Doppler Imaging in Congenital Heart Disease

6/23/2009. Inversion Recovery (IR) Techniques and Applications. Variations of IR Technique. STIR, FLAIR, TI and TI Null. Applications of IR

Fratz et al. Journal of Cardiovascular Magnetic Resonance 2013, 15:51

Use of Cardiac Computed Tomography for Ventricular Volumetry in Late Postoperative Patients with Tetralogy of Fallot

Single and Biplane TrueFISP Cardiovascular Magnetic Resonance for Rapid Evaluation of Left Ventricular Volumes and Ejection Fraction

Normal LV Ejection Fraction Limits Using 4D-MSPECT: Comparisons of Gated Perfusion and Gated Blood Pool SPECT Data with Planar Blood Pool

Coronary artery disease (CAD) risk factors

Open Access. 90 The Open Cardiovascular Medicine Journal, 2011, 5, 90-98

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

On the feasibility of speckle reduction in echocardiography using strain compounding

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

How I do it: Non Contrast-Enhanced MR Angiography (syngo NATIVE)

Full terms and conditions of use:

Myocardial tagging by Cardiovascular Magnetic Resonance: evolution of techniques pulse sequences, analysis algorithms, and applications

Automated Image Analysis Techniques for Cardiovascular Magnetic Resonance Imaging

Real-Time MRI of Joint Movement With TrueFISP

3D-stress echocardiography Bernard Cosyns, MD, PhD

8/4/2016. Optimizing Pediatric Cardiovascular MRI. Disclosure. Outline. Jie Deng, PhD, DABMP, Cynthia Rigsby, MD

Black Blood Gradient Echo Cine Magnetic Resonance Imaging of the Mouse Heart

Transcription:

JOURNAL OF MAGNETIC RESONANCE IMAGING 28:773 777 (2008) Technical Note 2D Free-Breathing Dual Navigator-Gated Cardiac Function Validated Against the 2D Breath-hold Acquisition Dana C. Peters, PhD, 1 * Reza Nezafat, PhD, 1 Holger Eggers,PhD, 3 Christian Stehning, PhD, 3 and Warren J. Manning, MD 1,2 Purpose: To develop and validate a free-breathing cardiac cine acquisition, with potential to simplify cardiac MR studies, provide registered slices, and increase spatial resolution. Materials and Methods: A 2D free-breathing (FB) navigator-gated cine radial acquisition for cardiac function was developed that used two navigators (one placed prior to the QRS, and another 500 msec after the QRS complex, after systole) to provide complete motion-compensated assessment of systole, without loss of end-diastole. Eleven subjects were studied. Results: The 2D FB method provided results visually and quantitatively similar to the 2D breath-hold (BH) methods. Comparison of volumes measured with FB to those measured by standard 2D BH cine resulted in mean bias 2 standard deviations of 1.0 ml 13.7 ml, 1.1 ml 7.6 ml, 3.0 g 18.8 g, and 0.3% 2.5%, for end-diastolic volume, end-systolic volume, left ventricular (LV) mass, and ejection fraction, respectively. Slice misregistration was identified in four (36%) of the BH studies, but none (0%) of the FB studies. In subjects with slice misregistration, there was greater discordance in LV volume measurements (P 0.05 for end-diastolic mass). Conclusion: The FB cine acquisition provided results qualitatively and quantitatively similar to 2D BH methods with improved slice registration. 1 Department of Medicine (Cardiovascular Division), Beth Israel Deaconess Medical Center, Boston, Massachusetts. 2 Department of Radiology, Beth Israel Deaconess Medical Center, Boston, Massachusetts. 3 Philips Research Europe, Sector Medical Imaging Systems, Hamburg, Germany. Contract grant sponsor: American Heart Association; Contract grant number: AHA SDG 0530061N; Contract grant sponsor: National Institutes of Health; Contract grant number: NIBIB K01 EB004434-01A1. *Address reprint requests to: D.C.P., Department of Medicine (Cardiovascular Division), 330 Brookline Ave., Beth Israel Deaconess Medical Center, Boston, MA 02215. E-mail: dcpeters@bidmc.harvard.edu Received July 13, 2007; Accepted March 14, 2008. DOI 10.1002/jmri.21417 Published online in Wiley InterScience (www.interscience.wiley.com). Key Words: myocardial function; free-breathing; navigatorgating; balanced SSFP; radial imaging; projection reconstruction J. Magn. Reson. Imaging 2008;28:773 777. 2008 Wiley-Liss, Inc. FREE-BREATHING (FB) cine 2D cardiac magnetic resonance (CMR) imaging of myocardial function is an important goal as an alternative to the standard breathheld (BH) approach to eliminate repeated breath-holding and to minimize slice misregistration. Such an approach has many applications, including dobutamine stress CMR (1), since repeated BH commands increase the complexity of this imaging test. Furthermore, imaging with multiple BHs commonly leads to slice misregistration, due to BH variability, resulting in unevaluated (ie, missed) and overevaluated segments of the myocardium (2,3). BH variability leads to throughplane and in-plane centroid shifts with a standard deviation (SD) of about 3 mm and 4 mm, respectively (3), for the short axis orientation. For basal slices, misregistration can effect volumetric measurements (eg, if a basal slice, which contributes significantly to volumes and mass, is not acquired or is reacquired). Therefore, we sought to develop an FB cine CMR method that provides registered slices. Many alternatives to the standard clinical 2D BH cine technique have been proposed. These include real-time 2D methods (4) and self-gating (5). At the same time, 3D approaches, mostly using BH, have been studied (6 11). 3D imaging is attractive because of the potential for higher signal-to-noise ratio (SNR) due to signal averaging, and because it provides registered slices, but has drawbacks including loss of inflow enhancement and the excitation of a large volume of magnetization, causing a greater possibility of dark band flow artifacts (12,13). Here we present an alternative approach using 2D balanced steady-state free precession (SSFP) and FB navigator-gated (NAV-gated) imaging for reducing respiratory motion artifacts. NAV-tracked BH cine has been described previously (2) for improving slice registration between BHs. In that study the navigator was 2008 Wiley-Liss, Inc. 773

774 Peters et al. 14 years) were studied on a 1.5T Philips Achieva (Philips Medical Systems, Best, Netherlands). All subjects provided written informed consent. In all subjects, both the conventional 2D BH cine and the 2D FB cine balanced SSFP methods were used to image the entire left ventricle in the short-axis orientation. 2D FB Cine With NAV-Gating Figure 1. Pulse sequence schematic for the 2D free-breathing NAV-gated cardiac function pulse sequence. The sequence is based on an ECG-triggered segmented multiphase balanced SSFP acquisition. The first respiratory navigator (NAV) is placed prior to the QRS, followed by time waiting for the QRS trigger. When the trigger is detected, a 5-pulse linear ramp (5LR) start-up sequence is performed, after which data are acquired for multiple phases. The second NAV is performed 500 msec later (after systole). The data from a heart-beat are used if the first and second NAV measurements are within the acceptance window. placed after each QRS, so that the initial image was delayed and did not represent end-diastole, resulting in smaller end-diastolic volumes. Use of NAVs is challenging for balanced SSFP cine because the NAVs interrupt steady-state and create about 25 msec of dead-time during which the cardiac function is not monitored. However, based on recent work using two NAVs (14 16) for phase-contrast, we have developed an efficient reliable NAV-gated method that does not interfere with imaging of systolic function. By placing the NAVs immediately prior to the QRS, and midway through the RR interval, end-diastole and all of systole can be imaged. The method allows assessment of global and regional systolic function, including wall motion abnormalities that are confined to the systolic phase. However, the loss of a frame at mid-diastole is a limitation for assessment of diastolic function. We hypothesized that this NAV-gated FB 2D approach would provide well-registered slices, because the registration is not reliant on the subject s BH capability, and have high NAV efficiency due to the short interval between the two NAVs. MATERIALS AND METHODS Eleven subjects (nine healthy subjects and two patients, one with atrial fibrillation; nine female, age 26 FB cardiac function data were acquired with a 2D balanced SSFP undersampled radial acquisition. Navigators (25 msec in duration each), placed on the right hemi-diaphragm with signals acquired just prior to the QRS and 500 msec after the QRS were used (Fig. 1). Data collected during a heartbeat were accepted if these two NAVs were both within the 5-mm acceptance window. The NAVs were surrounded by /2-TR/2 store/ restore sequences (17). The first cardiac phase was prepared using a series of 5 8 ramped flip angles, resulting in a 15 25 msec delay in data acquisition (18). The scan parameters were: TR/TE/ 3.1 msec/1.5 msec/40 45. The flip angle was less than the standard flip angle of 60 used in BH cine (see below) in order to lower the specific absorption rate (SAR) during the 3 4-minute scan. Sixteen cardiac phases were acquired with prospective ECG-gating (with dead time at the end of each RR interval, see Fig. 1), with 20 views per segment acquired, providing a 50-msec temporal resolution. The method requires about 60 heartbeats with navigator measurements within the 5-mm acceptance window. With typical navigator efficiency of 20% 40%, the actual scan time was 2.5 5 minutes. Spatial resolution was 2 2 8 mm (2 mm gaps), using 160 80 Np (acquired with UNFOLD interleaving which alternated with each cardiac phase (19)); 320 cm field of view (FOV); 9 12 slices acquired, dependent on the chamber size. FB images were reconstructed offline, using gridding reconstruction, FFT, and UNFOLD (19) processing. The UNFOLD filter was a lowpass filter applied in the temporal frequency domain (10). 2D BH Cine For comparison, conventional Cartesian 2D BH cine imaging was performed in analogous short-axis orientation using the following sequence: balanced SSFP, TR/TE/ 3.1 msec/1.5 msec/60, 30 cardiac phases reconstructed using retrospective ECG-gating, with Figure 2. Comparison of standard 2D breath-hold cine imaging with 2D freebreathing cine in one subject, showing all slices from base to apex at end-systole.

2D Navigator-Gated Cardiac Cine 775 Figure 3. Comparison of standard 2D BH cine with 2D FB cine in one subject, showing the first two phases and the systolic phase for a mid-wall slice. For the FB method, the first and second phase are of similar quality, demonstrating an acceptable transition to steady state. Some improvement in epicardial border delineation is observed in subsequent cardiac phases for the FB method (arrows). about 12 views per segment acquired, 12 heartbeats per slice, 2 2 8 mm (2 mm gaps) spatial resolution, 160 160 Ny, 320 cm FOV, 9 12 slices in 9 12 breathholds. Measurements SNR was measured from a single basal end-systolic slice in each subject. Noise was estimated as the SD of an air space region of interest (ROI) ( 40 pixels), anterior to the chest wall, with an attempt to avoid regions with artifacts. The left ventricular (LV) end-diastolic (EDV) and end-systolic (ESV) volumes were measured from the short-axis data using ROIs and computeraided planimetry with custom tools (MatLab 7.1, Math- Works, Natick, MA). End-diastolic myocardial volumes (EDMV) were measured and converted to end-diastolic masses (EDM) by multiplication by 1.05 g/ml. Endsystole was defined as the mid-ventricular frame with the smallest LV cavity, and end-diastole was defined as the first cardiac phase in the dataset. The 2D BH and FB datasets were reformatted in a long-axis view (using MatLab) to visualize slice misregistration. Slice misregistration was classified as large if the septal wall was observed to shift by a distance comparable to its width, otherwise small. and BH methods for the first, second, and systolic cardiac phases. For the FB scans, the quality of the initial cardiac phase (Fig. 3) was visually acceptable using dummy RF pulses and a linear ramp to catalyze the transition to steady state. For the FB scans, the epicardial border delineation improved somewhat in later phases, especially the contrast between myocardium and epicardial fat. Total scan time was 10 minutes (including resting periods) with the BH method and 3.7 minutes (including a NAV efficiency of 30 9.8%) with the FB method. The average blood and myocardial SNR were 70 16 and 23 7 for 2D BH, and 51 15, and 30 9 for 2D FB, respectively. Blood-myocardium contrast to noise ratio (CNR) was 47 14 and 21 10 for the BH and FB, respectively. The lower flip angle used for FB method to reduce SAR (40 45 vs. 60 for BH) theoretically results in higher myocardial signal and thus lower CNR between blood and myocardium signal, as observed. Slice misregistration was visible in the long-axis reformats of all 2D BH stacks, to a greater or lesser extent, with much less misregistration present in the 2D FB scans (Fig. 4). Of the 11 studies, the slice misregistration was large in 4 (36%) of the 2D BH short-axis stacks, versus 0 (0%) of the 2D FB stacks. Figure 5a plots LV volumes (end-diastolic and end-systolic) and end-diastolic myocardial volume of 2D FB versus 2D BH. The correlation coefficients were R 0.97, 0.97, and 0.85 for LVEDV, LVESV, and LVEDM, respectively. The data are also shown in a Bland Altman plot in Fig. 5b. Table 1 provides the results of comparison of LVESV, LVEDV, EDM, and LV ejection fraction (LVEF), including the Bland Altman limits of agreement. The differences between the BH and FB measurements were not significant. Figure 5c compares discordance between BH and FB measures of LVEDV, LVESV, and LVEDMV, showing greater discordance for studies with large slice misregistration versus small misregistration, which was significant only for LVEDM (P 0.01 for EDMV, P 0.35 for EDV, P 0.30 for ESV). DISCUSSION In this study we demonstrate that 2D FB cardiac function imaging is possible using NAVs for respiratorycompensation. The two NAVs provide efficient respiratory motion compensation for complete assessment of the first half of the cardiac cycle, ie, systolic function. Quantitative comparisons of LV volumes with the 2D Statistics All data are presented as mean SD. The results from the 2D FB method and 2D BH method were compared using a two-tailed Student s t-test. P 0.05 was considered significant. The limits of agreement between the two techniques were estimated as the mean difference 2 SD, following the Bland Altman analysis. RESULTS Figure 2 compares a 2D stack of short-axis slices with BH and with FB, showing similar image quality from apex to base. Figure 3 compares images from the FB Figure 4. Comparison in one subject of long-axis reformats of a short-axis stack of slices for 2D BH and FB studies, showing the slice misregistration of the BH study, which is readily observed in the septal wall.

776 Peters et al. gold-standard show excellent correlation, with limits of agreement similar to that of reported inter- and intraobserver variability of 2D BH SSFP (20). Discordance between BH and FB measurements was greatest for myocardial mass (EDM) (Table 1), although Table 1 Bland Altman comparison of 2D breath-hold and free-breathing measurements of left ventricular function 2D FB 2D BH Bias 2 SD p-value EDV (ml) 127.2 126.2 1.0 13.7 0.64 ESV (ml) 50.3 49.5 1.1 7.6 0.49 EDM (g) 86.4 83.4 3.0 18.8 0.31 EF (%) 59.1 58.8 0.3 2.5 0.59 SD standard deviation, FB free-breathing, BH breath-hold, EDV end-diastolic volume, ESV end-systolic volume, EDM end-diastolic mass, EF ejection fraction. the difference was not significant (P 0.31) in this small study. The data show that this was caused by larger end-diastolic epicardial contours using 2D FB radial. Potentially, this may relate to steady-state effects in the FB method, which reduce the contrast between myocardium and epicardial fat in the first cardiac phase (see Fig. 3, arrow). The FB method provides registered slices and high NAV efficiency, both important for improving assessment of myocardial function, for example, during a stress exam. This investigation showed that the common presence of slice misregistration resulted in less agreement between 2D BH and 2D FB measures of LV function, which reached statistical significance for LV mass. This important finding suggests that the 2D FB method may be more accurate than the 2D BH method. The SNR measurements demonstrate that 2D SNR is sufficient, and the benefits of 3D for SNR are not required at this spatial resolution, or even greatly increased spatial resolution. Furthermore, past studies show that the expected SNR increases due to averaging have been offset by reduced inflow signal with the 3D approach (8,10). Since SNR and scan time are sufficient, increased spatial resolution is possible with this 2D FB cine method. This would contribute to visualization of valves, right ventricular wall motion, and trabeculation. Limitations Figure 5. a: Direct comparison of end-diastolic LV volume (EDV), end-systolic LV volume (ESV), and end-diastolic myocardial volume (EDMV) for the 2D FB and 2D BH methods, in all 11 subjects. b: Bland Altman analysis, showing bias and 2 SDs. c: The extent of disagreement between the 2D BH and FB methods, plotted for studies with registered versus misregistered BH images. Greater disagreement was observed for studies with misregistration, but significance was only found for myocardial volume. This study compared 2D BH and FB methods. However, there were other differences, including different temporal resolution (50 msec vs. 36 msec), use of radial with undersampling, and a smaller flip angle for FB (40 45 vs. 60 ). The differences in total scan time (10 minutes for BH vs. 4 minutes for FB) were due to radial undersampling by a factor of 2 and a temporal resolution that was 1.4 times reduced for the FB method. Parallel imaging, which is used clinically at our center for cine scans, can reduce the scan time of the BH cine study by about 2, but was not used here in order to permit simple SNR measurements. Our employment of the radial technique for the FB study was motivated by its well-known superior motion blurring and ghosting properties. However, the conclusions that NAV-gated FB cardiac function is feasible, and that it reduces the common occurrence of slice misregistration that may result in inaccurate volumetric measurements, is independent of our choice of trajectory and temporal resolution.

2D Navigator-Gated Cardiac Cine 777 In conclusion, we have developed a 2D FB dual navigator-gated method using two NAVs, permitting complete assessment of systole. This method provided results qualitatively and quantitatively similar to 2D BH methods with improved slice registration, and may facilitate acquisition of images with higher spatial resolution. REFERENCES 1. Kwong RY, Schussheim AE, Rekhraj S, et al. Detecting acute coronary syndrome in the emergency department with cardiac magnetic resonance imaging. Circulation 2003;107:531 537. 2. Chuang ML, Chen MH, Khasgiwala VC, McConnell MV, Edelman RR, Manning WJ. Adaptive correction of imaging plane position in segmented k-space cine cardiac MRI. J Magn Reson Imaging 1997; 7:811 814. 3. Swingen C, Seethamraju RT, Jerosch-Herold M. An approach to the three-dimensional display of left ventricular function and viability using MRI. Int J Cardiovasc Imaging 2003;19:325 336. 4. Plein S, Smith WH, Ridgway JP, et al. Qualitative and quantitative analysis of regional left ventricular wall dynamics using real-time magnetic resonance imaging: comparison with conventional breath-hold gradient echo acquisition in volunteers and patients. J Magn Reson Imaging 2001;14:23 30. 5. Larson AC, Kellman P, Arai A, et al. Preliminary investigation of respiratory self-gating for free-breathing segmented cine MRI. Magn Reson Med 2005;53:159 168. 6. Alley MT, Napel S, Amano Y, et al. Fast 3D cardiac cine MR imaging. J Magn Reson Imaging 1999;9:751 755. 7. Barger AV, Grist TM, Block WF, Mistretta CA. Single breath-hold 3D contrast-enhanced method for assessment of cardiac function. Magn Reson Med 2000;44:821 824. 8. Jung BA, Hennig J, Scheffler K. Single-breathhold 3D-trueFISP cine cardiac imaging. Magn Reson Med 2002;48:921 925. 9. Kozerke S, Tsao J, Razavi R, Boesiger P. Accelerating cardiac cine 3D imaging using k-t BLAST. Magn Reson Med 2004;52:19 26. 10. Peters DC, Ennis DB, Rohatgi P, Syed MA, McVeigh ER, Arai AE. 3D breath-held cardiac function with projection reconstruction in steady state free precession validated using 2D cine MRI. J Magn Reson Imaging 2004;20:411 416. 11. Uribe S, Muthurangu V, Boubertakh R, et al. Whole-heart cine MRI using real-time respiratory self-gating. Magn Reson Med 2007;57: 606 613. 12. Storey P, Li W, Chen Q, Edelman RR. Flow artifacts in steady-state free precession cine imaging. Magn Reson Med 2004;51:115 122. 13. Markl M, Alley MT, Elkins CJ, Pelc NJ. Flow effects in balanced steady state free precession imaging. Magn Reson Med 2003;50: 892 903. 14. Jung B, Zaitsev M, Hennig J, Markl M. Navigator gated high temporal resolution tissue phase mapping of myocardial motion. Magn Reson Med 2006;55:937 942. 15. Nezafat R, Stehning C, Gharib A, et al. Improved spatial-temporal resolution MR coronary blood flow imaging at 3T. J Cardiovasc MR 2005;7:424. 16. Stehning C, Nezafat R, Gharib A, et al. Dual navigators for timeresolved MR coronary blood flow. In: Proc 13th Annual Meeting ISMRM, Miami Beach, 2005; p 1616. 17. Scheffler K, Heid O, Hennig J. Magnetization preparation during the steady state: fat-saturated 3D TrueFISP. Magn Reson Med 2001;45:1075 1080. 18. Deshpande VS, Shea SM, Laub G, Simonetti OP, Finn JP, Li D. 3D magnetization-prepared true-fisp: a new technique for imaging coronary arteries. Magn Reson Med 2001;46:494 502. 19. Madore B, Glover GH, Pelc NJ. Unaliasing by Fourier-encoding the overlaps using the temporal dimension (UNFOLD), applied to cardiac imaging and fmri. Magn Reson Med 1999;42:813 828. 20. Alfakih K, Plein S, Thiele H, Jones T, Ridgway JP, Sivananthan MU. Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences. J Magn Reson Imaging 2003;17:323 329.