Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Similar documents
Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract.

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 1290 Infinity LC system

Analysis of bisphenol A leaching from baby feeding bottles

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

2D-LC as an Automated Desalting Tool for MSD Analysis

Edgar Naegele. Abstract

Comparison of a UPLC Method across Multiple UHPLC Systems

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

The One Minute Chromatographer

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

New HPLC Column Options for UHPLC

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Journal of Chemical and Pharmaceutical Research

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

Extended Application Note

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Online 2D-LC Analysis of Complex N-Glycans in Biopharmaceuticals Using the Agilent 1290 Infinity 2D-LC Solution

APPLICATIONS Improving Intact Biogeneric Protein Separations with Aeris WIDEPORE Core-Shell Columns

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Agilent 1260 Infinity Analytical SFC System

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases

Am I getting the very best value from my UHPLC analyses?

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

Impurity Profiling of Carbamazepine by HPLC/UV

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Aeris. Precision Engineered Core- Shell Particles for Ultra-High Resolution BioSeparations. Aeris PEPTIDE. Aeris WIDEPORE

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Pesticide Analysis using an Agilent 1290 Infinity LC System with an Agilent 6140 Single Quadrupole LC/MS System

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Application Note. Separation of clindamycin phosphate and process impurities. Summary. Introduction. Category Pharmaceutical analysis Matrix

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Two-Dimensional LC/MS/MS to Reduce Ion Suppression in the Determination of Cannabinoids in Blood Plasma

Determination of enantiomeric purity of Timolol Maleate by Supercritical Fluid Chromatography

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

ACE Excel UHPLC Columns

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Agilent 1260 Infinity Analytical Hybrid SFC/UHPLC system

Application Note. Abstract. Authors. Pharmaceutical

A Novel HILIC Column for High Speed N-linked Glycan Analysis

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

LC/MS Method for Comprehensive Analysis of Plasma Lipids

ISSN: ; CODEN ECJHAO E-Journal of Chemistry 2011, 8(3),

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Evaluation of UHPLC System Performance Under High Throughput Conditions. William Hedgepeth, Masatoshi Takahashi, Shimadzu Scientific Instruments

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Transfer and Speed-up of Methods to Fused-Core Particle Columns EPA Method 1694

Ultra High Definition Optimizing all Analytical Dimensions

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Authors. Abstract. Introduction. Environmental

Pelagia Research Library

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Authors. Abstract. Food

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

THIN LAYER CHROMATOGRAPHY

Journal of Chemical and Pharmaceutical Research

APPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit

Sue D Antonio Application Chemist Cedar Creek, TX

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Sepax Technologies, Inc.

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

A Validated Chiral Liquid Chromatographic Method for The Enantiomeric Separation of Dapoxetine Hydrochloride

Agilent 1220 Infinity LC System Performance Specifications

Quality Analysis of Extra Virgin Olive Oils Part 7

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

Application Note. Authors. Abstract. Petrochemical

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms

Transcription:

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system A practical guide on how to use parameters to increase peak capacity Application Note Pharmaceutical and Chemical Author Siji Joseph Agilent Technologies India Pvt. Ltd. Bangalore, India 100 80 Phenol N,n-diethyl-m-toluamide Peak A Edgar Naegele Agilent Technologies Waldbronn, Germany mau 60 40 20 0 Uracil Methyl paraben Ethyl paraben Propyl paraben Butyl paraben Toluene Heptyl paraben Abstract 0 2 4 6 8 10 12 Minutes Peak capacity can be used to evaluate the performance of a chromatographic separation. In this Application Note, the effects of gradient parameters, such as flow rate on peak capacity are measured and discussed to provide UHPLC instrument users with guidance for easily maximizing peak capacity. An Agilent 1290 Infinity LC was used for the study because it can deliver a broad power range, which is the integral on flow rate and pressure. The unique design of the Agilent 1290 Infinity LC Binary Pump tolerates up to 1200 bar column backpressure. This Application Note discusses the variation of peak capacity values with change in gradient parameters such as flow rate, gradient time, column length and slope of the gradient.

Introduction Peak capacity is a performance measure describing the number of peaks that can be separated during a gradient run with a certain resolution 1. The importance of higher peak capacity values is highlighted by the increasing demand for high throughput gradients for the separation of complex samples with an unknown number of various types of analytes. Generally speaking, the higher the peak capacity the higher the probability of separating all peaks in different samples. For example, the peak resolution is compromised when the number of components exceeds 1/3 of the peak capacity. Specifically, 98% of the components can only be resolved if the peak capacity exceeds the number of components by a factor of 100. 2,3 Peak capacity has an impact on: Starting point in method development, because a method with higher peak capacity can separate compounds in an unknown sample with higher probability. Impurity scouting, because a method with higher peak capacity unravels all impurities with higher probability. Generic method, because higher peak capacity can separate compounds in an unknown sample with higher probability. The following equation was used to calculate peak capacity: t P = 1+ n 1 n G 1 w Where, n is the number of peaks used for the calculation, t G is the gradient time, w P is the average peak width measured at 4σ peak height. P This Application Note, evaluates the variation of peak capacity values with change in gradient parameters such as flow rate, gradient time, column length and slope of the gradient. In addition, practical guidelines for the maximization of peak capacity are provided. A more detailed discussion of this topic is presented in Reference 4. Experimental Instrument configuration Agilent 1290 Infinity LC controlled by ChemStation (Version B.04.02) and equipped with an Agilent 1290 Infinity Binary Pump with integrated vacuum degasser, Agilent 1290 Infinity Autosampler, Agilent 1290 Infinity Thermostatted Column Compartment and an Agilent 1290 Infinity Diode Array Detector with 10 mm flow cell was used for data acquisition. Chemicals and standards Super gradient grade acetonitrile (ACN) was purchased from Lab-Scan (Bangkok, Thailand) and formic acid was purchased from Sigma-Aldrich (India) as a modifier. HPLC grade water was freshly taken from a Milli-Q water purification system. Samples for analysis were prepared by mixing the HPLC gradient system diagnostic mix (Supelco, USA. Cat No: 48271) and reverse phase test mix (Supelco, USA. Cat No: 47641-U). The individual components of the sample were uracil, phenol, methyl paraben, ethyl paraben, propyl paraben, butyl paraben, heptylparaben, toluene and N, N-diethyl-mtoluamide. Columns Agilent ZORBAX Eclipse Plus C18, 50 mm 2.1 mm, 1.8 µm Agilent ZORBAX Eclipse Plus C18, 100 mm 2.1 mm, 1.8 µm Agilent ZORBAX Eclipse Plus C18, 150 mm 2.1 mm, 1.8 µm LC Parameters The LC method parameters are tabulated in Table 1 and the gradient used for the study is tabulated in Table 2. Parameter Details Mobile phase A 0.1% Formic acid in water Mobile phase B 100% Acetonitrile Flow rate Variable (0.2 ml/min, 0.4 ml/min, 0.6 ml/min, 0.8 ml/min, 1.0 ml/min, 1.2 ml/min, 1.4 ml/min, 1.6 ml/min, 1.8 ml/min or 2.0 ml/min. Injection volume 2 µl Needle wash Flush port activated for 6 seconds using mobile phase B Column temp. 50 C Detection 254/4 nm; Reference off Post run time 3 minutes Table 1 LC method details for experiment. %B Time (min) 20 to 100 Variable (5, 10, 15, 20, 25, 30, 35, 40, 45 or 50 min) -gradient time 100 For 2 minutes column rinsing 20 For 3 minutes column reconditioning Table 2 Gradient used for experiment. 2

The binary pump was operated at variable flow rates and gradient times with a gradient range of 20 to 100 % B. The detector was operated at a sampling acquisition rate of 80 Hz for all runs in order to avoid compromise of peak shape and number of data points (response time 0.062 seconds, >0.003 m). 100 80 Phenol N,n-diethyl-m-toluamide Peak A Procedure Samples were injected and separated using 10 different gradients (Table 2). Each gradient was run with 10 different flow rates (Table 1). The use of higher flow rates was restricted by a high backpressure limit of 1200 bar when longer 1.8 µm columns were used. An Agilent ZORBAX Eclipse Plus 50 mm 2.1 mm, 1.8 µm column was used for a flow rate range of 0.2 to 2.0 ml/min (10 different trials), a 100 mm 2.1 mm, 1.8 µm column was used up to 1.4 ml/min (seven trials) and a 150 mm x 2.1 mm, 1.8 µm column was used up to 1.0 ml/min (five different trials). 4σ peak width for N,N-diethyl-m-toluamide (peak A) was measured for each gradient and used for peak capacity calculation. A representative chromatogram for the analyte using an Agilent ZORBAX Eclipse Plus C18, 100 mm 2.1 mm, 1.8 µm column is shown in Figure 1, with peak A highlighted. Results and discussion General considerations for peak capacity In order to assess the effect of flow rate on peak capacity the analyte mix was first separated using several flow rates. For the lower end of the flow rate range a peak capacity of about 100 was achieved. By increasing the flow rate up to 2 ml/min a peak capacity of about 250 was achieved for the short 5-min gradient. A flow rate of mau 60 40 20 0 Uracil Methyl paraben Ethyl paraben 0 2 4 6 8 10 12 Minutes 2 ml/min using a 5-min gradient can return approximately the same peak capacity as a 10 min gradient with a 0.8 ml/min flow rate (Figure 2). This pattern was also evaluated as peak capacity versus gradient time. The results shows that 2.0 ml/min flow rate gave a peak capacity value of Propyl paraben Figure 1 Chromatographic representation of analyte separation using an Agilent ZORBAX Eclipse Plus C18, 100 mm 2.1 mm, 1.8 µm column. Gradient time: 10 min and flow rate: 0.2 ml/min. Peak capacity 800 700 600 500 400 300 200 100 0 Butyl paraben Toluene Flow rate and gradient time versus peak capacity 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 Flow rate (ml/min) Heptyl paraben about 250 when using a 5-min gradient while the peak capacity value was 500 when the gradient time was increased to 30 min. Figure 2 illustrates that the peak capacity of a 30-min gradient at 0.4 ml/min can be similar to the peak capacity using 5-min gradient at 2.0 ml/min with the same column. 5.00 500 Gradient time (min) 30.00 Peak capacity 700.00-800.00 600.00-700.00 500.00-600.00 400.00-500.00 300.00-400.00 200.00-300.00 100.00-200.00 0.00-100.00 Figure 2 Variation of peak capacity with change in gradient time and flow rate for N, N-diethyl-m-toluamide using an Agilent ZORBAX Eclipse Plus 50 mm 2.1 mm, 1.8 µm column. 3

The combined effects of gradient time and flow rate can be illustrated in a three-dimensional plot as shown in Figure 2. The peak capacity values for peak A (N, N-diethyl-m-toluamide) with changing gradient time and flow rate are tabulated in Table 3. The results show that the highest peak capacity values were observed for the higher flow rates for each examined gradient time. Typical gradients for 50 mm 2.1 mm, 1.8 µm columns are about 5 or 10 min long and peak capacities of about 250 and 330 can be achieved, respectively. The effect of column length on peak capacity at the same flow rates but various gradient times on 50 mm, 100 mm and 150 mm 2.1 mm, 1.8 µm ZORBAX Eclipse Plus columns shows that at low flow rates (for example, 0.2 ml/min), peak capacity values with a 150 mm column for a short gradient time (5 min) was marginally better than that for a 50 mm column. With longer gradient times the longer columns achieved higher peak capacities. For example, a 150 mm column (50-min gradient at 0.2 ml/min) achieved a peak capacity value of 308 while a 100 mm column achieved a peak capacity of 264, and a 50 mm column achieved a peak capacity value of only 211. The results are tabulated in Tables 3A, 3B and 3C. For a given gradient volume, peak capacity values are higher using a shorter gradient with higher flow rates than a longer gradient at lower flow rates. For example, the peak capacity value for a 50-min gradient at 0.2 ml/min flow rate was 211, whereas for a 5-min gradient with a 2.0 ml/min flow rate the value was 246 (approximately 17% more). This proves that the efficiency for short gradients at higher flow rates can resolve the compounds in a sample by increasing the corresponding peak capacity. A) Column: 50 mm 2.1 mm, 1.8 µm particles Flow rate (ml/min) Gradient time (min) 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 5.00 90.20 134.80 164.50 185.00 211.30 211.30 227.40 227.40 227.40 246.30 10.00 126.30 179.40 219.10 246.30 268.60 281.40 295.40 310.90 320.90 328.10 15.00 145.80 211.30 253.30 285.90 316.40 340.70 354.30 369.00 385.00 385.00 20.00 162.30 231.90 288.20 328.10 357.80 380.80 407.00 421.50 437.10 453.90 25.00 174.20 250.50 307.60 351.10 388.30 421.50 447.00 475.80 491.60 508.50 30.00 185.00 264.60 334.30 385.00 421.50 465.80 491.60 520.50 536.20 570.80 35.00 190.00 279.40 350.30 413.10 458.90 503.60 529.40 573.40 589.80 625.40 40.00 200.60 295.40 374.80 437.80 491.60 536.20 575.40 620.70 655.20 693.60 45.00 204.80 309.10 396.40 465.80 520.50 577.00 617.10 663.30 698.20 736.90 50.00 211.30 321.00 415.60 491.60 556.40 614.30 670.00 719.00 755.80 796.60 B) Column: 100 mm 2.1 mm, 1.8 µm particles Flow rate (ml/min) Gradient time (min) 0.20 0.40 0.60 0.80 1.00 1.20 1.40 5.00 102.50 164.50 197.30 211.30 227.40 246.30 246.30 10.00 144.60 219.10 268.60 295.40 310.90 310.90 328.10 15.00 174.20 253.30 305.50 340.70 354.30 369.00 385.00 20.00 194.00 281.40 337.40 369.00 393.50 407.00 437.10 25.00 211.30 301.40 360.00 398.80 433.90 447.00 461.00 30.00 224.60 322.10 385.00 431.80 465.80 491.60 505.60 35.00 237.90 338.80 405.00 449.00 491.60 516.20 543.30 40.00 246.30 352.50 429.20 472.00 513.00 548.70 575.40 45.00 255.70 369.00 442.60 500.90 541.70 577.00 617.10 50.00 263.80 383.00 416.00 517.40 567.10 601.80 640.90 C) Column: 150 mm 2.1 mm, 1.8 µm particles Flow rate (ml/min) Gradient time (min) 0.20 0.40 0.60 0.80 1.00 5.00 99.10 164.50 185.00 211.30 211.30 10.00 152.00 227.40 268.60 291.40 305.40 15.00 188.90 277.00 316.40 340.70 340.70 20.00 219.10 310.90 357.80 380.80 380.80 25.00 238.40 335.50 388.30 409.90 421.50 30.00 257.00 361.50 411.80 442.60 453.90 35.00 272.10 382.60 439.40 458.90 469.30 40.00 284.70 400.20 453.90 481.60 502.10 45.00 295.40 415.00 474.10 510.50 520.50 50.00 307.60 427.60 491.60 526.70 536.20 Table 3 Tabulated data for peak capacities dependend on flow rates, gradient times and change in column length. A: Agilent ZORBAX Eclipse Plus, 50 mm 2.1 mm, 1.8 µm column. Some examples with the same gradient slope are color coded. B: Agilent ZORBAX Eclipse Plus, 100 mm 2.1 mm, 1.8 µm column. C: Agilent ZORBAX Eclipse Plus, 150 mm 2.1 mm, 1.8 µm column. 4

Adjusting the gradient time and flow rate obtains the same gradient slope and selectivity. The peak capacity using a 50 mm 2.1 mm ZORBAX Eclipse Plus column is given in Table 3A. In this study, the slope of a long gradient with low flow rate [80%/(40 min 0.2 ml/min) = 10%/mL] is equal to a short gradient with a high flow rate [80%/(5 min 1.6 ml/min) = 10%/mL]. The second gradient provided a shorter analysis time at a higher flow rate compared to the first long gradient at a lower flow rate without a change in selectivity. However, the second gradient achieved a higher efficiency. Practical considerations to achieve highest peak capacities The following section explains how to develop a method with the maximum peak capacity in six steps. 1. Consider the number of peaks and required peak capacity. The numbers of probable peaks (compounds) in the sample, that are to be separated require the peak capacity necessary to resolve all compounds as described in the introduction. As a rule of thumb the simple equation P=n 1.5 (where P is the sample peak capacity and n is the number of compounds) can be used to get an estimation of the required peak capacity. For example, a peak capacity of 300 would be required for 45 peaks 5. This is roughly achieved within a 5-min gradient at 2 ml/min (Table 3A). 2. Consider the choice of gradient length. The length of the gradient for a given separation problem depends on several factors. If high sample throughput is required shorter run times are chosen. The same is true if the results of the sample analysis is required in a short time to make a decision. Conversely, the required separation time depends on the complexity of the sample. In reality, a compromise between these requirements may be necessary. 3. Consider the choice of column. The proper column length must be selected depending on the chosen run time and the sample complexity. If a short run time below 5 min is required, a column length of 50 mm is appropriate, and will allow separations with required peak capacities of up to 250. If a moderate separation time between 5 min and 30 min is chosen, a column length of 100 mm is the appropriate choice. This allows separations of moderately complex samples with required peak capacities of up to 500. For samples of highest complexity where the longest gradient times above 30 min are required, a column length of 150 mm is the best choice. This allows separations where a peak capacity of more than 500 is required. Tables 3A, 3B and 3C display the results of the described experiments. 4. Consider the choice of the right flow rate according to system and column pressure limitations. The maximum flow rate that can give the maximum peak capacity for a chosen column and gradient length depends on the limitations of the HPLC or UHPLC system and the column used. For example, with a safety margin of 10%, the maximum available backpressure of the Agilent 1200 Series HPLC is 360 bar; the maximum achievable backpressure of the Agilent 1200 Series RRLC is 540 bar; and the maximum achievable backpressure of the Agilent 1290 Infinity LC is 1080 bar. This clearly demonstrates that the maximum peak capacity can be achieved with the Agilent 1290 Infinity LC. The column also limits the system pressure. For example, an Agilent RRHT 1.8 µm column can be used with a backpressure of up to 600 bar and an Agilent RRHD 1.8 µm column can be used for backpressures of up to 1200 bar. This makes the RRHT column the right choice for the standard Agilent 1200 Series HPLC and RRLC systems for highest resolution and peak capacity. The right solution for the Agilent 1290 Infinity LC is given with the RRHD column. The change in pressure during the gradient must also be considered to choose the right flow rate. For example, the maximum viscosity and backpressure is reached for a mixture of 10% acetonitrile in water or for a mixture of 50% methanol in water. This is where the highest pressure occurs during the gradient. 5

5. Consider the choice of temperature on peak capacity. The effect of temperature on peak capacity depends on the individual compound. In general, at elevated flow rates of more than 0.8 ml/min, elevated temperatures above 40 C deliver higher peak capacities 1. When elevating the temperature, the stability of the analyte and of the column phase must be considered. For example, ZORBAX Stable Bond (SB) phases are stable up to 90 C and ZORBAX Eclipse phases are stable up to 60 C. 6. Achieve the highest efficiency for the separation. After these decisions are made, the flow rate can be increased to work at the maximum to achieve the best separation at the highest possible peak capacity. If selectivity is important it might be necessary to work at lower flow rates to separate a given set of peaks under a desired selectivity. In this case the flow rate can still increase to its possible maximum while reducing the gradient time to preserve the gradient and the required selectivity. This means that the most efficient separation can be achieved. Table 3A shows some examples of flow rate and gradient times with the same gradient slope and peak capacity using color coding. Conclusion The results discussed in this Application Note provide a course of action for maximizing the peak capacity and improving chromatographic separation performance by varying method parameters for small molecule separations. In general, peak capacity value increases with gradient time and flow rate. The highest separation efficiency can be achieved by increasing the flow rate using the Agilent 1290 Infinity LC. This system can operate up to 1200 bar, providing the highest possible flow rate for smaller particle sized columns. A simultaneous decrease in the gradient run time while maintaining the gradient and selectivity for a given separation can be easily achieved with the Agilent 1290 Infinity LC. References 1. Patrik Petersson and et al, Maximizing peak capacity and separation speed in liquid Chromatography, J. Sep. Sci. 31 (2008) 2346-2357. 2. J.M. Davis, J.C. Giddings, Statistical theory of component overlap in multicomponent chromatograms, Anal. Chem. 55 (1983) 418. 3. J.C. Giddings, Sample dimensionality: a predictor of order-disorder in component peak distribution in multidimensional separation, J. Chromatogr. A, 703 (1995). 4. "Maximization chromatographic peak capacity with Agilent 1290 Infinity LC system by using gradient parameters" Agilent publication, Publication number 5990-6933EN. 5. Uwe Dieter. Neue, Peak Capacity in unidimensional chromatography, J. Chromatogr. A, 1184 (2008) 107-130. 6

7

www.agilent.com/chem/lc Agilent Technologies, Inc., 2010 Published December 1, 2010 Publication Number 5990-6932EN