Bilateral Total Knee Replacements using Two Different Implant Designs - Preliminary Report

Similar documents
Both Knee Re-revision Operations with Different Types of Endoprosthesis after Septic Complications

Comparison of high-flex and conventional implants for bilateral total knee arthroplasty

Analysis of factors affecting range of motion after Total Knee Arthroplasty

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Kinematics Analysis of Different Types of Prosthesis in Total Knee Arthroplasty with a Navigation System

CLINICAL AND OPERATIVE APPROACH FOR TOTAL KNEE REPLACEMENT DR.VINMAIE ORTHOPAEDICS PG 2 ND YEAR

itotal Customized POSTERIOR-STABILIZED knee replacement system

A study of functional outcome after Primary Total Knee Arthroplasty in elderly patients

Early Results of Total Knee Replacements:

Evolution. Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee. Key Aspects

Functional and radiological outcome of total knee replacement in varus deformity of the knee

Midterm results of cemented Press Fit Condylar Sigma total knee arthroplasty system

TOTAL KNEE ARTHROPLASTY (TKA)

STIFFNESS AFTER TKA PRE, PER AND POST OPERATIVE CAUSING FACTORS

Total Knee Replacement

SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY

Anterior knee pain following total knee replacement caused by isolated Paget's disease of patella

Bicruciate-Retaining or Medial Pivot Total Knee Prosthesis Pritchett 225 Fig. 3. The MP total knee prosthesis. Fig. 1. An anteroposterior radiograph o

Stephen R Smith Northeast Nebraska Orthopaedics PC. Ligament Preserving Techniques in Total Knee Arthroplasty

Unicondylar Knee Vs Total Knee Replacement: Is Less Better In the Middle Aged Athlete

Knee kinematics after TKA depends on preoperative kinematics

Presented By Dr Vincent VG An MD BSc (Adv) MPhil Dr Murilo Leie MD Mr Joshua Twiggs BEng Dr Brett A Fritsch MBBS FRACS (Orth) FAOrthA.

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year

Dora Street, Hurstville 160 Belmore Road, Randwick

ATTUNE KNEE SYSTEM: SOFCAM CONTACT

Over 20 Years of Proven Clinical Success. Zimmer Natural-Knee II System

Total Knee Original System Primary Surgical Technique

ATTUNE Knee System: Stability in Total Knee Replacement

Saiph Knee System. Technical Dossier

Conversion of unicompartmental knee arthroplasty to total knee arthroplasty : The challenges and need for augments

Functional Outcome of Uni-Knee Arthroplasty in Asians with six-year Follow-up

BIOMECHANICAL EVALUATION OF DEGREE OF FREEDOM OF MOVEMENTS OF AN NOVEL HIGH FLEXION KNEE FOR ITS SUITABILITY IN EASTERN LIFESTYLES

Complications of Total Knee Arthroplasty

EARLY CLINICAL RESULTS OF PRIMARY CEMENTLESS TOTAL KNEE ARTHROPLASTY

Abramsohn Retractor 1

Computer aided analysis of total knee replacement

BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS IN TOTAL KNEE REPLACEMENTS

THE PERSONALIZED KNEE

Frontal Plane Kinematics After Mobile- Bearing Total Knee Arthroplasty

There have been conflicting results reported in the

Management of Periprosthetic Fracture in Unicompartmental Knee Arthroplasty Patients: A Case Series

Reprint requests: Dr MA Hassaballa Avon Orthopaedic Centre, Southmead Hospital Westbury-on-Trym Bristol BS10 5NB, UK

Unicompartmental or total knee replacement

Total Knee Arthroplasty in a Patient with an Ankylosing Knee after Previous Patellectomy

Mid-Term Results of Oxford Medial Unicompartmental Knee Arthroplasty

Medium term results of the Miller-Galante unicompartmental knee arthroplasty with 10 year survivorship

Biomechanics of. Knee Replacement. Mujda Hakime, Paul Malcolm

The Egyptian Journal of Hospital Medicine (January 2019) Vol. 74 (7), Page

Section of total knee replacement. Total Knee Replacement System. Knieendoprothesen System. Système de prothèse totale de genou

The Result of PFC (press-fit condylar) Total Knee Replacement Arthroplasty*

YOU MAY NEED A KNEE REPLACEMENT

Implant Overhang after Unicompartmental Knee Arthroplasty: Oxford Prosthesis versus Miller-Galante II Prosthesis

ConforMIS, Inc. 28 Crosby Drive Bedford, MA Phone: Fax:

Stefan Rahm MD University Hospital Balgrist

Evaluation of soft-tissue balance during total knee arthroplasty

Full Function, Faster Medial-Pivot

The Cruciate Ligaments in Total Knee Arthroplasty

Mid-term Results of Total Knee Arthroplasty Using PFC Sigma RP-F

Zimmer FuZion Instruments. Surgical Technique (Beta Version)

CONTRIBUTING SURGEON. Barry Waldman, MD Director, Center for Joint Preservation and Replacement Sinai Hospital of Baltimore Baltimore, MD

Total knee replacement with rotational proximal tibial osteotomy for osteoarthritis with severe external tibial torsion and patellar instability

Mædica - a Journal of Clinical Medicine

Clinical Results of Genesis-I Total Knee Arthroplasty

Knee Revision. Portfolio

Range of Motion of Standard and High-Flexion Posterior Stabilized Total Knee Prostheses A PROSPECTIVE, RANDOMIZED STUDY

STABILITY & MOTION VS. THE PERSONA KNEE SYSTEM. The Zimmer Persona Knee

why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS

Kinematic vs. mechanical alignment: What is the difference?

Intramedullary Tibial Preparation

SPECT/CT Imaging. Knee & Hip Prosthesis

Knee Replacement Implants

PARTIAL KNEE REPLACEMENT

Soft Tissue Releases in Valgus Knees

Simplifying the Most Clinically Proven Partial Knee in the World. Oxford Partial Knee with Microplasty Instrumentation

Partial Knee Replacement

15-Year Follow-up Study of Total Knee Arthroplasty in Patients With Rheumatoid Arthritis

EMPOWERING LIFE THROUGH NATURAL MOTION

Mid- to long-term outcomes of a medialpivot system for primary total knee replacement

Clinical evaluation of 292 Genesis II posterior stabilized high-flexion total knee arthroplasty: range of motion and predictors

Hospital for Joint Diseases

Gold standard of a TKA. Conflicting goals? POLYETHYLENE WEAR THE SOLUTION: MOBILE BEARING KNEES. MOBILE BEARING A totally new approach (1977)

Knee Surg Relat Res 2013;25(1): pissn eissn Knee Surgery & Related Research

Mr Aslam Mohammed FRCS, FRCS (Orth) Consultant Orthopaedic Surgeon Specialising in Lower Limb Arthroplasty and Sports Injury

1 Introduction. 2 Data and method

Surgical Technique. Hinge Disassembly and Rebuild Technique

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination.

Think isometry Feel balance

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

The information contained in this document is intended for healthcare professionals only.

Why precision is powerful

Influence of Hip Position on Knee Flexion Angle in Patients Undergoing Total Knee Arthroplasty

Augmented Glenoid Component for Bone Deficiency in Shoulder Arthroplasty

Mark Clatworthy Middlemore Hospital Auckland New Zealand

The Evolution of Modern Total Knee Prostheses

Sulzer Orthopedics Joint & Fracture Care

Massive Varus- Overview. Massive Varus- Classification. Massive Varus- Definition 07/02/14. Correction of Massive Varus Deformity in TKR

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

High flexion knee arthroplasty: the relationship between rotational angles and flexion angle after total knee arthroplasty

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

THE KNEE SOCIETY VIRTUAL FELLOWSHIP

Transcription:

Bahrain Medical Bulletin, Vol. 27, No. 1, March 2005 Bilateral Total Knee Replacements using Two Different Implant Designs - Preliminary Report Ali Redha, MD* Abdalla A Malki FRCS Ed, MD Orth** Faisal Al-Mousawi, FRCS I, FRCS Ed** Yasser Al-Sawad, MD* Mona MY Booz, FFR-RCSI*** Objectives: To evaluate the results and advantages of using two different implant designs for total knee replacement arthroplasties. Design: Prospective randomized study. Setting: Orthopaedic Department, SMC, Bahrain. Methods: Between January 2000 and December 2002, nine patients with advanced osteoarthritis, with the indication of bilateral total condylar replacements, had one knee replaced with design 'A' and the other side with design 'B'. Results: The nine patients with designs A and B were critically analyzed to compare the results of the two designs in the same group of patients by the same medical team. The nine patients compromised of six women and 3 men with a mean age of 66.5 at the time of replacing the first knee. The follow-up period ranged from 18 to 42 Months. All patients were satisfied and showed improvement of pain, deformity and function. The mean pre-operative scores were 48.5 and 48.3 while the follow-up scores were 89.1 and 88.7 for knees with designs A and B respectively. Conclusion: The results of the knees with either design A or B are similar. Purchasing two designs will not affect the results, but it will reduce cost and improve technical support. Bahrain Med Bull 2005;27(1): Significant evolution of total knee replacement arthroplasty (TKR) took place during the last three decades. Due to improvement of the instruments, materials and implant designs, it became possible to obtain appropriate implant positioning and fixation provided accurate surgical technique is followed 1,2. * Senior Resident, Orthopaedic Department ** Consultant Orthopaedic Department *** Consultant, Radiology Department Salmanyia Medical Complex Kingdom of Bahrain 1

Many of the currently marketed designs provide painless, stable and mobile knee for more than 10 and 15 years 3,4. Some of the early-introduced designs were further developed, modified over the years and dominated the market. Domination of few manufacturers created several problems. One of the problems is a state of monopoly that in turn increased the cost and on occasions delayed the delivery of implants 5. With hundreds of different implant designs available in the medical market, surgeons, especially in the developing countries may find it difficult to choose an implant design. The appropriate selection should be based on implant s developmental history, instruments reliability, independent published results and price 6. In our department a total condylar design A has been in use since the early 1990 s. A state of monopoly occurred and reflected on the cost, availability and technical support. In 1999 a second total condylar design B was selected and introduced to our unit. Both designs A and B have documented histories of development and published results 7-9. The aim of this paper is to compare the results of the two implants ( A & B ) when used for bilateral knee replacement arthroplasties by the same surgical team. METHODS Between January 2000 and December 2002, all patients with osteoarthritic knees with the indication of bilateral TKR were randomized for using two different implant designs A and B. During that period, nine patients had design A on one knee and design B on the other. The decision for surgery was based on severity of pain, functional disability and radiological changes (Figure 1). Figure 1. Radiograph of osteoarthritic knees showing decreased joint spaces, osteophyes, sub-chondral sclerosis and varus deformity All the knees were explored through straight mid-line skin incision and curved parapatellar incision for the muscles and capsule. The preparation of the tibial and femoral condyles was performed with the instrument systems for either design A or B. The appropriate size trial implants were positioned and tested for alignment, range of movement and stability. The appropriate pre-sterile components were selected and cemented. The patella was not replaced in any of the knees. Wound was closed in layers and drained. Prophylactic antibiotics and anticoagulants started on the day of surgery and continued for three to five days. Standard postoperative programme was followed for mobility and rehabilitation of the patients. 2

All the patients were assessed before and after surgery according to the Knee Function Assessment Chart of the British Orthopaedic Association 10. For the sake of comparison on reporting the results of this study, the assessment was transformed to the 100 points Bristol Scoring System that resembles that of the Hospital for Special Surgery 11. RESULTS The nine patients with design A on one knee and design B on the other were six women and three men. The mean age at the time of replacing the first and second knees was 66.5 (59 to 73) and 67.1 (60 to 74) respectively. The sizes of the implants required for the patients were mostly of small and medium sizes. One patient required large size. The thickness of the tibial inserts required was 8 and 10 mm. At a follow-up period of 18 to 42 months, the nine patients were satisfied with the results on both knees. On assessment, there were improvements in pain; movement and deformities.the improvements in the overall scores were remarkable as demonstrated in Table 5. Pre-operatively, the pain was severe in all patients. Post-operatively, one knee with B design had moderate pain while the other knees from both groups were either pain free or with mild ache (Table 1). Table 1. Pre- and post-operative pain Pain Pre-operative Post-operative None/ mild 0 0 9 8 Moderate 1 0 0 1 Severe 8 9 0 0 Total 9 9 9 9 Most of the patients were satisfied with the range of movement obtained following surgery. Pre-operatively, two knees from group A and three from group B had range of flexion more than 100 degrees while the other knees had flexion less than 100 degrees. Post-operatively eight knees from each group had flexion more than 100 degrees (Table 2). Fifty per cent of the knees reached 120 degrees of flexion. Table 2. Pre- and post-operative flexion Flexion/degrees Pre-operative Post-operative 100 or more 2 3 8 8 81 to 100 6 6 1 1 60 or less 1 0 0 0 Total 9 9 9 9 3

Pre-operatively, most of the knees had flexion deformity more than 10 degrees. Postoperatively, one patient from each group had flexion deformity more than 10 degrees (Table 3). Table 3. Pre- and post-operative flexion deformity Flexion deformity Pre-operative Post-operative 30 degrees or more 1 1 0 0 21 to 30 1 1 0 0 11 to 20 6 5 1 1 Less than 10 1 2 8 8 Total 9 9 9 9 Figure 2. Radiograph of knees following total knee replacement that demonstrates appropriate positioning of implants in valgus. The right knee shows slight varus tilt associated with pre-existing rotational deformity All the knees showed improvement of the alignment after operation. Pre-operatively all the knees were in varus. After operation (Figure 2), the knees became within the required valgus angle (Table 4). Table 4. Pre- and post-operative varus/ valgus angle Angle Pre-operative (varus) Post-operative (valgus) 20 degrees or more 1 1 0 0 11 to 20 6 5 0 0 Less than 10 2 3 9 9 Total 9 9 9 9 All patients admitted improvement after surgery. However, patients with flexion less than 100 degrees expressed concern as it restricted their socio-religious requirements, especially with preying and squat setting. Clinical deep vein thrombosis occurred once in each group. A knee from group B developed wound haematoma that required aspiration and another knee from group 4

A showed delayed wound healing. None of these complications affected the followup scores. Scores out of 100 pints: the mean pre- and post-operative follow-up scores were 48.5 and 89.1 for group A, and 48.3 and 88.7 for group B respectively (Table 5). Table 5. Pre- and post-operative scores (out of 100 points) Knee score Pre-operative Post-operative 1 46 47 90 91 2 50 50 91 91 3 52 51 93 92 4 48 46 84 83 5 47 47 92 90 6 51 50 82 81 7 48 50 84 86 8 46 46 94 93 9 49 48 92 92 -- -- -- -- Mean score 48.5 48.3 89.1 88.7 DISCUSSION The objectives of surgical treatment of osteoarthritic knee are to relieve pain, correct deformity and to restore a mobile knee. In advanced stages of the disease, these objectives are achieved by knee replacement surgery. Total replacement of both femoral and tibial articulating surfaces with a simple metal hinge implant developed in the 1950 s. The design showed early satisfactory results. However, the knee joint is more complex than a simple hinge. The normal knee permits few degrees of rotation, sliding, and abduction-adduction, besides the wide range of flexion-extension. Therefore, the concept of simple hinge implants failed to provide long-term satisfactory results especially in active patients 12-14. Significant evolution of knee replacement arthroplasty took place in the 1970 s when the concept of low friction arthroplasty was applied by using a metal femoral component and a polyethylene tibial component. Further development provided metal-backed ultrahigh density polyethylene tibial component 15-17. Soon after the evolution of TKR in the 1970 s, the importance of reproducible correct positioning of the implants with balanced ligaments was recognized. Improvement in surgical techniques and developments of instrumentations for current implants were of great help to achieve these objectives. Both design A and B used in this study have reliable instrumentation systems. Many of the currently marketed implants give acceptable results. However, it is possible that many of the implants are selected and purchased without sound basis. 5

There are criteria for selection of implants including results, reliability and cost 6,18. To facilitate comparing the results of different implants, there is a need for universal system for assessment and scoring. Despite this requirement, there are more than thirty different systems for assessments identified in the literature 19. In this study an easy to use system of scoring was used. Many published reports demonstrated comparison between different replacement arthroplasties in different patients and few were compared in the same patient 20-23. In our department we have been using the same implant design for bilateral TKA until this study started. The current study compared the results of two different implant designs used in the same group of patients where the operations were performed by the same surgical team. Since our early days of performing knee replacement arthroplasties, we have been adopting the Knee Function Assessment Chart of British Orthopaedic Association that was introduced in 1978 10,24. For the sake of comparing results of different implant designs, we transformed our assessment to a system with scoring out of 100 points 11. The replaced knees in this study showed acceptable tibio-femoral valgus angle within 10 degrees. The recommended tibio-femoral valgus angle following TKR range is 7 degrees 25. However, there are variations related to the gender, reliability of the instruments and surgeons judgment. A valgus angle between 4 and 10 degrees is acceptable 26,27. Besides achieving acceptable alignment, the results showed reduction of pain, improved movement and function. The mean pre-operative score of approximately 49 points improved on follow-up to 89.1 for design A and 88.7 for design B. The follow-up scores are comparable to other published reports from specialized centers 8,9,28. CONCLUSION Total Knee Replacement arthroplasty is a well-established technique for treatment of joints with advanced osteoarthritic changes. Many of the currently marketed implant designs if appropriately selected can give equally good results. Based on appropriate selection criteria, a second total condylar implant design was introduced to our department. That achieved a state of competition between the manufacturers with positive impact on technical support, delivery and reduction of cost. In this study, we demonstrated that the results and scoring of both designs are equally good. The drawbacks of this study are the small number of patients and the short-term follow-up period. REFERENCES 1. Robertsson O, Knutson K, Lewold S, et al. The Swedish knee Arthroplasty Register 1975-1997. Acta Orthop Scand 2001;72:503-13. 6

2. Uehara K, Kadoya Y, Kobayashi A, et al. Bone anatomy and rotational alignment in total knee arthroplasty. Clin Orthop 2002;402:196-201. 3. Khaw FM, Kirk LM, Morris RW, et al. A randomized controlled trial of cemented versus cementless press-fit condylar knee replacement. Ten-year survival analysis. J Bone Joint Surg 2002;84-B:658-66. 4. Parker DA, Rorabeck CH, Bourne RB. Long-term follow-up of cementless versus hybrid fixation for total knee arthroplasty. Clin Orthop 2001;388:68-76. 5. Malki A, Dhaif B, Al-Mousawi F. Reducing hospital expenditures: Experience with orthopaedic implants. Bahrain Med Bull 2003;25:64-7. 6. Sharkey PF, Sethuraman V, Hozack WJ, et al. Factors influencing choice of implants in total hip arthroplasty and total knee arthroplasty. J Arthroplasty 1999; 14:281-7. 7. Walker S. Design of kinemax total knee replacement. Bearing surfaces. Acta orthop Belg 1991;57 Suppl 2:108-13. 8. Back DL, Cannon SR, Hilton A, et al. The kinemax total knee arthroplasty. Nine years experience. J Bone Joint Surg 2001;83-B:359-63. 9. Martin SD, McManus JL, Scott RD, et al. Press-fit condylar knee arthroplasty. 5-9 year follow-up evaluation. J Arthroplasty 1997;12:603-14. 10. Aichroth P, Freeman MAR, Smillie IS, et al. A knee function assessment chart. J Bone Joint Surg 1978; 60-B:308-9. 11. Ackroyed CE, Whitehouse SL, Newman JH, et al. A comparative study of the medial St George Sled and Kinematic total knee arthroplasties. J Bone Joint Surg 2002;84-B:667-72. 12. Shiers LG. Arthroplasty of the knee; preliminary report of new method. J Bone Joint Surg 1954;36-B:553-60. 13. Watson JR, Wood H, Hill RC. The Shiers arthroplasty of the knee. J Bone Joint Surg 1976;58-B:300-4. 14. Cameron HU, Hu C, Vyamont D. Hinge total knee replacement revisited. Can J Surg 1997;40:278-83. 15. Rittman N, Kettelkamp DB, Pryor P, et al. Analysis of knee motion walking for four types of total knee implants. Clin Orthop 1981;155:111-7. 16. Papachristou G. Experience with knee arthroplasty. Clin Orthop 1981;246:81-5. 17. Sorrells RB. The clinical history and development of low contact stress total knee arthroplasty. Orthopaedics 2002;25:207-12. 18. Liow RYL, Murray DW. Which primary total knee replacement. A review of currently available TKR in the UK. Ann R Coll Surg Engl 1997;79:335-40. 19. Drake BG, Callahan CM, Dittus RS. et al. Global rating systems used in assessing knee arthroplasty outcome. J Arthroplasty 1994;9:409-17. 20. Title CI, Rodriguez JA, Ranawat CS. Posterior cruciate-sacrificing versus posterior cruciate-substituting total knee arthroplasty: a study of clinical and functional outcomes in matched patients. J Arthroplasty 2001;16:409-14. 21. Walker PS, Manktelow AR. Comparison between a constrained condylar and a rotating hinge in revision knee surgery. Knee 2001;8:269-79. 22. Pritchett JW. Patient preferences in knee prostheses. J Bone Joint Surg 2004; 86:979-82. 23. Ranawat CS, Rossi R, Loreti I, et al. Comparison of the PFC Sigma fixed-bearing and rotating platform total knee arthroplasty in the same patient: short-term results. J Arthroplasty 2004;19:35-9. 24. Malki A, Wrighton JD. Compartmental knee replacement. J Bone Joint Surg 1982; 64-B:246. 7

25. Jonsson B, Astrom J. Alignment and long-term clinical outcome of semiconstrained knee prosthesis. Clin Orthop 1988;226:124-8. 26. Mahaluxmivala J, Bankes MJ, Nicolai P, et al. The effect of surgeons experience on component positioning in 673 press-fit condylar posterior cruciate-sacrificing total knee arthroplasties. J Arthroplasty 2001;16:635-40. 27. Bankes MJ, Back DL, Cannon SR, et al. The effect of component mal-alignment on the clinical and radiological outcome of the kinemax knee replacement. Knee 2003;10:55-60. 28. Maruyama S, Yoshiya S, Matsui N, et al. Functional comparison of posterior cruciate-retaining versus posterior stabilized total knee arthroplasty. J Arthroplasty 2004;19:349-53. 8