Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Similar documents
Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

ICD: Basics, Programming and Trouble-shooting

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

Tachycardia Devices Indications and Basic Trouble Shooting

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

Pause-induced Ventricular Tachycardia: Clinical Characteristics

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Troubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network

ICD leads survival and troubles in the last 10 years

Primary prevention ICD recipients: the need for defibrillator back-up after an event-free first battery service-life

Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure

Clinical and Electrocardiographic Characteristics of Patients with Brugada Syndrome: Report of Five Cases of Documented Ventricular Fibrillation

Destructive Device Removal - Sparks and Deletion of Therapy History From an Implantable Cardioverter Defibrillator. Case Reports

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

Supplemental Material

New generations pacemakers and ICDs: an update

Ventricular tachycardia Ventricular fibrillation and ICD

Inappropriate electrical shocks: Tackling the beast

Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis

Arrhythmias Focused Review. Who Needs An ICD?

A 'shocking' case rectified

Shock Reduction Strategies Michael Geist E. Wolfson MC

Chapter 14. Risk of Failure of Transvenous Implantable Cardioverter Defibrillator Leads.

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

Advanced ICD Concepts

Long-Term Prognosis in Recipients of Implantable Cardioverter-Defibrillators for Secondary Preventions in Taiwan A Multicenter Registry Study

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Tech Corner. ATP in the Fast VT zone

Introduction. CLINICAL RESEARCH Clinical Trial Design. Mohammad Saeed 1 *, Mehdi Razavi 1, Curtis G. Neason 2, and Simona Petrutiu 2. Aims.

MEDICAL POLICY Cardioverter Defibrillators

The Immediate Reproducibility of T Wave Alternans During Bicycle Exercise

A case of Brugada syndrome coexisting with vasospastic angina: Caution should be taken when using calcium channel blockers

Cover Page. The handle holds various files of this Leiden University dissertation.

Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads

Subcutaneous implantable cardioverter-defibrillator (S-ICD)

Pacing Lead Implant Testing. Document Identifier

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

Journal of the American College of Cardiology Vol. 37, No. 2, by the American College of Cardiology ISSN /01/$20.

Update on Device Innovation (S-ICD, Wearable, Leadless)

Management of RT patients with implanted cardiac devices: From recommendation to implementation

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany

Practice Questions.

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Teaching Rounds in Cardiac Electrophysiology

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Pacemaker-ICD/Drug Interaction

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm

The Multicenter Unsustained Tachycardia Trial (MUSTT)

Case presentation: A 14-yearold

Riata Lead Extraction. Thomas D. Callahan, MD, FACC, FHRS

Different indications for pacemaker implantation are the following:

John H. Ip, M.D FACC Associate Professor of Medicine Michigan State University Medical Director, EP Service

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

QRS Duration Does Not Predict Occurrence of Ventricular Tachyarrhythmias in Patients With Implanted Cardioverter-Defibrillators

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

Experience with pectoral versus abdominal implantation of a small defibrillator

Sudden cardiac death (SCD) is a

Supplementary Online Content

Troubleshooting pacemakers and implantable cardioverter-defibrillators David L. Scher

SPORTS AND EXERCISE ADVICE IN PATIENTS WITH ICD AND PPM

ICD Diagnostics: Making the most of it

Atrial Defibrillation With a Transvenous Lead A Randomized Comparison of Active Can Shocking Pathways

Defibrillation threshold testing should no longer be performed: contra

CIEDs Infection: Lead Extraction, First or Last option?

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm

Dual-Chamber Implantable Cardioverter-Defibrillator

ICD THERAPIES: are they harmful or just high risk markers?

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement

Case Report. Faculty of Medicine, Oita University 2 Department of Cardiology, Hakuaikai Hospital

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Riata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Advanced ICD Troubleshooting: Part II REVIEW

Continuous ECG telemonitoring with implantable devices: the expected clinical benefits

Original Article Complications and Mortality of Single Versus Dual Chamber Implantable Cardioverter Defibrillators

Reduction in Defibrillator Shocks With an Implantable Device Combining Antitachycardia Pacing and Shock Therapy

Riata lead extraction- a single centre experience

20 ng/ml 200 ng/ml 1000 ng/ml chronic kidney disease CKD Brugada 5 Brugada Brugada 1

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012

G Lin, R F Rea, S C Hammill, D L Hayes, P A Brady

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

PARAD/PARAD+ : P and R Based Arrhythmia Detection

CLINICAL INVESTIGATIONS. Methods

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

Device Update Implantable Cardioverter Defibrillator (ICD) 박상원

Several studies of the primary prevention. Original Research

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

Diagnostic capabilities of the implantable therapeutic systems

Panagiotis N. Margos MD, Rolf Schomburg MD, Jorg Kynast MD, Ahmed A. Khattab MD, Gert Richardt MD.

HRC Carole Joyce. Bradford Royal Infirmary. Senior Chief Cardiac Physiologist. Pacing & Invasive Services.

Implantable Cardioverter Defibrillator Shocks: A Troubleshooting Guide

The implantable cardioverter defibrillator is not enough: Ventricular Tachycardia Catheter Ablation in Patients with Structural Heart Disease

Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision

When to implant an ICD in systemic right ventricle?

Transvenous Pacemaker Implantation 22 years after the Mustard Procedure

Transcription:

Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE, 1 MD, Satoru KOMURA, 1 MD, Hirotaka SUGIURA, 1 MD, Yasutaka TANABE, 1 MD, Hiroshi FURUSHIMA, 1 MD, Satoru FUJITA, 1 MD, Yuji OKURA, 1 MD, and Yoshifusa AIZAWA, 1 MD SUMMARY We describe here the case of a 58-year-old female patient who experienced inappropriate shocks from her implantable cardioverter-defibrillator (ICD). Stored electrograms from her ICD showed high frequency noise preceding the shock. Although the pacing threshold was normal and lead fracture was not found in chest X-rays, pacing lead impedance decreased to 480Ω. Moreover, such high frequency noise was observed by electrogram telemetry, but not by routine evaluation every 3 months. ICD lead dysfunction was suspected, so we elected to replace the ICD lead system. At the time of the operation, lead impedance was 410Ω and pacing threshold was the same as it was at the time of the ICD implantation, and no lead insulation disturbances were observed in the generator pocket. However, manipulation of the lead system produced high frequency noise reproducibly. Since some of the ICD lead dysfunction initially was clinically silent at rest, dysfunction was difficult to detect before serious problems occurred. Therefore, more careful evaluation of the ICD lead system is needed during long-term follow-up of ICD implants. (Int Heart J 2005; 46: 909-913) Key words: Implantable cardioverter defibrillator, Inappropriate shock, Lead failure THE implantable cardioverter-defibrillator (ICD) is widely used for patients with malignant ventricular tachyarrhythmias and can prevent sudden cardiac death. 1,2) Transvenous lead systems have decreased the length of hospital stay, morbidity, and mortality in patients with ICD. However, they are imperfect and ICD lead dysfunction has been reported. 3-7) ICD lead dysfunction may result in failure of the ICD therapy for ventricular tachycardia (VT) or fibrillation (VF) and lead to syncope or sudden death. Lead dysfunction may also cause inappropriate shocks and subsequent psychological distress. In addition, some of the lead dysfunction initially is clinically silent, so it is difficult to find lead dysfunction From the 1 Division of Cardiology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, Japan. Address for correspondence: Takashi Washizuka, MD, Division of Cardiology, Niigata University Graduate School of Medical and Dental Sciences, Asahimachi 1-754, Niigata 951-8510, Japan. Received for publication February 2, 2005. Revised and accepted July 4, 2005. 909

910 WASHIZUKA, ET AL Int Heart J September 2005 before serious problems occur. Thus, ICD lead dysfunction is one of the most important problems in patients with ICD. Nevertheless, reports of ICD lead dysfunction are rare in Japan. We report here a case of inappropriate shocks caused by oversensing due to ICD lead dysfunction. CASE REPORT The patient was a 58-year-old female who suffered a myocardial infarction in 1994. In February 1997, she had near syncope and sustained VT at a rate of 210 bpm. At electrophysiological study, 5 morphologically distinct monomorphic VTs including clinical VT were induced, and they were still inducible after administration of procainamide, flecainide, or amiodarone. In April 1997, a transvenous ICD system (Medtronic Micro Jewel, 7223, model 6936 right ventricular screw-in lead, Medtronic Inc., Minneapolis, MN, USA) was implanted through the cephalic vein. At the operation, shocks at 20 J were twice confirmed to be capable of terminating VF induced by T-wave shocks, and delivery of 5 J was set as the first treatment of VT and 30 J was set as the first treatment of VF. The amplitude of the ventricular electrogram from the tips of the defibrillation lead was 8.2 mv during sinus rhythm and the pacing threshold at the site was 4.0 V at a pulse width of 0.20 msec with an impedance of 809Ω. In October 1997, she Figure 1. Stored electrocardiogram recordings on admission. Intracardiac electrograms 1 (Vtip to Vring) showed high frequency noise during sinus rhythm. The patient received an inappropriate shock in response to oversensing (DC1), then received a total of 4 inappropriate shocks in response to oversensing. EGM = intracardiac electrograms.

Vol 46 No 5 INAPPROPRIATE SHOCK DUE TO LEAD FAILURE 911 Figure 2. Real-time telemetered intracardiac electrogram recordings. High frequency noise seen in intracardiac electrograms (Vtip to Vring) is not observed in surface electrocardiograms. experienced one spontaneous shock and VT was successfully detected and terminated by the device. In January 2001, the device had reached its elective replacement indicator, a charge time of 20 seconds. In February 2001, the device was replaced by a GEM ICD (7227Cx, Medtronic Inc.). At the time of the operation, the pacing threshold was 4.0 V at a pulse width of 0.50 msec with an impedance of 694Ω, so the lead system was not changed. Shocks at 25 J were twice confirmed to be capable of terminating VF induced by T-wave shock. After the operation, a routine evaluation every 3 months at our out-patient ICD clinic revealed no spontaneous shock, and the pacing threshold and amplitude of the sensed R wave did not change. In November 2002, she experienced several shocks during sitting. Interrogation of the device revealed 4 shocks of 35 J, and retrieval of the stored electrogram showed that high frequency noise preceded the shock (Figure 1). Although the pacing threshold did not change and lead fracture was not obvious in chest X-rays, the pacing lead impedance decreased to 480Ω. Moreover, such high frequency noise was observed on the electrogram telemeter (Figure 2). Since dysfunction of the sensing lead was suspected, we elected to implant a new lead system. At the time of the operation, lead impedance was 410Ω and pacing threshold did not change, and no lead insulation disturbances were observed in the generator pocket. However, manipulation of the lead system produced high frequency noise reproducibly. Thus, we implanted a new ICD lead system (Medtronic model 6943 right ventricular screw-in lead). Using the new lead system, the pacing threshold was 0.5 V at a pulse width of 0.50 msec with an impedance of 1339Ω, and the amplitude of the ventricular electrogram from the tips of the defibrillation lead was 9.0 mv during sinus rhythm. Although we attempted to extract the old lead system, it could not be extracted because of adhesion to the vessel wall. Therefore, we were unable to determine the exact site of lead failure. After the operation, neither inappropriate discharge nor high frequency noise was observed by telemetry during the ensuing 6 months.

912 WASHIZUKA, ET AL DISCUSSION Int Heart J September 2005 In this patient, oversensing of high frequency noise was observed and it caused inappropriate shocks. Lead dysfunction was suspected as the cause of oversensing since the lead impedance decreased and high frequency noise was observed by electrogram telemetry in real time, although the site of lead fracture was undetermined on the chest X-rays or during ICD lead replacement. Lead failure is one of the most serious complications of defibrillator systems since it may result in sudden cardiac death if VT or VF is not detected or not terminated by the ICD. In addition, lead problems often result in oversensing and inappropriate shocks. The relatively high failure rate of ICD leads has been recently reported by others, including the manufacturer. 5-9) Luria, et al 5) and Hauser, et al 6) have both reported failure rates for the ICD lead of close to 20% at four years. More recently, Ellenbogen, et al 7) reported that the failure probability of the polyurethane lead, which is the same as the lead that failed in our patient, was 37% after a mean follow-up period of 68.6 months. The subclavian crush syndrome, which was caused by compression of a lead between the first rib and clavicle, has been reported as one of the major causes of lead failure in pacemakers and ICDs. In previous reports, it occurred in 0.9-1.8% of ICD patients, 10,11) and the occurrence of subclavian crush syndrome increases significantly when subclavian venipuncture was performed compared to cephalic venous cutdown at lead insertion. In this case, lead insertion was performed by cephalic venous cutdown, so it would not seem that subclavian crush syndrome was the cause of the lead failure. The main consequence of lead failure found in all studies was oversensing, as we found in this case. The manufacturer's 8) analysis also showed that the polyurethane lead is vulnerable to structural damage. The outer and middle insulation layers and the high voltage coil account for 79% of the confirmed defects. This finding suggests that the 80A polyurethane insulation is the cause of low lead durability. However, routine follow-up may not detect some lead failures because 1) the defect is partial and therefore the lead problems may be intermittent, and 2) routine follow-up (usually every 3 months) may be insufficient. As previously reported, 9) ICD lead insulation failure was usually associated with normal values for impedance, sensing and pacing threshold at the time of lead failure. To detect these insulation failures, Dorwarth, et al 9) showed the importance of short RR interval (< 140 msec) nonsustained VT in stored intracardiac electrograms. They reported that 31 of 261 ICD patients experienced a lead related sensing failure during a mean follow-up period of 4.0 ± 2.6 years, and in 21 of 31 patients device interrogation showed short RR interval nonsustained VT 40 ± 43 days before clinically relevant lead failure. Therefore, if short RR interval nonsustained VT is

Vol 46 No 5 INAPPROPRIATE SHOCK DUE TO LEAD FAILURE 913 recorded at routine ICD follow-up, the device system should be checked more carefully. Moreover, because the survival rate of polyurethane leads decreases 4 years after implantation, routine follow-up should be performed every month in these patients, although reports of ICD lead failure are rare in our country. In conclusion, we have described a case of inappropriate shocks due to lead failure. In ICD patients who have polyurethane leads, more frequent evaluation of the ICD lead system is needed. REFERENCES 1. AVID investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Med 1997; 337: 1576-83. 2. Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med 1999; 341: 1882-90. 3. Lawton JS, Wood MA, Gilligan DM, Stambler BS, Damiano RJ Jr, Ellenbogen KA. Implantable transvenous cardioverter defibrillator leads: the dark side. Pacing Clin Electrophysiol 1996; 19: 1273-8. (Review) 4. Degeratu FT, Khalighi K, Peters RW, Shorofsky SR, Gold MR. Sensing lead failure in implantable defibrillators: a comparison of two commonly used leads. J Cardiovasc Electrophysiol 2000; 11: 21-4. 5. Luria D, Glikson M, Brady PA, et al. Predictors and mode of detection of transvenous lead malfunction in implantable defibrillators. Am J Cardiol 2001; 87: 901-4. 6. Hauser RG, Cannom D, Hayes DL, et al. Long-term structural failure of coaxial polyurethane implantable cardioverter defibrillator leads. Pacing Clin Electrophysiol 2002; 25: 879-82. 7. Ellenbogen KA, Wood MA, Shepard RK, et al. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol 2003; 41: 73-80. 8. Medtronic Inc. ICD Lead Survival Analysis. Tachyarrhythmia Product Performance Report. 2nd ed. Fridley, MN: Medtronic; 2001. 9. Dorwarth U, Frey B, Dugas M, et al. Transvenous defibrillation leads: high incidence of failure during longterm follow-up. J Cardiovasc Electrophysiol 2003; 14: 38-43. 10. Roelke M, O'Nunain SS, Osswald S, Garan H, Harthorne JW, Ruskin JN. Subclavian crush syndrome complicating transvenous cardioverter defibrillator systems. Pacing Clin Electrophysiol 1995: 18: 973-9. 11. Antonelli D, Rosenfeld T, Freedberg NA, Palma E, Gross JN, Furman S. Insulation lead failure: is it a matter of insulation coating, venous approach, or both? Pacing Clin Electrophysiol 1998; 21: 418-21.