AbstractID: 8073 Title: Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Seattle, WA May 2008

Similar documents
Dosimetry of the Gamma Knife TM

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Gamma Knife Radiosurgery A tool for treating intracranial conditions. CNSA Annual Congress 2016 Radiation Oncology Pre-congress Workshop

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Strategies and Technologies for Cranial Radiosurgery Planning: Gamma Knife

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS

Dosimetry, see MAGIC; Polymer gel dosimetry. Fiducial tracking, see CyberKnife radiosurgery

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

Overview of MLC-based Linac Radiosurgery

Otolaryngologist s Perspective of Stereotactic Radiosurgery

ph fax

Gamma Knife Radiosurgery

Staged-Volume Radiosurgery of Large AVMs

Stereotactic Radiosurgery

AAPM TG178: Towards the Recommendation of a Standard Dosimetry Protocol for Gamma Knife Stereotactic Radiosurgery

Chapter 5 Section 3.1

SRS Uncertainty: Linac and CyberKnife Uncertainties

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

Quality assurance in external radiotherapy

Leksell Gamma Knife Icon. Treatment information

Gamma. Knife PERFEXION. care. World-class. The Gamma Knife Program at Washington Hospital

Credentialing for the Use of IGRT in Clinical Trials

Concepts of Uncertainty and their Application in Gamma Knife SRS. David J. Schlesinger, Ph.D. University of Virginia

Advanced Tumor Treatment

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Work partially supported by VisionRT

Background Principles and Technical Development

Let s speak the same language: Standardization in Terminology

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Leksell Gamma Knife Icon A New User s Perspective

Gamma Knife radiosurgery with CT image-based dose calculation

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies

Specialised Services Policy: CP22. Stereotactic Radiosurgery

Introduction to Cranial Neurosurgery and Advanced Radiosurgery

SRS Plan Quality and Treatment Efficiency: VMAT vs Dynamic Conformal ARCs

Stereotactic Radiosurgery for Intracranial Metastases

AAPM Annual Meeting. ACR Accreditation Update in CT

Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Unrivaled, End-to-End

Selected radiosurgery cases from the Rotating Gamma Institute Debrecen, Hungary

Original Date: April 2016 Page 1 of 7 FOR CMS (MEDICARE) MEMBERS ONLY

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Stereotactic Radiosurgery/Fractionated Stereotactic Radiotherapy for Acoustic Neuroma (Vestibular Schwannomas)

Risk of extracranial secondary cancer after radiosurgery: comparison of Gamma Knife Perfexion, Cyberknife and Novalis

Quality Assurance for Gamma Knives

EORTC Member Facility Questionnaire

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

X-Ray Guided Robotic Radiosurgery for Solid Tumors

Stereotactic Diffusion Tensor Tractography For Gamma Knife Stereotactic Radiosurgery

Use of the Leksell gamma knife C with automatic positioning system for the treatment of meningioma and vestibular schwannoma

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Eric E. Klein, Ph.D. Chair of TG-142

Quality Assurance in Stereotactic. Radiotherapy. Swedish Cancer Institute Seattle, WA

Image Fusion, Contouring, and Margins in SRS

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

pat hways Medtech innovation briefing Published: 16 May 2018 nice.org.uk/guidance/mib147

IMAGE-GUIDED RADIOSURGERY USING THE GAMMA KNIFE

THE GOLD STANDARD IN INTRACRANIAL RADIOSURGERY

Effect of skull contours on dose calculations in Gamma Knife Perfexion stereotactic radiosurgery

Leksell Gamma Knife Icon contributes to significant growth, greater efficiency and an enhanced patient experience

Dose differences between the three dose calculation algorithms in Leksell GammaPlan

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Collection of Recorded Radiotherapy Seminars

Public Statement: Medical Policy Statement:

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Quality Assurance in Radiation Therapy

brain SPINE 2 SRS Matures into breast lung spine LUNG Dr. Robert Timmerman Discusses SBRT for Inoperable Lung Cancer BRAIN

Proton Stereotactic Radiotherapy: Clinical Overview. Brian Winey, Ph.D. Physicist, MGH Assistant Professor, HMS

Guideline & Reports 医学物理学会教育委員会資料

SRS/SBRT Errors and Causes

FLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system

CT Quality Control Manual FAQs

Nuclear Medicine and PET. D. J. McMahon rev cewood

Image Guided Stereotactic Radiotherapy of the Lung

LONG-TERM FOLLOW-UP OF ACOUSTIC SCHWANNOMA RADIOSURGERY WITH MARGINAL TUMOR DOSES OF 12 TO 13 Gy

Clinical Quality Assurance for Particle Therapy Plus ça même Plus

Global Radiosurgery 2011 All Rights Reserved 1

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA

Brain Tumor Radiosurgery. Gabor Fichtinger, PhD

Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Cyberknife Radiotherapy for Vestibular Schwannoma

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Cyberknife Stereotactic Treatment

SUPPLEMENTAL MATERIAL

iviewdose Confidence and assurance in dose delivery

Stereotactic Radiotherapy for Acoustic Neuromas (CyberKnife) UHB is a no smoking Trust

Radiosurgery of the Brain (stereotactic)

Transcription:

Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery Steven J. Goetsch, Ph.D., FAAPM ACMP 25 th Annual Meeting Seattle, WA May 2008 1. Brief Overview of Gamma Knife First invented in 1968 in Stockholm Commercially introduced in US in 1987 Cobalt-60 sources (roughly 30Ci each) at 40cm SSD with 4, 8, 14 and 18mm diameters APS introduced in 2001 Gamma Knife Perfexion introduced in 2007 350,000 patients treated worldwide thru 2005 Review of Models Gamma Knife Model 4C Rotating Gamma Units New manufacturers: GammaARM Orbiter GammaStar Cancer Care International Shenzen, China 42 Co-60 sources, 6000Ci 6500 Ci Co-60 Total body capability 4D chair 2. Traditional Gamma Knife QA Numerous clinical papers over the years See special QA of Radiation Therapy issue of Red Journal for detailed list AAPM Report 54 Stereotactic Radiosurgery (Task Group 42): Published in 1995, covered Gamma Knife, linac and protons

Knife Radiosurgery S. Goetsch, Ph.D. Page 2 TG 54 report needs to be updated Other Gamma Knife QA Resources ASTRO/AANS Consensus Statement on stereotactic radiosurgery quality improvement, 1993 RTOG Radiosurgery QA Guidelines, 1993 European Quality Assurance Program on Stereotactic Radiosurgery, 1995 DIN 6875-1 (Germany) Quality Assurance in Stereotactic Radiosurgery/Radiotherapy AAPM Task Group 68 on Intracranial stereotactic positioning systems, 2005 This is what you don t want: Patient Speaks Out After Wrong Side Of Brain Treated Headline from newspaper in Midwest, October, 2007 The Radiosurgery process See patient in clinic, assess usefulness of GK radiosurgery Pre-op patient visit Day of Treatment: apply the frame Obtain thin slice high resolution imaging Construct treatment plan Administer treatment Follow up over long term How it works: Axial Imaging Study High resolution CT and/or MRI scan required Fiducial box allows image registration Planning with proprietary planning system High speed workstation Hospital ethernet connection RS 232 link to computerized treatment console

Knife Radiosurgery S. Goetsch, Ph.D. Page 3 Same Day Treatment Treatments last 5 minutes to 5 hours Acceptance Testing and Commissioning Qualified Medical Physicists needs to perform rigorous acceptance testing Manufacturer recommends output factors for helmets/collimators: difficult for typical user to measure reliably Manufacturer also does radiochromic film analysis Typical users can obtain radiochromic film and attempt to scan in flatbed scanner Imaging QA Essential Must validate accuracy for MRI and angio or digital angiography Quality assurance phantom essential CT can serve as Gold Standard Encourage hospital to obtain ACR certification for all modalities. How accurate do you need to be? Must hit 3mm diameter nerve consistently with 4mm beam diameter Massager: correlated FU MRI with treatment planning and found deviation of 0.91mm; Int J Radiation Oncol Biol Phys, Dec. 2007 (69) 1514-20 MRI Imaging Phantoms Manufacturers: Elekta Inc. CIRS Radiosurgery Head Phantom Standard Imaging LUCY phantom Phantom Laboratory RSVP Phantom Head

Knife Radiosurgery S. Goetsch, Ph.D. Page 4 Elekta Plastic Phantom Problems with Dosimetry Gamma Knife is one of several modalities that does not comply with TG51 AAPM Working Group on Dosimetry Calibration Protocol for Beams that are Not Compliant with TG-51, chaired by Jan Seuntjens Charge is to offer input into a future IAEA dosimetry protocol Therapy Physics Committee considering creation of new GSR Task Group 3. Gamma Knife Clinical Results Over 400,000 patients treated worldwide at 250 Gamma Knife Centers thru 2006 from Leksell Society: 54,000 vascular (14%) 139,000 benign tumors (35%) 172,000 malignant tumors (43%) 30,000 functional (8%) A. Pain relief from Trigeminal Neuralgia ¾ of San Diego Gamma Knife Center patients had Grade 1 or 2 results Grade 1: relief of pain while still on meds Grade 2: relief of pain without meds 34 of 426 (8%) patients had previous surgery (microvascular decompression) Longhi (Neurosurgery, 2007): 61% Grade 1 and 29% Grade 2 B. Non Malignant Acoustic Neuromas Acoustic neuromas (acoustic neurinomas, acoustic schwannomas) are tumors of 8 th cranial nerve Open surgery extremely complex and hazardous Chopra (IJROBP July 2007) reported 10 year control rate in 216 patients of 98.3%, w 71% hearing preservation Marginal dose decreased in 1990 to reduce risk of complications

Knife Radiosurgery S. Goetsch, Ph.D. Page 5 C. Metastatic Brain Tumors Paradigm changing for some primary tumors Radiation oncology dogma: whole brain for every brain met patient Neurosurgery dogma: WBRT causes dementia Local tumor control of 87% reported by U of Pittsburgh (Neurosurgery 2004) D. Now you see it, now you don t Obliteration of AVMs Greater than 30 year history of treatment w Gamma Knife: Steiner at Karolinska Highly successful with both operable and inoperable lesions Flickinger quoted 75% obliteration rate for 250 pts w 3 to 11 year follow-up Open surgery may be risky 4. Elekta Perfexion Controversy New Elekta unit Moving sources No helmets Couch is precision movement device Perfexion at work NRC Advisory Committee on Medical Use of Isotopes (ACMUI) Elekta obtained 510 (k) from FDA Also listed Perfexion on NRC Sealed Source and Device Registry NRC still put Perfexion into Part 35.1000 (essentially for experimental devices); not in Part 35.600 for GSR devices ACMUI objected to this

Knife Radiosurgery S. Goetsch, Ph.D. Page 6 AAPM Rides to Rescue S. Goetsch and Lynne Fairobent created questionnaire and sent to first nine Perfexion sites Asked whether sites complied with NRC License guidance for Perfexion New AAPM Task Group 172 formed in February, 2008 New Part 35.600 ACMUI proposed re-writing Part 35.600 to include Perfexion and rotating gamma units Discussed at semi-annual meeting April 28-29 AAPM TG 172 had a chance to review and comment on proposed new regs Rotating gamma units also included in new proposed Part 35.600 4. Conclusions The price is freedom is eternal vigilance Thomas Jefferson Radiosurgery is a one time event: no second chance Get outside audit before you begin Much work remains to be done with new technologies and dosimetry ISRS 9 th International Congress May or June, 2009 Seoul, South Korea www.isrs2009.org