Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT

Similar documents
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

Limits of Precision and Accuracy of Radiation Delivery Systems

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Unrivaled, End-to-End

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Activity report from JCOG physics group

IMRT QA: Point Dose Measurements or 2D Array?

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

ph fax

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

doi: /j.ijrobp

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

Basic dosimetric data for treatment planning system

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Volumetric Modulated Arc Therapy - Patient Specific QA

Leila E. A. Nichol Royal Surrey County Hospital

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Protection of the contralateral breast during radiation therapy for breast cancer

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Preliminary dosimetric evaluation of a designed head and neck phantom for intensity modulated radiation therapy (IMRT)

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Advanced film dosimetry for a UK brachytherapy audit

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

Pre-treatment verification for IMRT / IGRT and Rotational therapy

Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Diode calibration for dose determination in total body irradiation

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy

QA for Clinical Dosimetry with Emphasis on Clinical Trials

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

IMRT point dose measurements with a diamond detector

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

IMRT pre-treatment verification basic approaches

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Radiochromic film dosimetry in water phantoms

Patient dosimetry for total body irradiation using single-use MOSFET detectors

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Statistical process control analysis for patient-specific IMRT and VMAT QA

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Small field diode dosimetry

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

Application(s) of Alanine

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Y FILMS DOSIMETR Nederland België / Belgique

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT Rajesh Kumar, 1,a S.D. Sharma, 1 Sudesh Despande, 2 Yogesh Ghadi, 2 V.S. Shaiju, H.I.Amols, 3 Y. S. Mayya 1 Radiological Physics and Advisory Division, 1 Bhabha Atomic Research Centre, CT&CRS, Anushaktinagar, Mumbai, India; P.D.Hinduja National Hospital and Research Centre, 2 Mumbai, India; Memorial Sloan-Kettering Cancer Center, 3 New York, USA. raj_resh@rediffmail.com Received 15 February, 2009; accepted 16 August, 2009 A novel IMRT phantom was designed and fabricated using Acrylonitrile Butadiene Styrene (ABS) plastic. Physical properties of ABS plastic related to radiation interaction and dosimetry were compared with commonly available phantom materials for dose measurements in radiotherapy. The ABS IMRT phantom has provisions to hold various types of detectors such as ion chambers, radiographic/radiochromic films, TLDs, MOSFETs, and gel dosimeters. The measurements related to pretreatment dose verification in IMRT of carcinoma prostate were carried out using ABS and Scanditronix-Wellhofer RW3 IMRT phantoms for five different cases. Point dose data were acquired using ionization chamber and TLD discs, while Gafchromic EBT and radiographic EDR2 films were used for generating 2D dose distributions. Treatment planning system (TPS) calculated and measured doses in ABS plastic and RW3 IMRT phantom were in agreement within ± 2%. The dose values at a point in a given patient acquired using ABS and RW3 phantoms were found comparable within 1%. Fluence maps and dose distributions of these patients generated by TPS and measured in ABS IMRT phantom were also found comparable both numerically and spatially. This study indicates that ABS plastic IMRT phantom is a tissue-equivalent phantom and, dosimetrically, it is similar to solid/ plastic water IMRT phantoms. Although this material is demonstrated for IMRT dose verification, it can also be used as a tissue-equivalent phantom material for other dosimetry purposes in radiotherapy. PACS number: 87.53Kn, 87.55Qr, 87.53Bn and 87.55Km Key words: IMRT, phantom, dose verification, ABS I. Introduction ICRU recommends that the dose delivery in radiotherapy should be accurate within ± 5%. (1) IMRT is an advanced high-precision radiotherapy technique which is used mainly for curative treatments. The principle of IMRT is to treat a patient from a number of different directions (or continuous arcs) with beams of nonuniform fluence, which have been optimized to deliver a high dose to the target volume and acceptably low dose to the surrounding normal structures. (2-6) MLCs are commonly used for delivery of prescribed IMRT treatment either in a static or in a dynamic manner, which further complicates the procedure. Due to complex nature of IMRT treatment technique, pretreatment dose verification is an important aspect of IMRT QA program. (7-12) An IMRT field often includes many small, irregular, off-axis fields, requiring that a Corresponding author: Rajesh Kumar, Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, CT&CRS, Anushaktinagar, Mumbai-400094, India; phone: +91-22-25598712; fax: +91-22-25519209; email: raj_resh@rediffmail.com 24 24

25 Kumar et al: ABS-based low cost IMRT phantom 25 a suitable IMRT phantom be used for dosimetric verification. (13-14) Pretreatment verification of IMRT fields, therefore, requires the use of a dedicated tissue-equivalent IMRT phantom with the provision of holding different types of detectors such as ionization chambers, TLDs, radiographic/radiochromic films, gel dosimeters, MOSFETs, and semiconductor diodes. A number of IMRT phantoms with the facility for holding different types of detectors are available commercially for this purpose. (15-18) The majority of these phantoms are made up of solid/plastic water material. Though these phantoms are suitable for pretreatment dose verification in IMRT, they are very costly and some have limited measurement options. Therefore there is a need to design and fabricate an IMRT phantom which is made up of tissue-equivalent material, with options to verify the dose at a point and obtain dose distribution in 2D and 3D. In addition, the phantom should be made available at a reasonable price. In light of these requirements, a versatile IMRT phantom was designed and fabricated from a low cost tissue-equivalent material. The tissue equivalency of a material for the dosimetry purpose depends on type and energy of the radiation. In the case of a photon beam, total attenuation coefficient; in the case of an electron beam, stopping power of the material should be comparable with tissue at the given energy of radiation. (19) Acrylonitrile Butadiene Styrene (ABS) plastic was used as tissueequivalent material for fabrication of the IMRT phantom. This ABS plastic phantom was used for the pretreatment dose verification measurement in IMRT using ionization chamber, TLDs, radiochromic and radiographic film to demonstrate its suitability. A few measurements were also carried out with a commercially available Scanditronix-Wellhofer IMRT RW3 phantom (IBA Dosimetry, Uppsala, Sweden) to compare the results obtained using an ABS IMRT phantom. This paper describes the design features of an ABS plastic IMRT phantom and the measurement results of pretreatment dose verification. II. Materials and Methods An IMRT phantom was designed and fabricated using tissue-equivalent material commercially known as Acrylonitrile Butadiene Styrene (ABS) plastic. Cost of a sheet of 1.2 by 2.4 m 2 of thickness 5 mm is about $200. ABS [(C 8 H 8 C 4 H 6 C 3 H 3 N) n ] is a copolymer made by polymerizing acrylonitrile and styrene in the presence of polybutadiene. The composition of ABS is: acrylonitrile 15% to 35%; butadiene 5% to 30%; and styrene 40% to 60%. Electron density of the phantom material relative to water was estimated by importing its CT scanned images to a recently calibrated treatment planning system for electron density, as well as by numerical evaluation using chemical composition of ABS. The total attenuation coefficient for the phantom material was also calculated using attenuation coefficient of its constituent elements. Table 1 presents dosimetry related physical parameters of commonly used phantom materials along with ABS for 60 Co gamma rays, and 6 and 15 MV X rays. These data indicate that physical parameters of the ABS are comparable to other standard phantom materials. The ABS phantom is elliptical in shape with dimensions sufficient to provide full scatter conditions similar to the irradiation of a patient. Figure 1 shows the schematic line diagram of ABS plastic IMRT phantom. This phantom has two parts. The first part has provision to incorporate ionization chambers, TLD discs, radiochromic films, and gel dosimeter; the second part has the provision to hold radiographic film only. The first part is designed in such a way that a cylindrical ionization chamber can be positioned within a 1 by 1 by 1 cm 3 grid in the central region of the phantom. The positioning of the chamber within this grid is possible by moving suitable compensating inserts. Three fiducial marks are also available in the first part of the phantom for its reproducible positioning. The second part of the phantom consists of elliptical slices of 1 cm thickness and each of these slices contains three fiducial marks to identify the film orientation. Figure 2(a) shows the cubical insert of size 15 by 15 by 15 cm 3 which has a cylindrical hole of 8 cm where the gel dosimeter container can be positioned. A large polymer

26 Kumar et al: ABS-based low cost IMRT phantom 26 Ta b l e 1. Physical properties of common phantom materials and ABS plastic. The data for ABS plastic was numerically calculated from its elemental composition. Energy Material Effective Z Electron Density Physical Density Attenuation Relative to Water Coefficient (µ/ρ) ( 10-2 cm 2 /g) Co-60 Water 7.420 1.000 1.00 6.323 PMMA 6.467 1.034 1.19 6.143 Polystyrene 5.690 0.969 1.06 6.123 Solid water (WT1) 7.54 0.973 1.02 6.230 RW-3-0.967 1.04 6.110 ABS 5.759 0.980 1.04 6.140 6MV Water 7.420 1.000 1.00 4.940 PMMA 6.467 1.034 1.19 4.796 Polystyrene 5.690 0.969 1.06 4.780 Solid water (WT1) 7.54 0.973 1.02 4.800 RW-3-0.967 1.04 4.767 ABS 5.759 0.980 1.04 4.793 15MV Water 7.420 1.000 1.00 3.030 PMMA 6.467 1.034 1.19 2.919 Polystyrene 5.690 0.969 1.06 2.890 Solid water (WT1) 7.54 0.973 1.02 2.920 RW-3-0.967 1.04 2.889 ABS 5.759 0.980 1.04 2.900 Fig. 1. Schematic line diagram of ABS plastic IMRT phantom. gel sample can be used for the acquisition of dose distributions for entire target volumes. The large gel volume allows a dose distribution measurement of a large target volume and sometimes neighboring critical structures as well. Figure 2(b) shows the line diagram and Figure 2(c) shows the photograph of radiochromic film holder where a radiochromic film sample of 15 by 15 cm 2 can be sandwiched between the plates to form a stack. This stack of radiochromic film can be used for the two-dimensional as well as three-dimensional dose distribution analysis. The ABS phantom has provision to hold a TLD tray in which TLDs of 4.5 mm in diameter and a thickness of 0.8 mm can be arranged at a spatial resolution of 7 mm (center to center). Figure 3 shows the final assembly of the fabricated ABS plastic IMRT phantom. The suitability of ABS plastic phantom in IMRT dose verification was tested in comparison with a Scanditronix-Wellhofer IMRT RW3 phantom. For this purpose, pretreatment dose verification was carried out for five different cases of carcinoma prostate treated by IMRT.

27 Kumar et al: ABS-based low cost IMRT phantom 27 Fig. 2(a). Holder for gel dosimeter container. Fig. 2(b). Schematic line diagram of radiochromic film holder.

28 Kumar et al: ABS-based low cost IMRT phantom 28 Fig. 2(c). Photograph of radiochromic film holder. Fig. 3. Final assembly of the locally fabricated IMRT phantom. Volumetric CT scan (GE Discovery, WI, USA) of the ABS plastic phantom, with ionization chamber (FC65G; IBA Dosimetry, Uppsala, Sweden) located at 5 cm from its proximal surface, using the slice width of 3 mm was taken for the point dose verification. The volumetric CT scan data of the ABS plastic phantom was transferred to the Brain SCAN v5.2 (BrainLAB AG, Germany) treatment planning system. IMRT treatment plans of the patient of carcinoma prostate were planned on the phantom assuming the center of ion chamber as the centre of the tumor. The optimized plan so generated was transferred to the Varis Vision network system (Varian Medical Systems, Palo Alto, CA) and planned doses were delivered to the phantom at the predefined position of the chamber with 6 MV photon beam using Varian Clinac 2300 CD (Varian Medical Systems, Palo Alto, CA). The reading of the ionization chamber was converted

29 Kumar et al: ABS-based low cost IMRT phantom 29 into absorbed dose to water using methodology described in the IAEA TRS 398. (20) Treatment plan and dose verification were carried out for five different cases of carcinoma prostate. The point of measurement was located in a low dose gradient zone for each of evaluated IMRT treatment plans. The experiment was repeated on a commercial IMRT phantom to compare the dose measurement carried out using the ABS plastic phantom. A number of IMRT treatment plans were transferred to ABS plastic and Scanditronix- Wellhofer RW3 IMRT phantom and doses at equal physical depth were calculated and measured to verify the equivalent of ABS plastic with RW3 as phantom material for dosimetry purpose. To demonstrate the use of various detectors in the ABS IMRT phantom, the dose verification measurements were carried out using radiographic film, radiochromic film, and TLD along with ionization chamber. Radiographic film (EDR2 film, Eastman Kodak Company, Rochester, NY) was placed in the second part of the phantom and was irradiated as required by the planned IMRT treatment. Readout of the EDR2 film was carried out using VIDAR Dosimetry Pro Advantage scanner (VIDAR Systems Corporation, Herndon, VA). This way the EDR2 film gave the dose distribution data in the transverse plane. The calibration of EDR2 was carried out in 6 MV X-ray beam. A 15 by 15 cm 2 Gafchromic EBT film (Gafchromic EBT; ISP Inc. NJ, USA) was positioned in the first part of the phantom at a depth of 5 cm and was irradiated as required by the planned IMRT treatment. The irradiated film was read by flatbed scanner (EPSON Expression 10000XL; EPSON, UK). Scanning was done using the EPSON SCAN software with all filters switched off. The images were scanned in transmission mode and saved in RGB uncompressed tagged image file format (TIFF). For EBT films, the absorption peak falls in the red region and therefore the red component of the image was extracted to maximize film readout using ImageJ software (ImageJ 1.41o; National Institute of Health, USA). This way the Gafchromic EBT film gave the dose distribution data in the coronal plane. Two-dimentional dosimetric analysis of scanned images was carried out using the IMRT dose verification software (OmniPro-I mrt version 1.5; IBA dosimetry, Uppsala, Sweden). To obtain a calibration curve for EBT film, the film samples of 4 by 4 cm 2 were positioned in a conventional solid phantom perpendicular to the radiation beam and irradiated with 6 MV X rays in the known dose ranges: 50, 100, 200, 300 and 400 cgy. The exposed film samples were scanned in similar manner as mentioned above. A dose response curve was plotted and best fit of these data was used to determine unknown dose values from the knowledge of OD of exposed films using the polynomial expression: D = a*od + b*od^c where a, b and c are free fitting parameters. (21) A TLD tray holding LiF: Mg, Ti (MTS-N; TLD, Poland) discs 4.5 mm in diameter and with a thickness of 0.8 mm was position in the first part of the phantom at a depth of 5 cm and was irradiated as required by the planned IMRT treatment. The readouts of the TLD discs were taken 24 h after irradiation using a TLD reader (Rexon UL320, USA) with programmable temperature profile. The TL readout was taken in integration mode. The standard annealing and analysis process for the TLDs were followed to obtain dose data from the exposed TLDs. The calibration of TLD discs was also carried out in 6 MV X rays. III. Results & DISCUSSION Table 2 shows the TPS calculated dose values at a point in the Scanditronix-Wellhofer IMRT RW3 phantom and the ABS plastic phantom for five different cases of prostate cancer treated by IMRT using 6 MV X rays. The table also includes ionization chamber measured dose values in these two phantoms at the same point where the dose was calculated by the TPS. A survey of data in this Table indicates that TPS calculated and ionization chamber measured dose values at corresponding points in Scanditronix-Wellhofer IMRT RW3 phantom and ABS plastic phantom agree within 2% for all the five cases. If we assume TPS calculated dose as reference dose values, then it can easily be concluded from these observations that these two phantoms are quite suitable for pretreatment dose verification in IMRT.

30 Kumar et al: ABS-based low cost IMRT phantom 30 Table 3 lists the TPS calculated and ionization chamber measured dose values at 5 cm depth in Scanditronix-Wellhofer IMRT RW3 phantom and ABS plastic phantom for five different cases of prostate cancer treated by IMRT using 6 MV X rays. The TPS calculated dose values in the two phantoms agree with each other within 0.5% for three cases, and shows a variation of 2.63% in one case. The ionization chamber measured dose values in these two phantoms agree within 1%. It can be concluded from the data in Table 3 that measured as well as calculated doses at a given point in these two phantoms are in agreement with each other. As Scanditronix-Wellhofer IMRT RW3 phantom is a tissue-equivalent phantom; the ABS plastic IMRT phantom can also be assumed to be a tissue-equivalent phantom. However, the cost of ABS plastic phantom is about 8 to 10 times lower than the cost of the Scanditronix-Wellhofer IMRT RW3 phantom and similar other commercial phantoms with equivalent features. Point doses measured using EBT Gafchromic film in ABS plastic IMRT phantom were found to be within 0.6% in comparisons of TPS calculated dose value. However, the dose measured using TLD discs at the same point were found to be within 2.8%. Figure 4 shows: (a) fluence map in coronal plane recorded on Gafchromic EBT film of an IMRT plan in ABS plastic phantom; (b) TPS-generated fluence map in coronal plane of an IMRT plan; (c) comparison of isodose lines recorded on Gafchromic EBT film in ABS plastic phantom and TPS-generated isodose line of an IMRT plan in coronal plane; and (d) gamma analysis of an IMRT plan. Figure 5 shows: (a) fluence map in transverse plane recorded on EDR2 film of an IMRT plan in ABS plastic phantom; (b) TPS-generated fluence map in coronal plane of an IMRT plan; (c) comparison of isodose lines recorded on EDR2 film in ABS plastic phantom and TPS-generated isodose line of an IMRT plan in transverse plane; and (d) gamma analysis of an IMRT plan. The correlation coefficient of plans recorded on Gafchromic EBT/ EDR 2 films in ABS plastic phantom were found to be better than 0.990 for the region of interest. The measured dose distribution by Ta b l e 2. TPS calculated and ionization chamber measured dose values in Scanditronix-Wellhofer RW3 and ABS plastic IMRT phantoms. Calculated and ionization chamber measured dose values and their deviations Scanditronix-Wellhofer RW3 IMRT Phantom ABS Plastic IMRT Phantom Case TPS (Gy) Ionization % Variation TPS (Gy) Measured % Variation Chamber (Gy) Dose 1 2.28 2.32-1.8 2.34 2.30 1.7 2 2.23 2.24-0.45 2.22 2.26 1.8 3 2.1 2.06 1.9 2.11 2.07 1.9 4 2.23 2.22 0.45 2.19 2.20-0.5 5 2.23 2.20 1.3 2.24 2.20 1.8 Ta b l e 3. TPS calculated and ionization chamber measured dose values at 5 cm depth in Scanditronix-Wellhofer RW3 and ABS plastic IMRT phantoms. TPS (Gy) Calculated and measured dose and their deviations Ionization Chamber (Gy) Case Scanditronix- ABS plastic % Scanditronix- ABS Plastic % Wellhofer RW3 IMRT Deviations Wellhofer RW3 IMRT Deviations Phantom Phantom Phantom Phantom 1 2.28 2.34-2.63 2.32 2.30 0.86 2 2.23 2.22 0.45 2.24 2.26-0.89 3 2.1 2.11-0.48 2.06 2.07-0.49 4 2.23 2.19 1.79 2.22 2.2 0.90 5 2.23 2.24-0.45 2.2 2.2 0

31 Kumar et al: ABS-based low cost IMRT phantom 31 Fig. 4. Graphic showing: (a) fluence map in coronal plane recorded on Gafchromic EBT film of an IMRT plan in ABS plastic phantom; (b) TPS-generated fluence map in coronal plane of an IMRT plan; (c) comparison of isodose lines recorded on Gafchromic EBT film in ABS plastic phantom and TPS-generated isodose line of an IMRT plan in coronal plane; and (d) gamma analysis of an IMRT plan. Fig. 5. Graphic showing: (a) fluence map in transverse plane recorded on EDR2 film of an IMRT plan in ABS plastic phantom; (b) TPS-generated fluence map in coronal plane of an IMRT plan; (c) comparison of isodose lines recorded on EDR2 film in ABS plastic phantom and TPS-generated isodose line of an IMRT plan in transverse plane; and (d) gamma analysis of an IMRT plan. EBT/ EDR2 films agreed in the region of interest with a planned dose distribution that passed 3%, 3 mm gamma-index evaluation. Figures 4 and 5 indicate close agreement between the information recorded on Gafchromic EBT/ EDR2 films in ABS plastic phantom and TPSgenerated data for the IMRT plan of a patient.

32 Kumar et al: ABS-based low cost IMRT phantom 32 IV. Conclusions A novel IMRT phantom was designed and fabricated using ABS plastic. The studies carried out on this phantom indicate that it is equivalent to plastic/solid water IMRT phantoms available commercially. The ABS phantom is a versatile tissue-equivalent phantom which can be used for pretreatment dose verification in IMRT using different types of radiation detectors. The phantom is suitable for dosimetry in 1D, 2D and 3D. Though this material is demonstrated for IMRT dose verification, it can also be used as a tissue-equivalent phantom material for other dosimetry purposes in radiotherapy. ACKNOWLEDGEMENT This work is partially supported by the International Atomic Energy Agency (IAEA). References 1. International Commission on Radiation Units and Measurements. Dose specification for reporting external beam therapy with photons and electrons. ICRU Report 29. Bethesda (MD): ICRU; 1978. 2. Webb S. Intensity-Modulated Radiation Therapy. Bristol, UK: Institute of Physics Publishing; 2001. 3. Galvin JM, Ezzell G, Eisbrauch A, et al. Implementing IMRT in clinical practice: a joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in Medicine. Int J Radiat Oncol Biol Phys. 2004;58(5):1616 34. 4. Palta JR, Mackie TR, editors. Intensity-Modulated Radiation Therapy: The State of the Art. Medical Physics Monograph No. 29. 2003 AAPM Summer School Proceedings. Madison (WI): Medical Physics Publishing; 2003. 5. Eisbruch A. Intensity-modulated radiotherapy of head-and-neck cancer: encouraging early results. Int J Radiat Oncol Biol Phys. 2002;53(1):1 3. 6. Zelefsky MJ, Fuks Z, Happersett L, et al. Clinical experience with intensity modulated radiation therapy (IMRT) in prostate cancer. Radiother Oncol. 2000;55(3):241 49. 7. Williams PC. IMRT: delivery techniques and quality assurance. Br J Radiol. 2003;76(911):766 76. 8. Manning M A, Wu Q, Cardinale RM, et al. The effect of setup uncertainty on normal tissue sparing with IMRT for head-and-neck cancer. Int J Radiat Oncol Biol Phys. 2001;51(5):1400 09. 9. Sandilos P, Angelopoulos A, Baras P, et al. Dose verification in clinical IMRT prostate incidents. Int J Radiat Oncol Biol Phys. 2004;59(5):1540 47. 10. Gillis S, De Wagter C, Bohsung J, Perrin B, Williams P, Mijnheer BJ. An inter-centre quality assurance network for IMRT verification: results of the ESTRO QUASIMODO project. Radiother Oncol. 2005;76(3):340 53. 11. van Zijtveld M, Dirkx MLP, de Boer HJC, Heijmen BJM. Dosimetric pre-treatment verification of IMRT using an EPID: clinical experience. Radiother Oncol. 2006;81(12):168 75. 12. Kung JH, Chen GT. Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracy. Med Phys. 2000;27(7):1617 22. 13. Molineu A, Followill DS, Balter PA, et al. Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the Radiation Therapy Oncology Group. Int J Radiat Oncol Biol Phys. 2005;63(2):577 83. 14. Low DA, Gerber RL, Mutic S, Purdy JA. Phantoms for IMRT dose distribution measurement and treatment verification. Int J RadiatOncol Biol Phys. 1998;40(5):1231 35. 15. Benchmark IMRT QA Phantom. Available from: http://civco.com/oncology/physics/phantom/benchmark/ 16. Computerized Imaging Reference Systems, Inc. Norfolk, VA, USA. Available from: http://www.cirsinc.com/ main_radiation.html 17. IBA Dosimetry AB, Uppsala, Sweden. Available from: http://www.iba-dosimetry.com/complete-solutions/ radiotherapy/imrt-igrt-rotational-qa/i-mrt-phantom 18. Standard Imaging; Middleton, WI, USA. Available from: http://www.standardimaging.com/product_home. php?id=37 19. International Commission on Radiation Units and Measurements. Tissue substitutes in radiation dosimetry and measurements. ICRU Report 44. Bethesda (MD): ICRU; 1989. 20. International Atomic Energy Agency. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA Technical Series Report 398. Vienna, Austria: IAEA; 2000. 21. Ferreira BC, Lopes MC, Capela M. Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry. Phys Med Biol. 2009;54(4):1073 85.