A Mechanism for "False" Inhibition of Demand Pacemakers

Similar documents
Figure 2. Normal ECG tracing. Table 1.

Radioelectrocardiography in Diagnosis of Rhythm

Electronic Analysis for Pacemaker Failure

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

The Influence of Electromagnetic

Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D.

SARASOTA MEMORIAL HOSPITAL

Pacemaker Concepts and Terminology*

Interpretation of complex demand pacemaker arrhythmias

Model 5392 EPG Temporary Pacer

Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Electrocardiography for Healthcare Professionals

Project Title Temporary Pacemaker Training Simulator

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

RN-BC, MS, CCRN, FAHA

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Differentiation of Ventricular Tachycardia from Junctional Tachycardia with Aberrant Conduction

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

The Results of Demand Pacing

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Cardiac Arrhythmias Simulated by Concealed Bundle of His Extrasystoles in the Dog

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

Surveillance and Follow-up of Patients with Implanted Cardiac Pacemaker by Telephone Transmission

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

Relation between QT interval and heart rate

Pediatric pacemakers & ICDs:

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Dr Randal McRoberts

Pacing Lead Implant Testing. Document Identifier

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Pacemaker Capture* MOSS AND SIDNEY GOLDSTEIN. and Greatbatch (1960). Permanent transvenous pacemakers

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

Reversion of ventricular tachycardia by pacemaker stimulation

QUICK SETUP GUIDE SECULIFE DFBASE

BEDSIDE ECG INTERPRETATION

Biomedical Instrumentation

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

CORONARY ARTERIES HEART

BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV.

MODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER

The Electrocardiogram

TIP CARD. Model Temporary External Pacemaker

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

The Electrocardiogram part II. Dr. Adelina Vlad, MD PhD

EHRA Accreditation Exam - Sample MCQs Invasive cardiac electrophysiology

Premature Ventricular Depolarization

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

3/26/15 HTEC 91. EKG Sign-in Book. The Cardiac Cycle. Parts of the ECG. Waves. Waves. Review of protocol Review of placement of chest leads (V1, V2)

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Ventricular Parasystole

Transtelephone Pacemaker Monitoring

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Cardiac arrhythmias. Janusz Witowski. Department of Pathophysiology Poznan University of Medical Sciences. J. Witowski

Principles and Applications of Electrical Circuits and Signal Theories in EP

Sample. Analyzing the Heart with EKG. Computer

UNDERSTANDING YOUR ECG: A REVIEW

An Official Journal of the American Heart

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES

Lab #3: Electrocardiogram (ECG / EKG)

Electrocardiography Abnormalities (Arrhythmias) 7. Faisal I. Mohammed, MD, PhD

Perioperative Management of Cardiac Implantable Devices

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

Supplemental Material

Birmingham Regional Emergency Medical Services System

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Lect.6 Electrical axis and cardiac vector Cardiac vector: net result Vector that occurs during depolarization of the ventricles Figure:

Mobitz II AV block within the His bundle, with

Effects of myocardial infarction on catheter defibrillation threshold

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES

Diagnostic capabilities of the implantable therapeutic systems

To investigate the safe levels of radiotherapy administered to patients who have an implanted cardiac device.

Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing.

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

BASIC CONCEPT OF ECG

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

ECG. Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

Specified Tolerance (ms)

Electrocardiography for Healthcare Professionals

ELECTROCARDIOGRAPHY (ECG)

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

THE CARDIOVASCULAR SYSTEM. Heart 2

Transcription:

A Mechanism for "False" Inhibition of Demand Pacemakers By KENNETH C. LASSETER, M.D., JACK W. BUCHANAN, JR., M.S.E.E., AND KARL F. YOSHONIS, M.D. SUMMARY Certain variations in discharge rate of demand pacemakers may be associated with intermittent fractures and resulting resistance changes in the electrode-lead system. These variations in rate are related to inhibition of the demand circuit of the pacemaker. Most pacemakers produce a biphasic output pulse which results in a low amplitude current during the interval between stimuli. Therefore, sudden changes of resistance in the electrode-lead system will produce a potential change at the pacemaker terminals which is of sufficient magnitude to be interpreted as spontaneous ventricular depolarization and inhibit the demand pacemaker. Such inhibition may occur in the absence of other signs of lead fracture and present a difficult differential diagnostic problem of pacemaker malfunction. Additional Indexing Words: Atrioventricular conduction disturbances Pacemaker failure Lead fracture Ventricular inhibited pacemaker Biphasic stimulation T HE use of ventricular inhibited (demand) pacemakers in the treatment of atrioventricular conduction disturbances has introduced problems of sensing in addition to those of adequate stimulation. Variations in discharge rate are known to occur if the pacemaker recognizes other ECG deflections' or extraneous electrical interference2-5 in addition to the QRS complex, or if there is circuit malfunction.3-6 Recently Kastor and associates7 reported that benign variations in the discharge rate of demand pacemakers may be due to variations in intraventricular conduction and the time of sensing. We have observed marked variations in the discharge rate of ventricular inhibited pacemakers in the From the Cardiovascular Division, Department of Medicine, University of Kentucky College of Medicine, Lexington, Kentucky. Supported in part by Training Grant HE 05598 from the National Institutes of Health, U. S. Public Health Service. Address for reprints: Departnent of Pharmacology, University of Miami School of Medicine, Miami, Florida 33152. Received June 11, 1970; revision accepted for publication August 10, 1970. 1093 absence of detectable extraneous interference or circuit malfunction in several patients. In one of these patients, such variations were found to be related to inhibition of the pacemaker by intermittent changes in lead resistance. This "false" inhibition of demand pacemakers has been reproduced in a laboratory model and the underlying mechanisms investigated. Report of Case A 46-year-old white female presented with a 3- day history of syncopal episodes. Physical examination revealed a heart rate of 30/min and the electrocardiogram revealed complete atrioventricular block. A temporary transvenous bipolar pacing catheter* was placed in the apex of the right ventricle and connected by an extension cable to the terminals of a Medtronic model 5840 External Demand Pacemaker. The patient improved clinically, and the ECG demonstrated normal function of the demand pacemaker in response to occasional extrasystoles (fig. 1A). The patient was monitored continuously in the coronary intensive care unit where all equipment is appropriately grounded. Six hours after inser- *Model 5651, U. S. Glens Falls, Catheter and Instrument Corp., New York.

1094 LASSETER ET AL. A E C i~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~44,a111ia1 B01g11 till 111itI'll fl1!10ii HITH1 11111F.111,! X i Figure 1 Serial strips of ECG lead II recorded during temporary transvenous pacing with Medtronic model 5840 demand pacemaker. Pacer rate remains at 82/min throughout the tracings. (A) Normal function shortly after insertion demonstrating inhibition of discharge by spontaneous beats. (B and C) Marked variations in pacemaker escape interval subsequently related to movement of extension cable. (D) Normal function of pacemaker in fixed rate mode. Note the pacemaker artifact after spontaneous beat. tion of the pacemaker, variations in the discharge rate were noted with some relatively long periods of electrical silence (fig. lb and C). When the pacemaker was switched to the fixed rate mode, normal pacing resulted (fig. ID). Fluoroscopy confirmed that the position of the electrode catheter was unchanged. It was noted that motion of one area of the extension cable resulted in these variations in rate. Resistance measurements of the disconnected cable with a Triplet 601 ohmmeter revealed intermittent changes in the resistance of one conductor suggestive of a lead fracture. Since no nonconducted stimuli were observed during fixed rate pacing, it is presumed that either the fracture was "incomplete" or that it was of shorter duration than the stimulus. Replacement of the connecting cable resulted in normal pacing in both modes of operation. Experimental Results Resistance measurements of the bipolar pacing catheters used in this institution revealed that several such catheters exhibited this same phenomenon of motion-related intermittent changes of resistance. Figure 2 illustrates an experiment with one such catheter. The output of the Medtronic 5840 Demand Pacemaker was recorded at the catheter tip using a 500fQ (ohms) resistor to simulate myocardial resistance.8 The intermittent changes in resistance did not interrupt conduction since no beats were lost during fixed rate pacing and since one beat was conducted with decreased amplitude during the high resistance phase in panel C (fourth from last beat). The circuit diagrammed in figure 3 was constructed as a model of this type of demand pacemaker failure. Resistor R1 had a constant Circulatson, Volume XLII, December 1970

FALSE INHIBITION OF DEMAND PACEMAKERS 1095 A B Figure 2 Recordings of pacemaker output at catheter tip with a Tektronix 564B oscilloscope with a 500o load. Pacemaker rate remains at 60/min throughout recording. (A) Normal pacing in demand mode. (B and C) Variations in discharge rate produced by manipulation of catheter. (D) No variation of rate during catheter manipulation in fixed rate mode. PACEMAKER F-/lo'-] ----ktv%r.l R2 RI TOCRO Figure 3 Circuit diagram of model to simulate changes in lead resistance. CRO = cathode ray oscilloscope. R1= 500o load simulating myocardial resistance at electrodes. R2 = variable resistance of lead fracture. Rapid opening and closing of switch simulates intermittent fracture. value of 500fQ and resistor R2 varied in value. Figure 4 illustrates representative experiments with this model using a typical value of 10,000 Ql for R2. In figure 4A, the eighth beat is delayed until 750 msec after switching resistor R2 into the circuit. This interval is the escape interval of the pacemaker, indicating that the pacemaker was "reset" by this event. If R2 is w Figure 4 Recordings of pacemaker output with the model diagrammed in figure 3. The upper trace in each panel records amplitude and rate of pacemaker output. In the lower trace in each panel the positive steps indicate that resistor R2 is in the circuit. R2 had constant value of 10,OOOo in these experiments. Pacemaker rate = 80/min. constantly left in the circuit, no change in rate is seen although there is diminution of the stimulus strength. It should be noted that if R. is switched into the circuit during the refractory period of the pacemaker, no delay is seen. In Figure 4B, R2 is intermittently and randomly switched in and out of the circuit to simulate intermittent fracture of a lead. Each delayed beat occurs 750 msec after a change of resistance, either switching R. into or out of

1096 the circuit. Various combinations of switching R2 in and out of the circuit can result in wide variations in pacemaker discharge rate. Figure 4C demonstrates that if such sudden changes in resistance occur at the proper frequency, long periods of pacemaker silence can be induced. All values of R2 greater than 1,OOOfl were found to inhibit the demand pacemaker. Similar inhibition has been demonstrated with the Medtronic Implantable Demand Pacemaker model 5841 and with the General Electric Implantable Demand Pacemaker model A2072A. Discussion Most commonly used cardiac pacemakers produce a biphasic output pulse which results in no net transfer of energy during each cycle.8 The large amplitude, short duration output pulse is delivered by a capacitor which is then recharged by a low amplitude, long duration current during the interval between stimuli. This current is shown in figure 5 as the voltage drop across the 500Qf load. The instantaneous magnitude of this current is dependent on the output stimulus setting and decreases exponentially with time. At a pacemaker setting of 3 ma, which might be used clinically, and at 500 msec, a time clearly after the refractory period of these pacemakers,9 the magnitude of this current is about 4 microamperes (/.a) (fig. 5). In the circuit diagram in figure 3, the potential across the pacemaker terminals (Ep) at any given time is dependent on the current (I) in the circuit and the series resistance. When the switch is closed (representing no lead fracture ), Ep = IR1 When the switch is open (representing lead fracture), Ep = I (R1 + R2). The potential change produced by opening the switch would be AEp =IR2 Assuming a current of 4,ua, a sudden resistance change of 1,000fl (the smallest value found to inhibit the pacemaker) would LASSETER ET AL. Figure 5 Recordings of current between stimuli of Medtronic 5840 Demand Pacemaker as the voltage drop across 500O load. At this sensitivity only the negative component of the biphasic stimulus pulse is seen, the positive spike being far off the screen. (A) Pacemaker stimulus setting of 3 ma. (B) Pacemaker setting of 25 ma. Vertical scale: 5 mv/division which corresponds to 10,ua/division. Horizontal scale: 500 msec/division. The solid line in each panel represents the zero potential. produce a potential difference of the order of 4 mv at the pacemaker terminals. The magnitude of the ventricular depolarization potential, recorded with an endocardial electrode, is of the order of 8 to 10 mv.9-11 To provide a margin of safety, the magnitude of the signal required to inhibit most demand pacemakers is less than this value and is of the order of 1.5 to 2.0 mv.10-'2 Any sudden potential change of this magnitude could be interpreted as a spontaneous ventricular depolarization and could inhibit the pacemaker.

FALSE INHIBITION OF DEMAND PACEMAKERS 1097 The potential difference produced by a sudden resistance change in the electrode system can, therefore, inhibit the pacemaker. In the case reported herein, such sudden resistance changes did inhibit the pacemaker. The basic escape interval of this pacemaker is 730 msec (fig. 1). The irregular beats exhibit R-R intervals from 900 to 3,000 msec, none of which were even multiples of the escape rate. The demonstration of the temporal relationship between these intervals and maneuvers which produced sudden changes in the resistance of the electrode-lead system, strongly supports this as the mechanism of pacemaker inhibition in this case. Zuckerman and associates"1 have reported a case, in which intermittent variable prolongation of the pacer escape interval developed 1 year after transthoracic implantation of a ventricular inhibited (demand) pacemaker. This was attributed to an intermittent electrode connection "which emitted a signal and produced false cancelling." This problem was overcome by removing the faulty lead from the circuit and pacing unipolarly. The same mechanism would seem to explain the pacemaker inhibition in each of these cases. Although lead fracture is a well-recognized complication of cardiac pacing, it would not generally be expected to change the escape interval of the pacemaker except by failure to pace. It has been postulated that body fluids may enter the fracture of an intracorporeal lead allowing adequate conduction of the stimulus at all times."' 13 In this situation the conductor could have two resistance states, either of which would conduct the stimulus. A change between these states might produce a potential change adequate to inhibit the pacemaker. Variation in the pacemaker discharge rate due to sudden resistance changes in the electrode system is a potential problem in all pacers using biphasic stimulation and common electrode systems for stimulation and sensing. This includes not only most ventricular inhibited (demand) pacemakers but ventricular-triggered pacemakers as well. This is not a problem with continuous asynchronous (fixed rate) pacemakers because of the absence of a sensing circuit, nor in the atrial- triggered pacemakers because of the separation of the sensing and stimulating electrode systems. When triggered pacemakers are used, the physician should be aware of the additional problems introduced and should be able to identify and correct the source of these problems. We have shown that at least one type of triggered pacemaker, the ventricular inhibited (demand) pacemaker is sensitive to defects in its lead system which would not be apparent with asynchronous pacing. This problem can be overcome by identifying and replacing the defective component of the electrode lead system. References 1. CASTELLANOS A JR, MAYTIN 0, LEMBERG L, ET AL: Pacemaker induced cardiac rhythm disturbances. Ann N Y Acad Sci 167: 903, 1969 2. FURMAN S, PARKER B, KRAUTHAMER M, ET AL: The influence of electromagnetic environment on the performance of artificial cardiac pacemakers. Ann Thorac Surg 6: 90, 1968 3. FURMAN S, ESCHER DJW: Choice of cardiac pacemaker. Ann N Y Acad Sci 167: 557, 1969 4. PARKER B, FURMAN S, ESCHER DJM: Input signals to pacemakers in a hospital environment. Ann N Y Acad Sci 167: 823, 1969 5. BILITcH M: Ventricular fibrillation and pacing. Ann N Y Acad Sci 167: 934, 1969 6. PARSONNETT V, MYERS GH, GILBERT L, ET AL: Prediction of impending pacemaker failure in a pacemaker clinic. Amer J Cardiol 25: 311, 1970 7. KASTOR JA, BERKOVITs BV, DESANCTIS RW: Variations in discharge rate of demand pacemakers not due to malfunction. Amer J Cardiol 25: 344, 1970 8. BOWERS DL: Engineering considerations of asynchronous pacer systems. Ann N Y Acad Sci 167: 835, 1969 9. CASTELLANOS A JR, SPENCE R: Pacemaker arrhythmias in context. Amer J Cardiol 25: 372, 1970 10. BERKOVITs BV, Demand pacing. Ann N Y Acad Sci 167: 891, 1969 11. ZUCKERMAN W, MATLOFF JM, HARKEN DE, ET AL: Clinical applications of demand pacing. Ann N Y Acad Sci 167: 1055, 1969 12. VAN DEN BERG J, THALEN HJT: Demand pacing, preferable with threshold control. Ann N Y Acad Sci 167: 896, 1969 13. THALEN HJT, VAN DEN BERG J, NIEVEEN J, ET AL: Photoanalysis of electrode function. Ann N Y Acad Sci 167: 807, 1969