Exploiting Selectivity in HPLC and UHPLC With Rational Stationary Phase Design

Similar documents
ACE Excel UHPLC Columns

ACE. For increased polar retention and alternative selectivity. Alternative selectivity for method development

ACE C18-AR. A C18 bonded phase with unique selectivity. Guaranteed reproducibility. Exceptional bonded phase stability

ACE C18-AR Use the Power!

ACE CN-ES. Extra Stability. Extended Spacer. Enhanced Selectivity ACE. Combining CN polar selectivity with enhanced hydrophobicity

Restek Ultra II HPLC Columns

Aeris. Precision Engineered Core- Shell Particles for Ultra-High Resolution BioSeparations. Aeris PEPTIDE. Aeris WIDEPORE

Fortis Method Development Options

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases

ACE. Intelligent Solutions for Method Development. Method Development Kits. ACE UHPLC and HPLC Columns. a a

Hyper-fast & Super-rugged

Am I getting the very best value from my UHPLC analyses?

ACE ACE. HPLC and UHPLC Columns. Product Catalogue

InertSustain Phenylhexyl. A New Second Choice Column to Change Chromatographic Behavior

Thermo Scientific Hypersil BDS Columns

2012 Waters Corporation 1

ACE ACE. HPLC and UHPLC Columns. Product Catalog

Dikma Technologies Inc.

Comparison of a UPLC Method across Multiple UHPLC Systems

For Far. Greater. UHPLC Column Performance You Must Have Kinetex Core-Shell UHPLC Columns.

ACE HILIC + + UHPLC and HPLC Columns. For Reproducible Polar Analyte Separations. Three HILIC Column Chemistries Available

APPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit

Dikma Technologies Inc.

Agilent Technologies Prep LC Columns

Analytical and Preparative SFC Columns

Chapter 22: Gas and Liquid Chromatography. Liquid Chromatography: Stationary Phases and Mobile Phases: Normal Phase: Reversed Phase:

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Sepax Technologies, Inc.

Ultimate Core Performance to Maximize Your Investment

Ultra Columns. also available. ordering note

Preparative Chromatography Meets UHPLC

The LC Column Playbook

Chromatogram Search Tool Search by compound name, synonym, CAS # or keyword Pinnacle II Columns

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

InertSustainBio C18. Rapid Separations of Proteins and Peptides

Eurospher II the logical choice. Physical properties of Eurospher II silica gel: Silica gel: ultra pure, > % Metal content: < 10 ppm

Ultra Columns HPLC COLUMNS

C30 ISOMERS HAVE MET THEIR MATCH

11:00 a.m. EST Telephone Number: Chair Person: Tim Spaeder

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Gerry Hendrickx Regional Sales Manager Central Europe. Developments in Waters Column Chemistries : BEH Technology

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

176 YMC Chiral Columns

SUMMARY, CONCLUSION & RECOMMENDATIONS

The high efficiency of sub-2 µm UHPLC columns combined with the low pressure and high speed of monolith columns.

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

SeQuant ZIC -HILIC For all who expect more...

Separation of Vitamin D and Vitamin D Metabolites on FLARE C18 MM (Mixed Mode) HPLC Column

LC Columns with Liquid Separation Cell Technology

Effective Surface Area (m 2 /g) Pore Size (Å)

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

SeQuant ZIC -HILIC For all who expect more...

New HPLC Column Options for UHPLC

APPLICATIONS TN A Comparison of Various Kinetex C18 Phases USP: L1

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Roc On with These Dependable LC Columns

UHPLC / HPLC Columns Developed for Mass Spectrometry

Columns The FIRST and ONLY sub-2 µm core-shell columns on the market

Chromatography on Immobilized Artificial Membrane

GlycanPac AXR-1 Columns

High-Throughput Separations for QA/QC, LC/MS, LC/MS/MS, Combinatorial Chemistry and Rapid Method Development

Reversed-Phase Columns (Other than ODS)

A New Era in SFC Column Technology

Column Stability and Performance under High Temperature HPLC Conditions

HICHROM. Chromatography Columns and Supplies NEW PRODUCTS. Catalogue 9. Hichrom Limited

Extended Temperature Range Impact on HPLC Column Stability and Performance

Core-Shell Technology for Proteins and Peptides

Impurity Profiling of Carbamazepine by HPLC/UV

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Choosing Preparative Columns

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Thermo Scientific. GlycanPac AXR-1. Column Product Manual. P/N: April, Part of Thermo Fisher Scientific

The One Minute Chromatographer

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns

The Raptor HILIC-Si Column

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

YMC-Triart C18. Multi-layer hybrid material for HPLC ph and temperature stability. Innovative micro-reactor technology

Studies on Stationary Phase Selectivity for Solid-Core Particles

UNISON UK - Amino For Aqueous Elutions and Exceptional Separation Balance

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

YMC-Actus. Columns for Semi-preparative HPLC. Selectivity Durability Loadability

Determination of enantiomeric purity of Timolol Maleate by Supercritical Fluid Chromatography

APPLICATIONS TN Dramatically Improve Existing 5 µm and 3 µm Fully Porous Methods with Kinetex 5 µm Core-shell Technology.

Unequalled durability against water elution. % tr

Protect your expensive HPLC and UHPLC columns from premature failure with MAC-MOD Pre-Column Filters

Meteoric Core. Core-Shell Columns for UHPLC & HPLC. Easy Method Optimization! Excellent Chromatography with Fast Separation!

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Yes You Can! New! Amazing EVO C18. Core-Shell Technology

SecurityGuard. Column Protection for UHPLC, HPLC to PREP. A Universal Guard Cartridge System. Increases HPLC Column Lifetime, Guaranteed!

HPLC Columns. HPLC Accessories HPLC Column Test Mixes Reference Books... HPLC Applications...

Nuevos Desarrollos en Columnas ACQUITY UPLC & HPLC

Prep Columns. and Bulk Media. and Bulk Media

Evaluation of UHPLC System Performance Under High Throughput Conditions. William Hedgepeth, Masatoshi Takahashi, Shimadzu Scientific Instruments

Core-Shell / Solid core Particles: A Practical Substitute to Sub-2 Micron Particles: An overview

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Transcription:

Exploiting Selectivity in HPLC and UHPLC With Rational Stationary Phase Design Alan P McKeown amckeown@ace-hplc.com Advanced Chromatography Technologies www.ace-hplc.com www.mac-mod.com

Outline 2 Chromatographic selectivity Stationary phase design concepts The unique C18-AR and C18-PFP phases Introducing the NEW Excel UHPLC products Examples Conclusions

Resolution (R s ) Chromatographic Peak Resolution 3 3.0 2.5 a Efficiency Selectivity Retention 2.0 R s = N a-1 a k 1+k 1.5 1.0 N k 0.5 0.0 1.00 1.05 1.10 1.15 1.20 1.25 0 5000 10000 15000 20000 25000 0 5 10 15 20 25 a N k Zhao, J.H. and P.W. Carr. Analytical Chemistry, (1999) 71, 2623-2632

The Importance of N, k and a For Resolution Typical separation: N = 10,000 plates k = 3.8 /.2 (.0 mean) a = 1.1 R S R S 1 1.8 1.11 10,000 1.1 1 R s = N a-1 a k 1+k Which looks like

The Importance of N, k and a For Resolution 5 Double Efficiency (eg 5 mm 2.5 mm): N = 10,000 20,000 plates R s = N a-1 a k 1+k R S R S 1 2.6 1.11 20,000 1.1 1 R s = 1.8 ~0% Increase R s = 2.6 Opportunity to optimise further eg reduce column length to speed up

The Importance of N, k and a For Resolution 6 Double Retention Factor (eg decrease solvent strength): k = 8 N R s = a-1 a k 1+k R S R S 1 2.0 1.11 8 10,000 1.1 1 8 R s = 1.8 ~10% Increase R s = 2.0 Slight improvement in resolution has led to increased analysis time

The Importance of N, k and a For Resolution 7 Increase Selectivity (eg change column): a = 1.1 1.2 R s = N a-1 a k 1+k R S R S 1 1.2 1 10,000 1.2 1 3.3 R s = 1.8 ~80% Increase R s = 3.3 Significant opportunity to speed up for modest change in selectivity

Resolution (R s ) Selectivity: The Key to Chromatographic Peak Resolution 8 Efficiency Selectivity Retention 3.0 2.5 a R s = N a-1 a k 1+k 2.0 ~0% ~80% ~10% 1.5 N 1.0 k 0.5 From the examples, selectivity has the greatest impact on increasing peak resolution 0.0 1.00 1.05 1.10 1.15 1.20 1.25 0 5000 10000 15000 20000 25000 0 5 10 15 20 25 a N k Zhao, J.H. and P.W. Carr. Analytical Chemistry, (1999) 71, 2623-2632

Which Factors a Affect Selectivity? 9 MOST Influence Isocratic Separations Gradient Separations Column stationary phase All parameters for isocratic Organic Modifier Gradient steepness ph (ionised analytes only) k* % Organic modifier Dwell volume Buffer selection Column dimensions Column temperature Buffer concentration LEAST Influence a Adapted from Introduction to Modern Liquid Chromatography, 3 rd Edition, Snyder, Kirkland, Dolan, 2010, p.29, Wiley & sons

Influencing Selectivity Bonded Phase Effects / Basic Analytes 10 C18 Increase Retention CN Elution Order C8 (start point) C Decrease Retention Phenyl Elution Order Use ultra high purity silica for good chromatography and reproducibility Column: 250 x.6mm 5mm Mobile phase: 80:20 MeOH/25mM KH 2 PO (ph6.0) Flow: 1.00ml/min Components; 1: Norephedrine, 2: Nortriptyline, 3: Toluene, : Imipramine, 5: Amitriptyline Wavelength: 215nm

HPLC End User Surveys a...listening To The Analyst 11 Column reproducibility and column lifetime are major factors for analysts Have been the top 2 feedback points since 2007 Critical in pharmaceutical and other major industries for method transfers / consistency and long term performance Reversed-phase is the dominant separation mode C18 & C8 = 60%; Phenyl = 16%; CN = 9.5%; Fluorinated = 5.9% 92% analysts use C18 at some time in their work...they typically meet the above criteria BUT limited selectivity a Current Trends in HPLC Column Usage, R.E. Majors, 1Jan12, Modern Medicine website, accessed on 15Feb12: http://www.modernmedicine.com/modernmedicine/article/articledetail.jsp?id=755108

16 Pharmaceutically Relevant Analytes C18 Columns 12 Excel 2 C18 P max : 309 bar Waters Acquity 1.7 BEH C18 P max : 78 bar Phenomenex Kinetex 1.7 C18 P max : 6 bar 1, 2 Agilent Zorbax Eclipse 1.8 XBD C18 P max : 396 bar 1 2 1 1 2 3, 2 5 3 3 3 5 6 5 6 7 6 7 5 6 7 8 7 8 9 8 8 10 9 10, 11 9 10 12 11 10 9 11 12 13 11 16 12 13 1 15 13 1 15 16 16 1 15 1 15 12 16 13 50x2.1mm A: 20 mm KH 2 PO, ph 2.7 B: 20 mm KH 2 PO, ph 2.7 in MeOH/H 2 O (65:35 v/v) Gradient: 3 100 %B in 5 min Flow rate: 0.6 ml/min Temperature: 60C Detection: 21 nm 1. N-Acetylprocainamide 2. 3-Hydroxybenzoic acid 3. Pindolol. Methylphenylsulfoxide 5. Benzyl alcohol 6. Quinoxaline 7. 1,-Dinitrobenzene 8. Phenacetin 9. 1,2-Dimethoxybenzene 10. Furosemide 11. Anisole 12. Methylbenzoate 13. Remacemide 1. Nimesulide 15. Ethylbenzoate 16. Diflunisal 1 2 3 5 C18 phases show similar selectivity... min All trademarks are recognised...comparative separations may not be representative of all applications

The Challenge... 13 To engineer new phases with alternative selectivity but with the robust properties of the C18 ligand Reproducible (column-to-column & batch-to-batch) Excellent column lifetime Superb efficiency provided by ultra-inert, ultra-pure silica particle Low MS bleed Usable in 100% aqueous eluents Available for HPLC & UHPLC separations Available as a Phase III Ready product family Globally available, supply chain, reproducible, multiple batches etc

Aromatic Functionality Engineering New Stationary Phases 1 Phases with aromatic functionality include phenyl and pentafluorophenyl (PFP) based ligands Advantages Aromatic functionality potentially offer unique interactions with analytes (c.f. C18) giving alternative selectivity Provides enhanced retention of polar compounds Many aromatic functionality-based phases can be used in 100% aqueous eluents Disadvantages Phenyl / PFP phases may suffer phase bleed Batch-to-batch reproducibility & robustness may be weak

Aromatic Functionality: π π Interactions 15 A type of electron donor-acceptor interaction Originates from π systems in unsaturated functional groups on analytes and the stationary phase Types of π-π interaction can be manipulated for maximum effect (orthogonality) in phase design eg phenyl: electron rich ring on the stationary phase also acts as π-base and interacts well with electron poor analytes eg PFP: electron poor ring on the stationary phase also acts as π-acid and interacts well with electron rich analytes

The Power of π...scientific Led Stationary Phase Design 16 Electron Donating Groups eg NH 2, NR 2, alkyl, OCH 3 OR, CH 3, Ar etc Si e.g. δ+ δ- Electron Rich Ring Si C 6 H 6 Classic π-π interaction Electron Withdrawing Groups eg NO 2, halides, NR 3+, CO 2 H, CN, CO 2 R, SO 3 H, COH etc Si e.g. F δ- δ- F δ+ F δ- δ- F F δ- Electron Deficient Ring Activity: π-donor (π-base) Activity: π-acceptor (π-acid) How do we exploit these properties for new stationary phases? C18+Phenyl = C18-AR C18+PFP = C18-PFP

Uniquely Designed Stationary Phases 17 C18-AR (USP L1) Ligand has C18 hydrophobic element PLUS phenyl character C18-PFP (USP L1) Ligand has C18 hydrophobic element PLUS PFP character Ultra-inert, ultra-pure silica particle technology as used in all products for high peak efficiency Available in 3, 5 & 10mm, ( ) and 2mm ( Excel TM ) Multi-mode interaction mechanisms result in enhanced chromatographic selectivity giving the analyst new options for method development

C18-AR: Multi-Mode Separation Mechanisms 18 Combining the character of C18+phenyl into a single individual phase harnesses the best of both phases for unique selectivity Separation mechanism Typical C18 Typical Phenyl C18-AR Hydrophobicity ++++ + / ++ ++++ - Interaction - +++ +++ Dipole - Dipole - + + Hydrogen Bonding - ++ ++ Shape Selectivity ++ ++ ++ / +++ The predominance of each retention mechanism will be dictated by the analyte s physicochemical properties, its structure and the chromatographic conditions applied Multi-Mode Interactions Offer the Chromatographer More

C18-AR Aromatic Selectivity 19 Illustrating hydrophobicity and π-base character / aromatic selectivity with a simple example using substituted aromatics 1. TNB 2. DNB 3. NB. Tol NO 2 NO 2 NO 2 NO 2 O 2 N NO 2 O 2 N NO 2 O 2 N NO 2 O 2 N NO 2 Log P* 1.2 1.6 1.9 2.7 π-acidity (order) 1 > 2 > 3 - mau 1 2 3 3 C18 3 2 1 3 Phenyl 150x.6 mm id 1 ml/min 0C, 210nm 1:1 v/v MeOH:H2O 2 3 1 0 2 6 8 10 12 1 16 18 3 C18-AR min * Predicted data from ACD Labs software, 30May12

C18-PFP: Multi-Mode Separation Mechanism 20 Combining the character of C18+PFP into a single individual phase harnesses the best of both phases for unique selectivity Separation mechanism Typical C18 Typical PFP C18-PFP Hydrophobicity ++++ + / ++ ++++ - Interaction - +++ +++ Dipole - Dipole - ++++ ++++ Hydrogen Bonding - +++ +++ Shape Selectivity ++ +++ ++++ The predominance of each retention mechanism will be dictated by the analyte s physicochemical properties, its structure and the chromatographic conditions applied Multi-Mode Interactions Offer the Chromatographer More

C18-PFP Selectivity* 21 Peak Number: 1 2 3 5 6 7 8 Log P: π-basicity (order): 1,2,3-1,2,- 1,2-1,- 1,3-1,3,5 TMB TMB DMB DMB MB DMB TMB Tol 1.7 1.6 1.7 2.1 2.2 2.2 1.6 2.7 1 1 2 2 3 2 1 - Elution / retention not simply a function of π-basicity and Log P Retention mechanism for C18-PFP multi-modal *Structures from www.chemspider.com Predicted data from ACD Labs software, 30May12

C18-PFP Selectivity 22 1,2 3 5,6 7 3 C18 Hypersil GOLD 3 μm PFP Hydrophobic reference 8 C18 or PFP mechanisms alone not enough to fully resolve the methoxybenzene isomers 1 3 5 2 1 8 6 7 3 2 5 6 7 8 Reduced Hydrophobicity 3 C18-PFP C18-PFP mechanism combines hydrophobicity, shape selectivity, dipole-dipole and π-π interactions Elution order, retention and selectivity all seen to differ 5 10 15 20 25 Minutes Powerful positional isomer and shape selectivity 1) 1,2,3-trimethoxybenzene, 2) 1,2,-trimethoxybenzene, 3) 1,2-dimethoxybenzene, ) 1,-dimethoxybenzene 5) methoxybenzene, 6) 1,3-dimethoxybenzene, 7) 1,3,5-trimethoxybenzene, 8) toluene (ref) Mobile phase 50:50 v/v MeOH / H 2 O; Column= 150 x.6 mm id; 1.00 ml/min; 0C; 25 nm All trademarks are recognised...comparative separations may not be representative of all applications

Rt C18-PFP t r column 1 Phase Comparisons With The Selectivity Descriptor* 23 Selectivity = 100 x (1 R 2 ) C18 x x x R 2 = 0.9987 S = 19 C18 S = 15 MeCN x t r column 2 MeOH S = 7 S = 8 6.000 5.000.000 3.000 2.000 1.000 0.000 >100 acidic, basic, neutral analytes y = 1.0019x + 0.1216 R² = 0.9888 0.000 2.000.000 6.000 Rt C18 Selectivity = 100 x (1 R 2 ) = 100 x (1 0.9888) = 10.6 S = 11 S = 18 C18-PFP C18-PFP S = 10 S = 12 S = 18 S = 8 C18-AR S = 15 C18-AR * Neue, O Gara, Méndez Selectivity in Reversed-Phase Separations: Influence of the Stationary Phase, J. Chromatogr. A 1127 (2006), 161-17

Ranking Phase Orthogonality With MeOH and MeCN 2 For the 102 acidic, basic and neutral analytes assessed MeOH MeCN Column 1 Column 2 Selectivity S Column 1 Column 2 Selectivity S C18 C18-AR 12 C18 C18-AR 8 C18 C18-PFP 11 C18-AR C18-PFP 8 C18-AR C18-PFP 10 C18 C18-PFP 7 MeOH MeCN Selectivity Value C18-PFP C18 19 C18-AR C18 18 C18-AR C18-PFP 18 C18-PFP C18-AR 18 C18-PFP C18-PFP 18 C18 C18-AR 17 C18 C18-PFP 17 C18 C18 15 C18-AR C18-AR 15

What Do I Use These Novel Phases For: C18-PFP? 25 Useful for analytes that contain electron donating moieites eg -NH 2, -NR 2, -OCH3, -OH, -alkyl, -Ar etc eg nucleotides, nucleosides, nucleobases, halogenated aryl / aromatics, catecholamines, tetracyclines, beta blockers, structural isomers, coumarins etc Excellent shape and positional isomer selectivity

What Do I Use These Novel Phases For: C18-PFP? 26 C18-PFP: chloroacetophenone halogenated isomers separation

What Do I Use These Novel Phases For: C18-AR? 27 Useful for analytes that contain electron withdrawing moieties eg -NO 2, -halides, -NR 3+, -SO 2, -CO 2 H, -SO 3 H, -CO 2 R, -CHO etc eg aromatic compounds, anthocyanins, steroids, analgesics, phenolics, water soluble vitamins, sulphur containing compounds, quinolones, positional isomers etc Moderate shape selectivity

What Do I Use These Novel Phases For: C18-AR? 28 C18-AR: steroids separation

Combining Selectivity With The NEW Excel Format 29 NEW high efficiency, ultra-inert 2mm silica particles suitable for UHPLC at 1000bar (15000psi) Nine selectivities including the unique C18-AR and C18-PFP High reproducibility: column-to-column and batch-to-batch Ultra-robust phases: NEW low dispersion column hardware and NEW High Stability Column (HSC ) packing technology Engineered with lower back pressures compared to other <2mm phases due to 2mm particle size and frit technology Fully scalable to 3mm, 5mm and 10mm phases Fully compatible with all commercial HPLC and UHPLC kit

Excel UHPLC Column Robustness 30 1000 bar for ~2000 gradient runs Isocratic efficiency assessments every ~100 runs more demanding! Excel 2 C18 (with HSC Packing Technology) Prototype 2 mm C18 (without HSC Packing Technology) 100x2.1mm; MPA 0.1% FA (aq); MPB: 0.1% FA in MeOH; 0.73mL/min; gradient: 20-90%B in 6 mins. NEW High Stability Column (HSC ) Packing Technology Significantly Improves UHPLC Column Robustness

Excel UHPLC Columns Scalability & Reproducibility 31 Excel 2 C18 150 x 2.1mm 0.21mL/min 1 2 3 5 6 2 µm UHPLC 3 C18 150 x 3.0mm 0.0mL/min 3 µm HPLC 5 C18 150 x.6mm 1.00mL/min 5 µm 10 C18 150 x 21.2mm 21.2mL/min 10 µm 0 1 2 3 5 6 7 8 Prep LC MP: 35:65 v/v MeCN:0.1% TFA (aq); 22C; 25nm; 1. uracil; 2. -hydroxybenzoic acid; 3. acetylsalicyclic acid;. benzoic acid; 5. 2-hydroxybenzoic acid 6. ethyl paraben

Excel Has Typically Lower Back Pressure For UHPLC 32 Specifically engineered for lower UHPLC backpressures Excel 2 C18-AR 1 0.95 2 1.22 3 1.69 1.79 5 2.2 6 2.51 8 2.81 7 2.91 9 3.55 10 3.81 11.19 12.7 13.53 P max : 36 bar Waters Acquity 1.7um BEH C18 1 2 0.90 0.87 3 1.8 1.51 5 2.06 7 2.17 6 2.29 8 2.50 10 3.30 9 3.3 12.02 11.09 13.32 P max : 581 bar Phenomenex Kinetex 1.7um C18 1, 2 0.80 3 1.3 5 1.36 1.91 7 2.03 6 2.15 8 2.38 10 12 3.91 13 3.18 9 11.19 3.22 3.9 P max : 50 bar Agilent Zorbax Eclipse 1.8 um XDB C18 2 0.90 1 0.93 3, 1.51 5 2.13 6 2.37 8 2.63 10 3.39 9 3.51 12.13 11.18 13.1 P max : 50 bar 0.5 1 1.5 2 2.5 3 3.5.5 5 min Conditions: A=5mM formic acid (aq); B=5mM formic acid in MeOH; tg= 3 to 100%B in 5 min; 0.6 ml/min; 0C; 25nm 1. Paracetamol 2. Hydrochlorothiazide 3. Methylphenylsulphoxide. Methylphenylsulphone 5. Aspirin 6. Phenacetin 7. 1,3-dinitrobenzene 8. 1,2,-trimethoxybenzene 9. Ethylbenzoate 10. Nimesulide 11. Ibuprofen 12. Indomethacin 13. Mefenamic acid All trademarks are recognised...comparative separations may not be representative of all applications

33 Selectivity, Speed & Scaling Isocratic & Gradient HPLC UHPLC

Excel C18-PFP Selectivity & Throughput (Isocratic) 3 Aim: obtain R s 1.7 in shortest possible time for mixture Waters XBridge 5µm C18 150 x.6 mm 1.00 ml/min 163 bar R s = 1.7 1 2 3 HPLC: 5µm C18 13.6 mins (L/d p = 3) 13.60 min Waters Acquity 1.7µm BEH C18 50 x 2.1 mm 0.21 ml/min 26 bar 3 R s = 1.1 1 2.69 min UPLC: <2µm C18.75 mins (L/d p = 2.9) To maintain R s and reduce run time, keep L / dp ratio constant ~ x23 Quicker Excel 2μm C18-PFP 30 x 2.1 mm 1.30 ml/min 92 bar UHPLC: Excel 2µm C18-PFP < 1 min 2 6 8 10 12 1 16 min 1 R s = 3. 2 3 0.65 min R s = 1.9 Using UHPLC and selectivity, it is possible to dramatically improve resolution allowing shorter columns & increased flow rates Sample: 1) 1,2,dimethoxybenzene, 2) 1,3-dimethoxybenzene, 3) 1,3,5-trimethoxybenzene, ) toluene (reference). Mobile phase 50:50 MeOH / H 2 O; Temperature 0 C; 25 nm All trademarks are recognised...comparative separations may not be representative of all applications

Excel C18-PFP Selectivity & Throughput (Gradient) 35 Aim: obtain R s 1.7 in shortest possible time for mixture 5µm C18 100 x.6 mm 1 ml/min, t G = 29 min max pressure: 92 bar 0 min cycle time R s = 1.8 1 3 2 5 7 6, 22.9 min HPLC: 5µm C18 22.9 mins Excel 2µm C18 50 x 2.1 mm 0.6 ml/min, t G = 5 min max pressure: 367 bar 9 min cycle time Excel 2 μm C18-PFP 30 x 2.1 mm 2.5 ml/min, tg = 0.7 min max pressure: 91 bar 1 min cycle time 0 0 R s = 1.8 0 R s = 1.5 1 3 2 7, 0.61 min R s = 1.9 1 2 3 5 R s = 1.7 6 0.5 1 5 2.5 5 7 R s = 1.7 UHPLC: 2µm C18-PFP 0.61 mins 6,.17 min UHPLC: 2µm C18.17 mins 1.5 2 2.5 3 3.5 25 ~ x25 Quicker min R s 1.7 0 5 10 15 20 25 1, aspirin; 2, phenacetin; 3, 1,3-dinitrobenzene ;, ethylbenzoate; 5, nimesulide; 6, ibuprofen; 7, indomethacin. min

36 Pressure Effects HPLC UHPLC

Background 37 Pressure is a complex physical parameter that affects many elements of a chromatography system Chromatographic selectivity and retention changes at elevated pressures have been investigated and reported a Observations are highly dependent upon the analytes and may be seen with any manufacturer phases operated under UHPLC conditions Changes are typically not helpful for HPLC UHPLC activities a Fallas et al. / J. Chromatogr. A 1209 (2008) 195 205

Effect of Pressure on Selectivity and Retention Factor 38 A B C D Initial 2µm and 3µm data are similar (A, B) Scalability looks good Retention and selectivity seen to change with pressure (BE) E Agilent 1290, 50 x 2.1 mm (constant flow and restrictor capillary used) Mobile phase: A=0.1% FA in water: B=0.1% FA in MeOH (51:9 v/v) Flow Rate: 0.21 ml/min, Temperature: 0 C K= Ketoprofen; S= Sulindac; N=Naproxen a Fallas et al. / J. Chromatogr. A 1209 (2008) 195 205

Summary: Unwanted Selectivity Changes 39 Pressure induced k and a changes may be seen for any manufacturer phases under UHPLC conditions Changes in selectivity and retention may be significant with ionised analytes and large MW analytes a, but the impact on neutral molecules is typically smaller Current discussions / theory focus on changes in analyte molar volume as the principle cause for changes in k and a observed Successful HPLC UHPLC possible...the analyst just needs to be vigilant a Fallas et al. / J. Chromatogr. A 1209 (2008) 195 205

0 Connections : Losses in N and A s Peak Dispersion

Background 1 UHPLC / optimised HPLC instruments are very sensitive to the introduction of extra column volume Any time you install a column (from any manufacturer) it is vital to ensure good connections Aim for a fresh connection every time to ensure a snug fit between tubing and column and reduce the likelihood of an unwanted gap and / or tubing slippage Free movement of the ferrule and nut when installing the column gives you a fresh connection

Losses in Performance Due to Incorrect Column Fitting 2 mau Correctly fitted column 300 1.536 3.17 5.665 8.226 250 200 150 100 50 N 5%h = 12,800 plates PW 5%h = 0.291 min A s = 0.9 Correctly fitted columns make the most of your column and system 0 2 6 8 10 min Incorrectly fitted column mau 300 1.537 3.160 5.713 250 200 8.313 < ~ 0.5mm gap Incorrectly connected columns lead to reduced efficiency, reduced asymmetry and possibly leaks 150 100 50 0 N 5%h = 9,850 plates PW 5%h = 0.335 min A s = 0.8 2 6 8 10 min Loss of ~23% for N Loss of ~11% for A s

Summary: Column Connections 3 Extra column volume reduces peak efficiency and asymmetry Make a fresh connection every time you install any column recommend reusable fittings for a fresh connection every time All Excel TM columns have a FREE Making Great UHPLC Connections leaflet in every box Also downloadable from the website: www.ace-hplc.com

Overall Summary & Conclusions Understanding the properties of building blocks in stationary phase design led to these unique products C18-AR and C18-PFP are powerful tools for method development due to unique but complementary selectivities These unique phases are available for HPLC as the range and also UHPLC as the NEW Excel 2 µm format These phases meet analyst demands of reproducibility, robustness & low phase bleed with excellent peak efficiency Operating at high pressures can deliver excellent results but remain vigilant - selectivity and retention may be affected...and even column connections become critical!

Full Information On All Products Available 5 Unique Selectivities Free Guides MACMOD Analytical = http://www.mac-mod.com/ ACT = http://www.ace-hplc.com

Thank You For Your Attention amckeown@ace-hplc.com www.ace-hplc.com www.mac-mod.com is a registered trademark of Advanced Chromatography Technologies Ltd. Excel and HSC are trademarks of Advanced Chromatography Technologies Ltd. UPLC, Xbridge, ACQUITY are trademarks of Waters Corporation; ZORBAX, Eclipse are trademarks of Agilent Technologies Inc.; Kinetex is a trademark of Phenomenex Inc.; GOLD is a trademark of Thermo Fisher Scientific.