Early catastrophic failure of Birmingham acetabular dysplasia cup in revision arthroplasty: a case report Manjunath Ramappa* and Andrew Port

Similar documents
The Treatment of Pelvic Discontinuity During Acetabular Revision

*smith&nephew CONTOUR

SURGICAL TECHNIQUE. Protrusio Cage A COMPREHENSIVE ACETABULAR REVISION SYSTEM

Clinical and radiological results of the stemmed Mc Minn cup in hip revision surgery

Principles of acetabular fixation in primary and revision hip arthroplasty Piotr Wojciechowski, Damian Kusz, Anna WAGNER

CIRCUMFERENTIAL PROXIMAL FEMORAL ALLOGRAFTS IN REVISION SURGERY ON TOTAL HIP ARTHROPLASTY: CASE REPORTS WITH A MINIMUM FOLLOW-UP OF 20 YEARS

RECOVERY. P r o t r u s i o

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.

Trabecular Metal Acetabular Revision System Buttress and Shim Augments Surgical Technique

Bone Bangalore

A 42-year-old patient presenting with femoral

Optimum implant geometry

Optimum implant geometry

Femoral Revision Algorithm. A practical guide for the use of the CORAIL Hip System in femoral revision surgery

Femoral neck fracture during physical therapy following surface replacement arthroplasty: a preventable complication?

Porous metal augments

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

Bone Preservation Stem

ACETABULAR CUP SURGICAL TECHNIQUE

The Leader in Orthopaedic Innovation

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

Revision Total Hip Replacement

Treatment of Periprosthetic Femoral Fractures Following Hip Arthroplasty

Short-term Results of a Custom Triflange Acetabular Component for Massive Acetabular Bone Loss in Revision THA

Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases

Peggers Super Summaries: THR and Revision Concepts

Optimizing function Maximizing survivorship Accelerating recovery

Fixed and Variable Geometry Total Shoulder Arthroplasty

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

Optimum implant geometry

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

CC TRIO VERSAFITCUP. Surgical Technique. each to their own. Hip Knee Spine Navigation

ADEPT Extra Fixation Cup. Operative Technique. Delivering Results Through Performance

CONGENITAL HIP DISEASE IN YOUNG ADULTS CLASSIFICATION AND TREATMENT WITH THA. Th. KARACHALIOS, MD, DSc PROF IN ORTHOPAEDICS

Metha Short Hip Stem System

Trilogy Acetabular System

Early results of Trabecular Metal augment for acetabular reconstruction in revision hip arthroplasty

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION

ESC. Enhanced Stability Liners. Design Rationale & Surgical Technique

DESIGN RATIONALE AND SURGICAL TECHNIQUE

The shape and size of femoral components in revision total hip arthroplasty among Chinese patients

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

pact SYSTEM Surgical Technique HEMISPHERICAL CEMENTLESS CUP SYSTEM MULTI-HOLE & RIM-HOLE Hip Knee Spine Navigation

ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT

One Stage or Two Stage

Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant stability

Pelvic discontinuity

Recently, the new generation of metal-on-metal total hip resurfacing. arthroplasty is well known for preserving the proximal femoral bone stock,

WHAT IS THE BEST BONE FIXATION TYPE? 2/11/2011

Comparative Study of Peripheral Rim Fixation Using Jumbo Cup in Revisional Hip Arthroplasty

FEMORAL REVISION IN TOTAL HIP ARTHROPLASTY

Dual Mobility Cups. Kris Govaers, MD, PhD Dendermonde Belgium

Hip Resurfacing System

28 Surgical Technique

Fracture of an Acetabular Component Inserted without Cement: A Case Report

A Proven Concept. Alloclassic Zweymüller Stem

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

TOTAL HIP REPLACEMENT:

Evaluating and Treating Acetabular Bone Loss with Pelvic Discontinuity

OPTIONAL/ADDITIONAL INSTRUMENTS

Trilogy Acetabular System

Trilogy Acetabular System

Tradition Hip Primary Surgical Technique

PRODUCT RATIONALE & SURGICAL TECHNIQUE

THE NATURAL FIT. Surgical Technique. Hip Knee Spine Navigation

PERIPROSTHETIC FRACTURES FOLLOWING TOTAL HIP ARTHOPLASTY

Salvage of failed dynamic hip screw fixation of intertrochanteric fractures

Surgical Technique r5.indd 1 12/8/10 10:36 AM

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Trabecular Metal Acetabular Restrictor and Augment

EVOLVING OUR HERITAGE, MEETING YOUR NEEDS. Surgical Technique

TRABECULAR METAL ACETABULAR RESTRICTOR AND AUGMENT. Surgical Technique

Design Rationale. ECHELON Primary Hip System

Technique Guide Hip Resurfacing System

operative technique Kent Hip

Introduction The oval cup MEDIN is used in clinic praxis since the 1999 year. During the period of two years it has been used for more than 1000 patie

Trinity. Advanced Bearing Acetabular System Surgical technique

Surgical Technique. Cup System

Posterolateral elbow dislocation with entrapment of the medial epicondyle in children: a case report Juan Rodríguez Martín* and Juan Pretell Mazzini

Outcomes of Surgical Treatment of Periprosthetic Femoral Fractures in Cementless Hip Arthroplasty

REVISING THE DEFICIENT PROXIMAL FEMUR

Revision. Hip Stem. Surgical Protocol

R/F. Can T-smart Tomosynthesis Improve Diagnostic Accuracy on THA Component Stability? 1. Abstract

Featuring. Technology. Product Rationale

Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Trabecular Metal Acetabular Restrictor and Augment

Revision Total Hip Arthroplasty Using Tantalum Augment in Patients with Paprosky III or IV Acetabular Bone Defects: A Minimum 2-year Follow Up Study

Case Report Bilateral Distal Femoral Nailing in a Rare Symmetrical Periprosthetic Knee Fracture

DIRECT ANTERIOR APPROACH

Hip Resurfacing.

Approach Patients with Confidence

PINNACLE REVISION CUP SYSTEM

Metasul LDH Large Diameter Head

CORAIL HIP SYSTEM DESIGN RATIONALE

ZMR CROSSOVER INSTRUMENTS AND SURGICAL TECHNIQUE. Surgical Technique for Revision Hip Arthroplasty

Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

Cotyloplasty in Cementless Total Hip Arthroplasty for an Insufficient Acetabulum

Department of Orthopaedic Surgery, University of Toledo Medical Center, Toledo, OH, US.

Transcription:

Open Access Case report Early catastrophic failure of Birmingham acetabular dysplasia cup in revision arthroplasty: a case report Manjunath Ramappa* and Andrew Port Address: Department Of Orthopaedics, James Cook University Hospital, Middlesbrough, TS4 3BW, UK Email: MR* - drbrm2004@yahoo.co.uk; AP - Andrew.port@stees.nhs.uk * Corresponding author Received: 17 August 2008 Accepted: 2 April 2009 Published: 30 July 2009 Cases Journal 2009, 2:8716 doi: 10.4076/1757-1626-2-8716 This article is available from: 2009 Ramappa and Port; licensee Cases Network Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Abstract Introduction: Revision acetabulum arthroplasty is one of the common procedures, which has been on the rise recently. Many implants are available in the market with variable results. Aseptic loosening is the most common indication for revision acetabulum arthroplasty. Birmingham dysplasia cup has been used occasionally in these complex procedures. We know that these implants have provided good results in resurfacing dysplastic hips with bone loss. Literature review failed to answer whether these implants were as effective in revision acetabulum arthroplasty. Case presentation: We herby, present a case of revision acetabulum arthroplasty performed with Birmingham dysplasia acetabular cup, which unfortunately failed within a period of eight months. Surgical technique appeared to be adequate. On further analysis, significant differences were identified between dysplastic hips with bone loss and revision hips with bone loss. Conclusion: Therefore results obtained with dysplasia cups in resurfacing dysplastic hips does not seem to be applicable to revision hip arthroplasty. Hence these cups should be restricted to primary arthroplasty. Introduction Revision acetabulum arthroplasty presents a challenge for arthroplasty surgeons. Over the years, choice of implants has constantly changed, in the quest for an ideal implant. Cementless acetabular components seem to be a favourable option at present. Its survival depends on its design, initial stability, any osteoconductive or osteoinductive properties and available host tissue. Many strategies have been utilised for stabilisation to improve long term outcome. They include medialisation, proximal positioning of the hip centre, bone grafting, supplementary screw fixation, metal mesh or reinforcement rings, capture cups and different combinations of these techniques. Birmingham acetabular dysplasia cup is one such implant, which has occasionally been utilised in revision surgery. It has provided good results in hip resurfacing in dysplastic hips with severe bone loss [1] and this seems to have created hope for utilisation in revision surgery. It consists of an uncemented hydroxyappatite coated porous surface Page 1 of 5

with two offset threaded lugs for supplementary neutralisation screws. The acetabular component and the screws combine into one solid composite 3 dimensional construct, where the stability in any plane is a function of the area circumscribed by the component and the screws as a whole [1]. It has been utilised effectively in dysplastic hips with bone defects. But does it provide similar effect when used in revision arthroplasty was the reason behind this report. Case presentation A 71 year old Caucasian lady of British origin, presented with aseptic loosening of left charnley total hip replacement, performed 20 years ago (Figure 1). Radiography revealed acetabular bone stock deficiency in the superolateral and inferior quadrants. She underwent Left revision total hip replacement with a Birmingham resurfacing dysplasia cup (HA coated size 52 mm) and femoral bone allograft in superolateral compartment, Echelon revision collared stem (190 mm) and calcar substitute, with head size measuring 42 mm. Acetabular cup was anchored with 2 suprolateral screws. Postoperative recovery was uneventful. Initial postoperative X-ray (Figure 2) showed a minimal interference gap between the acetabular cup and bone. Cup and screw positions were satisfactory. Initial mobilisation was with 2 elbow crutches for 6 weeks, 1 elbow crutch for another 6 weeks and finally onto 1 stick. Patient s recovery was satisfactory at 6 weeks, 3 and 6 months follow up. At 8 months follow up she complained of worsening left hip pain especially with mobilisation. X-ray (Figure 3) confirmed broken acetabular screws with minimal migration of cup. Sepsis was ruled out after thorough clinical, radiological, biochemical Figure 2. Initial postoperative radiograph showing a minimal interference gap between the acetabular cup and bone. Cup and screw positions are satisfactory. and microbiological investigations. At 12 months there was a progressive posterior inferior migration of acetabular cup (Figures 4, 5). Figure 1. Pre operative radiograph showing loosening of left charnley total hip replacement, performed 20 years ago. Discussion One of the major causes for aseptic loosening in the cementless acetabular cup is its insufficient initial stability [2]. This in turn leads to fibrous tissue interposition at the metal bone interface and subsequent mechanical fatigue, due to loss of osseointegration. Birmingham dysplasia cup was produced for utilisation in dysplastic hips. Uncemented HA coated Birmingham dysplasia acetabular cups rely initially on a stable interference fit, maintained by fixation of peripheral screws, followed by biologic fixation. They have provided good results in resurfacing dysplastic hips with extensive bone loss [1]. Page 2 of 5

Cases Journal 2009, 2:8716 Figure 3. Radiograph at 8th month postoperative period showing broken acetabular screws. But is primary arthroplasty in a dysplastic acetabulum with bone stock deficiency equivalent to revising an acetabular component with bone stock deficiency and aseptic loosening. If we considered patients with dysplastic hips to be in group A and patients with aseptic failure of acetabular cups to be in group B, a comparison of these two groups would identify the suitability of this dysplasia cup in the two groups. Dysplastic patients (group A) are usually young and active in contrary to the patients in the revision group (group B) Figure 5. Radiograph at 12th month postoperative period showing a posterior inferior migration of acetabular cup in Lateral view. who are more often in the elderly population. Postoperative rehabilitation is much faster and easier in younger patients. Subsequently the complication rates are much lower in younger patients. Osteoporosis, especially post menopausal, is a common finding in group B, which is a relative contraindication for resurfacing with Birmingham dysplasia cup, both due to risk of neck fracture and acetabular loosening due to poor bone quality (B). The use of any revision acetabular component must optimize the use of whatever bone remains after primary hardware is removed. Variation in the bony and soft tissue anatomy, especially soft tissue contractures and muscle weakness in the two groups can make comparisons difficult. Figure 4. Radiograph at 12th month postoperative period showing a posterior inferior migration of acetabular cup in AP view. Though acetabular bone deficiency can be seen in both groups, in group A there is enough sclerotic bone in the false acetabulum for the two neutralisation screws. The screws obtain good purchase in sclerotic bone of false acetabulum, rather than osteopenic bone of in the acetabular roof. (A) Whereas in group B, aseptic loosening and biological effects of wear debris provoke peri prosthetic osteolysis Page 3 of 5

and loosening of the implant. Sclerotic bone is more often than not, inadequate in this group. In group A, as the first screw advances through the lug and reaches the prepared screw hole on the bony surface, further advancement should be very meticulous to gain initial purchase without disengaging the acetabular component from its bed. After initial purchase, the screw advances without difficulty and tightens the cup against the bone towards the end by locking. Once this screw is inserted the second screw can be inserted with relative ease. Whereas in group B, initial purchase, which is so essential can be difficult due to lack of sclerotic bone. Therefore cup has the possibility of relatively more micromovement during insertion of both first and second screws. If stable interface fit between the acetabulum cup and bone is not achieved, screws can act as potential pivot points leading to an increased micromotion which will eventually smoothen the cup surface. Smooth surface HA coated cups are known [3,4] to be associated with increased rate of aseptic loosening due to osteolysis and unstable fixation. So the acetabular component now relies heavily on the screws for positional stability. During weight transmission this can also cause differential movements between the screw, flange and cup. This not only causes fatigue failure of the screw but also loss of biologic fixation. Excess removal of acetabular bone is usually not a problem when these cups are used in group A [5]. In revision surgery (group B), poor bone quality and bone loss is a common finding [6]. Bone loss during cup insertion or reaming [7] could further compromise cup fixation. We know that it is difficult to estimate accurate bone loss preoperatively by radiographs. Also, bone allograft has influence on failure rates of uncemented sockets [8] in the revision setting. Theoretically, screw fixation can close any gap created by non congrous reaming [9]. Eccentric drill holes can theoretically lift the cup out of its bony bed during screw insertion, especially in revision surgery [2]. The ilium provides the least amount of support to immediate acetabular fixation, while the pubis (anterior column) and ischium (posterior column) provide more stability [10]. Eccentric screws in the ilium, in acetabuar cups with insufficient initial stability, like in group B, can therefore fail to decrease motion at the pubis or ischium significantly [10]. This can lead to fatigue failure. In addition, locking screws will not provide compression to decrease any persistent gap between the acetabular cup=and its bed. Therefore, it is obvious that dysplastic acetabulum with bone deficiency is a completely different problem to tackle with, as compared to a revision acetabulum. Hence, good results obtained from Birmingham dysplasia cups in resurfacing dysplastic hips with severe acetabular insufficiency may not indicate its suitability for use in revision acetabulum arthroplasty. Conclusion Birmingham acetabular dysplasia cup does not seem to be as forgiving a tool in revision hip arthroplasty, as compared to dysplastic hips with bone loss. Obtaining stable initial purchase during screw insertion can be difficult in revision acetabulum arthroplasty. HA coating and screw fixation does not compensate for interference gap and subsequent micromotion between the dysplasia cup and its bony bed. Therefore this cup should be restricted to primary surgery alone. Consent Written informed consent was obtained from the patient for publication of this case report and accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal. Competing interests I can confirm that there are no competing interests declared by either author. Authors contributions MR and AP have made substantial contributions to conception and design, or acquisition of data, or analysis and interpretation of data; have been involved in drafting the manuscript or revising it critically for important intellectual content; and have given final approval of the version to be published. References 1. McMinn DJ, Daniel J, Ziaee H, Pradhan C: Results of the Birmingham Hip Resurfacing dysplasia component in severe acetabular insufficiency: A six- to 9.6-year follow-up. J Bone Joint Surg Br 2008, 90:715-723. 2. Hsu JT, Chang CH, Huang HL, Zobitz ME, Chen WP, Lai KA, An KN: The number of screws, bone quality, and friction coefficient affect acetabular cup stability. Med Eng Phys 2007, 29:1089-95. 3. Kim SY, Kim DH, Kim YG, Oh CW, Ihn JC: Early failure of hemispheric hydroxyapatite-coated acetabular cups. Clin Orthop Relat Res 2006, 446:233-238. 4. Cheung KW, Yung SH, Wong KC, Chiu KH: Early failure of smooth hydroxyapatite-coated press-fit acetabular cup-7 years of follow-up. J Arthroplasty 2005, 20:627-631. 5. McBryde CW, Shears E, O Hara JN, Pynsent PB: Metal-on-metal hip resurfacing in developmental dysplasia: A case-control study. J Bone Joint Surg Br 2008, 90:708-714. 6. Nixon M, Taylor G, Sheldon P, Iqbal SJ, Harper W: Does bone quality predict loosening of cemented total hip replacements? J Bone Joint Surg Br 2007, 89:1303-1308. 7. Loughead JM, Starks I, Chesney D, Matthews JN, McCaskie AW, Holland JP: Removal of acetabular bone in resurfacing arthroplasty of the hip: a comparison with hybrid total hip arthroplasty. J Bone Joint Surg Br 2006, 88:31-34. 8. Brady OH, Masri BA, Garbuz DS, Duncan CP: Use of reconstruction rings for the management of acetabular bone loss during revision hip surgery. J Am Acad Orthop Surg 1999, 7:1-7. Page 4 of 5

9. Hsu JT, Chang CH, An KN, Zobitz ME, Phimolsarnti R, Hugate RR, Lai KA: Effects of screw eccentricity on the initial stability of the acetabular cup. Int Orthop 200731:451-455. 10. Perona PG, Lawrence J, Paprosky WG, Patwardhan AG, Sartori M: Acetabular micromotion as a measure of initial implant stability in primary hip arthroplasty. An in vitro comparison of different methods of initial acetabular component fixation. J Arthroplasty 19927:537-547. Do you have a case to share? Submit your case report today Rapid peer review Fast publication PubMed indexing Inclusion in Cases Database Any patient, any case, can teach us something www.casesnetwork.com Page 5 of 5