I-gel vs cuffed tracheal tube during volume controlled ventilation in elective laparoscopic cholecystectomy

Similar documents
I - Gel Versus Cuffed Tracheal Tube in Elective Laparoscopic Cholecystectomy A Clinical Comparative Study

RESPIRATION AND THE AIRWAY Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation

Deposited on: 24 February 2009

Deposited on: 11 February 2010

ISSN X (Print) Research Article. *Corresponding author Dr. Souvik Saha

A Comparative Study of Two Disposable Supraglottic Devices in Diagnostic Laparoscopy in Gynecology

Comparison of clinical performance of i-gel with laryngeal mask airway pro-seal in elective surgery in adults

Dr. Ranjeet Rana De 1, Dr. Saurav Shekhar 2, Dr. D G Pathak 3, Dr. Harshwardhan 4, Dr. Shashank Dhiraj 5 1,2,4,5

A randomised comparison between Cobra PLA and classic laryngeal mask airway and laryngeal tube during mechanical ventilation for general anaesthesia

A comparative study between i-gel and classical laryngeal mask airway in elective surgery under general anaesthesia

Citation British journal of anaesthesia, 104. pp ; 2010 is available onlin

Other methods for maintaining the airway (not definitive airway as still unprotected):

Use of the Intubating Laryngeal Mask Airway

Original Contributions

Comparison of the i-gel and the LMA-Unique laryngeal mask airway in patients with mild to moderate obesity during elective short-term surgery

A Comparative Study of Classic LMA and Proseal LMA in Paralyzed Anaesthetized Patients

Evaluation of Baska Mask Performance in Laparoscopic Cholecystectomy

ORIGINAL ARTICLE. Shashank Chitmulwar, MD, Charulata Deshpande, MD, DA ABSTRACT. ANAESTHESIA, PAIN & INTENSIVE CARE

The Laryngeal Mask and Other Supraglottic Airways: Application to Clinical Airway Management

Comparison of the LMA Supreme vs the i-gel TM in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation*

COMPARATIVE CLINICAL EVALUATION OF PROSEAL LMA WITH I - GEL AIRWAY IN PATIENTS UNDERGOING ELECTIVE SURGICAL PROCEDURES UNDER GENERAL ANAESTHESIA

International Journal of Health Sciences and Research ISSN:

Comparison of I-gel with Baska Mask Airway for Controlled Ventilation in Obese Patients Undergoing Ambulatory Surgery: A Prospective Randomized Trial

Observation of ventilation effects of I-gel TM, Supreme TM and Ambu AuraOnce TM with respiratory dynamics monitoring in small children

Recent Advances in Airway Management HA Convention 2014

Cricoid pressure: useful or dangerous?

Materials and Methods

AIRWAY MANAGEMENT AND VENTILATION

LMA Supreme Second Seal. Maintain the airway. Manage gastric contents. Meet NAP4 recommendations.

Tibe ProSeal Laryngeal Mask Airway

ISPUB.COM. M Roberts, M Mani, A Wilkes, E Flavell, N Goodwin INTRODUCTION

The LMA CTrach TM, a new laryngeal mask airway for endotracheal intubation under vision: evaluation in 100 patients

Anatomy and Physiology. The airways can be divided in to parts namely: The upper airway. The lower airway.

Comparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients.

The Laryngeal Tube. An Evaluation of the Laryngeal Tube During General Anesthesia Using Mechanical Ventilation

Comparative evaluation of Ambu AuraGain with ProSeal laryngeal mask airway in patients undergoing laparoscopic cholecystectomy

A Clinical Comparative Study Of Evaluation Of Proseal LMA V/S I-GEL For Ease Of Insertion And Hemodynamic Stability; A Study Of 60 Cases

Comparison of Ease of Insertion and Hemodynamic Response to Lma with Propofol and Thiopentone.

Positive pressure ventilation with the laryngeal mask airway in non-paralysed patients: comparison of sevoflurane and propofol maintenance techniques

JMSCR Vol 4 Issue 02 Page February 2016

Comparison of the Hemodynamic Responses with. with LMA vs Endotracheal Intubation

Angkana Lurngnateetape,, MD. Department of Anesthesiology Siriraj Hospital

A prospective randomized study comparing the efficacy of the LMA Classic TM, the

Advanced Airway Management. University of Colorado Medical School Rural Track

DIFFICULT AIRWAY MANAGMENT. Dr.N.SANTHOSH KUMAR MD ANESTHESIA (2 nd Yr)

Equipment: NRP algorithm, MRSOPA table, medication chart, SpO 2 table Warm

Anesthesia Final Exam

Comparison of Three Disposable Extraglottic Airway Devices in Spontaneously Breathing Adults

Unanticipated difficult tracheal intubation - during routine induction of anaesthesia in an adult patient

Emergency Department/Trauma Adult Airway Management Protocol

Received 1 November 2009 Revised 8 December 2009 Accepted 10 December 2009

Comparison of efficacy of the Laryngeal tube with the Laryngeal mask airway in securing the upper airway

Comparison of the clinical performance of i-gel, LMA Supreme and LMA ProSeal in elective surgery

Airway Anatomy. Soft palate. Hard palate. Nasopharynx. Tongue. Oropharynx. Hypopharynx. Thyroid cartilage

Use of the Aintree Intubation Catheter with the Laryngeal Mask Airway and a Fiberoptic Bronchoscope in a Patient with an Unexpected Difficult Airway

A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures

CONTINUING EDUCATION IN HONOR OF NORMAN TRIEGER, DMD, MD

ADVANCED AIRWAY MANAGEMENT

Optimal Size AMBU Laryngeal Mask Airway Among Asian Adult Population

Waitin In The Wings. Esophageal/Tracheal Double Lumen Airway (Combitube ) Indications and Use for the Pre-Hospital Provider

Randomized Comparison of Actual and Ideal Body Weight for Size Selection of the Laryngeal Mask Airway Classic in Overweight Patients

Original Research Article. Amol P. Singam 1, Arpita A. Jaiswal 2 *, Ashok R. Chaudhari 1

Airway Management. Teeradej Kuptanon, MD

Overview. Chapter 37. Advanced Airway Techniques. Sellick Maneuver 9/11/2012

Dr.Bharghavi.M 2 nd year post graduate Dept of Anaesthesia

Adult Intubation Skill Sheet

Airway/Breathing. Chapter 5

Difficult Airway. Department of Anesthesiology University of Colorado Health Sciences Center (prepared by Brenda A. Bucklin, M.D.)

The Pro-Seal LMAtm And The Tracheal Tube: A Comparison Of Events At Insertion Of The Airway Device

This interdisciplinary clinical support document provides guidelines for the safe establishment of an artificial airway.

Randomised comparison of the LMA Supreme with the I-Gel in spontaneously breathing anaesthetised adult patients

LEVITAN S FIBREOPTIC STYLET: BEYOND BARRIERS. - Our Perspective.

Advanced Airway Management

Airway Management & Safety Concerns Experience from Bariatric Surgery

We will not be using the King LTS-D in our system!

Comparison of the Baska â mask with the single-use laryngeal mask airway in low-risk female patients undergoing ambulatory surgery

Airway/Breathing. Chapter 5

Pharyngolaryngeal Morbidity with the Laryngeal Mask Airway in Spontaneously Breathing Patients

Veena Patodi*, Maina Singh, Surendra K. Sethi, Vini N. Depal, Neena Jain, Vijay Kumar

REVISTA BRASILEIRA DE ANESTESIOLOGIA

Airway Workshop Lecture. University of Ottawa

University of Groningen

Comparison of the air-q ILA and the LMA-Fastrach in airway management during general anaesthesia

Saumya Biswas *, Soumen Mandal **, Tapobrata Mitra ***, Suprio De Ray ****, Ranabir Chanda ***** and Debajyoti Sur ******

1 Chapter 40 Advanced Airway Management 2 Advanced Airway Management The advanced airway management techniques discussed in this chapter are to

Z.U.M.J.Vol.19; N.5; September; 2013

Clinical Study Performance of Size 1 I-Gel Compared with Size 1 ProSeal Laryngeal Mask in Anesthetized Infants and Neonates

Cronicon ANAESTHESIA. Khaled Mohammed Elnaghy 1 and Javier Yuste 2 * Abstract. Keywords: Intubation; Video laryngoscopes; GlideScope; LMA CTrach

GENERAL ANAESTHESIA AND FAILED INTUBATION

Use of laryngeal mask airway compared to endotracheal tube with positive-pressure ventilation in anesthetized swine

General Medical Procedure. Emergency Airway Techniques (General Airway Protocol)

The effect of head rotation on efficiency of ventilation and cuff pressure using the PLMA in pediatric patients

A Comparative Study Between Proseal Laryngeal Mask Airway And Endotracheal Intubation.

Airway Management Essentials Self-Study Guide

Chapter 40 Advanced Airway Management

J of Evolution of Med and Dent Sci/ eissn , pissn / Vol. 3/ Issue 64/Nov24, 2014 Page 13923

Airway Management and The Difficult Airway

Airway Management. Key points. Rapid Sequence Intubation. Rapid Sequence Intubation Recognizing difficult airway Managing difficult airway

Original Article INTRODUCTION. Abstract

J. Brimacombe, 1 L. Holyoake, 2 C. Keller, 3 J. Barry, 4 D. Mecklem, 4 A. Blinco 5 and K. Weidmann 5

Transcription:

Egyptian Journal of Anaesthesia (2011) 27, 7 11 Egyptian Society of Anesthesiologists Egyptian Journal of Anaesthesia www.elsevier.com/locate/egja www.sciencedirect.com Research Article I-gel vs cuffed tracheal tube during volume controlled ventilation in elective laparoscopic cholecystectomy Mohamed Ibrahim a, *, Ashraf Ragab b, Hossam ElShamaa b a Department of Anesthesiology, Faculty of Medicine, Zagazig University, Egypt b Department of Anesthesiology, Faculty of Medicine, Cairo University, Egypt Received 28 November 2010; accepted 5 December 2010 Available online 18 February 2011 KEYWORDS I-gel; Leak fraction; Pneumoperitoneum; Cholecystectomy Abstract Background: In this study we tested the new, single-use supraglottic airway device; i-gel with a n\on-inflatable cuff as an effective airway as conventional endotracheal tube. Gas leak was measured and compared with that of cuffed tracheal tube during volume controlled ventilation in elective laparoscopic cholecystectomy. Methods: Sixty patients, ASA I II, were randomly selected to the study. Standard anaesthetic technique was used for all patients. The i-gel was then inserted. The lungs were ventilated at three different tidal volumes (6, 8 and 10 ml kg 1 ) using volume controlled ventilation (VCV). The leak volume was calculated as the difference between the inspired and expired tidal volumes. The leak fraction was also calculated as the leak volume divided by the inspired tidal volume. These observations were recorded with every tidal volume before and after pneumoperitoneum with the i-gel and the conventional tracheal tube. * Corresponding author. E-mail address: mibrahim72@hotmail.com (M. Ibrahim). 1110-1849 ª 2011 Egyptian Society of Anesthesiologists. Production and hosting by Elsevier B.V. Open access under CC BY-NC-ND license. Peer review under responsibility of Egyptian Society of Anesthesiologists. doi:10.1016/j.egja.2010.12.003 Production and hosting by Elsevier

8 M. Ibrahim et al. Results: Before pneumoperitoneum there was no significant difference in leak fraction between i-gel and tracheal tube at 6 and 8 ml kg 1 tidal volume. Significant differences were found after pneumoperitoneum at 8 and at 10 ml kg 1 before and after pneumoperitoneum. Conclusion: We suggest that i-gel can be used as an alternative device to endotracheal tube during VCV for laparoscopic cholecystectomy provided that peak pressure does not exceed leak pressure. ª 2011 Egyptian Society of Anesthesiologists. Production and hosting by Elsevier B.V. Open access under CC BY-NC-ND license. 1. Introduction Supraglottic airway devices are used now for airway control during spontaneous and controlled ventilation under general anesthesia. The i-gel (Intersurgical Ltd., Wokingham, UK) is a new supraglottic airway device (SAD) made of thermoplastic elastomer which is soft, gel-like and transparent [1]. The i-gelä is an anatomical device achieving a mirrored impression of those structures without causing multidirectional forces of compression or displacement trauma to the tissues and structures in the vicinity. A drain tube is placed laterally to the airway tube that allows the insertion of a gastric tube (maximum 14 F gauges). Studies on Cadaver showed that i-gels effectively conformed to the perilaryngeal anatomy and consistently achieved proper positioning for supraglottic ventilation [2]. Manikins studies and patients have shown that the insertion of the i-gel was significantly easier when compared with insertion of other SADs [3,4]. Few studies had been done to evaluate the use of i-gel during controlled ventilation but they did not evaluate its use during procedures with airway pressure more than 25 cm H 2 O [5]. Our study was designed to evaluate the i-gel sealing pressure and as effective airway as cuffed tracheal tube during volume controlled ventilation in laparoscopic cholecystectomy. 2. Method After ethics committee of New Jeddah Clinic Hospital (Jeddah, Saudi Arabia) approval and written informed consent 60 patients, 18 60 years old, ASA I II, Mallampati class I or II, male and female were selected for elective laparoscopic cholecystectomy. Patients were randomly selected for airway management by either i-gel (I-group) or cuffed endotracheal tube (T-group) and data reading done before (B) and after (A) pneumoperitoneum to 14 mmhg in supine position. Patients with body mass index >35 kg m 2, difficult airway possibility, cervical spine problems, mouth opening <2.5 cm, full stomach and gastroesophageal reflux disease, were excluded. All patients received premedication midazolam 1 mg, ranitidine 50 mg and dansetron 4 mg intravenously, 45 min before surgery. We used Datex-Ohmeda; Aisys (GE healthcare) with standard monitor. Anesthesia protocol was made the same for all patients. Preoxygenation then induction of anesthesia was done by propofol 2 2.5 mg kg 1, fentanyl 1 1.5 lg kg 1 and neuromuscular relaxation achieved by rocuronium 0.8 mg kg 1 with increments of 0.15 mg kg 1 boluses to maintain TOF >1. Ventilation by face mask was done till adequate depth of anesthesia and relaxation. Anesthesia was maintained by oxygen/nitrous oxide, sevoflurane mixture. I-gel device (Fig. 1) was carefully selected according to manufacturer recommendations. Patient weighing 50 90 kg, size 4 was used and patients weighing above 90 kg, size 5 was appropriate for them. Proper preparation, lubrication with water soluble lubricant was done. Senior anesthetist inserted the i-gel by continuous introduction into the mouth against hard palate till resistance felt as recommended by the manufacturer. Correct placement of the device was confirmed by observation of proper chest expansion, square shape of end tidal CO 2 waveform and absence of audible leak sounds. In case the airway and ventilation were not established properly; gentle pushing, pulling, head extension, jaw thrust, or neck flexion manipulations were tried before considering failed attempt. Failed attempt was recorded when the device removed from the mouth. Three attempts were allowed before considering failed attempt. Device was then tapped over the chin and connected to anesthesia machine. Gastric tube is lubricated and inserted down the gastric drainage port. In T-group the trachea of the participant was intubated with an appropriate size tracheal tube: size 8.5 was used for the male participants and size 7.5 was used for the female participants. Fresh gas flow was adjusted at 3 l min 1. Leak pressure then measured by closing the expiratory valve and recording airway pressure (not allowed to be more than 40 cm H 2 O) at which equilibrium was achieved. Equilibrium point was identified by either: (1) plateau on pressure time curve or digitally displayed pressure gauge. (2) Pressure at which audible gas was heard by auscultation of lateral aspect of thyroid cartilage. The patient was then ventilated by volume controlled ventilation with three tidal volumes (6, 8 and 10 ml kg 1 ), inspiratory to expiratory ratio 1:2 without positive end expiratory pressure. Respiratory rate was adjusted to maintain 35 45 mmhg Figure 1 Components of the i-gel (adopted from i-gel user guide, Intersurgical Co.).

I-gel vs cuffed tracheal tube during volume 9 EtCO 2. Gastric distension was detected by auscultation of epigastrium or as noted by surgeon intra-operative. Inspiratory (ITV) and expiratory (ETV) tidal volumes were recorded and leak volume (LV) was calculated as the difference between both volumes (LV = ITV ETV). The leak fraction was calculated as leak volume divided by ITV (i.e. leak fraction LV/ ITV). Ventilation was started by tidal volume 6 ml kg 1 for 3 min with data recording then the tidal volume increased to 8 and 10 ml kg 1 for the same period with same data recording. After pneumoperitoneum same ventilation and data collection was repeated. Ventilation was maintained at the parameters that achieve best oxygenation and ventilation. Goals were to maintain SpO 2 > 95% and EtCO 2 < 45 mmhg. Oxygenation was considered sub-optimal if SpO 2 was 90 94% and failed if SpO 2 < 90%. Suboptimal ventilation was considered when EtCO 2 45 53 mmhg and failure if EtCO 2 > 53 mmhg. Peak airway pressure was recorded pre- and post-pneumoperitoneum. If oxygenation or ventilation failed during procedure, surgeons were asked to hold and release of pneumoperitoneum then preoxygenation and endotracheal tube insertion. The primary outcome of our study was difference in the leak fraction between two airway devices under investigation before and after pneumoperitoneum (B and A), with different tidal volumes. Secondary outcomes include: number of insertion attempts; presence of gastric insufflations; leak pressure; leak volume; and ventilatory variables. Also complications occurring during insertion, and removal were noted for each patient. Sore throat, coughing, gastric distension, regurgitation, aspiration, and dysphonia were evaluated after recovery. The sample size was determined by considering a difference in the leak fraction of more than 20% for the i-gel when compared with the tracheal tube to be significant. Previous studies had used a difference of 20% and 25% in the leak fraction for power calculation according to their institute s anesthetist concepts [5]. Parametric data were analyzed by paired and unpaired t-tests. Non-parametric data were analyzed with the v 2 -test. Data are mean (SD, range) or number of patients unless otherwise mentioned. A standard deviation value of (0.15) was used for the leak fraction from a previous study performed with conventional laryngeal mask airways (LMAs) [1]. A two-sample study design, used a t-test for comparison of group means, would therefore require a total of 60 patients for 90% power at a significance level of 5%. Table 1 Demographic data (mean ± SD or n). I group (i-gel) T group (ETT) Age 40 ± 18.1 38.9 ± 16.3 Sex Male 15 17 Female 15 13 Weight 73 ± 10.2 74.2 ± 9.3 Height 169 ± 8 170 ± 5 BMI 26 ± 1.4 27 ± 0.2 difference between i-gel and tracheal tube both before and after pneumoperitoneum (P = 0.002 and 0.006). Leak volume at 6 and 8 ml kg 1 showed no significant difference before pneumoperitoneum but we found statistically significant difference after pneumoperitoneum at 8 ml kg 1 (P < 0.05). Median difference in leak volume (Fig. 3) showed significance both before and after pneumoperitoneum at 10 ml kg 1. The median differences in leak volume at 10 ml kg 1 tidal volume were 28.5 and 33.5 ml before and after pneumoperitoneum, respectively. Four cases were found to have leak volumes of more than 200 ml at the three set tidal volumes. Their leak fractions were more than 20%. At 6 ml kg 1 ventilation was optimum in all cases of both groups. Before pneumoperitoneum, ventilation was also optimum in all cases of both I and T groups at 8 ml kg 1. It was suboptimal in seven cases in I-group A, of which three cases ventilation failed and required tracheal intubation. The remaining four cases ventilation improved after increasing respiratory rate. At 10 ml kg 1 ventilation was sub-optimum in three and nine cases of I-group B and A, respectively. Five cases ventilation failed and required tracheal intubation of I- group A. Oxygenation was optimum in all cases. None of our patients was positive for gastric distension by auscultation over epigastric area. The i-gel was inserted within three attempts without failure. Only four patients of I-group needed minor manipulations after insertion. None of the cases needed more than one manipulation. On removal, no visible blood was noticed. Two cases had a minor trauma to the lip. After recovery, there was no hoarseness, spasm or regurgitation (Table 2). 3. Results Sixty patients were selected for the study; as regarding demographic data, there was no statistically significant difference between groups. The mean (SD) age, sex, weight, height and BMI of the participant are shown in Table 1. The median airway leak (sealing) pressure with i-gel was 25.75 (20 31.5) cm H 2 O. I-gel was inserted without failure in all cases within three attempts and four patients required manipulations. Before pneumoperitoneum there was no statistically significant difference in leak fraction (Fig. 2) between i-gel (I group B) and tracheal tube (T group B) at tidal volume 6 and 8 ml kg 1 (P = 0.59 and 0.60), respectively. The mean difference in leak fraction was significant between (I group A) and (T group A) after pneumoperitoneum at 8 ml kg 1 tidal volume (P = 0.03). At 10 ml kg 1 there was statistically significant Figure 2 Leak fractions using i-gel and tracheal tube, before (B) and after (A) pneumoperitoneum.

10 M. Ibrahim et al. Figure 3 Leak volumes using i-gel and tracheal tube, before (B) and after (A) pneumoperitoneum. Table 2 Comparison of other parameters. I group (i-gel) T group (ETT) Gastric insufflations 0 0 No. of attempts 1st 26 29 plus 4 2nd 2 1 3rd 2 0 Manipulations Jaw thrust (n) 1 0 Chin lift (n) 1 0 Increasing the depth of insertion 2 0 Blood staining 0 0 Oral trauma 0 0 Bronchospasm/laryngospasm 0 1 Hoarseness 0 0 Distension, regurgitation and aspiration 0 0 4. Discussion Supra glottic airway devices have several advantages including lower incidence of sore throat [6], less hemodynamic upset during induction and maintenance of anesthesia [7,8] and better oxygenation during emergence [9]. I-gel is a relatively new disposable supraglottic airway device that has no inflatable cuff. It has an integral bite block, wide bore lumen, and an additional distal lumen that allows for the passage of a gastric tube. These features may give the i-gel an advantage over the LMA and even the ProSeal LMA. In our study we found that i-gel airway can be used safely and effectively during volume controlled ventilation with low and moderate tidal volumes. There was too much debate between anesthesiologists about using these devices during procedures requiring positive pressure ventilation. During high tidal volume ventilation and laparoscopic procedures peak airway pressure rises and exceeds airway sealing (leak) pressure leading to increase in leak volume and fraction. These findings explain difficulties found to maintain optimum ventilation. Uppal et al. [5] found leak pressure for i-gel 28 (20 35) cm H 2 O by both auscultation and manometer stabilization methods. In our study we concluded that airway leak pressure for i-gel was 25.75 cm H 2 O. Ishwar et al. [10] concluded that airway leak pressure for i-gel was 25.27 cm H 2 O using same methods. Lopez-Gil et al. [11] compared four kinds of measurements of the airway sealing pressure, which involved detection of audible noise by listening over the mouth, detection of exhaled carbon dioxide by placing a gas sampling line for the capnograph inside the mouth, detection of a steady value airway pressure while occluding the expiratory valve of the circle system and detection of an audible noise using a stethoscope placed just lateral to the thyroid cartilage. They concluded that all four tests were excellent. We suspected from previous study, Uppal et al. [5], that there may be some cases with laryngeal anatomical variation that may lead to high leak volumes and leak fraction. This leads to unequal distribution of data. Therefore data were analyzed using non-parametric tests. The four cases with high leak volume with all tested tidal volumes, were included in T group. In our study we wanted to evaluate the i-gel during procedures with higher pressure like laparoscopic procedures. Lu et al. [12] compared ProSeal laryngeal mask airway (PLMA) with Classic laryngeal mask airway (LMA) for positive pressure ventilation during laparoscopic cholecystectomy. They concluded that PLMA is more effective ventilator device for laparoscopic cholecystectomy than classic LMA. This was attributed to higher leak pressure due to large cuff size (leak pressure was 29 ± 6 cm H 2 O). We thought that i-gel could be used during such procedures but unfortunately during our study we found leak pressure for i-gel was 25.75 cm H 2 O which is less than peak pressure during pneumoperitoneum especially at moderate and high tidal volumes [13]. Donaldson and Michalek [14] had inserted i-gel airway to control airway during laparoscopic cholecystectomy with volume controlled ventilation in patient with subglottic stenosis. They ventilated the patient with 500 ml tidal volume and a rate of 12 b min 1. The peak pressures were 14 and 18 cm Hg before and after pneumoperitoneum, respectively. We considered 10 ml kg 1 as a high tidal volume because more that 10 ml kg 1 are probably seldom used during anesthesia [15]. We did not correlate anatomical position of i-gel with clinically evident leaks by using fibreoptic bronchoscope [16]. No gastric distension, regurgitation, or aspiration while using the i-gel for VCV during our study. There were no cases of failed insertions. Visible blood was observed on the i-gel after removal, in our study, was 12% (4/30). This is similar to those reported with other SAD. The occurrence of visible blood with the use of other SAD has been reported from 12% to 18%, depending upon the type of SAD, technique of insertion, and ease of insertion [17,18]. Our study supports the use of i-gel during VCV in elective laparoscopic cholecystectomy using low to moderate tidal volumes provided that peak airway pressure not more than device leak pressure. Although leak volume was significant, ventilation and oxygenation were optimal in most cases. Tracheal tube should be inserted if failed ventilation and oxygenation. References [1] Devitt JH, Wenstone R, Noel AG, O Donnell MP. The laryngeal mask airway and positive-pressure ventilation. Anesthesiology 1994;80:550 5. [2] Levitan RM, Kinkle WC. Initial anatomic investigations of the i-gelä airway: a novel supraglottic airway without inflatable cuff. Anesthesia 2005;60:1022 6.

I-gel vs cuffed tracheal tube during volume 11 [3] Jackson KM, Cook TM. Evaluation of four airway training manikins as patient simulators for the insertion of eight types of supraglottic airway devices. Anaesthesia 2007;62:388 93. [4] Wharton NM, Gibbison B, Gabbott DA, Haslam GM, Muchatuta N, Cook TM. i-gel insertion by novices in manikins and patients. Anaesthesia 2008;63:991 5. [5] Uppal V, Fletcher G, Kinsella J. Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation. Br J Anaesth 2009;102(2):264 8, ISSN: 0007-0912. [6] Higgins PP, Chung F, Mezei G. Postoperative sore throat after ambulatory surgery. Br J Anaesth 2002;88:582 4. [7] Dyer RA, Llewellyn RL, James MF. Total i.v. anaesthesia with propofol and the laryngeal mask for orthopaedic surgery. Br J Anaesth 1995;74:123 8. [8] Cork RC, Depa RM, Standen JR. Prospective comparison of use of the laryngeal mask and endotracheal tube for ambulatory surgery. Anesth Analg 1994;79:719 27. [9] Webster AC, Morley-Forster PK, Dain S, et al. Anaesthesia for adenotonsillectomy: a comparison between tracheal intubation and the armoured laryngeal mask airway. Can J Anaesth 1993;40:1171 7. [10] Ishwar S, Monika G, Mansi T. Comparison of clinical performance of i-gelä with LMA ProSealä in elective surgeries. Indian J Anaesth 2009;53(3):302 5. [11] Lopez-Gil M, Brimacombe J, Keller C. A comparison of four methods for assessing oropharyngeal leak pressure with the laryngeal mask airway (LMA) in paediatric patients. Paediatr Anaesth 2001;11:319 21. [12] Lu PP, Brimacombe J, Yang C, Shyr M. ProSeal versus the Classic laryngeal mask airway for positive pressure ventilation during laparoscopic cholecystectomy. Br J Anaesth 2002;88(6): 824 7. [13] Brimacombe J, Keller C, Fullekrug B, et al. A multicenter study comparing the ProSeal and Classic laryngeal mask airway in anesthetized, nonparalyzed patients. Anesthesiology 2002;96: 289 95. [14] Donaldson W, Michalek P. The use of an i-gel supraglottic airway for the airway management of a patient with subglottic stenosis, a case report. Minerva Anesthesiol 2010;76. [15] Schultz MJ, Haitsma JJ, Slutsky AS, Gajic O. What tidal volumes should be used in patients without acute lung injury? Anesthesiology 2007;106:1226 31. [16] Van Zundert A, Brimacombe J, Kamphuis R, Haanschoten M. The anatomical position of three extraglottic airway devices in patients with clear airways. Anaesthesia 2006;61:891 5. [17] Parker MR, Day CJ. Visible and occult blood contamination of laryngeal mask airways and tracheal tubes used in adult anaesthesia. Anaesthesia 2000;55:388 90. [18] Tordoff SG, Scott S. Blood contamination of the laryngeal mask airways and laryngoscopes what do we tell our patients? Anaesthesia 2002;57:505 6.