Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Similar documents
Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Verification of Advanced Radiotherapy Techniques

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA

How effective is current clinical trial QA

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

iviewdose Confidence and assurance in dose delivery

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Leila E. A. Nichol Royal Surrey County Hospital

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

EORTC Member Facility Questionnaire

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

National Institute for Health and Clinical Excellence. Health Technology Appraisal. IIntensity modulated radiotherapy for prostate cancer

Credentialing for Clinical Trials -IGRT. Evidence Based Radiation Oncology. Levels of Clinical Evidence. Why Credentialing?

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

Elekta Synergy Digital accelerator for advanced IGRT

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

Advances in external beam radiotherapy

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

TomoTherapy. Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA. May 2016

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation. Research and development to advance radiotherapy

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Corporate Medical Policy

IMAT: intensity-modulated arc therapy

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

Elekta - a partner and world-leading supplier

IMRT vs Tomoterapia - Le nuove sfide

IMRT Planning Basics AAMD Student Webinar

Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Horizon Scanning Technology Briefing National Helical Tomotherapy Horizon Hi-ART System for Scanning external cancer Centre radiotherapy August 2006

Elekta Infinity Digital accelerator for advanced treatments

Sarcoma and Radiation Therapy. Gabrielle M Kane MB BCh EdD FRCPC Muir Professorship in Radiation Oncology University of Washington

Future upcoming technologies and what audit needs to address

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates

Elekta Infinity. Digital accelerator for advanced treatments. Redefining treatment precision, speed and control

Variable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10

Quality assurance in external radiotherapy

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

A national dosimetric audit of VMAT and Tomotherapy in the UK

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

Research Article An IMRT/VMAT Technique for Nonsmall Cell Lung Cancer

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

RPC s Credentialing Programs for Clinical Trials

Clinical experience with TomoDirect System Tangential Mode

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

SunCHECK Patient Comprehensive Patient QA

Overview of Advanced Techniques in Radiation Therapy

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

Radiotherapy Advances

Intensity Modulated RadioTherapy

I. Equipments for external beam radiotherapy

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

Effective and efficient radiotherapy dosimetry audit: Where to next?

Corporate Medical Policy

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Flattening Filter Free beam

Clinical Trial Credentialing:

Defining Target Volumes and Organs at Risk: a common language

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Address for Correspondence: Department of Medical Physics, Khwaja Yunus Ali University, Enayetpur, Sirajgonj ,

A review of RTTQA audit. Karen Venables

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Implementation of advanced RT Techniques

Cell survival following high dose rate flattening filter free (FFF) and conventional dose rate irradiation

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center

NEWER RADIATION (3 D -CRT, IMRT, IGRT) TECHNIQUES FOR CERVICAL CANCERS (COMMON PELVIC TUMORS)

SBRT Credentialing: Understanding the Process from Inquiry to Approval

Economic Implications of Advancements in Radiation Technology. Dramatic Growth of RT Capacity in the US

Treatment Planning for Lung. Kristi Hendrickson, PhD, DABR University of Washington Dept. of Radiation Oncology

Techniques and Technologies in Radiation Oncology 2018

Transcription:

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015

Table of Contents Abbreviations & Glossary... 3 Executive Summary... 4 1. Inversely planned treatments in TROG Clinical Trials... 5 1.1 Scope... 5 1.2 Introduction... 5 2. Credentialling of inversely planned treatments... 6 2.1 Facility questionnaire... 6 2.2 Trial-specific Benchmarking exercise... 6 2.3 Dosimetric Quality Assurance... 7 3. Information for trial investigators... 10 4. References... 11 Acknowledgements... 11 Version 2.1 April 2015 Page 2 of 11

Abbreviations & Glossary CQMS IMRT VMAT IGRT QA TMC 3D CRT CT MLC RANZCR ACDS TROG COO Central Quality Management System Intensity Modulated Radiation Therapy; here using static gantry Volumetric Modulated Arc Therapy Image Guided Radiation Therapy Quality Assurance Trial Management Committee Three-dimensional Conformal Radiation Therapy Computerised Tomography Multi Leaf Collimator Royal Australian and New Zealand College of Radiologists Australian Clinical Dosimetry Service TROG Central Operations Office Version 2.1 April 2015 Page 3 of 11

Executive Summary Inversely planned treatment techniques, including IMRT, VMAT and Tomotherapy, have become increasingly the standard of care since the first issue of TROG guidelines. This review aims to update requirements for trials participation to ensure robustness of delivered dose across participating centres, to support staff at participating centres, and to enhance trials accrual. The uses of inversely planned techniques in trials fall into three categories: 1. No RT question, or RT dose not critical 2. RT is part of the study question, but IMRT is not considered likely to impact on the study question. 3. Inversely planned treatments form all or part of the study question; or the study cannot be conducted without inversely planned treatments. All centres will be asked to provide: a. Facility questionnaire, in two parts: TROG QA has prepared a generic questionnaire, with results held on the website; a short, additional trial-specific IMRT questionnaire. b. Benchmarking exercise for the trial: for a provided CT dataset, preparation and submission of treatment plan using CQMS. c. Dosimetric quality assurance exercise, with requirements depending on the category of the trial question. TROG QA staff are happy to advise centres and discuss any practical issues. Please contact qa@trog.com.au. These guidelines have been prepared by TROG QA New Technologies and Techniques Committee and approved by TROG Scientific Committee. Version 2.0 replaces the guidelines issued in September 2008 and revised in October 2009. Version 2.1 April 2015 Page 4 of 11

1. Inversely planned treatments in TROG Clinical Trials 1.1 Scope These guidelines cover the following treatment delivery techniques: IMRT, ie static gantry IMRT, using either step and shoot MLC fields, or dynamic MLC fields. VMAT, ie dynamic gantry IMRT. Manufacturers trade names are Varian RapidArc [1]; Elekta VMAT [2]. TomoTherapy with HiArt planning system [3]. Forward planned field-in-field techniques are subject to 3D conformal treatment guidelines. IGRT requirements are trial-specific and are not part of this document. The TROG protocol template refers to these guidelines, which are specifically applicable to current trials introducing inversely planned techniques as a treatment option and to trials under development when QA procedures are being developed. 1.2 Introduction Static gantry IMRT has been shown to improve outcomes [4] and longer term toxicity [5, 6] in prostate treatments compared with previous techniques. The TROG project Assessment of New Radiation Oncology Techniques and Technologies (ANROTAT) concluded that IMRT has been shown to be both more effective and less costly than its comparator three- dimensional conformal radiation therapy (3DCRT) in three disease sites: naso-pharynx, post-prostatectomy and anal canal [7]. RANZCR have reported that Dynamic gantry inversely planned treatment (VMAT) provides significant evidence that volumetric modulated arc therapy, when performed with multiple arcs, achieves superior target dose homogeneity and adjacent normal tissue sparing in a range of tumour sites when compared to 6 field 3D conformal radiotherapy [8]. It is evident that these techniques are being generally adopted in routine clinical practice and becoming the standard of care. Version 2.1 April 2015 Page 5 of 11

The challenge for clinical trials groups is to incorporate these new techniques into currently open trials, and developing trials, while maintaining robustness and validity of delivered dose data. Clinical dose reporting follows ICRU recommendations [9]. US and UK trials groups have published guidelines with IMRT credentialling requirements [10, 11] and details of VMAT audits [12, 13]. The ANROTAT project included dosimetric site visits with phantoms, with protocols and tolerances developed by a committee of experienced medical physicists. This experience has enhanced the development and update of the TROG guidelines, which aim to ensure robustness of delivered dose across participating centres, to support staff at participating centres, and to enhance trials accrual. 2. Credentialling of inversely planned treatments 2.1 Facility questionnaire TROG QA have centralised the generic part of the facility questionnaire, with information on equipment and facilities being held on CQMS, and covering 3D conformal treatments. The facility questionnaire requires centres to submit details of their local patient-specific and machine-specific QA practices. Additional technical aspects will be covered by trial-specific short questionnaires, eg IMRT in TROG 08.08 TOPGEAR. Information will be held centrally, and with the centre s permission, may be made available for other trials. The TROG 08.08 TOPGEAR IMRT module asks for information on treatment planning system algorithms; treatment unit and MLC performance; IGRT techniques used; patient immobilisation; patient-specific QA procedures and dose verification; and participation in external IMRT audits. 2.2 Trial-specific Benchmarking exercise The trial QA process will include benchmarking exercises, to ensure the centre s ability to comply with the trial protocol and the QA review process. The benchmarking exercise will include a contouring and planning exercise with external review by independent reviewers (appointed by the TMC) using TROG plan review software including MIM [14] or SWAN [15] Contouring of targets and normal tissue is critical for all complex volumes/plans required in inverse planning, and especially for uncommon sites/primaries, and can be the weakest link in the treatment delivery chain. The external review of the contours is fundamental to the validity of the treatment plan. Version 2.1 April 2015 Page 6 of 11

2.3 Dosimetric Quality Assurance 2.3.1. All centres will be asked to provide evidence, by report, of an external level 1 audit by a recognised body, such as Australian Clinical Dosimetry Service (ACDS), within the last 2 years. Reports from other bodies may be submitted to TROG QA office for consideration. 2.3.2. Clinical trial data robustness relies on the equivalence of delivered dose in each participating centre. Traditionally this has been validated by a level 3 dosimetric audit site visit by a trial representative, using the same phantom and equipment at each centre. It has been the responsibility of each trial to cover the QA procedures as required. In some instances, it has been possible for other trials to accept credentialling reports. TROG QA regards the validity of the level 3 audit report ideally as up to 5 years, and will discuss any issues with individual centres. 2.3.3. Due to limited resources, it is not always possible to make the credentialling site visits. TROG is working on alternative approaches, which would enable efficient timescales for the credentialling process. 2.3.4. Some centres are satellite centres to a principal centre, or part of a network, and may be considered in conjunction with the principal centre on a case-by-case basis, provided that: Equipment is the same: manufacturer, matched linacs, treatment planning system and beam models, oncology information management system Common QA protocols across centres Staff rotate across centres. Each geographical centre is required to provide a level 1 audit report as in 2.3.1. TROG QA will discuss the position with centres. 2.3.5. International centres wishing to take part in TROG trials are invited to submit evidence of dosimetry audits, from recognised bodies, to TROG QA for consideration. 2.3.6. All centres using TomoTherapy shall have an external dosimetry audit to assess the delivered dose and ensure consistency of delivered dose for comparison with linear accelerator Version 2.1 April 2015 Page 7 of 11

treatment deliveries. TomoTherapy treatments have been introduced in Australia since 2011, and as such are a new technique which TROG must validate. 2.3.7. VMAT, ie dynamic gantry modulated delivery with inverse planning, is regarded as a new technique and each centre wishing to use VMAT in TROG trials must have an external dosimetry audit. There are differences in planning and delivery techniques between the manufacturers, eg in gantry and dose modulation prioritising. The external audit is therefore only valid for the specified equipment. 2.3.8. Other new treatment delivery techniques, including Cyberknife, will require the centre to have an external dosimetry audit contact TROG QA for advice. 2.3.9. Each trial will set the tolerance criteria for acceptance, for both point dose accuracy and gamma analysis of measured dose planes [16, 17]. 2.3.10. The uses of inversely planned techniques in trials fall into three categories with increasing levels of QA requirements: A. No RT question, or RT dose not critical, or <20% of prescription dose delivered using inversely planned techniques TROG minimum requirements: Facility questionnaire detailing IMRT delivery and QA procedures, fulfilling tolerance criteria specified by the Trial Management Committee and TROG QA. B. RT is standard in both arms TROG minimum requirements: Facility questionnaire detailing IMRT delivery and QA procedures. Evidence of relevant and appropriate IMRT level 3 dosimetric audit by recognised body OR evidence of activities and data that provides a similar level of confidence in the delivery process as a level 3 audit. This could be a remote process. Details to be discussed with the TMC and TROG QA. Version 2.1 April 2015 Page 8 of 11

Centres using VMAT, TomoTherapy or other new techniques will require an external dosimetry audit (see 2.3.5 to 2.3.7 above). C. RT is the research question, specifically the technology/technique being tested TROG minimum requirements: Evidence of relevant and appropriate IMRT level 3 dosimetric audit by recognised body. All patients treated with IMRT, VMAT or Tomo must have individual plan QA. Dosimetric audit site visit by TMC/TROG may be required, as specified by the TMC physicist and radiation therapist. Centres using VMAT, TomoTherapy or other new techniques will require an external dosimetry audit (see 2.3.5 to 2.3.7 above). Version 2.1 April 2015 Page 9 of 11

3. Information for trial investigators TROG QA are anxious to support trial investigators, for current trials introducing inversely planned techniques, and for trials under development when QA procedures are being developed. This checklist is intended to support investigators. Area Comments TROG protocol template The protocol template covers all areas to be addressed, and is available from TROG QA Multi-professional trial QA team Each team member should have good clinical experience using the inversely planned technique. The team may be drawn from different centres and cover all professional disciplines. QA programme [ to assess compliance with trial protocol] Contact TROG QA office for advice and information on requirements and documentation specific to the trial, including facility questionnaire and benchmarking case. Discuss QA requirements including dose-volume constraints and definitions of minor/major violations. Centre credentialling requirements Contact TROG QA office to discuss the category of credentialling required, and any previous audits which may be relevant for this trial Notification to centres: current trials only Inform centres of specific requirements, eg external dosimetry audit for new technique Additional requirements Specify any requirements, eg IGRT Version 2.1 April 2015 Page 10 of 11

4. References 1. RapidArc Radiotherapy Technology one revolution is all it takes. 2013 26 Nov 2013]; Available from: http://www.varian.com. 2. Comprehensive image-guided radiation therapy system with VMAT. 2013 26 Nov 2013]; Available from: http://www.elekta.com/. 3. The TomoTherapy Hi Art treatment system uses a patented multileaf collimator to modulate the intensity of the radiation beam 2013 26 Nov 2013]; Available from: http://www.tomotherapy.com/. 4. Zelefsky, M., et al., Clinical experience with intensity modulated radiation therapy (IMRT) in prostate cancer. Radiother Oncol, 2000. 55: p. 241-249. 5. Alicikus, Z., et al., Ten-year outcomes of high-dose, intensity-modulated radiotherapy for localized prostate cancer. Cancer, 2011. 117: p. 1429 1437. 6. Michalski, J., et al., Preliminary toxicity analysis of 3-dimensional conformal radiation therapy versus intensity modulated radiation therapy on the high-dose arm of the Radiation Therapy Oncology Group 0126 prostate cancer trial. Int J Radiat Oncol Biol Phys, 2013. 87: p. 932-938. 7. TROG, The Assessment of New Radiation Oncology Technology and Treatments (ANROTAT) Project: FINAL REPORT. 2012. 8. RANZCR, Position paper: The evidence base for multiple volumetric modulated arc therapy (VMAT)- a quality perspective. 2011. 9. ICRU, Report 83: Prescribing, recording and reporting Photon beam Intensity Modulated Radiation Therapy (IMRT), I.C.o.R.U.a. Measurements, Editor. 2010. 10. ATC. Credentialing for Treatment Techniques of Dose Painting (including IMRT). 2009 27 Nov 2013]; Available from: http://atc.wustl.edu/credentialing/atc_reports/imrt_credentialing_09jan09.pdf. 11. RTTQA. RTTQA IMRT credentialing program. 27 Nov 2013]; Available from: http://www.rttrialsqa.org.uk/. 12. Hussein, M., et al., A methodology for dosimetry audit of rotational radiotherapy using a commercial detector array. Radiotherapy and Oncology, 2013. 108(1): p. 78-85. 13. Tsang, Y., et al., Development of a novel treatment planning test for credentialing rotational intensitymodulated radiotherapy techniques in the UK. The British journal of radiology, 2013. 86(1022): p. 20120315. 14. Comprehensive Solutions for Radiation Oncology. 2013 26 Nov 2013]; Available from: http://www.mimsoftware.com/. 15. Ebert, M., et al., Detailed review and analysis of complex radiotherapy clinical trial planning data: Evaluation and initial experience with the SWAN software system. Radioth Oncol 2008. 86: p. 200-210. 16. Low, D., et al., A technique for the quantitative evaluation of dose distributions. Med Phys 1998. 25: p. 656-661. 17. Hussein, M., et al., A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems. Radiotherapy and Oncology, 2013. Acknowledgements TROG appreciates the support of the members of the IMRT Working party who contributed to version 1; the members of TROG New Technologies and Techniques Committee; the TROG Scientific Committee and the Clinical Liaison leader. Special thanks to Tomas Kron, Peter Greer, Jarad Martin, Michael Bailey, Rob McDowell, David Willis, Martin Ebert, Melissa Crain and Joan Hatton. Version 2.1 April 2015 Page 11 of 11