Factors affecting the efficiency of aerosol therapy with pressurised metered-dose inhalers through plastic spacers

Similar documents
A profile of journals of complementary and alternative medicine 1

ADDITIONAL TECHNOLOGIES FOR PRESSURIZED METERED DOSE INHALERS. Steve Newman Scientific Consultant Nottingham, UK

An update on inhalation devices

Perceptual reasons for resistance to change in the emergency department use of holding chambers for children with asthma

Pharmacokinetics, pharmacodynamics, and the delivery of pediatric bronchodilator therapy

Quality of life of Do-Not-Resuscitate (DNR) patients: how good are physicians in assessing DNR patients quality of life?

Gemcitabine-related pulmonary toxicity

University of Groningen. Technology in practice Lexmond, Anne

The effects of static charge in spacer devices on glucocorticosteroid aerosol deposition in asthmatic patients

Delay Between Actuation and Shaking of a Hydrofluoroalkane Fluticasone Pressurized Metered-Dose Inhaler

Anxiety in health care workers after exposure to potentially HIV-contaminated blood or body fluids

Inhalational drug delivery from seven different spacer devices

Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi)

Hydrofluoroalkane Pressurized

Tracheobronchial foreign body aspiration in children diagnostic value of symptoms and signs

AEROSOL THERAPY: THE PRACTICALITIES

Promoting hospital-based smoking cessation services at major Swiss hospitals: a before and after study

Prevalence rate and reasons for refusals of influenza vaccine in the elderly

Paediatric pulmonary drug delivery: considerations in asthma treatment

Evaluation of a population-based. prevention program against Influenza, among Swiss elderly people

Comparison of different methods for the measurement of serum testosterone in the aging male

Response to Albuterol MDI Delivered Through an Anti-Static Chamber During Nocturnal Bronchospasm

Special Problems in Aerosol Delivery: Neonatal and Pediatric Considerations

Clinical recognition and treatment of atrial ectopic tachycardia in newborns

facemask Drug delivery from holding chambers with attached Mark L Everard, Andrew R Clark, Anthony D Milner kilogram from each chamber was relatively

Valved holding chamber drug delivery is dependent on breathing pattern and device design

L. Borgström*, E. Bondesson*, F. Morén**, E. Trofast*, S.P. Newman +

The majority of asthmatic patients continue to use pressurized

THE NEW ZEALAND MEDICAL JOURNAL

Neuromotor development from kindergarten age to adolescence: developmental course and variability

Advanced Inhaler Technique. Learning Outcomes. Disclosure 1.1. Belgian Chocolate, French Champagne and Inhaled Medication: Too Good To Waste?

The inhaled route is preferred for the

Metered Dose Inhalers with Valved Holding Chamber: A Pediatric Hospital Experience

Pitfalls in the emergency department triage of frail elderly patients without specific complaints. RUTSCHMANN, Olivier Thierry, et al.

Patricia KP Burnell Inhalation Product Development

Combination of hydrocolloid dressing and medical compression stocking versus Unna s boot for the treatment of venous leg ulcers

CLINICAL RELEVANCE OF IN VITRO PARTICLE SIZING DATA. Steve Newman, PhD Nottingham, UK November 2006

Safety and efficiency of the Ottawa ankle rule in a Swiss population with ankle sprains

Appropriateness of serum level determinations of antiepileptic drugs

Metered Dose Inhaler Technology

Higher lung deposition with Respimat Soft Mist Inhaler than HFA-MDI in COPD patients with poor technique

Results of conservative treatment for perforated gastroduodenal ulcer in patients not eligible for surgical repair

News on lung volume reduction surgery

Aerosol Delivery Devices

Population-based epidemiology of rotavirus hospitalisations in Switzerland

Sleep and wakefulness disturbances in Swiss pharmacy customers

Coronary artery disease screening in diabetic patients: how good is guideline adherence?

nebulised budesonide in preschool children

Correspondence: C-G. Löfdahl Dept of Respiratory Medicine and Allergology University Hospital S Lund Sweden.

Mechanisms of alveolar epithelial repair in acute lung injury a translational approach

Paediatric Infectious Diseases Ward, Boali-Cina Hospital, Mazandaran University of Medical Sciences, Pasdaran Boulevard, Sari-Iran

B-type natriuretic peptide (BNP): can it improve our management of patients with congestive heart failure?

Inhaled corticosteroid delivery systems: clinical role of a breath-actuated device

Pulmonary deposition of inhaled drugs

Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics

Patient. Device Clinician. Safety & efficacy

Intussusception as a cause of bowel obstruction in adults

I. Subject: Medication Delivery by Metered Dose Inhaler (MDI)

Assessing Quality of Inhaled Products And Links to Efficacy and Safety

Use of Respimat Soft Mist Inhaler in COPD patients

2008 IPAC-RS Conference Doing the Right Thing Science, Quality and Patient Focus

Diagnostic value of lung auscultation in an emergency room setting

Inh l a t a i tional D i ev ces for the Ou O t u pat p ie i nts Dr Sunil Sharma Senior Resident

Your Inhaler Devices & You

Appendix M: Device Technique

Inhalation Therapy. Inhalation Therapy

Affective distress and fibromyalgia

What you need to know about inhalers and how to use them Henry Chrystyn PhD, FRPharmS and David Price MA, MRCGP, DRCOG

Ventamol CFC-Free 100 micrograms, Pressurised Inhalation, Suspension (Salbutamol)

Hydrofluoroalkane- vs Chlorofluorocarbon-Beclomethasone Delivery in Neonatal Models. Tufts University School of Medicine, Boston, MA, and

Scintigraphic comparison of budesonide deposition from two dry powder inhalers

What You Need to Know about Metered-Dose Inhalers and the HFA Propellant

Delivering Aerosol Medication in ICU

The ideal inhaler: design and characteristics to improve outcomes

Nebulizer versus inhaler with spacer for beta-agonist treatment in acute bronchospastic disease

Respiratory Care in PICU Aerosol Therapy ส พ ชชา ชา แสงโขต โรงพยาบาลสมเด จพระป นเกล า

Metered Dose Inhalers (MDIs) In Vitro Measures to Confirm Patient Perceptions: HFA vs. CFC

Are internists in a non-prescriptive setting favourable to guidelines?

Diagnostic yield of flexible bronchoscopy in current clinical practice

Beclometasone dipropionate (BDP) Prophylactic management of mild, moderate or severe asthma in adults

21/03/2011 AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY. Fundamentals of aerosols

PRODUCT INFORMATION INTAL CFC-FREE INHALER AND INTAL FORTE CFC-FREE INHALER

Silent coronary artery disease in patients with diabetes mellitus

Diagnostic accuracy of exercise electrocardiogram in patients with left anterior hemiblock

NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS

Smart asthma therapy. Patient information Spacer inhaling aid for metered-dose aerosol inhalers

This leaflet can be made available in other formats including large print, CD and Braille and in languages other than English, upon request.

following inhalation using metered dose

Optimizing Aerosol Delivery by Pressurized Metered-Dose Inhalers

COMPARISON OF THE RESPIRABLE FRACTION FROM THREE DIFERENT DPI DEVICES

Asthma and the competitive swimmer

Concurrent Use Of Metered Dose Inhalers Without Spacer And Dry Powder Inhalers By Asthmatic Children Adversely Affect Proper Inhalation Technique

A multitude of devices

Information for Patients. Inhaler devices for routine treatment of chronic asthma in older children (aged 5-15 years)

Optimisation of inhaled corticosteroid delivery to asthmatic children: integration of in vitro and in vivo methods

The Novolizer s : overcoming inherent problems of dry powder inhalers

Archives of Pulmonology and Respiratory Care

COPD Education for Clinic RNs and MA s 6/28/18

VENTOLIN EVOHALER. Salbutamol

Inhalation devices, proper technique and cleaning

Transcription:

Original article Peer reviewed article SWISS MED WKLY 2001;131:14 18 www.smw.ch 14 Factors affecting the efficiency of aerosol therapy with pressurised metered-dose inhalers through plastic spacers Anouk A Chuffart a, Felix H Sennhauser a, Johannes H Wildhaber a, b Swiss Paediatric Respiratory Physiology Research Group a Department of Respiratory Medicine, University Children s Hospital, Zürich, Switzerland b Department of Paediatrics, Princess Margaret Hospital for Children, Subiaco, Australia Summary Aim: The main objective of this study was to compare the in vitro delivery of salbutamol from a chlorofluorocarbon(cfc)-propelled pressurised metered-dose inhaler (pmdi) versus a newly developed hydrofluoroalkane(hfa)-propelled pmdi through various spacers. In addition, we aimed to study the effect on bronchodilator response when using an optimal pmdi/spacer combination for aerosol delivery compared to a suboptimal combination. Methods: Particle size distribution and output from salbutamol pmdis containing either CFC propellants (Ventolin ) or HFA propellants (Airomir ) were measured using a multistage liquid impinger (MSLI) and compared to that through both detergent-coated (non-static) or untreated (static) large volume (Nebuhaler, Volumatic ) and small volume (Aerochamber ) plastic spacers. Flow-volume curves (FEV 1 ) were obtained from twelve asthmatic children with known significant bronchodilator response (8 males), aged 13 17 years, randomly inhaling salbutamol from a CFC-pMDI through a static spacer (Nebuhaler ) and from an HFA-pMDI through a non-static spacer (Nebuhaler ). Results: In vitro output of particles in the respirable range (<6.8 µm) from HFA-pMDIs was significantly higher than that from CFC-pMDIs using various spacers. Removal of electrostatic charge increased output from CFC- and HFApMDIs through all spacers by 17 82%. The mean (SD) bronchodilator response after inhalation of salbutamol from a CFC-pMDI through a static spacer was 7.1% (6.3%) compared to 17.5% (7.9%) after inhalation from an HFA-pMDI through a non-static spacer (p = 0.002). Conclusions: Use of a newly developed HFApropelled pmdi greatly improves drug delivery through spacers compared to a CFC-propelled pmdi. However, electrostatic charge in plastic spacers remains the key determinant limiting delivery of salbutamol from a pmdi through spacers, and can be reduced by soaking the spacer in a household detergent. Using an optimal pmdi/spacer combination leads to a significantly improved bronchodilator response. Keywords: spacers; holding chambers; electrostatic charge; salbutamol; HFA; CFC; aerosol therapy Introduction Inhalation therapy plays a major role in the management of asthma [1]. The choice of an optimal inhalation device is essential in achieving effective, predictable and consistent dose delivery to the airways of asthma patients. Pressurised metered-dose inhalers (pmdis) are the most widely used form of inhalation therapy for asthmatics. However, there are a number of problems associated with their use. After actuation, the aerosol cloud leaves the canister at very high speed, and thus leads to high aerosol deposition in the oropharynx when actuated directly into the mouth. To overcome these problems, holding chambers (spacers) are widely used [2]. The main function of a spacer device is to act as a chamber reservoir where the actuated aerosol cloud can be held prior to inhalation by the patient. The use of spacers is specifically recommended in children, in order to prevent difficulties in coordinating actuation and inhalation [3].

SWISS MED WKLY 2001;131:14 18 www.smw.ch 15 However, it has been shown that the actuated dose from a pmdi through a spacer reaching the lungs of children is still relatively low, suggesting that additional factors influence drug delivery. A radio-labelled deposition study in children has shown that only 5% at most of the total actuated dose from a pmdi through a plastic spacer actually reaches the lungs [4]. Several in vitro and in vivo studies have shown that various factors influence the dose delivered from a holding chamber and may be responsible for this poor efficiency [5 14]. These include chamber size, shape, resistance of the valve, dead volume, the use of multiple actuations, inhalation delay, and construction materials which affect the levels of electrostatic charge in the chamber. Electrostatic charge is inherent to all plastic devices, including plastic spacers, owing to their non-conducting properties. The charge varies in a random manner. The net effect of electrostatic charge is absorption of aerosol particles onto plastic surfaces of the spacers, leading to a significant reduction in the initial dose available for inhalation and hence the lung dose. The most widely marketed spacer devices are manufactured from plastic. Extensive in vitro and in vivo work has been carried out to determine the effect of static charge on drug delivery from pmdis through plastic spacers, and it has been found that the output from these spacers can be increased by reducing static [5, 15]. Reducing electrostatic charge, and hence increasing the ratio between dose delivered to the patient and dose delivered to the spacer, not only increases drug delivery but also reduces variability in delivery [15]. This has major implications, especially for children, in whom drug delivery is already highly variable due to age-specific breathing patterns such as low tidal volumes and low inspiratory flows. It has been shown that drug delivery to children is much higher and less dependent on breathing patterns when using non-electrostatic spacers [12, 16]. Much of the work on the effect of detergent coating of spacers has been carried out using chlorofluorocarbon(cfc)-propelled pmdis. Because of the harmful effect of CFC propellants on the ozone layer, these pmdis are being phased out and replaced by pmdis containing hydrofluoroalkane (HFA) propellants, which have different aerosol characteristics [17 22]. The main objective of this study was therefore to compare in vitro delivery of salbutamol from CFC-pMDIs versus the newly developed HFApMDIs through various spacers and under different conditions. In addition, based on our in vitro findings, the clinical impact of using an optimal pmdi/spacer combination on bronchodilator response was analysed. Material and methods In vitro The particle size distribution and output from salbutamol pmdis (nominal dose: 100 µ g/actuation), containing either CFC propellants (Ventolin, Glaxo Wellcome, UK) or HFA propellants (Airomir, 3M Pharmaceuticals, UK) were measured and compared with those through both large volume (750 ml; Nebuhaler, Astra-Zeneca, UK and Volumatic, Glaxo-Wellcome, UK) and small volume (165 ml; Aerochamber, Trudell, Canada) plastic spacers. The effect of electrostatic charge was tested by comparing new spacers carrying a static charge (3.3 6.7 µc/m 2 ) with spacers coated with detergent to remove static (0 1.2 µc/m 2 ). The impact of the valves on particle size distribution and output from non-static CFC-pMDIs was also analysed by removing the valves. To coat spacers with detergent, they were soaked for 20 30 minutes in a commercially available ionic detergent (Palmolive, UK) and allowed to drip dry. Particle size distribution and total drug delivery was measured using a multistage liquid impinger (MSLI; Copley, Nottingham, UK) with an inhalation flow of 60 l/min. The pmdi was attached to the spacer and the spacer was inserted into the MSLI throat. In each case, a complete seal was ensured by using an appropriate adaptor at the throat. The pmdi was shaken vigorously for 30 seconds prior to actuation, and the first two actuations were wasted. In total, ten single actuations were introduced into the MSLI. The pmdi was shaken vigorously for five seconds between each actuation. The aerosol generated by the pmdi was drawn immediately through the MSLI with the entraining airflow. Droplets were deposited on the actuator, spacer, throat and stages 1 to 4. The location of particle deposition was determined by the aerodynamic size of the particle. The sizes of particles depositing on stages 1, 2, 3 and 4 were >13 µ m, 6.8 13 µ m, 3.1-6.8 µ m and <3.1 µ m. The actuator, spacer, throat and stages of the MSLI were washed with approximately 40 ml of methanol. Five ml of 0.1M NaOH was added to each wash, and the volume was then made up to a total of 50 ml with methanol. The absorbance (246 nm) of each sample was measured in duplicate on a spectrophotometer. The concentration of salbutamol in the samples was obtained by using the absorbance of a solution containing a known concentration of salbutamol. The standard curve for salbutamol was linear (r 2 = 1.00) for concentrations between 0 and 27 µ g/ml. Each experiment was repeated four times and the temperature, relative humidity and barometric pressure were recorded. In vivo Twelve asthmatic children (8 males) under regular inhaled steroid therapy, with known airflow obstruction (mean [SD] % predicted FEV 1: 68% [9%]) and known significant bronchodilator response, aged 13 17 years, were studied in a randomised, double-blind, placebo-controlled, crossover trial twice. The children inhaled on two visits (one week apart) once 200 µg salbutamol from a CFC-pMDI (Ventolin, Glaxo-Wellcome, UK) through a new, static spacer (Nebuhaler, Astra Zeneca, UK) and 200 µg salbutamol from an HFA-pMDI through a detergent coated, non-static spacer (Nebuhaler, Astra Zeneca, UK). The Nebuhaler was chosen for its superior in vitro performance in delivering salbutamol. Flow-volume curves were obtained before and five minutes after in-

Factors affecting the efficiency of aerosol therapy with pressurised metered-dose inhalers through plastic spacers 16 halation, and FEV 1 measurements were compared. Subjects were eligible if they had bronchodilator responsive airflow limitation, defined as an improvement in FEV 1 from baseline of at least 10%. The study was approved by the Hospital Ethics Committee and informed consent was obtained from the parents. Statistical analysis In vitro Results were calculated as mean (SD) % of the total dose. Statistical analysis (StatView 512+; Abacus Concepts Inc., Berkely, CA, USA) was carried out using analysis of variance (ANOVA) for unmatched data. Post-hoc analysis was performed using the Fisher protected least significant difference (Fisher PLSD), with a significance level of 95% (p <0.05), unless otherwise stated. In vivo The mean and standard deviation (SD) are reported for baseline spirometry and following inhalation of the bronchodilator. Paired t-test was used to determine the differences between responses with each spacer. Significance was accepted at the 0.05 level. Results In vitro Output of particles <6.8 µm (stage 3 and 4; respirable range) from HFA-pMDIs was significantly higher than from CFC-pMDIs, using either static spacers or non-static spacers (p <0.0001) (table 1). Removal of static increased output from both CFC- and HFA-pMDI through all spacers by 17 82% (p <0.0001) (table 1). Drug delivery from CFC-pMDIs through all spacers with static was similar and not significantly different (table 1). However, drug delivery from HFA-pMDIs through the static Nebuhaler was higher compared to the static Volumatic and the static Aerochamber (p <0.0001). Among the non-static spacers, the Nebuhaler delivered a significantly greater amount of particles smaller than 6.8 µ m from CFC-pMDIs (p = 0.0001) and HFA-pMDIs (p <0.0001) as compared to the Volumatic and the Aerochamber. There was no significant difference in delivery between the Volumatic and the Aerochamber (CFC-pMDI; p = 0.07; HFApMDI; p = 0.1). With the valves removed, all spacers had a greater output than with the valve in place (table 1). There was a smaller though significant increase in output from the non-static Nebuhaler (p = 0.0008) and Aerochamber (p = 0.002) without the valves, as compared to the remarkable increase in output after removal of the valves from the Volumatic (p <0.0001). Output from both large volume spacers with removed valves was significantly higher than from the small volume spacer (p <0.0001) (table 1). In vivo The mean (SD) FEV 1 before and after inhalation of salbutamol from a CFC-pMDI through a static spacer was 2.78 L (0.53) and 2.98 L (0.55) respectively, with a mean (SD) bronchodilator response of 7.1% (6.3%). The mean (SD) FEV 1 before and after inhalation of salbutamol from an HFA-pMDI through a non-static spacer was 2.67 L (0.42) and 3.13L (0.51) respectively, with a mean (SD) bronchodilator response of 17.5% (7.9%) (fig. 1). Table 1 Influence of various factors, such as the propellant used, electrostatic charge on the surface and design of the spacer, on the mean percentage (SD) drug output* of salbutamol through large (Nebuhaler and Volumatic ) and small (Aerochamber ) volume spacers. * Cutoffs of particles >13, 6.8 to 13, 3.1 to 6.8, and <3.1µm for stage 1, 2, 3, and 4 respectively. Stages 3 and 4 contain particles <6.8µm, which are in the respirable range. Nebuhaler Volumatic Aerochamber CFC-pMDIs Static actuator 5.1 (1.0) 7.1 (0.9) 6.6 (1.6) spacer 57.0 (1.3) 55.0 (2.1) 54.2 (2.7) throat 0.9 (0.4) 1.2 (0.5) 1.5 (0.7) stages 1 and 2 3.8 (0.8) 4.1 (1.0) 2.6 (0.9) stages 3 and 4 33.2 (1.9) 32.6 (1.8) 35.1 (3.5) Non-static actuator 6.8 (1.8) 8.2 (1.8) 7.1 (1.6) spacer 23.6 (4.2) 38.4 (2.0) 41.5 (1.7) throat 1.1 (0.5) 0.9 (0.4) 1.0 (0.3) stages 1 and 2 8.2 (1.1) 8.1 (1.4) 7.9 (1.4) stages 3 and 4 60.3 (2.9) 44.4 (2.5) 42.5 (2.0) Valve removed actuator 8.4 (1.3) 6.6 (1.4) 9.2 (1.6) spacer 17.9 (2.3) 21.1 (2.2) 37.5 (0.7) throat 0.7 (0.5) 1.4 (0.4) 0.9 (0.3) stages 1 and 2 8.2 (1.6) 7.2 (0.7) 7.5 (1.7) stages 3 and 4 64.8 (2.2) 63.7 (4.2) 44.9 (0.6) HFA-pMDIs Static actuator 4.1 (1.2) 5.2 (1.2) 4.1 (0.6) spacer 36.1 (1.6) 44.1 (3.0) 42.0 (2.5) throat 0.2 (0.2) 1.4 (0.8) 1.2 (0.3) stages 1 and 2 1.0 (0.5) 2.1 (0.4) 2.7 (1.2) stages 3 and 4 58.6 (2.3) 47.2 (2.6) 50 (2.8) Non-static actuator 4.1 (0.5) 5.2 (1.6) 4.0 (1.0) spacer 19.0 (2.6) 30.8 (3.7) 32.5 (1.5) throat 2.0 (1.0) 1.4 (1.1) 1.2 (1.3) stages 1 and 2 5.9 (1.2) 6.2 (1.6) 4.0 (2.2) stages 3 and 4 69.0 (1.2) 56.4 (2.5) 58.3 (2.4)

SWISS MED WKLY 2001;131:14 18 www.smw.ch 17 FEV1 (L) 3.2 3 2.8 2.6 (SD 0.53) (SD 0.42) (SD 0.51) (SD 0.55) static/cfc-pmdi non-static/hfa-pmdi Figure 1 Flow-volume curves before and five minutes after inhalation of 200 µ g salbutamol from a CFC-pMDI through a new, static spacer (Nebuhaler ) and from an HFA-pMDI through a detergent-coated, non-static spacer (Nebuhaler ). 2.4 before after Discussion The results of this study indicate that the delivery of salbutamol from newly developed HFApropelled pmdis through spacers is significantly improved compared to CFC-propelled pmdis. However, electrostatic charge on the surface remains the key determinant limiting delivery of salbutamol from either CFC- or HFA-pMDIs through both large and small volume plastic spacers. In addition, we have demonstrated a significant improvement in bronchodilator response when salbutamol was delivered from an HFApMDI through a non-static spacer, compared to a CFC-pMDI through a static spacer. These findings highlight the importance of using a non-static spacer for inhalation therapy with pmdis, as well as the use of newly developed HFA-propelled pmdis. When CFC-pMDIs using various static spacers were compared, the output was uniformly low despite varying volume and design. This finding suggests that the buildup of sufficiently high levels of static in plastic spacers overrules the differences in output from CFC-pMDIs through spacers, owing to differing volume and design. However, when HFA-pMDIs using various static spacers were compared, the drug delivery through the static Nebuhaler was higher compared to the static Volumatic and the static Aerochamber, indicating that the design of the spacer plays a role when using HFA-pMDIs. When static is reduced in different large and small volume spacers, there is still a difference in aerosol delivery and hence performance with both CFC- and HFA-pMDIs, due to differences in volume and design. Without static, the output from the Nebuhaler was greater than from another large volume spacer, the Volumatic, probably owing to resistance of the valve. This observation shows the importance of the valve in drug delivery from spacers and confirms the results of other studies [23]. When the valve from the Volumatic was removed the output increased by 43%, indicating that this valve does indeed significantly impede drug delivery [23]. Only a small increase of 5.6 7.5% was observed when the Nebuhaler or Aerochamber valve was removed. Electrostatic charge, volume and design of the spacer affect dose delivery differently depending on the pmdi used. Different pmdis have different vapour pressures, and therefore different aerosol cloud velocities and volumes. Output of particles in the respirable range from CFC-pMDIs was significantly lower than from HFA-pMDIs, using either static or non-static spacers. In addition, static appears to have a much greater effect on the output from CFC-salbutamol pmdis than from HFA-salbutamol pmdis. These findings are due to differences in the way the aerosol cloud is emitted from the actuator. The aerosol cloud from an HFA-pMDI is emitted at a slower speed and occupies a smaller volume than the conventional CFC formulation, leading to a much lower impaction on the spacer surface. Several studies have shown these additional benefits of HFA-pMDIs as delivery is increased not only in vitro but also in vivo [24 26]. Conducting materials carry no electrostatic charge, so a metal spacer solves the problem of reduced drug delivery due to electrostatic charge [27]. A previous study has shown that a metal spacer is superior to plastic spacers, even if the electrostatic charge is reduced on the surface of a plastic spacer [27]. But in view of the present results and the issues of cost and availability, a simple plastic spacer may remain the device of choice worldwide. Washing in detergent does build up a conducting layer on the surface of the plastic, thus reducing static and hence the attraction of the aerosol particles to the spacer surface. Previous studies have already shown that reducing electrostatic charge on plastic spacers by coating them with an ionic detergent significantly improves in vitro and in vivo drug delivery [5, 15]. It has been shown that all commercial household detergents offer a simple, practical and cheap way of avoiding static. In contrast, priming a plastic spacer with multiple actuations has been shown not to be effective in reducing static [15]. The present study has demonstrated that the major improvement in delivery efficiency of inhalation devices brought about by a simple method for static reduction, such as detergent coating, has considerable impact on

Factors affecting the efficiency of aerosol therapy with pressurised metered-dose inhalers through plastic spacers 18 bronchodilator response. In addition, this improvement is likely to have a notable economic impact. Improving delivery efficiency may also greatly increase efficacy and reduce treatment costs. However, important questions regarding the need to change current aerosol dosing practices have been raised in connection with previous deposition studies using spacers with static. Increased lung deposition by improved inhalation devices needs careful monitoring for potential systemic side effects, especially when using inhaled steroids. Framing of new guidelines will need to take these aspects into account. In practice, plastic spacers should be soaked in a household detergent once a week, and subsequently drip dried. Alternatively, a metal spacer could be used. In addition, spacers should be used in conjunction with HFA-pMDIs. In summary, this study offers new insights into the influence of the propellant used on the respirable range of drug output and its relation to spacer design. These findings highlight the importance of standardized inhalation methods for efficient and efficacious therapy. In addition, our results underline the difficulty of comparing drug deposition studies using different drugs, propellants, spacers and valve systems. Correspondence: Johannes H Wildhaber, MD Department of Respiratory Medicine University Children s Hospital Steinwiesstrasse 75 CH-8032 Zürich E-mail johannes.wildhaber@kispi.unizh.ch References 1 Matthys H. Inhalation delivery of asthma drugs. Lung 1990;168:645-52. 2 Konig P. Spacer devices used with metered-dose inhalers, breakthrough or gimmick? Chest 1985;88:276-84. 3 Bisgaard H. Delivery of inhaled medication to children. J Asthma 1997;34:443-67. 4 Tal A, Golan H, Grauer N, Aviram M, Albin D, Quastel MR. Deposition pattern of radiolabelled salbutamol inhaled from a metered-dose inhaler by means of a spacer with mask in young children with airway obstruction. J Pediatr 1996;128:479-84. 5 Wildhaber JH, Devadason SG, Hayden MJ, James R, Dufty AP, Fox RA, et al. Electrostatic charge on a plastic spacer device influences the delivery of salbutamol. Eur Respir J 1996;9:1943-6. 6 Everard ML, Clark AR, Milner AD. Drug delivery from holding chambers with attached facemask. Arch Dis Child 1992;67:580-5. 7 Barry PW, O Callaghan C. Multiple actuations of salbutamol MDI into a spacer device reduce the amount of drug recovered in the respirable range. Eur Respir J 1994;7:1707-9. 8 O Callaghan C, Lynch J, Robertson C. Improvement in sodium cromoglycate delivery from a spacer device by use of an antistatic lining, immediate inhalation, and avoiding multiple actuations of drug. Thorax 1993;48:603-6. 9 Barry PW, O Callaghan C. The effect of delay, multiple actuations and spacer static charge on the in vitro delivery of budesonide from the Nebuhaler. Br J Clin Pharmacol 1995;40:76-8. 10 Barry PW, Robertson CF, O Callaghan C. Optimum use of a spacer device. Arch Dis Child 1993;69:693-4. 11 Hindle M, Chrystyn H. Relative bioavailability of salbutamol to the lung following inhalation using metered-dose inhalation methods and spacer devices. Thorax 1994;49:549-53. 12 Wildhaber JH, Devadason SG, Eber E, Hayden MJ, Everard ML, Summers QA, et al. Effect of electrostatic charge, flow, delay and multiple actuations on the in vitro delivery of salbutamol from different small volume spacers for infants. Thorax 1996;51:985-8. 13 Melchor R, Biddiscombe MF, Mak VHF, Short MD, Spiro SG. Lung deposition patterns of directly labelled salbutamol in normal subjects and in patients with reversible airflow obstruction. Thorax 1993;48:506-11. 14 Newman SP, Moren F, Pavia D, Little F, Clarke SW. Deposition of pressurized suspension aerosols inhaled through extension devices. Am Rev Resp Dis 1981;124:317-20. 15 Piérart F, Wildhaber JH, Vrancken I, Devadason SG, Le Souef PN. Washing plastic spacers in household detergent reduces electrostatic charge and greatly improves delivery. Eur Respir J 1999;13:673-8. 16 Wildhaber JH, Devadason SG, Hayden MJ, Eber E, Summers QA, LeSouef PN. Aerosol delivery to wheezy infants: a comparison between a nebulizer and two small volume spacers. Pediatr Pulmonol 1997;23:212-6. 17 Newman SP. Metered dose pressurised aerosols and the ozone layer. Eur Respir J 1990;3:495-7. 18 Manzer LE. The CFC-ozone issue: progress on the development of alternatives to CFCs. Science 1990;249:31-5. 19 Barry PW, O Callaghan C. In vitro comparison of the amount of salbutamol available for inhalation from different formulations used with different spacer devices. Eur Respir J 1997;10:1345-8. 20 Leach CL, Davidson P. Breath-actuated MDI delivers 60% airway deposition with a new CFC-free beclomethasone formulation. Eur Respir J 1996;9:255. 21 Seale JP, Harrison LI. Effect of changing the fine particle mass of inhaled beclomethasone dipropionate on intrapulmonary deposition and pharmacokinetics. Respir Med 1998;92 (Suppl):9-15. 22 Schultz D, Carlson S, Ross D. In vitro performance characteristics of two CFC-free metered dose inhalers with large and small volume spacers. Eur Respir J 1996;9(Suppl 23):255s. 23 Everard ML, Milner AD. Pressure flow characteristics of the valve in spacer devices (letter to the Editor). Arch Dis Child 1990;65:159. 24 Lipworth BJ, Clark DJ. Lung delivery of non-cfc salbutamol via small volume metal spacer and large volume plastic spacer devices compared with an open vent jet nebulizer. Br J Clin Pharmacol 1998;45:160-3. 25 Bisgaard H. A metal aerosol holding chamber devised for young children with asthma. Eur Respir J 1995;8:856-60. 26 Anhoj J, Bisgaard H, Lipworth BJ. Effect of electrostatic charge in plastic spacers on the lung delivery of HFA-salbutamol in children. Br J Clin Pharmacol 1999;47:333-6. 27 Bisgaard H, Anhoj J, Klug B, Berg E. A non-electrostatic spacer for aerosol delivery. Arch Dis Child 1995;73:226-30.

Swiss Medical Weekly: Call for papers Swiss Medical Weekly Official journal of the Swiss Society of Infectious disease the Swiss Society of Internal Medicine the Swiss Respiratory Society The many reasons why you should choose SMW to publish your research What Swiss Medical Weekly has to offer: SMW s impact factor has been steadily rising, to the current 1.537 Open access to the publication via the Internet, therefore wide audience and impact Rapid listing in Medline LinkOut-button from PubMed with link to the full text website http://www.smw.ch (direct link from each SMW record in PubMed) No-nonsense submission you submit a single copy of your manuscript by e-mail attachment Peer review based on a broad spectrum of international academic referees Assistance of our professional statistician for every article with statistical analyses Fast peer review, by e-mail exchange with the referees Prompt decisions based on weekly conferences of the Editorial Board Prompt notification on the status of your manuscript by e-mail Professional English copy editing No page charges and attractive colour offprints at no extra cost Editorial Board Prof. Jean-Michel Dayer, Geneva Prof. Peter Gehr, Berne Prof. André P. Perruchoud, Basel Prof. Andreas Schaffner, Zurich (Editor in chief) Prof. Werner Straub, Berne Prof. Ludwig von Segesser, Lausanne International Advisory Committee Prof. K. E. Juhani Airaksinen, Turku, Finland Prof. Anthony Bayes de Luna, Barcelona, Spain Prof. Hubert E. Blum, Freiburg, Germany Prof. Walter E. Haefeli, Heidelberg, Germany Prof. Nino Kuenzli, Los Angeles, USA Prof. René Lutter, Amsterdam, The Netherlands Prof. Claude Martin, Marseille, France Prof. Josef Patsch, Innsbruck, Austria Prof. Luigi Tavazzi, Pavia, Italy We evaluate manuscripts of broad clinical interest from all specialities, including experimental medicine and clinical investigation. We look forward to receiving your paper! Guidelines for authors: http://www.smw.ch/set_authors.html Impact factor Swiss Medical Weekly 2 1.8 1.537 1.6 1.4 1.2 1.162 1 0.8 0.770 0.6 0.4 0.2 0 1995 1996 1997 1998 1999 2000 2002 2003 2004 Schweiz Med Wochenschr (1871 2000) Swiss Med Wkly (continues Schweiz Med Wochenschr from 2001) Editores Medicorum Helveticorum All manuscripts should be sent in electronic form, to: EMH Swiss Medical Publishers Ltd. SMW Editorial Secretariat Farnsburgerstrasse 8 CH-4132 Muttenz Manuscripts: Letters to the editor: Editorial Board: Internet: submission@smw.ch letters@smw.ch red@smw.ch http://www.smw.ch