Lesson 4 ECG 2 Electrocardiography

Similar documents
Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Analysis Procedure. Rev

Lesson 2 EMG 2 Electromyography: Mechanical Work

Physiology Lessons for use with the Biopac Student Lab

Lesson 10 Aerobic Exercise Physiology

Biopac Student Lab Lesson 5 ELECTROCARDIOGRAPHY (ECG) I Procedure. Rev

Lesson 5 BIOPAC Systems, Inc. Manual Revision PL3.7.5

Lesson 5 EEG 1 Electroencephalography: Brain Rhythms

Lesson 2 ELECTROMYOGRAPHY II Motor unit recruitment Fatigue

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease

Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG

Physiology Lessons for use with the Biopac Student Lab. Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG

iworx Sample Lab Experiment HH-4: The Six-Lead Electrocardiogram

Biopac Student Lab Lesson 3 ELECTROENCEPHALOGRAPHY (EEG) I Analysis Procedure. Rev

INTRODUCTION TO ECG. Dr. Tamara Alqudah

Biopac Student Lab Lesson 20 SPINAL CORD REFLEXES Analysis Procedure. Rev

Electrocardiography I Laboratory

Richard Pflanzer, Ph.D. J.C. Uyehara, Ph.D. William McMullen David Pittman, Ph.D. Manual Revision PL3.7.0 M3.0.6 Modified by David Pittman

Lab 7. Physiology of Electrocardiography

iworx Sample Lab Experiment HH-4: The Six-Lead Electrocardiogram

Biopac Student Lab Lesson 8 RESPIRATORY CYCLE I Analysis Procedure. Rev

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation

LABORATORY INVESTIGATION

BSL PRO Lesson H03: Nerve Conduction Velocity: Along the Ulnar Nerve of a Human Subject

#4 Cardiovascular I The Heart & EKG

Science in Sport. 204a ECG demonstration (Graph) Read. The Electrocardiogram. ECG Any 12 bit EASYSENSE. Sensors: Loggers: Logging time: 10 seconds

Overview. Page 1 of 9. Impedance Cardiography

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation

Exercise 5: Muscle Physiology I - Electromyography

Experiment HH-2: The Electrocardiogram and Heart Sounds

Cardiovascular Effects of Exercise. Background Cardiac function

Each student should record the ECG of one of the members of the lab group and have their own ECG recorded.

Sample. Analyzing the Heart with EKG. Computer

Lect.6 Electrical axis and cardiac vector Cardiac vector: net result Vector that occurs during depolarization of the ventricles Figure:

Warning: Avoid putting the tip of the syringe needle too deeply into the hole to avoid damage to the heart tissue.

The Normal Electrocardiogram

Figure 1 The respiratory cycle

5- The normal electrocardiogram (ECG)

Cardiovascular Physiology

Figure 1 muscle tissue to its resting state. By looking at several beats you can also calculate the rate for each component.

Lesson 9. Measuring Emotional Responses to the Acoustic Startle Reflex: Stimulus Habituation & Sensitization

Lab 5: Electromyograms (EMGs)

iworx Sample Lab Experiment HC-3: Pulse Wave Velocity

Analyzing the Heart with EKG

Exercise 6: Muscle Physiology II Twitch & Summation

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

ELECTROCARDIOGRAPHY (ECG)

Lab #3: Electrocardiogram (ECG / EKG)

Axis. B.G. Petty, Basic Electrocardiography, DOI / _2, Springer Science+Business Media New York 2016

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent?

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

EKG. Danil Hammoudi.MD

Lab 4: Introduction to Physiological Measurements - Cardiovascular

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals

HUMAN ANATOMY AND PHYSIOLOGY

BIO 360: Vertebrate Physiology Performing and analyzing an EKG Lab 11: Performing and analyzing an EKG Lab report due April 17 th

Sheet 5 physiology Electrocardiography-

= BIOPAC. J ~ ~ t. Lesson 6 ELECTROCARDIOGRAPHY Bipolar Leads (Leads 1, II, III) Einthoven's Law Mean Electrical Axis on the Frontal Plane

BASIC CONCEPT OF ECG

ELECTROCARDIOGRAPHY, ECG

Lesson 1: Types of ECG s

Lab #3: Electrocardiogram (ECG / EKG)

ECG. Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13

ECG INTERPRETATION MANUAL

Introduction to ECG Gary Martin, M.D.

ECG and Cardiac Electrophysiology

Overview Active Learning

Electrocardiography negative zero LA/VL RA/VR LL/VF recording electrode exploring electrode Wilson right arm right arm, left arm left arm

- why the T wave is deflected upwards although it's a repolarization wave?

How Do We Sense, Think, and Move? -- Lab #11 Bioelectronics Measuring Electrical Properties of the Body

Exercise 2: Effects of Cold Temperature Aim: To record changes in heart rate after the heart is bathed in cold Ringer s solution.

Experiment HM-11: Electromyograms (EMG) for Paired Arm Wrestling

Birmingham Regional Emergency Medical Services System

iworx Sample Lab Experiment HP-7: Hypothesis-driven Biofeedback Lab/Research Study

Experiment HM-7: Electromyogram (EMG) Activity in Antagonistic Muscles and Range of Motion

This presentation will deal with the basics of ECG description as well as the physiological basics of

2The Concept of Periodic Functions

Teacher s Guide. Physiology Lessons for use with the Biopac Science Lab MP40

The Electrocardiogram part II. Dr. Adelina Vlad, MD PhD

Educators have placed an emphasis on the development of laboratory materials that

8/20/ Identify the functions of common ECG machines. 3.3 Explain how each ECG machine control is used. 3.4 Recognize common electrodes.

EKG Sensor Product Number: ENEKG189

ECE ECE PRINCIPLES OF BIOMEDICAL SYSTEMS & DEVICES LAB 1 - ELECTROCARDIOGRAM

SS11LA Pneumotach Transducer

iworx Sample Lab Experiment AN-5: Cockroach Leg Mechanoreceptors

Biology 212: Anatomy and Physiology II. Lab #5: Physiology of the Cardiovascular System For Labs Associated With Dr. Thompson s Lectures

Electrocardiography Normal 5. Faisal I. Mohammed, MD, PhD

Determining Axis and Axis Deviation on an ECG

Lab Activity 24 EKG. Portland Community College BI 232

ECG-PS ECG SIGNAL MEASUREMENT MODULE. Operation Manual. March 2005

ELECTROMYOGRAM ANALYSIS OF MUSCLE FUNCTION INTRODUCTION

PROBLEM SET 3. Assigned: February 19, 2004 Due: February 26, 2004

CARDIOVASCULAR PHYSIOLOGY ECG. Dr. Ana-Maria Zagrean

3/26/15 HTEC 91. EKG Sign-in Book. The Cardiac Cycle. Parts of the ECG. Waves. Waves. Review of protocol Review of placement of chest leads (V1, V2)

1724 Lab: Frog Skeletal Muscle Physiology (Marieb Exercise 16A) Marieb/iWorx / Ziser, 2002

Relax and Learn At the Farm 2012

CARDIOCOMP Electrocardiography Primer

Humans make voluntary decisions to talk, walk, stand up, or sit down. The

By the end of this lecture, you will be able to: Understand the 12 lead ECG in relation to the coronary circulation and myocardium Perform an ECG

Farah Khreisat. Raghad Abu Jebbeh. Faisal Mohammad. 1 P a g e

12 Lead ECG Interpretation: Color Coding for MI s

Transcription:

Physiology Lessons for use with the Biopac Science Lab MP40 Lesson 4 ECG 2 Electrocardiography PC running Windows XP or Mac OS X 10.3-10.4 Lesson Revision 1.20.2006 BIOPAC Systems, Inc. 42 Aero Camino, Goleta, CA 93117 (805) 685-0066, Fax (805) 685-0067 info@biopac.com www.biopac.com

Page 2 Biopac Science Lab Lesson 4 ELECTROCARDIOGPAHY 2 I. SCIENTIFIC PRINCIPLES Recall from Lesson 3, ECG 1, that each cardiac cycle, or heartbeat, begins with a bioelectric impulse generated by the sinoatrial (SA) node, the pacemaker of the heart. The bioelectric impulse is conducted through the heart by the pacemaker system and travels as a wave of depolarization followed in a few thousandths of a second by a wave of repolarization. Ventricular muscle depolarization and repolarization are represented in the electrocardiogram (ECG) as the QRS complex and the T wave, respectively. The body contains fluids with ions that allow for electric conduction. This makes it possible to use electrodes on the surface of the skin to detect electrical activity in and around the heart and use an electrocardiograph to record the activity. Conveniently, the legs and arms act as simple extensions of points in the torso, allowing the recording and ground electrodes to be placed on the wrists and ankles (Fig. 4.1). - Lead I + - - Lead II + Lead III + right arm left arm right leg (ground) left leg Fig. 4.1 Electrode placement options Fig. 4.2 Einthoven s triangle and vectorgraph The ECG is a record of the overall spread of electric current through the heart as a function of time in the cardiac cycle. The direction of polarity (+ or -) of the recorded waveforms depends upon the location of the recording electrodes on the surface of the body and whether the electrical activity during the cardiac cycle is coming toward or going away from the surface electrode. In general, as a wave of depolarization approaches a positive electrode, a positive voltage is seen by that electrode. If the wave of depolarization is traveling toward a negative electrode, a negative voltage will be seen. The term lead is defined as a spatial arrangement of two recording electrodes on the body. One lead is labeled positive (+) and the other negative (-). The electrode placements define the recording direction of the lead, which is called the lead axis or angle. The axis is determined by the direction when going from the negative to positive electrode. The electrocardiograph computes the voltage difference (magnitude) between the positive and negative electrodes and displays the changes in voltage difference with time. The standard bipolar limb leads are: Lead I = right arm (-), left arm (+) Lead II = right arm (-), left leg (+) Lead III = left arm (-), left leg (+) The three standard bipolar limb leads may be used to construct an equilateral triangle, called Einthoven s triangle, at the center of which lies the heart (Fig. 4.2, left). Each side of the triangle represents one of the bipolar limb leads, and each lead forms a 60 angle with the two opposite leads. By placing Lead I through the center of the heart and moving the axes of Lead II and Lead III horizontally until they intersect Lead I at the center of the heart, a vectorgraph can be constructed (Fig. 4.2, right). The vectorgraph is used to plot vectors and aid in visualizing the electrical axes of the heart. At any given moment during the cardiac cycle (during the QRS complex, for example), the net electrical activity seen by a lead may be represented by a vector. A vector is an entity that can be represented by an arrow of specific length pointing in a specific direction. A vector has both magnitude and direction. The length of the arrow, divided equally into 0.1 mv units, represents the vector s magnitude, and is expressed as a mean potential. The direction of the vector, in degrees, is represented by its position relative to the horizontal 0º to 180º line (Lead I) on the vectorgraph. The dominant waveform in all three bipolar limb leads is the QRS complex signifying ventricular depolarization. The mean axis of ventricular depolarization is called the mean QRS axis. The mean QRS axis is diagnostically important

Lesson 4: ECG 2 Page 3 because it represents the majority of the heart s electrical activity occurring during a cardiac cycle. When the bipolar limb leads are used to obtain data for plotting the vector, the mean QRS axis lies in the frontal plane because the electrical picture seen by the recording electrodes placed on the extremities lies in the frontal plane. In normal persons, the mean QRS axis is between -30 and +90 (Fig. 4.2). In the frontal plane, the mean QRS axis is often between 0 and +90. A mean QRS axis between +90 and +180 indicates right axis deviation (RAD). RAD occurs, for example, in persons with right ventricular hypertrophy (over-sized right ventricle). A mean QRS axis between -30 and -90 indicates left axis deviation (LAD). LAD is found in patients with left anterior blockage in the conduction system of the ventricle. To define the mean QRS axis precisely, the cardiologist defines it in three dimensions (X, Y and Z planes). This is done in practice by using a standardized set of 12 leads. In this lesson, you will sequentially record two of the bipolar limb leads (Lead I and Lead II) and extract the data from the ECG necessary to determine the mean QRS axis in the frontal plane. A simple way to approximate the mean QRS axis in the frontal plane is to plot the magnitude of the R wave as measured from Lead I and Lead II (Fig. 4.3). magnitude of R wave from leads 0,0 I - - - - + + + + + + Mean Electrical Axis III Mean Electrical Axis Fig. 4.3 Mean electrical axis In this lesson, you will record data from Lead I and Lead III connections with the Subject lying down and sitting up. II. EXPERIMENTAL OBJECTIVES 1) To record ECG from Leads I and III in the following conditions: lying down, sitting up, and breathing deeply while sitting. 2) To compare the direction of the QRS complex (+ or ) with the direction of the lead axis. 3) To estimate the mean ventricular potential. 4) To estimate the mean electrical axis of the QRS complex using two methods. III. MATERIALS Computer system (running Windows XP or Mac OS X) Biopac Science Lab system (MP40 and software) Electrode lead set (40EL lead set) Disposable vinyl electrodes (EL503), four electrodes per subject Cot or lab pillow Protractor Two different color pens or pencils

Page 4 Biopac Science Lab IV. EXPERIMENTAL METHODS A. Set Up EQUIPMENT SUBJECT - Lead I + right forearm WHITE lead left forearm RED lead Fig. 4.4 right leg BLACK lead (Ground) Fig. 4.5 Electrodes for Leads I & III/Lead I configuration FAST TRACK 1. Turn the computer ON. 2. Set the MP40 dial to OFF. 3. Plug the equipment in as follows: Electrode leads (40EL) MP40. 4. Attach four electrodes to the Subject as shown in Fig. 4.5. 5. Connect the electrode leads (40EL) for a Lead I configuration, matching lead color to electrode position as shown above. Details Attach four electrodes to the Subject to allow Lead I and Lead III configurations, as shown in Fig. 4.5: one electrode on the right leg, just above the ankle bone one electrode on the left leg, just above the ankle bone one electrode on the left wrist (same side of arm as the palm of hand) one electrode on the right wrist (same side of arm as the palm of hand). For optimal electrode response, place electrodes on the skin at least five minutes before starting the Check section. Lead I will be recorded first, so connect the leads for a Lead I configuration: white: right wrist; red: left wrist; black: right ankle. IMPORTANT Clip each electrode lead color to its specified electrode position. 6. Start the Biopac Science Lab software. 7. Choose lesson L04-ECG-2 and click OK. 8. Type in a unique file name. No two people can have the same file name, so use a unique identifier, such as the Subject s nickname or student ID#. 9. Click OK. This ends the Set Up procedure.

Lesson 4: ECG 2 Page 5 B. Check MP40 Check FAST TRACK Details 1. Set the MP40 dial to ECG/EOG. 2. Press and hold the Check pad on the MP40. 3. Click when the light is flashing. 4. Wait for the MP40 Check to stop. 5. Let go of the Check pad. 6. Click Continue. Signal Check 7. Click Check Signal. 8. Wait for the Signal Check to stop. 9. Review the data. If correct, go to the Record section. Continue to hold the pad down until prompted to let go. The MP40 check procedure will last five seconds. The light should stop flashing when you let go of the Check pad. When the light stops flashing, click Continue. Lead I configuration is used for the signal check. Subject should be supine, relaxed, and still. The electrocardiograph is very sensitive to small changes in voltage caused by contraction of skeletal muscles, and the Subject s arms and legs need to be relaxed so that the muscle (EMG) signal does not corrupt the ECG signal. The eight-second Signal Check recording should resemble Fig. 4.6. If incorrect, click. Fig. 4.6 If the recording does not show a recognizable ECG waveform with R-wave peaks and no large baseline drift, click Redo Signal Check.

Page 6 Biopac Science Lab C. Record FAST TRACK Details 1. Prepare for the recording and have the Subject lie down and relax. SEGMENT 1 Supine, Lead I Watch the Help menu videos to prepare for the recording. You will record two segments for each lead configuration, one with the Subject lying down, and another with the Subject sitting up. In order to work efficiently, read this entire section so you will know what to do before recording. Check the last line of the journal and note the total amount of time available for the recording. Stop each recording segment as soon as possible so you don t use an excessive amount of time (time is memory). Tips for obtaining optimal data To minimize muscle (EMG) corruption of the ECG and baseline drift: a) The Subject should not talk or laugh during any of the recording segments. b) When lying down or sitting up, the Subject should be completely relaxed. c) When sitting up, the Subject s arms should be supported on an armrest (if available). d) The recording should be suspended before the Subject prepares for the next recording segment. e) The Subject should breathe normally during the recording to minimize EMG from the chest area. f) Make sure electrodes do not peel up. 2. Click. When you click Record, the recording will begin and an append marker labeled Supine, Lead I will automatically be inserted. 3. Record for 30 seconds. Subject should be supine, relaxed, and still (seconds 0-30). 4. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 5. Review the data. Your data should resemble Fig. 4.7. If correct, go to Step 6. If incorrect, click Redo. Recording continues Fig. 4.7 Segment 1 (Supine, Lead I) A little baseline drift when the Subject breathes in and out does not indicate incorrect data. The data is incorrect if: a. The Suspend button was pressed prematurely b. An electrode peeled up, giving a large baseline drift c. The Subject has too much muscle (EMG) artifact If the data is incorrect, click Redo and repeat Steps 1-5; the last data segment you recorded will be erased.

Lesson 4: ECG 2 Page 7 SEGMENT 2 Supine, Lead III 6. Disconnect the leads from the Lead I configuration and reconnect to establish a Lead III configuration (Fig. 4.8). - Lead III + left forearm WHITE lead white: left wrist red: left ankle black: right ankle right leg BLACK lead (Ground) left leg RED lead Fig. 4.8 Lead III Configuration 7. Click Resume. When you click Resume, the recording will begin and an append marker labeled Lying down, Lead III will automatically be inserted. 8. Record for 30 seconds. Subject should be supine, relaxed, and still (seconds 30-60). 9. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 10. Review the data on the screen. If correct, go to Step 11. Your data should resemble Fig. 4.9. If incorrect, click Redo. Fig. 4.9 Segment 2 (Supine, Lead III) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a. The Suspend button was pressed prematurely b. An electrode peeled up, giving a large baseline drift c. The Subject has too much muscle (EMG) artifact If data was incorrect, click Redo and repeat Steps 7-10; the last data segment you recorded will be erased. Recording continues

Page 8 Biopac Science Lab SEGMENT 3 Seated, Lead III 11. Have the Subject quickly get up and sit in a chair, with arms relaxed. 12. Click Resume as soon as possible after the Subject sits up. 13. After 10 seconds of recording, the Subject should breathe in and out once such that the breath is audible Note Do not insert markers. Subject should sit with arms relaxed, preferably in a chair with armrests. In order to capture the heart rate variation, it is important that you resume recording as quickly as possible after the Subject sits. However, it is also important that you do not click Resume while the Subject is in the process of sitting or you will capture motion artifact. When you click Resume, the recording will continue and an append marker labeled Seated, Lead III will be automatically inserted. After 10 seconds of recording, the Director should instruct the Subject to breathe in and out once such that the inhale and exhale are audible. Note Do not insert markers in this segment because the automatic processing for this lesson splits the recorded segments into separate channels (after Done is clicked), which makes the time base for markers meaningless so they are not retained. The recording should run for 30 seconds (seconds 60-90). 14. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 15. Review the data on the screen. If correct, go to Step 13. Your data should resemble Fig. 4.10. If incorrect, click Redo. Fig. 4.10 Segment 3 (Seated, Lead III) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a) The Suspend button was pressed prematurely. b) An electrode peeled up, giving a large baseline drift. c) The Subject has too much muscle (EMG) artifact. If data was incorrect, click Redo and repeat Steps 11-15; the last data segment you recorded will be erased. Recording continues

Lesson 4: ECG 2 Page 9 SEGMENT 4 Seated, Lead I 16. Disconnect the leads from the Lead III configuration and reconnect to establish a Lead I configuration (Fig. 4.11) while Subject remains seated and relaxed. - Lead I + right forearm WHITE lead left forearm RED lead white: right wrist red: left wrist black: right ankle right leg BLACK lead (Ground) Fig. 4.11 Lead 1 configuration 17. Click Resume. When you click Resume, the recording will continue and an append marker labeled Seated, Lead I will be automatically inserted. 18. After 10 seconds of recording, the Subject should breathe in and out once such that the breath is audible. Note Do not insert markers. After about 10 seconds of recording, the Director should instruct the Subject to breathe in and out once such that the inhale and exhale are audible. Note Do not insert markers in this segment because the automatic processing for this lesson splits the recorded segments into separate channels (after Done is clicked), which makes the time base for markers meaningless so they are not retained. The recording should run for 30 seconds (seconds 90-120). 19. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data. 20. Review the data. If correct, go to Step 21. Your data should resemble Fig. 4.12. If incorrect, click Redo. Recording continues Fig. 4.12 Segment 4 (Seated, Lead I) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a) The Suspend button was pressed prematurely. b) An electrode peeled up, giving a large baseline drift. c) The Subject has too much muscle (EMG) artifact. If data was incorrect, click Redo and repeat Steps 17-20; the last data segment you recorded will be erased.

Page 10 21. Optional: Click Resume to record additional segments. 22. Click Done. 23. Click Yes. 24. Choose an option and click OK. 25. Remove the electrodes. END OF RECORDING Biopac Science Lab Optional: You can record additional segments by clicking Resume instead of Done. A time marker will be inserted at the start of each added segment. A pop-up window with options will appear. Click Yes (or No if you want to redo the last segment). When you click Yes, a dialog with options will be generated. Make your choice, and click OK. If you choose Analyze current data file, go to the Analyze section for directions. Unclip the electrode lead connectors and peel off the electrodes. Throw out the electrodes.

Lesson 4: ECG 2 Page 11 V. ANALYZE In this section, you will examine ECG components of cardiac cycles and measure amplitudes (mv) and durations (msecs) of the ECG components. Note: Interpreting an ECG is a skill that requires practice to distinguish between normal variation and those arising from medical conditions. Do not be alarmed if your ECG is different than the examples shown or from the tables and figures in Section I (Scientific Principles). Because ECGs are widely used, basic elements have been standardized to simplify reading ECGs. ECGs have standardized grids of lighter, smaller squares and, superimposed on the first grid, a second grid of darker and larger squares (Fig. 4.13). The smaller grid always has time units of 0.04 seconds on the x-axis, and the darker vertical lines are spaced 0.2 seconds apart. The horizontal lines represent amplitude in mv. The lighter horizontal lines are 0.1 mv apart, and the darker grid lines represent 0.5 mv. Fig. 4.13 Standard ECG grid FAST TRACK Details 1. Enter the Review Saved Data mode and choose the correct data file. Note Channel Number (CH) designations: To review saved data, choose Analyze current data file from the Done dialog after recording data, or choose Review Saved Data from the Lessons menu and browse to the required file. Channel Displays CH1 CH2 CH3 CH4 Supine, Lead I Supine, Lead III Seated, Lead III Seated, Lead I Important For this lesson analysis, segments that were recorded sequentially have been split into channels for relative comparison and do not share a time base. 2. Set up measurement boxes for: Channel SC Measurement Max Fig. 4.14 IMPORTANT: Channels do not share a time base The measurement boxes are above the marker region in the data window. Each measurement has three sections: channel number, measurement type, and result. The first two sections are pull-down menus that are activated when you click them. SC SC Delta Mean Max: finds the maximum amplitude value within the selected area (including the endpoints) by the I-Beam tool. Delta: computes the difference in amplitude between the last point and the first point of the selected area. It is particularly useful for taking ECG measurements because the baseline does not have to be at zero to obtain accurate, quick measurements. Note The polarity of Delta is based on (first point-last point). First point is where you begin your selection; last point is where you release the mouse to end your selection. 3. For each of the leads, note whether the R wave is positive or negative. A This is just a visual check as to whether the R wave goes up (positive) or down (negative).

Page 12 Biopac Science Lab Analysis continues 4. Set up the display window for optimal viewing of one cardiac cycle in the supine channels. 5. Using the I-Beam tool, select and measure Max from a cycle in the center of each channel. B Fig. 4.15 The following tools help you adjust the data window: Autoscale Horizontal Autoscale Waveforms Zoom Your window should resemble Fig. 4.15. Horizontal (Time) Scroll Bar Vertical (Amplitude) Scroll Bar Zoom Previous/Back You ll use the recorded measurements to construct the graphical estimate of the Mean QRS Potential in the Data Report. 6. Place an event marker to indicate where the QRS measurement was taken you will add four markers. Insert the event marker directly above the selected R wave. To place an event marker after the data has been recorded, click the cursor in the event marker region under the marker label bar (above the top data channel). An inverted triangle should appear, which, when selected, will be yellow. Type marker labels: Max Lead I down, Max Lead I up, Max Lead III down, and Max Lead III up. 7. Repeat Steps 5 and 6 for a cardiac cycle in the Seated channel. B 8. Select the breathe in area of a cardiac cycle in the CH3 and repeat Steps 5 and 6. B 9. Select the breathe in area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B Breathe in was recorded in the Seated, Lead I and Seated, Lead III segments, now displayed as CH3 and CH4. The breathe change should occur as an upward spike in the data. 10. Select the breathe out area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B Breathe out was recorded in the Seated, Lead I and Seated, Lead III segments, now displayed as CH3 and CH4. The breathe change should occur as a downward spike in the data. 11. Select the breathe out area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B 12. Go back to the Max Lead I down marker created in Step 6. This is the same QRS region you measured and marked in Steps 5 and 6. It represents a QRS cycle in the supine, relaxed

Lesson 4: ECG 2 Page 13 13. Measure and record the amplitudes of the Q, R, and S waves individually for both Lead I and Lead III for each segment. C state. You will be taking a total of six measurements. You may wish to look at one channel at a time by hiding the other channel(s) using the channel boxes. To measure a peak, go from the baseline (isoelectric line) to the peak of the wave. Fig. 4.16 Sample measurement of the R-wave peak 14. Save or print the data file. You may save the data, save notes that are in the journal, or print the data file. 15. Exit the program. 16. Set the MP40 dial to Off. END OF LESSON 4 Complete the Lesson 4 Data Report that follows.

Page 14 Biopac Science Lab The Data Report starts on the next page.

Biopac Science Lab Page 15 Lesson 4 ELECTROCARDIOGRAPHY 2 DATA REPORT Student s Name: Lab Section: Date: These are sample questions. You should amend, add, or delete questions to support your curriculum objectives. I. Data and Calculations Subject Profile Name Gender: Male / Female Height Weight Age Direction of R-Waves for Different Leads A. Put a check in the column to indicate whether the R-wave is positive (+) or negative ( ) for each lead configuration: Table 4.1 Lead Lead I Lead III R-wave + Mean QRS Potential and Axis Graphical Estimate Using R-wave Amplitude B. Use Table 4.2 to record measurements from the Data Analysis section: Table 4.2 QRS Potential and QRS Axis Condition Potential QRS Supine Seated Breathe In Breathe Out Lead I Max Lead III Max Mean QRS Potential Mean QRS Axis

Page 16 Biopac Science Lab A simple way to approximate the mean QRS axis in the frontal plane is to plot the magnitude of the R wave as measured from Lead I and Lead III (Fig. 4.3, repeated below from Section 1). magnitude of R wave from leads 0,0 I - - - - + + + + + + Mean Electrical Axis III Mean Electrical Axis Fig. 4.3 1) Determine the R-wave amplitude (mv) in Lead I and Lead III. 2) Plot the Lead I R-wave amplitude value on the Lead I axis. Draw an arrow from the center of the vectorgraph to the amplitude value. 3) Plot the Lead III R-wave amplitude value on the Lead III axis. Draw an arrow from the center of the vectorgraph to the amplitude value. 4) Draw a line perpendicular to the tip of each arrowhead 5) Determine the point of intersection of these two perpendicular lines. 6) Draw a new vector from point 0,0 at the center of the vectorgraph to the point where the perpendicular lines intersect with one another. The direction (in degrees) of this resulting vector approximates the mean QRS axis. The length (mv) of this vector approximates the mean QRS potential. To determine Mean QRS Potential from the plotted vector, subdivide it into 0.1 mv increments exactly as the leads are in the vectorgraph. Alternatively, the length of the vector can be measured (with a compass, string, notepaper, etc.) and then transferred to a lead line to determine the mv value. One way to approximate the mean electrical axis in the frontal plane is to plot the magnitude of the R wave from Lead I and Lead III, as shown in Fig 4.2 in Section I (Scientific Principles). 1. Draw a perpendicular line from the ends of the vectors (right angles to the axis of the Lead) using a protractor or right angle guide. 2. Determine the point of intersection of these two perpendicular lines. 3. Draw a new vector from point 0,0 to the point of intersection. The direction of this resulting vector approximates the mean electrical axis of the heart. The length of this vector approximates the mean potential of the heart.

Lesson 4: ECG 2 Page 17 Create two plots on each of the following graphs, using data from Table 4.2. Use a different color pencil or pen for each plot. Graph 1: Lead I Lying Down and Sitting Up 0,0 mv -.1.1.2.3.4.5 1 1.5 2 0 Lead I.5 1 1.5 2 mv 50 40 30 20 10 60 120 Lead III 110 100 2 90 80 70 Lead II From the above graph, find the following values: Condition Mean QRS Potential Mean QRS Axis Lying Down Sitting Up Explain the difference (if any) in Mean QRS Potential and Axis under the two conditions:

Page 18 Biopac Science Lab Graph 2: Breathing In /Breathing Out or Lead III Lying Down/Sitting Up 0,0 mv -.1.1.2.3.4.5 1 1.5 2 0 Lead I.5 1 1.5 2 mv 50 40 30 20 10 60 120 Lead III 110 100 2 90 80 70 Lead II From the above graph, find the following values: Condition Mean QRS Potential Mean QRS Axis Lying Down Sitting Up Explain the difference in Mean QRS Potential and Axis for the Lying Down data in this plot (Graph 2) and the first plot (Graph 1). C. Mean QRS Potential and Magnitude More Accurate Approximation Add the Q, R, and S potentials to obtain net potentials. Lying down: Lead I Q R S Lead III Q R S QRS Net 1 QRS Net 2

Lesson 4: ECG 2 Page 19 II. Data Summary and Questions D. Define ECG. E. Define Einthoven s Triangle. F. What factors affect the orientation of the mean electrical axis? G. Refer to Table 4.2: How did the amplitudes of Lead I and Lead III change between inhalation and exhalation? Did the heart s axis and magnitude change? H. What factors affect the amplitude of the R wave recorded on the different leads? I. Compare the mean electrical axis and magnitude obtained when i. using just the amplitude measurement of the R wave vs. net potentials ii. lying down vs. sitting up End of Biopac Science Lab Lesson 4 Data Report

Biopac Science Lab Page 20 VI. ACTIVE LEARNING LAB Design a new experiment to test or verify the scientific principle(s) you learned in the Biopac Science Lab recording and analysis segments. For this lesson, you might examine how gender, duration of exercise, and changing body position influence the ECG and/or time to return to baseline. Design Your Experiment Use a separate sheet to detail your experiment design, and be sure to address these main points: A. Hypothesis Describe the scientific principle to be tested or verified. B. Materials List the materials will you use to complete your investigation. C. Method Describe the experimental procedure be sure to number each step to make it easy to follow during recording. See the Set Up section or Help > About Electrodes for electrode placement guidelines. Run Your Experiment D. Setup Set up the equipment and prepare the subject for your experiment. E. Recording Use the Record, Resume, and Suspend buttons in the Biopac Science Lab program to record as many segments as necessary for your experiment. Click Done when you have completed all of the segments required for your experiment. Analyze Your Experiment F. Set measurements relevant to your experiment and record the results in a Data Report.