Small Field Dosimetry

Similar documents
Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy

Much of what we will say

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Detector comparison for small field output factor measurements with flattening filter free photon beams

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Limits of Precision and Accuracy of Radiation Delivery Systems

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Small-Field Dosimetry: Its Importance In Clinic

Application(s) of Alanine

Dosimetry of Small Static Fields Used in External Beam Radiotherapy

Quality Assurance of Helical Tomotherapy Machines

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Basic dosimetric data for treatment planning system

Applications of Monte Carlo simulations to radiation dosimetry

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR.

How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues

SRS/SBRT Errors and Causes

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

Small field diode dosimetry

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainLAB m3 micro-multileaf collimator

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

INTRODUCTION. Material and Methods

The Journey of Cyberknife Commissioning

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)

Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Dosimetry and QA of proton and heavier ion beams

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Small Field Dosimetry & Radiochromic film

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Ionization chamber dosimetry of small photon fields: a Monte Carlo study on stopping-power

IMRT point dose measurements with a diamond detector

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

AAPM TG178: Towards the Recommendation of a Standard Dosimetry Protocol for Gamma Knife Stereotactic Radiosurgery

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Implantable MOSFET dosimeter response to 192 Ir HDR radiation

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.

Recent advances in dosimetry in reference conditions for proton and light-ion beams

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Contributors to this Talk

PGY-1. Resident Review Session Schedule

Intensity Modulated RadioTherapy

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Patient and Linac QA - present and the future

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams.

EORTC Member Facility Questionnaire

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Progress Report on Radiation Dosimetry at NPL

Expectations of Physics Knowledge for Certification

Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Unrivaled, End-to-End

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Quality assurance in external radiotherapy

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

Eric E. Klein, Ph.D. Chair of TG-142

Outline. Clinical Implementation of Monte Carlo Methods for External Photon Therapy. Outline. Indrin J. Chetty Henry Ford Hospital, Detroit MI

Detectors assessment for stereotactic radiosurgery with cones

Overview of Advanced Techniques in Radiation Therapy

Clinical Implementation of SRS/SBRT

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Transcription:

Small Field Dosimetry Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin Madison WI Conflict of Interest Statement: Part-time consultant to Accuray Outline Relevant to IMRT, SRS, and SBRT Main Small Field Dosimetry Issues Charged Particle Equilibrium Temporal Non-Constancy Partial Volume Effects Recombination/Charge Multiplication Calibration of Non-Standard Fields IMRT as a Large Collection of Small Non- Standard Fields IAEA Calibration Initiative for Non- Standard Fields Model-Based Calculation Methods Are Need for IMRT Treatment Planning IMRT = Σ small fields Dose = function(penumbra+leakage+head scatter) Need accurate treatment head model to get this right IMRT XRT IMRT beam profile XRT beam profile From Michael Sharpe, U. of Toronto 1

Intensity Profile possibly unconstrained intensity levels IMRT Is All About Using Small Fields Intensity Grouping limit delivery to a few discrete intensity levels Reconstituted from Beam Segments includes MLC constraints From Michael Sharpe, U. of Toronto IMRT Is All About Using Small Fields Accuracy of dose model at small field sizes is a consideration Convolutionsuperposition or Monte Carlo desirable Measurement Conventional RTP IMRT From Michael Sharpe, U. of Toronto Leaf Penumbra is Important Lines Defining One Half Value Layer of Beam Attenuation Lines Defining Light Field Boundary Central Axis It is important to have an accurate model of the curved leaf ends. More important for summation of small fields. 2

Gap Error is Fundamental fo Conventional MLCs Gap error Dose error 20.0 % Dose error 15.0 10.0 5.0 Range of gap width Gap error (mm) 2.0 1.0 0.5 0.2 0.0 0 1 2 3 4 5 Nominal gap (cm) From Tom Losasso, Memorial Sloan Kettering Potential Dosimetry Issues Charged particle equilibrium Different spectrum for collection of small fields Non-uniform dose Temporal non-constancy A very small effect for ion chambers May not be true for other dosimeters Partial volume effect Most important effect especially when measuring output factors for small fields Perturbation effects for small chambers Charge multiplication (failure of electronegative gas assumption) Electron emission from electrodes Non-Uniform Intensity Changes the Energy Spectrum and Intensity More High Energy Electrons Photon Beamlet Photon Beamlet More Low Energy Electrons 3

Dose per Incident Energy Fluence As a Function of Field Diameter A=Adipose, M=Muscle, B=Bone, L=Lung 4 MV, Parallel Beam Ahnesjo and Asparadakis, 1999 Phys Med Biol 44:R99-R155 Change in Stopping Power Ratio Comparison of Spencer-Attix (Δ=10 kev) restricted mass collisional stopping powers ratios (water/air) at 5 cm depth in water for various 6 MV beams with stereotactic and MLC beams. 6 MV beams Beam quality, TPR(20,10) Elekta SL-18 10 x 10 cm 2 1.0 cm diameter stereo field 0.3 cm diameter stereo field Siemens Primus 10 x 10 cm 2 Andreo, (1994) Sánchez- Doblado, (2003) 0.690 1.1187 1.1188 1.000 1.1155 0.997 1.1153 0.997 0.677 1.1213 1.1221 1.001 Column4/Column3 2 x 2 cm 2 irregular on-axis beamlet 2 x 2 cm 2 irregular 8 cm off-axis 1.1203 0.999 1.1250 1.003 Temporal Delivery of IMRT Delivery of Dose to a Single Voxel IMRT Delivery Signatures Step & Shoot HIART 12000 10000 8000 Signal (fc) 6000 4000 2000 0 0.16 46.9 93.7 141 188 235 282 329 376 423 470 517 564 613 658 702 747 792 836 881 926 971 1015 1060 1105 1149 1194 1239-2000 Elapsed Time (sec) From Tim Holmes, St. Agnes Baltimore 4

Temporal Delivery of IMRT Delivery of Dose to a Single Voxel Normalized Cumulative Dose 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 5 % 90 % 5 % Step and Shoot Helical Tomo 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Elapsed Time (minutes) From Tim Holmes, St. Agnes Baltimore 0.9 0.8 Partial Volume Effect Dose to Water For Small Fields Output Factor 0.7 0.6 0.5 0.4 0.3 Monte Carlo Diamond Diode LAC+film Film PTW 31014 (Pinpoint) PTW 31010 (Semiflex) PTW 30006 (Farmer) 0.2 0.5 1.0 1.5 2.0 2.5 3.0 Side of Square Field (cm) From Roberto Capote, IAEA High Uncertainty in Output Factors Example: Statistics of 45 Output Factors for 6 mm and 18 mm square fields (Novalis, SSD = 1000 mm, depth = 50 mm, various detectors) Factor of Two in Beam Calibration! From Wolfgang Ullrich, BrainLab 5

Radiation Bills Raise Question of Supervision By WALT BOGDANICH and REBECCA R. RUIZ Published: February 25, 2010 Dose in 76 patients exceeded prescription by 50% Beam measurement error by physicist Ion chamber Detector too large blurs measured dose profile for small beams True dose profile Measured dose profile 50% underestimate in peak dose 50% overestimate in beam on time needed Reasons for Drop in Output with Small Field Size Backscatter into monitor unit from beam defining jaws Reduced scatter (phantom and head) Electronic disequilibrium Obscuration of the source Backscatter into Monitor Chamber The effect is due to backscattered photons entering the monitor and resulting in feedback to the linac to lower its output Varian 2100 10 MV. Results with other jaw completely open Liu et al., Med. Phys 2000;27:737-744 6

Problems with Measuring Conventional Output Factors Small field openings obscure the source which is difficult to measure and error prone. Amount of phantom scatter changes as well as lateral disequilibrium. Partial volume effects can mask machine output factor. Standard Reference Conditions Small Field Conditions Measured and MC Output Factor (Scp) as Function of Electron-Beam FWHM Point source assumption starts breaking down for small fields. Chamber Selection For Beams without Field Flattening Filters Dose flatness insufficient for Farmer-type chamber Cavity length should not be greater than 1 cm Option: cross calibrate a short chamber with Farmer-type chamber Kawachi et al, Med Phys (2008) 4591 7

Normalized Chamber Response Courtesy Jessica Snow and Larry DeWerd, UW ADCL 0.6 cc Courtesy Jessica Snow and Larry DeWerd, UW ADCL 0.057 cc Clear sign of charge multiplication Courtesy Jessica Snow and Larry DeWerd, UW ADCL 8

0.007 cc Courtesy Jessica Snow and Larry DeWerd, UW ADCL Which physical effects results in the greatest perturbation for small field dosimetry for ion chambers? Pick One: 26% 1. Charged particle disequilibrium 23% 26% 25% 2. Temporal non-constancy 3. Partial volume effects 4. Recombination/charge multiplication effects Which physical effects results in the greatest perturbation for small field dosimetry for ion chambers? Pick One: 1. Charged particle disequilibrium 2. Temporal non-constancy 3. Partial volume effects 4. Recombination/charge multiplication effects 9

True or False? Small field dosimetry is important for good quality assurance, but typical inaccuracies encountered can never put patients at grave harm. 1. True 50% 2. False 51% True or False? Small field dosimetry is important for good quality assurance, but typical inaccuracies encountered can never put patients at grave harm. 1. True 2. False 2. False: Output factors for the smallest fields are difficult to measure and can change the dose the patient receives by many 10s of percent IMRT Is Far from Reference Conditions 0.05 0.30 0.55 0.80 0.10 0.35 0.60 0.85 0.15 0.40 0.65 0.90 0.20 0.45 0.70 0.95 0.25 0.50 0.75 1.00 From Art Boyer, Stanford 10

Audit for TRS 398 Reference Dosimetry Relative Deviation (%) 10 8 6 4 2 0-2 -4-6 -8-10 CASE NUMBER Farmer Semiflex PinPoint 1 2 3 4 5 6 7 From Roberto Capote, IAEA IMRT Audit Relative deviation (%) 10 8 6 4 2 0-2 -4-6 -8-10 Farmer Semiflex PinPoint Step-and-shoot Dynamic 1 2 3 4 5 6 7 8 9 10 11 12 13 CASE NUMBER Sánchez-Doblado F, Hartmann G., Pena J., Capote R. et al IJROBP 68 (2007) 301-310 IAEA/AAPM Meeting for the Dosimetry Code of Practice: Small Fields and Novel Beams Outside Participants Jan Seuntjens Hugo Palmans Karen Rosser Saiful Huq Wolfgang Ullrich Warren Kilby Rock Mackie IAEA Pedro Andreo Ken R. Shortt Stanislav Vatnitskiy Roberto Capote Joanna Izewska Ahmed Meghzifene Rodolfo Alfonso Alfonso et al., A new formalism for reference dosimetry in small and non-standard fields, Med. Phys. 35: 5179 (2008) 11

Examples of Small and Novel Fields GammaKnife - 1.8 cm diameter collimator (1.6 cm on the Perfexon) is the largest collimator intrinsically composite field not 100 cm SSD Linac SRS beams - extrapolate to small field conditions Accuray - 6 cm diameter collimator is the largest collimator no field flattening filter TomoTherapy - 5 cm is the largest slice width no field flattening filter IMRT - made up of numerous small fields IAEA TRS-398 IAEA TRS-398 & TG-51 Nomenclature D = M N k k fmsr fmsr f msr,fref w,qmsr Qmsr D,w,Q0 Q,Q0 Q msr,q IAEA TG-51 D = M N k k Qmsr Co 60 w D,w Q Qmsr Overview of Static Field Dosimetry REFERENCE DOSIMETRY D = M N k k Qmsr Co 60 w D,w Q Qmsr 10 cm x 10 cm Reference Field Linac Stereotactic Field D = D Ω Q clin Q msr Q clin w w Qmsr N Co 60 D,w k Q k Qmsr Hypothetical 10 cm x 10 cm Reference Field BrainLab MiniMLC (10 cm x 10 cm) CyberKnife (6 cm Diameter) Q Ω clin Qmsr k Qmsr Ion Chamber = TomoTherapy (5 cm x 10 cm ) 12

Static Field Calibration Uses a machine-specific reference field, fmsr D = M N k k k Qmsr Co 60 w D,w Q Qmsr Qmsr Q (D/ M) w = (D/ M) Corrects for the differences between the conditions of field size, geometry, and beam quality of the conventional reference field and the machine-specific reference field. msr Q w Q msr D w fmsr D w Calculate Using MC Using method of Sempau et al 2004 PMB 49;4427-44 k Qmsr ( D / D ) w = ( D / D ) w Qmsr a is the dose at the reference point in water D a is the average dose in the ion chamber a Q D a D w Adapted from Edmond Sterpin Composite Field Calibration Uses a plan-class specific reference field, fpcsr D = M N k k Qpcsr Co 60 w D,w Q Qpcsr k Qpcsr = (D/ M) (D/ M) Qpcsr w Q w fpcsr Corrects for the differences between the conditions of field size, geometry, and beam quality of the conventional reference field and the plan-class specific reference field. 13

Effective kq for Tomo 1. Determine the %dd(10)x[ht Ref] for a tomotherapy unit for a 5 cm x 10 cm field at 85 cm SSD. 2. Using the graph at the right look up the value of %dd(10)x[ht TG-51]. 3. Using the graph on the left and the derived value %dd(10)x[ht TG-51] determine the abscissa and this effective kq value is then called k k Q Q msr in the IAEA protocol. Thomas et al (2005) Static and Composite Field Calculations for Tomo Static Field Calibration Composite Field Calibration (Section III.A.1) (Section III.A.2) Jeraj et al 2005 Duane et al 2006 k Qmsr 5 cm x 10 cm 0.997 Unmodulated Helical Delivery 5 cm Slice Width k Qpcsr 1.000 2 cm x 10 cm 0.993 Unmodulated Helical Delivery 2.5 cm Slice Width 1.000 2 cm x 2 cm 0.990 Unmodulated Helical Delivery 1 cm Slice Width 0.997 TomoTherapy: Static Field KQeff = KQ x KQmsr (Standard Uncertainty) Other References (10 cm x 5 cm) Exradin A1SL 0.997 (+- 0.001) Exradin A12 Farmer PTW-31014 Pinpoint PTW-31010 Semiflex PTW-31006 Farmer PTW-31018 (MicroLion) 1.000 (+-0.003) 0.992 (+- 0.003) Sterpin et al. (10 cm x5 cm) 0.996 (+-0.002) 1.001 (+-0.004) 0.993 (+-0.002) 0.996 0.995 (+-0.002) 0.995 0.997 (+-0.002) 0.993 (+-0.002) NE2571 0.995 0.997 (+-0.002) Sterpin et al. Med. Phys. 39 (7), July 2012 14

Two 6 MV Linacs: Nearly Same KQeff = KQ x KQmsr TomoTherapy CyberKnife Exradin A12 (0.65 cc) NE2571 (0.67 cc) Exradin A1SL (0.057 cc) 1.001 1.006 0.997 1.003 0.996 0.997 To within about a half percent, TomoTherapy and CyberKnife are like Co-60 units. Pantellis et al. Med. Phys. 39 (8), August 2012, 4875-4885 Which is more correct? The IAEA and the AAPM are: 25% 25% 25% 26% 1. Developing an entirely new dosimetry protocol to handle small fields. 2. Fundamentally modifying the TG-51 or IAEA protocols to be applicable for small fields. 3. Providing a methodology to obtain an appropriate calibration factor correction that is applicable for small fields. 4. Wringing their hands, unable to act. 15

Which is more correct? The IAEA and the AAPM are: 1. Developing an entirely new dosimetry protocol to handle small fields. 2. Fundamentally modifying the TG-51 or IAEA protocols to be applicable for small fields. 3. Providing a methodology to obtain an appropriate calibration factor correction that is applicable for small fields. 4. Wringing their hands, unable to act. Conclusions IMRT, SRS and SBRT uses complex field boundaries and/or one or many small circular fields. Partial volume effects can result in severe error in output factor measurements. Small chambers exhibit unusual charge collection behavior. IMRT deliveries are far from the measurement conditions of calibration. IAEA/AAPM has developed a formalism to account for small and novel beams in more realistic beam conditions. 16