1 x 5067B Instruction 1 x 9198B Sample Tube, Graduated (25 ml) w/ cap & blue dot, plastic 1 x R-0619B-C Hardness Buffer, 2 oz w/ blue cap, DB 1 x R-0620B-I Hardness Indicator Powder, 10 g w/ blue dot 1 x R-0683-C Hardness Reagent, 2 oz, DB TOTAL HARDNESS (1 drop = 10 ppm) 5. Multiply drops of R-0683 Hardness Reagent by 10. Record as parts per million (ppm) total hardness as calcium carbonate. Instr. #5067B NOTE: When dispensing reagents from dropper bottles, always hold bottle in a vertical position. Total Hardness Test 1. Rinse and fill 25 ml sample tube (#9198B) to 25 ml mark with water to be tested (Fig. 1). Fig. 1 2. Add 5 drops R-0619B Hardness Buffer. Swirl to mix. Fig. 2 3. Add 1 dipper R-0620B Hardness Indicator Powder. Swirl until dissolved. Sample will turn red (Fig. 2) if hardness is present. 4. Add R-0683 Hardness Reagent dropwise, swirling and counting after each drop, until color changes from red to blue (Fig. 3). 800-TEST KIT (837-8548) 410-472-4340 7/16 Fig. 3
1 x 5067W Instruction 1 x 9198W Sample Tube, Graduated (25 ml) w/ cap & white dot, plastic 1 x R-0638W-C Phenolphthalein Indicator, 2 oz w/ white cap, DB 1 x R-0699-C Iodide Iodate Reagent, 2 oz, DB 1 x R-0725-I Acid Starch Indicator Powder, 10 g NOTE: When dispensing reagents from dropper bottles, always hold bottle in a vertical position. Sodium Sulfite Test NOTE: Sample must be cooled to less than 100ºF (38ºC) to prevent high test results. Sample must be protected from air contact while cooling to prevent low test results. 1. Collect water to be tested in a clean, preferably large-mouthed, bottle to overflowing. Immediately cap and cool to room temperature. SODIUM SULFITE (1 drop = 10 ppm) 3. Add 1 drop R-0638W Phenolphthalein Indicator. Swirl to mix. Sample will turn pink (Fig. 2). 4. Add R-0725 Acid Starch Indicator Powder a dipper at a time, swirling after each dipper, until color changes from pink to colorless. Add 2 more dippers. Swirl until dissolved. 5. Add R-0699 Iodide Iodate Reagent dropwise, swirling and counting after each drop, until sample changes from colorless to a faint but permanent blue (Fig. 3). 6. Multiply drops of R-0699 Iodide Iodate Reagent by 10. Record as parts per million (ppm) sodium sulfite. NOTE: For 14.6 ml sample, record drops as grains per gallon (gpg) sodium sulfite. NOTE: For results as sulfite, multiply sulfite concentration by 0.64. Fig. 1 Fig. 2 Instr. #5067W 2. Rinse and fill 25 ml sample tube (#9198W) to 25 ml mark with cooled (room temperature) water to be tested (Fig. 1). NOTE: For results in grains per gallon (gpg), fill to 14.6 ml mark. 800-TEST KIT (837-8548) 410-472-4340 8/16 Fig. 3
1 x 5067pH Instruction 1 x 5425 Color Card Comparator/Instruction 1 x 9017 Test Cell, Calibrated (5 ml) w/ cap, plastic 1 x R-1003U-C-DB Long Range Indicator, 2 oz, DB COLOR COMPARISON TEST ph (LONG RANGE) (3.0-11.0) Instr. #5067pH NOTE: When dispensing reagents from dropper bottles, always hold bottle in a vertical position. ph (long range) Test 1. Rinse and fill 5 ml test cell (#9017) to 5 ml mark with water to be tested. 2. Add 5 drops R-1003U Long Range Indicator. Cap and mix. 3. Place test cell on white area of Color Card Comparator (#5425). Match color in test cell with a color standard. Record as ph units. 800-TEST KIT (837-8548) 410-472-4340 8/16
2 x 2269 Cap, Dispenser, 18 mm, white 1 x 4027 Funnel, 58 mm, plastic 1 x 5067OP Instruction 1 x 6003 Brush, Test Tube 1 x 6009 Filter Paper, 9.0 cm, #610, 100/box 1 x 9021 Test Tube, Mixing, Calibrated (5, 10, 14, 15, 17.5 ml), glass w/ stopper 1 x 9025 2-Standard Comparator, Orthophosphate, Stannous Chloride, 30 & 60 ppm 2 x R-0601-C Molybdate Reagent*, 2 oz 1 x R-0602P-I Stannous Chloride Powder, 10 g Orthophosphate Test 1. Filter water to be tested to clarify. COLOR COMPARISON TEST ORTHOPHOSPHATE (30 & 60 ppm) Instr. #5067OP 6. Match color in test tube with a color standard. Record as parts per million (ppm) orthophosphate. *WARNING: Molybdate Reagent (R-0601) contains 5-10% (w/w) sulfuric acid, a corrosive acid. 2. Rinse and fill test tube (#9021) to 5 ml mark with filtered sample. 3. Cap R-0601 Molybdate Reagent with dispenser cap (#2269). Add R-0601 Molybdate Reagent to 15 ml mark. Stopper and mix. Remove stopper. 4. Add 2 level dippers R-0602P Stannous Chloride Powder. Stopper and mix. 5. Wipe dry and place in comparator. WAIT 1 MINUTE. Read before 5 minutes. 800-TEST KIT (837-8548) 410-472-4340 9/16
1 x 5067O Instruction 1 x 9198O Sample Tube, Graduated (25 ml) w/ cap & orange dot, plastic 1 x R-0630-C Chromate Indicator, 2 oz, DB 1 x R-0638O-A Phenolphthalein Indicator,.75 oz w/ orange cap, DB 1 x R-0687O-C Sulfuric Acid.12N, 2 oz w/ orange cap, DB 1 x R-0706-C Silver Nitrate Reagent, 2 oz, DB CHLORIDE (1 drop = 10 ppm) 3. Add 5 drops R-0630 Chromate Indicator. Swirl to mix. Sample will turn yellow (Fig. 2). 4. Add R-0706 Silver Nitrate Reagent dropwise, swirling and counting after each drop, until color changes from yellow to milky salmon (brick) red (Fig. 3). Instr. #5067O NOTE: When dispensing reagents from dropper bottles, always hold bottle in a vertical position. Chloride Test NOTE: When sulfite content of sample water to be tested exceeds 10 ppm, the sulfite should be oxidized to prevent interference in test. A 25 ml water sample is first adjusted to the appropriate ph, then 1 ml (or 25 drops) of R-0649 Hydrogen Peroxide Solution (sold separately) is added and thoroughly mixed. Continue with the rest of the procedure. NOTE: Do not add enough R-0706 Silver Nitrate Reagent to give a brown color. First change from yellow to a milky salmon (brick) red is the endpoint. 5. Multiply drops of R-0706 Silver Nitrate Reagent by 10. Record as parts per million (ppm) chloride. Fig. 1 Fig. 2 1. Rinse and fill 25 ml sample tube (#9198O) to 25 ml mark with water to be tested. 2. Add 2 drops R-0638O Phenolphthalein Indicator. Swirl to mix. If sample turns pink (Fig. 1), add R-0687O Sulfuric Acid.12N dropwise, swirling after each drop, until color changes from pink to colorless. 800-TEST KIT (837-8548) 410-472-4340 8/16 Fig. 3
1 x 5067G Instruction 1 x 9198G Sample Tube, Graduated (25 ml) w/ cap & green dot, plastic 1 x R-0637-C Methyl Orange Indicator, 2 oz, DB 1 x R-0638G-A Phenolphthalein Indicator,.75 oz w/ green cap, DB 1 x R-0645-C Total Alkalinity Indicator, 2 oz, DB 1 x R-0687G-C Sulfuric Acid.12N, 2 oz w/ green cap, DB P/M & P/T ALKALINITY (1 drop = 10 ppm) 4. Add 5 drops R-0637 Methyl Orange Indicator. Swirl to mix. Sample will turn yellow (Fig. 2). 5. Add R-0687G Sulfuric Acid.12N dropwise, swirling and counting after each drop, until color changes from yellow to orange (salmon pink) (Fig. 3). Record total drops (Steps 3 & 5) as M reading. Instr. #5067G NOTE: When dispensing reagents from dropper bottles, always hold bottle in a vertical position. P/M Alkalinity Test 1. Rinse and fill 25 ml sample tube (#9198G) to 25 ml mark with water to be tested. NOTE: For results in grains per gallon (gpg), fill to 14.6 ml mark. 6. Multiply P reading by 10. Record as parts per million (ppm) P alkalinity as calcium carbonate. Multiply M reading by 10. Record as ppm M alkalinity as calcium carbonate. NOTE: For 14.6 ml sample, record P reading as grains per gallon (gpg) P alkalinity as calcium carbonate. Record T reading as gpg T alkalinity as calcium carbonate. Fig. 1 Fig. 2 2. Add 3 drops R-0638G Phenolphthalein Indicator. Swirl to mix. Sample will turn pink (Fig. 1) if P alkalinity is present proceed to Step 3. If sample is colorless, proceed to Step 4. 3. Add R-0687G Sulfuric Acid.12N dropwise, swirling and counting after each drop, until color just changes from pink to colorless. Record drops as P reading. (OVER) Fig. 3
P/T Alkalinity Test 1. Rinse and fill 25 ml sample tube (#9198G) to 25 ml mark with water to be tested. P/M & P/T ALKALINITY (1 drop = 10 ppm) Instr. #5067G NOTE: For results in grains per gallon (gpg), fill to 14.6 ml mark. 2. Add 3 drops R-0638G Phenolphthalein Indicator. Swirl to mix. Sample will turn pink (Fig. 4) if P alkalinity is present proceed to Step 3. If sample is colorless, proceed to Step 4. 3. Add R-0687G Sulfuric Acid.12N dropwise, swirling and counting after each drop, until color just changes from pink to colorless. Record drops as P reading. Fig. 4 4. Add 5 drops R-0645 Total Alkalinity Indicator. Swirl to mix. Sample will turn green (Fig. 5). 5. Continue adding R-0687G Sulfuric Acid.12N dropwise, swirling and counting after each drop, until color changes from green to red (Fig. 6). Record total drops (Steps 3 & 5) as T reading. Fig. 5 6. Multiply P reading by 10. Record as parts per million (ppm) P alkalinity as calcium carbonate. Multiply T reading by 10. Record as ppm T alkalinity as calcium carbonate. NOTE: For 14.6 ml sample, record P reading as grains per gallon (gpg) P alkalinity as calcium carbonate. Record T reading as gpg T alkalinity as calcium carbonate. 800-TEST KIT (837-8548) 410-472-4340 8/16 Fig. 6
Instruction Number: 5425 Component Description > COLOR CARD TEST > LONG RANGE ph (3 11) 1. Rinse and fill 5 ml test cell (#9017) to 5 ml mark with water to be tested. 2. Add 5 drops R 1003U Long Range Indicator. Cap and mix. 3. Lay test cell flat along white area of Color Card (#5425). Match color in test cell with a color standard. Record as ph units. 2/00 2016 Taylor Technologies, Inc. All Rights Reserved View Privacy Policy