Mass Spectrometry Laboratory AIMS Instrumentation & Sample Report Documentation AB Sciex QStar XL chemistry UNIVERSITY OF TORONTO AIMS Mass Spectrometry Laboratory Department of Chemistry, University of Toronto 80 St. George Street, Toronto ON, M5S 3H6
QStar XL Q-TOF MS Manufacturer: AB Sciex, Concord, ON Features Mass Analyzer: Ion Source: Source Interface: Ion Polarity: Mass range: Resolution: Mass Accuracy: HPLC: Quadrupole time- of- flight (Q- TOF) Ionspray source (ESI) HSID interface (IONICS, Bolton, ON) Positive and negative 20-40 000 Da Up to 8 000 mass resolution +/- ~2.5 mda Agilent 1100 Series HPLC QStar XL Sample Results Format Low Resolution ESI Spectrum The low resolution QStar XL ESI- MS report consists of a profile mass spectrum showing m/z versus intensity. Spectral peaks are labeled with m/z values rounded to one decimal place. The standard data acquisition range is m/z 90-1000 Da but printed mass spectra are normally expanded to display an appropriate region of interest. Accurate Mass Report The QStar XL accurate mass report is generated using the Elemental Composition Calculator as implemented in the Analyst QS software package. The report provides the high resolution mass spectrometry (HRMS) value and a list of possible molecular formulae with calculated exact m/z values. All of the values contained in the report correspond to the ionic species of interest and the given ionic formula includes the charging agent e.g. [M+H + ]+. Therefore, both the measured and calculated values are corrected for the mass of the electron (0.00054 Da) and should be reported as m/z. Suggested reporting format for compound characterization: HRMS (ESI- QTOF+): m/z [M+H + ]+ calc d for C21H35N10O5+ 507.2786 Da, measured 507.2793 Da.
Sample Name: mb-2-22 * ESI - QStar XL Acq. Date: Thursday, July 11, 2013 Acq. Time: 10:15 +TOF MS: 1.300 to 1.367 min from 130711_0874.wiff a=3.56624206142460140e-004, t0=5.28843480409968980e+001 R;, subtracted (0.783 to 0.867 min) Max. 370.4 counts. 370 360 507.3 340 320 300 280 260 240 220 Intensity, counts 200 180 160 140 120 100 80 60 40 254.1 20 98.0 156.0 174.0 264.9 285.0 372.2 415.2 529.3 561.2 0 100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000 m/z, amu Acq. File: 130711_0874.wiff Polarity/Scan Type: Positive
* ESI - QStar XL Elemental composition calculator Target m/z: +507.2793 amu Tolerance: +10.0000 ppm Result type: Elemental Max num of results: 100 Min DBE: -0.5000 Max DBE: +50.0000 Electron state: Even Num of charges: 1 Add water: N/A Add proton: File Name: N/A 130711_0874.wiff Elements Min Number Max Number 1 2 3 4 5 6 7 8 9 10 11 12 B 0 0 C 0 500 Cl 0 0 F 0 0 Fe 0 0 H 0 1000 K 0 0 N 0 12 Na 0 0 O 0 12 P 0 0 S 0 0 Formula Calculated m/z (amu) mda Error PPM Error DBE 1 2 3 4 5 6 7 8 C37 H35 N2 507.2794-0.1-0.3 21.5 C21 H35 N10 O5 507.2786 0.6 1.2 9.5 C24 H43 O11 507.2799-0.6-1.3 3.5 C20 H39 N6 O9 507.2773 1.9 3.9 4.5 C25 H39 N4 O7 507.2813-2.0-3.9 8.5 C26 H35 N8 O3 507.2826-3.3-6.6 13.5 C32 H35 N4 O2 507.2754 3.8 7.5 17.5 C16 H35 N12 O7 507.2746 4.6 9.2 5.5
QStar XL Biomolecule Compound Report The protein compound report summarizes results from ESI- MS analyses of intact proteins or peptides for the purpose of molecular weight determination. The raw mass spectral data is processed using the Protein or Peptide Reconstruct algorithms as implemented in the BioAnalyst v1.1.5 software package. The report is composed of three sections: (1) Mass Spectrum Panel (1) shows the averaged and background- subtracted mass spectrum in a plot of m/z versus intensity. Spectral peaks are labeled with m/z values rounded to one decimal place. The standard m/z acquisition range for biomolecular analysis is 200-2000 Da; however, the printed mass spectrum is often expanded to highlight the appropriate region of interest. (2) Mass List Panel (2) contains the compound mass list generated by the Protein or Peptide Reconstruct algorithms and includes the following columns: Mass (Avg) the centroid average neutral mass in Da Mass (mono) the monoisotopic neutral mass in Da (peptides only) Apex Mass the apex neutral mass in Da Area the peak area Start Scan start of peak Stop Scan end of peak Score reconstruction fit score Evidence C indicates compound molecular weight determined via identification of consecutive charge states I indicates compound molecular weight determined via identification of charge state from a resolved isotopic envelope (3) Reconstructed Mass Spectrum Panel (3) shows the reconstructed mass spectrum generated by the Protein or Peptide Reconstruction algorithms in a plot of neutral mass (Da) versus intensity. Spectral peaks are labeled with mass values rounded to one decimal place. Mass Accuracy The mass accuracy associated with protein molecular weight determinations varies according to several factors including molecular weight, charge states observed, signal intensity and signal complexity (e.g. adducts, multiple proteins, etc.). The rule of thumb for a relative measure of protein mass accuracy is +/- 0.01% of the protein molecular weight; however, in practice the observed mass accuracy is typically less than 1 Da for proteins up to ~30 kda.
QStar XL Protein Compound Report Sample Name: L197E 10x * ESI - QStar XL Acq. Date: Thursday, March 07, 2013 Acq. Time: 10:16 +TOF MS: 4.934 to 5.184 min from Sample 1 (L197E 10x) of 130307_... a=3.56589330169289460e-004, t0=5.19759967633031080e+001, subt... Max. 24.8 counts. Intensity, counts 24.8 20.0 15.0 10.0 5.0 788.3 767.6 810.2 747.9 729.3 833.3 883.7 940.7 972.0 711.5 1005.5 914.4 1121.4 694.6 1166.2 0.0 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800 m/z, amu Mass reconstruction of +TOF MS: 4.934 to 5.184 min from Sample 1 (L1... Max. 739.0 cps. 700 29130.0 600 Intensity, cps 500 400 300 200 100 0 29114.0 29229.0 29187.0 29092.0 29326.0 2.86e4 2.88e4 2.90e4 2.92e4 2.94e4 2.96e4 2.98e4 3.00e4 Mass, amu Acq. File: 130307_CHM379_Intact.wiff Polarity/Scan Type: Positive