Technical Topics. Where did my gain go? Thin Tube Open Fit BTE Verification. History of Thin Tubes: Summary of presentation by C. Staples and S.

Similar documents
Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial

AccuQuest Spotlight: Successful Fittings with Oasis. Fitting Range

Building Practice Success Through Speechmap Real Ear Measurements. Chris Stokes-Rees

A comparison of manufacturer-specific prescriptive procedures for infants

Validation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method

Mistakes on Real Ear Measures Clinicians Often Make. Michael Valente

Manufacturers NAL-NL2 Fittings Fail Real-ear Verification

The Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2

pep Step up with pep Pep Specification Guide _Pep_Specification_Guide_UK.indd 1

Audiology Today MarApr2011

Sonic Spotlight. The Acoustic Flexibility of Sonic BTEs and minibtes: A Size, Sound and Set-up for Every Patient

Clinical Tips for Matching Real-Ear Prescriptive Targets

The REM Cookbook How to Correctly Perform Real Ear Measurements

Prescribe hearing aids to:

Feelin The Squeeze? 10/17/2016. Any of these been on your mind lately? Key Part of NAS 12 Recommendations

RECD-Measurements with Connexx 6.3 and Unity 2

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies

Phonak Baseo Q (Q15) Product information

Paediatric Amplification

Top 10 ideer til en god høreapparat tilpasning. Astrid Haastrup, Audiologist GN ReSound

ChildFit. Widex Baby. Compass quick guide

New CTA PSAP Quality Standard (Starting with Some Interesting Questions)

Using digital hearing aids to visualize real-life effects of signal processing

FONIX FP35. Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Visible Speech

The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February :50

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04

Optimising Outcomes for Speech Mapping

Industry Emphasis. Hearing Aids. Basic Design and Function. Hearing Aid Types. Hearing Aid Types. Hearing Aid Types.

DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015.

Step-by-Step RECD Guide

Audiological Bulletin no. 46

WIDEXPRESS THE WIDEX FITTING RATIONALE FOR EVOKE MARCH 2018 ISSUE NO. 38

personalization meets innov ation

Insio: A new standard in custom instruments

How to use AutoFit (IMC2) How to use AutoFit (IMC2)

Verifying the Lyric Audibility Advantage

Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit

Digital. hearing instruments have burst on the

Fitting Decisions and their Impact on Hearing Aid User Benefit. Mallory Maine, AuD Audiologist, GN ReSound

An assessment of the validity of a subjective hearing aid verification procedure

Flip Specification Guide. flip your expectations

Re/Habilitation of the Hearing Impaired. Better Hearing Philippines Inc.

Elements of Effective Hearing Aid Performance (2004) Edgar Villchur Feb 2004 HearingOnline

Hearing Aids and Assistive Listening Devices

Quick Guide Binaural REM

DREAM feature and product Quick guide

Otolens Invisible In Canal Hearing Aids

Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: /jaaa

Product Portfolio Supplement

Contents #

The Situational Hearing Aid Response Profile (SHARP), version 7 BOYS TOWN NATIONAL RESEARCH HOSPITAL. 555 N. 30th St. Omaha, Nebraska 68131

User groups Most sensorineural, conductive and mixed hearing losses in the categories of severe and profound.

Verification of soft speech amplification in hearing aid fitting: A comparison of methods

Supplementary Online Content

2/16/2012. Fitting Current Amplification Technology on Infants and Children. Preselection Issues & Procedures

AND9020/D. Adaptive Feedback Cancellation 3 from ON Semiconductor APPLICATION NOTE INTRODUCTION

Hearing Loss & Hearing Assistance Technologies

Single channel noise reduction in hearing aids

ipod Noise Exposure Assessment in Simulated Environmental Conditions

Candidacy and Verification of Oticon Speech Rescue TM technology

Why? Speech in Noise + Hearing Aids = Problems in Noise. Recall: Two things we must do for hearing loss: Directional Mics & Digital Noise Reduction

Technical Data. Orion RIC. usa.siemens.com/hearing. Life sounds brilliant. Description. Standard Features

How to select the optimum BTE fitting option

Adaptive Feedback Cancellation for the RHYTHM R3920 from ON Semiconductor

This Page Intentionally Left Blank

Using the FONIX 7000 to Verify Coupler and Real-Ear Performance to Adhere to the AAA Guidelines for the Audiologic Management

Max. Powered by Era. Hearing matters

binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way

Sonic Spotlight. Frequency Transfer Providing Audibility For High-Frequency Losses

Testing FM Systems on the 7000 Hearing Aid Test System

Abstract.

Best practice protocol

THE OCCLUSION EFFECT: PHYSICAL MEASUREMENT AND SELF-REPORT. Karen Jarvis Advisor: Robyn Cox, Ph.D. April 2004

The WIDEX DREAM family has just got even better.

CROS System Initial Fit Protocol

Product information. For more information visit

Opening the door to ultimate clarity, performance and convenience

Hearing. istockphoto/thinkstock

Beltone Electronics 2601 Patriot Boulevard Glenview, IL U.S.A. (800)

DEVELOPMENT OF NAL-NL2

Hearing Solutions Catalog Winter Supplement January 2009

Electro-Acoustic Stimulation (EAS) Naída CI Q90 EAS System. Naída CI Q90 EAS System Components. Naída CI Q90 Acoustic Earhook

In the days before programmable hearing

how we hear. Better understanding of hearing loss The diagram above illustrates the steps involved.

Best Practice Protocols

Testing Digital Hearing Aids

ASR. ISSN / Audiol Speech Res 2017;13(3): / RESEARCH PAPER

Hearing Solutions Catalog

Make the world louder!

Counseling Patients on Fitting Options. Michael Valente

Clinical fitting guide

DSM PRO. Software Training Manual. Copyright November 2003

Differences in added stable gain between manufacturers, audiometric configurations, earmold styles, and frequency bands

Digital hearing aids are still

Now, Delta can fit 4 out of 5 people!

FALL Product Portfolio

Introducing ReSound LiNX Quattro Unprecedented Layers of Sound and the world s most advanced rechargeable hearing aid solution.

International Journal of Health Sciences and Research ISSN:

Phonak Naída Q. Product information

Transcription:

Technical Topics Where did my gain go? Thin Tube Open Fit BTE Verification Summary of presentation by C. Staples and S. Aiken History of Thin Tubes: Growth of new BTE market: First quarter of 2006 market: 42% BTE instruments Second quarter of 2006 market: 46% BTE hearing instruments 17% of hearing instruments sold were Thin Tube Open Fit BTE hearing instruments U.S. hearing aid sales by type (%) 70 60 50 40 30 Total number of Thin Tube Open Fit BTE hearing instruments available in Canada 38 (as of October 1, 2006) - 12 designated Open Fit - 26 hybrid design Reasons for the growth Minimal to no occlusion Cosmetically appealing - BTE is small - Thin Tube blends well with skin and hair - Fashionable designs Instant fits - Simplified process for Hearing Care Professional and customer 20 10 0 2001 2002 2003 2004 2005 2006 CIC Other Custom BTE U.S. hearing and sales by type (total net units), indicating growth in BTE market share from 23% in 2001 to nearly 45% in the first quarter of 2006. Source: Hearing Industries Assn. Concerns with Thin Tube Open Fit BTEs Limited fitting range Potential for feedback with increased hearing losses Uncertainty of extended fitting ranges

Acoustic effects of Open Fit Thin Tube BTEs The vent-out effect The vent is a two-way street; sound travels in through the vent and travels out the vent as well. The larger the vent the greater the possibility for more sound to escape through the vent More gain is needed to account for vent loss (below 0 Hz) 20 10 0-10 -20-30 db Gain (Open Ear) Vent loss -40 0 00 Thin Tube effect The thin tube (because of its smaller diameter) reduces the resonance peaks that would normally be present in standard tubing. This creates a situation where there is more electronic gain required by the hearing aid to meet the same output More gain is needed to reach the same output (above 0 Hz) db SPL (Coupler) 90 2cc output with regular tube 85 80 75 70 65 2cc output with.9mm thin tube 60 0 00 page 2 Amplified sound and vent-in interactions Open Fit BTEs provide a pleasing mix of natural and amplified sound; however at similar levels these sounds do not interact well Where the amplified sound is dominant (usually high frequencies) only the amplified sound is heard Where the vent-in sound is dominant (usually the low frequencies) only the vent-in sound is heard When the amplified sound and vent-in sound are at the similar levels, a conflict occurs. The combination of these two sounds at similar levels can lead to a doubling or complete cancellation of sound More/less gain can be needed to account for phase interactions 90 80 70 60 50 40 db SPL (Real Ear) 0 00 Doubling and cancelling occur at frequencies where the vent-in and amplified sounds interact at similar levels; a doubling is seen when in phase and a cancelling of sound is seen when out of phase.

Verification Recommendations Reference microphone contamination Due to the large vent with Open Fit Thin Tube BTEs, sound can escape out of the canal and possibly contaminate the input to the reference microphone - If contamination occurs, the reference microphone measures an increase in sound volume and the hearing aid analyzer reduces the level of the test stimulus This might result in an underestimation of hearing aid output. The result of not taking this into account could lead to over amplification Underestimation of output due to reference microphone contamination True Response Contaminated Response True Response Contaminated Response The solution The contamination of the reference microphone can be avoided by muting the hearing aid during the equalization process. Upon completion of the equalization process, turn the hearing microphone on and proceed with the hearing aid analysis. It is critical that the patient does not move during this process. Should the patient move restart the fitting/equalization process page 3

Real Ear Verification Always verify Open Fit Thin Tube BTEs on the Real Ear - Coupler based measurements do not account for venting effects Green = Real Ear Measurement (correct) Purple = Simulated Real Ear Measurement with HA2 coupler (incorrect) Simulated REM REM Simulated REM is not appropriate for Open Fit BTEs because the vent cannot be evaluated accurately Ensure adaptive feedback cancellation system is active and is not reducing gain to control for feedback STOP! Summary Know what you are verifying - Since digital hearing aids delay the signal, amplified and vent-in sounds will likely be in-phase and out-of-phase at different frequencies. When the vent-in sound and amplified sound are approximately the same level, this creates a peaky REAR. If this is a problem, try minimizing or shifting the area of interaction by varying the hearing aid settings - Check final REAR to ensure that response is appropriate - Preferred method of verification is Real Ear SPL with speech or speech-like stimulus page 4

Thin Tube Verification Study Method - 14 Hearing Aids from 9 manufacturers - Set Client Experience level to maximum - Set to manufacturer recommended, and NAL-NL1, DSL [i/o] if available - Feedback Manager ran if recommended by manufacturer - Verified on the Audioscan Verifit System - Real Ear Verification completed as suggested in Verifit Help menu - REM calculated using Standard Speech sentence #1 at input levels of 55 and 70dB SPL - Estimated Speech Intelligibility Index (SII) was collected for each input level and converted to a predicted percent intelligibility for Connected Speech Test dbspl 45 DSL[i/o] NAL-NL1 0 00 Manufacturer Recommended First Fit compared to NAL-NL1 and /DSL[i/o] NAL-NL1 dbspl 45 0 00 NAL-NL1 First Fit compared to NAL-NL1 DSL[i/o] dbspl page 5 45 0 00 DSL[i/o] First Fit compared to DSL[i/o]

Scores for Connected Speech Test vs SII Key Messages - Speech Intelligibilty Index converted to a predicted percent intelligibility for Connect Speech test Significant decrease in predicted intelligibility when SII is reduced from 45 to 25 90% Predicted CST score 40% Relationship between SII and CST 25 45 SII Value CST/SII Score for Loud Speech (70 db) - 98% was the predicted score of the best Thin Tube Open Fit BTE performer - 85% was the predicted score of the worst Thin Tube Open Fit BTE performer - All Thin Tube Open Fit BTEs performed quite well with Loud Speech (Best) 98% (Worst) 85% Predicted CST score Range of SII/CST values with 70 db speech input 41 67 SII Value CST/SII Score for Soft Speech (55 db) - 88% was the predicted score of the best Thin Tube Open Fit BTE performer - The 88% score for Soft Speech was better than the worst performer for Loud Speech - 8% was the predicted score of the worst Thin Tube Open Fit BTE performer (Best) 88% Predicted CST score Range of SII/CST values with 55 db speech input Ironically, the hearing aid with a predicted score of 8% costs three times more than the best Open Fit BTE performer (Worst) 8% page 6 13 43 SII Value