Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Similar documents
High-Dose Standardization Study For γ-rays radiation processing at NIM

Process Control Methods in Radiation Technologies

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

MAY NOT BE SAME AS THE PUBLISHED VERSION

Supplementary comparison CCRI(I)-S2 of standards for absorbed dose to water in 60 Co gamma radiation at radiation processing dose levels

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

GEX Technical Information Report

IAEA RADIATION TECHNOLOGY SERIES No. 3. Guidelines for development, validation and routine control of industrial radiation processes

INTERNATIONAL STANDARD

High Precision Dose Delivery from Electron & X-ray Beam Lines

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

Is IQ/OQ/PQ part of irradiation process control?

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

NPL REPORT CIRM 29. Guidelines for the Calibration of Routine Dosimetry Systems for use in Radiation Processing. Peter Sharpe and Arne Miller

World-Wide Radiation Metrology: The BIPM, the CIPM MRA, and ICRU

Radiation Dosimeters for Foods. Ashish Anand Department of Biological and Agricultural Engineering Texas A&M University

DOSIMETRY PRINCIPLES, DOSE MEASUREMENTS AND RADIATION PROTECTION

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

2015 APMP Meeting TCRI Report

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects

Y FILMS DOSIMETR Nederland België / Belgique

DOSIMETRY PRODUCTS & SERVICES 2018 PRICE LIST

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Good manufacturing practices (GMP) utilized on human blood irradiation process

A Method for Statistical Process Control of Radiation Sterilization Facilities

Recent advances in dosimetry in reference conditions for proton and light-ion beams

B3 RADIOCHROMIC FILM DOSIMETRY

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

Progress Report on Radiation Dosimetry at NPL

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Standardization of Nuclear Component Off-Carrier Processing. ASTM E61 New Orleans, LA January 27, 2015 Ryan Tracy, STERIS Isomedix Services

REVISED CODEX GENERAL STANDARD FOR IRRADIATED FOODS CODEX STAN , REV

IAEA RADIATION TECHNOLOGY SERIES No. 4. Guidelines for the Development, Validation and Routine Control of Industrial Radiation Processes

The use of polyvinyl chloride films dyed with methyl red in radiation dosimetry

The APLMF Guide. Guide to the application of ionizing radiation metrology in. medicine under legal metrological system

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

Accreditation of Dosimetry Services

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

CHEMICAL DOSIMETRY OF GAMMACELL WITH FERROUS SULFATE

IAEA-TECDOC-1564 Intercomparison of Personal Dose Equivalent Measurements by Active Personal Dosimeters

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

CIRMS: A Retrospective & Look Forward

Strengthening Activity Measurement Quality in Radiation Protection from Metrological Science to Reliable End-user Application

REPORT TO THE CCRI SECTION I ON THE ACTIVITY CARRIED OUT AT ENEA-INMRI ON PHOTON AND CHARGED PARTICLE DOSIMETRY IN THE PERIOD MAY MAY 2003

Highlights from TC-IR

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

ISO INTERNATIONAL STANDARD

Dosimetry and QA of proton and heavier ion beams

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Medical Physics 4 I3 Radiation in Medicine

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry

Application(s) of Alanine

DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Properties of Commercial PVC Films with Respect to Electron Dosimetry

Activity report from JCOG physics group

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Publishable JRP Summary Report for JRP HLT11 MetroMRT Metrology for Molecular Radiotherapy Background

Chapter 4: Radiation Monitoring Instruments

THE IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES SSDL NETWORK CHARTER

The determination of timer error and its role in the administration of specified doses

Radiolysis-induced Dosimetric Characterization of Aqueous Solutions of a Synthetic Dye for Gamma Dosimetry

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

First National Intercomparison of Individual Dosimetry of Photon Radiation in Ukraine

Feng Xiujuan National Institute of Metrology (NIM),China

Neutral Density Filter OD Measurements

CPGAN #002. FTIR Quantification of Absorbed Radiation Dose in Polyethylene

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS

PREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY

Applications of Monte Carlo simulations to radiation dosimetry

Hampton University Proton Therapy Institute

Zakithi Msimang SAAPMB/SARPA 2011

MONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

IORT with mobile linacs: the Italian experience

Progress Report Department 'Dosimetry for Radiation Therapy'

Technical Reports Series No.409. Dosimetry for Food Irradiation

SSDL Newsletter. Contents. From the editor. No. 51 September IAEA/WHO Network of Secondary Standards Dosimetry Laboratories

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

Radiation Monitoring Instruments

Transcription:

Dosimetry Standards and Dissemination Systems for Radiation Processing in China Zhang Yanli National Institute of Metrology, China April 2017 in Vienna, Austria

Organization Founded in 1955, National Institute of Metrology (NIM) is the national metrology institute of China, subordinate to General Administration of Quality Supervision, Inspection and Quarantine (AQSIQ), China. It is the nation s highest technical authority in the field of metrology, and responsible for providing metrology services of the highest accuracy and reliability to customers. 2

NIM s main tasks Research, establish, maintain and improve national primary standards & national standards Participate in international comparisons to ensure traceability and international equivalence of national standards; Disseminate values to its clients through verification and calibration services, ensure all unit values are traceable and reliable within the nation; Carry out fundamental and frontier research in metrological science, study measurement theory and technique; develop value disseminating and tracing methods, etc. Div. of Metrology in Ionizing radiation : Three section: Radioactivity, Dosimetry, Accelerator and Neutron Primary standards: 18 ( 1 for radiation processing) Secondary standards: 15 ( 4 for radiation processing) Comparison: 34 ( 1 for radiation processing ) CMCs:195 ( 8 for radiation processing ) 3

Dosimetry system for high-dose In 1983, a project of high dose standardization was initiated in China. To establish the high-dose standards system for γ-rays and EB irradiation. To establish the dissemination of high-dose value through intercomparison, calibration and verification. Implementation of NDAS using alanine/epr dosimetry system. To Calibrate routing dosimeters and measure radiation field dose mapping of radiation processing facility. To train the technical personal in dosimetry for irradiation facilities and to help them to set up their own routine dosimetry system. To enact administrative regulation and technical standards concerned absorbed dose for radiation processing. 4

Dosimetry system for high-dose In 1986, the high-dsoe standards and dissemination system has been established at NIM. Primary standard : Fricke liquid chemical dosimeter Transfer standards: Ceric-cerous sulfate liquid chemical dosimeters Dichromate liquid chemical dosimeters Alnine/EPR dosimeter Fricke dosimeter Silver Dichromate dosimeter Potassium Dichromate dosimeter Alanine/EPR Dosimeter Absorbed dose range 40~400 Gy 0.4~5 kgy 5~40 kgy 0.1~40 kgy Uncertainty at k=2 2.0% 4% 4% 4% 5

Dosimetry system for high-dose liquid chemical dosimeter Standards at NIM A type of Cary 4E Uv/vis spectrophotometer has been used for measuring optical absorbance of dosimetric solution, the preirradiation absorbance : Ao and the post-irradiation absorbance: Ai. ΔA = Ai Ao Cary 4E Uv/vis spectrophotometer Liquid chemical dosimeters 6

Dosimetry system for high-dose The Fricke dosimetry system as primary standard to measure the absorbed dose in water of 60 Coγrays in processing level with uncertainty of less than 2% (k=2) The absorbed doses are calculated as follows : D W = 1.004 D F =1.004 K F ΔΑ (1) K F =1 / ε G l ρ = 279 (GyAbs) (2) ---K is the dose-response conversion factor. ---G is radiation chemical yield (=1.61 10-6 mol J -1, ICRU 34# report) ---εis molar linear absorption coefficient of Fe 3+ (=219.2±0.3 m 2 mol ) --- l is optical pathlength of the dosimetric solution in the cuvette (=10.0mm) ---ρ is density of the dosimetric solution, (=1.024 g/cm 3 ) 7

Alanine/EPR dosimetry system It has been developed as a transfer standard used for NDAS program similar to IDAS of IAEA, and used to unify quantity of absorbed dose for irradiation at radiation processing dose level with uncertainty of less than 4% (k=2). Dosimetry system for high-dose alanine/epr dosimetry system at NIM

Dosimetry system for high-dose Routine (working) dosimeters The Fricke, potassium/silver dichromate, several kinds of radiochromic films (RCD), cellulose triacetate (CTA) and PMMA dosimeters have been recommended as routine dosimeters for radiation processing in China. 9

Dosimetry system for high-dose The schema of dosimetry tractability to national standards for radiation processing in china 10

γ-rays reference irradiated facility Dosimetry system for high-dose The pool 60 Co source with an activity of 770 TBq on Oct 1st, 2016 opened beam and barrel source The source was calibrated in terms of absorbed dose to water by a set of Fricke dosimeter. The K of transfer standard and routing dosimeter are calibrated through the Fricke dosimeter standard at the dose rate of about 2 kgyh -1 in this irradiation filed, respectively. 60 Coγrays high-dose irradiation facility at NIM 11

EB reference irradiated facility Dosimetry system for high-dose Nominal energy of EB: 9 10 11 12 MeV Maximum power of electron beam: 10 kw Repetition rate of EB current pulses: 50, 100, 150, 300 1/s Scanning length in 100 cm from extraction window : 60cm Speed of conveyor system: 10-200mm/s Minimum absorbed dose (50 Hz, 200mm/s ): 40 Gy linear electron accelerator conveyor system Graphite and polystyrene calorimeter 12

The CCRI(I)-S2 comparison In 1998 and 2008, NIM s Fricke dosimeter participated in CCRI (I)-S2 comparisons of standards for absorbed dose to water in 60 Co γ-rays at radiation processing dose levels using alanine/epr dosimeters of the NIST and the NPL as the transfer dosimetrs. Results in 1998 Results in 2008 Di 0.10 0.08 0.06 0.04 0.02 0.00 CMI-IIR ENEA-INMRI LNE-LNHB NIM VNIIFTRI Riso-HDRL NPL NIST -0.02-0.04-0.06-0.08-0.10 0 1 2 3 4 Dose level (1 kgy, 5 kgy, 15 kgy, 30 kgy) 13

NDAS Program in China From 1988, the project of NDAS was initiated by using NIM alanine/epr and dichromate dosimeters transfer standards in China, which is used to unify value of absorbed dose for 60 Coγ-rays radiation processing all over China. a set of 3~5 dosimeters with information sheet every six months by mail. Irradiation dose in the range of 0.5 to 40 kgy. Up to now, more then 70% γ-rays or EB radiation facilities used for food processing and sterilization participated in the NDAS program in China. From 2007 to 2016, altogether 988 dose checks have been carried out. 929 of 988 dose checks fall on the deviation range less than ±5%. 14

Dosimetric Process of Control Characterization of the irradiation facility and Validation process Dose Mapping in simulated product units: to determine location of D min and D max, and timer setting for achieving the desired D min for actual product. For EB: to measure beam and depth profiles, and to determine conveyor speed, beam current, electron energy, scan width for D min and maximise the beam penetration capability. Routine product processing To Control, monitor and document the operating parameters to ensure that each process load that passes through the irradiator is processed in accordance with specifications. To place dosimeters either within or on the selected process load at predetermined locations of D min (and D max if a prescribed limit) or at the reference positions to provide an independent check that absorbed dose is still within the desired D min and D max during the processing. 15

Standard and technical norm In order to guaranteeing the product quality of radiation processing, General Administration of Quality Supervision, Inspection and Quarantine (AQSIQ) has issued 20 national standards or technical norms concerned to the field of radiation processing with reference to ISO /ASTM Standards since 1990. Include as follows: GB 139-2008 Standard method for using the ferrous sulfate (Fricke) dosimeter to measure absorbed dose in water. JJG 1018-90 Standard Method for Using the Potassium (Silver) Dichromate Dosimeter to Measure γ-rays Absorbed Dose in Water. GB/T 16639-2008 Alanine-ESR dosimetry system for radiation processing. GB/T 15053-2008 Standard method for using the radiochromic film and PMMA dosimeters to measure absorbed dose GB/T 16640-2008 Standard guides for selection and calibration of dosimetry systems for radiation processing. GB/T 16334-2008 Practical guide of dosimetry in a gamma irradiation facility for food processing. GB/T 16841-2008 Guide for dosimetry in an electron beam facility for radiation processing at energies between 300kev and 25 Mev GB/T 16509-2008 Standard guides for estimating uncertainties in dosimetry for radiation processing. 16

Conclusion The Dosimetry Standards and Dissemination Systems for Radiation Processing has already provide a useful method to ensure the routine dosimeters tracing to the national standard and take an important role to control quality of radiation processing in China.