Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer

Similar documents
METHOD VALIDATION: WHY, HOW AND WHEN?

Method Comparison Report Semi-Annual 1/5/2018

METHOD VALIDATION CASE

METHOD VALIDATION CASE

The Virtues and Pitfalls of Implementing a New Test

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

SCATTER PLOTS AND TREND LINES

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

CRITERIA FOR USE. A GRAPHICAL EXPLANATION OF BI-VARIATE (2 VARIABLE) REGRESSION ANALYSISSys

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

The analytical phase

Total Thyroxine ELISA (T4)

Pitfalls in Linear Regression Analysis

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

Simple Linear Regression the model, estimation and testing

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

Dr. Kelly Bradley Final Exam Summer {2 points} Name

In Vitro Diagnostic Glucose Test System

Chapter 1: Exploring Data

Understandable Statistics

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan

Farouk AL Quadan lecture 3 Implentation of

Application Note 201

Statistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES

For In Vitro Diagnostic Use

2.75: 84% 2.5: 80% 2.25: 78% 2: 74% 1.75: 70% 1.5: 66% 1.25: 64% 1.0: 60% 0.5: 50% 0.25: 25% 0: 0%

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Validation Case studies: the good, the bad and the molecular

GUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS

Human Neurology 3-Plex A

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

SCIEX Vitamin D 200M Assay for the Topaz System

Validity and Reliability of a Glucometer Against Industry Reference Standards

Chapter 2. High-sensitivity C-reactive protein methods examined

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

1.4 - Linear Regression and MS Excel

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

Simple Linear Regression

Chapter 3 CORRELATION AND REGRESSION

LIPASE liquicolor. Design Verification. Multipurpose Reagent

Determination of hemoglobin is one of the most commonly

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Guide to Fulfillment of Measurement Uncertainty

liquicolor (AMP Buffer, IFCC) Design Verification

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

WDHS Curriculum Map Probability and Statistics. What is Statistics and how does it relate to you?

Applications of Freezing Point Osmometry

Business Statistics Probability

Supplemental Data. Methods- Different concentrations of substrate solutions (final concentrations during incubation- 10, 3,

Bias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

Table of Contents. Plots. Essential Statistics for Nursing Research 1/12/2017

Introduction to Statistical Data Analysis I

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

Unit 1 Exploring and Understanding Data

Rethinking Traditional Lab QA/QC:

Lecture Outline. Biost 517 Applied Biostatistics I. Purpose of Descriptive Statistics. Purpose of Descriptive Statistics

Technical Bulletin No. 161

25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL

New York State Soluble Tumor Markers Proficiency Test

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

UNIVERSITY OF TORONTO SCARBOROUGH Department of Computer and Mathematical Sciences Midterm Test February 2016

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

bivariate analysis: The statistical analysis of the relationship between two variables.

Research Article Derivative Spectrophotometric Method for Estimation of Metformin Hydrochloride in Bulk Drug and Dosage Form

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels

Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2000i system

Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up

Evaluation of the LIAISON 25 OH Vitamin D TOTAL Assay on the new LIAISON XL analyser

Evaluation and comparison of automated analysers on hepatic enzymes

EVALUATING THE UNCERTAINTY OF MEASUREMENT ON BLOOD'S GLUCOSE LEVEL

Evaluation of the QBC Star centrifugal three-part differential haematology system

SODIUM/NA. See below for information on factors affecting results. Certain substances, such as drugs, may affect analyte levels in vivo.

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

BAM Monitor Performance. Seasonal and Geographic Variation in NC

POTASSIUM/K. 3.6 mmol/l

From Biostatistics Using JMP: A Practical Guide. Full book available for purchase here. Chapter 1: Introduction... 1

Hemostasis Test Validation, Performance and Reference Intervals

Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens

Lecture 12: more Chapter 5, Section 3 Relationships between Two Quantitative Variables; Regression

Laboratory Utilization and Analytical Validation of Fecal Electrolyte Tests

A PROPOSED REFERENCE MATERIAL FOR OXIDATIVE INDUCTION TIME BY DIFFERENTIAL SCANNING CALORIMETRY

READ HIGHLIGHTED CHANGES

Individual Lab Report Ci-Trol Nov,2016. Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC Your Lab

Performance of Median and Least Squares Regression for Slightly Skewed Data

Performance Characteristics of Eight Estradiol Immunoassays

REGRESSION MODELLING IN PREDICTING MILK PRODUCTION DEPENDING ON DAIRY BOVINE LIVESTOCK

Describe what is meant by a placebo Contrast the double-blind procedure with the single-blind procedure Review the structure for organizing a memo

Statistical analysis and comments on the USI-IDD survey Azerbaijan 2007

ACCESS hstni SCIENTIFIC LITERATURE

Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS

The Adiponectin Turbidimetric Immunoassay Reagent Kit

CHAPTER 3 Describing Relationships

Ecological Statistics

Transcription:

AACC, July 2007 Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer E. Garry, M. Pesta, N. Zampa Advanced Instruments, Inc., Norwood, MA 02062 AbstracT The Advanced Model 3320 is an automated, single-sample freezing point depression (FPD) microosmometer that determines the total solute concentration (osmolality) of biological fluids, such as serum or urine. Osmolality measurements are commonly used by clinicians to assist in diagnosing and monitoring certain fluid and electrolyte imbalances in patients (i.e., hyponatremia, polyuria). FPD osmometers have been used in clinical chemistry laboratories for more than 40 years and provide a direct measurement of osmolality. This study was conducted to evaluate linearity and imprecision performance parameters of the new Advanced Model 3320 Micro-Osmometer using 20 μl samples of commercially available calibrators, reference solutions, controls, and normal human serum. Five replicates of 5 different osmolality levels ranging from 100 to 2,000 mosm/kg H 2 O were tested to assess linearity. Short and long-term replication experiments were conducted to evaluate imprecision. For the short-term study, 5 replicates of the following solutions were tested over a period of 20 days: 290 mosm/kg H 2 O reference solution; 850 mosm/kg H 2 O standard; urine matrix 300 mosm/kg H 2 O; urine matrix 800 mosm/kg H 2 O; serum matrix 240 mosm/kg H 2 O; serum matrix 320 mosm/kg H 2 O; and normal human serum. The long-term study lasted 547 days and was conducted by testing 10 replicates of the 290 mosm/kg H 2 O reference solution at various time intervals, for a total of 37 days of testing. The calculated slope and y-intercept of the linearity test were 0.99 and 1.5, respectively. The following short-term total standard deviation (SD) results were obtained: 1.3SD, 290 mosm/kg H 2 O reference solution; 2.8SD, 850 mosm/kg H 2 O standard; 1.6SD, urine matrix 300 mosm/kg H 2 O; 3.0SD, urine matrix 800 mosm/kg H 2 O; 1.5SD,

serum matrix 240 mosm/kg H 2 O; 1.5SD, serum matrix 320 mosm/kg H 2 O; and 2.2SD, normal human serum. In the long-term study, the mean value and total SD for the 290 mosm/kg H 2 O reference solution were 290.6 and 1.5, respectively. The linearity results of the osmometer indicate a strong linear response with a slope close to 1 and an intercept close to zero. The total standard deviations for all samples tested in both the short and longterm studies were within the manufacturer s imprecision claims of ± 2 mosm/kg H 2 O (1SD) between 0 and 400 mosm/kg H 2 O, and ± 0.5% (1SD) between 400 and 2,000 mosm/kg H 2 O. The linearity and imprecision results generated on the Advanced Model 3320 Micro-Osmometer correlate well to its specifications and demonstrate excellent instrument performance. Introduction One of the major changes in the CLIA Final Rule is that laboratories must now perform validation studies on all non-waived methods implemented after April 24, 2003. According to the Code of Federal Regulations, Title 21, Volume 8, 21 CFR Section 862.2730, An osmometer for clinical use is a device intended to measure the osmotic pressure of body fluids. Osmometers determine osmotic pressure by methods such as the measurement of the freezing point. Measurements obtained by this device are used in the diagnosis and treatment of body fluid disorders. Osmolality is considered a moderate complexity analyte, and both moderate and high complexity methods are now combined under the new term non-waived. CLIA requirements for method validation of clinical osmometers include the verification of performance specifications, as outlined in section 493.1253 (b) (1), prior to reporting patient test results. In this study, we assessed linearity and imprecision performance characteristics on the Advanced Model 3320 Micro-Osmometer following CLSI protocol guidelines. MATERIAL AND METHODS Linearity and short and long-term replication experiments to evaluate imprecision were conducted by testing 20 μl samples on the Advanced Model 3320 Micro-Osmometer (Advanced Instruments, Inc.) following manufacturer s instructions. Five replicates of 5 different osmolality levels ranging from 100 to 2,000 mosm/kg H 2 O (Osmolality Linearity Set, 3LA028, Advanced Instruments, Inc.) were tested to assess linearity. For the short-term study, 5 replicates of the following commercially available solutions from Advanced Instruments, Inc., were tested over a period of 20 days: 290 mosm/kg H 2 O reference solution (3MA029); 850 mosm/kg H 2 O standard (3MA085); urine matrix 300 mosm/kg H 2 O (3LA085); urine matrix 800 mosm/kg H 2 O (3LA085); serum matrix 240 mosm/kg H 2 O (3MA028); serum matrix 320 Advanced Instruments, Inc. Page 2 July 2007

mosm/kg H 2 O (3MA028); and normal human serum. The long-term replication experiment lasted 547 days and was conducted by testing 10 replicates of the 290 mosm/kg H 2 O reference solution (3MA029) at various time intervals, for a total of 37 days of testing. Linearity (slope and intercept) and short-term precision parameters (mean and total standard deviation) were evaluated statistically using EP Evaluator Release 7.0.0.251 (David G. Rhoads Associates, Inc.). The total SD is a composite of within-run, between-run, and between-day SD. Long-term precision data was evaluated using Minitab (Minitab, Inc.). A total of 369 data points were included and one excluded due to a sample did not freeze error. The long-term precision data was also evaluated using ControlChart!Pro Plus version 7.13.02 (ChemSW, Inc.) by plotting the average values of 10 data points for each day, using limits based on the mean ± 3 standard deviations. One outlier was removed from this data set. results Fig. 1. Linearity summary using 5-level osmolality set, showing measured versus assigned values, slope and intercept. 2,500 Scatter Plot 20 Residual Plot Measured (mosm/kg) 2,000 1,500 1,000 500 1:1 Line Fitted Overall Residual (mosm/kg) 10 0-10 0-20 0 500 1,000 1,500 2,000 2,500 0 500 1,000 1,500 2,000 2,500 Assigned (mosm/kg) Assigned (mosm/kg) Linearity Summary N Slope Intercept Error Overall 5 0.999 1.5 0.3% LINEAR within Allowable Systematic Error of 0.5% 3320 Osmometer Page 3 Advanced Instruments, Inc.

Statistical Analysis and Experimental Results Assigned Est Mean Resid Linear? Measured Concentrations a 100 101.5 101.2-0.3 Pass 101 100 103 102 100 b 500 501.3 502.2 0.9 Pass 502 503 501 503 502 c 900 901.0 903.4 2.4 Pass 907 902 898 905 905 d 1500 1500.7 1496.8-3.9 Pass 1499 1491 1497 1493 1504 e 2000 2000.4 2000.8 5.2 Pass 2003 1989 2011 2006 2019 Table. 1. Short-term total standard deviation results, n=20 Sample Mean (SD) 290 mosm/kg H 2 O reference solution 290.7 (1.3) 850 mosm/kg H 2 O standard 851.2 (2.8) urine matrix 300 mosm/kg H 2 O 300.4 (1.6) urine matrix 800 mosm/kg H 2 O 803.9 (3.0) serum matrix 240 mosm/kg H 2 O 240.4 (1.5) serum matrix 320 mosm/kg H 2 O 319.0 (1.5) normal human serum 286.9 (2.2) Advanced Instruments, Inc. Page 4 July 2007

Fig. 2. Long-term precision data descriptive statistics. Variable: Result Anderson-Darling Normality Test A-Squared 7.867 P-Value 0.000 Mean 290.569 StDev 1.490 Variance 2.21872 Skewness 0.363742 Kurtosis 0.470413 N 369 Minimum 287.000 1st Quartile 290.000 Median 290.000 3rd Quartile 291.000 Maximum 296.000 Descriptive Statistics 287 289 291 293 295 95% Confidence Interval for Mu 290.0 290.5 291.0 95% Confidence Interval for Mu 290.417 290.722 95% Confidence Interval for Median 95% Confidence Interval for Sigma 1.389 1.606 95% Confidence Interval for Median 290.000 291.000 Fig. 3. Control chart depicting average values for each day of long-term testing using limits based upon the mean ± 3 standard deviations. 292.187 291.635 +3.0SD +2.0SD 290.531 Mean 289.426 288.874-2.0SD -3.0SD 5 10 15 20 25 30 35 n = 36 mean = 290.5305 SD = 0.5523 %CV = 0.19% min = 289.70000 max = 291.6000 3320 Osmometer Page 5 Advanced Instruments, Inc.

conclusion NIST-traceable reference salt solutions and CLSI-recommended matrix control materials formulated to mimic unknown specimens were used in this study. The linearity results of the osmometer indicate a strong linear response with a slope close to 1 and an intercept close to zero (Fig. 1). The total standard deviations for all samples tested in both the short (Table 1) and long-term (Fig. 2) studies were within the manufacturer s imprecision claims of ± 2 mosm/kg H 2 O (1SD) between 0 and 400 mosm/kg H 2 O, and ± 0.5% (1SD) between 400 and 2,000 mosm/kg H 2 O. The instrument was calibrated only once at the beginning of the long-term drift study, indicating that this instrument is extremely stable (Fig. 3). The linearity and imprecision results generated on the Advanced Model 3320 Micro-Osmometer correlate well to its specifications (Fig. 4) and demonstrate excellent instrument performance. Fig. 4. Model 3320 Single-Sample Micro-Osmometer specifications. Model 3320 Single-Sample Micro-Osmometer Specifications Sample Volume Test Time Sample Capacity Units Resolution Range Communications 20 μl 60 seconds Single Sample mosm/kg H 2 O 1 mosm/kg H 2 O 0 to 2000 mosm/kg H 2 O DTE RS-232 serial port, parallel printer port, and optional barcode scanner Performance at Reference Conditions 1 Linearity Less than ± 1% from a straight line between 0 and 2000 mosm/kg H 2 O Repeatability ± 2 mosm/kg H 2 O (1 S.D.) between 0 and 400 mosm/kg H 2 O; ± 0.5% (1 S.D.) between 400 and 2000 mosm/kg H 2 O Drift Performance at Operating Conditions Less than 1 mosm/kg H 2 O per month Temperature Effects Less than 1 mosm/kg H 2 O per 5ºC (9ºF) ambient temperature range 1 Reference Conditions 20 to 25ºC (68 to 77ºF); 40 to 60% relative humidity; tolerances of reference or calibration solutions excluded Advanced Instruments, Inc. Page 6 July 2007

references 1. CLSI. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline Second Edition, CLSI document EP05-A2 (ISBN 1-56238-542-9). CLSI, 2004. 2. CLSI. Statistical Quality Control for Quantitative Measurement Procedures; Approved Guideline Third Edition, CLSI document C24-A3 (ISBN 1-56238-613-1). CLSI, 2006. 3. Westgard J. Basic Method Validation, Second Edition, 2003. 4. Westgard J, Ehrmeyer, S, Darcy, T. CLIA Final Rules for Quality Systems, First Edition, 2004. 3320 Osmometer Page 7 Advanced Instruments, Inc.

Two Technology Way Norwood, MA 02062 www.aicompanies.com (800) 225-4034 +1(781) 320-9000 info@aicompanies.com