ReShape Integrated Dual Balloon System Procedure Manual

Similar documents
ReShape Integrated Dual Balloon System Instructions for Use

ReShape Integrated Dual Balloon System Instructions for Use

DISCOVER NEW HORIZONS IN FLUID DRAINAGE. Bringing Safety and Convenience to Fluid Drainage Management

Initial placement 20FR Guidewire PEG kit REORDER NO:

Ponsky * PEG Safety System - "Pull" Bard * PEG Safety System - "Guidewire" Information for Use

Initial placement 24FR Pull PEG kit REORDER NO:

How to perform: HALO 360 Radiofrequency Ablation of Barrett s s Esophagus

Single Use Curlew TM Multiple Biopsy Forceps

Title: EZ-IO. Effective Date: January SOG Number: EMS Rescinds:

PRODUCTS FOR THE DIFFICULT AIRWAY. Courtesy of Cook Critical Care

Division 1 Introduction to Advanced Prehospital Care

Policies & Procedures. RNSP - RN Procedure. I.D. Number: 1097

Information Technology Solutions

AXIOS Stent and Electrocautery Enhanced Delivery System. Quick Reference Guide

Danis Procedure Pack - BASIC

ThruPort systems ProPlege peripheral retrograde cardioplegia device

CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE

DIGESTIVE HEALTH ENTERAL FEEDING PRODUCTS

1 Chapter 40 Advanced Airway Management 2 Advanced Airway Management The advanced airway management techniques discussed in this chapter are to

Chapter 40 Advanced Airway Management

RECOMMENDED INSTRUCTIONS FOR USE

Quick Reference Guide

CATHETER ACCESS KIT. For use with Prometra Programmable Infusion Systems

Placing PEG and Jejunostomy Tubes in Dogs and Cats

ARROW EZ-IO Intraosseous Vascular Access System Procedure Template

A-Tube Instructions for Use

Home Health Foundation, Inc. To create more permanent IV access for patients undergoing long term IV therapy.

ASEPT. Pleural Drainage System INSTRUCTIONS FOR USE REF LOT STERILE EO. Manufactured for: 824 Twelfth Avenue Bethlehem, PA

TJF-160F/VF Cleaning and Disinfection Checklist

EL DORADO COUNTY EMS AGENCY FIELD PROCEDURES

INNOVATIVE PRODUCTS G, J AND GJ-TUBES WITH ENFit TM CONNECTION

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

RD180 SP TECHNOLOGY GUIDE READ THIS PRODUCT INSERT THOROUGHLY BEFORE USE

Part 1 Principles and Routes of Medication Administration

CLARIVEIN INFUSION CATHETER

CCTC Minnesota Procedure: Minnesota Tube, Assisting with Insertion and Care of Patient

Patient Information Publications Warren Grant Magnuson Clinical Center National Institutes of Health

Endorectal Balloon (ERB) Endorectal Balloon (ERB) Instructions for Use

Integra. Endoscopic Gastrocnemius Release System SURGICAL TECHNIQUE

Managing Foreign Objects in the GI Tract Tips & Tricks

Distal Radius Surgical Technique Guide

RX Biliary System. Start

FAST1 Intraosseous Infusion System. Training Session

Vascu-PICC WITH CUFF PERIPHERALLY INSERTED CENTRAL VEIN ACCESS CATHETER INSTRUCTIONS FOR USE

Instruction for use of Foley Catheter

ER REBOA Catheter. Instructions for Use

Z-MED II BALLOON AORTIC AND PULMONIC VALVULOPLASTY CATHETER. Instructions for Use

Instruction For Use for All Silicon Foley Catheter

Tools of the Gastroenterologist: Introduction to GI Endoscopy

ASEPT. Pleural Drainage System INSTRUCTIONS FOR USE. Rx only REF LOT. STERILE EO Sterilized using ethylene oxide

Chapter 7, Medication Administration Part 1 Principles and Routes of Medication Administration Caution: Administering medications is business Always

Balloon in Balloon (BIB )

Balloon in Balloon (BIB )

limbsandthings.com Advanced Catheterisation Trainer User Guide For more skills training products visit Limbs & Things Ltd.

Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL

TRACHEOBRONCHIAL FOREIGN BODY REMOVAL ADVICE IN DOGS AND CATS

Itemized Billing and Procedure Description for the AspireAssist

GASTRO-INTESTINAL (ENTERAL)

ATI Skills Modules Checklist for Central Venous Access Devices

Procedure for removal and reinsertion of a supra pubic catheter

Peritoneal Drainage System

Aspira* Peritoneal Drainage Catheter

MULTIFIX S Knotless Implants

Technique Guide. *smith&nephew N8TIVE ACL Anatomic ACL Reconstruction System

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER

ASEPT Pleural Drainage System

Endoscopic Retrograde Cholangiopancreatography (ERCP)

SARASOTA MEMORIAL HOSPITAL. NURSING PROCEDURE INTRAOSSEOUS NEEDLE: INSERTION, CARE, AND REMOVAL (inv08) 12/18 12/18 1 of 7 RESPONSIBILITY:

The Spiral Enteroscopy Experience in 101 Consecutive Patients: Safety and Efficacy Using the Discovery SB

Adult Intubation Skill Sheet

Part 1: General requirements

Waitin In The Wings. Esophageal/Tracheal Double Lumen Airway (Combitube ) Indications and Use for the Pre-Hospital Provider

BULKAMID STANDARD OPERATING PROCEDURE

Procedure. P/N Rev B

SOP: Urinary Catheter in Dogs and Cats

CrackCast Episode 30 GI Bleeding

Overview. The Team Concept. Chapter 7. Assisting the ALS Provider 9/11/2012. The Team Concept ALS Procedures and Equipment

Pulmonary. Pulmonary Endoscopy. Alair Bronchial Thermoplasty System. Transbronchial Aspiration Needles. Cytology Brushes.

Educational Session: Evaluation and Management of the Difficult Airway

AIRWAY MANAGEMENT PRODUCTS

Rhinology Products. Rhinology Products. Superior solutions for superior patient care.

INSTRUCTIONS FOR PREPARING AND GIVING AN INJECTION OF ENBREL POWDER

Introducing the Fastrach-LMA. Prepared by Jim Medeiros, NREMT-P Regional Field Coordinator Lord Fairfax EMS Council

fassier DUVal TelescoPIc IM system s RESCUE SYSTEM 2.0

Gastric Balloon for Weight Loss. Karol A Gutowski, MD, FACS Hot Topics

Phoenix Atherectomy System Product Overview /LB

SHOULDER PATIENTS. Diagnostic Shoulder Arthroscopy Technique Guide

Instruction Guide to Sterile Self-Intermittent Catheterization For Men Using the Cure Catheter Closed System

How to Change a. Foley catheter. Patient Education Rehabilitation Nursing. Step-by-step instructions for the caregiver

Conventus CAGE PH Surgical Techniques

Look For These Other Products by PARÉ Surgical, Inc. INNOVATIVE MEDICAL DEVICES

Instructions for Use 1

SYNFLATE SYSTEM SURGICAL TECHNIQUE

IO considerations. Daniel Dunham

LESSON ASSIGNMENT. Urinary System Diseases/Disorders. After completing this lesson, you should be able to:

Technique Guide. *smith&nephew SPEEDSCREW Fully Threaded Knotless Implant

PREPARING FOR REFLUX TESTING. Bravo Reflux Testing System. A simple way to evaluate your gastroesophageal reflux symptoms

Helping to protect you from exposure to hazardous drugs

mild Devices Kit - Instructions for Use

Transcription:

ReShape Integrated Dual Balloon System Procedure Manual 1

TABLE OF CONTENTS Introduction... 3 Product Description... 3 Overview... 3 Component description... 4 Intragastric Balloon... 4 Placement Catheter... 4 Removal Catheter... 5 Procedure Supplies... 6 Balloon Placement Procedure... 6 Balloon Removal Procedure... 6 ReShape Dual Balloon Placement Procedure... 7 Step 1. Device preparation... 7 Step 2. Patient sedation... 8 Step 3. Diagnostic endoscopy and guidewire positioning... 8 Step 4. ReShape Dual Balloon insertion... 9 Step 5. Balloon inflation... 9 Step 6. Disengaging the Balloon from the catheter... 10 ReShape Dual Balloon Removal... 12 Step 1. Procedure preparation... 12 Step 2. Patient sedation... 12 Step 3. Balloon drainage... 13 Step 4. Capture and removal of the balloon... 14 References... 15 2

Introduction This procedure manual is designed to give the reader essential background information on the ReShape Integrated Dual Balloon System. The information in this manual provides a technical description of the devices and also serves as a guide on preparation, placement and removal procedures. Note: this manual does not replace the instructions for use (IFU), and as such should be used in combination with the IFUs supplied with the products. Product Description Overview The ReShape Integrated Dual Balloon is a temporary implant designed to facilitate weight loss by occupying space in the stomach and producing a sensation of satiety. The device is constructed from medical grade silicone rubber. The balloon is introduced orally via a placement catheter under endoscopic guidance and direct visualization. The balloon is designed to be placed in the stomach during an outpatient procedure using conscious sedation (MAC). It remains in the stomach for up to six months and is then removed. Fig.1 - ReShape Integrated Dual Balloon 3

Component description The ReShape Integrated Dual Balloon System contains (a) Intragastric Balloon / Placement Catheter Assembly, and the following accessories: (b) two Prefilled Valve Sealant Syringes, (c) Guidewire and (d) Patient ID Card. The ReShape Integrated Dual Balloon System is not made with natural rubber latex. (a) Intragastric Balloon (a) Placement Catheter (b) Prefilled Valve Sealant Syringes (c) Guidewire 4

(d) Patient ID Card The ReShape Removal Catheter is packaged separately 5

Procedure Supplies Balloon Placement Procedure Equipment/Supplies Single-use items Quantity per procedure Brand Part Number ReShape Dual Balloon 1 ReShape Medical RSM110 Guidewire 0.035in (0.89mm) diameter / 260cm length ReShape Infiltration Pump single-spike tubing Methylene blue (injectable, concentration 10mg/mL) 500cc flexible saline bags (0.9% NaCl solution) Prefilled Valve Sealant syringes (USP grade) Reusable items 1 ReShape Medical (AdvancedCath) 2 ReShape Medical ITS-10-RS Included with balloon 4 cc Clinic supply Clinic supply 2 Clinic supply Clinic supply 2 ReShape Medical Included with balloon Sterile 5cc or 6mL syringe & needle 1 Clinic supply Clinic supply Sterile graduated containers (size = min. 100 cc) Sterile adult endoscopy bite block opening size: 20mm(H) x 27mm(W) Balloon Removal Procedure Single-use items 2 Clinic supply Clinic supply 1 Olympus MAJ-168 ReShape Infiltration Pump 1 ReShape Medical KIP-II-RS Diagnostic gastroscope single-channel, 2.8mm channel 8.6-11.3mm insertion tube 1400mm total length Equipment/Supplies 1 Olympus GIF-Q160 GIF-Q180 Quantity per procedure Brand Part Number ReShape Removal Catheter 1 ReShape Medical RSM210 Large hexagonal endoscopic snare Minimum opening size: 20mm Sterile adult endoscopy bite block opening size: 20mm(H) x 27mm(W) Suction pump tubing Standard ¼ inch diameter Reusable items Suction pump (Removal Catheter independent from endoscope suction) Rat tooth graspers (wide jaw) Working Length: 165mm Opening Width: 19.5mm Minimum Channel Size: 2.8mm 1 1 Cook Medical Olympus ASH-1-S MAJ-168 2 Clinic Supply Clinic supply 2 Clinic Supply Clinic Supply 1 Olympus FG-49L-1 Laryngoscope and Magill forceps 1 Clinic supply Clinic supply Diagnostic gastroscope single-channel, 2.8mm channel 8.6-11.3mm insertion tube 1400mm total length 1 Olympus GIF-Q160 GIF-Q180 6

ReShape Dual Balloon Placement Procedure While the patient is under conscious sedation (MAC), an endoscope is inserted through the mouth into the stomach and a guidewire is placed. The un-inflated balloons are advanced over the guidewire and placed in the stomach. Each balloon is inflated independently with a salinemethylene blue solution using an automated pump. The device is then released and remains in the stomach for up to six months. The balloons do not change the stomach s anatomy; they simply occupy space to reduce the capacity for food. In the event of a complication requiring the removal of an inflated ReShape Dual Balloon, a ReShape Removal Catheter should be available at each placement procedure. Recommended patient diet instructions to minimize the amount of food and liquid in the stomach during the procedure: - 48 hours prior to procedure: Soft foods only, no meat in any form. - 24 hours prior to procedure: Clear liquids only. - 12 hours prior to procedure: NPO - No food or liquids by mouth. Step 1. Device preparation Materials: - ReShape Integrated Dual Balloon - ReShape Guidewire - Lubricant - Adult-sized endoscopic bite block - ReShape Infiltration Pump and ReShape Pump Tubing - Two (2) prefilled valve sealant syringes - Two (2) 500cc saline bags - 4cc of methylene blue (10mg/mL concentration) - Syringe and needle to transfer the methylene blue into the saline bags 1. The balloon is supplied preloaded on the placement catheter. Remove the catheter from the package and inspect for any visible damage. Do not use device if damage is noted. 2. Prepare two 500cc bags of saline solution. a. Inject and mix 2cc of methylene blue (10mg/mL) into each 500cc bag of saline b. Attach the infiltration pump tubing to the saline bag c. Determine the final inflation volume for each balloon. Hold the saline bag upside down and use the pump to remove any air and excess fluid from the 500cc saline bag in order to prevent over/under-inflation of the ReShape Dual Balloon. For example, if the intended balloon size is 450cc, remove all air and 50cc of solution from the saline bag such that only 450cc of fluid remains. Repeat this step for the saline bag intended for the second balloon. i. When emptying the saline bags into the graduated cylinder, place the pump speed between 4 7 to avoid splashing Note: Maintain cleanliness of the saline and tubing. Do not allow the pump tubing to contact the floor. Do not handle the tubing with dirty gloves. 7

3. Refer to the ReShape Infiltration Pump Instructions for Use for setup and operating information for the ReShape Pump and its accessories. a. Set the infiltration pump flow on the back of the pump to Momentary Flow b. Increase the infiltration speed to 10 for filling the balloons c. Alternate manual inflation method: The manual inflation method requires (1) 20mL Luerlock syringe, (1) 122cm IV tubing with spike, (1) 3-way valve, and (1) pressure tubing set. Assemble the inflation instruments and connect them to the saline/methylene blue solution. Verify that the instruments are connected such that the solution is pumped from the saline bag to the distal end of the pressure tubing intended for connection to the balloon catheter. Step 2. Patient sedation Prepare and position the patient as customary for a diagnostic endoscopy (i.e. left lateral decubitis). Administer an oral anesthetic spray as necessary. Administer conscious sedation per the institution s protocol for monitored anesthesia care (MAC). Step 3. Diagnostic endoscopy and guidewire positioning 1. The placement procedure requires a single-channel gastroscope with a diameter of 8.6 11.3mm, instrument channel diameter of 2.8mm, and a maximum working length of 140cm. 2. Position the bite block in the patient s mouth. Ensure that the bite block is large enough for both the ReShape Dual Balloon and the endoscope to pass through in parallel. 3. Perform an esophagogastroduodenoscopy (EGD) to identify any significant gastrointestinal pathology that would contraindicate implantation per protocol. If excessive food or liquid is present in the stomach, the procedure should be postponed. 4. Advance the guidewire through the endoscope s instrument channel port. Place the distal portion of the guidewire into the duodenum and position the guidewire along the greater curvature of the stomach. 5. Remove the endoscope from the patient while leaving the guidewire in place. While removing the endoscope, take note of the distance from the patient s mouth to the gastro-esophageal junction. This measurement will be used when introducing the Dual Balloon. Guidewire is placed in the duodenum and positioned along the greater stomach curvature 8

Step 4. ReShape Dual Balloon insertion 1. Prior to insertion, apply an ample amount of lubricant to the entire surface of the balloon. 2. Apply lubricant to the proximal segment of the guidewire and pass the guidewire through the ReShape Dual Balloon s guidewire lumen. The main operating physician should maintain control of the guidewire and provide countertraction when necessary. Insert the placement catheter into the patient s mouth, down the esophagus, and into the stomach. The depth markers on the placement catheter may be used as a reference for the distance between the patient s mouth to the proximal end of the implant. a. In the event of reduction in oxygen saturation levels, anesthesia should increase the oxygen flow rate to the patient, attempt to suction the back of the throat, and hold a jaw thrust to help regain normal oxygen saturation levels. 3. Reinsert the endoscope into the stomach alongside the placement catheter. Verify that the ReShape Dual Balloon is properly positioned as described below: a. The delivery catheter detent cap of the placement catheter is distal to the GE junction. b. The balloon is lying along the greater curvature of the stomach. c. Confirm the distal balloon is at or before the angle of incisura, not extending past the pylorus. d. Reposition the ReShape Dual Balloon as necessary. 4. After verifying the balloon s placement, position the endoscope superior to the proximal balloon and withdraw the guidewire from the ReShape Dual Balloon. Step 5. Balloon inflation 1. Remove the Luer-caps from each fill tube and inflate the balloons sequentially. Note: Maintain cleanliness of the saline and tubing. Do not allow the pump tubing to contact the floor. Do not handle the tubing with dirty gloves. 2. The proximal balloon is to be inflated first. Connect the infiltration pump tubing to the proximal balloon fill tube (catheter lanyard 1) and inflate to the desired volume. Monitor inflation under endoscopic visualization. a. Confirm that the saline/methylene blue solution bag contains only the intended inflation volume for the balloon and any air and excess fluid has been removed. b. Inflate the proximal balloon. Monitor its inflation through the endoscope. 9

i. Monitor the balloon during filling and verify that the fill tube does not disconnect. ii. Observe and make sure the orientation and location of the balloon does not change during inflation. iii. Upon completion of balloon inflation, endoscopically inspect the balloon and confirm: 1. The intended inflation volume was completely transferred from the saline bag to the balloon. 2. No leakage of fluid from balloon. c. Once inflation is complete, attach the first valve sealant syringe to the proximal balloon fill tube (catheter lanyard 1) and inject 6cc of sealant into the proximal balloon i. Leave the valve sealant syringe attached to the fill tube to serve as a confirmation that this step has been completed. 3. Then, reposition the endoscope so the distal balloon can be viewed. Connect the infiltration pump tubing to the distal balloon fill tube (catheter lanyard 2) and inflate to the desired volume. Monitor inflation under endoscopic visualization. a. Confirm the saline/methylene blue solution bag contains only the intended inflation volume for the balloon and any air and excess fluid has been removed. b. Inflate the distal balloon. Monitor its inflation through the endoscope. i. Monitor the balloon during filling and verify that the fill tube does not disconnect. ii. Upon completion of balloon inflation, endoscopically inspect the balloon and confirm: 1. The intended inflation volume was completely transferred from the saline bag to the balloon. 2. No leakage of fluid from balloon. c. Once inflation is complete, attach the second valve sealant syringe to the distal balloon fill tube (catheter lanyard 2) and inject 6cc of sealant into the distal balloon. i. Leave the valve sealant syringe attached to the fill tube to serve as a confirmation that this step has been completed. 4. DO NOT DISCONNECT THE PLACEMENT CATHETER FROM THE BALLOON UNTIL VALVE SEALANT HAS BEEN INJECTED INTO EACH BALLOON. Step 6. Disengaging the Balloon from the catheter 1. Position the endoscope proximal to the detent cap on the placement catheter. It is important to keep the endoscope proximal to the detent cap during removal of the placement catheter. If the endoscope is alongside the detent of the balloon catheter, the combined diameter of the detent cap and the endoscope may be too large to be pulled into the esophagus. 2. Hold both the Dual Balloon Catheter and the endoscope together and pull back in tandem. Take care to keep the endoscope proximal to the detent cap. Pull the inflated balloons against the GE junction; continue to pull the catheter and the endoscope with a slow and smooth constant force to release the inflated balloons from the placement catheter. 10

3. Once released from the Dual Balloon Catheter, continue to withdraw the placement catheter and endoscope from the patient. 4. Reinsert the endoscope to inspect the upper GI tract and the ReShape Dual Balloon. 5. Provide the patient with a Patient ID Wallet Card and instruct the patient on potential postprocedure discomfort and symptoms. 11

ReShape Dual Balloon Removal Attention: Appropriate precautions must be taken to ensure that the patient returns for balloon removal six months after insertion. The device is removed via endoscopy under conscious sedation (MAC). Each balloon is completely drained in a controlled manner. A snare captures the deflated dual balloon around the proximal cap and the device is removed through the mouth. Recommended patient diet instructions to minimize the amount of food and liquid in the stomach during the procedure: - 48 hours prior to procedure: Soft foods only, no meat in any form. - 24 hours prior to procedure: Clear liquids only. - 12 hours prior to procedure: NPO - No food or liquids by mouth. Note: If the patient does not follow these instructions, consider cancelling the removal procedure as food in the stomach during removal could increase the risk of aspiration. Step 1. Procedure preparation Equipment and Materials: - ReShape Removal Catheter - Adult-sized endoscopic bite block - Two (2) suction pumps or equivalent with ¼ inch tubing; one dedicated to the Removal Catheter. - Lubricant - Endoscopic snare - Endoscopic rat-tooth grasping forceps - Laryngoscope and Magill forceps 1. Remove the ReShape Removal Catheter from its packaging. Inspect the Removal Catheter for any damage. 2. Prepare two (2) suction canisters or equivalent with empty 1-liter canisters. 3. Have a laryngoscope and Magill forceps on hand. In case of difficulty while removing the drained balloons through the upper esophageal sphincter, a laryngoscope and Magill forceps may be used to retrieve the ReShape Dual Balloon device. Step 2. Patient sedation Prepare and position the patient as customary for a diagnostic endoscopy (i.e., left lateral). Administer conscious sedation, per the institution s protocol for monitored anesthesia care (MAC). Note: In case of emergency or complication during removal, the patient should be intubated. 12

Step 3. Balloon drainage 1. The removal procedure requires a single-channel gastroscope with a diameter of 8.6 11.3mm, instrument channel diameter of 2.8mm, and a maximum working length of 140cm. 2. Position an adult-size endoscopic bite block into the patient s mouth. 3. Perform an endoscopic inspection of the esophagus and stomach. If excessive food or liquid is present, the procedure should be postponed. 4. With the endoscope in the stomach, insert the Removal Catheter through the endoscope s instrument channel. 5. The order in which the balloons are drained is at the discretion of the physician. Position the endoscope adjacent to the first balloon to be drained and orient the Removal Catheter perpendicular to the surface of the balloon. 6. Advance the Removal Catheter up to the balloon surface, apply light pressure and turn the crank handle clockwise until it stops to advance the needle and puncture the balloon. a. Insert the Removal Catheter into the balloon up to the second depth marking, and then fully retract the needle by turning the crank handle counter clockwise. It is important to ensure that the needle has been retracted before inserting the catheter beyond the second marking. b. Insert the catheter deeper into the balloon up to the third depth marking. c. Open the Removal Catheter s valve and apply suction to drain the balloon. d. Hold the endoscope steady and ensure that the catheter remains positioned within the balloon until the balloon is completely drained. 13

e. After draining the balloon, withdraw the catheter from the balloon. Note: If necessary, temporarily leave the Removal Catheter s valve open and disconnect the suction pump tube. This will relieve the vacuum within the deflated balloon and allow the catheter to be pulled out of the balloon. 7. Close the Removal Catheter s valve and reconnect the suction pump tube. Reposition the Removal Catheter perpendicular to the surface of the next balloon and repeat the same steps as was done to drain the first balloon. 8. Completely withdraw the Removal Catheter from the endoscope when the second balloon has been drained. 9. Insert endoscopic rat-toothed grasping forceps to tear a hole in each of the balloons. This allows trapped fluid and air to completely drain from the balloon and reduce resistance during removal. 10. Before removing the graspers, search for the proximal cap of the balloon device. If necessary, use the graspers to reposition the balloon device such that the proximal cap is easily seen. Note: If the proximal cap is unreachable or sitting too far into the fundus, grasp the central shaft and push the entire balloon device distally. This will allow snaring of the proximal cap in an antegrade position. Alternatively, flip the balloon to place the proximal cap in the distal portion of the stomach where the distal cap can be snared by retroflexing the endoscope. 11. Remove the grasper and any remaining fluid from the stomach using endoscopic suction. Step 4. Capture and removal of the balloon 1. Insert an endoscopic snare and prepare to capture the drained ReShape Dual Balloon. a. Position the endoscope as necessary in order to access the proximal cap. b. Capture the proximal cap with the endoscopic snare. 14

Note: Snaring the proximal cap provides a secure hold of the drained ReShape Dual Balloon. Do not capture the device by the balloon skin or along the shaft. 2. With the ReShape Dual Balloon proximal cap firmly grasped within the snare, remove the endoscope and the captured balloon from the patient. Make sure to hold both the snare and endoscope when pulling the balloon up the esophagus. The snared balloon should remain under direct endoscopic visualization during the entire removal. a. In case of difficulty during removal, use the following technique: i. If the balloon releases from the snare in the lower esophagus: 1. Push the balloon with the tip of the endoscope until the balloon has been pushed from the esophagus back down into the stomach. Note: this should be done with light and steady pressure. Too much force could cause the scope to bend instead of pushing the balloon. 2. Reposition the balloon device as necesary and snare the proximal cap, again. ii. If the balloon releases from the snare in the upper esophagus, use a laryngoscope and Magill forceps to retrieve the ReShape Dual Balloon device. 3. Reinsert the endoscope to inspect the patient s esophagus and stomach. References Instructions for Use ReShape Integrated Dual Balloon System Instructions for Use ReShape Gastroenterology Guidewire Instructions for Use ReShape Removal Catheter Instructions for Use ReShape Infiltration Pump 15

ReShape Medical, Inc. 100 Calle Iglesia San Clemente, CA 92672 844-YES-RESHAPE 844-937-7374 www.pro.reshapeready.com 2016 ReShape Medical, Inc. All rights reserved. RESHAPE and the RESHAPE MEDICAL Logo are registered trademarks of ReShape Medical, Inc. 04-0070 Rev. C 16