The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses

Similar documents
Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology,

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population

Studies with electron beam computed tomography (EBCT) Imaging

Coronary revascularization treatment based on dual-source computed tomography

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Non-invasive Coronary Angiography: the Role, Limitations and Future of 64-Slice Spiral Computed Tomography Coronary Angiography

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT

Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control

Accuracy of Multislice Computed Tomography in the Preoperative Assessment of Coronary Disease in Patients With Aortic Valve Stenosis

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital

Low-dose CT coronary angiography in the step-andshoot mode: diagnostic performance

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications

EXPERIMENTAL AND THERAPEUTIC MEDICINE 5: , 2013

Coronary CT Angiography

M Marwan, D Ropers, T Pflederer, W G Daniel, S Achenbach

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography

Zurich Open Repository and Archive

Radiation dose of cardiac CT what is the evidence?

THE ROLE OF HIGH END MULTI DETECTOR CT IN CORONARY IMAGING ESSAY

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats

Diagnostic Accuracy of Computed Tomography Angiography in Patients After Bypass Grafting

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination

Cardiac CT imaging in coronary artery disease: Current status and future directions

What every radiologist should know about cardiac CT: A case-based pictorial review

2004;77:800 4 MSCT OF CORONARY ARTERY BYPASS GRAFTS. Results. CABG With Adequate Diagnostic Quality

Chapter. Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006

Cardiac Computed Tomography

Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients

Adapted Transfer Function Design for Coronary Artery Evaluation

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

Angio-CT: heart and coronary arteries

TITLE: Multi-Slice Computed Tomography Coronary Angiography for Coronary Artery Disease: A Review of the Clinical Effectiveness and Guidelines

Computed Tomography of the Coronary Arteries

A Noninvasive Assessment of CAD

Dr Felix Keng. Imaging of the heart is technically difficult because: Role of Cardiac MSCT. Current: Cardiac Motion Respiratory Motion

Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT

Coronary angiography is the standard way of visualizing

Low Dose Era in Cardiac CT

Value of dual-source computed tomography in evaluating left ventricular function in patients with coronary heart disease

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease

Diagnostic Value of 64-Slice Dual-Source CT Coronary Angiography in Patients with Atrial Fibrillation: Comparison with Invasive Coronary Angiography

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16.

The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6

Purpose. Methods and Materials

Computed Tomography Imaging of the Coronary Arteries

Disclosure Information

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time

Diagnostic Accuracy of Multidetector Computed Tomography Coronary Angiography in Patients With Dilated Cardiomyopathy

Cardiopulmonary Imaging Original Research

Recent developments in coronary computed tomography imaging

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING

1. Department of Medical Radiation Sciences, Curtin University, Perth, Western Australia, 6845,

Routine low-radiation-dose coronary computed tomography angiography

Non-invasive coronary angiography with high resolution multidetector-row computed tomography

A systematic approach for analysis, interpretation, and reporting of coronary CTA studies

Cardiac computed tomography: indications, applications, limitations, and training requirements

Index. radiologic.theclinics.com. Note: Page numbers of article titles are in boldface type.

Detection of Coronary Artery Stenoses with Thin Slice Multi Detector Row Spiral Computed Tomography Angiography

Recent developments in cardiac CT

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Dual-source computed tomography for evaluating coronary stenosis and left ventricular function

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA

Impact of 64-Slice Multidetector Computed Tomography on Other Diagnostic Studies for Coronary Artery Disease

Clinical Medicine Insights: Cardiology

General Cardiovascular Magnetic Resonance Imaging

Coronary artery bypass grafting has been a historically. Multislice CT Evaluation of Coronary Artery Bypass Graft Patients SYMPOSIA

CT Versus MR for the Runoff

PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT OF VASCULAR OCCLUSION EXTENT AND LOCALIZATION OF EMBOLI 1. BACKGROUND

Automatic Ascending Aorta Detection in CTA Datasets

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography

Pak Heart J ORIGINAL ARTICLE ABSTRACT Vol. 45 (02) : Farhan Tuyyab, Muhammad Yahya Naeem, Ghulam Rasool Maken, 4 5

The assessment of cardiac chamber dimensions, ventricular

Diagnosis of coronary stenosis with CT angiography comparison of automated computer diagnosis with expert readings.

Interpreting CT Angiography: Three-Dimensional Reconstruction Techniques

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease

Sixty-four-slice multidetector computed tomography: the future of ED cardiac care

QUANTITATIVE ASSESSMENT OF IMAGE QUALITY IN 64-SLICE-COMPUTED TOMOGRAPHY OF CORONARY ARTERIES IN SUBJECTS UNDERGOING SCREENING

Cardiac Imaging Tests

ECG-Gated 16-MDCT of the Coronary Arteries: Assessment of Image Quality and Accuracy in Detecting Stenoses

Sang Ho Lee, Byoung Wook Choi, Hee-Joung Kim*, Member, IEEE, Haijo Jung, Hye-Kyung Son, Won-Suk Kang, Sun Kook Yoo, Kyu Ok Choe, Hyung Sik Yoo

X-ray coronary angiography is considered the diagnostic

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference.

Transcription:

Original Article Open Access The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Ziqiao Lei 1, Jin Gu 2, Qing Fu 3, Heshui Shi 4, Haibo Xu 5, Ping Han 6, Jianming Yu 7 ABSTRACT Objective: To evaluate the diagnostic accuracy of dual-source computed tomography (DSCT) in the diagnosis of coronary artery stenoses by comparing with conventional coronary angiography (CCA). Methodology: CCA and DSCT were performed in 64 patients with suspected coronary artery disease (CAD) respectively (46 male, 18 female, age from 48 to 82 years old, mean 68.18 years). Various postprocessing reconstructions of coronary artery and its branches, such as volumetric imaging, multi-planar reconstruction, curved planar reconstruction, maximum intensity projection were used. The coronary segments, with statistical evaluations combined with its diameter 1.5mm were collected to analyze the diagnosis accuracy of DSCT on coronary artery stenoses, with CCA as the gold standard. Results: About 4.14% of coronary artery segments could not be evaluated, while 95.86% were evaluable arteries, the sensitivity, specificity, positive and negative predictive value of DSCT for detecting coronary artery stenoses were, 93.58%, 99.61%, 95.31% and 99.48% respectively. There were no significant differences in the diagnostic accuracy of coronary artery stenoses between DSCT and CCA. Conclusion: DSCT is a reliable tool that is accurately appropriate for patients with CAD, as it has a higher accuracy and specificity, which is valuable in the screening of CAD. KEY WORDS: Coronary artery stenoses, Dual-source CT, Computed tomography, Coronary angiography. doi: http://dx.doi.org/10.12669/pjms.291.2645 How to cite this: Lei Z, Gu J, Fu Q, Shi H, Xu H, Han P, Yu J. The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses. Pak J Med Sci 2013;29(1):107-111. doi: http://dx.doi.org/10.12669/pjms.291.2645 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. 1. Ziqiao Lei, 2. Jin Gu, 3. Qing Fu, 4. Heshui Shi, 5. Haibo Xu, 6. Ping Han, 7. Jianming Yu, 1-7: Department of Radiology, Union Hospital, Wuhan, China. Correspondence: Ziqiao Lei, Department of Radiology, Union Hospital, Wuhan, China. E-mail: kevin@asiastm.org * Received for Publication: July 2, 2012 * Revision Received: October 31, 2012 * Revision Accepted: November 2, 2012 INTRODUCTION Multi-slice computed tomography (MSCT) has gained a great acceptance in clinical application as a reliable and non-invasive method to detect coronary heart disease. 1,2 Dual-source computed tomography (DSCT) is one of the latest innovations in multi-slice CT technology, which is composed of two x-ray tubes and two corresponding detectors. The gantry rotation time is 0.33s, the temporal resolution of 83ms is faster than the other MSCT before. It has a great display of coronary artery and the evaluation of coronary artery stenoses with diameter more than 1.5mm. 3-5 We retrospectively evaluated subjects who prospectively underwent SCA and DSCT in order to evaluate the application value of DSCT in patients Pak J Med Sci 2013 Vol. 29 No. 1 www.pjms.com.pk 107

Ziqiao Lei et al. with coronary artery stenoses as well as systematic analysis of its value in the diagnosis of varying coronary artery stenoses. METHODOLOGY Patients: From December 2008 until April 2011 we included 64 patients who were referred to our hospital for DSCT (Somatom Definition, Siemens Medical Solutions, Forchheim, Germany) for suspected coronary artery disease (CAD), Exclusion criteria were: unstable angina, allergic to iodinated contrast material, renal insufficiency, severe respiratory function impairment and heart failure. METHODS DSCT data acquisition: All the patients cooperated with the DSCT examination at the time of scan, without taking any anti-arrhythmic drugs. The heart rate ranged from 58 to 96beats/min, and averaged 70.3beats/min. We used artificial intelligence trigger system (bolus tracking mode) and the scan was started with a delay time of 5s after the density in the aortic root exceeded a density value of 100HU. We used an injector (Medrad Stellant) with high-pressure syringe of 3.8-4.5 ml/s rate of 65-75ml of the contrast medium injection, then a saline flush followed (50ml at 4.0ml/s of 0.9%). The coronary angiography was performed with DSCT at 0.6mm collimation, 120kVp.330ms gantry rotation time, automatic pitch or manual one (0.2 to 0.44), 0.75 reconstruction slice thickness. During the scan, patients were in deep inspiration breath-hold status with a scan range from tracheal subcarinal to diaphragmatic dome in a craniocaudal direction. ECG was recorded simultaneously in order to make image reconstruction. DSCT image post-processing: The images were transferred to separate workstations for postprocessing. The initial axial images reconstruction was set with the reconstruction window starting at 75% (equivalent to end of diastolic phase) and 45% (equivalent to end of systolic phase) of the cardiac cycle. The images were reconstructed with a section thickness of 1mm and an increment of 0.5mm by using a medium smooth reconstruction kernel (B35f). Then some post-processing work was analyzed such as MPR (multi-planar reformation), CPR (curved planar reformation), MIP (maximum intensity projection), VRT (volume rendering techniques) and optimized analysis software for coronary vessels. VRT was used to observe three dimensional structures of coronary artery while MPR, CPR and MIP were used to manifest coronary artery wall, the lumen changes and the relationship with the adjacent vascular and other organizations. Coronary artery section and evaluation standard: The American College of Cardiology (ACC) and the American Heart Association (AHA) classification of coronary segments were adopted as guidelines and basis for imaging evaluation. 6 Right coronary artery (RCA) is divided into proximal, middle and distal segments (otherwise the right atrium branch, right ventricle branch, right acute marginal branch, posterior interventricular branch and so on). The left coronary artery is divided into left main coronary artery (LM) and left circumflex branch (LCX). LAD is divided into proximal, middle and distal segments, LCX into proximal and distal segments (otherwise the interval branch, the first, second, third diagonal branches, the first, second left obtuse marginal branch). All the coronary segments were evaluated gradually in segmentation if the vessel diameter 1.5mm. The evaluation was possible even with a few artifacts but still met the demands of the diagnosis. The vessel was not evaluated if any segment of the coronary main branch could not be evaluated. CCA examination: CCA was successfully performed in all 64 patients before or after DSCT, 3-28 days interval between the two examinations, mean 16 days. Various perspectives were applied to observe the coronary artery, at least four perspectives to left coronary artery and two to right coronary artery. Assessment methods: Double-blind analysis between two independent investigators of DSCT and SCA output were compared. For coronary vessel diameter 1.5mm, analysis of coronary artery stenoses was done with the use of internationally accepted method of visual diameter, calculated as follows: Degree of stenosis = (normal vessel diameter proximal to stenosis-diameter of stenosed part)/ normal vessel diameter proximal to stenosis x 100%. Grade of coronary stenosis 7 : mild <50% diameter stenosis, moderate 50% to <75%, severe 75% and vascular occlusion 100%. Statistical analysis: Statistical analysis was done using SPSS 15.0 software. Comparing with the CCA results, we calculated the sensitivity, specificity, positive and negative predictive value of the DSCT coronary artery pathological lesions; Paired χ2 test was used to detect the differences between DSCT and SCA in their ability to manifest coronary artery lesions. For all the data, P values of 0.05 were considered to be significant. 108 Pak J Med Sci 2013 Vol. 29 No. 1 www.pjms.com.pk

DSCT in the diagnosis of coronary artery stenoses RESULTS Sixty four patients successfully underwent the DSCT and SCA examinations (46 men, 18 women, aged 48-82 years, mean age 61 years). There were 845 segments in SCA and 810 segments in DSCT of the coronary artery that could be assessed with the diameters>1.5mm, while, 4.14% (35/845) segments could not be evaluated separately for following reasons: 19 segments of motion artifact, 12 segments of dense calcification and 4 segments of negative lumen manifestation. These cases were excluded from our study. In the 36 segments of mild stenosis which was manifested in SCA, 29 segments were correctly diagnosed, 3 segments were misdiagnosed and 4 segments were overestimated (3 was moderate stenoses and 1 was severe) in DSCT. In the 30 segments of moderate, diagnosis which is manifested in SCA, 26 segments were correctly diagnosed (Fig.1-4), 3 segments were miss diagnosed, and one segment was overestimated in DSCT. In the 65 segments of severe stenoses which was manifested in SCA, 61 segments were correctly diagnosed, 4 segments were misdiagnosed and 3 segments were overestimated in DSCT (Table-I). According to the results mentioned above, the sensitivity, specificity, positive and negative predictive value of the DSCT evaluation for the severe coronary artery stenoses were 93.85%, 99.61%, 95.31%, 99.48% respectively. Paired χ2- test found P value >0.05, showing no statistically significant difference between DSCT and CCA on manifestation of severe coronary artery stenosis. DISCUSSION As one of motives of the development of multislice CT technology, non-invasive cardiac imaging has been limited by low resolution. Therefore, the imaging results of the 16-slice or 64-slice CT still have some discordance with CCA. 8 DSCT coronary angiography has more advantages over SCA. DSCA has dual X-ray tubes, a gantry rotation time of 0.33s thus resulting in a temporal resolution of 82.5ms. The 90 rotation can generate excellently higher imaging quality and scan speed compared to single-source CT, and a significant improvement of time resolution and clinical application as well as patient comfort. Secondly, DSCT possesses higher spatial resolution, which combines with the spatial resolution of <0.4mm that can clearly demonstrate 2 to 3 grade branches of coronary artery with diameter 1.5mm, making the detection rate increased substantially to display accurate small plaques or calcification of the coronary artery. 9,10 Several international studies suggest that DSCT has a good consistency than SCA in the diagnosis Fig.1: The right coronary showing a middle occluded right coronary artery as well as a distal severe stenosis. a.vrt b.cpr c.axial images Fig.2: Invasive coronary angiogram of the right coronary artery showing a middle and distal severe stenosis. Pak J Med Sci 2013 Vol. 29 No. 1 www.pjms.com.pk 109

Ziqiao Lei et al. Table-I: DSCT analysis of 64 patients. Comparative indicators Mild Moderate Severe True positive (segments) 29 26 61 True negative (segments) 679 731 764 False positives (segments) 4 1 3 False negative (segments) 3 3 4 Sensitivity [n (%)] 29/32(90.63) 26/29(89.66) 61/65(93.85) Specificity [n (%)] 679/683(99.41) 731/732(99.86) 764/767(99.61) Positive predictive value [n (%)] 29/33(87.88) 26/27(96.30) 61/64(95.31) Negative predictive value [n (%)] 679/682(99.56) 731/734(99.59) 764/768(99.48) of CAD. DSCT can display most of coronary arteries and could accurately assess the degree of stenosis. 10 By comparing the results of DSCT with that of SCA, it was found that DSCT can clearly show coronary distribution and 2 to 3 grade branches of coronary artery with diameter 1.5mm. The diameter 1.5mm of coronary artery was chosen as an evaluation object because arteries of this much size can be revascularized. 11 In this study, DSCT image has a good quality rate of 95.86%, and the sensitivity, specificity, positive and negative predictive value of coronary artery stenoses at severe levels. DSCT has a higher reliability and diagnostic value on severe coronary artery stenoses. It has a high sensitivity of 93.8%, a specificity of 99.61%, 95.31% positive predictive value and negative predictive value of 99.48%. The results are in accordance with that of Scheffel et al 12 research output whereby 30 cases of CAD patients underwent DSCT coronary angiography without controlling heart rate, the sensitivity, specificity, positive and negative predictive value of coronary artery segmental stenoses are 96.4%, 85.7%, 85.7 and 99.4% Heart rate control is not mandatory prior to the DSCT scan. Johnson et al 8 concluded that DSCT coronary angiography can provide high quality diagnostic value of coronary arteries, heart valves and heart imaging after re-evaluation of them on 24 patients with 44-92 beats/min heart rate. Scheffel et al 12 study and Achenbanh et al 10 study all came to the similar conclusions. In our study, Patient's heart rate before the test was 58 to 96 / min, an average of 70.3 / min without taking β- blockers or other drugs to reduce heart rate. While the single source 64-slice CT requires the patients heart rate to be maintained at 60 to 65 beats/min. 13 The study found that DSCT for coronary artery stenoses had a high negative predictive value is important to rule out coronary artery lesions. This means that invasive examination of CCA could be avoided for the patients with high negative predictive value in DSCT. However, positive predictive value of mild stenoses is low. Nineteen segments of cardiac motion artifacts, 12 segments of dense calcification and 4 segments of poor lumen enhancement were analyzed retrospectively. The cardiac motion artifacts were caused by arrhythmia (2 patients with atrial fibrillation, one patient with sinus arrhythmia), indicating that although there are no significant requirements on heart rate speed, an irregular cardiac rhythm is the key restrictions. When the coronary walls are calcified, it is difficult to accurately determine the degree of stenosis or it could be overestimated because of the partial volume effects in DSCT. Diffuse mixed calcified coronary artery plaques are often accompanied by vascular positive remodeling and no significant stenoses of the artery lumen, so the DSCT examination shows overestimation of luminal stenoses, the accuracy of coronary artery angiography results on both DSCT and single source 64-slice CT are affected by coronary artery calcification. 14 It is difficult to distinguish the vessel lumen with density extensive calcification, leading to an incorrect evaluation of the degree of luminal stenosis. 15-17 Several studies have shown that the sensitivity, specificity, positive and negative predictive value of 64-slice CT in detecting severe calcification (score>400) are 93%, 67%, 93% and 67% respectively. 13 In Scheffel et al 12 study on DSCT the sensitivity, specificity, positive and negative predictive values are 96.1%, 94.8%, 86% and 98.7% respectively which are better than the 64-slice CT. Although diagnostic accuracy of DSCT is affected by severe calcification, it has been greatly improved in comparison to other non-invasive coronary artery examination methods. Poor lumen enhancement has been correlated with the amount of contrast agent, scan delay time and the degree of proximal segment coronary artery stenoses. The 4 segments in the study which couldn t be evaluated were caused by proximal coronary artery stenoses. 110 Pak J Med Sci 2013 Vol. 29 No. 1 www.pjms.com.pk

DSCT in the diagnosis of coronary artery stenoses A total of 8 stents in the group were assessed. DSCT coronary angiography can affirm a clear indication to display the stent lumen and plaques at both sides. DSCT and CCA examination have the same capability to check for re-stenosis detection and classification of diagnosis, which suggests that DSCT has a high accuracy in the diagnosis of stent re-stenosis, and it is the same with reports of Das et al. 18 In short, DSCT coronary angiography has solved problems such as time resolution. 9,10 We can get better image quality without the need for patients to heart rate control. DSCT for noncalcified coronary arteries show a higher reliability and diagnostic value. The sensitivity, specificity, positive and negative predictive values, for evaluating the degree of stenosis by DSCT, all meet the clinical requirements. So DSCT can be used as the preferred method in preoperative screening for CCA in patients with clinical suspicion of CAD or postoperative coronary stent implantation or patients on followup after coronary artery bypass grafting surgery. REFERENCES 1. Dodd JD, Rieber J, Pomerantsev E. Quantification of nonculprit coronary lesions: comparison of cardiac 64- MDCT and invasive coronary angiography. AJR Am J Roentgenol. 2008;191(2):432-438. 2. Ropers D, Rixe J, Anders K. Usefulness of multidetector row spiral computed tomography with 64- x 0.6-mm collimation and 330-ms rotation for the noninvasive detection of significant coronary artery stenoses. Am J Cardiol. 2006;97(3):343-348. 3. Lell MM, Panknin C, Saleh R. Evaluation of coronary stents and stenoses at different heart rates with dual-source spiral CT (DSCT). Invest Radiol. 2007;42(7):536-541. 4. Brodoefel H, Burgstahler C, Tsiflikas I. Dual-source CT: effect of heart rate, heart rate variability, and calcification on image quality and diagnostic accuracy. Radiology. 2008;247(2):346-355. 5. Ropers U, Ropers D, Pflederer T. Influence of heart rate on the diagnostic accuracy of dual-source computed tomography coronary angiography. J Am Coll Cardiol. 2007;50(25):2393-2398. 6. Scanlon PJ, Faxon DP, Audet AM. ACC/AHA guidelines for coronary angiography. A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (Committee on Coronary Angiography). Developed in collaboration with the Society for Cardiac Angiography and Interventions. J Am Coll Cardiol. 1999;33:1756-1824. 7. Nieman K, Oudkerk M, Rensing BJ. Coronary angiography with multi-slice computed tomography. Lancet. 2001;357:599-603. 8. Johnson TR, Nikolaou K, Wintersperger BJ. Dual-source CT cardiac imaging: initial experience. Eur Radiol. 2006;16(7):1409-1415. 9. Flohr TG, McCollough CH, Bruder H. First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol. 2006;16:256-268. 10. Achenbach S, Ropers D, Kuettner A. Contrast-enhanced coronary artery visualization by dual-source computed tomography-initial experience. Eur J Radiol. 2006;57:331-335. 11. Ropers D, Baum U, Pohle K. Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation. 2003;107:664-666. 12. Scheffel H, Alkadhi H, Plass A. Accuracy of dual-source CT coronary angiography: First experience in a high pretest probability population without heart rate control. Eur Radiol. 2006;16:2739-2747. 13. Leschka S, Alkadhi H, Plass A. Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J. 2005;26(15):1482-1487. 14. Raff GL, Gallagher MJ, O Neill WW, Goldstein JA. Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol. 2005;46:552-557. 15. Zheng M W, Li J Y. Dual-source computed tomographic coronary angiography. Clinical Invest. 2009;36(2):117-124. 16. Dev D, Lee C J, Ohba M. Image quality and artifacts in coronary CT angiography with dual-source CT: initial experience. J Cardiovasc Computed Tomogr. 2008;2(2):105-114. 17. Rajab MI, Eskandar AA. Enhancement of radiographic images in patients with lung nodules. Thoracic Cancer. 2011;2(3):109-155. 18. Das KM, El-Menyar AA, Salam AM. Contrast-enhanced 64-section coronary multidetector CT angiography versus conventional coronary angiography for stent assessment. Radiology. 2007;245:424-432. Pak J Med Sci 2013 Vol. 29 No. 1 www.pjms.com.pk 111