Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Similar documents
Biomedical Instrumentation E. Blood Pressure

A GUIDE TO OUR NIBP TECHNOLOGY

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

JAMES R. KNIGHT

Blood Pressure Estimation Using Photoplethysmography (PPG)

1 Non-invasive measurement of arterial pressure

External Oscillatory Blood Pressure - EOBPTM

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

BP-600 Noninvasive Blood Pressure Sensor

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Blood Pressure Determination Using Analysis of Biosignals

Blood Pressure Laboratory

Permanent City Research Online URL:

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Hybrid EEG-HEG based Neurofeedback Device

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

WIRELESS PATIENT HEALTH MONITORING SYSTEM

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Quick Doc: Design of Wireless Blood Pressure Device. Doralice Ferreira

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

Estimation of Haemoglobin Using Non-Invasive Technique

Optical sensor for Indian Siddha Diagnosis System

EEL 6586, Project - Hearing Aids algorithms

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013

LABVIEW based expert system for Detection of heart abnormalities

Literature Review Aarthi Sivasankar

VENTRICULAR DEFIBRILLATOR

Automated Blood Pressure Monitor

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Heart Rate Monitoring System Using Lab View

5.2. MEASUREMENT OF BLOOD PRESSURE

HUMAN ANATOMY AND PHYSIOLOGY

PCA Enhanced Kalman Filter for ECG Denoising

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

Glabella. Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device. Christian Holz Edward Wang Microsoft Research

Heart Abnormality Detection Technique using PPG Signal

Lab 4: Introduction to Physiological Measurements - Cardiovascular

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Various Methods To Detect Respiration Rate From ECG Using LabVIEW

Adaptive Non-invasive Blood Pressure Measurement

Design of Remote Monitoring System for the Detection of Vital Signs

PHONOCARDIOGRAPHY (PCG)

AFFORDABLE TECHNOLOGY

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Heart Rate, Blood Pressure, and Exercise. Evaluation copy

Lab #3: Electrocardiogram (ECG / EKG)

ECG Acquisition System and its Analysis using MATLAB

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Blood Pressure and Exercise Lab

Raw BP and Eight Hemodynamic Measurements Displayed in Real Time

Research Journal of Pharmaceutical, Biological and Chemical Sciences

POWER EFFICIENT PROCESSOR FOR PREDICTING VENTRICULAR ARRHYTHMIA BASED ON ECG

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart

Blood Pressure Monitor PASESA

SONIMAGE HS1. Premium Portable Ultrasound

COMPUTER PLAY IN EDUCATIONAL THERAPY FOR CHILDREN WITH STUTTERING PROBLEM: HARDWARE SETUP AND INTERVENTION

How To Measure Vital Signs

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT

Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms

VLSI Implementation of the DWT based Arrhythmia Detection Architecture using Co- Simulation

A bioimpedance-based cardiovascular measurement system

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

iworx Sample Lab Experiment HC-5: Body Position, Exercise, and Cardiac Output

DETECTION OF HEART ABNORMALITIES USING LABVIEW

Feasibility Evaluation of a Novel Ultrasonic Method for Prosthetic Control ECE-492/3 Senior Design Project Fall 2011

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date:

Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating Arrhythmic Heart-Beat

Author's copy - produced at Lamar University

Measurement of Hemodynamic Parameters: Design of Methods and Hardware

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

Detection of Foot Ulcer using Pressure Kit M. Prem Anand 1, J. Sri Arunaa 2 1 Assistant Professor 2 M.E Student

Artificial Neural Networks in Cardiology - ECG Wave Analysis and Diagnosis Using Backpropagation Neural Networks

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Wrist Energy Fuzzy Assist Cancer Engine WE-FACE

Team 11: Device to Monitor and Control IHD for Painful Stiff Shoulder Treatment

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

Project: Feedback Systems for Alternative Treatment of Obstructive Sleep Apnea

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System!

Ultrasound in Medicine

INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

ISSN: Volume 2 Number 7 (July-2014) pp

Impedance Cardiography (ICG) Method, Technology and Validity

A Simple Portable ECG Monitor with IOT

II. PROCEDURE DESCRIPTION A. Normal Waveform from an Electrocardiogram Figure 1 shows two cycles of a normal ECG waveform.

Transcription:

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive method of measuring systolic and diastolic pressures with the use of an automated occluding cuff which results in valid estimates of mean pressure, but questionable estimates of systolic and diastolic pressure due to inappropriate algorithms and proper hardware. It is also sensitive to differences in pulse pressure and artery stiffness. To improve the accuracy of systolic and diastolic pressure using cuff pressure oscillation remains an open challenge. It can be made more deterministic by using LED s of proper wavelength considering the absorption of light by the blood, photodiodes, considering the elasticity of the cuff and its inflation and deflation parameters along with the selection of proper pressure sensor. The method involved to execute these is oscillometric method which is a more advanced method for blood pressure monitoring. This model reproduces the realistic cuff pressure oscillations. The systolic and diastolic pressures are then extracted using regression analysis in face of varying pulse pressure and arterial stiffness. Proper hardware and accurate input yields in précised and deterministic results. Keywords- Oscillometric, non-invasive, arterial stiffness, pulse pressure, automated cuff, regression analysis. I. INTRODUCTION Blood Pressure (BP), sometimes referred to as arterial blood pressure which is one of the principal vital signs is the pressure exerted by circulating blood upon the walls of blood vessels. The rate of mean blood flow is dependent on both, the BP and resistance to flow provided by the walls of blood vessels. BP is usually expressed in terms of systolic pressure (SP) over diastolic pressure (DP) and is measured in millimetres of mercury (Hg). The normal BP of an adult is 120/80 mm of Hg. This is 120mm of Hg over the atmospheric pressure of 760 mm of Hg. So, the overall pressure is 880 mm of Hg. The Mean Arterial Pressure (MAP) is the average over a cardiac cycle and is determined by the cardiac output (CO), system vascular resistance (SVR), and central venous pressure (CVP). Pulse Pressure up-down fluctuations of the arterial pressure results in a pulsatile nature of CO. It can be simply calculated from the difference of the measure of SP and DP. Non-invasive Blood Pressure measurements (NBP) are quick and easier to execute and also cause minimum stress to the patient. This technique uses an automated cuff and a pressure sensor to detect the pressure signal when the cuff is deflated. The pressure at which the first pressure signal is received is recorded as the SPand similarly DP is recorded with the last pressure signal. As with all the practical research devices, this device also has merits and some challenges which need to be faced and accordingly revamped, such as, (1) high motion tolerance of NBP devices, (2) the pressure in the cuff must be steadily scrutinized so as to avoid any mishap, (3) the devices need a yearly calibration for deterministic results, (4) some errors that are accounted which must be keenly guarded are, (i) the rapid deflation of the cuff, (ii) cuff s elasticity to restrain the rated pressure. 151

In this paper, a mathematical model for NBP measurement is presented which could capture the systolic and diastolic pressure as well as varying pulse pressure along with artery stiffness. Pressure sensors are designed for proper operations and the output response is compared with the known medical devices to verify SP and DP obtained through this device and measuring the accuracy and precision achieved. II. OSCILLOMETRIC METHOD The oscillometric method of BP measurement takes into account the intra-arterial pulses to the occluding cuff which surrounds the limb or artery (finger in this case). The measurement is done by placing the cuff upon the artery and inflating it till the blood flow in the artery stops. The pressure in the cuff is then released slowly in steps of 4-8mm of Hg using a pressure release valve. The blood starts flowing and its flow changes from turbulent to laminar creating a korotkoff sound, which is then measured by a pressure sensor. The data is then recorded using data acquisition system and resulting oscillogram is detected and processed at each pressure step. Fig.1. Oscillogram obtained by releasing the cuff pressure The pulsations or the korotkoff sound starts at the systolic pressure and ends at diastolic pressure while releasing the pressure from the cuff. Due to difficulties in measuring the SBP because of oscillating amplitudes, a filter of order 3, with a cut-off frequency of 50 Hz along with curve fitting of second order is used to filter the signal and noise and enabling the ripples in the passband and stopband thereby reducing the loss of signal power and data. III. MATHEMATICAL MODELLING A. Non-Invasive Blood Pressure Measurement model NIBP mathematical model involves inflation and deflation of the cuff by the use of motor for a précised time upto a pressure of 300 mm of Hg and this pressure needed to be controlled using a pressure switch which will cut-off the circuitry of the motor as soon as the pressure is reached to its set-point which in this case is 300 mm of Hg. As soon as the cuff starts deflating, blood start flowing in the arteries which in turn results in a turbulent flow also causing a kortokoff sound which is detected by the pressure sensor, and the pressure sensor gives the systolic pressure reading. Similarly, when the blood starts flowing freely with the full diameter of the artery, then its flow changes to laminar flow and that kortokoff sound stops, that reading is then sensed by the pressure 152

sensor is recorded as diastolic pressure. The mean arterial pressure (MAP) is then calculated which is the average over a cardiac cycle and is determined by cardiac output (CO), system vascular resistance and central venous pressure (CVP). (1) (2) The pulse pressure is then calculated from the difference of measure of systolic and diastolic pressure. B. Signal Conditioning of Input Signal The input signal obtained is a voltage signal which is converted to the pressure signal with the use of equations specified in the specification sheet of the sensor which is used to sense the input data. The signal obtained as the pressure signal is weak and noisy. An Infine Impulse Response (IIR) filtering technique is used to eliminate the noise. The filter is a combination of analog filter, its prototype and the digital filter. The filter we are using is the Inverse Chebyshev filter. The main feature of this approximation is the ripple in the stop-band. The passloss in this type of filter is a monotonic function of frequency, as in the case of Butterworth filters. Fig.2. Magnitude response of inverse chebyshev filter IV. GRAPHS &CIRCUIT The scaling frequency of inverse chebyshev filter is defined equal to stop-band edge frequency. The response of Inverse Chebyshev filter is given as A. Mean Arterial Pressure The MAP and Pulse pressure calculated from systolic and diastolic pressures is shown below in Fig. 3, the graph gives a clear overview and explanation of the above equations no. (2) and (3). 153 Fig. 3. Mean Arterial Pressure

B. Motor Driving Circuit An air pumping motor is used for inflating the cuff and creating the required pressure which is to be sensed by the pressure sensor or a LED- Photodiode coupler. This circuit is used for efficient working of the motor keeping in mind the current rating. It uses two MOSFET s that are n channel and p channel along with IRF510 and diode IN4007 which resists the motor to operate in one direction only. C. Device Circuitry It is a combination of NI myrio, which is a microprocessor based system along with FPGA i.e. field programmable gate array, thereby, making the complete device standalone and provides high determinism.the device is then connected to various digital and analog inputs or outputs such as the motor, a circuitry of infrared led and photodiode along with the pressure valve and a LCD display. The input as shown in Fig. 3, is provided by the device to the motor for its operation. It is a pulse width modulation signal commonly known as PWM signal for a period of time required to inflate the cuff to a pressure of 300 mm of Hg. Fig. 4. Sample PWM Signal As shown in Fig. 2, the device is programmed using the graphical programming using NI LabVIEW, and the program is then dumped into NI myrio s memory. Fig. 5. Sample LabVIEW Block diagram Window The wavelength of LED is selected on the basis of the diameter of the red blood cell which is 7-8 microns and size of the artery which is 2mm in diameter. The wavelength absorbed by the red blood cells lies in the IR region and the suitable photodiode is selected to generate the output voltage which is then calibrated using proper specification sheet and the algorithm. The pressure valve used deflates the cuff and allows the blood to flow again which creates a pressure signal that is received by the photodiodeand the value so obtained is compared to the data sheet and the diastolic and systolic pressures are obtained. The wavelength so chosen is then changed according to resolution to obtain more accurate readings and reducing the scanning timeand moving towards beat to beat pressure measurement. Proper filters are used with their cut-off frequencies along with the feedback so as to minimize the errors and curve fitting of second order is carried out for the proper plotting of the waveform generated through the photodiode. 154

To obtain the value of systolic and diastolic pressure, the waveform is compared by overlapping the waveform over an ECG waveform or a blood pressure waveform obtained from any other device using pressure sensor to calculate the peak heights which in turn will determine the value of SBP and DBP. The values are stored and recorded to calculate a mean error which is obtained as the difference in readings of syphygmo-feno manometer and the normal blood pressure meter. Fig.6. In-beam Prototype V. CONSEQUENCES There is an alarming need of highly accurate and precised Blood Pressure monitors to substantiate the monitoring done through mercury whose accuracy is also not good and it also cannot be used by an average person. The BP monitor market is growing rapidly due to increasing number of hypotensive and hypertensive patients, who require time to time BP monitoring. WHO has confirmed that approximately 12.8% of total deaths in a year occur due to BP problems and its improper or careless monitoring. There is an alarming need of a product through which we can monitor the blood pressure with high accuracy and high precision which is commonly called as deterministic property of the product, along with it, this product also provides high comfort level to the patients because the inflation and deflation of the cuff on the arm sometimes causes irritation, so shifting the cuff from arm to second phalynx will be a huge improvement along with improved scanning time and its portable nature. Therefore, this device will provide a great market, finding its efficient use in ICU s, CCU s along with the high determinism, high comfort level, alarms, emergency stop buttons which will make the device very much handy and will not ask for too much knowledge for its operation. The LCD used in the device makes the device even more user friendly providing a great user interface. As with all the practical research devices, this device also has merits and some challenges which need to be faced and accordingly revamped, such as, (1) high motion tolerance of NBP devices, (2) the pressure in the cuff must be steadily scrutinized so as to avoid any mishap, (3) the devices need a yearly calibration for deterministic results, (4) some errors that are accounted which must be keenly guarded are, (i) the rapid deflation of the cuff, (ii) cuff s elasticity to restrain the rated pressure. VI. CONCLUSION The method involved to execute these, is an oscillometric method which is a more advanced method for blood pressure monitoring. This model reproduces the realistic cuff pressure oscillations. The systolic and diastolic pressures are then extracted using regression analysis in face of varying pulse 155

pressure and arterial stiffness. Proper hardware and accurate input yields in précised and deterministic results. ACKNOWLEDGMENT I would like to thank my parents for their constant support and blessings.i would also like to dedicate a sincere thanks to our guide Mr. Sandeep Panwar Jogi for his useful contribution in the paper by providing me the proper guidance and knowledge. REFRENCES [1]. GRAPHIC COMM. CENTRAL, Definition of Graphic Communications,2008, accessed Feb 25, 2009. [2]. K. Elleithy (ed.), Advances and innovations in systems, computing sciences and software engineering. [3]. National Instruments, White paper on NI myrio. [4]. Pulse Width Modulation [5]. NI myrio Handbook, www.ni.com/myrio [6]. Saif Ahmad, IEEE Transactions on Biomedical Engineering, Electrocardiogram- Assisted blood pressure estimation, Accessed March 2012 [7]. Matheonics Technology, Inverse Chebyshev approximation method for designing of filters. [8]. EulaliaBalestrieri, IEEE Transactions on Instrumentation and Measurement, Instruments and methods for calibration of oscillometric blood pressure measurement devices Accessed September 2010 [9]. Saif Ahmad, IEEE Transactions on Instrumentation and Measurement, Measurement of heart rate variability using an oscillometric blood pressure monitor Accessed October 2010 [10]. Wendy Van Moer, IEEE Transactions on Instrumentation and Measurement, Linearizing oscillometric blood pressure measurements [11]. Ya-Li Zheng, IEEE Transactions on Biomedical Engineering, An armband wearable for overnight and cuff-less blood pressure measurement 156