Patient Safety During Cardiac Catheterization

Similar documents
Biomedical Instrumentation

The Influence of Electromagnetic

Potassium Efflux from Myocardial Cells Induced by Defibrillator Shock

A Mechanism for "False" Inhibition of Demand Pacemakers

Effects of myocardial infarction on catheter defibrillation threshold

Cardiovascular Collapse Caused by Electrocardiographically Silent 60-Hz Intracardiac Leakage Current. Implications for Electrical Safety

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

08-Electrical Safety. EE570 Energy Utilization & Conservation Professor Henry Louie

APPLICATIONS OBJECTIVES PRE-LECTURE

A New Technique for Repeated Measurement of Cardiac Output During Cardiopulmonary Resuscitation

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

An Official Journal of the American Heart

Transcoronary Chemical Ablation of Atrioventricular Conduction

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Biomedical Instrumentation

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Efficacy and safety of the reciprocal pulse defibrillator current waveform

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

Electrodes (for TransAeris System)

VALLIAMMAI ENGINEERING COLLEGE

Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current

Part 1: General requirements

Endotracheal Versus Intravenous Epinephrine During Electromechanical Dissociation with CPR in Dogs

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Citation Acta medica Nagasakiensia. 1984, 29

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent?

High Voltage Safety. Dangers of Electricity. W. Maes. Department of Marine Engineering Antwerp Maritime Academy HV, 2015

The Heart 1 of 34 Boardworks Ltd 2012

Percutaneous atrial septal defect closure with the Occlutech Figulla Flex ASD Occluder.

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

The Normal Electrocardiogram

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

Pacing Lead Implant Testing. Document Identifier

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

CT angiography techniques. Boot camp

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

Model 5392 EPG Temporary Pacer

Effects of Pericardial Effusates of Various Conductivities on Body-Surface Potentials in Dogs

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER

Ventricular Arrhythmias

Acute dissections of the descending thoracic aorta (Debakey

Prevention of Acetylcholine-Induced Atrial Fibrillation. Shigetoshi CHIBA, M.D. and Koroku HASHIMOTO, M.D.

Cardiac Electrical Therapies. By Omar AL-Rawajfah, PhD, RN

2

Julie A. Thompson, MSE, and Janice M. Jenkins, PhD

ICD Implantation Patient Information

SCOPE OF PRACTICE PGY-4 PGY-6 (or PGY-5 PGY-7 if Medicine/Pediatrics resident)

Cardiothoracic Fellow Expectations Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center

SARASOTA MEMORIAL HOSPITAL

Biphasic Clinical Summaries

WHILE it is generally agreed that elevation

Diagnostic approach to heart disease

Technique. Technique. Technique. Monitoring 1. Local anesthetic? Aseptic technique Hyper-extend (if radial)

TASER and In-Custody Deaths

Analysis of Human Cardiovascular System using Equivalent Electronic System

L: Cardiovascular. Saskatchewan Association of Licensed Practical Nurses, Competency Profile for LPNs, 3rd Ed. 107

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition

CITY OF WAUPACA, WISCONSIN WATER DEPARTMENT WATER SERVICE LATERAL THAWING RULES AND PROCEDURES (prepared February 24, 2015)

Pacemaker and AV Node Ablation Patient Information

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS

RECOMMENDED INSTRUCTIONS FOR USE

Patient Resources: Arrhythmias and Congenital Heart Disease

Chapter 20: Test Administration and Interpretation

Version Effective date Changes Prepared By CPR + AED

University of Wisconsin - Madison Cardiovascular Medicine Fellowship Program UW CICU Rotation Goals and Objectives

Understanding Atrial Fibrillation A guide for patients

V4, V5 and V6 follow the 5 th intercostal space and are NOT horizontal as indicated in the image. Page 1 of 7

ECG SENSOR ML84M USER S GUIDE. CENTRE FOR MICROCOMPUTER APPLICATIONS

Cardiac Arrhythmias Simulated by Concealed Bundle of His Extrasystoles in the Dog

Radioelectrocardiography in Diagnosis of Rhythm

Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh

Shock-induced termination of cardiac arrhythmias

Clinical Data Summary: Avoid FFS Study

DOWNLOAD OR READ : VALIDITY OF CARDIAC IMPLANTABLE ELECTRONIC DEVICES IN ASSESSING DAILY PHYSICAL ACTIVITY PDF EBOOK EPUB MOBI

Design of Low-Power CMOS Cell Structures Using Subthreshold Conduction Region

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

Autopsy Errors with Electronic Control Devices: A Cardiovascular Perspective

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

Mitral Regurgitation

ER REBOA Catheter. Instructions for Use

Mechanics of Cath Lab Support Devices

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals

MODERATOR Felix Rapaport, other members of this

ELECTROCARDIOGRAPHY (ECG)

The Facts about Biphasic Defibrillation

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

About your tunnelled dialysis catheter. Information for patients Sheffield Kidney Institute (Renal Unit)

Shock-induced termination of cardiac arrhythmias

HealthCare Training Service

CORESTA RECOMMENDED METHOD N 22

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

Cardiovascular Effects of Exercise. Background Cardiac function

Cardiovascular System

Transcription:

119 CIRCULATION VOL 58, No 6, DECEMBER 1978 32. Neff MS, Kim KE, Persoff M, Onesti G, Swartz C: Hemodynamics of uremic anemia. Circulation 43: 876, 1971 33. Cangiano JL, Ramirez-Mux, Ramirez-Gonzalez R, Trevino A, Campos JA: Normal renin uremic hypertension. Arch Intern Med 136: 17, 1976 34. Capelli JP, Kasparian H: Cardiac work demands and left ventricular function in end-stage renal disease. Ann Intern Med 86: 261, 1977 35. Rubin JE, Friedman EA: Dialysis and transplantation of diabetics in the United States. Nephron 18: 39, 1977 36. Weinrauch LA, Healy RW, Leland OS Jr, Goldstein HH, Kassissieh SD, Libertino JA, Takacs FJ, D'Elia JA: Coronary angiography and acute renal failure in diabetic azotemic nephropathy. Ann Intern Med 86: 56, 1977 A Conductive Catheter to Improve Patient Safety During Cardiac Catheterization MARTIN J. LIPTON, M.D., ALLEN K. REAM, M.D., AND BRUCE H. HYNDMAN, M.S. Downloaded from http://ahajournals.org by on October 9, 218 SUMMARY A 6 Hz current, as small as 2,uA (rms) is capable of causing ventricular fibrillation when directly applied to the heart. Significant cost and engineering effort has been spent to construct monitoring equipment which satisfies the safety regulations requiring maximum leakage currents below this value. Patients undergoing cardiac catheterization are particularly at risk from electrical hazards, primarily because catheters are made from nonconductive materials. A conductive catheter should allow externally applied currents to leak through its walls before reaching the catheter tip. A new electrically conductive catheter was compared with a standard nonconductive catheter. Five dogs were studied, with 81 attempts to cause fibrillation. Sixty-hertz voltage between the catheter and an external electrode was increased until fibrillation occurred or 13 V was reached. Eight states were studied in randomized sequence: conductive or nonconductive catheter, guidewire or saline-filled and tip touching wall, or free in left ventricle (verified by fluoroscopy and cineangiography). The saline-filled conductive catheter was safer in that fibrillation never occurred, while fibrillation nearly always occurred with the nonconductive catheter. A conductive guidewire negates the protection of the conductive catheter. The application of conductive catheters could reduce instrumentation costs in laboratories and intensive care units and improve patient safety. ADVANCES IN BIOMEDICAL ENGINEERING have resulted in a progressive increase in the number and complexity of hospital electrical monitoring devices. These devices are particularly available in specialized centers, such as intensive care and cardiac catheterization units. Such devices are important in the provision of health care, but have brought about an increased danger of accidental electrocution. This hazard has received attention in the past few years, but the incidence of electrical complications in hospitals remains a very controversial issue. Such electrocutions may be far more common than realized,1 2 From the Departments of Radiology, Anesthesiology and Clinical Engineering, Stanford University Medical Center, Palo Alto, California, and Palo Alto Veterans Administration Hospital. Supported in part by a gift from Medtronic, Inc., Minneapolis, Minnesota and the research service of the Veterans Administration Hospital, Palo Alto, California. Presented in part at the 49th Annual Scientific Session of the American Heart Association, Miami Beach, Florida, November 1976. Dr. Lipton is the recipient of a US Public Health Service Research Career Development Award, Grant 1K4 HL36-1 from the National Heart, Lung, and Blood Institute. Address for reprints: Martin J. Lipton, M.D., Department of Radiology, M-38, University of California, San Francisco, California 94143. Received May, 8, 1978; revision accepted August 17, 1978. Circulation 58, No. 6, 1978. because proof of such incidents is very difficult to obtain. However, incidents of ventricular fibrillation and hospital deaths from electrocutions have been documented during cardiac catherization and also in patients with myocardial pacemakers.3" Studies have shown that 6 Hz alternating currents of as little as 8 MA, with typical values below 1 ma, can produce ventricular fibrillation in humans."2-15 Considerable effort has been spent to construct monitoring equipment which draws maximum leakage currents of less than the established safety standard threshold of 1 MA.'9 2 The increasing importance of such safety is immeasurable. One common opportunity for fibrillation with minimum current is provided by intracardiac catheters used during cardioangiographic studies and monitoring.2 15, 21 Macroshock and Microshock There are two modes of possible electric shock in hospitals. Macroshock is due to externally applied currents and can cause ventricular fibrillation with currents > 1 ma. Hence, direct or commercial alternating power frequencies with 75-12 V are dangerous. All hospital patients may be exposed to such macroshock hazards, just as they would be in their own homes - for example, from a defective electrical appliance.

CONDUCTIVE CATHETERS/Lipton et al. 1 191 Downloaded from http://ahajournals.org by on October 9, 218 Catheter Ground FIGURE 1. Diagram illustrating an electrically sensitive patient. The current path is provided by a catheter's contents, normal saline or a guidewire placed from the arm, with its tip lying within the heart. The current source is provided by a piece ofelectrical equipment, such as an ECG machine or other monitoring device. A return pathway to the source, such as an electrode wire, is necessary for current to flow. Microshock results from small-magnitude currents (MA) which involve electrode contact within the body and, in particular, the heart. The two requirements for the conduction of electric current through the heart are a current path and a current source. A pathway is provided by a saline- or other fluid-filled cardiac catheter or pacemaker electrode. The walls of cardiac catheters are excellent insulators of electricity. The current source is most often a leakage from electrical equipment, a static discharge, or currents due to physical contact with the catheter by a nurse or physician. Currents as small as 5,A produced by voltage levels as small as 5 mv can produce ventricular fibrillation under these circumstances. Such an electrically sensitive patient is schematically represented in figure 1. Most safety requirements limit the amount of current which could flow into the patient. This problem would not exist, however, if a method could be devised to prevent the current from reaching the heart via the catheter.22 Methods and Materials A new conductive catheter material developed by Medtronic, Inc., of Minneapolis, Minnesota, incorporates polyvinylchloride and carbon black with added barium sulfate for radiopacity. The material can be extruded to form standard-sized catheters. Such a catheter with a conductive wall should enhance patient protection against electrically induced ventricular fibrillation by leaking away externally applied current before it reaches the catheter tip. Mongrel dogs weighing 15-25 kg, were anesthetized with sodium pentobarbital (25 mg/kg body weight). Intravenous supplemental doses were given to assure a depth of anesthesia permitting stable spontaneous ventilation. The experimental sequence lasted 2-4 hours. An intravenous catheter was maintained with 5% dextrose in water. A catheter was placed in the left femoral artery, and mean pressure was measured via an aneroid manometer. Typical mean blood pressure was 13 torr. The ECG was monitored via external electrodes on the shaved skin. The dogs were intubated with a cuffed endotracheal tube and were permitted to breathe spontaneously in room air. Controlled ventilation for four to eight breaths using 1% oxygen was imposed after defibrillation. The current path was completed through a subcutaneous electrode in the left hind limb of the dog. This terminal was connected to a variable current source, as shown in figure 2. The other side of the NCTION FIGURE 2. Schematic representation of JERATOR electrical connections employed in in vivo o studies. Current was applied via a 6-Hz voltage applied between the external end of the catheter (through a metal luer connector) and a subcutaneous electrode in the left ero hind leg. Current delivered to the animal A\PLIFIER was varied by altering the output ofthefunc- 111 tion generator. The multimeter was used to monitor delivered current. None of the instrument connections shown were connected to powerline ground or chassis ground.

1192 CIRCULATION ~~~~~~~~~~VOL 58, No 6, DECEMBER 1978 Downloaded from http://ahajournals.org by on October 9, 218 current source was connected to the proximal end of the catheter. The current level applied to the catheter could be controlled and monitored using a digital multimeter (model 3, Data Technology). The catheter was introduced through a right femoral artery approach and advanced until the tip position was satisfactory, as verified by fluoroscopy and cineangiography in two planes. Two tip positions were studied: free in the left ventricle, and wedged against the left venticular wall. When spontaneous arrhythmias were observed with wall contact, the catheter tip was moved to a new location before applying voltage to the catheter. The catheters were filled with normal saline (.9% NaCl) or with a guidewire. Two #7 French 1 cm catheters were used: a conductive catheter (Medtronics model 3X5A-39 diagnostic catheter) and a standard nonconductive catheter (Cook Inc. polyethylene catheter, model 6673-2). Sterile catheters were used in each animal and for each bench study. Baseline pressures and ECG recordings were obtained, and then the current was steadily increased for each event. Fibrillation was induced via a 6 Hz voltage applied between the external end of the catheter and the return path through the left hind limb. The applied voltage was slowly increased with continuous measurement of applied current until fibrillation occurred (as determined by ECG activity and fall in mean blood pressure) or until a maximum of 13 V was reached. After fibrillation current was noted, the current was immediately disconnected and the heart defibrillated, using a Hewlett-Packard DC defibrillator with levels set in a range between 4-1 watt/sec stored. The next episode was initiated after blood pressure, cardiac rhythm, and ventilation had returned to normal. Defibrillation was usually successful on the first attempt. Eight states were studied in a randomized sequence: conductive or nonconductive catheter, with the catheter tip free in the ventricle or against one of its walls, and saline or guidewire filled. Eighty-one attempts to cause fibrillation were made in five dogs. The catheters were examined in vitro and demonstrated to conform to theory. Previously unused catheters were used in each animal and also for each bench study. TABLE 1. Incidence of Ventricular Fibrillation (Incidents/ Attempts) for the Possible Combination of Catheter Composition, Contents and Tip Location Tip free Tip touching wall A) Catheter containing normal saline Conductive /1 /1 Nonconductive 4/11 1/1 B) Catheter containing metal guidewire Conductive 9/9 1/1 Nonconductive 1/1 11/11 Results Table IA shows the results for the saline-filled catheters as the number of fibrillations per number of attempts with a saline-filled catheter. Table lb shows the same results for the catheters with guidewire. Table 2 presents the fibrillation current threshold average in gaa for each of the eight states studied. Episodes in which fibrillation did not occur are not included in the averages. While these catheters did not exhibit ideal memory and torque control, the necessary modifications appear feasible. The catheters have been evaluated for thrombogenicity and appear satisfactory. Discussion A schematic diagram of a presently used nonconductive catheter is illustrated in figure 3. The current entering the catheter remains within the lumen because the catheter wall is an insulator, and the current emerges distally at the exit hole. The resistance of the patient and the wires is so low compared with the resistance of the catheter under normal conditions that one can ignore all other resistance values. A schematic diagram for the conductive catheter is shown in figure 4. In this case, the electric current can leak away through the catheter wall. The current leaks away more rapidly near the skin because the current is the result of voltage, and the voltage is highest near the skin and falls off toward the catheter's tip; the form of the decay is approximately exponential. Hence, very small current values can be rapidly achieved. The value of lumen current at a distance x from the catheter tip is represented by the equation: i(x) where R longitudinal )1/2 R(i. )J cosh(kx) catheter resistance, G lateral catheter conductance, K VRU, total catheter length, L catheter length beneath skin, Vmx source voltage, i(x) current in catheter at point x ( x 1), and i () current at catheter tip. TABLE 2. Average Fibrillation Current (SD) in Microamperes for Each State Tip free Tip touching wall A) Catheter containing normal saline Conductive Nonconductive 133 (66) 71 (51) B) Catheter containing metal guidewire Conductive 3,71 ( 3,9) 13,15 (26,38) Nonconductive 9,69 (1,91) 9,65 (2,57)

CONDUCTIVE CATHETERS/Lipton et al. 1193 :... :::::...: Catheter Wall vw Electric Current ::::::::::... :: i: fi: : ::.:-Skin FIGURE 3. Schematic representation of a nonconductive catheter. A regular nonconductive catheter is seen entering the tissues. All the current traveling within the saline-filled lumen emerges at the tip, unable to leak through its insulated walls. Downloaded from http://ahajournals.org by on October 9, 218 Tip Current C sinh (KL) where C expresses the reduction in current due to the resistance of the external portion of the catheter and internal catheter characteristics. The current, therefore, leaks away along the length of the catheter. Furthermore, because leakage current is proportional to the driving voltage (which decreases along the catheter), leakage current is largest near the skin and least near the tip, providing additional protection. In practical terms, the importance of this is shown in figure 5, where catheter length is plotted against skin resistance. Two results are graphically represented, one for the nonconductive and one for the conductive catheter. The curve for the conductive catheter shows that resistance becomes constant after the length exceeds a certain value. This means that the tip current is negligible compared with the total. This is strikingly different from the normal, nonconducting catheter where resistance rises and all the current must pass through the tip. This is better illustrated by plotting the logarithm of tip current against catheter length (fig. 6); this curve is obtained from the equations given earlier. After a minimum length, the graph becomes a straight line, which means that for a given catheter a distance can be specified such that increasing the catheter by that length decreases tip current by a power of 1, no matter how long the catheter. This result is based upon the assumption that G, the conductivity of the catheter wall, is constant, which is true for the clinical range studied experimentally. Normal saline was used within the catheter because this has the greatest conductivity of any common in-.......... Elcti.Cu..rre..n.t.............. FIG RE ~~~...... ductivewall and rapidly diminishes... iv et r. ur as ak hr 2l sen toeme...f.m.he.t...ichlie.witin.he.hart th

1194 CIRCULATION VOL 58, No 6, DECEMBER 1978 1 NON-CONDUCTIVE 8 a) u c -6 a) a) 4-' a1) c- -4-' - 6 4 2- CONDUCTIVE SALINE FILLED CATHETER 1 2 3 4 5 Downloaded from http://ahajournals.org by on October 9, 218 Length Beneath Skin (cm) FIGURE 5. The relationship of catheter length beneath the skin to catheter resistance. The resistance (R) (thousandths ofohms), seen at the skin is plotted on the y axis against length, 1, beneath the skin (cm) on the x axis. The curve for the conductive catheter shows that resistance becomes constant after the length exceeds a certain value; hence, tip current is then negligible compared with the total. The resistance for a conventional catheter progressively rises and all the current must pass through the tip. L- a1) a) Q E 3 L.6 a, Q.. ~ )". 4- x w a1) CO) 4- cn 1 1-1 - Theory: GConstant Data: Single Catheter I.II i~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~ I I I I I I I 1 2 3 4 5 Length Beneath Skin (cm) FIGURE 6. The relationship between tip current and length of immersed catheter for a fixed applied voltage. The logarithm of tip current is plotted on the ordinate and the length beneath the skin is given on the abscissa. The solid line represents the theoretical calculation. The points are direct measurements taken from an actual conductive catheter. The dotted line is 2gA threshold.

CONDUCTIVE CATHETERS/Lipton et al. 1195 Downloaded from http://ahajournals.org by on October 9, 218 travenous solution. One hundred thirty volts was chosen as a practical maximum in our animal studies because the highest voltage to which a hospital patient is likely to be exposed is powerline, which is 12 V. The condition of greatest clinical interest is the unanesthetized human subject. This study involved anesthetized dogs, but it has been suggested that, at least in one dog, there is no significant difference between the conscious and anesthetized fibrillation thresholds using sodium pentobarbital.23 Furthermore, the intracardiac fibrillation thresholds for the dog appear similar to the threshold in man,13 and may even be smaller.'2' 18, 24 Details of the mathematical analysis of the catheters have been documented.25 Conclusion The data from these experiments have shown that a suitably designed conductive catheter can substantially reduce patient risk of microshock-induced ventricular fibrillation. As the fibrillation threshold for patients could be as low as 2 ata,2 a catheter which only allows a tip current lower than this amount should be safe. It is possible to design a saline-filled catheter for any specified maximum voltage which, for practical purposes, will not electrically cause fibrillation when the tip is within the heart. The specifications for such a catheter can be met easily for both infants and adults, so that the dimensions are similar to those catheters presently in routine clinical use. As expected, the safety of a conductive catheter is negated by a highly conductive guidewire (table 1B), although in practice this lies within the heart for a much shorter time than the catheter, and often the guidewire is not placed within the heart. Ideally, conductive catheters should be used with nonconductive guidewires; their manufacture should not be technically difficult. If conductive catheters could replace presently used nonconductive ones, they should assure better protection against cardiac fibrillation from failed or noncurrentlimited electrical sources. References 1. Dalziel CF: Lethal electrical currents. IEEE Spectrum 6: 44, 1969 2. Roy OZ: Hz ventricular fibrillation and rhythm thresholds and the nonpacing intracardiac catheter. Med Biol Eng 13: 228, 1975 3. Medicine and the law: fatal shock from a cardiac monitor. Commentary. Lancet 1: 872, 196 4. Furman S, Schwedel JB, Robinson G, Hurwitt ES: Use of an intracardiac pacemaker in the control of heart block. Surgery 49: 98, 1961 5. Noordijk JA, Oey FTI, Tebra W: Myocardial electrodes and the danger of venticular fibrillation. Lancet 1: 975, 1961 6. Pengelly LD, Klassen GA: Myocardial electrodes and the danger of ventricular fibrillation. Lancet 1: 1234, 1961 7. Bousvaros GA, Conway D, Hopps JA: An electrical hazard of selective angiocardiography. Can Med Assoc J 87: 286, 1962 8. Lee WR: The nature and management of electric shock. Br J Anaesth 36: 572, 1964 9. Bruner JMR: Hazards of electrical apparatus. Anesthesiology 28: 396, 1967 1. Lee WR: The hazards of electrocution during patient monitoring. Postgrad Med 46: 355, 197 11. Kugelberg J: Accidental ventricular fibrillation of the human heart. Scand J Thorac Cardiovasc Surg 9: 133, 1975 12. Geddes LA, Baker LE: Response to passage of electric current through the body. Med Instrum 5: 13, 1971 13. Raftery EB, Green H, Gregory I: Electrical safety: fibrillation thresholds with 5 Hz leakage currents in man and animals. Br Heart J 35: 864, 1973 14. Starmer CF, Whalen RE: Current density and electrically induced ventricular fibrillation. Med Instrum 7: 158, 1973 15. Lee WR, Scott JR: Thresholds of fibrillating leakage currents along intracardiac catheters: an experimental study. Cardiovasc Res 7: 495, 1973 16. Cairns JA, Holder DA: Ventricular fibrillation due to the passage of a Swan-Gantz catheter. Am J Cardiol 35: 589, 1975 17. Snider D: Electrical threshold of cardiac fibrillation in humans. Med Instrum 7: 8, 1973 18. Watson AB, Wright JS, Loughman J: Electrical thresholds for ventricular fibrillation in man. Med J Aust 1: 1179, 1973 19. Green HL, Raftery EB, Gregory IC: Ventricular fibrillation threshold of healthy dogs to 5-Hz current in relation to earth leakage currents of electromedical equipment. Biomed Eng 7: 48, 1972 2. Editorial: leakage current limits. Health Devices 2: 31, 1972 R. 1974, pp 21-23 21. Weinberg DI, Artley JL, Whalen RE, McIntosh HD: Electric shock hazards in cardiac catherization. Circ Res 11: 14, 1962 22. Kahn AR, Stroebel FA: Catheter apparatus. U.S. Patent 3,659,588, 1972 23. McIntosh HD, Starmer F, Whalen RE: A comparison of the electrical ventricular fibrillation threshold with and without anesthesia. Am Heart J 72: 419, 1966 24. Kugelberg J: Accidental ventricular fibrillation of the human heart. Scand J Thorac Cardiovasc Surg 9: 133, 1975 25. Ream AK, Lipton MJ, Hyndman BH: Reduced risk of cardiac fibrillation with use of a conductive catheter. Ann Biomed Eng 5: 287, 1977