Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Similar documents
Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Blood Pressure Estimation Using Photoplethysmography (PPG)

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

Optical sensor for Indian Siddha Diagnosis System

Biomedical Instrumentation E. Blood Pressure

DEVELOPMENT OF NON-INVASIVE STEERING-TYPE BLOOD PRESSURE SENSOR FOR DRIVER STATE DETECTION. Received December 2017; revised April 2018

Estimation of Haemoglobin Using Non-Invasive Technique

Heart Abnormality Detection Technique using PPG Signal

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Blood Pressure Determination Using Analysis of Biosignals

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

MCP PMT with high time response and linear output current for neutron time-of-flight detectors

Vital Signs. Vital Signs. Vital Signs

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG

Hybrid EEG-HEG based Neurofeedback Device

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Design and Dynamic Modeling of Flexible Rehabilitation Mechanical Glove

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

Blood Pressure Laboratory

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Optical Sensor System for Hemoglobin Measurement

Author's copy - produced at Lamar University

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

External Oscillatory Blood Pressure - EOBPTM

Design of the HRV Analysis System Based on AD8232

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

Permanent City Research Online URL:

PART I: HEART ANATOMY

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion

MX3 Plus English. Introduction. Table of contents. 1 How to obtain meaningful blood pressure readings

LAB 9: Metabolic Rates

APG (Accelerated Photoplethysmography) clinical manual

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

A New Approach on Digital Blood Pressure Measurement Method for u-healthcare Systems

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

Measurement of Cuffless Blood Pressure by Using a Magnetoplethysmogram Pulsimeter

Chapter 15: Measuring Height, Weight, and Vital Signs. Copyright 2012 Wolters Kluwer Health Lippincott Williams & Wilkins

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Competitive Intelligence Report Technology & Patents

Smart Wearable Body Sensors for Glucose Monitoring

Design of Remote Monitoring System for the Detection of Vital Signs

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood

QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION

M1 Classic. Digital Blood Pressure Monitor. Instruction manual page 1 SYS. mmhg DIA. mmhg. PULSE/min

Choosing Mindray SpO2

How To Measure Vital Signs

Data Collection Worksheet

AFFORDABLE TECHNOLOGY

IB TOPIC 6.2 THE BLOOD SYSTEM

Chapter 12 - Vital_Signs_and_Monitoring_Devices

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Make a Smart IT & BT Devices for Future Warning for Heart Failure, Hypoglycemia, and Hyperglycemia Smart Watch. LP Lab LP Watch

Special Lecture 10/28/2012

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR

Physiology of the Circulatory System modified from

World s Fastest Ankle-Brachial Index Screening Device

Clinical Research Coordinator Skills Program

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS

EGR1200 Freshman Mini Project Biomedical Signal Analysis

Impedance Cardiography (ICG) Method, Technology and Validity

IB TOPIC 6.2 THE BLOOD SYSTEM

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System!

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS

Keun-Ho Kim and Sang-Suk Lee* Department of Oriental Biomedical Engineering, Sangji University, Wonju , Korea

Results of geometric ballistocardiography

Lab #10 Physiology of the Circulatory System

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

Wrist Energy Fuzzy Assist Cancer Engine WE-FACE

Female Populations. By: Kelsey Joyce. Group Members: Sooriya Jhanagan, Nelson Ferreira, Lindsay Hummell. TA: Jingjie Hu. Biology 240W Section 002

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM

Physical Fitness Biology Concepts of Biology 13.1

and Technology I. INTRODUCTION METHOD & MATERIAL

Literature Review Aarthi Sivasankar

Lab #3: Electrocardiogram (ECG / EKG)

Cardiac Output Monitoring - 6

Blood Pressure Competency Training

TEPZZ _ 849 A_T EP A1 (19) (11) EP A1. (12) EUROPEAN PATENT APPLICATION published in accordance with Art.

Development of an Electronic Glove with Voice Output for Finger Posture Recognition

a central pulse located at the apex of the heart Apical pulse Apical-radial pulse a complete absence of respirations Apnea

The Circulatory System

Chapter 24 Vital Signs. Copyright 2011 Wolters Kluwer Health Lippincott Williams & Wilkins

EFFECTS OF MECHANICAL PUMPING ON THE ARTERIAL PULSE WAVE VELOCITY : PERIPHERAL ARTERY AND MICRO-VESSELS

EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION DANIEL JOHN BADGER. A thesis submitted to the

Application of PPG Signal Processing to Patients undergoing Angioplasty Procedure

A non-invasive tongue controlled wireless assistive device for quadriplegics.

Glabella. Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device. Christian Holz Edward Wang Microsoft Research

A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD

World s Fastest Ankle-Brachial Index Screening Device

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device

Non-invasive Hemoglobin Measurement for Anemia Diagnosis

Transcription:

Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062 View the article online for updates and enhancements. This content was downloaded from IP address 148.251.232.83 on 10/10/2018 at 06:17

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Valerie Tan 1, Audrey Huong 1, Xavier Ngu 1 1 Faculty of Electrical and Electronic Engineering University Tun Hussein Onn Malaysia E-mail: audrey@uthm.edu.my Abstract. This paper aims to demonstrate the use of a finger-cuff system in the approximation of one's systolic and diastolic value based on the acquired heart pulse data. The prediction of the required systolic and diastolic value of blood pressure is via the use of two-element Windkessel Model. The results are compared for three groups of individuals of different clinically measured blood pressure range. A comparison with the result given from clinically available arm-cuff blood pressure monitor revealed a relatively low mean difference of 1.6 ± 3 mmhg predicted for the systolic values, while a considerably larger difference of 25.3 ± 6.3 mmhg is found for the estimated diastolic values. This work concluded that the proposed strategy that used heart pulse in its analysis may be used as an alternative means for continuous measurement of one s blood pressure, but a system with high sampling resolution may be required to improve the accuracy of the predicted value. 1. Introduction Continuous measurement of blood pressure is vital especially for aged population, pregnant women and individuals who either have just undergone surgery or with critical medical conditions. It is crucial to continuously monitor the blood pressure of these groups so that prompt action can be taken following the detected sudden change in their blood pressure as the latter may result in death. The conventional sphygmomanometer technique, which required the use of a blood pressure cuff and a stethoscope, is inconvenient for continuous measurement of blood pressure. The arm-cuff blood pressure monitor is a more convenient alternative as it provides continuous measurement of the value and it does not require the use of a stethoscope. Nonetheless both of these approaches may potentially cause harm especially to those who suffered from hypertension and hypotension. The exertion of pressure on the arm of these populations via the use of a cuff narrows the size of local blood vessels; this increases the chances of one getting stroke due to a possible insufficiency in the supply of oxygen to the brain. Therefore, cuffless Photoplethysmography (PPG) [1] and the use of a finger-cuff [2] were previously introduced to overcome these problems. Both of these techniques are wearable technologies that do not require the occlusion of artery in their measurement. While the implementation of the latter technique is limited to fingertip, the former may include a wide range of wearable devices such as wrist-watch, ring and clothing [3, 4]. PPG is a non-invasive optical technology that measures relative blood volume changes in vessels close to the skin surface. Many works have been done in the past to predict one s blood pressure using the PPG data and via the use of Windkessel model [5, 6], wherein it was reported that the accuracy of value predicted using this model increases with the order of the element used. Meanwhile other study Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

[7] emphasized the importance of PPG signals measurement from the wrist. According to this study, only blood pressure predicted from the wrist radial artery can be validated via the sphygmomanometer. Many of these inventions allow portable and non-invasive measurement of blood pressure. It is the objective of this work to explore the feasibility of using heart pulse collected from the developed circuit for continuous and real time prediction of one s blood pressure value via the measurement from fingertip. 2. Materials and Methods 2.1. Finger-cuff blood pressure system The finger-cuff system developed in this study worked in transmission mode for real-time measurement of heart pulses from one s fingertip. The schematic diagram of the constructed circuit is shown in figure 1, while figure 2 showed the overall operation diagram of the system. This work used infrared (IR) ray in the wavelength of 940 nm for the illumination of the selected skin site. The detection of the PPG signals was via the use of market available finger pulse sensor (HRM-2511E), which wrapped and held in position around the selected finger during the experiment, to minimize the possibility of motion artifacts. The measured analog PPG signals were converted to digital pulses via a 10-bit analog to digital (A/D) converter (MCP3008). These pulses were processed in real time for the prediction of the required systole and diastole blood pressure; these values were then sent to a computer via Raspberry Pi 3 for online monitoring of one s blood pressure value. Figure 1. Heart pulse detection circuit diagram HRM-2511E (Sensor) PPG circuit Raspberry Pi 3 Display and monitoring Figure 2. The operation diagram of the constructed finger-cuff PPG system 2

2.2. Two-element Windkessel model The Windkessel model was previously used in the past to describe the in-vivo arterial system with respect to the blood flow and blood pressure. The use of this model in the prediction of one s blood pressure was extensively demonstrated in the work by Choudhury et al. [5]. This study employed a two-element Windkessel model shown in equation (1) due to the simplicity of the model. This model has an analogy of an electrical circuit, wherein a resistor was placed in parallel with a capacitor. It was reported that the resistance, R, represents total peripheral resistance while capacitance, C, represents the arterial compliance. Both of these values were taken to be 1 [8]. (1) where I and P represent blood flow and blood pressure, respectively. Solving for equation (1) yields an expression of blood pressure during systolic and diastolic cycle. This is viable with the assumption that I is a sinusoidal wave during systole with peak blood flow represented by I o, and is zero during diastole. The expression of I o from ventricles to artery is given by [5]: (2) where C o and T c represent cardiac output and cardiac time, respectively. The cardiac output is defined as the product of stroke volume and average heart rate in a minute. The stroke volume in the unit of millilitre was assumed to have the same value as the body weight of the investigated individual in kilogram. In this study T c was taken as the total duration of the high of the pulse as shown in figure 3. Figure 3. Measurement of cardiac time from a pulse It was also assumed that the systolic time, T s, is one-third of the total cardiac time, while the diastolic time, T d, is two-third of the total cardiac time as followed: (3) From equation (3) and (4), the systolic blood pressure value is expressed as: while the diastolic blood pressure is given by:. (4) exp 1 exp (5) exp. (6) 3

where P ts and P td is the value of P s and P d calculated from its preceding pulse. In this work the initial value of P ts is set as 80 mmhg; the P s value calculated from equation (5) is then used as the initial value of P td in the calculation of P d from equation (6). 2.3 Experimental subject and method This study recruited three groups of individuals (aged between 20 and 30 years) from University Tun Hussein Onn Malaysia (UTHM) for a comparison of their blood pressure value measured from the conventional cuff technique and the developed approach. The device used for the cuff technique, which was used as the gold standard in this study, is OMRON arm-cuff blood pressure monitor (model no. HEM 907). These volunteers were grouped according to their blood pressure value measured from the arm-cuff technique. There were ten subjects from each group of blood pressure value range shown in Table 1. All self-declared healthy with no known underlying illness, and the recruited females were not pregnant during the study. These recruits gave their consent prior to the study. During the experiment, the subjects were told to relax in a chair and breathe normally for five minutes before blood pressure measurement was collected from either their right or left arm using the arm-cuff blood pressure monitor. This is repeated for measurement using the finger-cuff system on their index finger. Table 1. Blood pressure value range defined in this study Blood Pressure range Systolic Value (mmhg) Diastolic value (mmhg) Low < 115 < 75 Normal < 140 and > 115 < 90 and > 75 High > 140 > 90 3. Results The systolic and diastolic blood pressure value measured for the recruited subjects of different blood pressure range tabulated in table 1 using arm-cuff blood pressure monitor and the developed system is shown in table 2. From table 1 the mean and standard deviation of systole pressure value measured using cuff blood pressure monitor and the developed system, respectively, for groups of individuals with high, normal and low blood pressure is calculated as 151.4 ± 4.51 mmhg, 128.5 ± 7.82 mmhg, 104.9 ± 6.72 mmhg and 151.3 ± 5.59 mmhg, 128.6 ± 5.93 mmhg, 106.7 ± 6.36 mmhg. Meanwhile the mean and standard deviation of the diastole pressure value measured using cuff blood pressure monitor and the developed system for these groups is given by 90.9 ± 6.41 mmhg, 81.9 ± 3.54 mmhg, 67.7 ± 3.83 mmhg and 123.9 ± 3.84 mmhg, 105.1 ± 4.77 mmhg, 87.3 ± 5.2 mmhg, respectively. The average of the mean and standard deviation of absolute difference in the systolic and diastolic value given from these systems is calculated as 1.6 ± 3 mmhg and 25.3 ± 6.3 mmhg, respectively. 4. Discussion From table 2, it can be clearly seen that the results for systolic value measured from the developed system for all groups of volunteers are similar to that from the arm-cuff blood pressure monitor with the average mean error of 1.6 ± 3 mmhg. Nonetheless, a comparatively larger mean difference was observed for diastolic measurement using these two techniques given by 25.3 ± 6.3 mmhg. This is likely due to the poor measurement of the cardiac time shown in figure 3 where the T d and T s value used in equation (5) and (6) were deduced. The heart pulses used in this study were obtained following 4

the conversion of analog PPG signals to digital pulses via a 10-bit converter, this converter provides a fast sampling rate at the price of a low resolution, and this poor sampling may misrepresent the shape of the PPG. Moreover, the foot of PPG pulse wave signals, which contains the latter part of the diastole, is likely masked by noise and may go undetected. This yields an underestimation of the T d value calculated from equation (4), the erroneous in this value propagates through equation (6) resulting in an overestimation of the diastolic blood pressure. This error is then manifested itself in the following diastole blood pressure values through the equation (6). This leads to a relatively consistent difference in the diastolic blood pressure value predicted using the arm-cuff blood pressure monitor and the developed system with the mean standard deviation given by 6.3 mmhg. This work demonstrated that the information deduced from a heart pulse may potentially be used to approximate one s systole and diastole blood pressure value. This approach allows a real time processing of the signals as it placed a lower demand on the system required for processing and sampling of the signals. However, an A/D conversion system with high sampling resolution may be needed to allow a more accurate measurement of the cardiac time, and to more accurately predict the required systole and diastole blood pressure. Table 2. The measured diastolic and systolic value from gold standard (arm-cuff blood pressure monitor) and the developed finger-cuff system Blood pressure range High Blood pressure Normal blood pressure 5

Low blood pressure Conclusion This study showed that the developed finger-cuff system is able to provide an online prediction of one s systolic and diastolic value based on the analysis of the heart pulse. The results showed high consistency in the systolic value given from the market available arm-cuff blood pressure monitor and the developed system for all the recruited groups. Nonetheless a relatively large difference is observed in the measured diastolic value due to a lack in the A/D conversion efficiency. This work concluded that the systolic and diastolic blood pressure value may be approximated from heart pulse signals through an accurate measurement of cardiac time and with the use of a two-element Windkessel model. Acknowledgement The authors would like to thank all volunteers who participated in this study. The authors are also grateful to Nooreis Shadilla Jarkasi from Department of Electronic Engineering, UTHM, for her help during data collection. This work was financially supported in part by the Ministry of Education Malaysia under FRGS grant no. 1581. References [1] Y Zheng, C C Y Poon, B P Yan and J Y W Lau 2016 Journal of Medical System 40 1 [2] K Matsumura, T Yamakoshi, P Rolfe and K Yamakoshi 2016 IEEE Transactions on Biomedical Engineering 64 1131 [3] C C Y Poo, Q Liu, H Gao, W H Lin and Y T Zhang 2011 Journal of Computing Science and Engineering 5 246 [4] Y T Zhang, C C Y Poon, C H Chan, M W W Tsang and K F Wu 2006 3 rd IEEE/EMBS International Summer school on medical devices and Biosensors [5] A D Choudhury, R Banerjee, A Sinha and S Kundu 2014 Conf Proc IEEE Eng Med Biol Soc p 4567 [6] E Maksuti, N Westerhof, B E Westerhof, M Broome, N Stergiopulos 2016 Plos one [7] S H Song, J S Cho, H S Oh, J S Lee and I Y Kim 2009 IEEE Explore: Computer in Cardiology 741 [8] M Hlaváč, Windkessel 2004 Proc 2004 Student Electrical Engineering, Information and Communication Technologies 1 6