THE SPIROMETERS TESTING

Similar documents
SPIROMETRY METHOD. COR-MAN IN / EN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark

Spirometry and Flow Volume Measurements

Chapter 3. Pulmonary Function Study Assessments. Mosby items and derived items 2011, 2006 by Mosby, Inc., an affiliate of Elsevier Inc.

UNIT TWO: OVERVIEW OF SPIROMETRY. A. Definition of Spirometry

EVect of breathing circuit resistance on the measurement of ventilatory function

Original Contributions

PULMONARY FUNCTION TEST(PFT)

Assessment of accuracy and applicability of a new electronic peak flow meter and asthma monitor

#7 - Respiratory System

Evaluation of the turbine pocket spirometer

Difference Between The Slow Vital Capacity And Forced Vital Capacity: Predictor Of Hyperinflation In Patients With Airflow Obstruction

DIAGNOSTIC ACCREDITATION PROGRAM. Spirometry Quality Control Plan

#8 - Respiratory System

Peak expiratory flow and the resistance of the mini-wright peak flow meter

Comparison of patient spirometry and ventilator spirometry

Respiratory Physiology In-Lab Guide

COMPREHENSIVE RESPIROMETRY

CIRCULAR INSTRUCTION REGARDING ESTABLISHMENT OF IMPAIRMENT DUE TO OCCUPATIONAL LUNG DISEASE FOR THE PURPOSES OF AWARDING PERMANENT DISABLEMENT

Spirometric protocol

Supramaximal flow in asthmatic patients

PREDICTION EQUATIONS FOR LUNG FUNCTION IN HEALTHY, LIFE TIME NEVER-SMOKING MALAYSIAN POPULATION

PULMONARY FUNCTION. VOLUMES AND CAPACITIES

D.J. Chinn*, J.E. Cotes

PULMONARY FUNCTION TESTING. Purposes of Pulmonary Tests. General Categories of Lung Diseases. Types of PF Tests

Int. J. Pharm. Sci. Rev. Res., 34(2), September October 2015; Article No. 24, Pages: Role of Spirometry in Diagnosis of Respiratory Diseases

Breathing and pulmonary function

INTERFACING ARRANGEMENT AND SOFTWARE DEVELOPMENT FOR ACQUISITION AND ANALYSIS OF RESPIRATORY DATA

Spirometry is the most frequently performed. Obstructive and restrictive spirometric patterns: fixed cut-offs for FEV1/FEV6 and FEV6

Physiology lab (RS) First : Spirometry. ** Objectives :-

BiomedicalInstrumentation

PULMONARY FUNCTION TESTS

Spirometry: an essential clinical measurement

PULMONARY FUNCTION TESTING. By: Gh. Pouryaghoub. MD Center for Research on Occupational Diseases (CROD) Tehran University of Medical Sciences (TUMS)

Annual lung function changes in young patients with chronic lung disease

You Take My Breath Away. Student Information Page 5C Part 1

6- Lung Volumes and Pulmonary Function Tests

Quality Assurance Mapping Your QC Program Equipment and Test Quality. Susan Blonshine RRT, RPFT, FAARC, AE-C

3.0 METHODS. 3.1 Participants

Credential Maintenance Program

S P I R O M E T R Y. Objectives. Objectives 2/5/2019

Getting Spirometry Right It Matters! Performance, Quality Assessment, and Interpretation. Susan Blonshine RRT, RPFT, AE-C, FAARC

Standardisation of spirometry

RESPIRATORY PHYSIOLOGY Pre-Lab Guide

Understanding the Basics of Spirometry It s not just about yelling blow

S P I R O M E T R Y. Objectives. Objectives 3/12/2018

Factors determining maximum inspiratory flow and

Biology 236 Spring 2002 Campos/Wurdak/Fahey Laboratory 4. Cardiovascular and Respiratory Adjustments to Stationary Bicycle Exercise.

Pulmonary Function Testing

Prediction equations for pulmonary function values in healthy children in Mashhad city, North East Iran

Content Indica c tion Lung v olumes e & Lung Indica c tions i n c paci c ties

GENERAL INFORMATION. Page 1 z 11

2.0 Scope: This document is to be used by the DCS staff when collecting participants spirometry measurements using the TruFlow Easy-On Spirometer.

Spirometry Training Courses

DESIGN, ANALYSIS AND IMPLEMENTATION OF SPIROMETER

Dependence of forced vital capacity manoeuvre on time course of preceding inspiration in patients with restrictive lung disease

Pulmonary Function Testing. Ramez Sunna MD, FCCP

Pulmonary Function Technologists (PFT) Detailed Content Outline

Spirometry. Obstruction. By Helen Grim M.S. RRT. loop will have concave appearance. Flows decreased consistent with degree of obstruction.

O.F. Pedersen*, M.R. Miller**, T. Sigsgaard*, M. Tidley +, R.M. Harding +

Teacher : Dorota Marczuk Krynicka, MD., PhD. Coll. Anatomicum, Święcicki Street no. 6, Dept. of Physiology

Spirometry Training Courses. Spirometry for. Thoracic Society of Australia and New Zealand. June Developed in partnership with

Technical Assessment of Spirometers Connected in Series

Testing Spirometers: Are the Standard Curves of the American Thoracic Society Sufficient?

Vitalograph Alpha Spirometer

CHAPTER 3. Correction factors for oxygen and flow-rate effects on neonatal Fleisch and Lilly pneumotachometers

Independent Health Facilities

BETTER SPIROMETRY. Marijke Currie (CRFS) Care Medical Ltd Phone: Copyright CARE Medical ltd

Purpose Of This Guide

SPIROMETRY. Marijke Currie (CRFS) Care Medical Ltd Phone: Copyright CARE Medical ltd

Computational Fluid Dynamics Modeling of Amsino OneMask Oxygen Mask

behaviour are out of scope of the present review.

Prevalence of undetected persistent airflow obstruction in male smokers years old

Spirometry: Introduction

Ventilator Waveforms: Interpretation

2.0 Scope: This document is to be used by the DCS staff when collecting participants spirometry measurements using the TruFlow Easy-On Spirometer.

Office Spirometry Guide

How to Perform Spirometry

MSRC AIR Course Karla Stoermer Grossman, MSA, BSN, RN, AE-C

Pulmonary Function Tests. Mohammad Babai M.D Occupational Medicine Specialist

Beta 2 adrenoceptor agonists and the Olympic Games in Athens

The Pressure Losses in the Model of Human Lungs Michaela Chovancova, Pavel Niedoba

SPIROMETRY. Performance and Interpretation for Healthcare Professionals. Faculty of Respiratory Physiology IICMS. Version 1 April 2015

Spirometry in primary care

SPIROMETRY TECHNIQUE. Jim Reid New Zealand

TECHNOLOGY - INNOVATION IN ULTRASONIC RESPIRATORY DIAGNOSTICS. Otthon. Ultrasonic handheld spirometer

THE SOUTH AFRICAN SOCIETY OF OCCUPATIONAL MEDICINE

Office Based Spirometry

Lung Function Basics of Diagnosis of Obstructive, Restrictive and Mixed Defects

The Role of Facemask Spirometry in Motor Neurone Disease (MND) Respiratory Sleep and Support Centre, Papworth Hospital NHS Trust, Papworth

Respiratory Mechanics

Angela Lorenzo MS, RRT, RPFT Adjunct Faculty, Respiratory Care Long Island University, Brooklyn

Outline FEF Reduced FEF25-75 in asthma. What does it mean and what are the clinical implications?

Comparison of Different Spirometry Reference Equations in Predicting Lung Function in Sri Lankan School Children

CARDIOVIT AT-10 plus THE BEST OF ITS CLASS!

Top 5 Pitfalls of Spirometry And How to Avoid Them

Pulmonary Function Test

What do pulmonary function tests tell you?

A mathematical reason for FEV1/FVC dependence on age

-SQA-SCOTTISH QUALIFICATIONS AUTHORITY HIGHER NATIONAL UNIT SPECIFICATION GENERAL INFORMATION MARCH

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP

Transcription:

THE SPIROMETERS TESTING B. K. Juroszek Division of Measuring and Medical Electronic Instruments Faculty of Basic Problems of Technology Wroclaw University of Technology, Smoluchowskiego 19 str., 50-370 Wroclaw, Poland Abstract: The spirometer is a measuring instrument which should give information about some mechanical parameters of the respiratory system in a synonymous and precise means. It consists of two main blocks: 1- the spirometric transducer, which converts non electric value, usually gas flow, into electric signal, 2- the electronic block, which measures, converts and presents the results. Their metrological parameters are usually tested by using a calibrating syringe, forced vital capacity simulator or standard waveforms gathered on a diskette. In this article the author proposes a new method of spirometer testing: two spirometers, i.e. the tested one and the standard one are serially connected into a course of calibrating signals. In this way the user can obtain quick information about tested instrument quality. Keywords: spirometry, spirometer testing. 1 INTRODUCTION The spirometer is an instrument, which should give the user complete information about some mechanical features of the respiratory system - in a precise way. The producer usually proposes the kind of measured values which will be measured precisely. Almost all metrological parameters are defined in a steady state and static conditions, which are far from reality. There are: pneumotachometers characteristics and its flow resistance. For this reason many instruments, even with similar technical parameters, can give for the same patient, extremely different results. To avoid such situations some calibrating procedures are realised, which allow us to compare measuring results obtained from different spirometers. 2 FUNDAMENTAL SPIROMETER BLOCKS AND THEIR FUNCTIONS Each spirometer consists of two fundamental blocks (Fig. 1): 1. non-electrical value sensor (of volume or flow velocity) co-operating with the transducer, with electrical signal on its output (voltage, pulses), 2. a block of processing electrical information and presenting the final results. P AAC MCS BFRP BEP Figure 1. General block diagram of electronic spirometer. P pneumotachometer (Fleish head)j, BEP - block of electrical processing, AAD amplifier and A/D converter, MCS - microcomputer with software, BFRP - block of final results presentation. All these blocks decide final preciseness of the spirometer. For this reason testing them is so important. The most popular spirometric transducer is a pneumotachometric Fleish head. It realises flow velocity conversion into pressure differences and then into voltage (Fig. 2). v P P L RESISTANCE INPUT (CAPILLARIES) NE NE GAS PRESSURE P1 PRESSURE P2 L R R U NE E E Ε OUTPUT Figure 2. The general scheme of pneumotachometric head (A) and steps of signal conversion (B) (NE - non electric value, E - electric value; v - flow velocity, P - pressure, L - membrane deflection, R - resistance, U - voltage).

The block of electrical processing must amplify the electric signal obtained from the pneumotachometer and match it into input signal level of A/D converter. A microcomputer with software, making many mathematical calculations, realises some tasks: 1. identification of the spirogram V(t) beginning, 2. correction all data, taking into consideration pneumotachometes ninlinearities, 3. discrete integration/differentiation of the signal, 4. finding signal samples needed for spirometric parameters calculation, 5. correction of the results with physical conditions (e.g. ambient temperature), 6. comparison of measured spirometric parameters with predicted (normal) values, 7. final results storage and presentation in the form of digital data and diagrams. 3 CHECKING OF THE PNEUMOTACHOMETER ACCURACY The pneumotachometric head plays a fundamental role in signal reception from the medical object through its converting accuracy (gas flow velocity into voltage conversion), characteristic nonlinearity and its flow resistance. The head characteristic is checked in static conditions. This means that when only flow remains constant output signal (usually voltage value) answers at a constant level. In practice this answer is frequently nonlinear. Figure 3 presents curvilinear pneumotachometric head characteristic, which can be software corrected, giving linearization effect. At the beginning this characteristic had the form: v = -2.76E-07U 2 + 4.35E-03U +1.7E-01. For this linearization some coefficients of this formulae must be taken into account. This decreased nonlinearity error from level 2.7 % to 1.2 %, diminishing the final error in this way. 12 12 Q i x i Q i QL i Figure 3. 0 160 VOLTAGE 3700 V i 0 160 VOLTAGE V i 3700 An example of pneumotachometric head characteristic (A) G02 connected with P/U transducer type MOTOROLA MPX12D. The effect of characteristic linearization (B) of the same head is the source of error. Head flow resistance takes the main role in checking the signal from the medical object. It must be very high (R z ), because head sensitivity should be high, according to the formulae: U = K P = R z v (1) where: U - head voltage answer, K - conversion constant, P - pressure difference like answer of gas flow v. On the other hand it should be minimal; the resistance R g which is noticed is the sum of reducing pipe resistance R z and tube resistance R p, which connects the patient with the pneumotachometer: R g = R z + R p. This resistance is serially connected with lung flow resistance (Fig. 4). RL RL CL P(t) Rg P(t) Rg RESPIRATORY SYSTEM Figure 4. PNEUMOTACHOMETER RESPIRATORY SYSTEM PNEUMOTACHOMETER The diagram illustrating the role of spirometer resistance R g in flow observation (A). This is analogue to the electric circuit. Theoretically, in steady state, flow velocity is defined like v t = P/R L. Whilst connecting the instrument with resistance R g the flow diminishes and then v r = P/(R L + R g ). The model for respiratory system in forced expiration 1 is presented in figure (B). 1 Forced expiration is the fundamental spirometric test: the patient inhales maximal gas volume into his lungs and then exhales it with maximal force and velocity.

In this way flow v i diminishes into v r value. During forced expiration this extra head resistance will change lung time constant (compare: Fig. 4B). According to electric analogue such measuring device resistance brings about the method error (analogue: the method error in current measurement) and can be eliminated with using the correction factor. But in spirometrical practice this way is not used. Rather it is better to agree with such an error, but only when it is very little. For this reason it is suggested the spirometer resistance R g ought be not bigger than 1.5 cm H 2 O/l/s for flow velocity v = 12 l/s [1]. But an extra non-measured reaction which exists between lung and instrument must be taken into account. Here appears such a phenomenon: during forced expiration for overcoming transducer resistance the patient must generate a higher (than normal) pressure in his lungs. This pressure is bigger if the spirometer resistance is bigger. This higher pressure appearing in the lungs opens usually blocked (especially in pathology) respiratory ducts and in result extra gas volume flows out, which normally (during normal breathing) does not exist. Then, expired volume is abnormally higher, which means that the measured results are doubtful. It is very difficult to define the value of this error. Therefore the results obtained from different spirometers, with different resistance R g can be discordant. 4 THE TEST OF ELECTRIC CONVERSION BLOCK The procedures of pneumotachometric signal conversion are dictated not only by head technical parameters but especially by medical object features. During forced expiration the volume changes exponentially. This change of velocity tells us about the state of the respiratory ducts (resistance R L ) and of the alveoli (compliance C L ) (compare Fig. 5). VOLUME TIME VOLUME Figure 5. Forced expiratory curve called TVC (timed vital capacity curve) (A). It presents expired volume changes in time: V(t) = V o (1 - e -t/(rl CL) ). The part of the respiratory loop called MEFV (maximal expiratory flow-volume curve) (B) presents only forced expiration: FEF max is the maximal value of flow velocity observed during forced expiration. On the basis of this test there different spirometric parameters are defined [2] which can be divided into three categories: 1. gas volumes (e.g. forced vital capacity FVC), 2. mean values of gas flow velocities (e.g. forced expiratory volume expired during the first second of forced expiration FEV 1 ), 3. instantaneous values of gas flow velocities (e.g. maximal forced expiratory flow FEF max ). The procedures of defining all these parameters are realised by easy arithmetic calculations with using signal samples V(t i ) or v(t i ). But most essential is to catch the starting moment of forced expiration. Not precisely established moment t o = 0 can be an essential source of error for all these calculations, where the time is taken into account. Perceptible time delay at the very beginning of the forced expiration signal V(t) is the result of the influence on the inertia system of both the patient and the pneumotachometer. For elimination such flow special extrapolation procedures are used which can help to define the synonymous signal of the starting moment (Fig. 6)

VOLUME BACK EXTRAPOLATION LINE TIME ZERO Figure 6. The typical spirogram form V(t) and extrapolation of the zero-point (starting point) [3]. Software testing, especially for calculation procedures, are realised by using special data which are stored on diskette (Fig. 7). TIME v FVC = 3.786 V FEV1 = 3.053 vmax = 4.368 TIME ZERO DEFECT v FVC = 3.419 FEV1 = 1.360 vmax = 2.868 V V t V t Figure 7. Some examples of testing spirometric curves no 14 (A) and no 23 (B) [3]. 5 THE REAL CONDITIONS OF SPIROMETER TESTING Both pneumotachometric head and electronic blocks decide about the spirometer preciseness. For this reason they should be tested together. In practice some testing methods are known. 5.1 One-point-direct testing method This direct testing is realised by using the respiratory system simulator. Usually it is a calibrating volume-constant syringe with stiff walls and output (orifice) in the form of a stiff tube. The gas flow depends on the direction of piston movement, about which the user decides. Such syringes are standard equipment of every spirometer. Typical volume is one or three litters, with preciseness better then 3 %. The user tests the spirometer himself. This way of testing can be called one-point, because only one precise value of volume is obtained [4]. Such a test is especially useful for quick definition of constant K for the spirometer, in accordance with the relation: 1 litre = K spirometer reading. 5.2 Many-points- direct testing method In this way the spirometer can be tested in some points: for different volumes and flow velocities [5]. The signal most similar to the reality generates the forced vital capacity simulator. Volume changes agree with the formulae: V(t) = V s (1 - e -t/τs ) (2) where: V s - the maximum volume, τ s - time constant of the generated signal. For such a form of testing signal it is easy to calculate all values of simulated spirometric parameters. This method realises the forced expiration simulator [6] or calibrator with special cartridge [7]. To be sure, it is easy to test read-out of some spirometers but is impossible to investigate pneumotachometer in the full range of working. 5.3 Indirect testing method The spirometer can also be tested by using a group of patients [8]. They are measured independently, in succession, on every instrument. This method was used to test two devices: S- Model (Vitalograph) and Stolberg (Zimmerman) (Fig. 7).

6 10 V i FEV1 V Y i v i FVC V y i 0 FEV1 S i Z 6 0 FVC s i Z 6 Figure 8. The results of the comparison of the two spirometers: S-Model (Vitalograph) and Stolberg (Zimmerman). The analysis was made for two parameters: FEV 1 (A) and FVC (B) in a group of 21 patients. In this way we can notice, that S-Model usually gives higher values of both parameters (+0.60 l for FEV 1 and + 0.53 l for FVC). Although both instruments co-operated with real medical object - the patient, if the pneumotachometers have different resistance the differences in value of spirometric parameters can lead to faulty conclusions. 6 CONCURRENT SPIROMETER TESTING It was noticed, that the pneumotachometric head is serially connected into a measured circuit (compare Fig. 4) which is an analogy to current measurement in electric circuit with using an ammeter. One of the methods in ammeters testing is to connect them serially into a current circuit. In this way current source resistance can be neglected during analysis and for this reason both ammeters should give the observer the same read-out. Analogously: spirometric results obtained form different instruments, serially connected, should be the same (in margin of measuring error). The source of testing signal can be generated both from the simulator or from the patient (or group of patients). 6.1 Test with using calibrating syringe A syringe with a volume of one litre (accuracy 1 %) was used for this test. The method of its connection into the tested spirometers is presented in Fig. 8. CALIBRATING SYRINGE HEAD OF TESTED SPIROMETER HEAD OF STANDARD SPIROMETER Figure 8. Serially connected pneumotachometric heads and calibrating syringe. The standard instrument was a spirometer Vitalograph Compact II. Because the volume changes were unrepeatable (the piston was moved by hand) only the final volume of one litre was controlled (see Tab. 1). Table 1. The results of checking K constant for the spirometers serially connected with the standard (only in couples: standard + one tested spirometer). The number of tests n = 20. kind of spirometer Vitalograph Compact II G02 G04 G05 G07 mean st. dev. mean st. dev. mean st. dev. mean st. dev. mean st. dev. 1.004 0.005 1.895 0.031 3.138 0.512 1.301 0.084 0.643 0.071 6.2 Tests made during patients measurements The spirometers were tested in couples as in p. 6.1. Two spirometric parameters were controlled: FEV 1 and FVC (Tab. 2), each of them 20 times. Before taking measurements, the devices were calibrated, for definition K constant, in 10 trials.

Table 2. The results of testing of four spirometers: G02, G04, G05, G07. Standard spirometer: Vitalograph Compact II. kind of spirometer FEV 1 FVC K FEV1 mean dispersion [%] correlation coefficient K FVC mean dispersion [%] G02 1.220 30.2 0.93 1.214 36.8 0.83 G04 3.406 25.5 0.96 4.269 44.2 0.56 G05 1.375 12.6 0.79 1.661 14.2 0.74 G07 0.716 24.7 0.97 0.795 11.6 0.94 K constant for each spirometer was defined in three ways: 1. with using calibrated syringe (K constant), 2. in comparison with a standard spirometer, for parameter FEV 1 (K FEV1 constant), 3. in comparison with a standard spirometer, for parameter FVC (K FVC constant). correlation coefficient 6.3 Results The suggested method of concurrent spirometers testing gives quick information about reliability of measurement results in devices particularly new for the user. The differences which appear in K constant (K FEV1, K FVC ) can prove nonlinearities in the pneumotachometric heads, which probably were not taken into consideration in software procedures. They suggest that it would be better at first to test these characteristics and then the flow resistance. Probably different values of K, K FEV1, K FVC constant may also be the result of great flow resistance (summarised resistance of connected pneumotachometers). 7 CONCLUSIONS Each spirometer is built of two extremely different blocks: 1. mechanical - pneumotachometer head, 2. electrical - block of analogic and discrete conversion. For this reason the procedure of its test can concern either mechanical or electrical part, or both of them simultaneously. Popularly used calibrating constant volume syringes can give information about one point only. It isn t possible to test an instrument to its full range. Such manner is useful only for a quick test of spirometer readings readiness. The testing signal collection proposed in [3] is destined only for software, excluding the pneumotachometric head. With regard to head flow resistance, which co-operates with the respiratory resistance, concurrent tests with serially connected spirometers can give complete information in full range. REFERENCES [1] C. H. Buess, U. Boutellier, E. A. Koller, Pneumotachometers. in: J. G. Webster (ed.), Encyclopaedia of medical devices and instrumentation. vol. 4, Willey N. Y. (1988) 2319-2324. [2] P. H. Quanjer, G. J. Tammeling, J. E. Cotes, L. M. Fabbri, H. Matthys, O. F. Pedersen, R. Peslin, J. Roca, P. J. Sterk, W. T. Ulmer, J. C. Yernault, Symbols, abbreviations and units. Eur. Respir. J. 6 (suppl. 16) (1993) 85-100. [3] R. M. Gardner, J. L Hankinson, J. L. Clausen, R. O. Crapo, R. L. Johnson, G. R. Epler, Standardization of spirometry - 1987 update. Am. Rev. Respir. Dis. 136 (1987) 1285-1298. [4] M. P. Yeh, R. M. Gardner, T. D. Adams, F. G. Yanowitz, Computerised determination of pneumotachometer characteristics using a calibrated syringe. J. Appl. Physiol. 53 (1) (1982) 280-285. [5] A. Shaw, J. Fisher, Calibration of same instruments for measuring peak expiratory flow. J. of Med. Eng. & Technol. 4 (6) (1980) 291-294. [6] B. Juroszek, Die Prüfung der Spirometer. in: Rechnergestutzte Diagnosefindung in der Lungendiagnostik. Komputerowe wspomaganie diagnostyki pluc. (Wroclaw, 22-25 listopada 1983). Prace Naukowe Instytutu Metrologii Elektrycznej Politechniki Wroclawskiej 1983 nr 23, Konferencje nr 9, 21-28. [7] H. W. Glindmeyer, S. T. Anderson, R. G. Kern, J. Hughes, A portable, adjustable forced vital capacity simulator for routine spirometer calibration. Am. Rev. of Resp. Dis. 121 (1980) 599-602. [8] B. M. Wright, C. B. McKerrow, Maximum forced expiratory flow rate as a measure of ventilatory capacity. Brit. Med. Journ. 21 (1959) 1041-1047.