Integra. Katalyst Bipolar Radial Head System SURGICAL TECHNIQUE

Similar documents
Bipolar Radial Head System

KATALYST. Bipolar Radial Head System. Surgical Technique. orthopedics. KATALYST English. PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

the shape, the size, the fit Ascension Modular Radial Head

Integra. Modular Radial Head System SURGICAL TECHNIQUE

Rehabilitation after Total Elbow Arthroplasty

modular RADIAL HEAD E VOLVE

rhead System Radial Head Arthroplasty Operative technique

rhead System Extended stems Operative technique

Locking Radial Head Plates

E-CENTRIX. Ulnar Head Replacement SURGICAL TECHNIQUE

Integra. Modular Radial Head System SURGICAL TECHNIQUE

SURGICAL TECHNIQUE GUIDE

UNI-Elbow System Radio-Capitellar Replacement. Operative technique

Integra. Silicone PIP Implant SURGICAL TECHNIQUE

Modular Ulnar Head surgical technique. Transforming Extremities

Ascension. Silicone MCP surgical technique. surgical technique Ascension Silicone MCP

UDHT08.1.qxd:UDHT /03/08 17:14 Page 1. Surgical. Technique. Elbow Prosthesis. RHS Radial Head System.

Surgical. Technique CRF II. Radial Head Prosthesis

Integra. Silicone MCP Joint Prosthesis SURGICAL TECHNIQUE

Elbow Elbow Anatomy. Flexion extension. Pronation Supination. Anatomy. Anatomy. Romina Astifidis, MS., PT., CHT

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

U2 PSA. Revision Knee. Surgical Protocol

Surgical Technique. DISCLOSURE: This device is not approved for sale in the U.S.A. Customer Service:

Recurrent subluxation or dislocation after surgical

Surgical Technique. Forearm Fracture Solutions

Radial Head Prosthesis System

2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1

Technique Guide Small Bone Fusion System

RADIAL HEAD PROSTHESIS SYSTEM For primary and revision joint replacement of the radial head

Integra. Spider and Mini Spider Limited Wrist Fusion System SURGICAL TECHNIQUE

Silicone PIP, MCP & MCP-X (PreFlex)

EXTERNAL FIXATION SYSTEM

AcUMEDr. FoREARM ROD SYSTEM

Case Presentation: Comminuted Radial Head Fracture

ACETABULAR CUP SURGICAL TECHNIQUE

Zimmer Trabecular Metal Ankle Interpositional Spacer and Trabecular Metal Ankle Fusion Spacer

LCP Distal Humerus Plates

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

THE P.F.C. SIGMA FEMORAL ADAPTER. Surgical Technique

Integra. DigiFuse Cannulated Intramedullary Fusion System SURGICAL TECHNIQUE

Simply Complete. VA-LCP Elbow Plating System.

Radial Head Fractures Save or Replace?

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Optimum implant geometry

Functional Anatomy of the Elbow

MICRONAIL. Intramedullary Distal Radius System SURGICAL TECHNIQUE

SURGICAL TECHNIQUE. Alpine Cementless Hip Stem

Terrible triad of the elbow

Anterior Elbow Capsulodesis

Total Knee Original System Primary Surgical Technique

Approach Patients with Confidence

Ovation Hip System. Surgical Technique

Anterior Approach Surgical Technique. Paragon Stem System. enabling people to enjoy life

Elbow Anatomy, Growth and Physical Exam. Donna M. Pacicca, MD Section of Sports Medicine Division of Orthopaedic Surgery Children s Mercy Hospital

WEEKEND 2 Elbow. Elbow Range of Motion Assessment

Visualize, stabilize, mobilize. Wristore * Distal Radius Fracture Fixator Abbreviated Surgical Technique

Case Presentation: Comminuted Fractures of the Proximal Ulna 11/28/2017. Disclosures. Surgical Strategy. Implant Choice. Melvin P.

Zimmer NexGen MIS Tibial Component. Cemented Surgical Technique IMAGE TO COME

Single Axis Revision Knee System

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

A Patient s Guide to Adult Radial Head (Elbow) Fractures

Surgical Technique. Distal Humerus Locking Plate

SURGICAL TECHNIQUE GUIDE

The Biomechanics of the Human Upper Extremity-The Elbow Joint C. Mirzanli Istanbul Gelisim University

Surgical Technique Guide

SURGICAL TECHNIQUE GUIDE

Distal Ulnar Locking Plate

Ascension. MCP surgical technique. surgical technique Ascension MCP PyroCarbon Total Joint

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

Conventus CAGE PH Surgical Techniques

Zimmer FuZion Instruments. Surgical Technique (Beta Version)

MIS Cemented Tibial Component

Surgical Technique. Hinge Disassembly and Rebuild Technique

Zimmer NexGen Trabecular Metal Tibial Tray

TRK REVISION KNEE Surgical Technique

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

MANAGEMENT OF INTRAARTICULAR FRACTURES OF ELBOW JOINT. By Dr B. Anudeep M. S. orthopaedics Final yr pg

Integra. ADVANSYS Plating System SURGICAL TECHNIQUE

Distal Cut First Femoral Preparation

System. Humeral Nail. Surgical Technique

Progeny Hip Stem. Surgical Protocol and Product Specifications

Integra. Freedom Wrist Arthroplasty System SURGICAL TECHNIQUE

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Wright Medical Technology, Inc Cherry Road Memphis, TN

Elbow dislocations represent 10% to 25% of all injuries. Elbow Fracture-Dislocations. The Role of Hinged External Fixation

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

Fractures and dislocations around elbow in adult

THE ELBOW. The elbow is a commonly injured joint in both children and adults.

Clinical Evaluation Surgical Technique

Bone Preservation Stem

Triathlon Knee System Universal Baseplate Surgical Protocol

VerSys LD/Fx Cemented and Press-Fit Hip Prostheses. Surgical Technique IMAGE TO COME. Versatile solutions for total and partial hip replacement

Knee Surgical Technique

Revolution. Unicompartmental Knee System

Pinit Plate Small Bone Fusion System Bone Plate & Screw System

Transcription:

Integra Katalyst Bipolar Radial Head System SURGICAL TECHNIQUE

Surgical Technique As the manufacturer of this device, Integra does not practice medicine and does not recommend this or any other surgical technique for use on a specific patient. The surgeon who performs any implant procedure is responsible for determining and using the appropriate techniques for implanting the device in each patient. Indications Comminuted radial head fracture with or without associated elbow instability Comminuted radial head fracture with associated failure of the interosseous membrane; the Essex-Lopresti lesion Post-traumatic or primary osteoarthritis involving the radiocapitellar or proximal radioulnar joint Rheumatoid arthritis involving the radiocapitellar joint or proximal radioulnar joints Revision of failed radial head implant Contraindications Inadequate bone and surrounding soft-tissues to accept and support the implant Neurological, metabolic or other medical conditions that might compromise the bone, soft tissues or proprioception about the elbow joint Open physes in the radius, ulna and distal humerus Infection Wear or extensive injury to the articular surface of the capitellum System Features The Katalyst Telescoping Bipolar Radial Head implant restores the support and bearing surface of the radial head in the face of fracture, arthritis or failed prior radial head implantation. The Katalyst System offers three unique features: Telescoping shaft allows correct length and tension of the soft tissues Bipolar neck provides optimal load transfer to the humerus and decreases the potential for capitellar wear Anatomic curvature of the head recreates normal anatomy 1 These features allow the surgeon to recreate appropriate tension in situ with an implant that articulates with the capitellum and sigmoid notch in an anatomic fashion. In addition, the modular design allows for radial head replacement without surgical disruption of the lateral ligament complex from the humerus. See Package Insert for full prescribing information.

Step 1 Assessment of Elbow Stability 1-1 Valgus stability, posterolateral stability and longitudinal stability are determined using commonly accepted practices such as preoperative radiographs, intraoperative inspection, and stress radiographs. Surgical Approach After the skin incision is performed (Figure 1-1), the radial head is exposed via a modified lateral approach designed to preserve the lateral ulnar collateral ligament and the posterior half of the radial collateral ligament. The approach splits the thick common extensor tendon origin in the midline of the radiocapitellar axis. 1-1 Step 2 Exposure 2-1 The appropriate plane can be identified by palpating the lateral condylar ridge and radial head. A full-thickness incision is made through the common tendon and ligament origin along the longitudinal axis of the radial head and neck (Figure 2-1). This is commonly through the center of the white common extensor tendon. The origin of the radial wrist extensors and the joint capsule may be elevated from the anterolateral aspect of the distal humerus as needed for exposure. Proximal extension is carried out anterior to the lateral epicondyle. The split of the lateral complex distally extends to the metaphysis of the radial neck. In the atraumatic setting, the exposure can be performed without destabilizing the elbow since the posterior half of the extensor origin and lateral ligament complex remain intact. 2-1 Note When the radial head has fractured in association with a dislocation of the elbow, the lateral complex will often be stripped from its origin on the lateral epicondyle. This should be noted during the exposure, and a plan should be formulated to repair or reconstruct the ligament complex at the end of the case. (Text on file) 2

Step 3 Radial Head Resection 3-1 Radial head resection at the proximal articular margin at this level will prevent impingement between the cut end of the proximal 3-1 radius and the sigmoid notch. The appropriate level of resection can be determined by using the Cutting Template (Figure 3-1); this corresponds with the resection of the head using the 1mm First Spacer. Note Once the cut in the neck is made, measure the distance from the coronoid to the cut edge of the proximal radius and re-create that distance with the head and neck (see Step 7). Step 4 Opening Intramedullary Canal 4-1 The Starter Awl is used to open the intramedullary canal of the proximal canal. (Figure 4-1) 4-1 Step 5 Broaching 5-1 The combination Broach with End Cutting Mill is used to create a smooth cut end of the proximal radius with a reamed intramedullary canal that is perpendicular to the long axis of the radius (Figure 5-1). It is recommended to start with the 6.5mm Broach prior to using the 7.5mm Broach. 5-1 3

Step 6 Choosing Trial Head 6-1 Once the canal is broached up to the appropriate size, the corresponding Trial Stem is placed into the intramedullary canal with the black line on the side of the trial stem facing the surgeon. The Broaches and trial stems are color coded for convenience. The appropriate diameter Trial Head is chosen based on pre-operative templating and/or reassembly of the fracture components using the Head Sizing Fixture as a template (Figure 6-1). In cases where there is a question between 2 head sizes, the smaller implant is chosen. Care must be taken to use the smallest diameter of the resected native radial head for measurement purposes to avoid placing too large an implant. 6-1 Step 7 Final Evaluation of Size and Length 7-1 Templates of the actual head sizes and lengths at the 1mm and the 3mm positions are used to document proper sizing (Figure 7-1). Head circumferences can be checked by seating the trial into the joint against the radial or sigmoid notch of the ulna. The concavity of the head should match the convexity of the capitellum when the trial is seated against the sigmoid notch of the ulna. Length of the radial head implant is determined by first making sure that the ulnohumeral joint is properly reduced (typically by looking at the lateral aspect of the ulnohumeral joint (the coronoid should sit flush with the trochlea). The forearm can be placed in neutral rotation for this purpose. The chosen size head circumference and length (1mm or 3mm position) should slide into the space between the cut end of the radial neck and the capitellum without significant force. This documents the appropriate radial neck resection for the length of the replacement. Determining the appropriate separation in the head-neck construct is more difficult when the lateral complex has torn. This can be accomplished by using the method above, taking care to ensure that the ulnohumeral joint is reduced. A provisional reduction of the tendon and ligament origin to the humerus with the ulnohumeral joint reduced (by flexion and pronation of the forearm) can also be performed. The proximal margin of the implant should line up with the articular surface of the coronoid when viewed from the lateral side. 7-1 NOTE: The forearm must be in 90 of flexion and at full pronation when determining the appropriate tension via the spacer forks. 4

Step 8 Using Spacers To Adjust Seperation 8-1 Once the Trial Head is snapped in place on the stem, the two components are extended to re-create the appropriate distance from the coronoid, to the edge of the proximal radius as determined in Step 3. This separation is achieved by inserting the First Spacer which will position the implant in the first extended position. For each additional extension required, 2mm Spacers are used. Step 9 Locking The Stem 9-1 Once the appropriate tension is achieved, the Trial Stem is locked in place by turning the Set Screw in a clockwise fashion for 3/4 of a turn (Figure 9-1). 9-1 Step 10 Implant Alignment 10-1 Intraoperative radiographs may be used to confirm appropriate size and position of the implant. NOTE: The set screw in the trial is captured in place and cannot be removed. 10-1a The stem should be aligned with the longitudinal axis of the radius. The head should match the capitellum and seat in a congruent fashion against the sigmoid notch. The medial joint space of the ulnohumeral articulation should be symmetrical and not converge on the anteroposterior projection (Figure 10-1a); if so, the elbow may have been forced into a valgus position by an overstuffed implant (Figure 10-1b). 10-1b The lateral ulnohumeral joint space may be slightly divergent. Comparison radiographs help define the appropriate alignment of implant and the joint. Note that joint spaces can not be adequately assessed unless the AP radiograph is performed with the elbow fully extended. Move the elbow through a full arc of flexion and forearm rotation. A lack of motion, particularly flexion, may be additional evidence of an oversized or over-distracted trial implant. Step 11 Stem Insertion 11-1 The Trial Implant is removed and the Permanent Implant of the same size is placed. Prior to placing the Stem in the surgical field, the Set Screw is removed from the side of the Stem. The Stem is fully inserted into the canal, again ensuring that the black line on the implant faces the surgeon (Figure 11-1). 11-1 5

Step 12 Attaching Radial Head 12-1 The appropriate sized Head is placed over the spherical end of the Stem. There is a hex feature on both the Head and the Stem. By placing one Hex Driver in each of these two holes and pinching the two Hex Drivers together, the Head will snap in place on the stem (Figure 12-1). The Hex Driver handle should be parallel with each other in order to ensure proper engagement. Pliers are provided to assist with pinching the two drivers together when snapping the Head onto the Stem. 12-1 NOTE: DO NOT DISASSEMBLE Once the head has been assembled to the stem, disassembly of these components will adversely affect retention strength between them and is not recommended unless the surgeon decides that the choice of head size was improper. Step 13 Set Screw Locking 13-1 Once in the correct position by use of the spacers (Figure 13-1), the Set Screw is placed back into the stem, taking care not to over-torque the screw. Finally, the Spacers are removed from the implant. Repeat radiographs and intra-operative stress testing are now performed. 13-1 Step 14 Repair of The Lateral Complex 14-1 If the posterior half of the lateral complex is intact, the anterior half is repaired to the intact posterior limb. If the lateral complex was avulsed or released, it is repaired to the lateral epicondyle. The skin is closed and the elbow is placed in a posterior splint. Step 15 Final Radiograph 15-1 In cases of unstable elbow fracture dislocations it is recommended that a lateral radiograph of the splinted elbow be obtained in the operating room before the patient is awakened to ensure that a concentric reduction of the elbow has been maintained. Post-Operative Care 16-1 Elbow motion and forearm rotation are begun in the early postoperative period. Rehabilitation is guided by intraoperative stability. Serial follow-up radiographs should be examined for evidence of implant malposition or wear. Sclerosis in the bone at the implant-bone interface is evidence of appropriate load transfer through the implant. Note: Revision Or Removal To disengage the head from the stem, place one hex driver in the head component and a second hex driver in the stem component and lever the hex driver in the head away from the driver in the stem until the head is disassembled. Reusing a head component that has been previously assembled to a stem is not recommended as the retention force will be compromised. 6

Integra Katalyst Bipolar Radial Head System Implants Catalog Number Description Size 22-1418 Katalyst Head 18mm 22-1421 Katalyst Head 21mm 22-1424 Katalyst Head 24mm 22-1665 Katalyst Stem 6.5mm 22-1675 Katalyst Stem 7.5mm 24mm 21mm 18mm Instruments Catalog Number Description Size Item 7.5mm 6.5mm Variable Sizes 225205KMI Katalyst Pliers A 225600 Cutting Template B 225900 Head Sizing Template C 226000 Starter Awl D 225375 Broach Long 7.5mm E 225365 Broach Long 6.5mm F 224175 Trial Stem Shaft 7.5mm G 224165 Trial Stem Shaft 6.5mm H 224324 Katalyst Trial Head 24mm I 224321 Katalyst Trial Head 21mm J 224318 Katalyst Trial Head 18mm K 225405KMI First Spacer L 225400KMI Spacer 2mm M, N, O, P 2.5 Hex Driver Q, R, Complete Instrument Set Catalog Number Description KATALYSTSET Katalyst Instrument Set (22-5000) 221001 Katalyst Case Base and Lid Component Materials Head and poly Assembly: Cobalt Chrome with polyethylene liner Stem Assembly: Stainless Steel Instrument Tray A B C D E F G H I J K L M N O P Q R For more information or to place an order, please contact: Integra n 311 Enterprise Drive, Plainsboro, NJ 08536 877-444-1122 USA n 609-936-5400 0utside USA n 866-800-7742 fax integralife.com Katalyst, Integra and the Integra logo are trademarks of Integra LifeSciences Corporation. 2011 Integra LifeSciences Corporation. All rights reserved. Printed in the USA ER4771-05/11