Interference and Point-of-Care Testing Devices

Similar documents
Point-of-Care Testing in Special Populations

Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

January Testing Delays and Spurious Results Caused by Improper Specimen Collection

Point of Care Testing (POCT)

Patient Safety: A Quality System Approach To POCT QC/QA

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Delta Check Calculation Guide

Relationship between glucose meter error and glycemic control efficacy

Individual Quality Control Plans (IQCP) Risk Assessment

Ellis Jacobs, Ph.D., DABCC. New York University School of Medicine Bellevue Hospital Center New York, New York

Patient Safety: A Quality System Approach To POCT QC/QA

Determination of hemoglobin is one of the most commonly

Compliance Issues in Off-Label Use of Blood Glucose Monitors

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D.

I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood

Mission Hb accurately detects both

It s Just a Waived Glucose, Isn t It?

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Aina Blood Monitoring System

Errors in POCT: When Improbable Situations Become Possible

Quality Control for Point-of-Care Testing

i-stat CHEM8 + Cartridge Quick Reference Guide

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

35.0. ug/ml Trough. 8.0 Carbamazepine 15.0 ug/ml Digoxin 2.5 ng/ml Gentamicin Peak Random Trough

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Use of Blood Lactate Measurements in the Critical Care Setting

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

Smart Probe for Uric Acid Detection

7/31/2018. Overview of Next Generation Cardiac Troponin T High Sensitivity. Disclosures. Course Objectives: high sensitive Troponin T assay

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient.

Validity and Reliability of a Glucometer Against Industry Reference Standards

Quality Specifications for POCT

MidMichigan Health LABORATORY POLICY Title: Effecti ve te: Key Words: places: Category: Applicability: reference laboratory

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

Chapter 14 Elderly, Home and Long-term Care Collections. Objectives:

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

GENERAL. Compulsory module GEN II Example questions (Acute and Routine Clinical Chemistry)

Empowering the RT with New Noninvasive Monitoring Capabilities

Biosensor and its electrochemical transducer

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Analytical and Clinical Performance of the epoc Blood Analysis System. Experience at a Large Tertiary Academic Medical Center

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

Coping with Analytical Interferences

Standard Operating Procedure

Monitoring of Blood Glucose Level

MEETING THE STANDARDS: FDA MANDATORY RAPID INFLUENZA DETECTION TEST (RIDTs) RECLASSIFICATION

BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S.

Quick Reference Guide

Introduction To Medical Technology. Hemoglobin Concentration Determination

CARE ABOUT YOUR HEALTH

Virtual Lectures Planning Committee Disclosure Summary

Patient: 55 y female (ambulatory)

Quick Reference Guide

Pulmonary Function Technologists (PFT) Detailed Content Outline

Quick Reference Guide

2008 CAP TODAY Q & A

PHSA Laboratories CW Site - Point of Care Title: CWPC_BGE_0105 istat1 Quick Reference Guide

The Whats and Hows of Reference Intervals. Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney

Millennium Technology for Critical Care

Use of PSI on CS instruments

Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine, THC and Opiates in Urine by EMIT

The good, the bad, the ugly A month in the life of a Point Of Care Coordinator

Use of PSI on CS instruments

Smart Diagnostics for Point-of-Care!

Red blood cell transfusions Risks, benefits, and surprises

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Human Hemoglobin Colorimetric Detection Kit

MADE IN KOREA. Advanced POCT. Automated 6 Channels. Automated Single Channel. POCT to Desktop. Our technology covers it all!

The analytical phase

Alere Triage Meterpro Alere

Multiple studies are available concerning the use of

Hematocrit a review of different analytical methods

BCH 471. Experiment (5) Hemoglobin Estimation, Erythrocyte Sedimentation Rate (ESR) and Hematocrit (HCT)

Why Old Blood is Bad. tales from the electronic perfusion record. Molly Marko, BS, BSE, CCP Geisinger Health System Danville, Pennsylvania

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

JOB AID CRITICAL VALUES AND TESTS W/O MICRO CRITICAL TESTS. Always call results for the following test(s):

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

Figure 2.1: Glucose meter

Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana

D- Xylose Absorption Test

HBA1C AN INACCURATE MEASURE OF GLYCEMIC CONTROL IN A FEMALE WITH HEREDITARY SPHEROCYTOSIS DESPITE NORMAL HEMOGLOBIN LEVELS

Utilizing Sysmex RET He to Evaluate Anemia in Cancer Patients

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC

Credential Maintenance Program

Waiting for High-Sensitivity POCT Cardiac Troponin Assays: Clinical and Analytical Needs I Have a Pain in My Chest That Hurts Very Bad

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support

PENN STATE MILTON S. HERSHEY MEDICAL CENTER TABLE OF CRITICAL LAB VALUES CURRENT AS OF: 1/2015

Transcription:

Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT 1

Learning Objectives Identify common interferences affecting POC testing Describe cases where interfering substances affected patient care. Describe solutions to mitigate the impact of interfering substances on POC testing. 2

POCT Device Formats Definition: POCT is defined as testing at or near the site of patient care 3

POCT Device Formats Examples: Disposable Handheld Portable Transportable Benchtop Monitoring 4

POCT Device Formats Examples: Disposable Handheld Portable Transportable Benchtop Monitoring Being FDA approved as a POCT device does not mean it is not susceptible to interfering substances!!! 5

Total Testing Process: Difference Phases Total Testing Process: Lab testing occurs over three critical phases: Pre-Analytical 6

Total Testing Process: Difference Phases Total Testing Process: Lab testing occurs over three critical phases: Pre-Analytical Analytical 7

Total Testing Process: Difference Phases Total Testing Process: Lab testing occurs over three critical phases: Pre-Analytical Analytical Post-Analytical 8

Total Testing Process: Difference Phases Total Testing Process: Lab testing occurs over three critical phases: Pre-Analytical Analytical Post-Analytical TREATMENT 9

Total Testing Process: Sources of Error Errors in the Pre-Analytical Phase: Most frequent source of errors (up to 70%). Incorrect Components Patient preparation Sample collection Transportation Accessioning Processing Pre-Analytical Analytical Post-Analytical TREATMENT 10

Total Testing Process: Sources of Error Errors in the Pre-Analytical Phase: Most frequent source of errors (up to 70%). Incorrect Components Sources of Error Patient preparation Sample collection Transportation Accessioning Processing Pre-Analytical Analytical Post-Analytical TREATMENT Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances 11

Total Testing Process: Sources of Error Errors in the Analytical Phase: Infrequent in laboratory tests, however may be higher in POCT due to non-lab trained personnel operating devices. Components Patient preparation Sample collection Transportation Accessioning Processing Testing Sources of Error Pre-Analytical Analytical Post-Analytical TREATMENT Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances 12

Total Testing Process: Sources of Error Errors in the Analytical Phase: Infrequent in laboratory tests, however may be higher in POCT due to non-lab trained personnel operating devices. Components Patient preparation Sample collection Transportation Accessioning Processing Testing Sources of Error Pre-Analytical Analytical Post-Analytical TREATMENT Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents 13

Total Testing Process: Sources of Error Errors in the Post-Analytical Phase: Second most common among laboratory-based results. Components Patient preparation Sample collection Transportation Accessioning Processing Testing Results interpretation Entry to LIS/EMR Contacting providers Sample archiving Sources of Error Pre-Analytical Analytical Post-Analytical TREATMENT Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents 14

Total Testing Process: Sources of Error Errors in the Post-Analytical Phase: Second most common among laboratory-based results. Components Patient preparation Sample collection Transportation Accessioning Processing Testing Results interpretation Entry to LIS/EMR Contacting providers Sample archiving Sources of Error Pre-Analytical Analytical Post-Analytical TREATMENT Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents Misinterpretation of results IT problems 15

Total Testing Process: Sources of Error Errors in the Post-Analytical Phase: Second most common among laboratory-based results. Components Patient preparation Sample collection Transportation Accessioning Processing Testing Results interpretation Entry to LIS/EMR Contacting providers Sample archiving Sources of Error Pre-Analytical Analytical Post-Analytical TREATMENT Incorrect patient ID Mislabeling of specimens Hemolysis Wrong specimen type Improper specimen collection Interfering substances QC/calibration Operator error Bad reagents Misinterpretation of results IT problems What is the significance of testing error in POCT? 16

Glucose Meter Paradigm to Highlight the Role of Testing Errors 17

Glucose Meter Paradigm to Highlight the Role of Testing Errors >12,000 glucose meter errors are reported to the FDA each year highest number of adverse events for any in vitro diagnostic device. 18

Glucose Meter Paradigm to Highlight the Role of Testing Errors >12,000 glucose meter errors are reported to the FDA each year highest number of adverse events for any in vitro diagnostic device. 12,672 serious injuries reported from 2004-2008 to the FDA. 19

Glucose Meter Paradigm to Highlight the Role of Testing Errors >12,000 glucose meter errors are reported to the FDA each year highest number of adverse events for any in vitro diagnostic device. 12,672 serious injuries reported from 2004-2008 to the FDA. Most of these reported errors are due to erroneous results from interfering substances and operator error. 20

Glucose Meter Paradigm to Highlight the Role of Testing Errors >12,000 glucose meter errors are reported to the FDA each year highest number of adverse events for any in vitro diagnostic device. 12,672 serious injuries reported from 2004-2008 to the FDA. Most of these reported errors are due to erroneous results from interfering substances and operator error. 21

Glucose Meter Paradigm to Highlight the Role of Testing Errors >12,000 glucose meter errors are reported to the FDA each year highest number of adverse events for any in vitro diagnostic device. 12,672 serious injuries reported from 2004-2008 to the FDA. Most of these reported errors are due to erroneous results from interfering substances and operator error. 22

Common Confounding Factors for Glucose Meters Anemia and polycythemia causes falsely high or falsely low results respectively. 23

Hematocrit Effects on BGMS Measurements 60 40 20 y = -0.9465x + 30.951 R² = 0.4171 BIAS (mg/dl) 0-20 0 10 20 30 40 50 60 70-40 -60 Sample Size: 60 Hematocrit Range: 19-60% Glucose Range: 90-296 mg/dl -80 Note: Bias = BGMS Plasma Glucose Hematocrit (%) 24

Common Confounding Factors for Glucose Meters 25

Common Confounding Factors for Glucose Meters Oxidizing and reducing substances interfere with electrochemical sensors causing falsely high or low results. 26

Drug Interferents (Oxidizing Substances) Tang Z, et al. Am J Clin Pathol 2000;113:75-86

The role of drug interferences in critical care BGMS accuracy Tran NK, et al. J Burn Care Res 2014;35:72-79 28

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE 300 250 History: Patient is a 21 y/o woman with 90% TBSA burns from MVA. BGMS B*** Blood Glucose (mg/dl) 200 150 100 50 Started Ascorbic Acid (66 mg/kg/hr) Stopped Ascorbic Acid 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Time Following Admission (Hours)

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE 300 BGMS B Blood Glucose (mg/dl) 250 200 150 100 50 Started Ascorbic Acid (66 mg/kg/hr) PLASMA GLUCOSE (Lab) Stopped Ascorbic Acid 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Time Following Admission (Hours)

CASE EXAMPLE: ASCORBIC ACID INTERFERENCE 300 BGMS A Blood Glucose (mg/dl) 250 200 150 100 50 Started Ascorbic Acid (66 mg/kg/hr) BGMS B*** PLASMA GLUCOSE (Lab) Stopped Ascorbic Acid 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Time Following Admission (Hours)

Common Confounding Factors for Glucose Meters 32

Common Confounding Factors for Glucose Meters Specimen temp alters biosensor enzyme kinetics. Hypotension/shock affect capillary specimens. 33

Common Confounding Factors for Glucose Meters Some glucose meters cannot differentiate between certain nonglucose sugars (e.g., maltose, galactose) 34

Non-Glucose Sugar Interferences Icodextrin is a dialysis drug. It is metabolized by the body to maltose. In some glucose biosensors, maltose is indistinguishable from glucose. 35

Maltose Related Deaths BGMS A BGMS B BGMS C Timeframe 1997-14 2013-14 2007-11 Adverse Events (Deaths) 28 (13) 5 (0) 0 (0) Erroneous Results 557 168 15 Non-Clinical Event 387 59 21 TOTAL 1094 232 36 FDA MAUDE Database website: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/ search.cfm, Accessed on August 20, 2014 37

Continuous Glucose Monitors? Similar sensor designs so susceptible to similar interferences (will vary based on manufacturer). CGM based on interstitial fluid measurements and not plasma or whole blood. Potential for many other sources of interferences. CGM does not fall under CLIA and most devices compared against obsolete or poor reference methods such as the YSI. Use WITH caution! 38

INTERFERENCES IN WHOLE BLOOD ANALYSIS 39

INTERFERENCES IN WHOLE BLOOD ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 40

INTERFERENCES IN WHOLE BLOOD ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 41

Air Contamination of Blood Specimens Background: Anesthesia reports impossible venous blood gas values in one patient where end tidal CO2 was greater than the venous blood gas (VBG). 42

Air Contamination of Blood Specimens Background: Anesthesia reports impossible venous blood gas values in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: ph = 7.54, pco2 = 17.5, po2 = 168.5 POC VBG#2: ph = 7.56, pco2 = 12.7, po2 = 165.9 End tidal CO2 = 28 43

Air Contamination of Blood Specimens Background: Anesthesia reports impossible venous blood gas values in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: ph = 7.54, pco2 = 17.5, po2 = 168.5 POC VBG#2: ph = 7.56, pco2 = 12.7, po2 = 165.9 End tidal CO2 = 28 Lab Venous Blood Gas: ph 7.54, pco2 = 19.2, po2 = 161.5 44

Air Contamination of Blood Specimens Background: Anesthesia reports impossible venous blood gas values in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: ph = 7.54, pco2 = 17.5, po2 = 168.5 POC VBG#2: ph = 7.56, pco2 = 12.7, po2 = 165.9 End tidal CO2 = 28 Lab Venous Blood Gas: ph 7.54, pco2 = 19.2, po2 = 161.5 Blood Gas Laboratory identified air bubbles in syringe 45

Air Contamination of Blood Specimens Background: Anesthesia reports impossible venous blood gas values in one patient where end tidal CO2 was greater than the venous blood gas (VBG). POC Venous Blood Gas: ph = 7.54, pco2 = 17.5, po2 = 168.5 POC VBG#2: ph = 7.56, pco2 = 12.7, po2 = 165.9 End tidal CO2 = 28 Lab Venous Blood Gas: ph 7.54, pco2 = 19.2, po2 = 161.5 Air bubbles can quickly (<5 mins) cause the specimen to equilibrate atmospheric air (1 atm = 760 mmhg = 0.21 x 760 = 150 mmhg for po2!!!) 46

INTERFERENCES IN BLOOD GAS ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 47

Specimen Processing Delays and Lactate Pre-Analytical Transportation delays Analysis should be performed within 20 to 30 minutes Faster is better! 48

Specimen Processing Delays and Lactate Pre-Analytical Transportation delays Seymour CW, et al. BMC Research Notes 2011;4:169 49

Specimen Processing Delays and Lactate Pre-Analytical Transportation delays Inadequate inhibition of glycolysis If delays are expected, using a grey top tube may be appropriate, however it may take up to 15 minutes to achieve inhibition! 50

Specimen Processing Delays and Lactate Pre-Analytical Transportation delays Inadequate inhibition of glycolysis Astles R, et al. Clin Chem 1994;404:1327 51

Specimen Processing Delays and Lactate Pre-Analytical Transportation delays Inadequate inhibition of glycolysis If delays are expected, using a grey top tube may be appropriate, however it may take up to 15 minutes to achieve inhibition! Astles R, et al. Clin Chem 1994;404:1327 52

Specimen Processing Delays and Lactate Pre-Analytical Transportation delays Inadequate inhibition of glycolysis Specimens not placed on ice False elevations of lactate could be mitigated by placing samples on ice. Iced samples exhibit similar results to those tested immediately at up to 6 hours. 53

Specimen Processing Delays and Lactate Pre-Analytical Transportation delays Inadequate inhibition of glycolysis Specimens not placed on ice Seymour CW, et al. BMC Research Notes 2011;4:169 54

INTERFERENCES IN WHOLE BLOOD ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 55

Contemporary Hemoglobinometric Techniques Spectrophotometric (Non-Cyanohemoglobin) Absorbance 500 550 600 650 700 nm White light Prism HHb O 2 Hb COHb MetHb Measurement of hemoglobin is based on the absorption spectra Oxy- and deoxyhemoglobin exhibit different absorption in the red to IR wavelengths. Measurement based on Beer s Law (A = elc). Some methods require lysis and reacting with non-cyanide-based reagents. 56

Contemporary Hemoglobinometric Techniques Conductance (Impendance) Electrode High Resistance VS. Low Resistance Red blood cell membranes are not conductive. 57

Contemporary Hemoglobinometric Techniques Conductance (Impendance) Electrode VS. Red blood cell membranes are not conductive. Resistance (Ω) Hematocrit (%) 58

Contemporary Hemoglobinometric Techniques Conductance (Impendance) Electrode VS. Red blood cell membranes are not conductive. The number of red blood cells is proportional to the change in conductance and conforms to Ohm s Law (V = IR) 59

Contemporary Hemoglobinometric Techniques Conductance (Impendance) Electrode VS. Red blood cell membranes are not conductive. The number of red blood cells is proportional to the change in conductance and conforms to Ohm s Law (V = IR) Conductance-based methods measure hematocrit. The hematocrit can then be used to calculate hemoglobin based on a conversion factor (estimated hemoglobin = hematocrit / 3.4)* 60

Contemporary Hemoglobinometric Techniques Conductance (Impendance) Electrode VS. 61

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. 62

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. 63

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs Specimens collected for chemistry and CBC testing. 64

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs Specimens collected for chemistry and CBC testing. 65

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs Specimens collected for chemistry and CBC testing. 66

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs Specimens collected for chemistry and CBC testing. Handheld Results Hct = 68% Hb = 21.9 g/dl 67

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs Specimens collected for chemistry and CBC testing. Handheld Results Hct = 68% Hb = 21.9 g/dl CBC Results Hct = 41% Hb = 13.2 g/dl 68

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs Specimens collected for chemistry and CBC testing. Handheld Results Hct = 68% Hb = 21.9 g/dl CBC Results Hct = 41% Hb = 13.2 g/dl RE-MIXING! Hct = 43% Hb = 13.8 g/dl 69

Case Study 2: Hemoconcentration Background: Patient with suspected Ebola Virus symptoms admitted for evaluation. Isolation protocols were in effect. A handheld blood gas chemistry analyzer served as the primary chemistry analyzer. 0853 hrs Specimens collected for chemistry and CBC testing. Handheld Results Hct = 68% Hb = 21.9 g/dl RE-MIXING! Hct = 43% Hb = 13.8 g/dl CBC Results Hct = 41% Hb = 13.2 g/dl Inadequate mixing may result in artificial changes in total hemoglobin measurements. 70

Contemporary Hemoglobinometric Techniques Conductance (Impendence) = Plasma Protein Electrode High Resistance Plasma protein content contributes to hematocrit measurements for conductance-based systems. 71

Contemporary Hemoglobinometric Techniques Conductance (Impendence) = Plasma Protein Electrode Low Resistance from low plasma protein concentration! Plasma protein content contributes to hematocrit measurements for conductance-based systems. Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin. 72

Contemporary Hemoglobinometric Techniques Conductance (Impendence) = Plasma Protein Electrode Plasma protein content contributes to hematocrit measurements for conductance-based systems. Conductance-based systems assumes a relatively fixed protein concentration. Therefore, during hemodilution, hematocrit may be falsely lower and causing an underestimation of total hemoglobin. UCDMC Study: Comparison of a handheld blood gas analyzer using conductance-based measurement of hemoglobin versus a benchtop blood gas analyzer using a spectrophotometricbased method for hemoglobinometry. 73

Clinical Impact of Hemodilution for Point-of- Care Hemoglobin Measurements Sixty patients requiring cardiac surgery were evaluated. Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory. Mean (SD) bias was -1.4 (1.1) g/dl, P = 0.011. Based on core laboratory results 12 patients would have received unnecessary transfusions. 74

Clinical Impact of Hemodilution for Point-of- Care Hemoglobin Measurements Sixty patients requiring cardiac surgery were evaluated. Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory. Mean (SD) bias was -1.4 (1.1) g/dl, P = 0.011. Based on core laboratory results 12 patients would have received unnecessary transfusions. 75

Clinical Impact of Hemodilution for Point-of- Care Hemoglobin Measurements Sixty patients requiring cardiac surgery were evaluated. Paired specimens were tested using a handheld POC analyzer and spectrophotometric methods through the core laboratory. = $219 Mean (SD) bias was -1.4 (1.1) g/dl, P = 0.011. Based on core laboratory results 12 patients would have received unnecessary transfusions. Toner RW, et al. Appl Health Econ Health Policy 2011;9:29-37 $219 x 12 = $2,628 POTENTIALLY WASTED 76

Case Study 3: Hemodilution 77

Case Study 3: Hemodilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. 78

Case Study 3: Hemodilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. POC device reported a hematocrit of 22%. Physician administered 7 ml of blood based on the POC result. 79

Case Study 3: Hemodilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. POC device reported a hematocrit of 22%. Physician administered 7 ml of blood based on the POC result. Transfusion was stopped halfway after the laboratory reported a hematocrit of 40% and hemoglobin of 11.7 g/dl. 80

Case Study 3: Hemodilution Background: FDA MAUDE database reports a case (03P76-25) of a neonatal patient with discrepant point-of-care (POC) hemoglobin values compared to the laboratory. The POC device used a conductance-based method of hemoglobin measurement, while the laboratory used a spectrophotometric method. POC device reported a hematocrit of 22%. Physician administered 7 ml of blood based on the POC result. Transfusion was stopped halfway after the laboratory reported a hematocrit of 40% and hemoglobin of 11.7 g/dl. Post-transfusion POC and lab hematocrit values were 45 and 50% respectively. 81

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry 11 Device #1 Hb (g/dl) 10 9 8 7 6 5 y = 0.5092x + 4.0176 R² = 0.5253 4 4 5 6 7 8 9 10 11 12 Central Laboratory Hb (g/dl) 82

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry 11 Device #2 Hb (g/dl) 10 9 8 7 6 5 y = 0.5249x + 3.9443 R² = 0.5407 4 4 5 6 7 8 9 10 11 12 Central Laboratory Hb (g/dl) 83

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Device #3 Hb (g/dl) 11 10 9 8 7 6 5 y = 0.9345x + 0.4057 R² = 0.9205 4 4 5 6 7 8 9 10 11 Central Laboratory Hb (g/dl) 84

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Notes: Reference Method = Beckman LH hematology analyzer 85

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Median (IQR) Bias: 0.78 (0.78) g/dl P < 0.001 N = 50 Notes: Reference Method = Beckman LH hematology analyzer 86

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Median (IQR) Bias: 0.73 (0.60) g/dl P < 0.001 N = 50 Notes: Reference Method = Beckman LH hematology analyzer 87

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Median (IQR) Bias: 0.22 (0.20) g/dl P = 0.510 N = 50 Notes: Reference Method = Beckman LH hematology analyzer 88

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Total Hemoglobin (g/dl) 10 9.5 9 8.5 8 7.5 7 6.5 *** Serial Testing Performance at 7 and 8 g/dl Serial testing revealed significant analytical bias between spectrophotometry vs. conductance-based measurements. Spectrophotometric-based Methods 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 89

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Total Hemoglobin (g/dl) 10 9.5 9 8.5 8 7.5 7 6.5 *** Serial Testing Performance at 7 and 8 g/dl Serial testing revealed significant analytical bias between spectrophotometry vs. conductance-based measurements. Spectrophotometric-based Methods Conductance-based Methods 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 90

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Total Hemoglobin (g/dl) 10 9.5 9 8.5 8 7.5 7 6.5 *** Unnecessary Transfusion Risk Serial Testing Performance at 7 and 8 g/dl Serial testing revealed significant analytical bias between spectrophotometry vs. conductance-based measurements. Conductance-based devices would have prompted unnecessary transfusions at time point #5 for patients using the 7 g/dl cutoff. 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 91

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Total Hemoglobin (g/dl) 10 9.5 9 8.5 8 7.5 7 6.5 *** Unnecessary Transfusion Risk Serial Testing Performance at 7 and 8 g/dl Serial testing revealed significant analytical bias between spectrophotometry vs. conductance-based measurements. Conductance-based devices would have prompted unnecessary transfusions at time point #5 for patients using the 7 g/dl cutoff. All serial conductance measurements were at risk for potential transfusions if the 8 g/dl cutoff was used. 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 92

Analytical Performance of Optical vs. Conductance-Based Hemoglobinometry Total Hemoglobin (g/dl) 10 9.5 9 8.5 8 7.5 7 6.5 *** Unnecessary Transfusion Risk Central Laboratory Method Serial Testing Performance at 7 and 8 g/dl Serial testing revealed significant analytical bias between spectrophotometry vs. conductance-based measurements. Conductance-based devices would have prompted unnecessary transfusions at time point #5 for patients using the 7 g/dl cutoff. All serial conductance measurements were at risk for potential transfusions if the 8 g/dl cutoff was used. 6 1 2 3 4 5 Time Point Notes: *** P<0.001, Central Lab = Spectrophotometric Method, n = 20 patients 93

Manufacturer and User Facility Device Experience (MAUDE) Database Summary Device 1 Device 2 Device 3 Timeframe 2011-2016 2011-2016 2014-2016* Erroneous Results 8 0 0 Improper Transfusions 5 0 0 https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/results.cfm, Accessed on July 19, 2016 94

INTERFERENCES IN WHOLE BLOOD ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration Hemolysis 95

INTERFERENCES IN WHOLE BLOOD ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration Pseudohyperkalemia Hemolysis 96

INTERFERENCES IN WHOLE BLOOD ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration Pseudohyperkalemia Pseudonormokalemia Hemolysis 97

INTERFERENCES IN WHOLE BLOOD ANALYSIS Air Contamination Delayed Testing Hemodilution/Hemoconcentration No current FDA approved integrated solutions for detecting hemolysis at the point-of-care Pseudohyperkalemia Pseudonormokalemia Hemolysis 98

Biotin: The Snake Oil of 2018? 99

Biotin and Cardiac Troponin Testing 101

Estimating the Probability of Biotin Interference 1,443 Gen 5 troponin T samples tested (0-hour, n = 797; 3-hour, n=646) from 850 patients. Mumma B, et al. AACC Poster Presentation 2018 102

Estimating the Probability of Biotin Interference 1,443 Gen 5 troponin T samples tested (0-hour, n = 797; 3-hour, n=646) from 850 patients. Biotin not detectable in 471 (59%) and 399 (62%) 3- hour samples. Mumma B, et al. AACC Poster Presentation 2018 103

Estimating the Probability of Biotin Interference 1,443 Gen 5 troponin T samples tested (0-hour, n = 797; 3-hour, n=646) from 850 patients. Biotin not detectable in 471 (59%) and 399 (62%) 3- hour samples. Only one 0-hour sample and one 3-hour sample had biotin >20 ng/ml (0.13% [95% CI: 0-0.7%]). Mumma B, et al. AACC Poster Presentation 2018 104

Estimating the Probability of Biotin Interference Mumma B, et al. AACC Poster Presentation 2018 105

Estimating the Probability of Biotin Interference 106

UC Davis Cardiac Troponin Patients Adult ED Patients with Unknown Biotin Status: 540 Average Plasma Biotin: 1.15 (0.97) ng/ml Specimens collected as part of clinical validation 107

UC Davis Cardiac Troponin Patients 160 140 Adult ED Patients with Unknown Biotin Status: 540 Average Plasma Biotin: 1.15 (0.97) ng/ml Gen 5 TnT Biotin Interference Threshold is 20 ng/ml Number of Patients 120 100 80 60 40 20 0 Biotin quantified by GC-TOF-MS Biotin Concentration (ng/ml) 108

UC Davis Cardiac Troponin Patients Adult ED Patients with Unknown Biotin Status: 540 Average Plasma Biotin: 1.15 (0.97) ng/ml Gen 5 TnT Biotin Interference Threshold is 20 ng/ml Number of Patients 160 140 120 100 80 60 40 20 0 BIOTIN IS LESS LIKELY TO BE A PROBLEM IN CARDIAC TROPONIN TESTING AND IS POPULATION SPECIFIC! Biotin quantified by GC-TOF-MS Biotin Concentration (ng/ml) 109

Biotin and Urine Pregnancy Testing 110

Biotin Interference with Urine Pregnancy Tests Recent studies show some point-ofcare urine pregnancy tests were affected by biotin. Biotin is cleared by the kidneys. In this study, the QuickVue urine pregnancy test exhibited interference as low as 6 microgram/ml of urine biotin! Williams G, et al. Clin Biochem 2018;53:168-170 111

Best POCT Practices for Mitigating Interfering Substances 112

POCT Best Practices for Interferences Education: The laboratory must be the leader in educating providers and patients of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge. 113

POCT Best Practices for Interferences Education: The laboratory must be the leader in educating providers and patients of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge. 114

POCT Best Practices for Interferences Education: The laboratory must be the leader in educating providers and patients of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge. Surveillance: Know your population! Collect data and determine if your local population may be be at risk for certain interferences (e.g., biotin, vitamin C, etc). MAUDE database is also helpful! 115

POCT Best Practices for Interferences Education: The laboratory must be the leader in educating providers of potential test interferences. Go to grand rounds, build partnerships, and provide multi-modality means to disseminate knowledge. Surveillance: Know your population! Collect data and determine if your local population may be be at risk for certain interferences (e.g., biotin, vitamin C, etc). MAUDE database is also helpful! Electronic Early-Warning Systems: Leverage electronic solutions. Ordering of susceptible tests could flag both on the provider and laboratory side certain substances are identified. 116

Conclusions Interfering substances are out there and impact POC testing as much as traditional lab testing! Interferences in common POC devices such as glucose meters have resulted in injury and death. Interferences in whole blood analysis have resulted in inappropriate treatment decisions. Medications and supplements may also affect POC immunoassays such as urine pregnancy tests. Education and awareness is critical to minimizing errors associated with interfering substances. 117

Questions? 118