Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Similar documents
An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

EPID dosimetry confi guration and pre-treatment IMRT verifi cation

Leila E. A. Nichol Royal Surrey County Hospital

Optimized Varian asi portal dosimetry: development of datasets for collective use

EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams

Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

Small field diode dosimetry

Feasibility of portal dosimetry for flattening filter-free radiotherapy

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

Portal dosimetry in wedged beams

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

iviewdose Confidence and assurance in dose delivery

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

REPLACING PRETREATMENT VERIFICATION WITH IN VIVO EPID DOSIMETRY FOR PROSTATE IMRT

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Limits of Precision and Accuracy of Radiation Delivery Systems

Volumetric Modulated Arc Therapy - Patient Specific QA

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

EPID dosimetry for pretreatment quality assurance with two commercial systems

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

Verification of Advanced Radiotherapy Techniques

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

IMRT QA: Point Dose Measurements or 2D Array?

Gamma analysis dependence on specified low-dose thresholds for VMAT QA

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

SunCHECK Patient Comprehensive Patient QA

Intensity Modulated RadioTherapy

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

PMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.

Effective and efficient radiotherapy dosimetry audit: Where to next?

Transmission portal in vivo dosimetry by means of the Monte Carlo method and the mathematical programming language MATLAB

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

Radiotherapy In-vivo Treatment Verification: Multi-center analysis for more than 6 million fractions delivered

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

A new transmission methodology for quality assurance in radiotherapy based on radiochromic film measurements

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

Breast in vivo dosimetry by EPID

Eric E. Klein, Ph.D. Chair of TG-142

A new geometric and mechanical verification device for medical LINACs

EORTC Member Facility Questionnaire

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Diode calibration for dose determination in total body irradiation

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Review of four novel dosimeters developed for use in radiotherapy

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

I. Equipments for external beam radiotherapy

Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate

Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy

IMRT point dose measurements with a diamond detector

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Intensity modulation techniques for spinal treatments and on-line VolumeView TM guidance

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Electron Beam ET - Reloaded Therapy - Reloaded

Collapsed Cone Convolution 2D illustration

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Mitsubishi Heavy Industries Technical Review Vol. 51 No. 1 (March 2014)

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

Hampton University Proton Therapy Institute

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

biij Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies Krishna Murthy K *

Commissioning and quality assurance of a commercial stereotactic treatment-planning system for extracranial IMRT

The impact of continuously-variable dose rate VMAT on beam stability, MLC positioning, and overall plan dosimetry

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

Transcription:

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor MI 48109-0010 jmmoran@med.umich.edu Abstract. As the complexity of radiation therapy has increased, the need for quantitative dosimetric evaluation of treatment delivery has also increased. A growing number of investigations have expanded the use of EPIDs from anatomic applications to dosimetric verification. This work focuses on the applications of EPIDs for pre-treatment dosimetric verification of IMRT and intensity modulated arc therapy techniques. The advantages and disadvantages of these techniques are discussed along with methods to extrapolate to 3D dose verification applications. 1. Introduction The evolution of radiation therapy treatment planning and delivery methods has required a similar evolution for the dosimetric methods for evaluation. Delivery techniques such as intensity modulated radiation therapy (IMRT) and intensity modulated arc therapy (IMAT) require more detailed spatial information. As the need for pre-treatment patient-specific quality assurance has increased, so have investigations into the use of electronic portal imager detectors (EPIDs). EPIDs offer advantages compared to other systems in that they are attached to the gantry and are sold with the majority of new accelerator purchases. They are also efficient to use and do not require the set up of any additional equipment once the system has been commissioned for such a purpose. However, the response of EPIDs to radiation is complex due to their composition. These systems were designed and optimized to require as little delivered dose as possible to obtain a high quality image that can be used for patient localization. EPIDs have been investigated for both pre-treatment and transit dosimetry applications. These different types of portal dosimetry methods have been comprehensively reviewed by van Elmpt et al.[1] Transit or in vivo dosimetry methods have been developed and reported on by investigators at the Netherlands Cancer Institute (see conference paper by Mijnheer et al).[1-5] These methods typically utilize the commercial configuration of the device. Additional software tools and acquisition methods have been developed with EPIDs to verify the accuracy of a delivered treatment pretreatment without a patient and with or without a phantom present. The majority of these pre-treatment investigations have been performed utilizing commercial EPID systems in an indirect detection configuration where the system includes a scintillation-layer as part of the manufacturer supplied configuration. Other investigators have used direct detection configurations where no scintillation layer is present. This work focuses on methods developed for amorphous silicon-based systems for pre-treatment quality assurance methods. 1 c 2010 Ltd

2. EPID Dosimetry with Standard Configurations Amorphous silicon EPIDs have been investigated for dosimetric applications for IMRT and arc therapy techniques. Reliable use of the system requires commissioning of the planning and delivery system prior to development of the EPID-based program. Then, it may be appropriate to establish a pre-treatment QA program based on calculations in the EPID compared with the EPID response. These methods typically rely on determining the appropriate parameters of operation of the EPID and accurate determination of correction factors and the accuracy of the calculation of a predicted portal dose image (PDI).[6-12] The features of the measurement are affected by the frequency of the data acquisition and communication speeds. The response itself can be affected by ghosting and lag. The response of these systems depends on the field size. The presence of the scintillator results in an increased sensitivity to low energies. For example, the effect of the different layers of material and backscatter on the detector response as a function of energy has been quantified using Monte Carlo calculations.[13] Therefore the response of the system is dependent on the energy spectrum of the delivered beam. In some investigations, additional buildup is added to provide additional attenuation of the beam. Initial investigations demonstrated a clear dose rate effect due to saturation of the detector.[8] 2.1. Calculating the Predicted Portal Dose Image (PDI) For pre-treatment QA, the EPID response is measured and then compared to the predicted portal dose image (PDI). This method is used to verify the accuracy of the intended fluence. The measured response is compared to a calculation in the imager rather than converting the imager response to dose in water. Initial methods used pencil beam methods to predict the portal dose image.[8] Other methods involved calculating separate fluences for the primary and scatter components.[6, 7] More recently, investigators have used Monte Carlo to create EPID-specific dose kernels. For example, one method uses Monte Carlo to incorporate details of the multileaf collimator design such as transmission, tongue-and-groove, and curved leaf ends.[9] In another study, Monte Carlo was used to generate imager-specific dose kernels to investigate different thicknesses of backscattering.[14] Other considerations included determination that the effect of backscatter due to the metallic components of the support arm can be as much as 6% for large fields.[15] With the corrections applied, the majority of these studies found agreement of >90% of the points to be within criteria 2%/2 mm for dose and distance, respectively for IMRT and IMAT deliveries. These studies demonstrate the complexity of modeling required for robust use of EPIDs for IMRT pretreatment QA. 2.2. Measuring the Portal Dose Image (PDI) When operated in an integrative mode during delivery, amorphous silicon EPIDs have been found to have a reproducible response which is essential for dosimetric applications.[7, 8] Investigators have characterized the response for a range of field sizes, doses, dose rates, and source-to-detector distances. Continuous acquisition mode has also been investigated for EPID verification for IMRT and IMAT delivery methods to allow for discrimination of time-based information. In this mode, some information is lost during the readout and the system was determined to have a non-linear response to dose for low doses.[12] The system was found to be reliable for the IMRT and arc therapy fields that were evaluated. For arc and IMRT applications, there are discrepancies for deliveries of low monitor units.[8, 10, 12] These need to be further evaluated to ensure that EPID techniques are only used in validated situations. 3. Dosimetry with Direct Detection Configurations Due to the difficulties in modeling scatter and the fact that the measured response differs substantially from dose in water, work has also been done using amorphous silicon-based systems in a direct detection configuration.[16-20] Because these systems do not have a scintillator, the overall signal is much lower. The measured signal is converted to dose more simply than with indirect detection 2

systems although there the dark field must be adequately measured. The devices can be operated in continuous or integrated modes with considerations similar to commercial systems about loss of data in the continuous acquisition mode.[16, 18] The configuration results in measured results that are much more similar to those in water, even with minimal backscatter.[20] 4. 3D Applications In vivo dosimetry methods rely on the backprojection of the measured response to a plane in the patient or phantom.[1] With cone beam CT data more readily acquired as part of a patient s treament, 3D verification of the actual delivered dose is becoming feasible. Methods for 3D dose verification have been developed for hypofractionated lung treatments[4, 21] and prostate, rectum, and head-andneck treatments[5]. Methods are being developed to account for artifacts in the cone beam CT images.[22, 23] Monte Carlo methods have been used to reconstruct the dose in a phantom as a function of the measured exit dose for an IMRT treatment.[24] These methods promise must more information about the accuracy of the delivered dose to patients. 5. Summary As radiation therapy continues to increase in complexity, it is crucial to have validated and easy to use methods to verify that the correct treatment plan was delivered to the patient. Information that has previously been obtained solely for adaptive therapy purposes can be used to develop more sophisticated models of the patient over the course of treatment. Methods to use commercial EPID systems for verification of IMRT and IMAT treatments have been developed and are in use in some centers. When used in the standard configuration (indirect mode), a number of correction factors must be applied to the measurement and to properly calculate the predicted portal dose image. Transit dosimetry methods have also been developed and the derived information is being coupled successfully with cone beam CT information to develop models of the 3D delivered doses. The accuracy of these methods must be validated with dosimetry systems of known performance. From a safety perspective, the use of EPID information to reconstruct 3D doses provides much more information about the quality of a patient delivery. This area is expected to mature over the next few years. Table 1. Summary of investigations into the use of EPIDs for pre-treatment verification in a standard configuration for advanced technology treatments. First Author/Year Method Conditions Limitations van Esch et al. 2004 Measurement: integrated acquisition mode; averaged gray scale image converted to portal dose image (PDI) Calculation: predicted PDI based on a pencil beam method 3x3 cm 2 to 25 x25 cm 2 for SDD=105 cm; SMLC/DMLC Chytyk et al. 2009 Improved calculation: Two-source fluence model with focal and extra-focal sources; more detailed modeling of jaws and MLC; Monte Carlo used to derive EPID-specific 3 Derived parameters with 1x1 cm 2 up to 20x20 cm 2 fields Tested: Fields for 10 prostate and 10 oropharyngeal patient plans Adversely affected by saturation of the detector at 600 mu/min dose rate Patient scatter not modelled and therefore not implemented for transit dosimetry

McCurdy et al. 2009 Iori et al. 2010 dose kernels Measurement: continuous acquisition mode Measurement: intensity-modulated arc therapy; accessory mount clamped between EPID and collimator housing to keep EPID position stable during rotation Calculation: pencil beam kernels derived from measurements IMRT and IMAT techniques 23 Treatment fields delivered with both IMRT and IMAT techniques Difficulties with verification of low MU; significant discrepancies from ion chamber measurements below 50 MU Compared to an ion chamber 10% disagreement for ~15 MU; (3.3% discrepancy 100 MU) References [1] W. van Elmpt, L. McDermott, S. Nijsten, M. Wendling, et al., "A literature review of electronic portal imaging for radiotherapy dosimetry". Radiother Oncol. 88: 289-309, 2008. [2] L.N. McDermott, M. Wendling, J. Nijkamp, A. Mans, et al., "3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT". Radiother Oncol. 86: 35-42, 2008. [3] L.N. McDermott, M. Wendling, J.J. Sonke, M. van Herk, and B.J. Mijnheer, "Replacing pretreatment verification with in vivo EPID dosimetry for prostate IMRT". Int J Radiat Oncol Biol Phys. 67: 1568-77, 2007. [4] W. van Elmpt, S. Nijsten, B. Mijnheer, A. Dekker, and P. Lambin, "The next step in patientspecific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations". Radiother Oncol. 86: 86-92, 2008. [5] M. Wendling, L.N. McDermott, A. Mans, J.J. Sonke, et al., "A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments". Med Phys. 36: 3310-21, 2009. [6] B.M. McCurdy and S. Pistorius, "A two-step algorithm for predicting portal dose images in arbitrary detectors". Med Phys. 27: 2109-16, 2000. [7] B.M. McCurdy, K. Luchka, and S. Pistorius, "Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device". Med Phys. 28: 911-24, 2001. [8] A. Van Esch, T. Depuydt, and D.P. Huyskens, "The use of an asi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields". Radiother Oncol. 71: 223-34, 2004. [9] K. Chytyk and B.M. McCurdy, "Comprehensive fluence model for absolute portal dose image prediction". Med Phys. 36: 1389-98, 2009. [10] M. Iori, E. Cagni, M. Paiusco, P. Munro, and A.E. Nahum, "Dosimetric verification of IMAT delivery with a conventional EPID system and a commercial portal dose image prediction tool". Med Phys. 37: 377-90, 2010. 4

[11] A. Kavuma, M. Glegg, M. Metwaly, G. Currie, and A. Elliott, "A novel method for patient exit and entrance dose prediction based on water equivalent path length measured with an amorphous silicon electronic portal imaging device". Phys Med Biol. 55: 435-52, [12] B.M. McCurdy and P.B. Greer, "Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT". Med Phys. 36: 3028-39, 2009. [13] H. Gustafsson, P. Vial, Z. Kuncic, C. Baldock, and P.B. Greer, "EPID dosimetry: effect of different layers of materials on absorbed dose response". Med Phys. 36: 5665-74, 2009. [14] S. Wang, J.K. Gardner, J.J. Gordon, W. Li, et al., "Monte Carlo-based adaptive EPID dose kernel accounting for different field size responses of imagers". Med Phys. 36: 3582-95, 2009. [15] P. Rowshanfarzad and B.M. McCurdy, "Measurement and modeling of the effect of support arm backscatter on dosimetry with a Varian EPID". Med Phys. 37: 2269-2278, 2010. [16] Y. Chen, J.M. Moran, D.A. Roberts, Y. El-Mohri, et al., "Performance of a direct-detection active matrix flat panel dosimeter (AMFPD) for IMRT measurements". Med Phys. 34: 4911-22, 2007. [17] Y. El-Mohri, L.E. Antonuk, J. Yorkston, K.W. Jee, et al., "Relative dosimetry using active matrix flat-panel imager (AMFPI) technology". Med Phys. 26: 1530-41, 1999. [18] J.M. Moran, D.A. Roberts, T.S. Nurushev, L.E. Antonuk, et al., "An Active Matrix Flat Panel Dosimeter (AMFPD) for in-phantom dosimetric measurements". Med Phys. 32: 466-72, 2005. [19] P. Vial, P.B. Greer, L. Oliver, and C. Baldock, "Initial evaluation of a commercial EPID modified to a novel direct-detection configuration for radiotherapy dosimetry". Med Phys. 35: 4362-74, 2008. [20] P. Vial, H. Gustafsson, L. Oliver, C. Baldock, and P.B. Greer, "Direct-detection EPID dosimetry: investigation of a potential clinical configuration for IMRT verification". Phys Med Biol. 54: 7151-69, 2009. [21] W. van Elmpt, S. Petit, D. De Ruysscher, P. Lambin, and A. Dekker, "3D dose delivery verification using repeated cone-beam imaging and EPID dosimetry for stereotactic body radiotherapy of non-small cell lung cancer". Radiother Oncol. 94: 188-94, 2010. [22] S.F. Petit, W.J. van Elmpt, P. Lambin, and A.L. Dekker, "Dose recalculation in megavoltage cone-beam CT for treatment evaluation: removal of cupping and truncation artefacts in scans of the thorax and abdomen". Radiother Oncol. 94: 359-66, 2010. [23] S.F. Petit, W.J. van Elmpt, S.M. Nijsten, P. Lambin, and A.L. Dekker, "Calibration of megavoltage cone-beam CT for radiotherapy dose calculations: correction of cupping artifacts and conversion of CT numbers to electron density". Med Phys. 35: 849-65, 2008. [24] I.J. Yeo, J.W. Jung, M. Chew, J.O. Kim, et al., "Dose reconstruction for intensity-modulated radiation therapy using a non-iterative method and portal dose image". Phys Med Biol. 54: 5223-36, 2009. 5