Quality Control of DXA System and Precision Test of Radio-technologists

Similar documents
9 Quality Assurance in Bone Densitometry section

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status

Documentation, Codebook, and Frequencies

Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p.

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7

The Bone Densitometry Examination

Bone Densitometry Equipment Operator

Quality Assurance and Control in Dual energy X ray Absorptiometry

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems?

DXA Best Practices. What is the problem? 9/29/2017. BMD Predicts Fracture Risk. Dual-energy X-ray Absorptiometry: DXA

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi

2013 ISCD Official Positions Adult

Standard Operating Procedure TCRC Dual-Energy X-ray Absorptiometry (DXA)

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302)

Purpose. Methods and Materials

The official position of the International Society for Clinical

ASJ. How Many High Risk Korean Patients with Osteopenia Could Overlook Treatment Eligibility? Asian Spine Journal. Introduction

Bone Densitometry. What is a Bone Density Scan (DXA)? What are some common uses of the procedure?

2013 ISCD Combined Official Positions

Bone Densitometry Radiation dose: what you need to know

CLINIQCT NO-DOSE CT BONE DENSITOMETRY FOR ROUTINE AND SPECIALIST USE.

Screening points for a peripheral densitometer of the calcaneum for the diagnosis of osteoporosis

Evaluation of Bone Mineral Status in Adolescent Idiopathic Scoliosis

QDR Series. Discovery and Explorer Advanced Health Assessment

Bone Densitometry at the Point of Care

Official Positions on FRAX

Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA

Concordance of a Self Assessment Tool and Measurement of Bone Mineral Density in Identifying the Risk of Osteoporosis in Elderly Taiwanese Women

Norland Densitometry A Tradition of Excellence

Annotations Part III Vertebral Fracture Initiative. International Osteoporosis Foundation March 2011

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Omnisense: At Least As Good As DXA

CT Quality Control Manual FAQs

X-ray (Radiography) - Bone

New Dual-energy X-ray Absorptiometry Machines (idxa) and Vertebral Fracture Assessment

Efficacy of risedronate in men with primary and secondary osteoporosis: results of a 1-year study

Clinical Densitometry

Mammography Quality Control: A Refresher

Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry

Prevalence of Sarcopenia Adjusted Body Mass Index in the Korean Woman Based on the Korean National Health and Nutritional Examination Surveys

Live Educational Programs

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT

Obesity Increases Precision Errors in Total Body Dual X-ray Absorptiometry Measurements.

Clinical Study Comparison of QCT and DXA: Osteoporosis Detection Rates in Postmenopausal Women

University of Groningen. Osteoporosis, identification and treatment in fracture patients de Klerk, Gijsbert

An audit of bone densitometry practice with reference to ISCD, IOF and NOF guidelines

Objectives: What is Osteoporosis 10/8/2015. Bone Health/ Osteoporosis: BASICS OF SCREENING, INTERPRETING, AND TREATING

The DXL Calscan heel densitometer: evaluation and diagnostic thresholds

DIAGNOSTIC ACCREDITATION PROGRAM. Spirometry Quality Control Plan

DEVELOPMENT OF A RISK SCORING SYSTEM TO PREDICT A RISK OF OSTEOPOROTIC VERTEBRAL FRACTURES IN POSTMENOPAUSAL WOMEN

pqct Measurement of Bone Parameters in Young Children

Available online at ScienceDirect. Osteoporosis and Sarcopenia 1 (2015) 109e114. Original article

Introducing the future of DXA. Powerful images. Clear answers. Horizon DXA System

Audit on follow-up of patients with primary Osteoporosis

QCT and CT applications in Osteoporosis Imaging

O. Bruyère M. Fossi B. Zegels L. Leonori M. Hiligsmann A. Neuprez J.-Y. Reginster

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers*

Foreword...v Preface...vii Acknowledgments... xi Dedication... xiii Continuing Medical Education... xxv. Chapter 1: Densitometry Techniques...

X-ray (Radiography) - Chest

DXA scanning to diagnose osteoporosis: Do you know what the results mean?

Bone Mineral Density and Its Associated Factors in Naresuan University Staff

Accuracy of DEXA scanning & other methods for determining BMD.

BONE DENSITOMETRY FOR TECHNOLOGISTS

Diagnostische Präzision von DXL im Vergleich zu DXA bei pmp Frauen mit Frakturen

Evaluation of measurement uncertainty of urine output using two kinds of urine bags

Identification of Early Osteoporosis Using Intensity Slicing method

Scintimammography. What is scintimammography?

Using GE Lunar DXA to Quantify, Visualize, and Trend Incipient Atypical Femoral Fractures

ReviewArticle. Frequently Asked Questions About Bone Mineral Density Test

ANNUAL REPORT OF POST-PRIMARY EXAMINATIONS 2014

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare.

Module 5 - Speaking of Bones Osteoporosis For Health Professionals: Fracture Risk Assessment. William D. Leslie, MD MSc FRCPC

Obesity Increases Precision Errors in Dual X-ray Absorptiometry Measurements.

International Journal of Health Sciences and Research ISSN:

Analysis of Clinical Features of Hip Fracture Patients with or without Prior Osteoporotic Spinal Compression Fractures

Temporal Trends in Bone Mineral Density, Body Mass Index and Fracture Rates: Implications for Osteoporosis Diagnosis and FRAX

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

Skeletal Sites for Osteoporosis Diagnosis: The 2005 ISCD Official Positions

This is a repository copy of Microarchitecture of bone predicts fractures in older women.

The Relationship between Prevalence of Osteoporosis and Proportion of Daily Protein Intake

nogg Guideline for the diagnosis and management of osteoporosis in postmenopausal women and men from the age of 50 years in the UK

Association between Sacral Slanting and Adjacent Structures in Patients with Adolescent Idiopathic Scoliosis

Portfolio Requirements and Instructions 2018

Pharmacy Management Drug Policy

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm

ACCURATE IDENTIFICATION of individuals at risk for

The New Jersey Radiographic Quality Assurance Program at 5 Years

Osteoporosis International. Original Article. Bone Mineral Density and Vertebral Fractures in Men

A FRAX Experience in Korea: Fracture Risk Probabilities with a Country-specific Versus a Surrogate Model

Lunar Prodigy Advance

Incidence and Mortality after Proximal Humerus Fractures Over 50 Years of Age in South Korea: National Claim Data from 2008 to 2012

TigerPrints. Clemson University. Nadia Marie Ghassan Najm Clemson University

Factors Associated with Treatment Initiation after Osteoporosis Screening

Clinical Application of Computed Radiography in Orthopedic Surgery

Inadequate Dietary Calcium and Vitamin D Intake in Patients with Osteoporotic Fracture

Transcription:

J Bone Metab 2014;21:2-7 http://dx.doi.org/10.11005/jbm.2014.21.1.2 pissn 2287-6375 eissn 2287-7029 Review Article Quality Control of DXA System and Precision Test of Radio-technologists Ho-Sung Kim 1, Seoung-Oh Yang 2 1 Department of Nuclear Medicine, Asan Medical Center, Ulsan University School of Medicine, Seoul; 2 Department of Nuclear Medicine (Radiology), Dongnam Institute of Radiological and Medical Science, Busan, Korea Corresponding author Seoung-Oh Yang Department of Nuclear Medicine (Radiology), DIRAMS, 40 Jwadong-gil, Jangan-eup, Gijanggunm, Busan 619-953, Korea Tel: +82-51-720-5060 Fax: +82-51-720-5913 E-mail: soyangmd@naver.com Received: October 14, 2013 Revised: October 22, 2013 Accepted: October 22, 2013 No potential conflict of interest relevant to this article was reported. The image quality management of bone mineral density (BMD) is the responsibility and duty of radio-technologists who carry out examinations. However, inaccurate conclusions due to the lack of understanding and ignorance regarding the methodology of image quality management can be a fatal error to patients. The accuracy and precision of BMD measurement must be maintained at the highest level so that actual biological changes can be detected with even slight changes in BMD. Accuracy and precision should be continuously preserved for image quality of machines. Those factors will contribute to ensure the reliability of BMD examination. The enforcement of proper quality control of radiologists performing BMD inspections which brings about the durability extensions of equipment and accurate results of calculations will help the assurance of reliable inspections. Key Words: Bone density, Densitometry, Dual-energy X-ray absorptiometry, Precision, Quality control INTRODUCTION Copyright 2014 The Korean Society for Bone and Mineral Research This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. Dual energy X-ray absorptiometry (DXA),[1] which is considered the most proper standard device for diagnosing osteoporosis, is the most commonly used due to its high accuracy and precision. In general, this device assesses bone mineral density (BMD) by examining the spine and femur. In addition, BMD examination using DXA is not a single-time examination, but requires follow-up for assessing BMD change or drug efficacy. Although BMD-measuring device requires higher precision than other medical devices, however, a further strict precision is required for the follow-up of osteoporotic patients as the range of BMD change is very narrow in osteoporotic patients.[2] Therefore, for the reproducibility of the examination, the quality control (QC) of DXA and precision test of radio-technologists should be accurately performed before applying examination results to treatment. For the maintenance of the consistency of BMD measurement, the QC of BMD measurement is essential.[3-5] Accordingly, the QC of BMD measurement using DXA is performed for minimizing the error range of precision and accuracy. For this end, error ranges from patient s condition, measuring device, and radio-technologist s experience should be minimized via quality control. QC is divided into device inspection and the assessment of radio-technologist s expertise. In the QC 2 http://e-jbm.org/

QC of DXA and Precision Test of Radio-technologists of device, it is very important to detect whether BMD value has been changed because BMD value could be changed depending on device condition. Daily QC is adequate for inspecting device condition on the same day, but inadequate for assessing the tendency of BMD change. Thus, changes in BMD value should be followed-up weekly and monthly. For the assessment of radio-technologist s precision (reproducibility), it is important to assess if radio-technologists conduct the examination or data analysis in a consistent manner. In other words, when the examination is repeatedly conducted on the same patient, the result should be same. Thus, adequate training and experience are required by radio-technologists for BMD examination. Accordingly, the QC of radio-technologist s precision is required, in which standard examination and analysis protocols are prepared to maintain the consistency of examination and data analysis conducted by radio-technologists. This paper reviews the methods of QC based on the International Society for Clinical Densitometry (ISCD) recommendation. 1. QC of DXA The QC of DXA varies depending on the features of each device manufactured by different companies. In daily QC which is conducted before BMD examination on a daily basis, a proper phantom, suitable for each device should be used. In addition, mechanical malfunction, radiation quality, absorption coefficient and tissue-equivalent materials are checked and calibrated. Only when the calibrated value is within the normal range, BMD examination can be conducted. As it is, however, not easy to detect the abnormality of BMD measurement, the tendency of the measured BMD values should be followed-up for maintaining the consistency of accuracy and precision unless malfunctioning is obviously found. 1) Selection of proper phantom Each phantom, recommended by device manufacturers, has its own features. In the case of Hologic, their device itself adjusts radiation quality, absorption coefficient and tissue-equivalent materials so that correction is made during the scanning of lumbar phantom for examining the tendency of BMD values. On the other hand, in the case of GE, their device adjusts radiation quality, absorption coefficient and tissue-equivalent materials using a block phantom and follows-up BMD values using a lumbar phantom. Therefore, in daily quality control, device-specified phantom should be used and various ranges of BMD should be measured using a general phantom to follow-up the tendency of the measured BMD values. General phantoms include European Spine Phantom (ESP) and Bona Fide Spine Phantom (BFP) (Fig. 1). 2) Management of single device After device installation, baseline BMD and intra system least significant change (LSC) should be set. According to the ISCD recommendation, the mean of BMD values, measured at the same location 20 times or measured daily for 20 days using a general phantom should be set as the baseline BMD and the control limit of upper and lower ranges should be set as baseline BMD±1.5%. Based on the range set, a pre-test is conducted daily or twice or more per week using a general phantom before actual use. The result of the test is presented in Shewhart chart or cumulative sum control chart (CUSUM), and assessed according to the corresponding rule in order to detect the shift or drift of BMD value.[6] The QC of BMD measuring device should be maintained at the highest condition for detecting delicate change and performed after calibration upon the translocation or installation of the device.[7-13] A B C D Fig. 1. General phantoms and scan images. (A) The European Spine Phantom (ESP; QRM GmbH Dorfstrasse 4. 91096, Möhrendorf, Germany). (B) ESP scan image. (C) Bio-imaging bone fide spine phantom (BFP; Bio-Imaging Technologies Inc., Newtown, PA, USA). (D) BFP scan image. http://dx.doi.org/10.11005/jbm.2014.21.1.2 http://e-jbm.org/ 3

Ho-Sung Kim, et al. 3) Cross calibration Cross calibration should be performed upon device replacement or the management of two or more devices. In the case of the replacement of device manufactured by the same company, cross calibration is conducted by a single radio-technologist, in which BMD is measured using the existing and replaced devices 10 times respectively to obtain mean BMD values and the mean BMD values are compared. If the mean BMD values are within the normal range, and their difference is within 1%, data before replacement can be used. However, if the difference is beyond 1%, calibration should be performed again to set the difference within 1%. In the case of the replacement of devices manufactured by different companies, if precision test has been conducted on the existing device, 30 patients should undergo scanning once using the existing device and then undergo scanning twice using the replaced device within 60 days after replacement. That is, those 30 patients undergo scanning three times, once using the existing device and twice using the replaced device. On the other hand, if precision test has not been conducted on the existing device, 30 patients should undergo scanning twice using the existing device and then undergo scanning twice using the replaced device within 60 days after replacement. That is, those 30 patients undergo scanning four times twice using the existing device and twice using the replaced device. Lumbar and proximal femur, which are commonly examined in clinical practices, are used as examination sites. Quantitative comparison of two devices can be performed by exanimation site via cross calibration. If cross calibration has not been conducted, previous scanning information cannot be directly compared with new information. In addition, if cross calibration has been improperly conducted, quantitative comparison between previous and new information cannot be performed. For the comparison of two devices, a calibration formula between the two devices is obtained using LSC and via regression analysis and then applied to obtain BMD values. Formula derivation is performed using a cross-calibration calculation tool provided by the ISCD. 4) DXA upgrade If software replacement is required due to DXA system upgrade, reasons for upgrade, change of boundary detecting algorithm, change of output system and the upgrade and characteristics of standard database should be inspected. In general, in the case of system upgrade, BMD is not affected though T-score and Z-score are affected. Thus, mean BMD before and after replacement should be compared and region of interest (ROI) change should be examined to ensure result consistency. 2. Measuring method and management of radio-technologist s precision In general, radio-technologist s precision is independently affected by factors including patient s movement, foreign materials, physical status, patient s location and scan analysis. In relation to significant factors that affect the overall precision of DXA, patients and radio-technologists are more important than the variability of the device. In the use of DXA, which is considered the most precise quantitative measurement in clinical practices, a precision test is performed. Precision, refers to the degree of obtaining consistent BMD from the same patient via repeated measurement for a short time, is an essential factor to determine whether the measured value means actual biological change or just error. As the variation of precision occurs even if the measurement is performed by the same person using the same device, re-measurement technique, via which variation range is calculated using statistics, is essentially required to minimize variation. Hospitals that conduct BMD examination should keep their own precision data for precision test.[14,15] 1) Assessment of radio-technologist s precision Assessment of radio-technologist s precision is performed after radio-technologists receive training repeatedly, through which radio-technologists conduct patient posturing, device manipulation and result analysis in a consistent manner and after they take practices on 100 patients. If sufficiently experienced, BMD measurement is performed on minimum 30 patients twice for each patient, or on 15 patients three times for each patient. And then precision error is calculated.[14] Patients should get off the device after first measurement and undergo second measurement after getting on the device again during BMD measurement in order to accurately assess measurement error. Short-term precision is measured within 2 weeks-one month and measurement error is obtained from each examination site including the spine and femur. In addition, if two or more radio-technologists participate in the measurement, data of 4 http://e-jbm.org/ http://dx.doi.org/10.11005/jbm.2014.21.1.2

QC of DXA and Precision Test of Radio-technologists Fig. 2. Precision test method. precision error should be obtained from individual radiotechnologists and the precision error of each hospital is defined as the mean of the data.[14] For the analysis of the obtained data, table of precision error calculation, which is an excel file, can be downloaded from the ISCD homepage[16] and precision error can be easily calculated by inputting the measured data Figure 2. If the calculated value is deviated from the acceptable range below, the measurement should be performed again after re-training. - Lumbar-spine: 1.9% (LSC=5.3%) - Total Hip: 1.8% (LSC=5.0%) - Femur neck: 2.5% (LSC=6.9%) The sum of precision errors of radio-technologists is the precision error of the corresponding hospital. Short-term precision error, coefficient of variation (CV) is obtained by dividing standard deviation by mean and then multiply 100%.[17] Precision error should be set using subjects whose age is similar to that of patients as much as possible. The assessment of precision error is performed by referring to the ISCD recommendation. If precision error is, however, deviated from the acceptable range, it should be measured again after re-training. It is important to understand that precision assessment is a process that is commonly conducted. It should be conducted in accordance with radiological regulations and after receiving patient s consent. It is unnecessary to repeatedly conduct precision assessment unless software and hardware of the system, or radio-technologist s skill is changed.[15,17] However, precision assessment could be independently conducted to determine whether additional training of BMD measurement is required.[18] 2) Main factors that affect precision test Proper positioning of patient s posture and scan analysis are the most important factors in precision assessment. Common factors causing the change of lumbar scanning include improper posture, acquisition of incomplete images of Lumbar1-4, change of spinal level, and improper setting of lumbar interval. In the scanning of proximal femur, factors such as improper rotation, improper positioning of legs, change of target size, and improper targeting of femoral neck cause the change. In addition, it is important to compare the interest site of the spine and femur as having the same size in scanning analysis. To do so, it is important to maintain patient s posture consistently. If rotation and location are inconsistent, improper setting of interest site or boundary commonly occurs. 3) Determination of LSC If the precision of the interest site of a certain bone has been identified, the nest step is to determine whether the degree of BMD change reflects biological significance. This is called LSC. Confidence interval (CI) should be statistically selected in order to determine LSC. Ninety-five percent confidence level is ideally used, but 80% confidence level is also used in some clinical practices. According to the recent ISCD recommendation, 95% confidence level is rec- http://dx.doi.org/10.11005/jbm.2014.21.1.2 http://e-jbm.org/ 5

Ho-Sung Kim, et al. 3% nologists and for the maintenance of expertise. REFERENCES Fig. 3. Calculation of least significant change (LSC) at 95% confidence level (in case of device error: 1%, operator error: 1%). 2.77 Table 1. Percent coefficient of variation (%CV) according to precision by confidence interval Confidence interval% LSC 95 =1.96 (PE) 1.414=2.77 (PE) 1. Equipment & Radio-technologists Error Area 2. LSC Area, 3. Precision Error ommended, in which LSC is calculated by multiplying 2.77 to precision error (Fig. 3, 4).[2,19] Table 1 presents %CV that represents BMD changes according to precision by CI after initial and follow-up examinations. CONCLUSION 2.77 2.33 1.81 LSC 95 Precision Error Area LSC 90 Precision (%CV) 1 1.25 1.5 1.75 2 99 ±3.65 ±4.56 ±5.48 ±6.39 ±7.30 95 ±2.77 ±3.46 ±4.16 ±4.85 ±5.54 90 ±2.33 ±2.91 ±3.50 ±4.08 ±4.66 85 ±2.04 ±2.55 ±3.06 ±3.57 ±4.08 80 ±1.81 ±2.26 ±2.72 ±3.17 ±3.62 %CV, percent coefficient of variation. LSC Area LSC 80 Fig. 4. Variation of precision error according to confidence level. Proper QC should be performed to provide reliability to BMD examination by conducting the examination based on the BMD change and stability of patients. Thus, QC of measuring devices and precision test of radio-technologists are mandatorily required. In addition, the maintenance of dedicated working system and the regular operation of educational programs are also required for the acquisition of examination and analysis skills by radio-tech- 1. Lenchik L, Leib ES, Hamdy RC, et al. Executive summary International Society for Clinical Densitometry position development conference Denver, Colorado July 20-22, 2001. J Clin Densitom 2002;5 Suppl:S1-3. 2. Kim DY. Clinical application of bone mineral density measurement. Korean J Nucl Med 2004;38:275-81. 3. Faulkner KG, McClung MR. Quality control of DXA instruments in multicenter trials. Osteoporos Int 1995;5:218-27. 4. Fuerst TP. Quality control of dual X-ray absorptiometry systems. In: Gould RG, Boone JM, editors. RSNA categorical course in physics. Oak Brook, IL: Radiological Society of North America; 1996. p.67-76. 5. Blunt BA, Jones D, Svensen RK, et al. Good clinical practice and audits for dual X-ray absorptiometry and X-ray imaging laboratories and quality assurance centers involved in clinical drug trials, private practice, and research. J Clin Densitom 1998;1:323-37. 6. Orwoll ES, Oviatt SK. Longitudinal precision of dual-energy x-ray absorptiometry in a multicenter study. The Nafarelin/Bone Study Group. J Bone Miner Res 1991;6:191-7. 7. Montgomery DC. Introduction to statistical quality control. 3rd ed. New York, NY: John Wiley & Sons; 1996. 8. Orwoll ES, Oviatt SK, Biddle JA. Precision of dual-energy x- ray absorptiometry: development of quality control rules and their application in longitudinal studies. J Bone Miner Res 1993;8:693-9. 9. Lu Y, Mathur AK, Blunt BA, et al. Dual X-ray absorptiometry quality control: comparison of visual examination and process-control charts. J Bone Miner Res 1996;11:626-37. 10. Faulkner KG, Gluer CC, Estilo M, et al. Cross-calibration of DXA equipment: upgrading from a Hologic QDR 1000/W to a QDR 2000. Calcif Tissue Int 1993;52:79-84. 11. British Standards Institution. Guide to data analysis and quality control using cusum techniques: BS 5703. London, UK: British Standards Institution; 1980. 12. Pearson D, Cawte SA. Long-term quality control of DXA: a comparison of Shewhart rules and Cusum charts. Osteoporos Int 1997;7:338-43. 13. Garland SW, Lees B, Stevenson JC. DXA longitudinal quality control: a comparison of inbuilt quality assurance, visual inspection, multi-rule Shewhart charts and Cusum 6 http://e-jbm.org/ http://dx.doi.org/10.11005/jbm.2014.21.1.2

QC of DXA and Precision Test of Radio-technologists analysis. Osteoporos Int 1997;7:231-7. 14. Bonnick SL, Johnston CC, Jr., Kleerekoper M, et al. Importance of precision in bone density measurements. J Clin Densitom 2001;4:105-10. 15. The Writing Group for the ISCD Position Development Conference. Technical standardization for dual-energy x- ray absorptiometry. J Clin Densitom 2004;7:27-36. 16. The International Society for Clinical Densitometry. The precision calculating tool. 2013 [cited by 2013 January 1]. Available from: http://www.iscd.org/ 17. Gluer CC, Blake G, Lu Y, et al. Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int 1995;5:262-70. 18. Miller PD, Bonmick SL. Clinical application of bone densitometry. In: Falus MJ, editor. Primer in the metabolic bone diseases and disorders of mineral metabolism. 4th ed. Philadelphia, PA: Lippincott Williams and Wilkins; 1999. p.158. 19. Lenchik L, Kiebzak GM, Blunt BA. What is the role of serial bone mineral density measurements in patient management? J Clin Densitom 2002;5 Suppl:S29-38. http://dx.doi.org/10.11005/jbm.2014.21.1.2 http://e-jbm.org/ 7