CLINICAL SCIENCES. The Effect of Texturing the Intraocular Lens Edge on Postoperative Glare Symptoms

Similar documents
The Effect of Polymethylmethacrylate and AcrySof Intraocular Lenses on the Posterior Capsule in Patients with a Large Capsulorrhexis

Comparison of the Long-term Clinical Results of Hydrophilic and Hydrophobic Acrylic Intraocular Lenses

go the distance NEW AcrySof IQ ReSTOR +2.5 D It s a wide world. Help your patients and everywhere in between.

Sulcoflex Trifocal Supplementary IOL. When expectations and outcomes align. An elegant solution for the correction of presbyopia MADE IN UK

The Efficacy of Questionnaire-based Evaluation in Determining the Incidence of Recent Pseudophakic Dysphotopsia

Electronic poster presentations

Postoperative refraction changes in phacoemulsification cataract surgery with implantation of different types of intraocular lens

November Sponsored by Allergan Surgical

Disorders of the. blood-aqueous barrier after. phacoemulsification in diabetic patients CLINICAL STUDY. Y Liu, L Luo, M He and X Liu

Every week, the John A. Moran Eye Center at the

Disclosure: Samuel Masket MD

International Journal of Current Research in Medical Sciences

Subnormal Vision in Uneventful Cataract Surgery after 6 Weeks Hospital Based Study

Comparison of posterior capsule opacification between fellow eyes with two types of acrylic intraocular lens

Experience with ARRAY multifocal lenses in a Singapore population

Visual Performance Results after Tetraflex Accommodating Intraocular Lens Implantation

COURSE DESCRIPTION BASIC FUNDAMENTALS

Start with ME. LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL

Late-onset secondary pigmentary glaucoma following foldable intraocular lenses implantation in the ciliary sulcus: a long-term follow-up study

White Paper. Etiology of Negative Dysphotopsia. Xiaolin Gu 1, MD, PhD Samuel Masket 2, MD Nicole Fram 2, MD George Pettit 3, MD, PhD

DRAFT CONSENSUS STATEMENT FOR COMMENT October

Introduction. Nikolaos Trakos 1 Elli Ioachim 2 Elena Tsanou 2 Miltiadis Aspiotis 1 Konstantinos Psilas 1 Chris Kalogeropoulos 1 ORIGINAL RESEARCH

Prospective Study of the New Diffractive Bifocal Intraocular Lens

Visual Outcome in Senile Cataract with Multifocal and Unifocal Intraocular Lens

Intrascleral-fixated intraocular lenses for aphakic correction in the absence of capsular support

Capsule fixation device for cataract surgery

Learn Connect Succeed. JCAHPO Regional Meetings 2017

P osterior capsule opacification (PCO) is the most common

FROM CATARACTS TO CLARITY

Management of Radial Tears During Capsulorhexis

New supplementary intraocular lens for refractive enhancement in pseudophakic patients

CONTROLLED PRECISION.

Clinical Evaluation of the BunnyLens IOL

Analysis of surface light scattering of hydrophobic acrylic intraocular lenses implanted 10 years vs. 1 month ago

Posterior capsule opacification following 20- and 23-gauge phacovitrectomy (posterior capsule opacification following phacovitrectomy)

Neodymium: Yag Laser Capsulotomy Rates Following Implantation of Silicone and Acrylic Intraocular Lenses

Visual performance of Acrysof ReSTOR compared with a monofocal intraocular lens following implantation in cataract surgery

You can see vivid colours again after cataract management at Sankar Foundation Eye Hospital

RayOne Trifocal IOL. The preloaded platform that performs again and again MADE IN UK

Refractive Dilemma. Challenging Case

1 Cathedral Eye Clinic, University of Ulster, Academy Street, Belfast, UK.

Relevant and Reliable Systematic Review Mapped to this Section. Relevance of Review to other sections of AAO PPP- 2011

Cataract Surgery Management in Eyes with Extensive Iridoschisis

Cataract Surgery in Patients with Uveitis

OPTOMETRY ORIGINAL PAPER. Post-capsulotomy dysphotopsia in monofocal versus multifocal lenses

Evaluation of long-term visual performance following AcrySof ReSTOR lens implantation

Outcome. Safety; Quality of Life; Pain, Ocular Akinesia; Acceptability of Block to patients; Need for Supplemental Injection

ASCRS 2016 Instructional Course Mastering Femtosecond Laser Assisted Phacoemulsification

Inadvertent trypan blue staining of posterior capsule during cataract surgery associated with "Argentinian flag" event

Premium Implant Options for the Cataract Patient. Justin Schweitzer, OD, FAAO Vance Thompson Vision Sioux Falls, South Dakota

Original article. Amon M Pediatric micro-incision cataract surgery Nepal J Ophthalmol 2011;3(5):3-8

Bringing astigmatism AND presbyopia into focus.

ORIGINAL ARTICLE. Aminollah Nikeghbali MD. Department of Ophthalmology, Rasoul-e-Akram Hospital, Iran University of Medical Sciences, Tehran, Iran

Original Article Capsular tension ring implantation after lens extraction for management of subluxated cataracts

Patient Information Brochure. Cataract

Clinical Efficacy and Complications of Intraocular Lens Exchange for Opacified Intraocular Lenses

Start with ME. LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR EACH PATIENT S LIFESTYLE. TECNIS MULTIFOCAL FAMILY OF 1-PIECE IOLs

Practical Care of the Cataract Patient with Retinal Disease

Clinical Study Is the Memory Effect of the Blind Spot Involved in Negative Dysphotopsia after Cataract Surgery?

Cataract Surgery: Patient Information

Nature and Science 2016;14(9)

WHAT SETS ACTIVEFOCUS DESIGN APART? THE DIFFERENCE IS IN THE DISTANCE.

COMPLICATION RATE IN PRELIMINARY EXPERIENCE IN PHACOEMULSIFICATION CATARACT SURGERY

CATARACT SURGERY IN UVEITIS. Professor Harminder Singh Dua

Cataract surgery in New Zealand

Optometric Postoperative Cataract Surgery Management

Late Intraocular Lens Subluxation in Patients with Uveitis

Innovative Iris Repair

Cataract Surgery by Phacoemulsification Surgical and Visual Outcome with Foldable and Non-Foldable Lenses

TOXIC ANTERIOR SEGMENT SYNDROME (TASS) WITH SEVERE PIGMENT DISPERSION K. Stephen Sudhakar 1, M. Loganathan 2, R. Panduragan 3, C.

Comparing Femtosecond Laser Assisted Cataract Surgery Before and After Phakic Intraocular Lens Removal

Cataracts are a normal feature of aging. About half of adults aged 65 to 74 have cataracts.

Issue 15 The following key clinical peer reviewed journals will be reviewed: MONTHLY RESEARCH UPDATE 151(3) American Journal of Ophthalmology 129(5)

Evaluation of Changes of Macular Thickness in Diabetic Retinopathy after Cataract Surgery

Ophthalmology. Cataract

Quantified Measurement of Intraocular Inflammation after Phacoemulsification Cataract Surgery

Ophthalmology. Cataract Surgery and IOL Implants

A Clinical Evaluation of Multifocal versus Monofocal Intraocular Lenses after Cataract Extraction in a Tertiary Care Hospital in India

Cataract and Refractive Surgery in ; a Survey of KSCRS Members From 1995~2006

In the developed world, cataract surgery has become a routine

Phacoemulsification with hydrodelineation and OVD-assisted hydrodissection in posterior polar cataract

Mr Vaughan Tanner BSc MBBS FRCOphth Consultant Ophthalmic Surgeon

ALTERNATIVES TO PHAKIC IMPLANT SURGERY

Appendix Table 1. Ophthalmic drugs approved by the US Food and Drug Administration,

The Visual Outcome between Foldable and Rigid Intraocular Lens Implantation in Phacoemulsification A Hospital Based Study

Complication and Visual Outcome after Peadiatric Cataract Surgery with or Without Intra Ocular Lens Implantation

MultifocalS: A Reference Guide

Background. Bj orn Johansson 1,2. Acta Ophthalmologica 2017

Anterior Chamber Depth Change Following Cataract Surgery in Pseudoexfoliation Syndrome; a Preliminary Study

Informed Consent For Cataract Surgery. And/Or Implantation of an Intraocular Lens INTRODUCTION

Complex cataract cases Managing catarocks : Better surgery on dense lenses, intumescent cataracts

2/6/2018. Andrew Siedlecki, M.D.

Cataract Surgery. This reference summary will help you understand what cataracts are and how they can be treated surgically.

Incidence of Early Onset Glaucoma after Infant Cataract Extraction With and Without

2/6/2018 RAPID FIRE PANEL: CO-MANAGEMENT OF UNUSUAL SITUATIONS IN CATARACT SURGERY. Andrew Siedlecki, M.D. Richard Orlando, M.D.

Enhancements After Premium IOL Cataract Surgery: Tips, Tricks, and Outcomes

A Study to Compare The Efficacy of 0.5% Timolol Eye Drops Versus 0.2% Brimonidine Eye Drops in Management of Rise In Iop Following Nd-Yag Capsulotomy

Management of Congenital Cataract Surgery. Dr. Vaishali Vasavada, MS. Dr. Abhay R. Vasavada, MS, FRCS (England) Raghudeep Eye Hospital, India

Clinical study of traumatic cataract and its management

Transcription:

CLINICAL SCIENCES The Effect of Texturing the Intraocular Lens Edge on Postoperative Glare Symptoms A Randomized, Prospective, Double-Masked Study Will R. Meacock, FRCOphth; David J. Spalton, FRCS, FRCOphth; Saarg Khan, MB, BS Objective: To determine whether texturing the optic edge of the intraocular lens reduces photic symptoms in patients with square-edged acrylic intraocular lense implants. Methods: Sixty patients underwent routine phacoemulsification performed by a single surgeon and were prospectively randomized to receive either the standard AcrySof MA intraocular lens (Alcon Laboratories, Ft Worth, Tex) (group 1) or an MA intraocular lens with a textured optic edge (group 2). Patients were seen at 1 month by an independent observer who was masked to treatment groups. A questionnaire on photic symptoms was completed, and attempts were made to elicit symptoms under mesopic and photopic conditions. Results: At 1 month postoperatively, (67%) of patients in group 1 noticed symptoms compared with 4 (13%) of patients in group 2 (P.1). The mean duration of symptoms was 3. weeks in group 1 and 1 week in group 2 (P=.1). By provocative testing under photopic conditions, symptoms could be elicited in 26 patients (87%) in group 1 and 6 (%) in group 2. No patients in either group were symptomatic at 3 months postoperatively. Under mesopic conditions, symptoms could be elicited in 27 patients (9%) in group 1 and 11 (37%) in group 2. There was no relationship between the incidence of symptoms and the degree of intraocular lens rhexis contact. Conclusion: Photic phenomena can be significantly reduced by texturing the edge of a square-edged profile AcrySof intraocular lens. Arch Ophthalmol. 2;1:1294-1298 From the Department of Ophthalmology, St Thomas Hospital, London, England. POSTOPERATIVE GLARE phenomena after surgical treatment for cataract were first reported by Masket et al. 1 They found that 4% of patients with ovoid (. mm 6. mm) 1-piece all polymethyl methacrylate (PMMA) intraocular lenses (IOLs) described photic glare symptoms compared with 22% of patients with round (. mm) PMMA IOLs. The optics of these 2 IOLs were studied using scatterometry, which measures the amount of light reflected off materials, and ray tracing analysis, which determines the fate of bundles of light rays passing through the IOL. Results showed that the thickened edge of the ovoid optic was associated with increased light scattering. Posterior capsule opacification has been shown to be significantly reduced by IOLs with a square-edged profile, 2-4 but in some patients, this has been associated with postoperative glare phenomena. Symptoms have been reported with squareedged acrylic (AcrySof; Alcon Laboratories, Ft Worth, Tex) and silicone IOLs.,6 Symptoms are most commonly noticed in the first few weeks after surgical treatment, usually in conditions of mesopic illumination when there is a peripheral light source from overhead or the temporal side. Using ray-tracing techniques, Holladay et al 7 have shown that these symptoms are due to internal light scattering from the IOL edge onto the retina. In most patients, these symptoms resolve over a period of a few weeks, probably due to fibrosis of the rhexis edge around the IOL. In some patients, symptoms persist, possibly due to a larger pupillary diameter in younger patients, a more posteriorly positioned IOL with a greater space between the IOL and the iris in myopic eyes, or from failure of the rhexis edge to adequately opacify with AcrySof IOLs. 8 On occasion, symptoms have been intractable, and it has been necessary to explant the IOL. 9 After it is replaced by a rounded-edged IOL, symptoms have re- 1294

A Textured B Smooth µm Figure 1. Scanning electron photomicrographs of an intraocular lens with a textured edge (A) and a lens of the same design but with a smooth edge (B). solved, providing circumstantial evidence that a squareedged optic is the cause of the symptoms. Recently, a variation of the AcrySof IOL has been developed in which the optic has a textured edge. On examination, this has a frosted appearance. The aim of this prospective, randomized study was to determine whether texturing the square edge of an AcrySof IOL reduces postoperative photic glare phenomena. METHODS Following approval from the hospital ethics committee and informed consent, patients were recruited from October 1999 through March. All patients were seen preoperatively by the same clinician (W.R.M.). Inclusion criteria were the presence of senile cataract in an otherwise normal eye in patients older than 6 years. Exclusion criteria were a history of any previous ocular disease, intraocular surgery, laser treatment, diabetes requiring medical control, glaucoma, previous uveitis, or posterior segment abnormalities that would preclude a postoperative visual acuity of /4 or better. All patients underwent phacoemulsification with continuous curvilinear capsulorrhexis, performed by a single surgeon (D.J.S.) using peribulbar anesthesia. The surgical technique and medication were standardized. A temporal clear corneal section was made, and the anterior chamber was re-formed with Healon (Pharmacia, Uppsala, Sweden). A continuous curvilinear capsulorrhexis of approximately mm was performed so that the anterior capsule flap lay on the implant optic. The nucleus was removed by a phaco-chop technique, and soft lens material was removed by irrigation-aspiration with balanced salt solution (Alcon Laboratories) containing vancomycin hydrochloride (concentration, mg/l) and epinephrine bitartrate (1:; 1 ml/l). No attempt was made to remove lens epithelial cells by polishing the anterior capsule. The bag was re-formed with Healon and the section enlarged to 3. mm. Preoperatively, patients were randomized to receive either a.-mm, 3-piece AcrySof IOL manufactured via the old process (MABM) (group 1) or a.-mm, 3-piece AcrySof IOL manufactured with the new process, resulting in a textured optic edge (designated for this study as MABT) (group 2) (Figure 1). The material and geometry of both IOLs were otherwise identical. The IOLs were inserted in the bag, and Healon was removed by irrigation aspiration with balanced salt solution. If there were any surgical complications, such as capsulorrhexis rim tear, zonular dehiscence, posterior capsule rupture, or vitreous loss, the patient was excluded from the study and replaced. Postoperatively, all patients used Maxitrol drops (neomycin sulfate, polymyxin B sulfate, and dexamethasone.1%; Alcon Laboratories) 4 times a day for 1 month. No nonsteroidal anti-inflammatory preparation was used preoperatively, perioperatively, or postoperatively. Postoperatively, all IOLs were confirmed to have in-the-bag placement. At 1 month postoperatively, all patients were assessed by an ophthalmologist (W.R.M. or S.K.) who was masked to which group the patients had been assigned. Pelli-Robson contrast sensitivity and ETDRS (Early Treatment of Diabetic Retinopathy Study) visual acuity were measured. All patients completed a questionnaire to determine the following: (1) presence of symptoms, (2) their postoperative duration, (3) the time of day they were most noticeable, (4) the relationship to the direction of light that induced them, () their impact on daily activities, and (6) the pattern of glare experienced, with reference to a sheet illustrating the various forms of glare phenomena (Figure 2). Patients were then examined under both photopic and mesopic conditions to provoke the symptoms. With the patient looking in the primary position at all times, a penlight was moved in an arc toward the peripheral visual field. The visual field was divided into 8 segments, and patients were asked to describe the presence and type of glare with reference to the pattern sheet. If symptoms were elicited in any sector, the test was regarded as positive. Following pupillary dilatation, slitlamp biomicroscopy was performed to determine the degree of IOL-rhexis contact and the presence of vertical striae through the center of the posterior capsule. Pupil diameter was not measured in the study. Patients with symptoms still present at 1 month postoperatively were reassessed at 3 months postoperatively using the same protocol. The Fisher exact test was used to analyze the proportion of patients with symptoms, the impact of symptoms on the patients, and the type of photic phenomena under mesopic and photopic conditions. An unpaired t test was used to compare the duration of symptoms and the means of visual acuity and contrast sensitivity data. RESULTS Of 6 patients, were allocated to each group. All patients in group 1 (untextured IOL) were available for postoperative assessment at 1 month and 3 months, when necessary. In group 2, (textured-edged IOL) 1 patient failed to attend the 1-month assessment because of illness. The mean age was 76 years in group 1 (range, 6-83 years) and 74 years in group 2 (range, 6-86 years). There was no difference between the 2 groups in age and sex distribution. At 1 month postoperatively, the mean ETDRS visual acuity was.2 for group 1 and. for group 2 (P=.7). Mean Pelli-Robson contrast sensitivity was 1.48 and 1.1 for groups 1 and 2, respectively (P=.). 129

1 2 3 4 6 7 8 9 11 12 13 14 Figure 2. Patterns of photic glare. 2 2 Day Night Night + Day None Aware Impact of Symptoms Nuisance Figure 3. The time of day when photic phenomena were noticed by symptomatic patients with textured-edged or smooth-edged intraocular lenses (IOLs). At 1 month postoperatively, patients (67%) in group 1 experienced photic glare symptoms compared with 4 (13%) in group 2 (P.1; relative risk,.; 9% confidence interval, 1.9-12.9). The mean duration of symptoms was 3. weeks for patients in group 1 and 1 week in group 2 (P=.1). No patients in either group had symptoms at 3 months postoperatively. The symptomatic patients in both groups noticed the photic phenomena more commonly from peripherally directed light, particularly from overhead light. In group 1, patients tended to experience symptoms more commonly at night (Figure 3). In group 1, the symptoms had no impact in 6 (%) of patients, 13 patients (6%) were aware of them, and in 1 patient (%) they were a nuisance. In group Figure 4. The impact of symptoms on patients with photic phenomena. IOL indicates intraocular lens. 2, symptoms had no impact in 3 (7%) of 4 patients, and 1 patient (2%) was aware of them (Figure 4). With provocative testing under photopic conditions it was possible to elicit symptoms in 26 patients (87%) in group 1 and 6 patients (%) in group 2 (P.1) (Figure A). Under mesopic conditions, photic phenomena were elicited in 27 patients (9%) in group 1 and in 11 patients (37%) in group 2 (Figure B). Flare (Figure 2, patterns 7-11) was the most commonly elicited glare pattern in group 1 under both light conditions. In group 2, there was no pattern bias. The relationship between the number of quadrants of IOL rhexis contact and the presence of symptoms is given in the Table. At 1 month postoperatively, the in- 1296

cidence of symptoms did not increase in those patients with more optic edge exposure (P =.2). There was no relationship between the presence of vertical striae in the posterior capsule and the incidence of photic symptoms. 2 A Photopic Conditions COMMENT The excellent visual outcome from modern phacoemulsification can be impaired by the unwanted images produced by photic glare phenomena, and because of the increasingly high expectations of patients, the phenomena need to be eliminated. Several features related to the design of IOLs have been associated with photic glare phenomena. Apple et al reported a case of a pilot who had monocular glare symptoms at night after the implantation of a 6-mm round PMMA lens with 4 positioning holes around the optic edge. Symptoms were abolished by an IOL exchange. In a separate report, Landry 11 indicated that IOL positioning holes were responsible for the unwanted photic images, and such observations have led to their subsequent removal from the majority of implants in the manufacturing process. In the late 198s and early 199s, IOLs with. mm 6. mm ovoid optics, constructed from round PMMA IOLs with the opposing sides truncated, were popular, but if the implant was positioned horizontally, patients experienced glare symptoms from superiorly located light sources, such as room lights and the sun. The reports of glare phenomena with ovoid IOLs has since made them unpopular. 12,13 Mamalis and Spencer 14 recently reported the results of a physician survey concerning the explantation of foldable IOLs. This survey reported that visual aberrations or optical phenomena were reported by 24% of patients with 3-piece monofocal silicone lenses, 27% of patients with 1-piece plate haptic silicone lenses, and 28% of patients with 3-piece acrylic lenses. A square edge is, therefore, not the only requirement for an IOL lens to produce photic phenomena. Tester et al conducted a telephone questionnaire to investigate the incidence of unwanted light images with different IOLs. They surveyed patients with 6 IOL types: AcrySof.-mm (MA) and 6.-mm (MA6) IOLs, the Allergan SI-4 (Allergan, Inc, Irvine, Calif), the Staar/Chiron plate haptic lens (Staar Surgical Co, Monrovia, Calif, and Chiron Corporation, Emeryville, Calif), and 2 PMMA IOLs (. mm and 6. mm) and compared them with controls with presbyopia. Patients were contacted between 6 and 12 months after surgical treatment for cataract to identify whether they were experiencing glare phenomena. Those with symptoms were asked to describe the pattern, the light conditions under which they occurred, and the effect they had on vision. Interestingly, the rates of glare and sensitivity symptoms in patients with AcrySof or SI-4 IOLs were similar to those of the controls. However, patients with IOLs had more unwanted images than did controls. Patients with AcrySof and SI-4 IOLs reported the highest incidence of unwanted images (MA, %; MA6, 3%; SI-4, 24%; compared with the control group, 4%). The SI-4 lenses have, in some cases, been successful in reducing photic symptoms after IOL exchange of 2 B Mesopic Conditions Flare Halos Pattern Streaks Figure. Patterns of elicited photic glare phenomena under photopic (A) and mesopic (B) conditions. IOL indicates intraocular lens. Photic Glare Symptoms and IOL Rhexis Contact in Patients With Smooth-Edged or s No. of Quadrants of IOL Rhexis Contact Smooth-Edged IOL (n = ) Textured-Edged IOL (n = ) 1 6 () 12 () 2 7 (8) 6 () 3 4 (7) 2 () 4 13 (67) (33) *Data are given as the number of patients (percentage of patients who were symptomatic). IOL indicates intraocular lens. an AcrySof lens. The authors could not explain the incidence of photic symptoms among patients with SI-4 IOLs in this survey. Similar rates of symptoms in the AcrySof and PMMA.-mm and 6.-mm optic IOL groups suggest that optic diameter is not particularly important in causing symptoms. Overall, this paper shows that lightinduced symptoms are common in both patients with presbyopia and patients with pseudophakia. It is difficult to identify their precise cause, and it is uncommon for them to be bothersome. Our study investigated photic phenomena in 2 square-edged.-mm 3-piece AcrySof IOLs, one with a smooth optic edge and the second with a textured edge. We have shown that patients who received the textured implant had a statistically significant reduction in the in- 1297

cidence, duration, and severity of edge glare symptoms. From Holladay s study using ray tracing in round-edged and square-edged IOLs, 7 the most likely explanation for this result is that light reflected off the internal IOL edge surface is more scattered by the textured-edged lens, and a focused internal image is not created. It was important to take into account the effect of IOL-rhexis contact on the incidence of symptoms in both groups of patients. The capsule on the anterior surface of the IOL optic undergoes fibrotic changes (anterior capsule opacification [ACO]) in the early postoperative period. This decreases the transparency of the anterior capsule and will increase the dispersion of light entering the eye, with the potential to reduce glare. In a histological study, the rate of ACO was found to be dependent on the type of implant. 8 Silicone plate haptic IOLs produced high ACO scores, whereas 3-piece acrylic optic/ PMMA haptic IOLs produced little ACO. Increased ACO might be beneficial in preventing photic symptoms, although it could lead to IOL decentration if the forces on the IOL are asymmetrical. The low rate of ACO with acrylic IOLs has the potential therefore to make these IOLs more susceptible to photic glare phenomena. In our study, the Table shows that there was no relationship between the number of quadrants covered by the rhexis and the presence of symptoms. This is not surprising because by 1 month postoperatively, when the patients were first questioned, ACO would have had little time to develop. No statistical differences in visual acuity or contrast sensitivity were detected between the 2 lens groups. Despite the high incidence of elicited symptoms with the smooth square-edged IOLs, only 1 patient reported any difficulties in daily visual tasks due to the presence of photic images, and none of the patients reported symptoms beyond 3 months postoperatively. This is not surprising given the relatively small group of patients studied and the fact that photic symptoms tend to be reported by younger patients with a high degree of critical awareness. This study has shown, however, that texturing the optic edge surface is very successful in reducing postoperative photic symptoms and has the potential to improve the quality of vision for patients. Submitted for publication August, 1; final revision received June 11, 2; accepted June 14, 2. This study was presented at the annual meeting of the European Society of Cataract and Refractive Surgery, Brussels, Belgium, September 6,. Corresponding author: David J. Spalton, FRCS, FRCOphth, Department of Ophthalmology, St Thomas Hospital, Lambeth Palace Road, London SE1 7EH, England. REFERENCES 1. Masket S, Geraghty E, Crandall AS. Undesired light images associated with ovoid intraocular lenses. J Cataract Refract Surg. 1993;19:69-694. 2. Nishi O, Nishi K, Sakanishi K. Inhibition of migrating lens epithelial cells at the capsular bend created by the rectangular optic edge of a posterior chamber intraocular lens. Ophthalmic Surg Lasers. 1998;29:87-94. 3. Nishi O, Nishi K, Wickstro K. Preventing lens epithelial cell migration using intraocular lenses with sharp rectangular edges. J Cataract Refract Surg. ; 26:43-49. 4. Kruger A, Schauersberger J, Abela C, Amon M. Two-year results: sharp vs rounded optic edges on silicone lenses. J Cataract Refract Surg. ;26:66-7.. Masket S, ed. Consultation section: cataract surgical problem. J Cataract Refract Surg. 1998;24:4-61. 6. Masket S, ed. Consultation section: cataract surgical problem. J Cataract Refract Surg. 1997;23:979-984. 7. Holladay JT, Ang A, Portney V. Analysis of edge glare phenomena in intraocular lens edge designs. J Cataract Refract Surg. 1999;2:748-72. 8. Werner L, Pandey S, Escobar-Gomez M, Apple DJ. Anterior capsule opacification: a histopathological study comparing different IOL styles. Ophthalmology. ;7:463-471. 9. Farbowitz A, Zabriskie N, Crandall AS, Olson R. Visual complaints associated with the AcrySof acrylic intraocular lens. J Cataract Refract Surg. ;26:1339-1346.. Apple DJ, Lichtenstein SB, Heerlein K, et al. Visual aberrations caused by optic components of posterior chamber intraocular lenses. J Cataract Refract Surg. 1987;13:431-43. 11. Landry RA. Unwanted optical effects caused by intraocular lens positioning holes. J Cataract Refract Surg. 1987;13:421-423. 12. Anderson CJ, Sturm RJ, Shapiro MB, Ballew C. Visual disturbances associated with oval-optic poly(methylmethacrylate) and round-optic silicone intraocular lenses. J Cataract Refract Surg. 1994;:29-298. 13. Waller S, Steinert RF. Symptomatic intraocular reflections from oval intraocular lens implants. Am J Ophthalmol. 1993;116:374-376. 14. Mamalis N, Spencer TS. Complications of foldable intraocular lenses requiring explantation or secondary intervention: survey update. J Cataract Refract Surg. 1;27:13-1317.. Tester R, Pace N, Samore M, Olson R. Dysphotopsia in phakic and pseudophakic patients: incidence and relation to intraocular lens type. J Cataract Refract Surg. ;26:817-824. 1298