Patient ventilator asynchrony and sleep disruption during noninvasive

Similar documents
A literature review was conducted using PubMed database. Search terms included patient

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients

Recognizing and Correcting Patient-Ventilator Dysynchrony

The Impact of Patient-Ventilator. Karen J Bosma, MD, FRCPC Critical Care Medicine and Respirology

Reasons Providers Use Bilevel

Noninvasive Mechanical Ventilation in Children ศ.พญ.อร ณวรรณ พฤทธ พ นธ หน วยโรคระบบหายใจเด ก ภาคว ชาก มารเวชศาสตร คณะแพทยศาสตร โรงพยาบาลรามาธ บด

Index. sleep.theclinics.com. Note: Page numbers of article titles are in boldface type.

Monitoring: gas exchange, poly(somno)graphy or device in-built software?

Ventilator Dyssynchrony - Recognition, implications, and management

Ron Hosp, MS-HSA, RRT Regional Respiratory Specialist. This program has been approved for 1 hour of continuing education credit.

Case 1. Level of difficulty: 2/5

Motor Neurone Disease NICE to manage Management of ineffective cough. Alex Long Specialist NIV/Respiratory physiotherapist June 2016

Potential Conflicts of Interest

Policy Specific Section: October 1, 2010 January 21, 2013

BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT

Patient Ventilator Interactions. Patient-Ventilator Interactions. Assisted vs Controlled MV. Ventilatory Muscle Fatigue Recovery

Ventilator Waveforms: Interpretation

Effect of expiratory positive airway pressure on tidal volume during non-invasive ventilation

Causes and Consequences of Respiratory Centre Depression and Hypoventilation

Chronic NIV in heart failure patients: ASV, NIV and CPAP

(To be filled by the treating physician)

Interdisciplinary Care of the Patient with Amyotrophic Lateral Sclerosis Respiratory Therapy Care

New Modes to Enhance Synchrony & Dietrich Henzler MD, PhD, FRCPC Division of Critical Care

An ideal ventilator for neuromuscular patients. F. Lofaso, Raymond Poincaré Hospital Inserm U 1179

How To Set Up A Ven.lator: Standard Versus High Pressure

RESPIRATORY PROBLEMS IN MND RICHARD HARRISON MND RESPIRATORY ASSESSMENT SERVICE LUNG HEALTH UNIVERSITY HOSPITAL OF NORTH TEES

Noninvasive ventilation: Selection of patient, interfaces, initiation and weaning

OSA and COPD: What happens when the two OVERLAP?

Approach to type 2 Respiratory Failure

Prepared by : Bayan Kaddourah RN,MHM. GICU Clinical Instructor

Home Mechanical Ventilation

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation

Non-invasive Ventilation protocol For COPD

Average volume-assured pressure support

INTRODUCTION Non-invasive ventilation (NIV) is recognised as an

INTRODUCTION Non-invasive ventilation (NIV) is recognised as an

Web Appendix 1: Literature search strategy. BTS Acute Hypercapnic Respiratory Failure (AHRF) write-up. Sources to be searched for the guidelines;

PEDIATRIC PAP TITRATION PROTOCOL

Ventilator curves. Fellowonderwijs 2 feb 2012

Rebecca Mason. Respiratory Consultant RUH Bath

Patient±ventilator interaction

WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35

Asynchrony and Dyspnea

Handling Common Problems & Pitfalls During. Oxygen desaturation in patients receiving mechanical ventilation ACUTE SEVERE RESPIRATORY FAILURE

Access to the published version may require journal subscription. Published with permission from: Blackwell Synergy

Mechanical Ventilation of the Patient with Neuromuscular Disease

THE EFFECTS OF MEDROXYPROGESTERONE ACETATE AND ACETAZOLAMIDE ON THE NOCTURNAL OXYGEN SATURATION IN COPD PATIENTS

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE SCOPE

Monitoring Respiratory Drive and Respiratory Muscle Unloading during Mechanical Ventilation

Noninvasive ventilation: modes of ventilation

Neuromuscular diseases (NMDs) include both hereditary and acquired diseases of the peripheral neuromuscular system. They are diseases of the

11/20/2015. Beyond CPAP. No relevant financial conflicts of interest. Kristie R Ross, M.D. November 12, Describe advanced ventilation options

GE Healthcare. Non Invasive Ventilation (NIV) For the Engström Ventilator. Relief, Relax, Recovery

Sleep Disordered Breathing: Beware Snoring! Dr T A McDonagh Consultant Cardiologist Royal Brompton Hospital London. UK

Measure #6: ALS Noninvasive Ventilation Treatment for Respiratory Insufficiency Discussed Amyotrophic Lateral Sclerosis

What is SDB? Obstructive sleep apnea-hypopnea syndrome (OSAHS)

Non-Invasive Ventilation

1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation.

Lumis Tx: the all-in-one hospital ventilation solution

Do Not Cite. For Public Comment Period DRAFT MEASURE #3: Evaluation of Pulmonary Status Ordered MUSCULAR DYSTROPHY

Identification and Treatment of the Patient with Sleep Related Hypoventilation

CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEFINITION

Competency Title: Continuous Positive Airway Pressure

Ventilator modes and settings during non-invasive ventilation: effects on respiratory events and implications for their identification

Obstructive Sleep Apnoea and COPD: Can we use auto-titrating ventilator devices?

Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo

Keywords: Non-invasive mechanical ventilation, Respiratory Failure, Respiratory muscles, Hypercapnia, Breathing pattern.

Mechanical Ventilation ศ.พ.ญ.ส ณ ร ตน คงเสร พงศ ภาคว ชาว ส ญญ ว ทยา คณะแพทยศาสตร ศ ร ราชพยาบาล

Name of Policy: Noninvasive Positive Pressure Ventilation

Basics of NIV. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC. Consultant, Critical Care Medicine Medanta, The Medicity

Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment

Objectives. Apnea Definition and Pitfalls. Pathophysiology of Apnea. Apnea of Prematurity and hypoxemia episodes 5/18/2015

Mechanical Ventilation 1. Shari McKeown, RRT Respiratory Services - VGH

Objectives. Health care significance of ARF 9/10/15 TREATMENT OF ACUTE RESPIRATORY FAILURE OF VARIABLE CAUSES: INVASIVE VS. NON- INVASIVE VENTILATION

Chronic hypoventilation and its management

Faculty Disclosure. Off-Label Product Use

NON INVASIVE LIFE SAVERS. Non Invasive Ventilation (NIV)

Non-Invasive Ventilation of the Restricted Thorax: Effects of Ventilator Modality on Quality of Life. The North Study

SHORT TERM EFFECTS OF NIV ON SLEEP IN END STAGE CF. Moran Lavie MD The national center for Cystic Fibrosis Sheba medical center

Nasal versus full face mask for noninvasive ventilation in chronic respiratory failure.

Multiple organ. support

Home Mechanical Ventilation:

Evaluation of Effect of Breathe Ventilation System on Work of Breathing in COPD patients. Matthew Cohn, M.D.

Monitor the patients disease pathology and response to therapy Estimate respiratory mechanics

Non-invasive ventilation (NIV)

Teacher : Dorota Marczuk Krynicka, MD., PhD. Coll. Anatomicum, Święcicki Street no. 6, Dept. of Physiology

Patient-Ventilator Interaction. David J Pierson MD FAARC

Sleep and non-invasive ventilation in patients with chronic respiratory insufficiency

Patients commonly arouse from sleep and experience awake states.

PEDIATRIC OBSTRUCTIVE SLEEP APNEA (OSA)

Respiratory Rate. Disclosure: Research grants and honoraria from Masimo Corp

Respiratory Failure how the respiratory physicians deal with airway emergencies

APRV Ventilation Mode

Combined effects of leaks, respiratory system properties and upper airway patency on the performance of home ventilators: a bench study

Breathing problems: and how to get on top of them

more than 50% of adults weigh more than 20% above optimum

Proportional Assist Ventilation (PAV) (NAVA) Younes ARRD 1992;145:114. Ventilator output :Triggering, Cycling Control of flow, rise time and pressure

Improving Care & Outcomes

MND Study Day. Martin Latham CNS Leeds Sleep Service

Transcription:

Review Article Patient ventilator asynchrony and sleep disruption during noninvasive ventilation Michelle Ramsay Lane Fox Unit, St Thomas Hospital, London, UK Correspondence to: Dr. Michelle Ramsay, MRCP (Resp), BSc, Consultant Physician. Lane Fox Unit, St Thomas Hospital, London, UK. Email: Michelle.Ramsay@gstt.nhs.uk. Abstract: Patient-ventilator asynchrony (PVA) are a common phenomenon affecting all patients receiving long-term domiciliary ventilation. The interruption of synchrony between the patient and the ventilator has been reported to be a possible cause of nocturnal sleep disruption leaking to arousals, awakenings and reduced periods of stage 3 and REM sleep overnight. The cause of PVA is multi-factorial driven frequently by leak at the interface, changing respiratory breathing patterns and changes in sleep stages. It currently remains unclear as to whether the PVA is the cause of sleep fragmentation or if PVA is purely a marker of unsuccessful ventilatory control, patient discomfort and underlying sleep disruption. Keywords: Patient-ventilator asynchrony (PVA); sleep disruption; non-invasive ventilation (NIV); leak Submitted Oct 11, 217. Accepted for publication Nov 2, 217. doi: 1.2137/jtd.217.11.31 View this article at: http://dx.doi.org/1.2137/jtd.217.11.31 Introduction Sleep disordered breathing is common in patients with chronic respiratory failure. Respiratory failure results from an imbalance between the loads placed upon the respiratory system and its capacity to compensate for that load. To balance the load on the respiratory system in patients with respiratory disease, neural respiratory drive is elevated to activate and recruit muscle to increase the operational capacity from the respiratory muscle pump. Overnight, the natural fall in neural respiratory drive during sleep, particularly in rapid eye movement (REM) sleep exacerbates the respiratory muscle load-capacity imbalance. This occurs in combination with muscle hypotonia of the upper airway musculature and a blunted hypercapnic ventilatory response predisposing to alveolar hypoventilation and respiratory failure (1,2). Sleep disordered breathing frequently develops as an early sign of underlying respiratory failure leading to sleep fragmentation as the patient awakes to augment neural respiratory drive and daytime somnolence (3,4). Nocturnal non-invasive ventilation (NIV) reduces the work of breathing, alleviates upper airways obstruction and assists the respiratory muscle pump to improve alveolar hypoventilation and gas exchange. As such, it is the treatment of choice in managing patients with chronic respiratory failure. Correcting sleep disordered breathing with NIV may therefore be expected to improve both nocturnal gas exchange and sleep quality by ameliorating the respiratory disturbance index. However, significant sleep disruption has been reported during the use of both invasive mechanical and NIV which may adversely impact on patients health, wellbeing and adherence to therapy (5-7). Patient-ventilator asynchrony (PVA) describes the poor interaction between the patient and the ventilator and is the consequence of the respiratory muscle activity of the patient being opposed to the action of the ventilator. A mismatch between the breaths demanded by the patient and those delivered by the ventilator has been proposed as a mechanism of leading to increased sleep disruption in patients requiring nocturnal NIV. PVA during NIV PVA is observed in all patients receiving NIV with a majority of patients experiencing >1% breaths affected by this phenomenon (8). A recent study in patients J Thorac Dis 218;1(Suppl 1):S8-S85

Journal of Thoracic Disease, Vol 1, Suppl 1 January 218 S81 Cycling from inspiration to expiration Premature expiration cycling/delayed expiratory cycling Trigger Time Ineffective triggering Autotriggering/ double triggering/ multiple triggering/ auto-cycling Figure 1 Demonstrating the different types of patient-ventilator asynchrony that can interrupt the harmony between the patient demands and ventilator delivered breaths. with amyotrophic lateral sclerosis receiving long term domiciliary NIV suggested that PVA affected up to 5% of the total sleep time (9). PVA can be characterised into those that interrupt the triggering of the ventilator triggering asynchrony and those that affect the cycling of the ventilator from inspiration to exhalation cycling asynchrony. This can be observed most clearly when examining the ventilator delivered breath waveform in more detail (Figure 1). Triggering asynchrony Triggering asynchrony are the most common type of asynchrony observed affecting almost a quarter of all breaths with ineffective triggering accounting for the majority (16% of breaths) (8). Ineffective triggering occurs when the patient attempts to exhibit inspiratory effort demanding a breath without a corresponding breath being delivered by the ventilator (Figure 2). This has the opposite desired effect for a patient in respiratory failure as it increases the work of breathing performed by the patient without any support or unloading of the respiratory muscles by the ventilator. By contrast, autotriggering occurs when an inappropriate breath is delivered by the ventilator that has not been requested by the patient. This phenomenon can result in the delivery of a single inappropriate breath autotriggering or the patient may request a single breath and the ventilator will deliver two breaths in quick succession double triggering or more than two breaths in quick succession multiple triggering. Cycling asynchrony Cycling asynchrony have been reported as less common than triggering asynchrony affecting approximately 5% of breaths in patients receiving long-term home mechanical ventilation (HMV) (8). However, in a study of patients requiring NIV for acute respiratory failure, cycling asynchrony were more prominent with delayed expiratory cycling the most frequent asynchrony detected, affecting 23% of patients studied (1). Cycling asynchrony occur when there is a mismatch between the inspiratory time that the patient demands termed neural inspiratory time and the inspiratory time provided by the ventilator. Either the patient may wish to breathe in for an extended period of time and the ventilator inappropriately cycles into expiration early premature expiratory cycling or the patient may wish to exhale and the ventilator inappropriately continues to deliver an inspiratory breath delayed expiratory cycling (Figure 3). Autocycling occurs when the ventilator is in free fall and continues to cycle between inspiration and expiration J Thorac Dis 218;1(Suppl 1):S8-S85

S82 Ramsay. PVA and sleep disruption during NIV A 2 1.8 1. 1.2.5 1. EMG para 1.5 2. 2.5.4.2..2 58:41 58:42 58:43 58:44 58:45 58:46 58:47 58:48 Ineffective effort B 2 1..5.2 EMG para. Rectified RMS signal..5 1.4.2 6 4 2 Ventilator delivered breath without patient effort (semg para ). Shorter Ti than set, indicating inappropriate PSV breath. 18:27 1:8:28 1:8:29 1:8:3 1:8:31 1:8:32 Figure 2 An example of (A) ineffective triggering and (B) autotriggering asynchrony., chest wall respiratory inductance plethysmography;, abdominal wall respiratory inductance plethysmography;, sum of both the chest wall and abdominal inductance plethysmography bands; EMGpara, surface second intercostal parasternal electromyography signal reflecting patient neural inspiratory drive; Rectified RMS Signal, rectified root mean square of the parasternal electromyography signal to better characterise onset and termination of neural inspiratory time; Ti, inspiratory time; PSV, pressure support ventilation delivered breath. delivering multiple breaths that are completely out of synchrony with any inspiratory effort made by the patient. PVA and sleep disruption In a study assessing patient experiences of mechanical ventilation, 35% retrospectively described significant sleep disruption (11). Thirty percent reported symptoms of agony/ panic during mechanical ventilation that was associated with difficulties in synchronising the patients breathing to the ventilator (11). This has led to the suggestion that patient ventilator asynchrony may be directly related with arousals and awakenings resulting in persistent sleep disruption. Previous studies that examined patients with chronic respiratory failure receiving HMV, demonstrated a relationship between increased PVA and poor sleep quality, J Thorac Dis 218;1(Suppl 1):S8-S85

Journal of Thoracic Disease, Vol 1, Suppl 1 January 218 S83 A 2 1.4.2 Ventilator cycling to expiration.8.6 1.2 1..8 Persistent chest excursion.2 EMG para..2 Rectified RMS signal 4 2 4:46:27 4:46:27.5 4:46:28 4:46:28.5 4:46:29 4:46:29.5 4:46:3 4:46:3.5 Persistent inspiratory parasternal muscle contraction (signal) Expiration (noise) B 2 1 5.6.4.2 1..5 1. EMG para. 2 Rectified 1 RMS signal 5 4:39:4 4:39:5 4:39:6 4:39:7 4:39:8 4:39 A persistent ventilator delivered breath continuing post the cessation of patient effort. Figure 3 An example of (A) premature expiratory cycling and (B) delayed expiratory cycling asynchrony., chest wall respiratory inductance plethysmography;, abdominal wall respiratory inductance plethysmography;, sum of both the chest wall and abdominal inductance plethysmography bands; EMGpara, surface second intercostal parasternal electromyography signal reflecting patient neural inspiratory drive; Rectified RMS Signal, rectified root mean square of the parasternal electromyography signal to better characterise onset and termination of neural inspiratory time. defined as greater sleep fragmentation with reduced stage 3 and REM sleep (5,12,13). Research by Fanfulla et al., also demonstrated an association between the number of ineffective efforts observed and the time spent overnight in REM sleep in patients with neuromuscular disease receiving NIV, suggesting that ineffective effort asynchrony may inhibit REM sleep. Guo et al., identified similar findings in 2 patients with obesity hypoventilation syndrome receiving nocturnal NIV, where PVA was associated with a reduction in stage 3 and REM sleep (12). PVA was also associated with arousals predominantly in lighter stage 1 & 2 sleep which may be related to enhanced neural respiratory drive to the respiratory muscles in light sleep facilitating opposition to the ventilator delivered breaths (12). Over time with increased experience of NIV there have been improvements in ventilatory strategy, most notably a move to higher levels of pressure support and rates of ventilator delivered breaths to unload respiratory muscles and control carbon dioxide levels. Newer modes of ventilation have also been adopted in an attempt to enhance ventilation, patient comfort and patient ventilator synchronisation. Bosma et al. compared proportional assist ventilation (which delivers a partial ventilatory support controlled by the patient s demand augmenting the flow and volume delivered to the patient) to pressure support ventilation in patients weaning from mechanical ventilation. J Thorac Dis 218;1(Suppl 1):S8-S85

S84 Ramsay. PVA and sleep disruption during NIV They reported that proportional assist ventilation improved sleep disruption as a consequence of improved patient ventilator interaction and a reduction in PVA (6). Despite improvements in the delivery of ventilation, the levels of PVA in patients receiving domiciliary NIV remain high (8,9). The clinical benefits of nocturnal NIV to patients in chronic respiratory failure remains clear and from experience, the majority of patients adhere to treatment without intolerable sleep disruption. Furthermore, recent studies have been challenging the earlier literature suggesting that a direct association between PVA and sleep disruption is far from clear (9). Changes between sleep stages overnight in patients with sleep disruption may impact on levels of neural respiratory drive and therefore disturb the interaction of the patient with the ventilator. Deleterious changes in gas exchange overnight may also lead to hypoxia induced arousals from events such as residual obstructive apnoea or hypercapnic blunting of neural respiratory drive. Also, obstructive apnoea may directly lead to ineffective effort asynchrony interfering with the ventilator detection of patient inspiratory effort due to the cessation of flow. Therefore, whether PVA occurs as a marker of underlying poor sleep disruption or is the driver of the sleep disruption is yet to be fully characterised. Conclusions In summary, PVA are frequent during domiciliary NIV. These asynchronies may impact on ventilator efficiency, respiratory muscle loading, patient comfort and sleep disruption, although the overall importance of PVA in NIV is becoming less certain. The current literature is conflicted in the evidence supporting PVA as a direct cause of significant sleep disruption. Similarly, recent studies have observed considerable PVA without any significant impact on gas exchange, challenging earlier findings (8,9,12). The relationship between sleep disruption resulting from PVA, leaks, gas exchange, and clinical outcomes such as adherence to therapy, health-related quality of life measures and survival require further evaluation. Acknowledgements The author acknowledges the lane fox respiratory research unit and King's College London University. Footnote Conflicts of Interest: The author has no conflicts of interest to declare. PVA and leak Leak is a further complexity that accompanies the noninvasive interface used to deliver ventilation and impacts on the interaction between the patient and ventilator. Leaks are intentional occurring via the exhalation valve to facilitate carbon dioxide clearance and unintentional leak occurring elsewhere in the ventilator circuit. Unintentional leak interferes with the ventilator detection of patient ventilatory efforts leading to both triggering asynchrony and delayed expiratory cycling asynchrony (13). These asynchrony were correlated with increased arousal and awakenings (13). However, more recent studies focusing on patients with amyotrophic lateral sclerosis in whom unintentional leak is a common phenomenon due to facial muscle weakness did not demonstrate a significant impact on sleep architecture (9). Unintentional leak also increases airflow across the skin and is associated with a sudden noise, which may itself have a deleterious impact on sleep quality. Identifying if PVA associated with leak is the driver of sleep disruption or just a consequence of external factors is challenging. References 1. Becker HF, Piper AJ, Flynn WE, et al. Breathing during sleep in patients with nocturnal desaturation. Am J Respir Crit Care Med 1999;9:112-8. 2. Elliott MW. Non-invasive ventilation during sleep: time to define new tools in the systematic evaluation of the technique. Thorax 211;66:82-4. 3. Ragette R, Mellies U, Schwake C, et al. Patterns and predictors of sleep disordered breathing in primary myopathies. Thorax 22;57:724-8. 4. Punjabi NM, O'Hearn DJ, Neubauer DN, et al. Modeling hypersomnolence in sleep-disordered breathing. A novel approach using survival analysis. Am J Respir Crit Care Med 1999;9:173-9. 5. Fanfulla F, Delmastro M, Berardinelli A, et al. Effects of different ventilator settings on sleep and inspiratory effort in patients with neuromuscular disease. Am J Respir Crit Care Med 25;172:619-24. 6. Bosma K, Ferreyra G, Ambrogio C, et al. Patientventilator interaction and sleep in mechanically ventilated J Thorac Dis 218;1(Suppl 1):S8-S85

Journal of Thoracic Disease, Vol 1, Suppl 1 January 218 S85 patients: pressure support versus proportional assist ventilation. Crit Care Med 27;35:148-54. 7. Parthasarathy S, Tobin MJ. Effect of ventilator mode on sleep quality in critically ill patients. Am J Respir Crit Care Med 22;166:1423-9. 8. Ramsay M, Mandal S, Suh ES, et al. Parasternal electromyography to determine the relationship between patient-ventilator asynchrony and nocturnal gas exchange during home mechanical ventilation set-up. Thorax 2;7:946-52. 9. Vrijsen B, Testelmans D, Belge C, et al. Patient-ventilator asynchrony, leaks and sleep in patients with amyotrophic lateral sclerosis. Amyotroph Lateral Scler Frontotemporal Degener 216;17:343-5. 1. Vignaux L, Vargas F, Roeseler J, et al. Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study. Intensive Care Med 29;35:84-6. 11. Bergbom-Engberg I, Haljamäe H. Patient experiences during respirator treatment--reason for intermittent positive-pressure ventilation treatment and patient awareness in the intensive care unit. Crit Care Med 1989;17:22-5. 12. Guo YF, Sforza E, Janssens JP. Respiratory patterns during sleep in obesity-hypoventilation patients treated with nocturnal pressure support: a preliminary report. Chest 27;131:19-9. 13. Crescimanno G, Canino M, Marrone O. Asynchronies and sleep disruption in neuromuscular patients under home noninvasive ventilation. Respir Med 212;16:1478-85. Cite this article as: Ramsay M. Patient ventilator asynchrony and sleep disruption during non-invasive ventilation. J Thorac Dis 218;1(Suppl 1):S8-S85. doi: 1.2137/jtd.217.11.31 J Thorac Dis 218;1(Suppl 1):S8-S85