Optical methods based on near-infrared spectroscopy. Tissue Oxygen Index Thresholds for Cerebral Ischemia Using Near-Infrared Spectroscopy

Similar documents
MONITOR CEREBRAL PERFUSION IN CAROTID ENDARTERECTOMY (CEA)

MONITORING CEREBRAL PERFUSION IN CAROTID ENDARTERECTOMY

The contribution of the external carotid artery to cerebral perfusion in carotid disease

The INVOS Cerebral/Somatic Oximeter Monitors Cerebral Perfusion in: Carotid Endarterectomy (CEA)

Measurements of oxygen saturation of brain, liver and heart areas in the supine and sitting position using near infrared spectrophotometry

Near-infrared spectroscopic cerebral oxygenation reading in neonates and infants is associated with central venous oxygen saturation

Current bedside monitors of brain blood flow and oxygen delivery

Materials and Methods

Local Haemodynamic Changes During Carotid Endarterectomy The Influence on Cerebral Oxygenation

Cerebral near infrared spectroscopy: emitter detector separation must be increased

Near-infrared spectroscopy for monitoring cerebral ischemia during selective cerebral perfusion

Oxygen Saturation Measurement of Calf Muscle During Exercise in Intermittent Claudication

On behalf of the education Committee of the SNACC

SdO 2. p Hypothermic circulatory arrest: HCA n = 6 Continuous retrograde cerebral perfusion: retrograde cerebral perfusion: IRCP

CEA and cerebral protection Volodymyr labinskyy, MD

perfusion pressure: Definitions. Implication on management protocols. What happens when CPP is too low, and when it is too high? Non-invasive CPP?

Continuous cerebral autoregulation monitoring

In vivo behaviour of near-infrared light Fig 1 Apparatus for measuring baseline recording, with both the emitter and detector on the skin surface. Fig

REVIEW ARTICLE Near infrared spectroscopy

SPECTRAL CHARACTERISTICS OF SPONTANEOUS OSCILLATIONS IN CEREBRAL HAEMODYNAMICS ARE POSTURE DEPENDENT

Absolute Cerebral Oximeters for Cardiovascular Surgical Cases

Tissue oxygenation index reflects changes in forearm blood flow after brief ischemia

Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution

1. INTRODUCTION. National Hospital for Neurology & Neurosurgery, London

Near Infrared Spectroscopy in Neonatal Intensive Care

Study on the Associating Factors with Regional Cerebral Oxygen Saturation Values during Cardiopulmonary Bypass

Lisa T. Hannegan, MS, CNS, ACNP. Department of Neurological Surgery University of California, San Francisco

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology

NIRS of the brain new diagnostic tool

Yuko Kondo, Kaoru Sakatani, Noriya Hirose, Takeshi Maeda, Jitsu Kato, Setsuro Ogawa, and Yoichi Katayama

Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy

Probing Brain Oxygenation with Near Infrared Spectroscopy

CEREBRAL oximetry using near-infrared spectroscopy

TCD in Anaesthesiology

Clinically Available Optical Topography System

In operations extending to the transverse aortic arch, selective antegrade cerebral

Near-Infrared Monitoring of Brain and Tissue Oxygenation: Is the Monitor on the Right Person?

Neuro Quiz 29 Transcranial Doppler Monitoring

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

ISPUB.COM. Intraoperative Neurophysiologic Monitoring. M Dinkel, U Beese, M Messner OBJECTIVES AND METHODS

IMPROVE PATIENT OUTCOMES AND SAFETY IN ADULT CARDIAC SURGERY.

Stanford Neuroanesthesia Syllabus: Journal Article Directory (click on section heading)

Some factors affecting cerebral tissue saturation during obstructive sleep apnoea

Chapter 37 Effects of Acupuncture on Anxiety Levels and Prefrontal Cortex Activity Measured by Near-Infrared Spectroscopy: A Pilot Study

Four-Wavelength Near-Infrared Imaging of Abdominal Aorta Blood Flow under Surgical Occlusion

19. Monitoring of CA using TCD- Mx and Sx

Disclosures. Set up audience participation. Test question. Outline. Neuromonitoring What and When? IP for monitoring technology licensed to Medtronic

Supporting Information

Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation

Chapter 19 Detection of ROSC in Patients with Cardiac Arrest During Chest Compression Using NIRS: A Pilot Study

CEREBRAL OXIMETRY IS FREQUENTLY A FIRST ALERT INDICATOR OF ADVERSE OUTCOMES

Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy

Cerebral Oximetry in Preterm Infants: Methods, Measurements and Evaluating Clinical Benefit

Case Report. Gholamreza Moradi, MD, Fereydoun Sabzi, MD, Abdolhamid Zokaei, MD *, Naser Hemati, MD

CARDIOVASCULAR Near-infrared spectroscopy as an index of brain and tissue oxygenation

Chapter 35 Gender and Age Analyses of NIRS/STAI Pearson Correlation Coefficients at Resting State

Posterior Cerebral Artery Aneurysms with Common Carotid Artery Occlusion: A Report of Two Cases

This quiz is being published on behalf of the Education Committee of the SNACC.

Physician s Vascular Interpretation Examination Content Outline

Vivek R. Deshmukh, MD Director, Cerebrovascular and Endovascular Neurosurgery Chairman, Department of Neurosurgery Providence Brain and Spine

Pre-and Post Procedure Non-Invasive Evaluation of the Patient with Carotid Disease

INVOS System Inservice Guide for Pediatric Use. INVOS System Inservice Guide for Pediatric Use

Carotid endarterectomy as complex intraoperative monitoring

The Clinical Utility of Cerebral Oximetry Monitoring

Neurovascular Physiology and Pathophysiology

Clinical application of near-infrared spectroscopy in perioperative assessment of cerebral and peripheral tissue oxygenation

Cerebral Perfusion Directly Revealed INVOS Cerebral/Somatic Oximeter Improving Patient Outcomes and Safety in Adult Surgery

Influence of external cardiac pacing on cerebral oxygenation measured by near-infrared spectroscopy in children after cardiac surgery

ESAT/SCD, Department of Electrical Engineering & IBBT Future Health Department, Katholieke Universiteit Leuven, Belgium

BRITISH BIOMEDICAL BULLETIN

noninvasive, nonionizing, portable, inexpensive, safe for serial or prolonged studies

Accuracy of Cerebral Monitoring in Detecting Cerebral Ischemia during Carotid Endarterectomy

Effect of Head and Neck Positioning on Cerebral Perfusion During Shoulder Arthroscopy in Beach Chair Position

State of the Art Multimodal Monitoring

ICP (Intracranial Pressure) Monitoring Brain Tissue Oxygen Monitoring Jugular Venous Bulb Oximetry

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Surgery for Congenital Heart Disease

The Change in Regional Cerebral Oxygen Saturation after Stellate Ganglion Block

Internal carotid artery near-total occlusions: Is it justified to operate on them?

Continuous Monitoring of Jugular Bulb Venous Oxygen Saturation for Evaluation of Cerebral Perfusion during Carotid Endarterectomy

Surgical Treatment of Aortic Arch Hypoplasia

Major Aortic Reconstruction; Cerebral protection and Monitoring

Surgical Procedures for. Symptomatic Post-CAS Carotid. Restenosis: Experiences and. Mid-Term Outcomes. Lefeng Qu M.D., Ph.D. Professor of Surgery

Title: Total Aortic Arch Replacement under Intermittent Pressure-augmented Retrograde Cerebral Perfusion

Relationship between muscle oxygenation by NIRS and blood lactate

Cerebral hyperperfusion syndrome after carotid angioplasty

Kaoru Sakatani, Yoichi Katayama, Takamitsu Yamamoto, Susumu Suzuki

Goals and Objectives of the Educational Program

A ssessment of regional cerebral blood flow (rcbf) might

Stroke & Neurovascular Center of New Jersey. Jawad F. Kirmani, MD Director, Stroke and Neurovascular Center

Role of Frequency Domain Optical Spectroscopy in the Detection of Neonatal Brain Hemorrhage A Newborn Piglet Study

GUIDELINE FOR RECOVERY ROOM MANAGEMENT OF PATIENTS AFTER CAROTID ENDARTERECTOMY

IOM at University of. Training for physicians. art of IOM. neurologic. injury during surgery. surgery on by IOM. that rate is.

Changes in middle cerebral artery blood flow after carotid endarterectomy as monitored by transcranial Doppler

Multimodal Brain Monitoring with NIRS and aeeg and their Clinical Significance

Carotid Stenting and Surgery in 2016 in Russia

Hyperperfusion syndrome after MCA embolectomy a rare complication?

Co chce/čeká neurochirug od anesteziologa během karotické endarterektomie?

Open heart surgery or carotid endarterectomy. Which procedure should be done first?

Near-infrared spectroscopy measurement of the pulsatile component of cerebral blood flow and volume from arterial oscillations

Transcription:

Tissue Oxygen Index Thresholds for Cerebral Ischemia Using Near-Infrared Spectroscopy Pippa G. Al-Rawi, BSc; Peter J. Kirkpatrick, FRCS (SN) Background and Purpose To date, the clinical application of near infrared spectroscopy in the adult brain has been limited. The NIRO 300 (Hamamatsu Photonics) provides a continuous measurement of tissue oxygen index (TOI) using spatially resolved spectroscopy. Although TOI reflects cerebral oxygenation to a high degree of sensitivity and specificity, to become a useful clinical tool, thresholds for cerebral ischemia need to be defined. This study has attempted to identify a quantifiable TOI threshold for ischemia in the adult brain. Methods One hundred sixty-seven patients undergoing carotid endarterectomy were studied. The NIRO 300 was incorporated into an established multimodal monitoring system enabling observations of cerebral hemodynamic changes under highly controlled conditions. Changes in TOI (% TOI) on clamping the internal carotid artery were compared with intracranial blood flow (middle cerebral artery flow velocity) and cerebral function monitoring to identify and quantify periods of cerebral ischemia. Results Significant correlation was seen between changes in middle cerebral artery flow velocity and TOI on clamping (r 0.74, P 0.0001). Thirty-one patients showed cerebral ischemia on internal carotid artery clamping as defined by a sustained fall in cerebral function monitoring. A threshold for % TOI of 13 was identified, above which no patients showed any evidence of ischemia on clamping. This threshold provided 100% sensitivity and 93.2% specificity for patients satisfying the preset criteria for cerebral ischemia. Conclusions These data demonstrate the potential to identify TOI-quantified thresholds for cerebral ischemia in the adult brain and thus improve the clinical use of near infrared spectroscopy. Our observations have defined a drop in TOI (13%) that can be adopted as a threshold for severe cerebral ischemia with high sensitivity and specificity. (Stroke. 2006;37:2720-2725.) Key Words: carotid endarterectomy cerebral ischemia near infrared spectroscopy spatially resolved spectroscopy tissue oxygen index Optical methods based on near-infrared spectroscopy (NIRS) can be used to monitor hemodynamic variables in the clinical setting. 1 Applications include monitors of fetal physiology, 2 4 during cardiac surgery, 5,6 muscle function, 7 10 and neonatal, 11 pediatric, 12 and adult brain. 13 19 A number of NIRS instruments are now available that detect changes in optical attenuation from several wavelengths of light, and although a variety of algorithms have been applied to attempt quantification, accuracy of these instruments remains a concern. The NIRO 300 (Hamamatsu Photonics) provides a continuous online measurement of tissue oxygen index (TOI) using spatially resolved spectroscopy. 20,21 We have previously determined that TOI reflects cerebral oxygenation to a high degree of sensitivity and specificity during carotid surgery, 20 an indication that NIRS can show clinically relevant estimations of cerebral hemodynamic and oxygen changes in the adult head (Figure 1). However, there are no data currently available to correlate TOI change with absolute measures of cerebral blood flow and/or cerebral oxygenation; hence, quantification is difficult. Defining thresholds for cerebral ischemia may be more fruitful. In this article, we have compared TOI with changes in signals from a cerebral function monitor (CFM) and transcranial Doppler ultrasonography during carotid artery crossclamping in an attempt to identify a threshold for ischemia in the adult brain. Materials and Methods Near-Infrared Spectroscopy The principle for NIRS is that near infrared light (delivered through optodes placed on the skin) penetrates the scalp and brain tissue whereby scatter and absorption occurs within the various tissues layers. The concentration changes of the chromophores oxy-, deoxy-, and total hemoglobin are measured by a modified Beer-Lambert method. Calculation of the TOI depends on spatially resolved spectroscopy; TOI is defined as the ratio of oxygenated to total tissue hemoglobin. 21,22 Received June 21, 2006; accepted July 20, 2006. From the University Department of Neurosurgery, Addenbrooke s Hospital, Cambridge, UK. Correspondence to Pippa G. Al-Rawi, BSc, University Department of Neurosurgery, Box 167, Level 4, A-Block, Addenbrooke s Hospital, Hills Road, Cambridge, CB2 2QQUK. E-mail pga20@medschl.cam.ac.uk 2006 American Heart Association, Inc. Stroke is available at http://www.strokeaha.org DOI: 10.1161/01.STR.0000244807.99073.ae 2720

Al-Rawi and Kirkpatrick TOI Threshold for Cerebral Ischemia 2721 Figure 1. Graphic display of data obtained from a patient during elective carotid endarterectomy. The vertical lines demonstrate time of application of vascular clamps. ABP indicates mean arterial blood pressure; LDF, frontal cutaneous laser Doppler flow. Method The NIRO 300 was incorporated into an established multimodal monitoring system for patients undergoing carotid endarterectomy enabling observations of cerebral hemodynamic change under highly controlled conditions. During surgery, brief and variable degrees of cerebral ischemia occur during crossclamping of the internal carotid artery. 23 26 Routine intraoperative monitoring of mean ipsilateral middle cerebral artery flow velocity (FV; SciMed), laser Doppler flowmetry (Moor Instruments Ltd), bilateral CFM (LectroMed UK Ltd), and mean arterial blood pressure has helped to characterize these changes. 23 The NIRO 300 probes were placed high on the ipsilateral forehead to avoid the temporalis muscle and sufficiently lateral from the midline to avoid the superior sagittal sinus. NIRS optodes were kept at a constant distance of 5 cm in a specialized rubber holder and secured to the skin with an adhesive sheet. In addition, the optode holder was secured still further by a crepe bandage around the head. The sampling time was set at 5 seconds. Sequential clamping of the external (ECA) and internal carotid arteries (ICA) was performed intraoperatively as previously described allowing a sufficient time interval (approximately 2 minutes) for stabilization of signals after each clamp application. 27 Severe cerebral ischemia was defined as a persistent fall in CFM and a drop in mean flow velocity to 40% of baseline. Selective shunting was performed in the presence of either of these criteria. Patients After Local Research Ethics Committee approval and with informed consent, 167 patients (121 male and 46 female) undergoing elective carotid endarterectomy were studied. The mean age was 69 years (range, 44 to 86 years). Monitoring was applied after induction of anesthesia, when the patient entered the operating room. All patients underwent direct microscopic closure of the endarterectomized vessel with the exception of those patients who had very tortuous or small arteries when a selective patch repair was undertaken. Anesthesia was standardized and patients were maintained physiologically stable throughout the procedure. Data Processing and Analysis Data signals from all the monitored parameters were digitized and collected using specialized multimodality software. 28,29 Maximum physiological stability is generally observed at the time of application of clamps rather than removal; therefore, data from this period were used for analysis. Values of mean arterial pressure, laser Doppler flowmetry, FV and TOI were averaged at baseline (before clamping but after completion of dissection) and over a 2-minute period immediately after ECA and ICA clamping. Changes in TOI ( TOI) on clamping the ICA were compared with intracranial blood flow (middle cerebral artery flow velocity [MCA FV]) and CFM to identify and quantify periods of cerebral ischemia. Data are given as mean SD.

2722 Stroke November 2006 TABLE 1. Summary of Baseline and Percent Change (% ) in Mean Flow Velocity (mfv; cm/sec), Mean Arterial Blood Pressure (mabp; mm Hg), and TOI (%) During Sequential Clamping Baseline (mean SD) Post ICA Clamping (mean SD) TOI (%) 68 8.96 (range, 46 93.6) 64.1 10.3 (range, 40.6 93) FV (cm/sec) 37.8 13.18 (range, 17.2 76) 28.8 13.7 (range, 4.54 92.4) mabp (mm Hg) 89.1 14.3 (range, 60.9 121) 89.3 17.2 (range, 53.4 114) Results Looking at the pooled data for all patients, the values obtained before and after clamping of the internal carotid artery are shown in Table 1. The mean baseline flow velocity (MCA FV) was 37.8 cm/second and the mean TOI was 68%. There was no significant change in mean arterial blood pressure seen with clamping. The mean percentage change in TOI from baseline to after ICA clamping (% TOI) was 8.2 ( 9.0). Significant correlation was seen between changes in MCA FV and TOI on clamping (r 0.74, P 0.0001). Of the 167 patients, 31 (19%) showed severe cerebral ischemia on ICA clamping as defined by a sustained fall in ipsilateral cerebral function monitoring (Figure 2) and underwent selective shunting. We defined a % TOI threshold of 13%, above which no patients showed any evidence of ischemia. Below the threshold of 13%, 10 false-positive cases occurred. Six showed a fall in both flow velocity and TOI on clamping the ECA with a further drop on clamping the ICA (mean arterial blood pressure remaining stable). In one case, review of the data showed that the ECA clamp time was too brief to assess the extracranial contribution to the signal. In a further case, the calculated value of % TOI ( 13.6%) was very close to the defined threshold value of 13%. Finally, in the remaining 2 patients, a significant increase in FV occurred shortly after clamping the ICA, one being associated with a significant increase in blood pressure at the time of ICA clamping. The threshold for % TOI of 13 provided 100% sensitivity and 93.2% specificity for patients satisfying the preset criteria for cerebral ischemia (Table 2). Discussion Many established forms of neuromonitoring require technical expertise and experience in interpreting the results. 30 Cerebral oximetry is an attractive alternative as a continuous, noninvasive measurement that is both safe and simple to apply. This study has addressed the issue of determining a threshold for ischemia when using NIRS for monitoring the adult brain. Our approach has been to sequentially clamp the ECA and ICA during carotid endarterectomy, allowing the opportunity to separate the intracranial and extracranial signal contribu- Figure 2. Scatterplot comparing percentage change in MCA flow velocity (% FV) with percentage change in tissue oxygen index (% TOI) from baseline to postapplication of ICA clamp for all 167 patients. Patients exhibiting a sustained fall in ipsilateral CFM marked as unfilled symbols.

Al-Rawi and Kirkpatrick TOI Threshold for Cerebral Ischemia 2723 TABLE 2. Contingency Table of Sustained Fall in CFM by Change in TOI > 13% on ICA Clamping* Fall in CFM No Fall in CFM Total TOI 13% 31 10 41 TOI 13% 0 126 126 Total 31 136 167 *Sensitivity of TOI to detect cerebral ischemia (at threshold of % TOI 13%) 100% and specificity 93.2%. tions and to identify those patients with severe cerebral ischemia. 23,27,31 Mean % TOI (including those patients where TOI increased on clamping), was 8.2 and changes were closely correlated with FV. In all patients who showed severe cerebral ischemia (n 31), TOI fell by 13%. The sensitivity of a % TOI threshold of 13 was 100% (Table 2). However some cases (n 10) showed a fall in TOI in the absence of satisfying the criteria for severe critical ischemia, leading to a lower specificity of 93.2%. These findings were on a background of a stable blood pressure; hence, the divergence between TOI and ischemic criteria requires more complicated anatomic and physiological considerations. Intermodal variability in the absolute optical measurements of hemoglobin concentration affects all types of noninvasive NIRS measurement. Interpatient variability in head geometry accounts for many model-related systematic errors in the absolute measurements. However, by focusing on relative hemodynamic and oxygenation changes, we believe that such systematic errors are largely canceled out. Sensitivity to superficial tissue layers is a common limitation of noninvasive NIRS for the adult brain. 32,33 However, we have previously shown that the NIRO 300 TOI measurement reflects adult intracranial brain tissue oxygenation to a high degree of sensitivity and specificity. 20 When a fall in both FV and TOI was seen on clamping the ECA, this probably reflects reverse flow from ECA to ICA through extracranial to intracranial collaterals. Thus, clamping the ECA will affect both cranial and intracranial blood compartments, augmenting TOI. When there is a lack of correlation between TOI and FV, which cannot be explained by the presence of simple EC-IC collaterals, or by arterial blood pressure instability, complex vascular anatomy may be accountable. A more robust anterior hemispheric contribution from the anterior cerebral artery with crossflow through the anterior communicating artery has already been shown to influence the chances of severe cerebral ischemia during carotid clamping 34 and serves to contaminate the relationship between TOI and FV, the latter being a pure MCA derivative. Cerebral oxygenation measured by NIRS may be influenced by changes in cerebral blood flow, cerebral metabolism, arterial saturation, and hematocrit. 35 However, arterial saturation was over 99% in each case and because there was no blood loss, it is unlikely that hematocrit changed during the surgery up until the point of crossclamping. Likewise, patients were maintained under stable anesthesia so we consider changes in cerebral metabolic rate unlikely to influence the observations. The difficulty in defining the normal range of cerebral oxygenation remains. In adults, normal TOI values have been reported to range from 65% to 85%. 14,36 41 Although a number of authors have compared NIRS with other techniques such as jugular bulb oxygen saturation (SjO 2 ), transcranial Doppler ultrasonography and somatosensory-evoked potential 15,37,42 56 or with neurologic events, few have attempted to define an absolute threshold for ischemia. 13,14,16,17,46,57 59 Most studies of this type using NIRS make no attempt to correct for extracranial contributions to signal changes. Of those who have tried to identify a threshold, the results show a low sensitivity and specificity. 17,46,58,59 Furthermore, given the low incidence of clamp-related ischemia in carotid surgery, a large number of patients need to be studied to obtain sufficient power. Very few studies have achieved this. In a study of 317 patients undergoing carotid endarterectomy, Beese et al compared somatosensory-evoked potential with regional oxygen saturation (rso 2 ; INVOS 3100A) to detect severe cerebral ischemia requiring shunt placement. 57 A significant decrease in rso 2 was seen in patients both with and without loss of cortical somatosensory-evoked potential. Although the difference between the decreases seen in the 2 groups was highly significant; they were unable to determine a threshold resulting from individual variability of both the rso 2 and the derived changes. A more recent study of 594 patients undergoing carotid endarterectomy aimed to determine whether NIRS could be used to determine the need for shunt placement. 60 Using the INVOS 3100-A and INVOS 4100-SSA, a drop of 20% rso 2 was defined as clinically relevant. The sensitivity and specificity for this threshold was 30% and 98%, respectively. In both these studies, however, no attempt was made to determine the ECA contribution to the signal. In conclusion, the NIRO 300 provides a measure of cerebral oxygenation that can be monitored during carotid surgery. Relative changes in TOI correlated well with FV. Our observations have defined a drop in TOI (13%) that can be adopted as a threshold for severe cerebral ischemia with high sensitivity and specificity. Confirmation that these findings apply in different clinical scenarios is needed, however, especially those where scalp swelling (eg, trauma, postcraniotomy) is present. Acknowledgments The authors thank K. Varty and M.E. Gaunt for assistance with sequential clamping during carotid endarterectomy and Hamamatsu Photonics K.K. for use of the NIRO 300 for the duration of this study. None. Disclosures References 1. Strangman G, Franceschini MA, Boas DA. Factors affecting the accuracy of near-infrared spectroscopy concentration calculations for focal changes in oxygenation parameters. Neuroimage. 2003;18:865 879. 2. Aldrich CJ, D Antona D, Wyatt JS, Spencer JAD, Peebles DM, Reynolds EOR. Fetal cerebral oxygenation measured by near infrared spectroscopy shortly before birth and acid-base status at birth. Obstet Gynecol. 1994; 84:861 866. 3. Parilla BV, Tamura RK, Cohen LS, Clark E. Lack of effect of antenatal indomethacin on fetal cerebral blood flow. Am J Obstet Gynecol. 1997; 176:1166 1171.

2724 Stroke November 2006 4. Nicklin SE, Hassan IA, Wickramasinghe YA, Spencer SA. The light still shines, but not that brightly? The current status of perinatal near infrared spectroscopy. Arch Dis Child Fetal Neonatal Ed. 2003;88:F263 F268. 5. Nollert G, Jonas RA, Reichart B. Optimising cerebral oxygenation during cardiac surgery: a review of experimental and clinical investigations with near infrared spectrophotometry. Thorac Cardiovasc Surg. 2000;48: 247 253. 6. Taillefer MC, Denault AY. Cerebral near-infrared spectroscopy in adult heart surgery: systematic review of its clinical efficacy. Can J Anaesth. 2005;52:79 87. 7. Nioka S, Moser D, Lech G, Evengelisti M, Verde T, Chance B, Kuno S. Muscle deoxygenation in aerobic and anaerobic exercise. Adv Exp Med Biol. 1998;454:63 70. 8. Boushel R, Langberg H, Olesen J, Gonzales-Alonzo J, Bulow J, Kjaer M. Monitoring tissue oxygen availability with near infrared spectroscopy (NIRS) in health and disease. Scand J Med Sci Sports. 2001;11:213 222. 9. McCully KK, Hamaoka T. Near-infrared spectroscopy: what can it tell us about oxygen saturation in skeletal muscle? Exerc Sport Sci Rev. 2000; 28:123 127. 10. Quaresima V, Lepanto R, Ferrari M. The use of near infrared spectroscopy in sports medicine. J Sports Med Phys Fitness. 2003;43:1 13. 11. Schulz G, Keller E, Haensse D, Arlettaz R, Bucher HU, Fauchère J-C. Slow blood sampling from an umbilical artery catheter prevents a decrease in cerebral oxygenation in the preterm newborn. Pediatrics. 2003;111:73 76. 12. Dullenkopf A, Frey B, Baenziger O, Gerber A, Weiss M. Measurement of cerebral oxygenation state in anaesthetized children using the INVOS 5100 cerebral oximeter. Paediatr Anaesth. 2003;13:384 391. 13. Hirofumi O, Otone E, Hiroshi I, Satosi I, Shigeo I, Masato NYS. The effectiveness of regional cerebral oxygen saturation monitoring using near-infrared spectroscopy in carotid endarterectomy. J Clin Neurosci. 2003;10:79 83. 14. Carlin RE, McGraw DJ, Calimlim JR, Mascia MF. The use of nearinfrared cerebral oximetry in awake carotid endarterectomy. J Clin Anesth. 1998;10:109 113. 15. Ferrari M, Zanette E, Sideri G, Giannini I, Fieschi C, Carpi A. Effects of carotid compression, as assessed by near infrared spectroscopy, upon cerebral blood volume and hemoglobin oxygen saturation. J R Soc Med. 1987;80:83 87. 16. Takeda N, Fujita K, Katayama S, Tamaki N. Cerebral oximetry for the detection of cerebral ischaemia during temporary carotid artery occlusion. Neurol Med Chir (Tokyo). 2000;40:557 563. 17. Samra SK, Dy EA, Welch K, Dorje P, Zelenock GB, Stanley JC. Evaluation of a cerebral oximeter as a monitor of cerebral ischaemia during carotid endarterectomy. Anaesthesiology. 2000;93:970. 18. Madsen PL, Secher NH. Near-infrared oximetry of the brain. Prog Neurobiol. 1999;58:541 560. 19. Obrig H, Villringer A. Beyond the visible; imaging the human brain with light. J Cereb Blood Flow Metab. 2003;23:1 18. 20. Al-Rawi PG, Smielewski P, Kirkpatrick PJ. Evaluation of a near infrared spectrometer (NIRO 300) for the detection of intracranial oxygenation changes in the adult head. Stroke. 2001;32:2492 2500. 21. Matcher SJ, Kirkpatrick PJ, Nahid K, Cope M, Delpy DT. Absolute quantification methods in tissue near infrared spectroscopy. Proc SPIE. 1993;2389:486 495. 22. Suzuki S, Takasaki S, Ozaki T, Kobayashi Y. A tissue oxygenation monitor using NIR spatially resolved spectroscopy. Proc SPIE. 1999; 3597:582 592. 23. Kirkpatrick PJ, Smielewski P, Whitfield PC, Czosnyka M, Menon D, Pickard JD. An observational study of near infrared spectroscopy during carotid endarterectomy. J Neurosurg. 1995;82:756 763. 24. Chan K-H, Miller JD, Dearden NM, Andrews PJD, Midgley S. The effect of changes in cerebral perfusion pressure upon middle cerebral artery blood flow velocity and jugular bulb venous oxygen saturation after severe brain injury. J Neurosurg. 1992;77:55 61. 25. Graham AM, Gewetz BL, Zarins CK. Predicting cerebral ischaemia during carotid endarterectomy. Arch Surg. 1986;121:595 598. 26. Jorgensen LG, Schroeder TV. Transcranial Doppler for detection of cerebral ischaemia during carotid endarterectomy. Eur J Vasc Surg. 1992;6:142 147. 27. Kirkpatrick PJ, Lam JMK, Al-Rawi PG, Smielewski P, Czosnyka M. Defining thresholds for critical ischaemia by using near-infrared spectroscopy in the adult brain. J Neurosurg. 1998;89:389 394. 28. Czosnyka M, Whitehouse H, Smielewski P, Kirkpatrick PJ, Guazzo EP, Pickard JD. Computer supported multimodal bed-side monitoring for neuro intensive care. J Clin Monit Comput. 1994;11:223 232. 29. Smielewski P, Czosnyka M, Zabolotny W, Kirkpatrick PJ, Richards H, Pickard JD. A computing system for the clinical and experimental investigation of cerebrovascular reactivity. Int J Clin Monit Comput. 1997;14: 185 198. 30. Cheng MA, Theard MA, Tempelhoff R. Anesthesia for carotid endarterectomy: a survey. J Neurosurg Anaesthesiol. 1997;9:211 216. 31. Kirkpatrick PJ, Smielewski P, Al-Rawi P, Czosnyka M. Resolving extraand intracranial signal changes during adult near infrared spectroscopy. Neurol Res. 1998;20:S19 S22. 32. Germon TJ, Evans DH, Barnett N, Wall P, Manara AR, Nelson RJ. Cerebral near infrared spectroscopy: emitter-detector separation must be increased. Br J Anaesth. 1999;82:831 837. 33. Lam JMK, Kirkpatrick PJ, Al-Rawi P, Smielewski P, Pickard JD. Internal and external carotid contribution to near infrared spectroscopy (NIRS) during carotid endarterectomy (CE). J Neurol Neurosurg Psychiatry. 1996;61:553. 34. Lam JMK, Smielewski P, Al-Rawi P, Griffiths P, Yu AL, Pickard JD, Kirkpatrick PJ. Prediction of cerebral ischaemia during carotid endarterectomy with preoperative CO 2 -reactivity studies and angiography. Br J Neurosurg. 2000;14:441 448. 35. Kurth CD, Uher B. Cerebral haemoglobin and optical path length influence near-infrared spectroscopy measurement of cerebral oxygen saturation. Anesth Analg. 1997;84:1297 1305. 36. Quaresima V, Sacco S, Totaro R, Ferrari M. Noninvasive measurement of cerebral hemoglobin oxygen saturation using two near infrared spectroscopy approaches. J Biomed Opt. 2000;5:201 205. 37. Samra SK, Dorje P, Zelenock GB, Stanley JC. Cerebral oximetry in patients undergoing carotid endarterectomy under regional anaesthesia. Stroke. 1996;27:49 55. 38. Misra M, Stark J, Dujovny M, Widman R, Ausman JL. Transcranial cerebral oximetry in random normal subjects. Neurol Res. 1998;20: 137 141. 39. Nielsen HB, Boushel R, Madsen P, Secher NH. Cerebral desaturation during exercise reversed by O 2 supplementation. Am J Physiol. 2001;277: H1045 H1052. 40. Thavasothy M, Broadhead M, Elwell C, Peters M, Smith M. A comparison of cerebral oxygenation as measured by the NIRO 300 and the INVOS 5100 near-infrared specrophotometers. Anaesthesia. 2002;57: 999 1006. 41. Yoshitani K, Kawaguchi M, Tatsumi K, Kitaguchi K, Furuya H. A comparison of the INVOS 4100 and the NIRO 300 near-infrared spectrometers. Anesth Analg. 2002;94:586 590. 42. Kragsterman B, Pärsson H, Bergqvist D. Local haemodynamic changes during carotid endarterectomy the influence on cerebral oxygenation. Eur J Vasc Endovasc Surg. 2004;27:398 402. 43. Williams IM, Picton A, Farrell A, Mead GE, Mortimer AJ, McCollum CN. Light reflective cerebral oximetry and jugular bulb venous oxygen saturation during carotid endarterectomy. Br J Surg. 1994;81: 1291 1295. 44. Williams IM, Vohra R, Farrell A, Picton AJ, Mortimer AJ, McCollum CN. Cerebral oxygen saturation, transcranial Doppler ultrasonography and stump pressure in carotid surgery. Br J Surg. 1994;81:960 964. 45. Cho H, Nemoto EM, Yonas H, Balzer J, Sclabassi RJ. Cerebral monitoring by means of oximetry and somatosensory evoked potentials during carotid endarterectomy. J Neurosurg. 1998;89:533 538. 46. Duffy CM, Manninen PH, Chan A, Kearns CF. Comparison of cerebral oximeter and evoked potential monitoring in carotid endarterectomy. Can J Anaesth. 1997;44:1077 1081. 47. Vets P, ten Broecke P, Adrianensen H, Van Schil P, De Hert S. Cerebral oximetry in patients undergoing carotid endarterectomy: preliminary results. Acta Anaesthesiol Belg. 2004;55:215 220. 48. Williams IM, Mead G, Picton AJ, Farrell A, Mortimer AJ, McCollum CN. The influence of contralateral carotid stenosis and occlusion on cerebral oxygen saturation during carotid artery surgery. Eur J Vasc Endovasc Surg. 1995;10:198 206. 49. Cuadra SA, Zwerling JS, Feuerman M, Gasparis AP, Hines GL. Cerebral oximetry monitoring during carotid endarterectomy: effect of carotid clamping and shunting. Vasc Endovascular Surg. 2003;37:407 413. 50. Samra SK, Chandler WF. Cerebral oximetry during circulatory arrest for aneurysm surgery. J Neurosurg Anaesthesiol. 1997;9:154 158.

Al-Rawi and Kirkpatrick TOI Threshold for Cerebral Ischemia 2725 51. Orihashi K, Sueda T, Okada K, Imai K. Near-infrared spectroscopy for monitoring cerebral ischemia during selective cerebral perfusion. Eur J Cardiothorac Surg. 2004;26:907 911. 52. Horie N, Kitagawa N, Morikawa M, Kaminogo M, Nagata I. Monitoring of regional cerebral oxygenation by near-infrared spectroscopy in carotid arterial stenting: preliminary study. Neuroradiology. 2005; 47:375 379. 53. Kaminogo M, Ochi M, Onizuka M, Takahata H, Shibata S. An additional monitoring of regional cerebral oxygen saturation to HMPAO SPECT study during balloon test occlusion. Stroke. 1999;30:407 413. 54. Yoshitani K, Kawaguchi M, Iwata M, Sasaoka N, Inoue S, Kurumatani N, Furuya H. Comparison of changes in jugular venous bulb oxygen saturation and cerebral oxygen saturation during variations of haemoglobin concentration under propofol and sevoflurane anaesthesia. Br J Anaesth. 2005;94:341 346. 55. Chen CS, Leu BK, Liu K. Detection of cerebral desaturation during cardiopulmonary bypass by cerebral oximetry. Acta Anaesthesiol Sin. 1996;34:173 178. 56. Ali MS, Harmer M, Vaughan RS, Dunne JA, Latto IP. Spatially resolved spectroscopy (NIRO-300) does not agree with jugular bulb oxygen saturation in patients undergoing warm bypass surgery. Can J Anaesth. 2001;48:497 501. 57. Beese U, Langer H, Lang W, Dinkel M. Comparison of near-infrared spectroscopy and somatosensory evoked potentials for the detection of cerebral ischaemia during carotid endarterectomy. Stroke. 1998;29:2032 2037. 58. Rigamonti A, Scandroglio M, Minicucci F, Magrin S, Carozzo A, Casati A. A clinical evaluation of near-infrared cerebral oximetry in the awake patient to monitor cerebral perfusion during carotid endarterectomy. J Clin Anesth. 2005;17:426 430. 59. Grubhofer G, Plöchl W, Skolka M, Czerny M, Ehrlich M, Lassnigg A. Comparing Doppler ultrasonography and cerebral oximetry as indicators for shunting in carotid endarterectomy. Anesth Analg. 2000;91: 1339 1344. 60. Mille T, Tachimiri ME, Klersy C, Ticozzelli G, Bellinzona G, Blangetti I, Pirrelli S, Lovotti M, Odero A. Near infrared spectroscopy monitoring during carotid endarterectomy: which threshold value is critical? Eur J Vasc Endovasc Surg. 2004;27:646 650.