Customized Cranial and Craniofacial Implants. Instructions For Use. 7000revD_Instructions_For_Use. Page 1 of 8

Similar documents
Pinit Plate Small Bone Fusion System Bone Plate & Screw System

For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM

Technique Guide Small Bone Fusion System

Technique Guide KISSloc Suture System

Precautionary Statement ( )

DISTAL RADIUS. Instructions for Use

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY

Low Profile Neuro Plating System. Surgical Technique

PRO-DENSE Bone Graft Substitute The following languages are included in this packet:

SYNPOR POROUS POLYETHYLENE IMPLANTS. For craniofacial and orbital augmentation and reconstruction

PRO-DENSE BONE GRAFT SUBSTITUTE

Cranial System. Neuro Plating and Screw System

The following languages are included in this packet:

TABLE OF CONTENTS. ShurFit Anterior Cervical Interbody Fusion (ACIF) System Overview 2. Implant Specifications 3. Instrument Features 4

SYMBOLOGY. = Manufacturer. = Manufacturing Serial Number. = Manufacturing Lot Number = Catalog Number LOT REF. = Do not resterilize Single Use Only

For use by an Accredited Orthopaedic Surgeon only

Talar Dome System Surgical Technique

WristMotion Wrist Hemiarthroplasty System Instructions for Use

OPERATIVE TECHNIQUE. CONSTRUX Mini PTC. Mini PTC Spacer System

ENDO-FUSE INTRA-OSSEOUS FUSION SYSTEM The following languages are included in this packet:

low ProfIle neuro PlaTIng system

INTELLIGENT SPINAL SYSTEM

D. The following factors are of extreme importance to the eventual success of the procedure.

RAPIDSORB RAPID ReSORBABLe CRANIAL CLAMP

Flexible Fragment Fixation. Surgical Technique

Lapidus Arthrodesis System Instructions for Use

Revision A Date:

METAL HEMI SYSTEM

Hip Resurfacing System

TC-PLUS Primary Knee IMPORTANT MEDICAL INFORMATION SPECIAL NOTE. This Package Insert is for product distributed in the US only.

ACP1 CERVICAL PLATE SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE II.

MEDPOR. Oral maxillofacial surgery

Document No Rev. B Pg. 1 of 5 Approvals / Date Title: INSTRUCTIONS FOR USE DJO SURGICAL MODULAR REVISION HIP SYSTEM AND ACETABULAR CAGE

Headless Compession Screw 2.5 / 3.0

Headless Compession Screw 4.5 / 6.5 SURGICAL TECHNIQUE. Titanium or Stainless Steel. Cannulated Headless Design. Multiple Thread Options.

BIOFOAM ANKLE SPACER BLOCK

Cranial idtm. Cranial restoration. id SolutionsTM. Individually designed. Personalized care.

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY

Technique Guide. SynPOR Porous Polyethylene Implants. For craniofacial and orbital augmentation and reconstruction.

Aesculap CeSpace TM PEEK and CeSpace TM XP Spinal Implant System Instructions for Use

CELLPLEX TCP SYNTHETIC CANCELLOUS BONE

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician.

STERILE IMPLANT PRODUCT INSERT PATIENT-FITTED TEMPOROMANDIBULAR JOINT RECONSTRUCTION PROSTHESIS SYSTEM

CRANIAL RECONSTRUCTION SOLUTIONS

REDUCT Headless Compression Screw System

Instructions for Use Reform POCT System Caution: Federal (USA) law restricts this device to sale by or on the order of a physician

Asnis. Micro Cannulated screw system. Xpress operative technique

Duraloc CONSTRAINED LINER

EVOLVE TRIAD SYSTEM

Surgical Technique Guide

Gallery Laminoplasty Spine System

ORTHOLOC 3Di ANKLE FUSION PLATING SYSTEM

CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE

Device Preparation (all steps to be performed per standard interventional technique)

Metallic Internal Fixation Devices The following languages are included in this packet:

EVOLVE TRIAD BONE SCREWS

JuggerLoc Bone-to-Bone System for Ankle Syndesmosis Fixation. Surgical Technique

Technique Guide Wrist Hemiarthroplasty System

C-THRU Anterior Spinal System

Important Information on ATLAS Hip Fracture Nail For use by an Accredited Orthopaedic Surgeon only

Olecranon Locking Plate II

Instructions for Use AccuFit Lateral Plate System Caution: Federal (USA) law restricts this device to sale by or on the order of a physician

Visit Linvatec.com today to learn more. Surgical Technique: Sequential Meniscal Running Stitch

Technique Guide. Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth.

Dentoalveolar Bone Fixation System. Bone fixation sets for the office.

Technique Guide Hip Resurfacing System

nvp Posterior Lumbar Interbody Fusion System

nvt Transforaminal Lumbar Interbody Fusion System

MetaFix Ludloff Plate

Table of Contents.

Apache Cervical Interbody Fusion Device. Surgical Technique. Page of 13. LC-005 Rev F

Instructions for use for Bone Plates, Bone Screws, Intramedullary Nails, Pins and Wires

ALIGN Radial Head System INSTRUCTIONS FOR USE

nva Anterior Lumbar Interbody Fusion System

CRANIOMAXILLOFACIAL. Rethinking Possibilities, Reshaping Lives

PPS P I N P O S I T I O N I N G S Y S T E M GMK EFFICIENCY VERSION. Hip Knee Spine Navigation

TABLE OF CONTENTS. Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview. Implants 3. Instruments 5. Surgical Technique 10

PediLoc Tibia Surgical Technique

Instructions for Use DENTAL IMPLANTS

Endorectal Balloon (ERB) Endorectal Balloon (ERB) Instructions for Use

OBSOLETE. Visit for the latest version. Ankle Plating System 3 PKGI-79-B EFFECTIVE

SIDEKICK EZ FRAME EXTERNAL FIXATION SYSTEM The following languages are included in this packet:

Hinged Laminoplasty System surgical technique

HawkeyeTM Peek. surgical technique

Mini External Fixator

TM TM Surgical Technique

Ceramic Femoral Heads and Acetabular Cup Liners

ACETABULAR CUP SURGICAL TECHNIQUE

EVOS MINI with IM Nailing

The following languages are included in this packet:

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

X-spine Aranax Anterior Cervical Plating System Instructions for Use / Package Insert

NEODENT GUIDED SURGERY

OPERATIVE TECHNIQUE. anterior cervical plating system

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Technique Guide. Midface Distractor System. For the temporary stabilization and gradual lengthening of the cranial or midfacial bones.

Page 1 of 7 Doc. No.: IU/15 Rev. No.: 05 Rev. Date:03MAY2018. Instructions for use for ATLAS Tibial Fracture(TF) Nail

KASM Knee Articulating Spacer Mold. Surgical Summary

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

Transcription:

Customized Cranial and Craniofacial Implants Instructions For Use Page 1 of 8

Table of Contents INDICATIONS... 3 DESCRIPTION OF DEVICE... 3 CONDITIONS OF USE... 3 CONTRAINDICATIONS... 3 KELYNIAM GLOBAL INC. CUSTOMIZED CRANIAL/CRANIOFACIAL IMPLANT... 4 POTENTIAL ADVERSE REACTIONS... 5 INTRAOPERATIVE AND EARLY POSTOPERATIVE COMPLICATIONS CAN INCLUDE:... 5 LATE POSTOPERATIVE COMPLICATIONS CAN INCLUDE:... 5 INSPECTION... 5 CLEANING... 6 STERILIZATION... 6 RECOMMENDATIONS FOR STEAM STERILIZATION... 6 STORAGE... 6 DO NOT USE IF THE PACKAGE IS OPEN OR DAMAGED... 6 MAGNETIC RESONANCE IMAGING... 6 CAUTION... 6 SYMBOLS CONTAINED IN DEVICE LABELING... 7 NOTES... 7 INFORMATION... 8 Page 2 of 8

INDICATIONS Patient Specific Implants are intended for the replacement of bony voids in the cranial/craniofacial skeleton. DESCRIPTION OF DEVICE The Kelyniam Global Inc. customized cranial/craniofacial implants are a non-load bearing, single use device and are: Manufactured using Polyether ether ketone (PEEK) implantable grade materials (i.e. PEEK-OPTIMA or Vestakeep i4 ). Designed individually for each patient to correct defects in cranial/facial bone. Individually sized and shaped implantable prosthetic cranioplasty plates intended to fill defects in a specific patient s cranial skeleton. Fabricated using the patient s CT imaging data. Provided non-sterile for sterilization prior to implantation. Attached to the native bone using commercially available, non-biodegradable fixation screws or any commercially available cranioplasty fixation system. CONDITIONS OF USE CAUTION - The Kelyniam Global Inc. customized cranial/craniofacial implants should only be implanted by surgeons who are fully trained and licensed in the use of such implants and the required specialized cranial surgical techniques. CONTRAINDICATIONS This device is contraindicated under any of the following conditions: Infection and sepsis; Degenerative bone disease which would render the device or the treatment unjustifiable; Distant foci of infection which can spread to the implant site; Uncooperative patients or patients with neurologic or psychiatric/psychological dysfunction who are incapable or unwilling to follow postoperative instructions. Page 3 of 8

KELYNIAM GLOBAL INC. CUSTOMIZED CRANIAL/CRANIOFACIAL IMPLANT These devices are used for augmenting and contouring bone. They are not intended or designed for full or partial load bearing. Do not use these devices for replacement of bone within articulating surfaces. Patients who engage in contact sports or other activities that risk facial injury are to be warned that facial injury may lead to damage of the implant and a subsequent failure of treatment. The patient is to be warned that the device does not replace normal healthy bone and that traumatic injury could necessitate surgical treatment. The patient must be advised of surgical risks and the possible adverse effects. CAUTION - This device has been designed to fit the defect existing at the time of the CT Scan and implant fabrication. changes in the patient s anatomy occurring after the CT Scan as well as the use of the implant after such changes may result in a sub-optimal fit within the defected area. Improper selection, placement, positioning, and fixation of the Kelyniam Global Inc. Customized Cranial/Craniofacial Implant can cause a subsequent undesirable result. The surgeon is to be familiar with the implant and the surgical procedure prior to performing the surgery. If a commercially available, non-biodegradable fixation system are to be used, a minimum of 3 fixation points is recommended. The screw holes must be pre-drilled (away from the surgical site) and should be at least 4mm from the edge of the implant. The screws should not protrude past the underside of the implant to avoid tissue damage. Due to the customized nature of the implant, the appropriate selection of FDA approved fixation devices are left to the surgeon s discretion. All instructions for fixation (drilling, tapping of holes, and insertion of screws) should be followed by FDA approved commercially available, non-biodegradable manufacturer s instructions and specifications. Optional pre-surgical models may have been provided to allow for detailed orientation and inspection of form, fit, & function (proper alignment and orientation of the device). Intra-operative shaping and sizing of the implant can be critical to the cosmetic success of the procedure. Improper shape and size can result in a noticeable undesirable prominence or possible disfigurement at or near the implant site. All intra-operative shaping and sizing should be performed away from the surgical site. Kelyniam Customized Cranial/Craniofacial Implants placed, positioned and fixated over or near air containing sinuses could result in infection. Kelyniam Global Inc. Customized Cranial/Craniofacial Implants are not recommended for patients that have had radiation therapy. Should the surgeon determine that use of this implant is necessary and appropriate for a patient that has had previous radiation therapy; the surgeon should be familiar with all applicable surgical techniques. To prevent dehiscence at the incision site, a firm primary closure of the incision is required. If the surgeon deems necessary, the re-shaping, sizing, or contouring of the Kelyniam Global Inc. Customized Cranial/Craniofacial Implant, it is best accomplished using high-speed rotating instruments. After the implant has been shaped, they must be rinsed in sterile saline to remove any loose particles. Particularly in instances where implants are shaped, intra-operative damage to the Page 4 of 8

molded implant may occur. It is recommended that the implant be examined for damage or disfigurement prior to implantation. Pediatric use is not recommended. Rapid remodeling of the pediatric skull may cause dehiscence of the incision, prominence or disfigurement at the implant site, or related complications that could result in the need to remove the implant. The surgeon should weigh the risks versus benefits when deciding to remove the implant. Implant removal should be followed by adequate postoperative management. Do not reuse Kelyniam Global Inc. Customized Cranial/Craniofacial Implants. Discard any unused portion. The device must be sterilized prior to use per ANSI/AAMI ST79. Surgical models, if provided, are to be used for pre-operative planning analysis only and are not intended to come in contact with the Kelyniam Customized Cranial/Craniofacial Implants or enter the Operating Room. POTENTIAL ADVERSE REACTIONS While rare, implantation of foreign materials may result in sensitivity reactions. Peripheral neuropathies have been reported in conjunction with surgical procedures involving implantation of various types of devices. Sub-clinical nerve damage occurs more frequently, usually as a result of surgical exposure/trauma. Kelyniam Global Inc. Customized Cranial/Craniofacial Implants can loosen/migrate due to loss of fixation or trauma. Infection can lead to failure of the procedure. INTRAOPERATIVE AND EARLY POSTOPERATIVE COMPLICATIONS CAN INCLUDE: Fracture of the implant; Fracture of bone or soft tissue damage; Dehiscence of the incision; Prominence or disfigurement at the implant site; and Infection. LATE POSTOPERATIVE COMPLICATIONS CAN INCLUDE: Fracture of the device due to traumatic injury; Loosening or migration due to loss of fixation or trauma; and Prominence or disfigurement over time at or near the implant site. INSPECTION All implantable devices are shipped NON-STERILE from Kelyniam Global Inc. in a sealed polyethylene bag to be repackaged at the hospital prior to steam sterilization. The seals along with the entire pouch should be inspected to ensure they have not been tampered with or damaged during shipping. All devices must be checked to include a twelve (12) digit alpha-numeric case number which matches the case number on the product label. Page 5 of 8

CLEANING This implant was produced in a non-sterile cleanroom environment and has been manually cleaned using a 70% alcohol solution. The implant MUST BE sterilized before use. STERILIZATION The Kelyniam Customized Cranial/Craniofacial Implants are supplied clean but non-sterile and should be repackaged before steam sterilization. Prior to sterilization, the Kelyniam Customized Implant together with an FDA approved biological Indicator should be double pouched in FDA approved steam sterilization pouches. All materials used should be applicable to the method of sterilization to be performed. The recommended parameters listed below are based on Laboratory testing conducted to establish sterility. These are in accordance with ANSI/AAMI ST79 - Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities RECOMMENDATIONS FOR STEAM STERILIZATION Minimum Exposure Method Cycle Time Minimum Exposure Temperature Drying Steam Gravity Displacement 10 minutes 132 C (270 F) 30 minutes Steam Pre-vacuum 4 minutes 132 C (270 F) 30 minutes Figure 7 - Sterilization Recommendations STORAGE Kelyniam Customized Cranial/Craniofacial Implants should be stored in a controlled environment at approximately 24 C (75 F) in accordance with AAMI ST79. DO NOT USE IF THE PACKAGE IS OPEN OR DAMAGED Following sterilization, the package should again be inspected to ensure the integrity of the packaging was maintained. Product not scheduled for immediate use should be stored in a cool dry place. MAGNETIC RESONANCE IMAGING The Kelyniam Customized Cranial/Craniofacial Implants described in this Instruction For Use (IFU) have not been evaluated for safety and compatibility in the MR environment by Kelyniam Global, Inc. These implants have not been tested for heating or migration in the MR environment by Kelyniam Global, Inc. **Kelyniam Global, Inc. shall not be liable for any data supplied by any third party.** CAUTION Federal (USA) law restrictions require these devices be sold by or on the order of a licensed physician. Page 6 of 8

SYMBOLS CONTAINED IN DEVICE LABELING REF Part Number Expiry Date Non-sterile LOT Part Number S/N Part Number Single Patient Use Prescription Manufactured By Instructions For Use Figure 7 - Symbols Referenced On Device Labeling. NOTES Page 7 of 8

INFORMATION For further information, please contact Customer Service at: Kelyniam Global Inc. 97 River Road, Suite A Canton, Connecticut 06019 Toll Free: (800) 280-8192 Fax: (501) 641-2000 Email: info@kelyniam.com URL: http:\\www.kelyniam.com Proudly Designed & Manufactured in Canton, Connecticut, USA Page 8 of 8