Intracardiac echocardiography: an ideal guiding tool for device closure of interatrial communications

Similar documents
Experience with Intracardiac Echocardiography for Guiding Transcatheter Device Closure of Atrial Septal Defects

Original Article Safe and effective guidance by intracardiac echocardiography for transcatheter closure in atrial septal defects

Contemporary echocardiographic guiding tools for device closure of interatrial communications

Patent Foramen Ovale Closure Without Echocardiographic Control: Use of Standby Intracardiac Ultrasound

Original Article Safe and efectivef guidance by intracardiac echocardiography for transcatheter closure in atrial septal defects

Fabien Praz, Andreas Wahl, Sophie Beney, Stephan Windecker, Heinrich P. Mattle*, Bernhard Meier

Atrial Septal Defect Closure. Stephen Brecker Director, Cardiac Catheterisation Labs

The earliest application of intracardiac imaging

Rahul Jhaveri, M.D. The Heart Group of Lancaster General Health

Feasibility of real-time three-dimensional transoesophageal echocardiography for guidance of percutaneous atrial septal defect closure

Eight-French Intracardiac Echocardiography

Anaesthesia for percutaneous closure of atrial septal defects Patrick A Calvert BCh MA MRCP Andrew A Klein MBBS FRCA

Percutaneous VSD closure

6). However, for cases in which an ASO cannot

Percutaneous occluder device closure through femoral vein guidance by transthoracic echocardiography in adult atrial septal defect patients

Transcatheter Closure of Atrial Septal Defects with Superior-anterior Rim Deficiency Using Amplatzer Septal Occluder

Echocardiographic Guidance During Placement of the Buttoned Double-Disk Device for Atrial Septa1 Defect Closure

Transcatheter closure of atrial septal defects with transthoracic echocardiography

Transcatheter closure of patent foramen ovale using the internal jugular venous approach

Hybrid Muscular VSD Closure in Small Weight Children

2D/3D in Evaluation of Atrial Septum

INTEGRATING ECHOCARDIOGRAPHY WITH CATHETER INTERVENTIONS FOR CONGENITAL HEART DISEASE. Krishna Kumar SevenHills Hospital, Mumbai, India

Transcatheter Closure of Secundum Atrial Septal Defect in Patients Over 60 Years Old

Γεώργιος Δ. Κατσιμαγκλής. Αν. Διευθυντής Καρδιολογικής ΚλινικήςΝΝΑ Διευθυντής Αιμοδυναμικού Εργαστηρίου ΝΝΑ

PFO (Patent Foramen Ovale): Smoking Gun or an Innocent Bystander?

Stroke and ASA / FO REBUTTAL

ATRIAL SEPTAL CLOSURE AND LEFT ATRIAL APPENDAGE OCCLUSION: INDICATIONS AND GUIDANCE ECHOCARDIOGRAPHY IN INTERVENTIONAL CARDIOLOGY

Transcatheter Closure of Secundum Atrial Septal Defects in the Elderly

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Device Closure of ASD in Children Using the Amplatzer Devices: Indications and catheter preparations, technique, and outcome [15]

Percutaneous atrial septal defect closure with the Occlutech Figulla Flex ASD Occluder.

IMAGES. in PAEDIATRIC CARDIOLOGY. Abstract

PERCUTANEOUS CLOSURE OF PATENT FORAMEN OVALE AND ATRIAL SEPTAL DEFECT: STATE OF THE ART AND A CRITICAL APPRAISAL

Role of Intracardiac Echocardiography (ICE) in Transcatheter Occlusion of Atrial Septal Defects

Atrial Septal Defects

Multidetector computed tomography in the evaluation of atrial septal defects

Index. cardiology.theclinics.com. Note: Page numbers of article titles are in boldface type.

Early and Late Complications Associated With Transcatheter Occlusion of Secundum Atrial Septal Defect

Jian Fang 1, Shaobo Xie 2, Lunchao Ma 2, Chao Yang 2. Original Article

Multifenestrated Septal Occluder Cribriform

RESPECT Safety Findings

Transcatheter aortic valve replacement with intracardiac echocardiography from the right internal jugular vein

BASIL D. THANOPOULOS MD, PhD Associate Professor Honorary Consultant, RBH, London, UK

CONGENITAL HEART DEFECTS IN ADULTS

Surgical Approach To An Unsuccessful Device Closure of Atrial Septal Defect. U Yetkin, B Özpak,? Yürekli, K Ergüne?, N Eren, O Ergene, A Gürbüz

For Personal Use. Copyright HMP 2013

Transcatheter Closure of Atrial Septal Defect in Young Children Results and Follow-Up

Paediatrica Indonesiana

AMPLATZER Septal Occluder Structural Heart Therapy. Over 15 years of. Demonstrated. Clinical Experience. We ll show you our data. Ask to see theirs.

Patent foramen ovale (PFO) is composed of

National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5

ΔΙΑΧΕΙΡΙΣΗ ΑΣΘΕΝΩΝ ΜΕ ΜΕΣΟΚΟΛΠΙΚΗ ΕΠΙΚΟΙΝΩΝΙΑ ΖΑΧΑΡΑΚΗ ΑΓΓΕΛΙΚΗ ΚΑΡΔΙΟΛΟΓΟΣ ΗΡΑΚΛΕΙΟ - ΚΡΗΤΗ

Case Report Hemostasis of Left Atrial Appendage Bleed With Lariat Device

T his article provides a review of the interventional

Corporate Medical Policy

Type Size AP/PS RP/RS Qp/Qs. Ia Resistive <0.3 < Ib Resistive <0.3 < IIa Resistive <0.5 >2

Clinical Commissioning Policy Statement: Patent Foramen Ovale (PFO) Closure. April Reference: NHSCB/A09/PS/a

Original Article Atrial Septal Defect (ASD) Pak Armed Forces Med J 2014; 64 (2):

Patent Foramen Ovale: Diagnosis and Treatment

INSTRUCTIONS FOR USE FOR:

GERIATRICS CASE PRESENTATION

Breakout Session: Transesophageal Echocardiography

For Personal Use. Copyright HMP 2013

Non-surgical extraction of right atrial mass by AngioVac aspiration device under fluoroscopic and transesophageal echocardiographic guidance

TRANSCATHETER CLOSURE OF ATRIAL SEPTAL DEFECT AND PFO

Transcatheter Closure of Atrial Septal Defects in a Center With Limited Resources: Outcomes and Short Term Follow-Up

Fermeture percutanée des CIA : matériels disponibles et résultats. Transcatheter closure of ASD : devices and results

Assessment of Atrial Septal Defect Role of Real-Time 3D Color Doppler Echocardiography for Interventional Catheterization

Percutaneous Mitral Valve Repair. Transesophageal Echo Acquisition Guide

Cryptogenic Stroke/PFO with Thrombophilia and VTE: Do We Know What To Do?

Transcatheter Closure of Cardiovascular Defects

Therapeutic Cardiac Catheterizations for Children with Congenital Heart Disease

AMPLATZER Septal Occluder

A challenging case of successful ASD closure without echocardiographic guidance in an 86-year old with severe kyphoscoliosis and platypnoeaorthodeoxia

Glenmark Cardiac Centre Mumbai, India

Treatment of Inter-MitraClip Regurgitation Due to Posterior Leaflet Cleft by Use of The Amplatzer Vascular Plug II device

Trans-septal Catheterization. December 8, Jonathan Tobis, MD Professor of Medicine Interventional Cardiology, UCLA

I, (Issam Moussa) DO NOT have a financial interest/arrangement t/ t or affiliation with one or more organizations that could be perceived as a real

Echocardiography Conference

Transcatheter closure of interatrial

Welcome Number 6. The background of the Cardiac Sonographer Network News masthead is a diagnostic image:

Transcatheter Closure of Secundum Atrial Septal Defects: Results in Patients with Large and Extreme Defects

UNMH Pediatric Cardiology Clinical Privileges. Name: Effective Dates: From To

Implantation of Cardioverter Defibrillator After Percutaneous Closure of Atrial Septal Defect

THINK OUTSIDE THE PILLBOX

Transcatheter Closure of Septal Defects

Pathology of Atrial Septal Defects. To Close or not to Close?: Atrial Septal Defects are NOT That Simple-Sometimes

JOINT MEETING 2 Tricuspid club Chairpersons: G. Athanassopoulos, A. Avgeropoulou, M. Khoury, G. Stavridis

Septal Occluder and Delivery System

Why Treat Patent Forman Ovale

Assessing ASDs prior to device closure using 3D echocardiography. Just pretty pictures or a useful clinical tool?

Concomitant procedures using minimally access

Atrial Septostomy in HLHS and ECMO: Indications, Technique and Outcome

Atrial septal defects (ASDs) are one of the most

Percutaneous Retrieval of Embolized Amplatzer Septal Occluder after Treatment of Double Atrial Septal Defect: A Case Report

Intracardiac EchoCardiography (ICE) Common Views

Description. Page: 1 of 23. Closure Devices for Patent Foramen Ovale and Atrial Septal Defects. Last Review Status/Date: December 2014

Nit-Occlud. Coil System for PDA Closure IMPLANTATION POCKET GUIDE. Rx only CV / B. Braun Interventional Systems Inc.

Transcription:

Eur J Echocardiography (2005) 6, 92e96 Intracardiac echocardiography: an ideal guiding tool for device closure of interatrial communications Thomas Bartel a, *, Thomas Konorza a, Ulrich Neudorf b, Tiko Ebralize a, Holger Eggebrecht a, Achim Gutersohn a, Raimund Erbel a a Cardiology Division, West-German Heart Center, University Clinic Essen, Hufelandstrasse 55, 45122 Essen, Germany b Cardiology Division, Department of Paediatrics, University Duisburg, Essen, Germany Received 12 December 2003; received in revised form 22 July 2004; accepted 27 July 2004 Available online 17 September 2004 KEYWORDS Intracardiac echo; Interatrial communications; Device closure Abstract Background This study sought to evaluate safety and radiation exposure when using intracardiac echocardiography (ICE) in comparison to transesophageal echocardiography (TEE) in order to guide transcatheter closure of interatrial communications. Methods Eighty patients (44 males, 36 females, mean age 46, SD 13 years) undergoing device closure of atrial septal defect (n Z 12) or patent foramen ovale (n Z 68) had the procedure guided by ICE (n Z 50, group 1) or TEE (n Z 30, group 2). In group 1, all procedural stages were completely guided by ICE, including imaging of the interatrial communication during balloon sizing, device unfolding and release, and during the final check for adequate positioning. In group 2, exclusive implantation of devices was guided by use of TEE. Results Especially, the spatial relationship between device and cardiac structures (e.g. the ascending aorta, the interatrial septum and the superior vena cava) was accurately demonstrated in group 1. Image resolution provided by ICE was superior to that of TEE. No severe complications, including any related to ICE, were seen. Fluoroscopy time (FT) and procedure time (PT) were shorter in group 1 than in group 2 (FT: 5.5G 1.5 min vs. 9.3 G 1.6 min, P! 0.0001; PT: 31.9 G 4.6 min vs. 38.8 G 5.8 min, P! 0.01). Neither sedation nor anesthesia was required in group 1. Conclusions ICE is a safe tool to guide device closure of interatrial communications. For the patient, procedural stress and radiation exposure are negligible. * Corresponding author. Tel.: C49 201 723 4805; fax: C49 201 723 5401. E-mail address: thomas.bartel@uni-essen.de (T. Bartel). 1525-2167/$30 ª 2004 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved. doi:10.1016/j.euje.2004.07.007

Intracardiac echocardiography: ideal guiding tool 93 ICE can be considered the guiding tool of choice for device closure, particularly when long or repeated echocardiographic viewing is required. ª 2004 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved. Introduction During the last decade, device closure of interatrial communication has become a well established technique in order to prevent paradoxical embolism in patients with patent foramen ovale (PFO) and to adequately treat severe left-to-right shunt associated with atrial septal defect (ASD). 1,2 Specific severe complications of transcatheter closure, the rate of which was reported to range between 1 and 4% depending on the device and the experience of the center, 2e5 can possibly be avoided by improved echocardiographic monitoring of interventional device closure. Transesophageal echocardiography (TEE) is considered the standard approach to guide those interventions, 6 but supine patients do not tolerate the transesophageal probe well, so that most centers use more or less extensive fluoroscopic imaging when performing device closure, although fluoroscopy cannot sufficiently depict the spatial relation between occluder device and cardiac structures next to the interatrial communication. Rapidly growing clots on the device are not shown either. The aim of the present investigation was to evaluate if intracardiac echocardiography (ICE) is superior to TEE as a safe guiding tool for transcatheter closure of interatrial communications. Methods A total of 80 patients (44 male, 36 female, mean age 46, SD 13 years) with PFO (n Z 68) or ASD (n Z 12) were randomly selected for either continuous guidance by ICE ( group 1; n Z 50) or TEE guidance ( group 2; n Z 30). An 11 F sheath was required to introduce the 10.5 F AcuNav-catheter (AcusoneSiemens Inc.) for ICE. After venous puncture and introduction of an 11 F access sheath into the left femoral vein, the catheter was advanced into the right atrium and the probe aimed at the interatrial septum (Fig. 1). Two standardised views were used to get a sufficient overview of the complete defect including its rims and its spacial relation to the aorta and the superior vena cava as well as the right-sided pulmonary veins: longitudinal view showing the extent of the interatrial septum (IAS) from cranial to its distal margins and a perpendicular short-axis view to visualise the anterior part of the IAS and the transition to the ascending aorta. 7 Group 2 patients were sedated with 5 mg of midazolam and continuous intravenous administration of propofol at 5e10 mg/h. Patients who did not tolerate the probe under this regimen were intubated. A Sequoia 256 ultrasound unit (AcusoneSiemens Inc.) was employed for ICE and TEE. Amplatzer and Starflex occluder systems (AGA Medical Corp. and NMT Medical Inc.) were implanted as previously described. 8 In group 1 patients, the procedure was primarily guided by ICE and was supplemented by fluoroscopy, whereas in group 2, the procedure was primarily guided by fluoroscopy and additionally monitored by TEE. Fluoroscopic time (FT), procedure time (PT) and image resolution were comparatively assessed. Parametric data were given as mean G 1 SD and tested with the unpaired 2-tailed Student s t test for group distinction. Nonparametric data were tested employing the c 2 test with one degree of freedom. A value P! 0.05 was considered significant. Results Age and sex ratio were not different between both groups. Group 1 included 8 ASDs and 42 PFOs and group 2 included 4 ASDs and 26 PFOs. In group 1, ICE depicted one early thrombus growing on the occluder device. Immediate treatment with Figure 1 The AcuNav-catheter is introduced into the right atrium. AC Z AcuNac-catheter, IAS Z interatrial septum, RA Z right atrium, SVC Z superior vena cava.

94 T. Bartel et al. 10,000 U of heparin became therefore possible. Further complications, especially cerebral embolization could be prevented in this patient. No severe complications, including any related to ICE, were seen. In group 2, 4 patients did not tolerate TEE and had to be intubated. The other 26 individuals tolerated the TEE probe for 3e18 min during device implantation. Neither sedation nor general anesthesia was required in any group 1 patient, and all procedural stages were completely guided by ICE (Figs. 2 and 3), including imaging of the interatrial communication during balloon sizing, device unfolding and release, and during the final check for adequate positioning. Other procedural steps were partially guided, e.g. catheter passage of the communication, placement of a wire into the left upper pulmonary vein, and placement of the long sheath with its distal opening inside the left atrium (Fig. 4). Especially, the spatial relationship between device and cardiac structures (i.e. the ascending aorta, the IAS and the superior vena cava) was accurately demonstrated. In addition, it was shown that the countercluders of the device had really engaged the IAS (Fig. 5). FT and PT were found to be shorter in group 1 than in group 2 (FT: 5.5 G 1.5 min vs. 9.3 G 1.6 min, P! 0.0001; PT: 31.9 G 4.6 min vs. 38.8 G 5.8 min, P! 0.01). Compared to TEE, ICE reduced patient stress and improved compliance. With respect to the image resolution, particular details, e.g. the guide wire inside the long sheath or the membrane inside the countercluders of an Amplatzer closure device, could be clearly depicted by ICE, but not by TEE. All patients underwent a follow-up TEE 6 months after device Figure 3 Balloon sizing of an atrial septal defect. LA Z left atrium, RA Z right atrium, SB Z sizingballoon. closure. There was no dislocation of a device. In all cases, the septum-primum parts as well as the septum-secundum parts of the IAS were correctly engaged in between the left-sided and the rightsided umbrellas. In both groups, no residual shunt could be detected by use of colour Doppler. After administration of non-transpulmonary contrast agent, 3 patients of group 1 and 4 patients of group 2 showed slight right-to-left shunt. There was no significant difference between both groups (P O 0.05). Discussion The present results demonstrate ICE monitoring in transcatheter device closure to be superior to Figure 2 Atrial septal defect with characteristic leftto-right shunt. ASD Z atrial septal defect and left-toright shunt, LA Z left atrium, RA Z right atrium. Figure 4 The long sheath does not interfere with the left atrial wall. LA Z left atrium, LS Z long sheath, RA Z right atrium.

Intracardiac echocardiography: ideal guiding tool 95 Figure 5 Occluder device connected to a delivery cable is being pushed towards the left atrium in order to make sure that it cannot embolize into the systemic circulation. AV Z aortic valve, LA Z left atrium, D Z device, RA Z right atrium. conventional TEE. They confirm and broaden the messages of previous investigations 7,9,10 showing the suitability of ICE to safely guide interventional therapy in interatrial communications. TEE requires general anesthesia with or without endotracheal intubation, whereas ICE permits unlimited echocardiographic viewing in fully conscious and compliant patients. This is of utmost importance, since malposition and migration of the device into the systemic or pulmonary circulation as well as perforation of the cardiac wall and rapid thrombus formation on the device are known to occasionally occur. 1,3,4,11 This new echocardiographic approach depicts the individual morphology of interatrial communications and the instrumentation needed for the closure procedure in fullest detail. It provides a much better image resolution than TEE and therefore also facilitates the procedure for the interventional cardiologist, particularly when long continuous or repeated echocardiographic viewing is required or when complications begin to develop. ICE results in much lower procedural stress to the patient, and FT as well as PT can be shortened. It also facilitates and shortens balloon sizing of interatrial communications (Fig. 3). Increasing the safety of device implantation and reducing radiation exposure and procedural stress for the patient can be considered major advantages of intra-procedural monitoring by ICE. Employment of ICE would also be desirable in the pediatric population. Especially in young individuals, reduction of radiation exposure is of utmost importance. Beside that, the risk of oesophageal injury from TEE probe insertion must also be considered an unresolved problem in small children. First experiences show that ICE represents a feasible and easy approach to also guide device closure in children and adolescents. 10 As a result, it is likely that ICE improves the safety of interventional device closure in interatrial communications, so that this technique should replace TEE as a guiding tool not only in adults but also in adolescents and children. To our knowledge, no complications related to ICE have been reported up to now. Nevertheless, the potential risks related to ICE cannot be finally estimated at this time. Although the risk potential seems to be low, it is mandatory that the ICEcatheter is handled with caution, since it is not wire-guided what makes it generally different from conventional intravascular ultrasound catheters. Thus, fluoroscopy is recommended to safely introduce the steerable ICE-catheter into the right atrium. An even smaller version of the ICE-catheter would probably lead to more frequent use of ICE guidance. Although the expenses for the AcuNav-catheter remain an important shortcoming, certified resterilization of the catheter is now being offered by different companies for the European Community. This way, the costs for one single application of ICE lessens to less than 1000 V, what may also improve acceptance. Nevertheless, some limitations inherent to this technique must be taken into account. That the catheter is not wire-guided makes its handling different from conventional catheter procedures. Therefore, a training including an animal experimental part offered by the manufacturer is strongly recommended in order to employ this catheter safely and effectively. In addition to that, the setting of the ultrasound unit must be particularly adjusted to optimal profit from ICE-imaging of the interatrial septum, the left atrium, and the pulmonary veins. Thus, the interventionalist handling the ICE-catheter must also be experienced in echocardiography. If this is not sufficiently the case, he should consult a colleague for interpretation of the echocardiographic images and even for adjustment of the ICE-catheter. Although large venous punctures are not necessarily expected to cause major complications, we recommend 2e6 h observation. In conclusion, ICE can be considered a safe tool for guiding device closure of PFO and ASD. Supine patients tolerate ICE much better than TEE. ICE also reduces FT and PT and seems to be advantageous, especially when extended echocardiographic viewing is required. The

96 T. Bartel et al. advantages of ICE seem to justify the expense for the catheter. References 1. Braun MU, Fassbender D, Schoen SP, et al. Transcatheter closure of patent foramen ovale in patients with cerebral ischemia. J Am Coll Cardiol 2002;39:2019e25. 2. Du ZD, Hijazi ZM, Kleinman CS, Silverman NH, Larntz K, Amplatzer Investigators. Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults: results of a multicenter nonrandomized trial. J Am Coll Cardiol 2002;39:1836e44. 3. Chessa M, Carminati M, Butera G, et al. Early and late complications associated with transcatheter occlusion of secundum atrial septal defect. J Am Coll Cardiol 2002;39:1061e5. 4. Nkomo VT, Theuma P, Maniu VC, et al. Patent foramen ovale transcatheter closure device thrombosis. Mayo Clin Proc 2001;76:1057e61. 5. Pinto FF, Sousa L, Fragata J. Late cardiac tamponade after transcatheter closure of atrial septal defect with cardioseal device. Cardiol Young 2001;11:233e5. 6. Sievert H, Horvath K, Zadan E, et al. Patent foramen ovale closure in patients with transient ischemic attack/stroke. J Interv Cardiol 2001;14:261e6. 7. Bartel T, Konorza T, Arjumand J, et al. Intracardiac echocardiography is superior to conventional monitoring for guiding closure of interatrial communications. Circulation 2003;107:795e7. 8. Bruch L, Parsi A, Grad MO, et al. Transcatheter closure of interatrial communications for secondary prevention of paradoxical embolism: single-center experience. Circulation 2002;105:2845e8. 9. Mullen MJ, Dias BF, Walker F, Siu SC, Benson LN, McLaughlin RR. Intracardiac echocardiography guided device closure of atrial septal defects. J Am Coll Cardiol 2003; 41:285e92. 10. Koenig P, Cao QL, Heitschmidt M, Waight DJ, Hijazi ZM. Role of intracardiac echocardiographic guidance in transcatheter closure of atrial septal defects and patent foramen ovale using the Amplatzer device. J Interv Cardiol 2003;16:51e 62. 11. Martin F, Sánchez PL, Doherty E, et al. Percutaneous transcatheter closure of patent foramen ovale in patients with paradoxical embolism. Circulation 2002;106: 1121e6.