Evaluation of the Quality of Thick Fibre Composites Using Immersion and Air- Coupled Ultrasonic Techniques

Similar documents
Ultrasonic Phased Array Testing of Complex Aircraft Structures

Application of Time Reversal Technique for the Inspection of Composite Structures

Descriptions of NDT Projects Fall 2004 October 31, 2004

10 Years Experience in Industrial Phased Array Testing of Rolled Bars

Non-Destructive Inspection of Composite Wrapped Thick-Wall Cylinders

Ultrasonic Testing Level I:

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

Ultrasonic Testing of Composite Structures

Ultrasound Measurements and Non-destructive Testing Educational Laboratory

PORTABLE AND HYBRID LASER GENERATION / AIR-COUPLED DETECTION SYSTEM FOR NON DESTRUCTIVE INSPECTION

Improved Inspection of Composite Wind Turbine Blades with Accessible Advanced Ultrasonic Phased Array Technology

Pipeline Technology Conference 2007

w. D. Jolly, F. A. Bruton, and C. Fedor

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics

Real Time 2D Ultrasound Camera Imaging: A Higher Resolution Option to Phased Array Bob Lasser Randy Scheib Imperium, Inc.

INSPECTION THROUGH AN OVERLAY REPAIR WITH A SMART FLEXIBLE ARRAY PROBE.

Basic Physics of Ultrasound and Knobology

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

ADVANCED PHASED ARRAY TECHNOLOGIES

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

ACOUSTORAPHY Acousto-Optical (AO) Ultrasonics & Its Applications

Developments in Ultrasonic Inspection II

Phased Array Ultrasonics and Eddy Current Examination for Graphite Components

TTU Phased Array: Quality and Productivity

Ultrasonic Testing of Rails Using Phased Array

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

Visualization of CFRP Structures Based on Acoustic Techniques for NDT

COWLEY COLLEGE & Area Vocational Technical School

STUDY OF PHASED ARRAY TECHNIQUES FOR CONCRETE INSPECTION

Q1. The diagram shows an ultrasound monitor being used to scan a fetus.

COURSE DESCRIPTION FOR NONDESTRUCTIVE TESTING

API. Defined Procedure. for. Ultrasonic Thickness Measurement API-UT-21

Shcherbakov O.N., Petrov A.E., Polevoy A.G., Annenkov A.S. ULTES LLC, Moscow, Russia

Ultrasonic Inspection of Adhesive Joints of Composite Pipelines

DEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS

Optimization of Phased-Array Transducers for Ultrasonic Inspection in Composite Materials Using Sliding Probes

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1

NONDESTRUCTIVE EVALUATION OF PLATE TYPE NUCLEAR FUEL ELEMENTS DURING MANUFACTURING STAGE USING ULTRASONIC TEST METHOD

Ultrasonic Thickness Procedure Pulse-Echo Thickness Gage TABLE OF CONTENTS. 1.0 Introduction Reference Documents Summary of Practice 2

Flip Chips and Acoustic Micro Imaging: An Overview of Past Applications, Present Status, And Roadmap for the Future

CORRELATION OF TRANSDUCER FREQUENCY AND SIGNAL/NOISE RATIO OF THIN WALLED FILAMENT WOUND CFRP-TUBES INSPECTED BY ULTRASONICS

Field inspection of Phase-Array ultrasonic for polyethylene electrofusion joints

Flaw Assessment Using Shear wave Phased array Ultrasonic Transducer

Dr Emma Chung. Safety first - Physical principles for excellent imaging

Routine Quality Assurance Cookbook

Case studies on uncertainties of ultrasonic weld testing interpretation

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING

1. SCOPE ELIGIBILITY EXAMINATION CONTENT RENEWAL & RECERTIFICATION PROCEDURE ESSENTIAL READING...

Introduction to Biomedical Imaging

Ultrasonic Testing of Adhesively Bonded Joints Using Air-Coupled Cellular Polypropylene Transducers

T.J. Meitzler US Army, TARDEC Warren, MI J. S. Steckenrider Illinois College Jacksonville, IL. L. P. Franks US Army, TARDEC Warren, MI

Employer s Unit of Competence Ultrasonic testing of materials, products and plant

Preamble (disclaimer)

Lesson 03: Sound Wave Propagation and Reflection. This lesson contains 15 slides plus 14 multiple-choice questions.

Detection. at first sight. One of the newest trends

MOTION ANALYSIS OF A LOW COST CUSTOMIZED A-SCAN NON DESTRUCTIVE TESTING UNIT. M.N. M. Zukri and E. A. Bakar

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions.

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Improved ultrasound-based navigation for robotic drilling at the lateral skull base

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

Ultrasonic arrays are now widely used in underwater sonar

Imaging spot weld inspection using Phased Array technology new features and correlation to destructive testing

Ultrasound Physics & Doppler

Table 1: Samples used in experimental program. Block ID/thickness [ mm ]

A new method of sonograph lateral resolution measurement using PSF analysis of received signal

Research on Digital Testing System of Evaluating Characteristics for Ultrasonic Transducer

Reliability Analysis of the Phased-Array Ultrasonic System used for the Inspection of Friction Stir Welds of Copper Canisters

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound

How NDT will underpin the certification of greener aircraft

Ultrasound in Medicine

Radiant NDT Services

Feng Xiujuan National Institute of Metrology (NIM),China

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

AN EXPERIMENTAL STUDY ON TRANSDUCER FOR THE DDS TECHNOLOGY DEMONSTRATOR ANDRZEJ ELMINOWICZ, WALDEMAR LIS*, WŁADYSŁAW MĘCIŃSKI

Ultrasound Physics & Terminology

NDT ON CONCRETE STRUCTURES UNDER LOAD TESTS Hazard: Extreme Loads

SUITABILITY OF NON DESTRUCTIVE TECHNIQUES FOR TESTING POLYETHYLENE PIPE JOINTS. Peter J. Postma, René J.M. Hermkens, Kiwa Technology, The Netherlands

Linear Ultrasonic Wave Propagation in Biological Tissues

Underwater Acoustic Measurements in Megahertz Frequency Range.

Precise defect detection with sensor data fusion

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS

NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment

Ultrasound Principles cycle Frequency Wavelength Period Velocity

DEVELOPMENT OF ELECTROMAGNETIC ACOUSTIC TRANSDUCER (EMAT) PHASED ARRAYS FOR SFR INSPECTION

NON-CONTACT ULTRASOUND (NCU)

Active Thermography Imaging - an Innovative Approach for NDT Inspection

Numerical Modelling of Ultrasonic Phased Array Transducers and Their Application

Multielement ultrasonic probes for projection imaging of biological media

Hand Held Ultrasonic Sizing of Stress Corrosion Cracking

Manual Ultrasonic Inspection of Thin Metal Welds

CONTENTS. Test Number cpd Tanya Reynolds (Nat. Dip. Diag. Rad., B. Tech. Diag. Rad., B. Tech. Ultrasound)

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals

High-Power Locomotive Solid Axle Defect on-line Detection Technique

Performance of phased array and conventional ultrasonic probes on the new ISO reference block

TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound

Physical Principles of Ultrasound

Transcription:

ECNDT 2006 - We.1.6.4 Evaluation of the Quality of Thick Fibre Composites Using Immersion and Air- Coupled Ultrasonic Techniques Kaj K. BORUM, Risø National Laboratory, Materials Research Department, Roskilde, Denmark Abstract. Two ultrasonic through-transmission techniques were evaluated: immersion and air-coupled technique. The immersion system was used for frequencies from 1 MHz to 10 MHz. The frequency range for the air-coupled system was 120 khz to 400 KHz. The purpose was to find the limitations of the techniques and compare their practical use for evaluation of the quality of up to 50 mm thick glass-fibre laminates. Additional results were obtained using single-sided immersion pulse-echo technique The glass-fibre/polyester specimens were made by resin transfer moulding. The shape is slightly curved as they are used as springs. If something goes wrong during the manufacturing process air can be trapped in the laminate and cause an unacceptable content of porosity. Both through-transmission techniques can find critical defects although the immersion technique has the best resolution and sensitivity. However, the air-coupled technique is preferable to use because it does not require water coupling but only atmospheric air as coupling media. Compared to pulse-echo techniques the through-transmission technique is more robust and gives a better overall picture of the areas with higher porosity content. Introduction The use of ultrasonic scanning is well-know for evaluation of composite materials [Ref. 1 and 2]. The increased use of thick composites as construction materials is however a challenge to evaluate for the ultrasonic techniques. The traditional pulse-echo techniques that work well for thickness up to about 10 mm are not well-suited for thickness of 30 mm and more. The purpose of this paper is to present and compare results from two different ultrasonic techniques for examination of the material quality of glass fibre composite springs produced by a vacuum assisted Resin Transfer Moulding (RTM) technique. The two ultrasonic techniques are both through-transmission technique: one is based on immersion technique and the other is using air-coupled technique. The air-coupled technique has the advantage of only using atmospheric air as coupling media, and thereby makes scanning easier. The specimens used for this examination are cut-out parts from springs. Description of ultrasonic techniques Two independent scanning systems were used for the two through-transmission techniques: Immersion and Air-coupled techniques. Supplementary, immersion pulse-echo technique has been used for comparison. 1

The physical principle is the same for both through-transmission techniques. An ultrasonic pulse is transmitted through the specimen perpendicular to the surface using separate transmitter and receiver transducers opposite to each other. The signal obtained from the receiver transducer is a measure of the total ultrasonic damping in the material. The ultrasonic damping is influenced by defects in the material such as high level of porosity or delaminations. Pores will cause the ultrasound to be reflected due to the high difference between acoustic impedance in the laminate and in air. The porosity are normally much smaller that the focal diameter and cannot be detected one by one. However, the sum of a large number of pores results in higher total damping and thus a smaller transmitted signal. The ultrasonic equipment is capable of measuring the peak amplitude of the transmitted signal. The scans are performed in a rectangular raster pattern. In each point the amplitude of the transmitted ultrasonic signal is measured. The result of the scanning is presented in a colour or grey-tone plot where the colour scale is in db, which gives a high dynamic range. In order to locate the depths of the defects immersion pulse-echo has been used as a supplementary technique. Only one transducer is used in this technique. The transducer works in this case both as sender- and receiver transducer. In order to inspect the whole cross section the ultrasonic signal has to travel twice through the specimen down to the backside of the specimen and back to the upper surface. If there are defects (i.e. pores or delaminations) present somewhere in the cross section the ultrasonic wave will be reflected by the defects and will be measured by the transducer. For the thick laminates (50 mm) it is difficult to examine the whole cross-section this way, and only a small echo is received from the back surface. The through-transmission technique gives a better and more reliable picture of where defects are located in the x-y plane but gives no information about the depths of the defects. The surface condition can influence the measurement. The surface should be clean and smooth. The ultrasound is send into the specimen perpendicular to the surface. If the surface is curved the transmitted pulse is damped more. For large curved specimens the transducers need to be adjusted for different sections of the specimen. However, that was not necessary for the scans presented in this paper. Description of equipment for immersion technique Transducers in the range of 0.5-50 MHz are available for this technique and examinations showed that the transducer pair of 0.5 MHz transducers gave the best results. Using higher frequencies gives too high damping and thus too small transmitted signal. If the surface is curved the sound beam will not be perpendicular to the surface and therefore the echo amplitude will be reduces. The sensitivity will normally also be reduced. The scanning resolution need to be sufficiently good to see details and it must be smaller than the beam diameter of the ultrasonic wave. For the scans made for this evaluation a resolution of 1 mm in both X- and Y direction was used. The scanning speed was 50 mm/s. 2

Figure 1. Photo of the immersion through-transmission setup. In the water tank the two transducers are mounted in a fork-fixture, which secure that the transducers are aligned correctly. Description of equipment for air-coupled technique The air-coupled through-transmission has the big advantage of only using atmospheric air as coupling media. The scans made on spring parts were made using two 120 khz transducers. Additional scans were made using two 400 khz transducers but the 120 KHz transducers gave better results. Due to the lower frequency and the air coupling the resolution is lower for the air-coupled technique. The scanner consists of two motor driven guides with a scanning area up to 1 x 1.5 m 2. The scanning speed is limited by the pulse repetition frequency, which is dependant on the distance between the transducer and the specimen. The longer distance the lower pulse frequency. Each pulse has to die out before a new pulse is sent into the specimen. Due to the much lower ultrasonic speed in air (340 m/s) compared to water (1500 m/s) the pulse repetition frequency - and thus the scanning speed - is much lower for air-coupled technique. The typical speed was app. 20 mm/s. The resolution was 1 mm in both X- and Y direction as for the immersion technique. If the purpose is just to find major defects in the specimen the resolution can be decreased and thus increasing the speed proportional. Figure 2. Photo of air-coupled techniques showing the scanner fixture and the transducer set-up. 3

Results and discussion Results from 4 specimens are included. These specimens were cut-out parts from large springs. All specimens were made of glass fibre reinforced polyester and manufactured by a vacuum assisted resin transfer technology. Results from scans are shown in figure 3-7. Generally, the immersion technique gives the best resolution and the sharpest picture. The contrast is also higher for immersion technique. However, the air-coupled technique has the important advantage of not using any coupling media other than air. The specimen holders can be seen in the plots of all the air-coupled scans whereas they are not visible on the immersion scan plots. The plots are in db where each grey-tone represents echo amplitude. Due to the nature of the measurements it is not possible to make measurements close to the edges. The size of the area around the edges where no valid measurements can be made depends on the transducer characteristics the higher frequency the smaller border. The borders are therefore larger for air-coupled technique than for the immersion technique. Figure 3. Specimen A. Dimension: 48 x 105 x 300 mm. Left picture: Immersion technique, right picture: Aircoupled technique. Both techniques detect a defect in the right side of the specimen but the contract is lower for the air-coupled technique. Figure 4. Specimen B. Dimension: 48 x 105 x 300 mm. Left: Immersion technique. Right: Air-coupled technique. Both techniques detect a defect in the middle of the specimen but the contract is lower for aircoupled technique. In Figure 5 the echo line profiles taken at the black vertical lines are shown. 4

Defect area Defect area Figure 5. Specimen B. Echo line profiles from the scans shown in Figure 4. Left: Immersion technique. Right: Air-coupled technique. The echo dynamic is higher for immersion technique than for air-coupled technique. For immersion technique the difference between area with defects and area with no defects is app. 11 db. For air-coupled technique the difference is app. 7 db. The scan plots for specimen A and B in Figure 3 and 4 both reveal a defect. Both throughtransmission techniques can detect them although the immersion technique gives the clearest indications. Figure 5 shows line profiles from the scans in Figure 4. The line profiles from the air-coupled technique is more scattered than for immersion technique and the sensitivity is larger for immersion technique. In Figure 6 the results from scanning of specimen C are show. As for specimen B the defects are detected more clearly and with a higher contrast using immersion technique. Figure 6. Specimen C Dimension: 48 x 105 x 300 mm. Left: Immersion technique. Right: Air-coupled technique. Both techniques detect the defect in the left part of the specimen but the contract is lower for aircoupled technique. In the right side of the specimen no signal is transmitted in immersion technique. In Figure 7 results from scanning of specimen D are show. The immersion technique detects a defect in the middle of the specimen, but this is not detected by the air-coupled technique. The reason is believed to be that the defect is too small compared to the diameter of the transducer. Although the defect also damp the ultrasonic wave generated by the aircoupled transducer the effect is not large enough to be detected by the receiver transducer. The main part of the sound beam is travelling outside the defect area. 5

Figure 7. Specimen D. Dimension: 48 x 105 x 300 mm. Left: Immersion technique. Right: Air-coupled technique. The defect seen in centre of the left scan is not visible on the right scan. This scanning indicates the limit for the air-coupled technique for these thick panels. The depth to the defects shown in the previous figures is not directly found by any of the through-transmission techniques. However, the depth can be found by pulse-echo technique where the time-of-flight from the surface to the defect echo is measured. In figure 8 the pulse-echo scan of specimen C is shown. From the time of flight measurement the depth of the defects were found to be in the range 20-29 mm from the surface, which is in the middle part of the cross section, which is typical for the RTM process if it is out of control. The main result of this evaluation is that it is possible to see the major defects with both the immersion technique and the air-coupled technique. Although there are differences in the details of the findings, the major defect areas can be found also with air-coupled technique. Figure 8. Pulse-echo measurement of specimen C. The scan plot shows the reflections from the defects. From the time-of-flight measurement the depth of the defects was found to be in the middle of the cross section. Some of the panels were examined by microscopy in order to verify the ultrasonic findings. Figure 9 shows a typical microstructure in an area with high ultrasonic damping. There are porosity both between bundles of Rovings and inside Rovings. 6

Porosity between Roving bundles Polyester Porosity inside a Roving Fibre Knitting threads (PET) Figure 9. Typical microstructure of material with many and larger imperfections, which cause a high damping of the ultrasonic signal. Conclusions The purpose of this investigation was to compare the two ultrasonic through-transmission techniques immersion and air-coupled technique for use on thick glass-fibre laminates (up to 50 mm thickness). Both techniques could find the major defects in the specimens, but two minor defects found by immersion technique could only barely be found by air-coupled technique. The immersion technique has a better resolution due to the higher frequency and the smaller focal spot of the immersion transducers. The air-coupled transducers have a larger focal beam and are therefore less sensitive to small defects that are smaller than the focal diameter. It can be seen from the scans that there is a clear edge effect where no valid measurement can be performed. The size of the border depends on the ultrasonic transducer. The higher frequency the smaller the border is. It has been shown that the air-coupled technique is sufficient sensitive to find critical defects in thick laminates such as the train bottom springs. The advantage of using the aircoupled technique is that no coupling media is needed which make the scanning much more simple and easier to perform. Acknowledgement The conducted work was a part of EUREKA project EU 1841 (EUROBOGIE), which was partly financed by the Danish Ministry of Science, Technology and Innovation. The EUROBOGIE project was managed by Sciotech (UK) and the spring material for this investigation was manufactured by EM Fiberglas A/S (DK). 7

References [1] W. Hillger, Inspection of CFRP components by ultrasonic imaging with air coupling, 8 th European Conference on Non-destructive Testing, Barcelona 2002, Conference Proc. on CD-ROM. [2] W. Hillger, Ultrasonic Testing of Composites From laboratory Research to In-field Inspections, 15 th World Conference on Non-destructive Testing, Roma, 15-21 October 2000, Conference proc. on CD-ROM. 8