Clinical Study Accuracy of Implant Placement Utilizing Customized Patient Instrumentation in Total Knee Arthroplasty

Similar documents
Computer Navigation in TKA The role of Robotic Surgery. Christos Yiannakopoulos, M.D., Ph.D.

PROPHECY INBONE. Preoperative Navigation Guides

Improving tibial component coronal alignment using clearly defined anatomical structures in total knee arthroplasty

CUTTING GUIDE SYSTEM PRODUCT RATIONALE. Custom-Made Patient Instruments for Total Knee Replacement

Ye-Ran Li, Yu-Hang Gao, Xin Qi *, Jian-Guo Liu, Lu Ding, Chen Yang, Zheng Zhang and Shu-Qiang Li

Presented By Dr Vincent VG An MD BSc (Adv) MPhil Dr Murilo Leie MD Mr Joshua Twiggs BEng Dr Brett A Fritsch MBBS FRACS (Orth) FAOrthA.

A Non-CT Based Total Knee Arthroplasty System Featuring Complete Soft-Tissue Balancing

Robotic-Arm Assisted Total Knee Arthroplasty Demonstrated Greater Accuracy to Plan Compared to Manual Technique

INVISION Total Ankle Replacement System with PROPHECY Preoperative Navigation Revision of a Failed Agility Total Ankle Replacement

Accelerometer-Based, Portable Navigation vs Imageless, Large-Console Computer-Assisted Navigation in Total Knee Arthroplasty

Personalized Solutions. Portfolio Brochure

CLINICAL PAPER / ORTHOPEDIC

Kinematic vs. mechanical alignment: What is the difference?

COMPUTER ASSISTED ROBOTIC TOTAL KNEE ARTHROPLASTY

Lateralization of Tibial Plateau Reference Point Improves Accuracy of Tibial Resection in Total Knee Arthroplasty in Patients with Proximal Tibia Vara

The Journal of Arthroplasty

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Masterclass. Tips and tricks for a successful outcome. E. Verhaven, M. Thaeter. September 15th, 2012, Brussels

VARIABILITY OF THE POSTERIOR CONDYLAR ANGLE

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

No radiographic difference between patient-specific guiding and conventional Oxford UKA surgery

TOTAL KNEE ARTHROPLASTY (TKA)

Case Report Navigation-Assisted Total Knee Arthroplasty for Osteoarthritis with Extra-Articular Femoral Deformity and/or Retained Hardware

RESECTION GUIDE SYSTEM. TRUMATCH Personalized Solutions Surgical Technique with ATTUNE Knee INTUITION Instruments

MEDICAL POLICY SUBJECT: COMPUTER ASSISTED NAVIGATION FOR KNEE AND HIP ARTHROPLASTY

Comparison of high-flex and conventional implants for bilateral total knee arthroplasty

In Vitro Analysis of! Foot and Ankle Kinematics:! Robotic Gait Simulation. William R. Ledoux

Introduction of FIREFLY Technology

ORTHOPAEDICS. Component position alignment with patient-specific jigs in total knee arthroplasty. Introduction. ANZJSurg.com

MEDICAL POLICY SUBJECT: COMPUTER ASSISTED NAVIGATION FOR KNEE AND HIP ARTHROPLASTY

Original Article Accuracy and efficacy of osteotomy in total knee arthroplasty with patient-specific navigational template

itotal Customized POSTERIOR-STABILIZED knee replacement system

STIFFNESS AFTER TKA PRE, PER AND POST OPERATIVE CAUSING FACTORS

Stiffness After Total Knee Arthroplasty: Does Component Alignment Differ in Knees Requiring Manipulation?

Accuracy of CT-based measurements of glenoid version for total shoulder arthroplasty

Case Report Bilateral Distal Femoral Nailing in a Rare Symmetrical Periprosthetic Knee Fracture

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

Total knee replacement in patients with significant femoral bowing in the coronal plane

HipNav: Pre-operative Planning and Intra-operative Navigational Guidance for Acetabular Implant Placement in Total Hip Replacement Surgery

Pin Guide System. Surgical Technique

CONTRIBUTING SURGEON. Barry Waldman, MD Director, Center for Joint Preservation and Replacement Sinai Hospital of Baltimore Baltimore, MD

SCIENTIFIC POSTER #28 Tibia OTA 2016

How coronal alignment affects distal femoral anatomy: an MRI-based comparison of varus and valgus knees

Innovations 2017 & 2018

Alignment of the femoral component in a mobile-bearing unicompartmental knee arthroplasty Kort, N. P.; van Raay, J. J. A. M.; Thomassen, B. J. W.

Corporate Medical Policy

Dora Street, Hurstville 160 Belmore Road, Randwick

TKA Gap Planning. Supporting healthcare professionals

Tibial Base Design Factors Affecting Tibial Coverage After Total Knee Arthroplasty: Symmetric Versus Asymmetric Bases

ConforMIS, Inc. 28 Crosby Drive Bedford, MA Phone: Fax:

Accuracy of bone resection in total knee arthroplasty using CT assisted-3d printed patient specific cutting guides

PIN GUIDE SYSTEM SURGICAL TECHNIQUE. with the SIGMA High Performance Instruments System. This publication is not intended for distribution in the USA.

PATIENT MATCHED TECHNOLOGY IN KNEE REPLACEMENT

The Bodycad Unicompartmental Knee System THE PURSUIT OF ORTHOPAEDIC PERFECTION

PPS P I N P O S I T I O N I N G S Y S T E M GMK EFFICIENCY VERSION. Hip Knee Spine Navigation

Sasaki E 1,2, Otsuka H 2, Sasaki N 2, and Ishibashi Y 1

BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS IN TOTAL KNEE REPLACEMENTS

A Kinematic Assessment of Knee Prosthesis from Fluoroscopy Images

Signature Personalized Patient Care*

Axial alignment of the lower extremity in Chinese adults. Journal Of Bone And Joint Surgery - Series A, 2000, v. 82 n. 11, p.

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

ANATOMIC. Navigated Surgical Technique 4 in 1 TO.G.GB.016/1.0

W. (2016) : - 98 (8) ISSN X,

Multiapical Deformities p. 97 Osteotomy Concepts and Frontal Plane Realignment p. 99 Angulation Correction Axis (ACA) p. 99 Bisector Lines p.

Robotic-arm-assisted Arthroplasty Advances Surgical Precision NEWYORK-PRESBYTERIAN QUEENS. Long Island

Variations of the grand-piano sign during total knee replacement

Calculating position of joint line of knee using various radiological parameters based on Indian population

Institutional review board approval was obtained prior to the start of this study.

)100( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY

THE KNEE SOCIETY VIRTUAL FELLOWSHIP

The Journal of Arthroplasty Vol. 22 No

Esperienza e primi risultati nei sistemi protesici custom made. Dott. Fernando Marcucci L. Buonocore, M.Greggi, E.D Antonio, N.

Abramsohn Retractor 1

)116( COPYRIGHT 2016 BY THE ARCHIVES OF BONE AND JOINT SURGERY

Case Report. Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology

LAMINA SPREADER SURGICAL TECHNIQUE

Reliability of patient specific instrumentation in total knee arthroplasty

Preliminary Outcome Using a New Free Motion Offloading KAFO for Postoperative Management of Hemiepiphysiodesis in Adolescent Tibia Vara

Kinematics Analysis of Different Types of Prosthesis in Total Knee Arthroplasty with a Navigation System

DePuy International Ltd St Anthony s Road Leeds LS11 8DT England Tel: +44 (0) Fax: +44 (0)

Patient-specific bicompart mental knee resurfacing system

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

High flexion knee arthroplasty: the relationship between rotational angles and flexion angle after total knee arthroplasty

Where Is the Natural Internal-External Rotation Axis of the Tibia?

Intramedullary Tibial Preparation

Computed Tomographic measurement of distal femor rotation in Iranian population

Predicting the Position of the Femoral Head Center

Total knee arthroplasty using computer-assisted navigation in patients with deformities of the femur and tibia: A report of 5 cases

)289( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY RESEARCH ARTICLE

BRAINLAB ORTHOPEDICS SOFTWARE SOLUTIONS TO IMPROVE PATIENT CARE

Case Report Reverse Segond Fracture Associated with Anteromedial Tibial Rim and Tibial Attachment of Anterior Cruciate Ligament Avulsion Fractures

Edinburgh Research Explorer

In-vivo Evaluation Of The Kinematic Behavior Of An Artificial Medial Meniscus Implant: A Pilot Study Using Open-mri

2017 Resident Advanced Trauma Techniques Course COMPLICATIONS / CHALLENGES MALUNIONS/DEFORMITY

Correspondence should be addressed to Thomas Kurien;

Mark Clatworthy Middlemore Hospital Auckland New Zealand

A Novel Free Form Femoral Cutting Guide

Distal Cut First Femoral Preparation

Transcription:

Hindawi Publishing Corporation Advances in Orthopedics Volume 2013, Article ID 891210, 6 pages http://dx.doi.org/10.1155/2013/891210 Clinical Study Accuracy of Implant Placement Utilizing Customized Patient Instrumentation in Total Knee Arthroplasty William D. Bugbee, 1 Hideki Mizu-uchi, 2 Shantanu Patil, 2 and Darryl D Lima 2 1 DivisionofOrthopaedicSurgery,ScrippsClinic,10666NorthTorreyPinesRoad,MS116,LaJolla,CA92037,USA 2 Shiley Center for Orthopaedic Research & Education at Scripps Clinic, 11025 North Torrey Pines Road, Suite 200, La Jolla, CA 92037, USA Correspondence should be addressed to William D. Bugbee; bugbee.william@scrippshealth.org Received 31 May 2013; Accepted 16 August 2013 Academic Editor: Masato Takao Copyright 2013 William D. Bugbee et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Customized patient instrumentation (CPI) combines preoperative planning with customized cutting jigs to position and align implants during total knee arthroplasty (TKA). We compared postoperative implant alignment of patients undergoing surgery with CPI to traditional TKA instrumentation for accuracy of implant placement. Twenty-five consecutive TKAs using CPI were analyzed. Preoperative CT scans of the lower extremities were segmented using a computer program. Limb alignment and mechanical axis were computed. Virtual implantation of computer-aided design models was done. Postoperative coronal and sagittal view radiographs were obtained. Using 3D image-matching software, relative positions of femoral and tibial implants were determined. Twenty-five TKAs implanted using traditional instrumentation were also analyzed. For CPI, difference in alignment from the preoperative plan was calculated. In the CPI group, the mean absolute difference between the planned and actual femoral placements was 0.67 in the coronal plane and 1.2 in the sagittal plane. For tibial alignment, the mean absolute difference was 0.9 in the coronal plane and 1.3 in the sagittal plane. For traditional instrumentation, difference from ideal placement for the femur was 1.5 in the coronalplaneand2.3 in the sagittal plane. For the tibia, the difference was 1.8 in the coronal plane. CPI achieved accurate implant positioning and was superior to traditional TKA instrumentation. 1. Introduction Accurate alignment and positioning of implants in total knee arthroplasty (TKA) is an important goal of the procedure. Numerous studies have demonstrated a high frequency of implant malalignment in TKA, regardless of the surgical techniques utilized [1 7]. The innovation cycle of TKA has mirrored this fundamental concept. Initially, free-hand surgical cuts were performed prior to the placement of implant components. Subsequently, mechanical alignment guides were devised based on bony or external landmarks, and predetermined angular or measured resections were performed. More recently, image-guided or imageless computer navigation systems have been developed to guide the surgical procedure and ultimate component alignment. The most recent innovation in TKA is customized patient instrumentation (CPI), which has been introduced as a next generation technology in an effort to further improve the accuracy and precision of surgical technique, implant placement, and alignment. The concept of CPI revolves around the use of preoperatively obtained imaging studies such as plain radiographs, magnetic resonance imaging (MRI), or computed tomography (CT) scans that are then manipulated in software programs to generate three-dimensional models of an individual patient s knee anatomy and limb alignment. This model is utilized to create a customized surgical plan, which defines the surgical cuts of the tibia and femur. This surgical plan is used to create customized cutting jigs that uniquely fit the individual patient s anatomy. The purpose of this study was to determine if a CTbased customized patient instrumentation system resulted in accurate implantation of tibial and femoral components

2 Advances in Orthopedics in TKA and to compare the CPI protocol to standard mechanical alignment guides used in traditional TKA with respect to femoral and tibial component alignment. 2. Materials and Methods 2.1. Patients. With the approval of our institutional review board, we retrospectively analyzed 25 consecutive patients who underwent TKA using CT-based CPI (TruMatch, Depuy, Warsaw, Indiana). Preoperative CT scans were obtained per the device manufacturer s protocol and were used to generate a surgical plan based on predefined surgeon preferences as well as restoration of ideal mechanical alignment (Figure 1). Customized cutting guides were then made and delivered sterile to the operating room (Figure 2). Standard TKAwas performed utilizing customcutting blocks rather than traditional mechanical jigs (Figures 3(a) and 3(b)). A second cohort of matched patients who underwent TKA using traditional instrumentation was analyzed for comparison. In the traditional TKA cohort, the targeted ideal placement was defined as 90 to the mechanical axis in the coronalandsagittalviewsforthefemurand90 in the coronal view for tibia. All patients underwent a standard postoperative rehabilitation protocol and were followed at 3 months, 6 months, and 1 year postoperatively. At the time of the latest clinical followup, postoperative radiographs (coronal and sagittal views) were obtained. 2.2. Accuracy of Implant Placement. Preoperative CT scans of the patient s lower extremity were segmented using MIMICS 13.0 (Materialise, Belgium) (Figure 4). Three-dimensional (3D) models of the knee joint and the bones were created from which limb alignment and mechanical axis were computed for each joint. The mechanical axis of the unoperated limb was calculated by a line joining the center of the femoral head to the middle of the talar dome. The models were then imported into a modeling and design software Rhino (McNeel North America, Seattle, WA). Based on the preoperative planning protocol, virtual implantation of the TKA computer-aided design (CAD) models was performed using Rhino (Figure 5). This constituted the ideal planned position of the femoral and tibial components. The digitized postoperative knee radiographs were analyzed to determine the final placement of the actual implants. To convert the two-dimensional radiographs to 3D, an open source image matching software JointTrack (University of Florida) was used. The relative positions of the femoral and tibial implant were thus obtained from the radiographs. This data was then imported into Rhino to accurately determine the position of implanted components. This position was compared with the virtual surgical placement. The difference in planned and actual placement of the implants was calculated in the sagittal and coronal planes for the CPI group (Figure 6). In the traditional TKA group, the targeted alignment was 90 in both planes for femur and 90 in coronal plane for the tibia in sagittal view. In the traditional TKA group, implant positions were calculated in the same manner as for the CPI group. 2.3. Statistics. Statistical analysis included t-test for equality of means and was determined using SPSS 13 (IBM software, Armonk, NY). 3. Results IntheCPIgroup,themeanabsolutedifferencefromthe planned femoral placement was 0.8 (±0.6 )inthecoronal plane and was 1.2 (±0.9 )inthesagittalplane(table 1). The single outlier in this group ( 4.4 difference) was a patient with posttraumatic femoral deformity, which necessitated a deviationfromtheplannedprotocolintraoperatively.inthe traditional TKA group, this difference from ideal placement was 1.5 (±1.6 )inthecoronalplaneand2.3 (±1.3 )inthe sagittal plane. For the tibial tray, the difference from planned placement was 1.0 (±1.0,range, 1.6 to 1.7 )inthecoronal planeforthecpigroup,whileinthetraditionaltkagroup, the difference from ideal placement was 1.8 (±1.6,range, 3 to 6.4 )inthecoronalplane.thecomparisonbetweenthecpi and the traditional TKA group was statistically significant forfemoralimplantpositioning(p < 0.01) butnotforthe tibial implant positioning. More outliers (±3 )occurredinthe traditional TKA group. 4. Discussion Interest in custom or patient-specific cutting guides is increasing. The potential for improved surgical efficiency with decreased operative times using fewer instruments and the possibility of improvements in surgical accuracy compared to conventional mechanical instruments are attractive features. Although computer-assisted techniques have demonstrated improved component alignment in TKA, the relative cost and increased operative time has led to resistance in the widespread adoption of this technique during TKA surgery [7, 8]. Few published clinical studies are available analyzing CT-based CPI for TKA. The results of our study validates the concept that using a CT-based protocol to create a three-dimensional model and subsequent cutting guides resulted in accurate surgical positioning of the TKA implants. The second question we asked was whether customized patient instrumentation was more accurate than traditional mechanical instrumentation utilizing intramedullary femoral alignment guides and extramedullary tibial guides. In our study, the CPI was more accurate than mechanical instrumentation on the femoral side and equal on the tibial side. CPI was associated with fewer outliers. A number of clinical studies have evaluated TKA alignment utilizing similar MRIbased patient-specific guides. Nunley et al. [9] evaluated 150 primary TKA using either conventional instrumentation, mechanical-axis-based patient-specific instrumentation, or kinematic-based patient-specific instrumentation. They found that these MR-based systems had a similar number of mechanical outliers as mechanical instruments

Advances in Orthopedics 3 Table 1: Mean absolute difference between the planned (CPI) and ideal (traditional TKA) placement of the femoral and tibial components. Asterisk denote statistical significance (paired t-testtwo tailed). CPI cohort Traditional TKA cohort Mean (SD, range) Mean (SD, range) P value Femur Coronal 0.8 (±0.6 ; 1.0 to 1.5 ) 1.5 (±1.6 ; 3.9 to 2.9 ) P < 0.01 Sagittal 1.2 (±0.9 ; 4.4 to 2.4 ) 2.3 (±1.3 ; 2.5 to 4.7 ) P < 0.001 Tibia Coronal 1.0 (±1.0 ; 1.6 to 1.7 ) 1.8 (±1.6 ; 3 to 6.4 ) S 5 Anatomical axis Femoral mechanical axis L M 0 Valgus to mechanical axis 7.4 Distal lateral femoral resection 0 Valgus to mechanical axis 11.0 Distal medial fermoral resection 0.0 Distal medial cartilage estimate I Figure 1: Portion of the surgical plan generated from preoperative CT scanning. Figure 2: Tibial (left) and femoral (right) customized cutting guides.

4 Advances in Orthopedics (a) (b) Figure 3: (a) Femoral cutting guide after placement on the distal femur. (b) Tibial cutting guide after placement on the proximal tibia. Figure 4: Segmenting process rendering CT scan into 3D model of the knee. and, therefore, questioned their clinical utility. Ng et al. [10] reviewed 569 TKA performed with an MRI-based patientspecific instrumentation system and found slightly better mechanical alignment and fewer outliers than in a matched group of 155 conventionally instrumented TKAs. The question of whether a CT-based or MR-based system is superior is not yet fully clarified. The relative advantage of a CT-based system is better bony landmark resolution than MR [11] and the ability to determine limb mechanical axis. Lower cost and shorter acquisition times are also potential advantages with CT. Relative disadvantages include the use of ionizing radiation. Certain limitations of this study should be discussed. We did not attempt to measure clinical outcome. To date, no conclusive evidence has demonstrated that either computer navigation or customized patient instrumentation leads to improved clinical outcome or implant longevity. However, many authors have demonstrated a correlation between coronal alignment and TKA failure [12, 13]. The clinical value of custom patient instrumentation has yet to be conclusively determined. Another limitation of the current study is the postoperative analysis using plain radiographs rather than CT scans. In this study, the precision of the data analysis would have been improved if patients underwent both preoperative and postoperative CT scans to obtain more accurate 2D to 3D modeling of the knee and limb. Ongoing advances in CT-based bone modeling protocols and subsequent manufacture of customized cutting blocks should lead to further improvements in surgical precision and accuracy. Additionally, the analytical methodology used in this study may be valuable in validating the accuracy of other implant alignment systems or image-based CPI protocols and devices [14, 15]. In conclusion, CPI is a novel technique that offers the potential of increased accuracy and

Advances in Orthopedics 5 Figure 5: Virtual implantation of TKA based on ideal planned component position. Figure 6: Comparison of actual versus planned surgical implantation. efficiency in TKA. The CT-based CPI used in this study (Tru Match, DePuy, Warsaw, Indiana) accurately positioned implants relative to the preoperative plan and achieved overall implant alignment better than traditional mechanical instrumentation, with fewer outliers. References [1] K. E. Teter, D. Bregman, and C. W. Colwell Jr., The efficacy of intramedullary femoral alignment in total knee replacement, Clinical Orthopaedics and Related Research,no.321,pp.117 121, 1995. [2] S. Patil, D. D. D Lima, J. M. Fait, and C. W. Colwell Jr., Improving tibial component coronal alignment during total knee arthroplasty with use of a tibial planing device, Journal of Bone and Joint Surgery A,vol.89,no.2,pp.381 387,2007. [3] H.Mizu-uchi,S.Matsuda,H.Miura,H.Higaki,K.Okazaki,and Y. Iwamoto, The effect of ankle rotation on cutting of the tibia in total knee arthroplasty, Journal of Bone and Joint Surgery A, vol. 88, no. 12, pp. 2632 2636, 2006. [4]M.R.Reed,W.Bliss,J.L.Sher,K.P.Emmerson,S.M.G. Jones, and P. F. Partington, Extramedullary or intramedullary tibial alignment guides: a randomized, prospective trial of radiological alignment, Journal of Bone and Joint Surgery B,vol. 84,no.6,pp.858 860,2002. [5]H.E.Cates,M.A.Ritter,E.M.Keating,andP.M.Faris, Intramedullary versus extramedullary femoral alignment systems in total knee replacement, Clinical Orthopaedics and Related Research, no. 286, pp. 32 39, 1993. [6]J.Y.Jenny,U.Clemens,S.Kohler,H.Kiefer,W.Konermann, and R. K. Miehlke, Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a casecontrol study of 235 cases compared with 235 conventionally implanted prostheses, Journal of Arthroplasty, vol. 20, no. 7, pp. 832 839, 2005. [7] G.Matziolis,D.Krocker,U.Weiss,S.Tohtz,andC.Perka, A prospective, randomized study of computer-assisted and conventional total knee arthroplasty: three-dimensional evaluation of implant alignment and rotation, Journal of Bone and Joint Surgery A,vol.89,no.2,pp.236 243,2007. [8]W.G.Blakeney,R.J.K.Khan,andS.J.Wall, Computerassisted techniques versus conventional guides for component alignment in total knee arthroplasty: a randomized controlled trial, JournalofBoneandJointSurgeryA,vol.93,no.15,pp. 1377 1384, 2011. [9] R. M. Nunley, B. S. Ellison, J. Zhu, E. L. Ruh, S. M. Howell, and R. L. Barrack, Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clinical Orthopaedics and Related Research,vol.470,no.3,pp.895 902,2012. [10] V.Y.Ng,J.H.DeClaire,K.R.Berend,B.C.Gulick,andA.V. Lombardi Jr., Improved accuracy of alignment with patientspecific positioning guides compared with manual instrumentation in TKA, Clinical Orthopaedics and Related Research,vol. 470, no. 1, pp. 99 107, 2012. [11] D. White, K. L. Chelule, and B. B. Seedhom, Accuracy of MRI vs CT imaging with particular reference to patient specific templates for total knee replacement surgery, International Journal of Medical Robotics and Computer Assisted Surgery,vol. 4, no. 3, pp. 224 231, 2008. [12]M.A.Ritter,P.M.Faris,E.M.Keating,andJ.B.Meding, Postoperative alignment of total knee replacement: its effect on

6 Advances in Orthopedics survival, Clinical Orthopaedics and Related Research, no. 299, pp.153 156,1994. [13] M. E. Berend, M. A. Ritter, J. B. Meding et al., Tibial component failure mechanisms in total knee arthroplasty, Clinical Orthopaedics and Related Research, no. 428, pp. 26 34, 2004. [14] H. Mizu-uchi, S. Matsuda, H. Miura, K. Okazaki, Y. Akasaki, and Y. Iwamoto, The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system, JournalofBoneandJointSurgeryB,vol.90,no.8,pp.1025 1031, 2008. [15] H. Mizu-uchi, C. W. Colwell Jr., S. Matsuda, C. Flores- Hernandez, Y. Iwamoto, and D. D. D Lima, Effect of total knee arthroplasty implant position on flexion angle before implantbone impingement, Journal of Arthroplasty, vol. 26, no. 5, pp. 721 727, 2011.