Directions For Use. All directions should be read before use. Page 1 of 8

Similar documents
Directions For Use. All directions should be read before use

The SplitWire Percutaneous Transluminal Angioplasty Scoring Device. Instructions for Use

Figure 1: Revolution TM Peripheral Atherectomy System Diagram. Table 1: Revolution TM Peripheral Atherectomy System Specifications Minimum Burr

CLARIVEIN INFUSION CATHETER

AXS Catalyst Distal Access Catheter

RECOMMENDED INSTRUCTIONS FOR USE

Cardiva Catalyst III INSTRUCTIONS FOR USE

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

ER REBOA Catheter. Instructions for Use

Endorectal Balloon (ERB) Endorectal Balloon (ERB) Instructions for Use

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER

Balloon in Balloon (BIB )

Radux StandTall Instructions for Use Sheath Extender and Securement Clasp

LeMaitre Embolectomy Catheter

Pruitt Irrigation Occlusion Catheter. English Instructions for Use. Pruitt Irrigation Occlusion Catheter

Z-MED II BALLOON AORTIC AND PULMONIC VALVULOPLASTY CATHETER. Instructions for Use

Instructions for Use Cordis PRECISE PRO Rx Nitinol Stent System

Balloon in Balloon (BIB )

Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.

Polyurethane Dual Lumen Occlusion Catheter English Instructions for Use. Polyurethane Dual Lumen Occlusion Catheter

Vascu-PICC WITH CUFF PERIPHERALLY INSERTED CENTRAL VEIN ACCESS CATHETER INSTRUCTIONS FOR USE

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter

PTCA Dilatation Catheter. Instructions for Use

PTA Balloon Dilatation Catheter

PTCA Dilatation Catheter. Instructions for Use

Single Use Curlew TM Multiple Biopsy Forceps

Instructions for Use Reprocessed 3D Diagnostic Ultrasound Catheters

Peel-Apart Percutaneous Introducer Kits for

SYMBOLOGY. = Manufacturer. = Manufacturing Serial Number. = Manufacturing Lot Number = Catalog Number LOT REF. = Do not resterilize Single Use Only

Instructions for Use Reprocessed ENDOPATH XCEL Bladeless Endoscopic Trocars and Cannulas. Reprocessed Device for Single Use.

FLEXIC ATH LTD. Peripherally Inserted. Instructions n For Use.

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Bladeless Trocar with OPTIVIEW Technology

Advanced Innovation for Exceptional Performance

Nit-Occlud. Coil System for PDA Closure IMPLANTATION POCKET GUIDE. Rx only CV / B. Braun Interventional Systems Inc.

Peripheral Intravascular Lithotripsy (IVL) Catheter Instructions for Use (IFU)

PTA Balloon Dilatation Catheter

Single Pass MCA Revascularization with Trevo

Peritoneal Drainage System

Instructions for Use 1

Pruitt-Inahara Carotid Shunt

Home Health Foundation, Inc. To create more permanent IV access for patients undergoing long term IV therapy.

Xcela Hybrid PICC with PASV Valve Technology

Azur CX 35 Peripheral Coil System (Detachable) Instructions for Use

Lutonix AV Clinical Trial

Document No. BMB/IFU/40 Rev No. & Date 00 & 15/11/2017 Issue No & Date 01 & 15/11/2017

Phoenix Atherectomy System Product Overview /LB

DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service

Allium Trans-hepatic Biliary Stent (BIS)

Advisor FL Circular Mapping Catheter, Sensor Enabled D-AVSE-D10-F20, D-AVSE-DF10-F20, D-AVSE-D10-F15, D-AVSE-DF10-F15. Instructions for Use

INSTRUCTIONS FOR USE

MULTILAYER FLOW MODULATOR (MFM ) FOR PERIPHERAL ANEURYSMS TREATMENT (LOWER LIMBS, VISCERALS AND SUBCLAVIAN)

Per-Q-Cath* PICC Catheters with Excalibur Introducer* System

The Ceterix NovoStitch Disposable Suture Passer. Do Not Reuse

ASEPT. Pleural Drainage System INSTRUCTIONS FOR USE REF LOT STERILE EO. Manufactured for: 824 Twelfth Avenue Bethlehem, PA

Advancing Lives and the Delivery of Health Care. The High-Flow Port Designed for Apheresis

ASEPT. Pleural Drainage System INSTRUCTIONS FOR USE. Rx only REF LOT. STERILE EO Sterilized using ethylene oxide

Quick Reference Guide

Ponsky * PEG Safety System - "Pull" Bard * PEG Safety System - "Guidewire" Information for Use

CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE

ARROW ENDURANCE. Extended Dwell Peripheral Catheter System. Rx only.

eptfe Vascular Graft Instructions for Use

BACTISEAL Endoscopic Ventricular Catheter (REF & )

Initial placement 24FR Pull PEG kit REORDER NO:

Initial placement 20FR Guidewire PEG kit REORDER NO:

TYSHAK DILATATION CATHETER. Instructions for Use CAUTION: Federal (USA) Law restricts this device to sale by or on the order of a physician.

Distal Radius Surgical Technique Guide

Device Preparation (all steps to be performed per standard interventional technique)

Aspira* Peritoneal Drainage Catheter

ASEPT Pleural Drainage System

Pocket Reference Guide For (CTO) Technologies

CATHETER ACCESS KIT. For use with Prometra Programmable Infusion Systems

Absorbable Woven Polyglycolic Acid Mesh Tube (Absorbable Nerve Conduit Tube) INSTRUCTIONS FOR USE 2 6

The High-Flow Port Designed & Indicated for Apheresis

Instructions for Use Reprocessed Arthroscopic Shavers

ROTABLATORTM. Peripheral. Rotational Atherectomy System. Quick Reference Cards

North Medical Endoscopic Biliary Stent (BilS)

Look For These Other Products by PARÉ Surgical, Inc. INNOVATIVE MEDICAL DEVICES

Groshong Catheters with Introcan Safety Peripheral IV Catheter Introducer

Transjugular liver access and biopsy

AXIOS Stent and Delivery System

for the ideal insertion of peritoneal dialysis catheter (patent pending)

DURABLE. CONSISTENT. SAFE. IN.PACT Admiral Drug-Coated Balloon

Groshong* PICC and Catheters

ASAHI Corsair INSTRUCTIONS FOR USE

LINEAR 7.5Fr. IAB Catheter

Instruction for use of Foley Catheter

BioFlo PICC. with ENDEXO Technology. Directions For Use A

Emerge. PTCA Dilatation Catheter ONLY. monorail. over-the-wire

Mechanical Thrombectomy Devices. Thomas M. Vesely, MD Vascular Access Services, LLC Saint Louis, Missouri

Assurant Cobalt Iliac BALLOON EXPANDABLE STENT SYSTEM

NEPHROCHECK Calibration Verification Kit Package Insert

Proven Performance Through Innovative Design *

INSTRUCTIONS FOR USE FOR:

Instruction For Use for All Silicon Foley Catheter

LONG-TERM HEMODIALYSIS CATHETER INSTRUCTIONS FOR USE

Interventional Treatment for Complete Occlusion of Arteriovenous Shunt: Our Experience in 39 cases

System.

Central Venous Catheter Insertion: Assisting

WallFlex Biliary RX Partially Covered Stent System Prescriptive Information

Transcription:

Directions For Use All directions should be read before use Page 1 of 8

WARNING: For single use only. Do not reuse, reprocess or re-sterilize. Reuse, reprocessing or re-sterilization may compromise the structural integrity of the device and/or lead to device failure which in turn may result in patient injury, illness or death. Reuse, reprocessing or re-sterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness or death of the patient. DEVICE DESCRIPTION: The Cleaner Rotational Thrombectomy System is a percutaneous, 6Fr catheter based system (single piece construction) that is compatible with a 6Fr introducer sheath. A disposable, hand-held battery operated rotator drive unit is attached to a wire which rotates at approximately 4000 RPM. The distal, sinuous shaped tip of the wire facilitates gentle mechanical declotting of occluded native vessel dialysis fistulae and synthetic dialysis access grafts. The wire is radiopaque for fluoroscopic visualization. INDICATIONS FOR USE: The Cleaner Rotational Thrombectomy System is indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts. CONTRAINDICATIONS: The Cleaner Rotational Thrombectomy System is contraindicated when: -In the medical judgment of the physician, such a procedure may compromise the patient s condition. -Existing hemodialysis access site infection. -Immature native vessel dialysis fistulae (fistulae that have not been used for at least one hemodialysis treatment). Page 2 of 8

WARNINGS AND PRECAUTIONS: -Prior to use, read all package insert warnings, precautions, and instructions. Failure to do so may result in severe patient injury and death. -These procedures should only be performed by physicians and staff familiar with the equipment and techniques involved. The device has been sterilized by EtO and is sterile unless the package is opened or damaged. The package should be examined before use, if damaged, DO NOT USE. The device is intended for single patient use only; DO NOT REUSE OR RE-STERILIZE. -Prior to use, carefully examine the Cleaner Rotational Thrombectomy System to verify that it has not been damaged during shipment. If product components show any sign of damage, DO NOT USE. -Due to the risk of exposure to HIV (Human Immunodeficiency Virus) or other blood borne pathogens, health care workers should routinely use universal blood and body fluid precautions in the care of all patients. Sterile technique must be strictly adhered to during any handling of the device. -Practitioners must be aware of potential complications associated with dialysis fistula and graft thrombolysis including: Hemorrhage Symptomatic pulmonary embolism Arterial embolization Allergic reaction to contrast media Pseudoaneurysm Pain and/or tenderness Vessel tear or disruption Infection Perforation of the artery or vein Hematoma Death -Caution should be used when dislodging the plug at the arterial anastomosis to minimize the risk of arterial embolization. -Due to the lack of excretion associated with hemodialysis patients, use of contrast should be kept to a minimum throughout this procedure. -Potential fatigue failure of the Cleaner sinuous wire may occur with prolonged activation of the Cleaner device. A withdrawal rate of 1-2 cm/second is recommended when sharp radii are encountered (i.e. radius of loop graft or fistula, radii < 3 cm). Page 3 of 8

A SUGGESTED PROCEDURE: Use sterile technique. Patient Preparation: 1. Premedicate with appropriate anxiolytic, analgesic and/or antibiotic per hospital protocol. Device Performance Testing: 2. Remove Cleaner Rotational Thrombectomy System from package. Depress ON/OFF switch to ensure sinuous wire spins freely (refer to Figure 2). Release switch to stop rotator. Precaution: Do not use device if rotator does not activate immediately when switch is depressed, and deactivate immediately when switch is released. 3. Flush Cleaner catheter with heparinized saline through the catheter lumen flush port. If stopcock is attached to device, return stopcock to off position prior to operation. Thrombolysis Procedure: 4. Complete the Cleaner Rotational Thrombectomy System procedure under continuous fluoroscopy. Do not initiate sinuous wire rotation (device activation) unless proper device positioning is confirmed within the fistula or graft. 5. Prepare and drape the puncture site as required. 6. Administer local anesthetic at puncture site for venous sheath insertion. Page 4 of 8

7. Select an appropriately sized sheath to accommodate the Cleaner catheter and other devices/catheters that may be used during the procedure. Maximum guidewire size will be dependent upon introducer sheath/dilator assembly chosen. 8. Prepare and place the venous introducer sheath per hospital protocol. The venous sheath should be placed in the venous limb of graft, and directed toward the venous anastomosis. In fistulae, the venous sheath placement can be optional depending on the clot burden in vessel. If a venous sheath is used, it should be placed in the venous limb of fistula and directed toward central venous outflow. Note: If no venous sheath is used in AV fistula, then go to step 16. 9. Place the device in the covered position by pushing the sliding lever to the distal position and rotating sliding lever to lock in covered position (refer to Figure 3). When in the covered position, only the flexible tip of the sinuous wire should extend from the catheter. The device should not be activated in the covered position. 10. Support flexible tip between thumb and index finger during insertion through sheath valve. Insert covered device through venous sheath into venous limb of fistula or graft. Page 5 of 8

11. In a graft, advance flexible tip up to venous anastomosis. Warning: Do not advance beyond anastomosis. In a fistula, advance flexible tip up to the central most extent of the clot. Uncover sinuous wire by unlocking, fully retracting sliding lever and rotating sliding lever until an audible "click" is heard (refer to Figure 4). Confirm device positioning within fistula or graft via fluoroscopy. Press ON/OFF switch to activate rotation. Note: Caution should be taken while uncovering wire to avoid advancing wire into clot and past anastomosis. 12. With device activated, slowly withdraw rotating sinuous wire along graft or fistula to break up clot. Warning: A withdrawal rate of 1-2 cm/second is recommended when sharp radii are encountered. When sinuous wire reaches tip of venous sheath, release switch to turn off rotator. 13. Cover device and remove from graft or fistula. Flush catheter lumen with heparinized saline and manually remove any accumulated fibrin from sinuous wire. 14. Aspirate macerated clot via sheath and discard aspirate. Precaution: Continued unsuccessful aspiration may collapse sheath and graft/fistula. 15. Inject small amount of contrast via venous sheath to assess degree of thrombus removal accomplished. Warning: Avoid over-injection of contrast to minimize risk of arterial embolization. Treat residual thrombus by repeating steps 11-14 until acceptable thrombus removal is achieved. 16. Administer local anesthetic at puncture site for arterial sheath insertion. Prepare and place the arterial introducer sheath per hospital protocol. The arterial sheath should be directed toward the arterial anastomosis. Precaution: Arterial and venous sheath tips must not overlap. 17. Support flexible tip between thumb and index finger during insertion through sheath valve. Insert covered device through arterial sheath into arterial limb of fistula or graft. Page 6 of 8

18. In a graft, advance flexible tip up to arterial anastomosis. Warning: Do not advance beyond anastomosis. In a fistula, advance flexible tip up to the central most extent of the clot. Uncover sinuous wire by unlocking, fully retracting sliding lever and rotating sliding lever until an audible "click" is heard. Confirm device positioning within fistula or graft via fluoroscopy. Press ON/OFF switch to activate rotation. 19. With device activated, slowly withdraw rotating sinuous wire, in uncovered position, along graft or fistula to break up clot. Warning: A withdrawal rate of 1-2 cm/second is recommended when sharp radii are encountered. When sinuous wire reaches tip of arterial sheath, release switch to turn off rotator. 20. Cover device and remove from graft or fistula. Flush catheter lumen with heparinized saline and manually remove any accumulated fibrin from sinuous wire. 21. Aspirate macerated clot using either sheath and discard aspirate. Precaution: Continued unsuccessful aspiration may collapse sheath and graft/fistula. 22. Pass an appropriate catheter through arterial sheath, and carefully feed past arterial anastomosis of graft or fistula. Inflate balloon, if balloon catheter. Pull arterial plug into middle of arterial limb. Deflate balloon and remove balloon catheter. 23. Reinsert covered Cleaner device through arterial sheath into arterial limb of graft or fistula. 24. Uncover sinuous wire and active device to break up arterial plug, using contrast to guide maceration. 25. Cover device and remove from graft or fistula. Flush catheter lumen with heparinized saline and manually remove any accumulated fibrin from sinuous wire. 26. Aspirate macerated clot via sheath and discard aspirate. 27. Inject contrast to assess degree of thrombus removal. Treat any residual thrombus using Cleaner via either sheath, as needed. 28. When thrombus removal is complete, treat any underlying disease or stenosis per hospital protocol. 29. Perform the final fistulogram. 30. Remove sheaths from the fistula or graft. 31. Achieve hemostasis at puncture site(s) per hospital protocol. Page 7 of 8

STORAGE Store at controlled room temperature. Do not expose to organic solvents, ionizing radiation or ultraviolet light. DISPOSAL Dispose of the catheter system in accordance with the Waste Electrical and Electronic Equipment Directive (WEEED) and according to standard institutional procedures for medical waste including single-use, blood contacting devices. DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY: There is no express or implied warranty, including without limitation and implied warranty of merchantability or fitness for a particular purpose, on the Rex Medical product(s) decribed in this publication. Under no circumstances shall Rex Medical be liable for any direct, incidental or consequential damages other than as expressly provided by specific law. No person has the authority to bind Rex Medical to any representation or warranty except as specifically set forth herein. Descriptions or specifications in Rex Medical printed matter, including this publication, are meant solely to generally describe the product at the time of manufacture and do not constitute any express warranties. Rex Medical will not be responsible for any direct, incidental or consequential damages from reuse of the product. Distributed by: Argon Medical Devices, Inc. 1445 Flat Creek Rd. Athens, TX 75751 Tel (800) 927-4669 www.argonmedical.com Manufactured by: 1100 East Hector St., Suite 245 Conshohocken, PA 19428 Tel (610) 940-0665 Fax (610) 940-1590 www.rexmedical.com P/N: P-1009-0136-00 Rev C mdi Europa GmbH Langenhagener Str. 71 30855 Hannover-Langenhagen Germany +49-511-39 08 95 30 0482 MEDICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK, FIRE, AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1 AND CAN/CSA- C22.2 No. 601.1-M90 4CR3 Page 8 of 8